Vaisala GMM111 Technical data

GMM111 Carbon Dioxide Module
www.vaisala.com
The Vaisala CARBOCAP® Carbon Dioxide Module GMM111 is a CO2 measurement module with flow-through aspiration.
Compact CO2 module with
flow-through aspiration
Ideal for control of CO2
concentrations in incubators
Incorporates Vaisala
CARBOCAP®, the silicon based NDIR sensor with unique internal referencing
Advanced single-beam, dual
wavelength measurement with no moving parts
Measurement range options
0 ... 5 %, 0 ... 10 % and 0 ... 20 % CO
Excellent long-term stability
2
GMM111
The Vaisala CARBOCAP® Carbon Dioxide Module GMM111 is designed especially for control of biological processes where high CO
concentrations are used. It has
2
3 optional measurement ranges 0 ... 5/10/20 % CO ow-trough model and has barbed connectors for attaching the in and out ow tubes. As the module is not mounted in the chamber, the chamber can be heatsterilized without removing the module.
The Vaisala CARBOCAP sensors have been proven to be accurate anddurable. They have an excellent long-term stability, which decreases maintenance. The superior performance of Vaisala
. The GMM11 is a
2
®
CO2
CARBOCAP from the stable reference provided by the electrically tunable Fabry-Perot Interferometer(FPI).
The tunable FPI lter measures CO absorption, and simultaneously a reference wavelength. This internal reference measurement compensates effectively for any changes in the optical path, such as light source intensity changes and contamination. In the HVAC market, this type of reference measurement is a unique feature to Vaisala CARBOCAP products.
The true internal reference measurement of Vaisala CARBOCAP CO
transmitters provides years of
2
stable CO
®
sensors results largely
measurements.
2
2
®
®
Technical data
TUBE I.D. Ø4
(TUBE O.D. Ø6)
mA
V 0 B A 0
24V
mA
V 0 B A 0
24V
25
65 6
30
60
96
Ø3.2
Performance
CO2 measurement range 0 ... 5 %, 0 ... 10 % or 0 ... 20 % Accuracy (including repeatability, non-linearity and calibration uncertainty) ±(1.5% of range + 3 % of reading) Long-term stability ±1 % CO Response time T
90
1 min at 0.5 l/min flow
/2 years
2
Flow rate dependence
< 1 l/min flow no effect
1 ... 10 l/min flow 4 % of reading/ l/min Temperature dependence, typical -0.3 % of reading/°C Pressure dependence, typical +0.15 % of reading/hPa Warm-up time 1 min, 10 min for full
specifications
Product lifetime
> 10 years
Operating environment
Temperature +5 ... +55 °C (+41 ... +131 °F) Humidity 0 ... 99 % RH non-condensing Pressure 700 ... 1200 hPa Gas flow
operating range < 10 l/min
recommended range 0.2 ... 0.8 l/min Electromagnetic compatibility
Complies with EMS Standard EN61326-1, Generic Environment
Dimensions
Dimensions in mm
Inputs and outputs
Outputs 4 ... 20 mA, 0 ... 10 V
RS485, 2-wire, non-isolated Operating voltage 24 V (±20 %) AC/DC Power consumption <2 W
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210566EN-B ©Vaisala 2009
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.