Vaisala HMI41 Technical data

HMI41 Indicator and HMP46 High Temperature Probe
www.vaisala.com
The Vaisala HUMICAP® Humidity Indicator HMI41 equipped with the Vaisala HUMICAP Temperature Probe HMP46 is an ideal combination for spot checking and field calibration.
®
Humidity and
The Vaisala HUMICAP® Humidity Indicator HMI41 tted with the Vaisala HUMICAP Temperature Probe HMP46 can be used for spot checking humidity and temperature in ducts or chambers. Typical applications are plant maintenance, installation and inspection of air conditioning systems, production and storage areas and production processes. In addition, it is suitable for eld checking Vaisala's humidity transmitters.
®
Withstands high temperatures
The HMP46 operates in full humidity range of 0 ... 100 %RH. The temperature range is from
-40 to +100 °C (-40 ... +212 °F). For short periods of time, the probe can withstand temperatures up to +180 °C (+356 °F).
The HMP46 probe is solid and rugged. Its stainless steel probe is made to withstand rough handling in mechanically demanding applications. The probe’s long shaft can also reach otherwise unreachable places.
Versatile and easy to use
In addition to displaying the humidity and temperature readings, the HMI41 indicator calculates dewpoint and wet bulb temperature, absolute humidity and mixing ratio.
The indicator has an easy-to-read two line liquid crystal display. The display units (metric or non-metric) are easily selected.
Features/Benefits
RH measurement range
0 ... 100 %RH
Temperature measurement
range -40 ... +100 °C (-40 ... +212 °F), for short periods up to +180 °C (+356 °F)
Calculates dewpoint, wet bulb
temperature, absolute humidity and mixing ratio
Versatile and easy-to-use
Incorporates Vaisala HUMICAP®
Sensor
Excellent stability
Data collection with serial line
NIST traceable (certificate
included)
Optional carrying case and
calibration cable
Available as a complete
calibration kit HMK41
These features, plus fast response time, high measurement accuracy and excellent stability, as well as the wide temperature range of the probe, make the HMI41/HMP46 combination an ideal choice for the most demanding applications.
High performance sensor
The HMP46 probe incorporates Vaisala HUMICAP sensor has high accuracy, excellent long-term stability and negligible hysteresis. In addition, the sensor is insensitive to dust, particulate dirt and most chemicals.
®
Sensor. This
Technical data
-40
0
40
16080
120
2
1
0
-1
-2
-20 20 60
100
140
180
1.5
0.5
-0.5
-1.5
error %RH
T/°C
for temporary use only
°C
-40 -20 0 20 40 60 80 100 120 140 160 T/°C
0.6
0.4
0.2
-0.2
-0.4
-0.6
0
180
HMI41 indicator
Calculated variables dewpoint temperature,
absolute humidity,
wet bulb temperature, mixing ratio Resolution 0.1 %RH; 0.1 °C/°F Power supply 4 batteries, type AA (LR 6) Battery operation time
(alkaline batteries) 72 h continuous use Auto-off function Operating temperature -20 ... +60 °C (-4 ... +140 °F) Storage temperature -40 ... +70 °C (-40 ... +158 °F) Display two line LCD Housing material ABS plastic Housing classification IP53 (with connectors blocked) Weight (incl. batteries) 300 g Maximum measurement error of indicator at +20 °C
humidity ±0.1 %RH
temperature ±0.1 °C (±0.18 °F)
HMP46 probe
HUMIDITY
Measurement range 0 ... 100 %RH, non-condensing Accuracy (incl. nonlinearity, hysteresis and repeatibility) at +20 °C
Maximum achievable accuracy when calibrated against
high quality, certified humidity standards:
0 ... 90 %RH ±1 %RH 90 ... 100 %RH ±2 %RH
When calibrated against salt solutions (ASTM E104-85)
0 ... 90 %RH ±2 %RH 90 ... 100 %RH ±3 %RH
Temperature Dependence
Typical long-term stability better than 1 %RH per year Response time (90%) at +20 °C in still air w/sintered filter 15 s Humidity sensor HUMICAP
®
180
TEMPERATURE
Continuous measurement -40 ... +100 °C (-40 ... +212 °F) Short-term measurement -40 ... +180 °C (-40 ... +356 °F) Accuracy at +20 °C (+68 °F) ±0.2 °C (±0.36 °F) Accuracy over the measurement range
Temperature sensor Pt 100 IEC 751 1/3 class B
General
Typical ranges of calculated variables
dewpoint temperature -20 ... +100 °C (-4 ... +100 °F) absolute humidity 0 ... 600 g/m wet bulb temperature 0 ... 100 °C (32 ... 212 °F)
mixing ratio 0 ... 600 g/kg d.a. The accuracies of these calculated variables are limited by the accuracies of the measured variables on which they are based, namely RH and T. The RH and T accuracies are stated under HMP46 probe.
Cable length 1500 mm; extended spiral cable Operating temperature range
for electronics -20 ... +60 °C (-4 ... +140 °F) Housing material
Electronics housing ABS plastic
Probe stainless steel Housing classification
electronics IP65 (NEMA 4) Sensor protection sintered filter, part no. 0195
optionmembrane filter, part no. 10159HM (max +80 °C / +176 °F)
plastic grid, part no. 6221 (max +80 °C / +176 °F) Weight 450 g Electromagnetic compatibility Complies with EMC standard
EN61326-1, Portable Equipment
3
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210940EN-A ©Vaisala 2010
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.