HMI41 Indicator and
HMP46 High Temperature Probe
www.vaisala.com
The Vaisala HUMICAP® Humidity
Indicator HMI41 equipped with the
Vaisala HUMICAP
Temperature Probe HMP46 is an ideal
combination for spot checking and field
calibration.
®
Humidity and
The Vaisala HUMICAP® Humidity
Indicator HMI41 tted with the
Vaisala HUMICAP
Temperature Probe HMP46 can be
used for spot checking humidity
and temperature in ducts or
chambers. Typical applications
are plant maintenance, installation
and inspection of air conditioning
systems, production and storage
areas and production processes.
In addition, it is suitable for eld
checking Vaisala's humidity
transmitters.
®
Humidity and
Withstands high
temperatures
The HMP46 operates in full
humidity range of 0 ... 100 %RH. The
temperature range is from
-40 to +100 °C (-40 ... +212 °F). For
short periods of time, the probe can
withstand temperatures up to
+180 °C (+356 °F).
The HMP46 probe is solid and
rugged. Its stainless steel probe is
made to withstand rough handling
in mechanically demanding
applications. The probe’s long
shaft can also reach otherwise
unreachable places.
Versatile and easy to use
In addition to displaying the humidity
and temperature readings, the HMI41
indicator calculates dewpoint and
wet bulb temperature, absolute
humidity and mixing ratio.
The indicator has an easy-to-read
two line liquid crystal display. The
display units (metric or non-metric)
are easily selected.
Features/Benefits
RH measurement range
▪
0 ... 100 %RH
Temperature measurement
▪
range -40 ... +100 °C
(-40 ... +212 °F),
for short periods up to +180 °C
(+356 °F)
Calculates dewpoint, wet bulb
▪
temperature, absolute humidity
and mixing ratio
Versatile and easy-to-use
▪
Incorporates Vaisala HUMICAP®
▪
Sensor
Excellent stability
▪
Data collection with serial line
▪
NIST traceable (certificate
▪
included)
Optional carrying case and
▪
calibration cable
Available as a complete
▪
calibration kit HMK41
These features, plus fast response
time, high measurement accuracy
and excellent stability, as well as
the wide temperature range of the
probe, make the HMI41/HMP46
combination an ideal choice for the
most demanding applications.
High performance sensor
The HMP46 probe incorporates
Vaisala HUMICAP
sensor has high accuracy, excellent
long-term stability and negligible
hysteresis. In addition, the sensor is
insensitive to dust, particulate dirt
and most chemicals.
®
Sensor. This
Technical data
-40
0
40
16080
120
2
1
0
-1
-2
-20 20 60
100
140
180
1.5
0.5
-0.5
-1.5
error %RH
T/°C
for temporary
use only
°C
-40 -20 0 20 40 60 80 100 120 140 160
T/°C
0.6
0.4
0.2
-0.2
-0.4
-0.6
0
180
HMI41 indicator
Calculated variables dewpoint temperature,
absolute humidity,
wet bulb temperature, mixing ratio
Resolution 0.1 %RH; 0.1 °C/°F
Power supply 4 batteries, type AA (LR 6)
Battery operation time
(alkaline batteries) 72 h continuous use
Auto-off function
Operating temperature -20 ... +60 °C (-4 ... +140 °F)
Storage temperature -40 ... +70 °C (-40 ... +158 °F)
Display two line LCD
Housing material ABS plastic
Housing classification IP53 (with connectors blocked)
Weight (incl. batteries) 300 g
Maximum measurement error of indicator at +20 °C
humidity ±0.1 %RH
temperature ±0.1 °C (±0.18 °F)
HMP46 probe
HUMIDITY
Measurement range 0 ... 100 %RH, non-condensing
Accuracy (incl. nonlinearity, hysteresis and repeatibility) at +20 °C
Maximum achievable accuracy when calibrated against
high quality, certified humidity standards:
0 ... 90 %RH ±1 %RH
90 ... 100 %RH ±2 %RH
When calibrated against salt solutions (ASTM E104-85)
0 ... 90 %RH ±2 %RH
90 ... 100 %RH ±3 %RH
Temperature Dependence
Typical long-term stability better than 1 %RH per year
Response time (90%)
at +20 °C in still air w/sintered filter 15 s
Humidity sensor HUMICAP
®
180
TEMPERATURE
Continuous measurement -40 ... +100 °C (-40 ... +212 °F)
Short-term measurement -40 ... +180 °C (-40 ... +356 °F)
Accuracy at +20 °C (+68 °F) ±0.2 °C (±0.36 °F)
Accuracy over the measurement range
Temperature sensor Pt 100 IEC 751 1/3 class B
General
Typical ranges of calculated variables
dewpoint temperature -20 ... +100 °C (-4 ... +100 °F)
absolute humidity 0 ... 600 g/m
wet bulb temperature 0 ... 100 °C (32 ... 212 °F)
mixing ratio 0 ... 600 g/kg d.a.
The accuracies of these calculated variables are limited by the
accuracies of the measured variables on which they are based,
namely RH and T. The RH and T accuracies are stated under
HMP46 probe.
Cable length 1500 mm; extended spiral cable
Operating temperature range
for electronics -20 ... +60 °C (-4 ... +140 °F)
Housing material
Electronics housing ABS plastic
Probe stainless steel
Housing classification
electronics IP65 (NEMA 4)
Sensor protection sintered filter, part no. 0195
optionmembrane filter, part no. 10159HM (max +80 °C / +176 °F)
plastic grid, part no. 6221 (max +80 °C / +176 °F)
Weight 450 g
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Portable Equipment
3
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Ref. B210940EN-A ©Vaisala 2010
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.