VACUUBRAND MV 10 NT User Manual

Part I

page 1 of 126
Technology for Vacuum Systems
Instructions for use
Part I of II
Speed controlled diaphragm pumps
ME 16 NT VARIO
MD 12 NT VARIO
MV 10 NT VARIO
page 2 of 126
Dear customer,
Your VACUUBRAND diaphragm pumps are designed to provide you with many years of trouble-free service with optimal performance. Our many years of practical experience allow us to provide a wealth of application and safety information. Please read these instructions for use before the initial operation of your pump. VACUUBRAND diaphragm pumps combine our many years of experi­ence in design, construction and practical operation, with the latest de­velopments in material and manufacturing technology.
Our quality maxim is the ”zero defect” principle: Every diaphragm pump, before leaving our factory, is tested intensively, including an endurance run of 14 hours. Any faults, even those which oc-
cur rarely, are identied and can be eliminated immediately.
After completion of the endurance run, every pump is tested, and must
achieve specications before shipment.
We are committed to providing our customers only pumps that meet this high quality standard. While our pumps cannot eliminate all of your work, we design, manufac­ture and test them to ensure that they will be an effective and trouble-free tool to assist you in that work.
Yours, VACUUBRAND GMBH + CO KG
After sales service: Contact your local dealer or call +49 9342 808-5500.
Trademark index:
VACUU•LAN® (US-Reg.No 3,704,401), VACUU•BUS®, VACUU•CONTROLTM,
chemistry-HYBRIDTM, Peltronic®, TURBO•MODETM, VARIO® (US-Reg.No 3,833,788), VARIO-SPTM, VACUUBRAND® (US-Reg.No 3,733,388) and also the shown company logos are trademarks of VACUUBRAND GMBH + CO KG in Germany and/or other countries.
page 3 of 126
DE Achtung: Die vorliegende Betriebsanleitung ist nicht in allen EU-Sprachen verfügbar. Der Anwender darf die beschriebenen Geräte nur dann in Betrieb nehmen, wenn er die vorliegende Anleitung versteht oder eine fachlich korrekte Übersetzung der voll­ständigen Anleitung vorliegen hat. Die Betriebsanleitung muss vor Inbetriebnahme der Geräte vollständig gelesen und verstanden werden, und alle geforderten Maß­nahmen müssen eingehalten werden. ”Sicherheitshinweise für Vakuumgeräte”
EN Attention: This manual is not available in all languages of the EU. The user must not operate the device if he does not understand this manual. In this case a technically correct translation of the complete manual has to be available. The manual must be completely read and understood before operation of the device and all required measures must be applied. ”Safety instructions for vacuum equipment”
FR Attention: Le mode d’emploi présent n’est pas disponible dans toutes les langues d’Union Européenne. L’utilisateur ne doit mettre le dispositif en marche que s’il comprend le mode d’emploi présent ou si une traduction complète et correcte du mode d’emploi est sous ses yeux. Le dispositif ne doit pas être mis en marche avant que le mode d’emploi ait été lu et compris complètement et seulement si le mode d’emploi est observé et tous les mesures demandées sont prises. «Avis de sécurité pour des dispositifs à vide»
