Vaisala HMI38, HMP36, HMP35 Operating Manual

HMI38 HUMIDITY DATA
PROCESSOR AND
HMP35/36/37E PROBES
Operating Manual
U063en-2.1
June 1998
Vaisala 1998
© Vaisala Oyj 1998 No part of this document may be reproduced in any form or by any
means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without a prior written notice of the copyright holder.
The instruction manuals may be changed without prior notice.
Yleisjäljennös 06/1998
HMI38
U063en-2.1 Operating Manual
Contents
1. PRODUCT DESCRIPTION..........................................................................................1
2. TO BE NOTED WHEN M EASURING HUMIDITY ........................................................2
3. TAKING THE HMI38 INTO USE..................................................................................3
4. COMMISSIONING .......................................................................................................4
4.1. Initializing the probe locations......................................................................4
4.2. Connecting the analogue outputs.................................................................4
4.3. Connecting the RS 232C or RS 485 serial bus .............................................5
4.4. Summary of the factory settings ...................................................................6
5. COMMANDS...............................................................................................................7
5.1. Display/keypad commands............................................................................7
5.1.1. Display mode....................................................................................7
5.1.2. Command m ode ...............................................................................7
5.1.3. Selecting and scaling the analogue outputs ......................................8
5.1.4. Setting the serial line parameters......................................................9
5.1.5. Setting the output mode..................................................................11
5.1.6. Others ............................................................................................12
5.2. Serial commands .........................................................................................16
5.2.1. Analogue output commands............................................................16
5.2.2. Output via the serial bus .................................................................18
5.2.3. Operating the HMI38 via the serial bus ...........................................20
6. DATA LOGGING.......................................................................................................22
6.1. Logging using display/keypad commands.................................................23
6.1.1. Start/stop data logging in the
6.1.2. Start/stop data logging in the
CONTINUOUS mode.............................23
CATCH mode ......................................24
6.1.3. Outputting the logged data to the serial line ....................................25
6.1.4. Outputting the logged data to the display ........................................26
6.2. Logging using serial line commands (
CONTINUOUS mode) .........................27
6.2.1. Data logging ...................................................................................27
7. CALIBRATION..........................................................................................................30
7.1. Humidity calibration.....................................................................................30
7.1.1. Calibration procedure......................................................................30
7.2. Temperature calibration ..............................................................................31
7.2.1. With display/keypad comm ands:.....................................................31
7.2.2. With serial line c om m ands..............................................................32
7.3. Calibrating the analogue outputs................................................................32
8. USING THE HMI38 AS A FIELD CALIBRATOR ........................................................ 33
8.1. HMD/W60/70 and HM P140 series transmitters ........................................... 33
8.1.1. Calibrating transmitters with voltage outputs ...................................34
8.1.2. Calibrating transmitters with current outputs ...................................36
8.2. HMP230 series transmitters ........................................................................37
8.2.1. One point calibration.......................................................................38
i
HMI38 Operating Manual U063en-2.1
8.2.2. Two point calibration.......................................................................39
8.3. HMD/W 20/30 and HM P130Y series transmit ters ........................................40
9. MAINTENANCE........................................................................................................42
9.1. Reference measurements............................................................................42
9.2. Self-diagnostics........................................................................................... 42
9.3. Adjusting the contrast of the display..........................................................43
10. TECHNICAL DATA...................................................................................................43
10.1. Relative humidity.........................................................................................43
10.2. Temperature.................................................................................................43
10.3. Calculated quantities...................................................................................43
10.4. Data logging.................................................................................................44
10.5. Analogue outputs.........................................................................................44
10.6. Serial interface.............................................................................................44
10.7. Electronics ...................................................................................................44
10.8. Mechanics ....................................................................................................45
10.9. Electromagnetic compatibility ....................................................................46
10.9.1. Emissions.......................................................................................46
10.9.2. Immunity.........................................................................................46
11. ACCESSORIES AND SPARE PARTS.......................................................................46
APPENDIX 1: SERIAL COMMANDS ..........................................................................47
APPENDIX 2: ERROR MESSAGES............................................................................69
APPENDIX 3: HMP35E PROBE..................................................................................75
APPENDIX 4: HMP36E PROBE..................................................................................81
APPENDIX 5: HMP37E PROBE..................................................................................87
WARRANTY.....................................................................................................................93
This manual is valid for programme version 2.01.
ii
HMI38
U063en-2.1 Operating Manual
1. PRODUCT DESCRIPTION
The microprocessor based HMI38 Humidity Data Processor is a light, portable instrument with an optional carrying case and built-in rechargeable battery which make it suitable for spot checking measurements or for use as a transfer standard.
The HMI38 is used together with Vaisala’s HMP35E, HMP36E or HMP37E humidity and temperature probes to measure relative humidity and tempera­ture quickly and accurately. Two probes can be connected to each HMI38. The HMI38 also calculates dewpoint, mixing ratio, absolute humidity and wet bulb temperature from the relative humidity and temperature it has measured.
The HMI38 has two analogue outputs (0...1V or 0...5 V) which can be selected and scaled freely. The HMI38 also has an RS 232C and an RS 485 interface.
The HMI38 Humidity Data Processor is used through the menu-based display. The user can choose between metric and non-metric units. Output modes are flexible: measured or calculated quantities can be displayed or they can be transferred to a computer or peripheral device using either RS 232C or RS 485 serial bus. Because the HMI38 units are addressable, there can be several units on the same RS 485 line.
The HMI38 can be used as a one point field calibrator for HMP140 series and HMD/W60/70 transmitters with standard or customized settings, as well as for calibrating their predecessors HMD/W20/30 and HMP130Y series. It can also be used as a one or two point calibrator for HMP230 series transmitters. The transmitters are calibrated using an appropriate reference cable between the HMI38 and the calibrated transmitter.
The HMI38 can also be used as a data logger. In continuous data logging, the user selects the interval and the starting time of logging; the HMI38 can store up to 254 readings from both probes in CONTINUOUS mode. In addition, the HMI38 can also be used for manual data logging. The CATCH mode enables the user to store readings manually at any given moment. In this mode, the HMI38 can store 127 readings from both probes. The HMI38 stores the measured relative humidity and temperature values. When the stored data is transferred to a computer or a peripheral through the serial bus, it calculates the other four quantities from them.
The HMI38 is powered with DC voltage, but with an AC adapter it can be powered from the mains. HMI38 has a built-in battery and a charger.
1
HMI38 Operating Manual U063en-2.1
2. TO BE NOTED WHEN MEASURING HUMIDITY
It is essential in the measurement of humidity and especially in calibration that temperature equilibrium is reached. Even a small difference in temperature between the measured object and the sensor causes an error. If the t emperature
is +20 °C (+68 °F) and the relative humidity 50 %RH, a difference of ±1 °C between the measured object and the sensor causes an error of ±3 %RH. When the humidity is 90 %RH, the corresponding error is ±5.4 %RH.
The error is at its greatest when the sensor is colder or warmer than the sur­roundings and the humidity is high. A temperature difference of a few degrees can cause water to condense on the sensor surface. In an unventilated space evaporation may take hours; good ventilation accelerates evaporation. The HUMICAP® sensor starts to function normally as soon as the water has evapo-
rated. If the condensed water is contaminated, the life span of the sensor may shorten and calibration may change.
