Vaisala HMDW6070, HMP140 User Manual

USER'S GUIDE
Calibration of Series HMD/W60/70 and
HMP140 Transmitters with
Vaisala HUMICAP® Indicator HMI41
U218EN11
PUBLISHED BY Vaisala Oyj Phone (int.): +358 9 8949 1
P.O. Box 26 Fax: +358 9 8949 2227 FIN-00421 Helsinki Finland
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2006 No part of this manual may be reproduced in any form or by any means, electronic or
mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual

Contents

1. INTRODUCTION...................................................................................................................... 1
1.1 HMI41 with calibration option ......................................................................................1
1.2 Selecting the HMI41 calibrator function ......................................................................1
2. OUTPUT SCALE SELECTIONS...............................................................................................3
2.1 Standard transmitter settings............................................................................... 3
2.2 Getting started ..............................................................................................................3
2.3 Scale selection modes ..................................................................................................4
2.3.1 Voltage output selection mode (U)...................................................................4
2.3.2 Current output selection mode (I) .....................................................................5
2.3.3 Customized voltage output selection mode (cU) ..............................................6
2.3.3.1 Humidity channel...........................................................................................7
2.3.3.2 Temperature channel.....................................................................................9
2.3.4 Customized current output selection m ode (cI) .............................................. 10
2.3.4.1 Humidity channel.........................................................................................11
2.3.4.2 Temperature channel...................................................................................12
3. CALIBRATION.......................................................................................................................13
3.1 Getting started ............................................................................................................13
3.2 Humidity calibration....................................................................................................15
3.3 Temperature calibration ............................................................................................. 16
APPENDIX 1: QUICK REFERENCE TO SCALE SELECTIONS AND CALIBRATION
1996-08-29 i
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
This page intentionally left blank.
ii 1996-08-29
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual

1. INTRODUCTION

1.1 HMI41 with calibration option

The HMI41 can be used as a field calibrator for various Vaisala humidity transmitters. There are three different calibration cables available:
19116ZZ for the calibration of HMD/W60/70 and HMP140
series analogue transmitters
19164ZZ for the calibration of HMP230 series digital
transmitters
19165ZZ for the calibration of the HMD/W20/30 and
HMP130 series analogue transmitters
In this manual, only the calibration of the HMD/W60/70 and HMP140 series analogue transmitters is dealt with. For other calibration options, please refer to corresponding manuals.
Calibration is quick and easy to perform. Yet, the HMI41 includes four output scale selection modes which cover all scalable features of analogue transmitters and provide for a simple and reliable on-site calibration of even customized transmitters. In calibration, the HMI41 measures relative humidity and temperature with a separate probe, and receives the RH and T information from the transmitter. These values are then compared on the HMI41 display and the transmitter is adjusted with calibration potentiometers inside the transmitter.

1.2 Selecting the HMI41 calibrator function

The desired calibrator function is selected in the HMI41 setup menu. To enter the setup mode, first press the ON/OFF button and the following appears:
RH
Td T
P setminmax hysthold LobatHi
% °C °F
°C
°F
Pahg/m
Then release the ON/OFF button and within 1...2 seconds press both ENTER and MODE buttons until the following text appears on the display:
After a few seconds, the text changes automatically to show the following:
1996-08-29 1
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
°C
set
If the basic settings of the HMI41 (display units, automatic power off function, display quantities and pressure) have to be changed, please refer to the HMI41 Operating Manual. Otherwise, press ENTER repeatedly until the following text appears:
set
The number on the first line of the display corresponds to the following HMI41 functions:
1 = HMI41 as a humidity and temperature indicator
2 = HMI41 as a calibrator for HMD/W60/70 and HMP140 series analogue transmitters
3 = HMI41 as a calibrator for HMP230 series digital transmitters 4 = HMI41 as a calibrator for the HMD/W20/30 and HMP130 series analogue transmitters
To calibrate the HMD/W60/70 or the HMP140 series transmitters, select number 2 with buttons (number up) and (number down) and then turn the HMI41 off. When the HMI41 is turned on again, it will automatically wake up as a calibrator for these analogue transmitters. If you wish to calibrate transmitters of some other type or use the HMI41 as an indicator, this selection has to be done accordingly; refer to corresponding manuals for further details.
Note that if the default settings in the HMI41 memory have not been changed (e.g. when it is used as a calibrator for the first time), the HMP140 series transmitters with standard settings and voltage outputs can be directly calibrated (see Chapter 3). In all other cases scale selections are necessary; please, consult Chapter 2. The default settings in the HMI41 memory (Table
1.2) are returned with U, I, cU or cI selections according to which changes have been made.
Humidity range Temperature range Voltage range
0...1V
0...100 %RH
-40...+60 °C
0...5V
0...10V
Table 1.2 Defa ult settin gs in the H MI41 memory
2 1996-08-29
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual

2. OUTPUT SCALE SELECTIONS

2.1 Standard transmitter settings

In Table 2.1 you will find the standard settings of the HMD/W60, HMD/W70 and the HMP140 series transmitters.
Transmitter Relative humidity
scaling range
HMP140
HMD60
HMD70
HMW60
HMW70
0...100 %RH
0...100 %RH
0...100 %RH
0...100 %RH
0...100 %RH
Table 2.1 Standard settings of analogue transmitters
Temperature scaling range
-40...+60 °C
-20...+80 °C
-20...+80 °C
-5...+55 °C
-5...+55 °C
NOTE
Voltage
Current output
output
0...1 V
0...5 V 0...20 mA
0...10 V
4...20 mA
0...1 V
0...5 V 0...20 mA
0...10 V
4...20 mA
0...1 V
0...5 V
0...10 V
When calibrating transmitters with current outputs, make sure that the output loop is terminated.

2.2 Getting started

After having selected the calibrator function 2 as explained in Chapter 1.2, turn the HMI41 on with the ON/OFF button and wait until the following text appears on the display:
RH
T
This message remains on the display for about 5 seconds; if no button is pressed, the HMI41 goes on to calibration. To enter the scale selections, press any button (except ON/OFF) within 5 seconds and the following appears:
1996-08-29 3
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
Scale selection mode is entered when this text is on the display.

2.3 Scale selection modes

The HMI41 has four scale selection modes: voltage output (U) and current output (I) selection modes are used for transmitters with standard settings. Customized voltage output (cU) and customized current output (cI) selection modes are meant for analogue transmitters with customized RH, T and/or output range settings. All these selections are stored in the HMI41 memory; therefore, if the transmitter settings or the selections in the HMI41 memory have not been changed by the next calibration, you can calibrate the transmitter directly. In the following, each selection mode is explained in its own section.

2.3.1 Voltage output selection mode (U)

If the transmitter has voltage outputs, its temperature range is given to the HMI41 with the voltage output selection mode (U). The humidity range is always 0...100 %RH. Select the U selection mode with buttons or and acknowledge the selection with ENTER. A text similar to the following is displayed:
C T
set
°C
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit with buttons (number up) or (number down) to correspond to that of the transmitter. Acknowledge the setting with ENTER:
4 1996-08-29
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual
C T
°C
set
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Hi. You can now set the upper limit with buttons (number up) or (number down) to correspond to that of the transmitter. Acknowledge the setting with ENTER and turn the HMI41 off with the ON/OFF button.You can now calibrate the transmitter (see Chapter
3).

2.3.2 Current output selection mode (I)

NOTE
When calibrating the transmitters with current outputs, make sure that the output loop is terminated.
If the transmitter has current outputs, its temperature range is given to the HMI41 with the current output selection mode (I). The RH range is always
0...100 %RH. Select the I selection mode with buttons or and acknowledge the selection with ENTER. A text similar to the following is displayed:
set
Numbers on the display are blinking, and you can now select the output (0...20 or 4...20 mA) with buttons or according to the transmitter to be calibrated. Acknowledge the selection with ENTER and a text similar to the following is displayed:
1996-08-29 5
Loading...
+ 21 hidden pages