BG
Внимание: Тези инструкции не са преведени на всички езици от ЕО. Потреби­телят не бива да работи с уреда, ако не разбира инструкциите за ползване. В този случай е необходимо да бъде предоставен пълен технически превод на инструкциите за ползване. Преди работа с уреда е задължително потребите­лят да прочете изцяло инструкциите за работа.
Указания за безопасност за вакуумни уреди
CN
注意:该操作手册不提供所有的语言版本。操作者在没有理解手册之前,不能操作 该设备。在这种情况下,需要有一个整个操作手册技术上正确的翻译。在操作该设 备前,必须完全阅读并理解该操作手册,必须实施所有需要的测量。 真空设备的安全信息
CZ
Upozornění :Tento návod k použití není k dispozici ve všech jazycích Evropské unie. Uživatel není oprávněn požít přístroj pokud nerozumí tomuto návodu. V tako­vém případě je nutno zajistit technicky korektní překlad manuálu do češtiny. Návod musí být uživatelem prostudován a uživatel mu musí plně porozumět před tím než začne přístroj používat. Uživatel musí dodržet všechna příslušná a požadovaná opatření. ”Bezpečnostní upozornění pro vakuové přístroje”.
page 4 of 126
DA Bemærk: Denne manual foreligger ikke på alle EU sprog. Brugeren må ikke be-
tjene apparatet hvis manualen ikke er forstået. I det tilfælde skal en teknisk korrekt
oversættelse af hele manual stilles til rådighed. Manual skal være gennemlæst og
forstået før apparatet betjenes og alle nødvendige forholdsregler skal tages.
»Sikkerhedsregler for vakuumudstyr«
EE
Tähelepanu! Käesolev kasutusjuhend ei ole kõigis EL keeltes saadaval. Kasutaja ei tohi seadet käsitseda, kui ta ei saa kasutusjuhendist aru. Sel juhul peab saadaval olema kogu kasutusjuhendi tehniliselt korrektne tõlge. Enne seadme kasutamist tu­leb kogu juhend läbi lugeda, see peab olema arusaadav ning kõik nõutud meetmed peavad olema rakendatud. ”Ohutusnõuded vaakumseadmetele”
ES
Atención: Este manual no está disponible en todos los idiomas de UE. El usuario no debe manejar el instrumento si no entiende este manual. En este caso se debe
disponer de una traducción técnicamente correcta del manual completo. El manual
debe ser leído y entendido completamente y deben aplicarse todas las medidas de seguridad antes de manejar el instrumento. ”Notas sobre la seguridad para equipos de vacío”
FI
Huomio: Tämä käyttöohje ei ole saatavilla kaikilla EU: n kielillä. Käyttäjä ei saa käyt­tää laitetta, jos hän ei ymmärrä tätä ohjekirjaa. Tässä tapauksessa on saatavilla ol­tava teknisesti oikein tehty ja täydellinen ohjekirjan käännös. Ennen laitteen käyttöä on ohjekirja luettava ja ymmärrettävä kokonaan sekä suoritettava kaikki tarvittavat valmistelut ja muut toimenpiteet. ”Vakuumilaitteen turvallisuustiedot”
GR
Προσοχή! : Οι οδηγίες αυτές δεν είναι διαθέσιµες σε όλες τις γλώσσες της Ευρω­παϊκής Ένωσης. Ο χρήστης δεν πρέπει να θέσει σε λειτουργία την συσκευή αν δεν κατανοήσει πλήρως τις οδηγίες αυτές. Σε τέτοια περίπτωση ο χρήστης πρέπει να προµηθευτεί ακριßή µετάφραση του ßιßλίου οδηγιών. Ο χρήστης πρέπει να διαßά­σει και να κατανοήσει πλήρως τις οδηγίες χρήσης και να λάßει όλα τα απαραίτητα µέτρα πριν θέσει σε λειτουργία την συσκευή. ”Υποδείξεις ασφάλειας για αντλίες κενού”
HR
