Vaisala DMT242 User Manual

Vaisala DRYCAP
®
Dewpoint Transmitter DMT242
USER'S GUIDE June 2008 | M010036EN-D
INTRODUCTION
DMT242 is a dewpoint transmitter for wide range of OEM applications. DMT242 measures dewpoint with excellent long term stability which is maintained automatically with the patented auto-calibration procedure. The Vaisala DRYCAP® polymer sensor technology used in DMT242 is also durable against dew in case condenced water exists in the process during system malfunction. The product mechanics have been designed for harsh environments requiring protection against dust, dirt and splashed water.
The disconnection and reconnection of the transmitter is easy with the connector where the output signal and supply voltage wires are connected. The unit also has a serial line for rescaling the analog output.
PRESSURE SETTING PROCEDURE FOR PRESSURIZED PROCESSES
For achieving the most accurate measurements in pressurized processes, set the process pressure to DMT242 according to Figure 1 by using the pressure switches (see Figure 3, item
8). As shipped from factory, the pressure switch setting is 1 bar, as in switch number 4 in the ON position.
MOUNTING
Switches Pressure (absolute)
2341
(bar) 0 … 2 2 … 4
4 … 6 6 … 8 8 … 10
10 … 12 12 … 14 14 … 16 16 … 18 18 … 20
THE SWITCH MARKED WITH A BLACK SQUARE IS IN
ON POSITION
THE
Figure 1 Pressure Setting Table
1. Insert the sealing washer (see Figure 1) on
the probe and set the probe through the fitting of the process pipe. The probe has G½" ISO 228/1 parallel thread.
2. Fasten the transmitter to the fitting of the
process pipe by tightening from the nut of the probe (24 mm).
1 = Cover 2 = Connector 3 = Flat gasket 4 = Counter connector 5 = Transmitter body 6 = Nut 7 = Sealing washer
WIRING
8 = Probe
1. Remove the cover.
2. Take out the connector.
3. Take out the screw terminal from the
connector by pushing it out, for example, with the fixing screw.
4. Use a three-wire cable. Suitable 2 m or
10 m optional cable is available from Vaisala (items: 221475 for 2 m cable and 221476 for 10 m cable). Connect the wires to the connector terminals as follows:
Terminal nr 1 = V supply + (VAC line) Terminal nr 2 = V supply - (VAC neutral)/signal ­Terminal nr 3 = Signal + Leave the ground terminal free
Figure 2 DMT242
­+
Wire colours in cables 221475 and 221476:
1 = brown 2 = blue 3 = black
Vaisala DRYCAP
®
Dewpoint Transmitter DMT242
USER'S GUIDE June 2008 | M010036EN-D
5. Insert the wired screw terminal back into the connector excatly in the position indicated in Figure 3.
Push the fixing screw through the connector. Fasten the cable clamp. Insert the wired connector into the counter connector.
NOTE
The connection is incompatible if the positioning of the screw terminal is NOT as indicated in Figure 3. Fasten the fixing screw.
6. Install the back cover allowing the cable to run through the hole in the cover. The transmitter is ready
for use.
NOTE
When the power is switched on, wait about 7 minutes before taking measurements. Start-up self-diagnostics freeze the output during the first operation minutes.
SCREW TERMINAL MUST BE INSTALLED TO THE CONNECTOR IN THE POSITION OF THE FIGURE
(TERMINALS 2 AND 3 AS UPPER EDGE)
1 = Connector 2 = Screw terminal of the connector 3 = Connector 4 = Cable clamp 5 = RS-232 serial port 6 = Terminals 2 and 3 7 = Three-wire cable 8 = Pressure setting switches 9 = Fixing screw
Figure 3 Wiring and Installation of Connector
CALIBRATION AND MAINTENANCE
Replacing the sintered filter
If the sintered filter is dirty it can prolong the response time of the measurement. If the filter needs to be changed unscrew the filter and replace the old filter with a new one.
Calibration
It is recommended to perform a calibration check every other year. A field check can be carried out by using a calibrated reference probe and comparing the readings measured with the transmitter and the reference probe. The Vaisala DRYCAP® Hand-Held Dewpoint Meter DM70 is ideal for confirming the performance of the transmitter in the field. By using a connection cable the readings of DMT242 and DM70 can be viewed simultaneously on the display of DM70. If there is need for adjustement contact Vaisala Instruments Service Centers or local Vaisala representative.