Vaisala GMM220 transmitter modules are versatile instruments for measuring CO2 in
industrial applications. The CARBOCAP® sensor is silicon based and its operation is
based on the NDIR Single-Beam Dual-Wavelength principle. The modules can be
configured for different CO2 measurement ranges and analog output ranges.
Figure 1 Dimensions of the Probes and Component Board of the GMM220
Series (in mm)
CO
Transmitter Modules
2
The GMM220 module consists of a component board, cable and a CO2 probe. The
shorter probe model (GMP221) is for higher and the longer model (GMP222) for lower
CO2 concentrations. A waterproof connector connects the probe to the cable. The probe
also features a non-volatile memory for storing the calibration parameters. This enables
interchangeability of the probes and allows easy field calibration by exchanging the
probe with a freshly calibrated one.
Mounting
For power supply and analog outputs, the board is available either with pinheaders on
bottom side or with screw terminals on top side; serial interface has always a pinheader.
Figure 2 Dimensions of the Component Board (in mm)
If the board has pinheaders, mount it by soldering or plugging it into host board's female
connectors.
GMM220_____________________________________________________ CO2 Transmitter Modules
LEDs:
- Normal operation: Led V8 (red) is not lit and Led V9 (green) is blinking.
- Measurement out of range: Led V8 (red) is blinking and Led V9 (green) is blinking,
maximum value is shown. GMM220 returns to normal operation when CO2 drops
within range.
- Error state: Led V8 (red) is blinking and Led V9 (green) is blinking, no readings. In
error state, check first that the probe is properly connected. In case of constant error,
please contact Vaisala Service Centers.
Serial Interface Protocol
The transmitter is connected to a computer with the optional serial COM adapter
(19040GM). A suitable terminal program such as WINDOWS® Hyper Terminal is
needed for the communication.
Accuracy (including repeatability, non-linearity and calibration uncertainty)
at 25 °C and 1013 hPa
GMM221 ±1.5% of range + 2 % of reading
(applies for concentrations above 2% of full scale)
GMM222 ±1.5% of range + 2 % of reading
Nonlinearity ± 0.5% FS
Typical output temperature
dependence -0.3% of reading / °C
Typical pressure
dependence +0.15 of reading/hPa
Long-term stability <±5 %FS / 2 years
Response time (63%) 20 seconds (GMM221)
30 seconds (GMM222)
Warm-up time 30 seconds, 15 minutes (full specifications)
Inputs and Outputs
Outputs 0 ... 20 or 4 ... 20 mA, 0 ... 1 V, 0 ... 2V
0 ... 2.5 V, or 0 ... 5 V
Analog output resolution 12 bits
Recommended external load:
current output max. 200 Ω
voltage output min. 1000 Ω
Operating voltage 11 … 20 VDC or 18 … 30 VDC
Connections Screw terminals, wire size 0.5 ... 1.5 mm2l
Power consumption < 2.5 W
Operating conditions
Operating temperature -20…+60 °C (-4 ... +140 °F)
Storage temperature -30 ... +70 °C (-22 ... +158 °F)
Operating pressure 700 ... 1300 hPa
Humidity range
probe 0 … 100 %RH (non-condensing)
mother board 0 … 85 % RH (non-condensing)
Mechanics
Probe:
housing material PC plastic
housing classification IP65
Weight:
GMM221(w/ 2m cable) max. 180 g
GMM222 (w/ 2m cable) max. 200 g
GMM220_____________________________________________________ CO2 Transmitter Modules
Order code
Description
GMP221, GMP222
Spare probe (use the order form to defi ne measurement range etc.)
25245GM
Clips (2 pcs) for attaching the probe
GMK220
CO2 Calibration Unit
GM45156
Mounting flange for the probe
GMM220Z100SP
1 m high temperature probe cable (180 °C / 365 °F)
GMM220Z200SP
2 m probe cable
GMP343Z600SP
6 m probe cable
GMP343Z1000SP
10 m probe cable
19040GM
Serial COM adapter
*M010115EN*
Compliance With EMC-Standards
- EN 61326-1:1997 + Am1:1998, Electrical equipment for measurement, control and
laboratory use - EMC requirements - Generic environment
- Testing:
- CISPR16/22 class B
- EN/IEC 61000-4-2
- EN/IEC 61000-4-3
- EN/IEC 61000-4-4
- EN/IEC 61000-4-5
- EN/IEC 61000-4-6.
Warranty
Visit our Internet pages for standard warranty terms and conditions:
http://www.vaisala.com/warranty.
Please observe that any such warranty may not be valid in case of damage due to normal
wear and tear, exceptional operating conditions, negligent handling or installation, or
unauthorized modifications. Please see the applicable supply contract or Conditions of
Sale for details of the warranty for each product.
Accessories
Technical Support
For technical questions, contact the Vaisala technical support by e-mail at
helpdesk@vaisala.com. Provide at least the following supporting information:
- Name and model of the product in question
- Serial number of the product
- Name and contact information of a person who can give information on the problem.
Product Returns
If the product must be returned for service, see www.vaisala.com/returns.
For contact information of Vaisala Service Centers, see