Pažnja:ove upute ne postoje na svim jezicima Europske Unije. Korisnik nemora ra­diti sa aparatom ako ne razumije ove upute.U tom slucaju tehnicki ispravni prijevod cijelih uputstava mora biti na raspolaganju. Uputstva moraju biti cijela procitana i razumljiva prije rada sa aparatom i sve zahtijevane mjere moraju biti primjenjene. ”Sigurnosne napomene za vakuumske uređaje
page 5 of 126
HU
Figyelem! Ez a kezelési utasítás nem áll rendelkezésre az EU összes nyelvén. Ha a felhasználó nem érti jelen használati utasítás szövegét, nem üzemeltetheti a készüléket. Ez esetben a teljes gépkönyv fordításáról gondoskodni kell. Üzembe helyezés előtt a kezelőnek végig kell olvasnia, meg kell értenie azt, továbbá az üzemeltetéshez szükséges összes mérést el kell végeznie. ”A vákuum�készü� ”A vákuum�készü�”A vákuum�készü­lékekkel kapcsolatos biztonsági tudnivalók
IT Attenzione: Questo manuale non è disponibile in tutte le lingue della Comunità Eu­ropea (CE). L‘utilizzatore non deve operare con lo strumento se non comprende questo manuale. In questo caso deve essere resa disponibile una traduzione tec­nicamente corretta del manuale completo. Il manuale deve essere completamente letto e compreso prima di operare con lo strumento e devono essere applicati tutti gli accorgimenti richiesti. ”Istruzioni di sicurezza per apparecchi a vuoto”
JP
注意:この取扱説明書はすべての言語で利用可能ではありません。 もしこの取扱 説明書を理解できないならば、ユーザーは装置を操作してはなりません。 この場 合、技術的に正しい翻訳がなされた完全なマニュアルを用意しなければなりませ ん。 装置を作動する前にマニュアルを完全に読み、そして理解されなくてはなり ません。そして、すべての要求される対策を講じなければなりません。 真空装置を安全に取り扱うために
KR
주의 : 이 매뉴얼은 모든 언어로 번역되지는 않습니다. 만약 이 매뉴얼의 내용을 충분 히 인지하지 못했다면 기기를 작동하지 마십시오. 매뉴얼의 내용을 기술적으로 정확 하게 번역한 경우에 이용하십시오. 기기를 사용하기 전에 이 매뉴얼을 충분히 읽고 이해하고 모든 요구되는 사항들을 적용해야 합니다. 진공 장비에 대한 안전 정보
LT
Dėmesio: šis vadovas nėra pateikiamas visomis ES kalbomis. Naudotojui drau­džiama eksploatuoti įtaisą, jeigu jis nesupranta šio vadovo. Tokiu atveju reikia turėti viso vadovo techniškai taisyklingą vertimą. Vadovą būtina visą perskaityti ir suprasti pateikiamas instrukcijas prieš pradedant eksploatuoti įtaisą, bei imtis visų reikiamų priemonių. ”Vakuuminės įrangos saugos informacija”
LV
Uzmanību: Lietotāja instrukcija nav pieejama visās ES valodās. Lietotājs nedrīkst lietot iekārtu, ja viņš nesaprot lietotāja instrukcijā rakstīto. Šādā gadījumā, ir ne­pieciešams nodrošināt tehniski pareizu visas lietotāja instrukcijas tulkojumu. Pirms sākt lietot iekārtu, un, lai izpildītu visas nepieciešamās prasības, iekārtas lietotāja instrukcija ir pilnībā jāizlasa un jāsaprot. ”Vakuuma iekārtu drošības noteikumi”
page 6 of 126
NL
Attentie: Deze gebruiksaanwijzing is niet in alle talen van de EU verkrijgbaar. De
gebruiker moet niet met dit apparaat gaan werken als voor hem/haar de gebruiks-
aanwijzing niet voldoende duidelijk is. Bij gebruik van deze apparatuur is het nood­zakelijk een technisch correcte vertaling van de complete gebruiksaanwijzing te hebben. Voor het in gebruik nemen van het apparaat moet de gebruiksaanwijzing volledig gelezen en duidelijk zijn en dienen alle benodigde maatregelen te zijn ge-
nomen. ”Veiligheidsvoorschriften voor vacuümapparaten”
PL