10
9 8 7 6 5
4
dRH (%RH)
3 2 1
0
-40 -20 0 20 40 60 80 100 Temperature (°C)
Figure 2.1 Measurement error at 100 % RH when the dif feren ce
between the ambient and sensor temperature i s 1 °C
2
HMI38
U063en-2.1 Operating Manual
3. TAKING THE HMI38 INTO USE
Please, note that when you take the HMI38 into use for the first time, both probe locations have the factory setting for the probe type HMP35E. If this is not the case, the HMI38 displays error message(s) after power up (see appendix 3). Error messages can be cleared by pressing ENT button several times. Should this happen, you must initialize the probe locations. During initialization, the HMI38 gets information on which probe locations have probes connected to them. Initializing is described in Chapter 4.1.
If you use the instrument as a data logger, you must enter the correct date and time; see Chapter 6.
In order to enable the HMI38 to calculate the mixing ratio and wet bulb temperature accurately, enter the ambient air pressure through the keypad or serial bus. This setting is described in Chapter 5.
The HMI38 has a rechargeable battery which has not been charged in the factory. Charge the battery overnight. You can use the instrument instantly when you connect it to the mains for charging. To maximize the battery lifetime, completely discharge and recharge the battery about once a month. You can discharge the battery by leaving the HMI38 on without an AC adapter until the low battery voltage switches the HMI38 automatically off.
3
HMI38
T
T
Operating Manual U063en-2.1
4. COMMISSIONING
4.1. Initializing the probe locations
You can use either keypad or serial bus commands for the initialization of the probe locations. If you use serial commands, see Appendix 1.
If the probe type setting in the HMI38 memory is not correct, some error messages may appear on the display after power up (see Appendix 3). You can clear these messages by pressing ENT button one or several times. When the humidity and temperature readings appear on the display, press CL key to enter the command mode. The first display is the main menu:
Use arrow keys to select Init which starts to blink, and then press ENT. The following appears with the indication of PROBE 1 blinking (in this example, HMP35E):
Use arrow keys to select the correct type for PROBE 1 (HMP35E, HMP36E, HMP37E, HMP35BM, HMP36BM) and press ENT. If there is no probe in location 1, select none and press ENT. Use arrow keys to select also the type for PROBE 2 and press ENT. The HMI38 returns to the display mode showing the current readings.
4.2. Connecting the analogue outputs
Figure 4.1 shows the connections to the analogue output connector. See Chapter 5.1.3 and Chapter 5.1.3.2 for selecting the quantities and for scaling the analogue outputs.
ANALOGUE OUTPUT
0-1V/0-5V
1 2 3 4 5 6 7 8 9
AGND CH1 OU NC CH2 OU AGND NC NC NC NC
D9S
Figure 4.1 Cable connections to the analogu e outpu t connector
4
HMI38
-
U063en-2.1 Operating Manual
4.3. Connecting the RS 232C or RS 485 serial bus
Figure 4.2 shows the pin configuration of the serial line connector and the RS 232C serial line connection of the HMI38 to a peripheral device.
DO NOT CONNECT
RS 232 C RXD
RS 232 C TXD
NC
GND
NC NC
RS 485 HI
RS 485 LO
HMI38
SER I/O
D9P
COMPUTER (IBM AT)
1 2 3 4 5 6 7 8 9
SER I/O
2 3 4 5 6 7 8
9 1
D9S
RXD TXD DTR GND DSR RTS CTS RI DCD
DO NOT CONNECT
RS 232 C RXD
RS 232 C TXD
RS 485 HI
RS 485 LO
NC
GND
NC NC
HMI38
SER I/O
D9P
SERIAL PRINTER
(RS 232 PANASONIC KX
1 2 3 4 5 6 7 8 9
SER I/O
2 3 4 5 6 7 8
20 22
D25P
TXD RXD NC NC DSR GND DCD DTR NC
Figure 4.2 Wiring between the HMI38 and a PC or a serial line
printer using RS 232C serial line.
Connect the serial bus between the PC and the HMI38 and switch the PC on. Use a terminal emulation programme (e.g. Procomm Plus, Datastorm or Windows terminal) when using a PC. The following chapter lists the factory settings for data transfer.
NOTE
When changing the serial bus settings, switch the instrument off and on again or give the serial line command RESET to activate the new settings.
Refer to the manuals of the PC and the terminal emulation programme when giving serial settings.
5
HMI38 Operating Manual U063en-2.1
4.4. Summary of the factory settings
The factory settings of the HMI38 are the following:
probe initializations:
-probe 1 HMP35E
-probe 2 HMP35E
auto power-off off
output units metric
pressure 1013.25 hPa (= mbar)
analogue outputs:
-channel 1 RH1: 0...100 %RH 0...1 V
-channel 2 T1: -40...160 °C 0...1 V
serial line:
-bauds, parity, data, stop 4800, even, 7, 1
-duplexity FDX
address 0
serial mode stop
echo on
output interval 0 s
measurement integration time16
filter off
6
HMI38
U063en-2.1 Operating Manual
5. COMMANDS
The HMI38 incorporates a microprocessor which allows the configuration to be set according to the user’s needs. This is done through commands, either utilizing the menus on the front panel display or giving commands through the serial interface. The commands can be used e.g. to choose the displayed quantities, to select and scale the analogue outputs, to calibrate the temperature channels as well as the analogue outputs and to set the serial interface.
5.1. Display/keypad commands
5.1.1. Display mode
In the display mode the HMI38 outputs measured and calculated quantities on the display. You can scroll the first line with button s and the second line with button t. After reset the instrument shows first the programme type and version and then goes on to the display mode.
You can scroll different quantities with arrow switches. Acknowledge each selection by pressing ENT; this way, they appear on the display even after power off. The display also shows error messages and alarms if they occur.
5.1.2. Command mode
Press the CL key to enter the command mode. The first display is the main menu:
You can scroll the commands with the arrow keys. The currently active command blinks; select the desired command with the ENT key. When a menu is on the display, either the first command or the currently valid setting blinks. The CL key takes the HMI38 back to the display mode.
7
HMI38 Operating Manual U063en-2.1
5.1.3. Selecting and scaling the analogue outputs
5.1.3.1. Selecting the output
Select Mode in the main menu and Analog outputs in the Mode menu:
Select Mode 1V / 5V. The current settings for channel 1 are displayed:
If the settings are correct, press ENT.
If you need to change the settings, press CL:
the lower limit starts blinking
acknowledge the lower limit with ENT or change it with the nu-
meric keys, then press ENT
the upper limit starts blinking
acknowledge the upper limit with ENT or change it with the nu-
meric keys, then press ENT
When channel 1 has been set, the programme goes on to channel 2;
the procedure is the same as with channel 1.
NOTE
The hardware has two basic voltage ranges: 0...1 V and
0...5 V. When scaling analogue outputs, please note that
the relative accuracy of the output may become too low if only a very small part of either range is used to output a quantity with a wide measurement range.
5.1.3.2. Selecting and scaling the analogue output quantities
Select Mode in the main menu and Analog outputs in the Mode menu:
Select Scale. The quantity and scaling for channel 1 are displayed:
8
HMI38
U063en-2.1 Operating Manual
If the settings are correct, press ENT.
If you need to change the settings, press CL:
the quantity (RHn, Tn, Tdn, xn, an, Twn; n = 1 or 2) starts blink-
ing; change it with the arrow keys and acknowledged with the ENT key
the lower limit starts blinking
acknowledge the lower limit with ENT or give a new value using
the numeric keypad and press ENT
the upper limit starts blinking
acknowledge the upper limit with ENT or give a new value using
the numeric keypad and press ENT
After channel 1 has been set the settings of channel 2 appear on the
display; follow the procedure described above for channel 1.
