Vaisala HMDW2030, HMP130 User Manual

USER'S GUIDE
Calibration of Series HMD/W20/30 and
HMP130 Transmitters with
Vaisala HUMICAP® Indicator HMI41
U220EN11
PUBLISHED BY Vaisala Oyj Phone (int.): +358 9 8949 1
P.O. Box 26 Fax: +358 9 8949 2227 FIN-00421 Helsinki Finland
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2006 No part of this manual may be reproduced in any form or by any means, electronic or
mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
HMI41 Calibration Option
19165ZZ-U220en-1.1 Operating Manual

Contents

1. INTRODUCTION...................................................................................................................... 1
1.1 HMI41 with calibration option ......................................................................................1
1.2 Selecting the HMI41 calibrator function ......................................................................1
2. CALIBRATION.........................................................................................................................3
2.1 Getting started ..............................................................................................................3
2.2 Calibration..................................................................................................................... 4
APPENDIX 1: QUICK REFERENCE GUIDE
1996-08-29 i
HMI41 Calibration Option Operating Manual 19165ZZ-U220en-1.1
This page intentionally left blank.
ii 1996-08-29
HMI41 Calibration Option
19165ZZ-U220en-1.1 Operating Manual

1. INTRODUCTION

1.1 HMI41 with calibration option

The HMI41 can be used as a field calibrator for various Vaisala humidity transmitters. There are three different calibration cables available:
19116ZZ for the calibration of Vaisala’s HMD/W 60/70 and
HMP140 series analogue transmitters
19164ZZ for the calibration of HMP230 series digital transmitters
19165ZZ for the calibration of the HMD/W 20/30 and
HMP130 series analogue transmitters
In this manual, only the calibration of HMD/W 20/30 and HMP130 series analogue transmitters is dealt with. For other calibration options, please refer to corresponding manuals.
Calibration is quick and easy to perform. In calibration, the HMI41 measures relative humidity with the reference probe and receives the RH information from the transmitter. These two values are then compared on the HMI41 display and the transmitter is adjusted with calibration potentiometers inside the transmitter.

1.2 Selecting the HMI41 calibrator function

The desired calibrator function is selected in the HMI41 setup menu. To enter the setup mode, first press the ON/OFF button and the following appears:
RH
Td T
P setminmax hysthold LobatHi
% °C °F
°C
°F
Pahg/m
Then release the ON/OFF button and within 1...2 seconds press both ENTER and MODE buttons until the following text appears on the display:
After a few seconds, the text changes automatically to show the following:
1996-08-29 1