GMM111 Carbon Dioxide Module
www.vaisala.com
The Vaisala CARBOCAP® Carbon Dioxide Module GMM111 is a CO2 measurement
module with flow-through aspiration.
Features/Benefits
Compact CO2 module with
▪
flow-through aspiration
Ideal for control of CO2
▪
concentrations in incubators
Incorporates Vaisala
▪
CARBOCAP®, the silicon based
NDIR sensor with unique
internal referencing
Advanced single-beam, dual
▪
wavelength measurement with
no moving parts
Measurement range options
▪
0 ... 5 %, 0 ... 10 % and
0 ... 20 % CO
Excellent long-term stability
▪
2
GMM111
The Vaisala CARBOCAP® Carbon
Dioxide Module GMM111 is
designed especially for control of
biological processes where high
CO
concentrations are used. It has
2
3 optional measurement ranges
0 ... 5/10/20 % CO
ow-trough model and has barbed
connectors for attaching the in and
out ow tubes. As the module is not
mounted in the chamber, the chamber
can be heatsterilized without
removing the module.
The Vaisala CARBOCAP
sensors have been proven to be
accurate anddurable. They have an
excellent long-term stability, which
decreases maintenance. The
superior performance of Vaisala
. The GMM11 is a
2
®
CO2
CARBOCAP
from the stable reference provided by
the electrically tunable Fabry-Perot
Interferometer(FPI).
The tunable FPI lter measures CO
absorption, and simultaneously a
reference wavelength. This internal
reference measurement compensates
effectively for any changes in the
optical path, such as light source
intensity changes and contamination.
In the HVAC market, this type of
reference measurement is a unique
feature to Vaisala CARBOCAP
products.
The true internal reference
measurement of Vaisala CARBOCAP
CO
transmitters provides years of
2
stable CO
®
sensors results largely
measurements.
2
2
®
®
Technical data
TUBE I.D. Ø4
(TUBE O.D. Ø6)
mA
V
0
B
A
0
24V
mA
V
0
B
A
0
24V
25
65 6
30
60
96
Ø3.2
Performance
CO2 measurement range 0 ... 5 %, 0 ... 10 % or 0 ... 20 %
Accuracy (including repeatability, non-linearity
and calibration uncertainty) ±(1.5% of range + 3 % of reading)
Long-term stability ±1 % CO
Response time T
90
1 min at 0.5 l/min flow
/2 years
2
Flow rate dependence
< 1 l/min flow no effect
1 ... 10 l/min flow 4 % of reading/ l/min
Temperature dependence, typical -0.3 % of reading/°C
Pressure dependence, typical +0.15 % of reading/hPa
Warm-up time 1 min, 10 min for full
specifications
Product lifetime
> 10 years
Operating environment
Temperature +5 ... +55 °C (+41 ... +131 °F)
Humidity 0 ... 99 % RH non-condensing
Pressure 700 ... 1200 hPa
Gas flow
operating range < 10 l/min
recommended range 0.2 ... 0.8 l/min
Electromagnetic compatibility
Complies with EMS Standard EN61326-1, Generic Environment
Dimensions
Dimensions in mm
Inputs and outputs
Outputs 4 ... 20 mA, 0 ... 10 V
RS485, 2-wire, non-isolated
Operating voltage 24 V (±20 %) AC/DC
Power consumption <2 W
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Ref. B210566EN-B ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.