Uwaga!! Ta instrukcja nie jest dostępna we wszystkich językach Unii Europejskiej. Użytkownik nie może rozpocząć pracy z urządzeniem dopóki nie przeczytał instruk­cji i nie jest pewien wszystkich informacji w niej zawartych. Instrukcja musi byc w całości przeczytana i zrozumiana przed podjęciem pracy z urządzeniem oraz nale­ży podjąć wszystkie niezbędne kroki związane z prawidłowym uzytkowaniem. ”Wskazówki bezpieczeństwa do urządzeń próżniowych”
PT
Atenção: Este manual não está disponível em todas as línguas da UE. O usuário
não deve utilizar o dispositivo, se não entender este manual. Neste caso, uma tra­dução tecnicamente correta do manual completo tem de estar disponível. O manu­al deve ser lido e entendido completamente antes da utilização do equipamento e
todas as medidas necessárias devem ser aplicadas. ”Informação de Segurança para Equipamento que funciona a Vácuo”
RO Atentie: Acest manual nu este disponibil in toate limbile EU. Utilizatorul nu trebuie
sa lucreze cu aparatul daca daca nu intelege manualul. Astfel, va 󰘰 disponibile o
traducere corecta si completa a manualului. Manualul trebuie citit si inteles in intre­gime inainte de a lucra cu aparatul si a luat toate masurile care se impun.
”Instrucţiuni de siguranţă pentru aparatele de vidare”
RU
Внимание: Эта инструкция по эксплуатации не имеется на всех языках. Потре­бителю не дозволенно эксплуатировать данный прибор, если он не понимает эту инструкцию. В этом случае нужен технически правильный перевод полной инструкции. Прежде чем использовать этот прибор, необходимо полностью прочитать и понять эту инструкцию и принять все не­обходимые меры. ”Указания по технике безопасности при работе с ваку­умными устройствами”
page 7 of 126
SE Varning: Denna instruktion är inte tillgänglig på alla språk inom EU. Användaren får inte starta utrustningen om hon/han inte förstår denna instruktion. Om så är fallet måste en tekniskt korrekt instruktion göras tillgänglig. Instruktionen måste läsas och förstås helt före utrustningen tas i drift och nödvändiga åtgärder göres. ”Säkerhetsinformation för vakuumutrustning”
SI
Pozor: Ta navodila niso na voljo v vseh jezikih EU. Uporabnik ne sme upravljati z napravo, če ne razume teh navodil. V primeru nerazumljivosti mora biti na voljo tehnično pravilen prevod. Navodila se morajo prebrati in razumeti pred uporaba naprave, opravljene pa moraja biti tudi vse potrebne meritve.
”Varnostni nasveti za vakuumske naprave”
SK
Upozornenie: Tento manuál nie je k dispozícii vo všetkých jazykoch EÚ. Užívateľ nesmie obsluhovať zariadenie, pokiaľ nerozumie tomuto manuálu. V takomto prípa­de musí byť k dispozícii technicky správny preklad celého manuálu. Pred obsluhou zariadenia je potrebné si prečítať celý manuál a porozumieť mu, a musia byť prijaté všetky opatrenia. ”Bezpečnostné pokyny pre vákuové zariadenia”
TR
Dikkat : Bu kullanım kitabı, tüm dillerde mevcut değildir. Kullanıcı, bu kullanım kita­bını anlayamadıysa cihazı çalıştırmamalıdır. Bu durumda, komple kullanım kitabı­nın, teknik olarak düzgün çevirisinin bulunması gerekir. Cihazın çalıştırılmasından önce kullanım kitabının komple okunması ve anlaşılması ve tüm gerekli ölçümlerin uygulanması gerekir. ”Vakumlu cihazlar için güvenlik uyarıları”
page 8 of 126

