Vaisala GMM20W User Manual

GMM20W CO
Sensor Systems Division U300en-1.1
module
2

INTRODUCTION

Vaisala’s GMM20W module uses a completely new sensor technology. The silicon based CARBOCAP sensor provides for excellent stability and reliability. The GMM20W module requires almost no maintenance: the recommended calibration interval is five years. This combined with the high performance makes the GMM20W modules an ideal choice for OEM applications related to ambient CO measurement.
72 mm

ELECTRICAL CONNECTIONS

The nominal 24 V supply must be connected between terminals + and - on the mother board. The analogue output is available at remaining terminals. The common wire is connected to terminal 0 and the other wire either to terminal V (voltage output) or to terminal mA (current output). The current output is chosen with the jumper 0/4mA (see Figure 3): 4...20mA is chosen by connecting the jumper (default) and the
2
0...20 mA is chosen by disconnecting the jumper. If the relay output of display models is used, note that the wiring is done at the back of the display board. The default relay trigger point has been set to 1000 ppm. This can be changed with the optional software kit 19222GM. For further details, contact your local Vaisala representative.
74 mm
Figure 1 Dimensions
Figure 2 Electrical connections and
leds V7, V10
Note that during normal operation the green led (V10) at the lower righthand corner is blinking. In case the self­diagnostics procedure comes across some abnormality, the red led (V7) lights up (see Figure 2 above).
1999-11-16
GMM20W CO
Sensor Systems Division U300en-1.1
module
2

SERVICE AND MAINTENANCE

The GMM20W module has an excellent stability and requires almost no mainte­nance. A full after sales calibration and service facility is naturally provided by Vaisala and its distributors.
In case the user prefers to do the calibra­tion himself, calibration gases and equipment are also available. Vaisala's portable CO2 meters together with the Calibration Kit provide for easy and convenient on-site calibration checks (see the corresponding manuals for detailed instructions).
GMM20W should be adjusted if the reading differs too much from the reference value during checking (note that in any case, only slight differences are expected). For this purpose, we recommend Vaisala’s Software Kit 19222GM which includes a floppy disk and a serial COM adapter.
Connect the cable to the connector marked “Serial Com” on the main PCB of the module and to the connector A on the serial COM adapter (see Figure 3).
In order to achieve full accuracy, the module has to be calibrated against accurate and traceable calibration gases in stable environmental conditions (temperature, pressure). Accurate cali­brations are usually performed in labora­tories; in these calibrations, temperature and pressure corrections have to be made. For further details, consult your local Vaisala representative.
1999-11-16
Figure 3 Connection of the serial communication cab le
Loading...
+ 3 hidden pages