5.1.4. Setting the serial line parameters
5.1.4.1. Turning the serial interface echo ON/OFF
Select More in the main menu, More in the More menu and Echo in
the second More menu.
Use the arrow keys to select the right alternative and press ENT.
5.1.4.2. Serial bus settings
Select Seri in the main menu; the currently valid serial int erface set-
tings are displayed:
If the settings are correct, press ENT; the programme returns to the
display mode.
If you need to change the settings, press CL:
9
HMI38 Operating Manual U063en-2.1
Select the parameter to be changed with the arrow keys and ENT key.
Selecting baud rate:
Selecting parity:
Selecting data bits:
Selecting stop bits:
Full duplex/half duplex:
NOTE
The serial bus settings become effective only after power off/on or the serial line command RESET.
5.1.4.3. Setting the instrument address
Address is used when more than one instrument is connected to one serial bus; this enables communication with one instrument at a time.
10
Select More in the Main menu and then Addr in the More menu; the
following appears on the display:
Change the address by entering a new value and pressing ENT.
Pressing ENT returns the programme to the More menu.
HMI38
U063en-2.1 Operating Manual
Pressing CL returns the programme to display mode and leaves the
address unchanged.
5.1.4.4. Selecting the output units
Select More in the Main menu and Unit in the More menu:
Use the arrow keys to select the right alternative and press ENT.
metric non-metric RH %RH %RH T °C °F
Td °C °F ag/m3gr/ft
3
x g/kg gr/lb Tw °C °F
5.1.5. Setting the output mode
The output modes only affect output through the serial interface: the instru­ment accepts all display/keypad commands irrespective of which serial output mode it is in. The HMI38 has three serial output modes: RUN, STOP and POLL.
In the RUN mode the instrument outputs measurements automatically through the serial interface to a PC or a peripheral. The only command that can be given through the serial interface is S (stop), which ends the RUN mode.
In the STOP mode serial commands are given to the instrument. Measure­ments can be output only by giving command SEND.
The POLL mode is used when more than one instrument is connected to the same serial bus; a single instrument can be addressed and communicated with. When the connection to one instrument is opened in the POLL mode, this instrument goes into STOP mode and can then accept commands normally. Closing the connection returns the instrument to POLL mode. In POLL mode the instrument outputs measurements only when requested (command SEND aa). If the user has forgotten the address of the instrument, it can be checked using display/keypad commands.
11
HMI38 Operating Manual U063en-2.1
5.1.5.1. Setting the serial interface operation mode
Select Mode in the main menu; the following appears on the display:
Select Serial:
The currently valid setting blinks. Select the desired mode with the ar-
row keys and press ENT. After this the programme returns to the previous Menu.
If you select the Run mode, the currently valid output interval appears
on the display (with number and unit blinking):
If required, change the output interval setting as follows:
press CL
the number starts blinking
if you need to change the interval, enter the new interval; otherwise
press ENT
the unit (seconds, minutes or hours) starts blinking
if necessary, change the unit with the arrow keys and acknowledge it
with ENT
after this the programme returns to the previous menu
5.1.6. Others
5.1.6.1. Setting the measurement integration time
Lengthening of the measurement integration time filters out any stray changes in the output: the instrument calculates the average of the number of measurement cycles defined by the user.
12
Select More in the main menu, More in the More menu and Mtim in
the second More menu:
HMI38
U063en-2.1 Operating Manual
Pressing ENT returns the programme to the previous menu without
changing the integration time.
If you need to change the integration time, enter the new time
(4...255) and press ENT.
5.1.6.2. Setting the measurement integration filter
If the integration filter is set on, the output reading is an average of four last measurements.
Select More in the main menu, More in the More menu and Filt in the
second More menu:
Choose the correct alternative (Filter on/off) with arrow keys and
press ENT.
5.1.6.3. Setting the pressure for mixing ratio and wet bulb temperature calculations
The atmospheric pressure has an effect on mixing ratio and on wet bulb tem­perature. To achieve accurate calculations, the instrument needs to take the ambient pressure into account.
Select More in the main menu:
Select Pres:
Pressing ENT returns the programme to the previous menu without
changing the pressure reading.
If you need to change the pressure, enter the new value and press
ENT.
13
HMI38 Operating Manual U063en-2.1
5.1.6.4. Setting the date
Select More in the main menu; select Date in the More menu:
If the date is correct, acknowledge it by pressing ENT; this takes the
programme back to the More menu.
If you need to change the date, press CL. Enter the new date in the
order the digits blink (year-month-day) and press ENT. Press ENT again to return to the previous menu. If there is no need to change the date, press CL.
5.1.6.5. Setting the time
Select More in the main menu; select Time in the More menu:
If the time is correct, acknowledge it by pressing ENT; this takes the
programme back to the More menu.
If you need to change the time, press CL. Enter the new time in the
order the numbers blink and press ENT. Press ENT again to return to the previous menu. If there is no need to change the time, press CL.
NOTE
As the HMI38 does not have a real-time clock with a backup battery, the date and time settings do not stay in the memory during power-off.
5.1.6.6. Setting the automatic shutdown
14
You can set the HMI38 to switch the power off automatically if no button is pressed for a certain time.
Select More in the main menu; select More in the first More menu
and then again in the second More menu:
Press ENT and select the correct alternative (on/off) for the power-off
mode with the arrow keys, then press ENT.
HMI38
U063en-2.1 Operating Manual
If you choose On, the current power-off time appears. If required,
change the time setting as follows:
press CL
the number starts blinking
if you need to change the time, enter the new time; otherwise press
ENT
the unit (seconds, minutes or hours) starts blinking
if required, change the unit with the arrow keys and acknowledge it
with ENT
after this the programme returns to Main menu
NOTE
Auto power-off is not operational during data logging and calibration procedures.
15
HMI38 Operating Manual U063en-2.1
5.2. Serial commands
Appendix 1 contains more detailed descriptions of the serial commands. The following chapters include only the most commonly used command sequences.
You can interrupt any serial command while giving it by pressing ESC. In the commands <cr> means carriage return.
5.2.1. Analogue output commands
5.2.1.1. Setting the analogue outputs
AMODE a.aaa b.bbb c.ccc d.ddd <cr>
a.aaa = l ower voltage limit of channel 1 b.bbb = upper voltage limit of channel 1 c.ccc = l ower voltage limit of channel 2 d.ddd = upper voltage limit of channel 2
Sets the voltage scale of the analogue outputs on channels 1 and 2 (anything between 0...5 V).
NOTE
The hardware has two basic voltage ranges: 0...1 V and
0...5 V. When scaling analogue outputs, please note that the relative accuracy of the output may become too low if only a very small part of either range is used to output a quantity with a wide measurement range.