Reset / Language selection

1 switch off
3 turn
2 press both
CVC 3000 V2.0
Deutsch
English Français Italiano Español Türkçe
한국어
中文
Portuguê
Ρyccкий
Polski Nederl.
日本語
Suomi
4 press
CVC 3000 V2.0
Deutsch English Français Italiano
Español
Türkçe
한국어
中文
Portuguê
Ρyccкий
Polski Nederl.
日本語
Suomi
page 9 of 126
Contents
Part I................................................................. 1
Reset / Language selection ............................................. 8
Safety information! ........................................................... 11
Important information! ........................................................................11
General information ........................................................................... 13
Intended use...................................................................................... 13
Setting up and installing the equipment ............................................ 14
Ambient conditions ............................................................................ 17
Operating conditions ......................................................................... 17
Safety during operation ..................................................................... 19
Maintenance and repair..................................................................... 22
`
Important information: Equipment marking (ATEX) ................... 24
Technical data ................................................................... 26
General technical data valid for all pumps ........................................ 26
Gas inlet temperatures ..................................................................... 27
Wetted parts ..................................................................................... 29
Abbreviations .................................................................................... 29
Pump parts ....................................................................................... 30
Use and operation ............................................................ 32
Installation ........................................................................................ 32
Vacuum connection (inlet) ................................................................ 33
Connecting the outlet ....................................................................... 34
Connecting the pressure transducer VSK 3000 ............................... 35
Electrical connection ........................................................................ 37
During operation ............................................................................... 38
Shutdown & storage ......................................................................... 41
Vacuum controller CVC 3000 ........................................................... 42
Menu guide ........................................................................ 47
Pump down function ........................................................ 49
Vac Control function ........................................................ 51
Auto mode ......................................................................... 54
Program function .............................................................. 56
Application example ......................................................................... 58
VACUULAN function ........................................................ 59
Application examples....................................................... 61
Vacuum for ltration and suction ...................................................... 61
page 10 of 126
Vacuum for gel dryer,
drying chambers and vacuum concentrators ................................... 62
Vacuum for distillation and evaporation (e.g., rotary evaporator) 62
Fore-vacuum for high vacuum pumps .............................................. 64
Conguration .................................................................... 65
Readjustment of VSK 3000 .............................................. 67
Calibration in the factory ................................................. 69
Cleaning the pressure transducer .................................. 70
Part II.............................................................. 71
Interface parameters ........................................................ 74
Setting of the interface ..................................................................... 75
Read commands ”CVC 2000” .......................................................... 76
Write commands ”CVC 2000” .......................................................... 77
Read commands ”CVC 3000” .......................................................... 79
Write commands ”CVC 3000” .......................................................... 81
Accessories / Spare parts................................................ 84
Troubleshooting ............................................................... 86
Replacing the device fuse ................................................................ 90
Replacing diaphragms and valves.................................. 91
Disassembling the pump housing .................................................... 93
Checking diaphragms and valves .................................................... 99
Assembling the pump head ............................................................ 103
Replacing the valve at the distributor ............................................ 109
Assembling the pump housing ........................................................113
Notes on return to the factory ....................................... 120
Warranty .......................................................................... 123
Health and safety clearance form ................................. 124
EC Declaration of Conformity of the Machinery ............. 125
page 11 of 126

Safety information!

Important information!

+ Keep this manual complete and accessible to per-
sonnel at all times!
+ Read this manual carefully before installing or op-
erating the equipment. Observe the instructions contained in this manual.
+ Do not modify the equipment without authoriza-
tion.
NOTICE
This manual is an integral part of the equipment de­scribed therein. It describes the safe and proper use of the vacuum pump. The manual consists of two parts. Make operating personnel aware of dangers arising from the pump and the pumped substances. VACUUBRAND disclaims any liability for inappropri­ate use of these pumps and for damage from failure to follow instructions contained in this manual.
This manual is only to be used and distributed in its com­plete and original form. It is strictly the users’ responsibility to check carefully the validity of this manual with respect to his product. Manual-no.: 999264 / 03/07/2014
The following signal word panels and safety symbols are used throughout this manual:
This is the safety alert symbol. It is used to alert you to po­tential personal injury hazards. Obey all safety messages that follow this symbol to avoid possible injury and death.
page 12 of 126
DANGER indicates a hazardous situation which, if not
avoided, will result in death or serious injury.
+ WARNING indicates a hazardous situation which, if
not avoided, could result in death or serious injury.
• CAUTION indicates a hazardous situation which, if not
avoided, could result in minor or moderate injury.
NOTICE
NOTICE is used to address practices not related to
personal injury.
Caution! Hot surface!
Disconnect equipment from AC power.
Dispose off electronic equipment according to regula-
tions.
Formatting used in this manual: Note: The signal word panels in all sections of this manual always refer
to all paragraphs of the same format ( / + / • / plain text) following each signal word panel.
page 13 of 126

