Vaisala HMK15 Technical data

HMK15 Humidity Calibrator
three step plugs to fit all chamber holes (12 mm,
13.5 mm and 18.5 mm)
transit covers for secure transportation and long-term storage
thermometer with calibration certificate
In addition to laboratory use, Vaisala Humidity Calibrator HMK15 is suitable for on­site checks of humidity probes and transmitters.
Features/Benefits
Easy and reliable calibration
of humidity probes and transmitters
Based on saturated salt
solutions
Fast temperature equilibration
No external power required
Suitable for laboratory use
and on-site checks
Chambers and transit covers
make HMK15 easy to transport
Pre-measured certified salts
available
Vaisala Service Centers oer
accredited calibrations for humidity, temperature and barometric pressure.
No measuring instrument stays accurate by itself. It is essential that the functioning of an instrument is periodically checked against a reference. Vaisala has developed the Vaisala Humidity Calibrator HMK15 to make calibration and spot checking of humidity probes and transmitters easy and reliable.
Method used by leading laboratories
The operating principle of the HMK15 is based on the fact that a saturated salt solution generates a certain relative humidity in the air above it. The reading of the humidity probe or transmitter can then be adjusted accordingly. Many leading laboratories use this generally
www.vaisala.com
accepted and reliable method to calibrate humidity instruments. Usually two or three different salt solutions are used. Salts are chosen according to the application.
Certified salts
The HMK15 can be ordered with certied and pre-measured salts. A sample calibration is made from each batch in Vaisala’s Measurement Standards Laboratory (MSL).
Calibrated thermometer
The HMK15 can be ordered with a thermometer, which is used for measuring the temperature during the calibration. It can also be used for checking the temperature measurement accuracy of the transmitter. The thermometer can contain either mercury (accuracy ±0.3 °C (±0.54 °F)) or red capillary uid (accuracy ±1.0 °C (±1.8 °F)).
FINAS approved Measurement Standards Laboratory
Vaisala’s Measurement Standards Laboratory is a FINAS accredited calibration laboratory. FINAS is a member of the EA (the European Cooperation for Accreditation).
Technical data
General
The standard HMK15 consists of the following parts:
Two salt chambers, chamber covers and transit covers
Base plate Choice of thermometers
mercury thermometer, Vaisala calibrated, order code 19728HM
thermometer with red capillary liquid, calibrated by manufacturer, order code 25130HM
Measurement cup and mixing spoon
Options
Certified and ready dosed salts:
LiCl salt 11 %RH 19729HM (±1.3% RH)
salt 33 %RH 19730HM (±1.2% RH)
MgCl
2
NaCl salt 75 %RH 19731HM (±1.5% RH)
salt 97 %RH 19732HM (±2.0% RH)
K
2SO4
*Uncertainties given at +20 °C
on exchanged water 19767HM Extra salt chambers 19766HM Carrying bag HM27032
Order code:
Total uncertainty*:
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210908EN-A ©Vaisala 2009
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.