Examples: Voltage output on channel 1 is set to 0...1 V and that on channel 2
is set to 2...5 V:
>AMODE 0 1 2 5 <cr> Ch1 0.000 ... 1.000 V Ch2 2.000 ... 5.000 V
The current settings are checked by entering AMODE without pa­rameters
>AMODE <cr> Ch1 0.000 ... 1.000 V Ch2 2.000 ... 5.000 V
16
HMI38
U063en-2.1 Operating Manual
5.2.1.2. Selecting and scaling the analogue output quantities
ASEL xxx yyy <cr>
xxx = quantity of channel 1 (RHn, Tn, Tdn, An, Xn or Twn) yyy = quantity of channel 2 (RHn, Tn, Tdn, An, Xn or Twn)
(n = 1 or 2)
Examples: Relative humidity of probe 1 is output on channel 1 and tempera-
ture of probe 2 on channel 2; the temperature range is scaled to be
0...100 °C:
>ASEL RH1 T2 <cr> Ch1 ( RH1 ) lo 0.000 %RH ? <cr> Ch1 ( RH1 ) hi 100.000 %RH ? <cr> Ch2 ( T2 ) lo -40.000 ’C ? 0 <cr> Ch2 ( T2 ) hi 160.000 ’C ? 100 <cr> >
After entering ASEL without parameters the instrument outputs its current settings
>ASEL <cr> Ch1 ( RH1 ) lo 0.000 %RH ? <cr> Ch1 ( RH1 ) hi 100.000 %RH ? <cr> Ch2 ( T2 ) lo 0.000 ’C ? <cr> Ch2 ( T2 ) hi 100.000 ’C ? <cr> >
5.2.1.3. Scaling the analogue outputs
ASCL <cr>
Scales the outputs selected on channels 1 and 2 (almost the same as ASEL but for scaling only).
Example: Scaling relative humidity on the range of 0...100 %RH and tem-
perature -40...+160 °C
>ASCL <cr> Ch1 ( RH1 ) lo 20.000 %RH ? 0 <cr> Ch1 ( RH1 ) hi 90.000 %RH ? 100 <cr> Ch2 ( T2 ) lo 0.000 ’C ? -40 <cr> Ch2 ( T2 ) hi 100.000 ’C ? 160 <cr> >
17
HMI38 Operating Manual U063en-2.1
5.2.2. Output via the serial bus
5.2.2.1. Starting the measurement output
R <cr>
Starts the measurement outputting of to peripheral devices (RUN mode); the only command that can be used is S (stop).
The output format can be changed with command FORM (see appendix 1).
5.2.2.2. Stopping the measurement output
S<cr>
Ends the RUN mode; after this command all other commands can be used.
5.2.2.3. Outputting the reading once
SEND <cr>
in STOP mode or
SEND aa <cr>
in POLL mode aa = address of the transmitter when more than one instrument is
connected to a serial bus (0...99)
The output format can be changed with command FORM (see appendix 1).
5.2.2.4. Setting the output interval for the RUN state
INTV xxx yyy <cr>
xxx = output interval (0...255) yyy = unit (s, min or h)
18
Sets the output interval for RUN mode and command R. If the interval is set to zero (s, min or h), the output interval will be determined by the frequency of measurements only.
Example: Changing output interval to 10 minutes
>INTV 10 min <cr> Output intrv. : 10 min
HMI38
U063en-2.1 Operating Manual
5.2.2.5. Serial bus settings
SERI b p d s x <cr>
b = bauds (300, 600, 1200, 2400, 4800, 9600) p = parity (n = none, e = even, o = odd) d = data bits (7 or 8) s = stop bits (1 or 2) x = duplex (H = half, F = full)
Examples: Giving new settings
>SERI O H <cr> changing parity and duplex only 4800 O 7 1 HDX
>SERI 600 N 8 2 F <cr> 600 N 8 2 FDX
changing all parameters
NOTE
New settings are effective only after switching the in­strument off/on or after the serial command RESET has been given.
If half-duplex setting is used with RS 485 bus, it will automatically turn the echo off. Even then the ECHO command can indicate that echo is on.
5.2.2.6. Selecting the output units
UNIT x <cr>
x = m(etric units)
n(on-metric units)
metric non-metric RH %RH %RH T °C °F
Td °C °F ag/m3gr/ft x g/kg gr/lb Tw °C °F
3
19
HMI38 Operating Manual U063en-2.1
5.2.2.7. Setting the address
ADDR aa <cr>
aa = address (0...99)
Example: The HMI38 is given address 99
>ADDR <cr> Address : 0 ? 99 <cr>
5.2.2.8. Resetting the HMI38
RESET <cr>
5.2.3. Operating the HMI38 via the serial bus
Setting the serial interface mode
SMODE xxxx<cr>
xxxx = STOP, RUN or POLL In STOP mode: measurements output only by command, all
commands can be used
In RUN mode: outputting automatically, only command S can be
used
In POLL mode: measurements output only with command SEND.
When in POLL mode, the output mode is changed as follows:
OPEN aa <cr>
SMODE xxxx<cr>
20
aa = address of the transmitter xxxx = STOP, RUN or POLL
The OPEN command sets the bus temporarily in STOP mode so that the SMODE command can be given.
Example:
>SMODE STOP <cr> setting STOP mode Serial mode : STOP
HMI38
U063en-2.1 Operating Manual
5.2.3.1. OPEN & CLOSE
OPEN nn <cr>
nn = address of the transmitter (0...99)
CLOSE <cr>
In STOP mode: OPEN has no effect, CLOSE sets the serial bus
temporarily to the POLL mode
In POLL mode: OPEN sets the serial bus temporarily to STOP
mode; CLOSE sets the bus back to POLL mode
The temporary mode changes are not stored in the memory of the transmitter; after reset the bus is in its original mode.
When more than one instrument is connected to the same serial bus, the POLL mode enables communication with an individual instrument.
Example: Transmitter 2 is temperature calibrated (currently in POLL mode)
>OPEN 2 <cr> this line is not visible on the screen
HMI 2 line opened for operator commands
>CT <cr> ...
>CLOSE <cr>
line closed
calibration procedure
21
HMI38 Operating Manual U063en-2.1
6. DATA LOGGING
The HMI38 has two data logging modes: the CONTINUOUS mode, i.e. automatic logging which can store up to 254 readings from both probes and the CATCH mode, i.e. manual logging with which the user can store up to 127 readings from both probes. The user can choose the logging interval and starting time of the logging when using the CONTINUOUS mode, and in the CATCH mode he can store readings at any given moment by pushing the button ENT. The HMI38 stores the measured relative humidity and temperature values and when the stored data is transferred to a computer or peripheral through the serial bus, it calculates the other four quantities – dewpoint temperature,
mixing ratio, absolute humidity and wet bulb temperature – from them.
NOTE
The CATCH mode is available only in units that have the software version 2.01 or more. You can easily check the version by reading the sticker on back of the device (SV2 indicates software version 2.01 or more).
Activate logging with either display/keypad or serial line commands. The logged data is output to serial line. In outputting the values for mixing ratio and wet bulb temperature, the currently valid pressure setting is used.
NOTE 1
To save battery, the analogue outputs cannot be used when data logging is on.
NOTE 2
When the HMI38 is powered from the battery and the logging memory becomes full, the instrument auto­matically turns itself off.
22
HMI38
U063en-2.1 Operating Manual
6.1. Logging using display/keypad commands
6.1.1. Start/stop data logging in the CONTINUOUS mode
NOTE
The HMI38 has no battery operated real-time clock. Give current date and time settings manually after power-on (see Chapter 5.1.6.5); this way, correct time and date indications are stored with each reading.
