Vaisala GMM220 transmitter modules are versatile instruments for measuring CO2 in
industrial applications. The CARBOCAP® sensor is silicon based and its operation is
based on the NDIR Single-Beam Dual-Wavelength principle. The modules can be
configured for different CO2 measurement ranges and analog output ranges.
Figure 1 Dimensions of the Probes and Component Board of the GMM220
Series (in mm)
CO
Transmitter Modules
2
The GMM220 module consists of a component board, cable and a CO2 probe. The
shorter probe model (GMP221) is for higher and the longer model (GMP222) for lower
CO2 concentrations. A waterproof connector connects the probe to the cable. The probe
also features a non-volatile memory for storing the calibration parameters. This enables
interchangeability of the probes and allows easy field calibration by exchanging the
probe with a freshly calibrated one.
Mounting
For power supply and analog outputs, the board is available either with pinheaders on
bottom side or with screw terminals on top side; serial interface has always a pinheader.
Figure 2 Dimensions of the Component Board (in mm)
If the board has pinheaders, mount it by soldering or plugging it into host board's female
connectors.
M010115EN-E _______________________________________________________________________ 1
GMM220_____________________________________________________ CO2 Transmitter Modules
Electrical Connections
Figure 3 Electrical Connections
Power supply:
Two supply voltage alternatives:
- +L 12 volt supply input (11 VDC - 20 VDC)
- +H 24 volt supply input (18 VDC - 30 VDC)
- GND common ground for power supply
Average current consumption is 100 mA (+ current output, max. 20 mA)
Analog outputs:
- 0 common for both analog outputs
- V voltage output (default 0-1 V, on request 0-2 V, 0-2.5 V or 0-5V)
- mA current output (default 4-20 mA, on request 0-20 mA)
Serial interface:
The interface is a 5 V (TTL) level full duplex serial port.
- GND signal ground (same potential as the supply GND)
- TX data transmitted by the CO2 module (low state corresponding mark)
- RX data received by the CO2 module (low state corresponding mark)
Leave other serial interface pins unconnected.
Probe wires: If the CO2 module is ordered with a separate probe cable, solder the wires
to the component board according to colors shown in Figure 3.