To start data logging:
Select Logg in the main menu. The LOGG menu is displayed:
Select Start and in the START menu, select Continuous:
The current delay (time before logging starts) and interval (time
between sequential data storings) settings appear on the display with the current delay blinking:
Press ENT to acknowledge the current delay time setting. To change
the setting press CL. Enter the new time and press ENT, then choose the correct time unit (s, min, h) with the arrow keys and press ENT. Acknowledge or change the time interval the same way.
After you have set the time interval, the data logging starts and the logging status appears on the display. The HMI38 displays the logging status and measurements alternately:
23
HMI38 Operating Manual U063en-2.1
NOTE
If the HMI38 has not output previous logging data, text Logged data is not transferred appears on the display. A new logging starts only after the old data has been transferred. To enable new logging, give command Print-Display or Print-Serial interface.
To stop data logging:
Press CL, select Logg in the main menu and then Stop in the LOGG
menu. Data logging ends.
The current status of the logging can be checked by selecting Status in the LOGG menu.
6.1.2. Start/stop data logging in the CATCH mode
NOTE
The HMI38 has no battery operated real-time clock. Give current date and time settings manually after power-on (see Chapter 5.1.6.5); this way, correct time and date indications are stored with each reading.
To start data logging in the CATCH mode:
Select Logg in the main menu. The LOGG menu is displayed:
Select Start in the LOGG menu and then Catch:
24
The measurement readings appear on the display alternately with the
following display:
HMI38
U063en-2.1 Operating Manual
You can now store readings at any given moment by pressing ENT
while the readings are on the display. The text DATA STORED appears to inform you that the readings as well as the time of measurement were correctly stored in the HMI38 memory:
Press ENT to return to the display showing the measurement
readings.
Note that to save battery life, you can turn the HMI38 off while not
storing readings and then turn it on again to continue. However, in this case remember to enter the correct date and time each time you turn the instrument on as it does not have a real-time clock. You can store max. 127 readings from both probes. To stop data logging, select Logg in the main menu and then Stop in the LOGG menu. Data logging ends.
6.1.3. Outputting the logged data to the serial line
Select Logg in the main menu and Print in the LOGG menu. Select
Serial interface:
The following appears:
The HMI38 outputs the logged data to the serial line:
> Data logging is ended
Delay to start : 0 s Interval : 10 min First data stored: 1994-10-18 13:45:55
Number of loggings : 1
18.1 23.5 -2.1 3.0 3.2 11.3
15.7 23.0 -4.4 3.2 2.7 10.6
The factory format for the data output is:
RH1 T1 Td1 a1 x1 Tw1 RH2 T2 Td2 a2 x2 Tw2
25
HMI38 Operating Manual U063en-2.1
You can change the format with serial line command LFORM and pause the output by pressing CL. If you wish to continue, press ENT; if you wish to stop the outputting completely, press CL:
The HMI38 stores the logged data in the memory until a new data logging starts.
6.1.4. Outputting the logged data to the display
Select Logg in the main menu and Print in the LOGG menu. Select
Display:
The first logging point stored appears on the display, e.g.:
The first number from the left on the first line (in this example, 1)
indicates which logging point is currently being observed. The second number from the left on the first line (in this example, 69) tells you how many logging points are stored in the HMI38 memory. Other numbers indicate the time and date of the measurement.
Press ENT and the readings of the first logging point appear on the
second line of the display, e.g.:
The abbreviation of the measurement quantity and the number of the
probe appear on the left of the second line. If you wish to see the readings of other measurement quantities or the readings of the second probe (if connected), you can scroll them with buttons s or t, for example:
26
Press the button CL to move on to the following logging point. When
the readings are not on the display, you can select any logging point for observation with buttons s or t.
You can stop outputting the data by pressing twice the button CL.
The HMI38 stores the logged data in the memory until a new data logging starts.
HMI38
U063en-2.1 Operating Manual
6.2. Logging using serial line commands (CONTINUOUS mode)
See Appendix 1 for a description of the format of the serial line commands.
6.2.1. Data logging
LOG xxx aa bbb cc ddd <cr>
xxx = functional mode: ON, OFF or P(rint) aa = logging start delay bbb = delay unit (h, min, s) cc = logging interval ddd = interval unit (h, min, s)
Examples:
start logging and continue with intervals of 10 minutes:
>log on 10 min <cr> Data logging is ON
Delay to start : 0 s
Interval : 10 min
First data stored: 0000-00-00 00:00:00
Number of loggings : 0
start logging after five hours and continue with intervals of 30 minutes:
>log on 5 h 30 min <cr> Waiting for the start of logging
Delay to start : 5 h
Interval : 30 min
First data stored: ????-??-?? ??:??:??
Number of loggings : 0
end logging:
>log off <cr> Data logging is ended
Delay to start : 0 s
Interval : 10 min
First data stored: 1994-10-18 13:45:55
Number of loggings : 1
output the logged data (defined by LFORM command):
>log p <cr> Data logging is ended
Delay to start : 0 s
Interval : 10 min
First data stored: 1994-10-18 13:45:55
Number of loggings : 1
18.1 23.5 -2.1 3.0 3.2 11.3
15.7 23.0 -4.4 3.2 2.7 10.6
27
HMI38 Operating Manual U063en-2.1
6.2.1.1. Setting the output format of the logged data
LFORM zzz...zzz <cr>
or
LFORM <cr>
xxx...xxx
? zzz...zzz <cr>
xxx...xxx = old format (blank = factory format) zzz...zzz = new format
Select output variables by variable name and the number of the measuring
probe:
RHn = relative humidity Tn = temperature Tdn = dewpoint temperature An = absolute humidity Xn = mixing ratio TWn = wet bulb temperature
where n is the number of the probe (1 or 2)
Give any other text you wish to appear in the output between inverted
commas (").
Give the number format in front of the output variable as:
<full digits>.<decimal digits>, for example: 3.2 or 2.1 etc. (max. 9.9).
Give the unit as:
Un where n specifi es the number of characters (m ax. 5), e.g. U3.
other parameters are:
28
\n or #n line feed <lf> \r or #r carriage return <cr> \t or #t horizontal tabulation <ht> or <tab> \nnn or #nnn ASCII-code nnn TIME current time (HH:MM:SS) DATE current date (YYYY-MM-DD)
HMI38
U063en-2.1 Operating Manual
If the input line is too short, continue the command parameter entry in the
following line as:
& <cr>.
Some examples:
command: >LFORM "rh1: " 3.2 RH1 " %RH" \r \n<cr> output: rh1: 29.62 %RH<cr><lf>
command: >LFORM "RH: " 2.2 RH1 U1 \r \n<cr> output: RH: 29.62%<cr><lf>
command: >LFORM date " " time #t RH1 #t T1 #r
#n<cr>
output: 1994-09-29 10:02:15 <tab> 29.6 <tab> 24.3
<cr><lf>
command: >LFORM "rh: " 2.2 RH2 " " U3 & <cr>
" t: " 2.2 T2 " " U2 #r #n<cr>
(line continued)
output: rh: 29.62 %RH t: 25.35 ’C <cr><lf>
The factory format can be recovered with command:
>LFORM * <cr>
Please, note that you can use the LFORM command for changing the format of the output of logged data only; otherwise it is functionally similar to the FORM command.
29
HMI38 Operating Manual U063en-2.1
7. CALIBRATION
As the HMI38 indicators and their probes have been fully calibrated at the fac­tory, there should be no immediate need for a recalibration. Calibrate them only if you have a reason to believe that the adjustments have changed. The adjustments of the temperature measurement channels and analogue output channels are particularly stable and in normal circumstances, need not be recalibrated. Perform humidity calibration for the probes at least once a year.
7.1. Humidity calibration
Perform the humidity calibration of the probes by adjusting the calibration po­tentiometers in the probe bodies (see Appendices 4, 5 or 6). Perform a two­point calibration with the HMK15 or HMK13B calibrator or send the probe to Vaisala. We recommend recalibration at least once a year. Please, note also that the probe needs recalibration every time the HUMICAP® humidity sensor is changed.
7.1.1. Calibration procedure
Leave the calibrator, the HMI38 and the probe for at least 4 hours in
the same space so that their temperatures have time to equalize. Remove the filter cap on the probe.
Insert the probe into a measurement hole of t he LiCl chamber in the
calibrator.
Wait at least 10 minutes, check the humidity reading from the HMI38
and adjust the reading to the correct value using the DRY potentiometer in the probe body.
Insert the probe into a measurement hole of the NaCl chamber i n the
calibrator.
Wait at least 10 minutes and adjust the humidity reading using the
WET potentiometer.
Table 7.1 Greenspan’s calibration table
Temperature °C 15 20 25 30 35
°F 59 68 77 86 95 LiCl %RH * 11.3 11.3 11.3 11.3 NaCl %RH 75.6 75.5 75.3 75.1 74.9
*) If the LiCl solution is used or stored at temperatures below +18 °C
(+64 °F), the equilibrium humidity of the salt solution changes permanently.
30
As the DRY and WET adjustments may affect each other, check again the humidity reading in the LiCl chamber (wait for 10 minutes after inserting the probe into the measurement hole). If the reading is not
HMI38
U063en-2.1 Operating Manual
correct, repeat the adjustments in both LiCl and NaCL chambers unt il the calibration is correct.
7.2. Temperature calibration
Temperature calibration should be done against an accurate temperature refer­ence. It can be done by using either the serial bus or the display/keypad com­mands.
7.2.1. With display/keypad commands:
Leave the reference instrument, the HMI38 and the probe in the same
space for at least 4 hours so that their temperatures have time to equalize. Remove the filter before calibration.
Select Cali in the main menu and select the probe:
Select either one-point or two-point calibration:
After the temperature of the instrument has stabilized, check the first
point reading against the reference instrument. Acknowledge the reading by pressing ENT or give the correct reading and press ENT:
In case of a two-point calibration, change the temperature and allow
enough time for temperature stabilization. Then check the transmitter against the reference, give the correct reading and press ENT.
NOTE
If you change the factory temperature calibration, the correction data stored in the HMI38 refers to the probe location of the calibrated probe. If you change the probe in this location, revert the measurement channel to its factory settings (see serial command LI in Appendix 1) or perform the calibration again with the new probe.
31
HMI38 Operating Manual U063en-2.1
7.2.2. With serial line commands
CT n <cr>
n = probe (1 or 2)
With this command, you can calibrate the instrument against an accurate ref­erence, such as a Pt 100 simulator.
Examples: two-point calibration for probe 1:
>CT 1 <cr> T1 : 0.9 Ref1 ? 0 <cr> Press any key when ready ... T1 : 119.8 Ref2 ? 120 <cr> >
one-point offset correction for probe 2 (<cr> is given after Ref2):
>CT 2 <cr> T2 : 25.9 Ref1 ? 26 <cr> Press any key when ready ... T2 : 25.9 Ref2 ? <cr> >
If you know the corrections at two points, you can also give the command di­rectly as:
CT n aa.aa bb.bb cc.cc dd.dd <cr>
where: n = probe location (1 or 2)
aa.aa = first point bb.bb = first point correction cc.cc = second point dd.dd= second point correction
Example: if you know the necessary corrections, perform the calibrations of
the previous examples as follows:
>CT 1 0.9 -0.9 119.8 0.2 <cr> >CT 2 25.9 0.1 <cr>
You can monitor the progress of the sensor temperature stabilization by enter­ing * <cr> at Ref1 or Ref2:
>CT 1 <cr> T1 : 0.80 Ref1 ? * <cr> T1 : 0.40 Ref1 ? 0.00 <cr> Press any key when ready ... T1 : 56.20 Ref2 ? 55.0 <cr> >
7.3. Calibrating the analogue outputs
32
You can perform the calibration with an accurate voltage meter using serial line command ACAL and make a test with command UTEST (see Appendix
1). Please, note that you should perform this calibration only if there is strong reason to believe that the adjustments of the analogue outputs have changed.
HMI38
U063en-2.1 Operating Manual
8. USING THE HMI38 AS A FIELD CALIBRATOR
You can also use the HMI38 as a field calibrator for HMD/W60/70 and HMP140 series and their predecessors HMD/W20/30 and HMP130Y series, as well as for the HMP230 series transmitters. Perform calibration by connecting an appropriate reference connection cable between the HMI38 and the transmitter to be calibrated (see chapter 11).
NOTE
The calibration of the HMD/W60/70 and HMP140 series is available only in units that have the software version 2.01 or more. You can easily check the version by reading the sticker on back of the device (SV2 indicates software version 2.01 or more).
8.1. HMD/W60/70 and HMP140 series transmitters
Use probe 1 as a reference measurement probe. Connect the D-connector of the reference connection cable to the connector of probe 2 in the HMI38. Connect the other connector of the cable to the test connector in the transmitter to be calibrated (refer to the transmitter manual for details). The connectors fit in one direction only.
NOTE
When you calibrate a transmitter with current output, terminate the output loop.
Make sure that the sensors of the transmitter and the reference probe are close to each other. If you are calibrating a duct type transmitter, push the probe through the support piece into the ready-made hole in the duct.
When the power of the HMI38 is off, press and hold down the NUMBER 2 button. Press the ON/OFF button and release it; keep the NUMBER 2 button still pressed. The following appears on the display:
Release the NUMBER 2 button and the following appears on the display:
33
HMI38 Operating Manual U063en-2.1
Select RH scale with buttons s or t and press ENT when the text RH scale is blinking. The following appears with the lower limit of the RH scale (0.00) blinking:
The default scale is 0.00 - 100.00 %RH. Acknowledge the lower limit with button ENT or enter the new value and press ENT. The upper limit (100.00) starts to blink. Acknowledge it with button ENT or enter the new value and press ENT. The display returns to show the following:
If you are calibrating a transmitter with current outputs, please, refer to Chapter 8.1.2.
8.1.1. Calibrating transmitters with voltage outputs
If voltage outputs are used, select Voltage with buttons s or t and press ENT when the text Voltage is blinking. The following text appears on the display:
Select Range with with buttons s or t and press ENT when the text Range is blinking. A text similar to the following appears on the display:
The lower limit (e.g. 0.00) is blinking. Acknowledge it with button ENT or change it by entering the new value and pressing then ENT. The upper limit (e.g. 1.00) starts to blink. Acknowledge it with button ENT or change it by entering the new value and pressing then ENT.
The display returns to show the following:
34
Select Jumper selection with with buttons s or t and press ENT when the text Jumper selection is blinking. A text similar to the following appears on the display:
HMI38
U063en-2.1 Operating Manual
The voltage range and the jumper selection are normally the same, but if the transmitter has customized ranges, these selections may differ from one another. Select the range according to the transmitter with buttons s or t and press ENT. You can return to voltage range and jumper selections by pressing the button CL. If you wish to return to the normal display mode, turn the HMI38 OFF and then ON again.
When you have acknowledged the jumper selection with ENT, the display changes to show a text similar to the following:
Numbers on the first line indicate the relative humidity measured by the transmitter, and numbers on the second line indicate the RH measured by the HMI38 reference probe. If you wish, you can now adjust the transmitter:
Wait until the readings have stabilized and adjust the calibration
potentiometers until the readings on the first and the second line are the same. It is recommended to use the offset potentiometer if the RH is <65 %RH and the gain potentiometer if the RH is >65 %RH.
After having completed the calibration, turn the HMI38 off
(disconnect the cable).
If you press the ENT button in the first RH calibration mode, you will go on to the second RH calibration mode which is an alternative for the first one. A text similar to the following is displayed:
Numbers on the first line indicate how much the RH measured by the transmitter differs from that measured by the HMI38 reference probe. Numbers on the second line indicate the relative humidity measured by the reference probe. If you wish, you can now adjust the transmitter:
Wait until the readings have stabilized and adjust the calibration
potentiometers until the numbers on the first line are at zero. It is recommended to use the offset potentiometer if the RH is <65 %RH and the gain potentiometer if the RH is >65 %RH.
After having completed the calibration, turn the HMI38 off
(disconnect the cable).
35
HMI38 Operating Manual U063en-2.1
8.1.2. Calibrating transmitters with current outputs
If current outputs are used, select Current with buttons s or t and press ENT when the text Current is blinking. A text similar to the following appears on the display:
Select the current range used with buttons s or t and press ENT. The display changes to show a text similar to the following:
Numbers on the first line indicate the relative humidity measured by the transmitter, and numbers on the second line indicate the RH measured by the HMI38 reference probe. If you wish, you can now adjust the transmitter:
Wait until the readings have stabilized and adjust the calibration
potentiometers until the readings on the first and the second line are the same. It is recommended to use the offset potentiometer if the RH is <65 %RH and the gain potentiometer if the RH is >65 %RH.
After having completed the calibration, turn the HMI38 off
(disconnect the cable).
If you press the ENT button in the first RH calibration mode, you will go on to the second RH calibration mode which is an alternative for the first one. A text similar to the following is displayed:
Numbers on the first line indicate how much the RH measured by the transmitter differs from that measured by the HMI38 reference probe. Numbers on the second line indicate the relative humidity measured by the reference probe. If you wish, you can now adjust the transmitter:
36
Wait until the readings have stabilized and adjust the calibration
potentiometers until the numbers on the first line are at zero. It is recommended to use the offset potentiometer if the RH is <65 %RH and the gain potentiometer if the RH is >65 %RH.
After having completed the calibration, turn the HMI38 off
(disconnect the cable).
HMI38
U063en-2.1 Operating Manual
8.2. HMP230 series transmitters
Use probe 1 as a reference measurement probe. Open the cover of the HMP230 to be calibrated. Connect the D-connector of the reference connection cable to the serial I/O connector of the HMI38 and the other connector of the cable to the connector X5 of the HMP230 (see Figure 8.1). This connector is also the other mounting connector for RS 485/422 serial port module and current loop module; if a module is mounted, it must first be removed.
RX GND TX
X17
X6
RX +
R2 R3
X5
R4
RX ­TX + TX -
SERIAL PORT
X4
Figure 8.1 Connector X5 on the main board of the HMP230
Make sure that the sensors of the transmitter and the reference probe are close to each other.
When the power of the HMI38 is off, press and hold down the NUMBER 0 button. Press and release the ON/OFF button; keep the NUMBER 0 button still pressed. When the programme type and version appear on the display, release the NUMBER 0 button. The following appears on the display:
If the connection fails, the following text appears:
Should this happen, check that the HMP230 is powered and that the reference cable is properly connected.
NOTE
The HMP230 must not be in the POLL mode and the serial line settings must be 4800, e, 7, 1 (factory set­tings, which can be restored using the factory settings jumper; see the HMP230 Operating Manual).
37
HMI38 Operating Manual U063en-2.1
When connection has been formed, select either one or two point calibration with the arrow keys and press ENT:
NOTE
If the programme version of the HMP230 is 1.06 or earlier (version is printed on the CPU hybrid), start the HMI38 as described above without connecting the reference connection cable. After the text ’no connection’ appears on the screen, connect the cable. This procedure is necessary for avoiding a possible overflow in the serial line buffer of the HMP230.
8.2.1. One point calibration
Select either offset or gain calibration using arrow keys and press
ENT. If the reference humidity is above 50 %RH, gain calibration is preferred. If the reference humidity is below 50 %RH, the HMP230 accepts only offset calibration .
The humidity readings of the HMP230 and the reference probe appear
on the display:
Wait until the readings have been stabilized and press ENT:
The correction factor is now stored to the memory of the HMP230 and the new readings of both the transmitter and the reference appear again on the display. In case of any drift in either of the readings, repeat the calibration by pressing the ENT button again. End the calibration procedure by pressing the CL button twice. This brings the following text on the display:
38
You have now completed the calibration; switch off the HMI38.
HMI38
U063en-2.1 Operating Manual
If the calibration fails (e.g. after an attempt of gain calibration below 50 %RH), an error message appears and no correction factor is stored into the HMP230:
8.2.2. Two point calibration
The HMI38 displays:
Select two-point calibration using arrow keys and press ENT. The
readings at the first humidity point appear on the display, e.g.:
Wait until the readings have stabilized and press ENT. The instrument
stores the correction at the first point and readings at the second humidity point appear:
Wait again until the readings have stabilized and press ENT. The
instrument stores the correction at the second point. End the calibration procedure by pressing the CL button twice. This brings the following text on the display:
Switch off the HMI38.
If the calibration fails, an error message appears on the display and no cor­rection factors are stored to the HMP230.
Two point calibration fails e.g. if the reference humidities at the first and the second point differ less than 50 %RH of each other.
39
HMI38 Operating Manual U063en-2.1
8.2.2.1. Two point calibration with reference salts
You can also give the reference humidities of the two point calibration manually (using the HMI38 as a display only). Insert the probe of the HMP230 into the low humidity bath. Follow the procedure of the two point calibration.
When the humidities at the first calibration point appear on the
display, press CL:
Enter the reference humidity with numeric keys and press ENT.
Insert the probe into the high humidity bath and repeat the procedure.
The correction factors are now stored to the HMP230. End the calibration procedure by pressing the CL button twice. Switch off the HMI38.
8.3. HMD/W 20/30 and HMP130Y series transmitters
Use probe 1 as a reference measurement probe. Connect the D-connector of the reference connection cable to the connector of probe 2 in the HMI38. Connect the other connector of the cable to the test connector in the transmitter to be calibrated (refer to the transmitter manual for details). The connectors fit in one direction only.
NOTE
When you calibrate a transmitter with current output, terminate the output loop.
Make sure that the sensors of the transmitter and the reference probe are close to each other. If you are calibrating a duct type transmitter, push the probe through the support piece into the ready-made hole in the duct.
When the power of the HMI38 is off, press and hold down the NUMBER 1 button. Press the ON/OFF button and release it; keep the NUMBER 1 button still pressed. When the programme type and version appear on the display, release also the NUMBER 1 button. The following appears on the display:
40
The humidity readings of the transmitter and the reference probe appear on the display:
HMI38
U063en-2.1 Operating Manual
You can change the display to show the difference between the transmitter and reference reading by pressing ENT; the reference reading appears on the second line:
You can change the display mode between these two modes by pressing the ENT button. You can make adjustments adjustment in all display modes. The potentiometer used for adjusting the transmitter varies depending on the transmitter type and the humidity level in which you perform the calibration (see the transmitter manual).
HMW20/30 UB/YB (wall mount transmitters):
use the "DRY" or one-point calibration trimmer potentiometer (see
the sticker inside the transmitter cover)
HMD20/30 UB/YB (duct mount transmitters):
if the reference humidity is below 65 %RH use the "DRY" trimmer
potentiometer
if the reference humidity is above 65 %RH use the "WET" trimmer
potentiometer
HMP130Y series:
if the reference humidity is below 65 %RH use the trimmer potenti-
ometer R5 (offset)
if the reference humidity is above 65 %RH use the trimmer potenti-
ometer R18
Wait until the readings stabilize and adjust the transmitter reading to corre­spond to the reference reading or adjust the difference reading to zero. Turn the power of the HMI38 off and on again to return to normal operation.
41
HMI38 Operating Manual U063en-2.1
9. MAINTENANCE
9.1. Reference measurements
Reference measurements help to verify whether the instrument readings are within specifications. This way the user can check if the probe or the indicator needs calibration or service.
The reference meter should preferably use the same technology as the instru­ment to be checked, i.e. equipment with capacitive sensors should be checked with instruments using capacitive sensors. This minimizes the risk of errors caused by different measurement techniques.
Whatever technique is used, make sure that the reference instrument and the instrument to be checked are at the same temperature in order to avoid errors caused by temperature differences. The reference measurements are most accurate when made in laboratories where the conditions are stable. Make the reference measurement as close to the sensor to be checked as possible and read the readings simultaneously, when possible.
9.2. Self-diagnostics
The HMI38 goes through a self-diagnostics procedure when the power is switched on. If the procedure does not reveal any errors or faults, the instru­ment the normal operation. If the procedure finds errors or faults, check first that the probes are connected, the initialization of the probes is correct and the humidity and temperature sensors are not damaged. If errors still appear, follow the troubleshooting instructions in Appendix 3 to find out whether the fault is in a probe or the HMI38. Send the faulty instrument to Vaisala or a Vaisala representative for repairs.
If any errors occur during operation, the error messages appear on the LCD display if the instrument displays measurements; if the menus are in use, error messages do not appear. During operation, however, the error messages will not output automatically through the serial interface. If there is any reason to doubt that there is something wrong with the instrument, use command ERRS:
If there are no error messages, only a prompt is displayed:
>ERRS <cr> >
When errors have occurred, the instrument outputs the error code (see Appen­dix 3 for all error messages):
ERRS <cr>
42
>ERRS <cr> E40 f ( all ) out of range >
HMI38
U063en-2.1 Operating Manual
9.3. Adjusting the contrast of the display
Adjust the contrast of the display with the trimmer R9 "CONTRAST". Release the six screws at the bottom of the HMI38 and remove the top cover of the indicator body. The trimmer is located at the front end of the main board.
10. TECHNICAL DATA
10.1. Relative humidity
Measuring range 0...100 %RH
Accuracy (at +20°C) 0.1 %RH (for overall system ac-
curacy see probe specifications in
Appendix 4, 5 or 6) Resolution 0.1 %RH Temperature dependence over the
entire operating temperature range ±0.2 %RH
10.2. Temperature
Measuring range -40...+120 °C (depending on the
Accuracy (error caused by HMI38 at ±0.1°C +20°C)
Resolution 0.1°C Typical temperature dependence of
electronics 0.004 °C/°C
10.3. Calculated quantities
Widest ranges dewpoint temperature -40...+100 °C (depending on the mixing ratio 0...500 g/kg dry air probe used) absolute humidity 0...600 g/m
wet bulb temperature 0...+100 °C
probe used, see Appendices 4, 5
and 6)
3
Accuracies of the calculated quantities depend on the accuracies of the meas­ured quantities; see Appendices 4, 5 or 6 for more details.
43
HMI38 Operating Manual U063en-2.1
10.4. Data logging
Storage capacity: continuous mode 254 measurements per probe catch mode 127 measurements per probe
Logging interval from 10 s to 99 h
10.5. Analogue outputs
Two analogue outputs
ranges 0...1 V and 0...5 V (freely adjust-
able between 0...5 V) Typical accuracy of analogue output
at +20 °C ±0.05 % full scale Typical temperature dependence of
analogue output 0.005 %/°C full scale
10.6. Serial interface
Serial port types RS 232C and RS 485 (only one usable at a time)
Communication parameters
bauds 300, 600, 1200, 2400, 4800, 9600 data bits 7 or 8 stop bits 1 or 2 parity even, odd or none
mode full or half duplex Connector D-type female D9S Data format ASCII characters
10.7. Electronics
Display 2 x 16 character alphanumeric
character height 3.85 mm (0.15")
high-contrast, wide view angle LCD
44
Keyboard 4 x 4 tactile membrane switches Probes (max. 2) HMP35E, HMP36E or HMP37E
(with some modifications also HMP35B or HMP36B - type changes into BM)
HMI38
U063en-2.1 Operating Manual
Rechargeable battery pack 8 NiCd 1.2 V cells, standard ’AA’
cell size
Battery operating time in
continuous use 8 hours
data-logging mode 24 hours
charging time 6 hours Supply voltage 12 VDC (11...16 VDC), 0.2 A Power consumption during charge 3 W maximum Recommended external load for
0...1 V voltage output >2 k (to ground)
Recommended external load for
0...5 V voltage output >10 k (to ground)
Operating temperature (electronics) 0...+50 °C
Storage temperature range -20...+70 °C Operating humidity range 95 %RH maximum, non-
Storage humidity range 90 %RH maximum
10.8. Mechanics
Housing material ABS plastic Housing dimensions (h x w x l) 90 x 200 x 295 mm
condensing
NOTE
The operating humidity range is higher than the storage humidity range because the air in a warm instrument has a lower relative humidity than the ambient air.
Accessories optional carrying case Weight 1090 g
45
HMI38 Operating Manual U063en-2.1
10.9. Electromagnetic compatibility
10.9.1. Emissions
Test: Setup according to:
Radiated interference EN55022 (class B)
10.9.2. Immunity
Test: Setup according to: Performance:
Electrostatic discharge EN61000-4-3:1995 criteria B Electrical fast transients IEC 801-4:1988 criteria B RF-radiated fields IEC 801-3:1984 criteria A * GSM field immunity ENV50204:1995 criteria A
* (additional test)
Vibration IEC 68-2-6 Fc (0.15 mm/10...500 Hz/2 g/20 sweep cycles)
11. ACCESSORIES AND SPARE PARTS
Order code Description 18326 Carrying case 17371 AC adapter for 230 VAC with EURO plug 18100ZZ Connection cable for the calibration of HMD/W20/30 and
HMP130Y series transmitters 18200ZZ Connection cable for the calibration of HMP230 transmitters 18300ZZ Connection cable for the calibration of HMD/W60/70 and
HMP140 series transmitters 0606 Rechargeable 1.2 V NiCd battery
must be ordered in batches of 8 pcs HMP35E Humidity and temperature probe for general measurement HMP36E Humidity and temperature probe for measurement of granu-
lates and concrete HMP37E Humidity and temperature probe for tight or remote places or
high temperatures (5 m cable)
46
Loading...