General information

NOTICE
Remove all packing material from the packing box. Re­move the product from its packing-box and retain all pack­aging until the equipment is inspected and tested. Re­move the protective caps from the inlet and outlet ports and retain for future use. Inspect the equipment promptly and carefully. If the equipment is damaged, notify the supplier and the carrier in writing within three days. Retain all packing ma­terial for inspection. State the item number of the product together with the order number and the supplier’s invoice number. Failure to check and give notice of damage will
void any and all warranty claims for those deciencies.
Replace the protective caps, if the equipment is not used immediately. Store the equipment in dry and non-corrosive conditions (see also “Technical data”, pg. 26).
+ Do not use any damaged equipment.
• Use the handles when moving the pump.

Intended use

+ Do not use the pump or any system parts on humans
or animals.
+ Ensure that the individual components are only con-
nected, combined and operated according to their de­sign and as indicated in the instructions for use. Use only original manufacturer’s spare parts and ac- cessories. Otherwise the safety and performance of the equipment, as well as the electromagnetic compat­ibility of the equipment might be reduced.
The CE mark may be voided if not using original man-
ufacturer’s spare parts.
+ Comply with all notes on correct vacuum and electri-
cal connections; see section “Use and operation”, pg.
32.
page 14 of 126
+ Do not use the pump to generate pressure.
+ The pumps are designed for ambient temperatures
during operation between +50°F and +104°F (+10°C and +40°C). Periodically check maximum tempera­tures if installing the pump in a cabinet or a housing. Make sure ventilation is adequate to maintain recom­mended operating temperature. Install an external au­tomatic ventilation system if necessary. If pumping hot process gases, make sure that the maximum permitted gas inlet temperature is not exceeded. The maximum permitted gas inlet temperature depends on several parameters like inlet pressure and ambient tempera­ture (see “Technical data”, pg. 26).
NOTICE
+ Particles and dust must not enter the pump. + Do not pump liquids.
• Ensure that the pump is chemically resistant to the
pumped substances prior to operation.
Use the equipment only as intended, that is, for genera­tion, measurement, and control of vacuum in vessels de­signed for that purpose. Any other use will automatically invalidate all warranty and liability claims. Remain aware of safety and risks.

Setting up and installing the equipment

Equipment must be connected only to a suitable elec-
trical supply and a suitable ground point. As such,
the plug must be plugged into an outlet that is properly grounded. Failure to connect the motor to ground may result in deadly electrical shock.
The supply cable may be tted with a molded Europe-
an IEC plug or a plug suitable for your local electrical supply. The cable contains wires color coded as fol­lows: green or green and yellow: ground; blue or white: neutral; brown or black: hot.
page 15 of 126
+ Due to the high compression ratio, the pump may gen-
erate overpressure at the outlet. Check pressure com­patibility with system components (e.g., exhaust pipe­line or exhaust valve) at the outlet.
+ Do not permit any uncontrolled pressurizing. Make
sure that the exhaust pipeline cannot become blocked. If there is an exhaust isolation valve, make sure that you cannot operate the equipment with the valve closed to avoid a risk of bursting!
+ Maximum permissible pressure at the pressure trans-
ducer: 21.8 psi (1.5 bar) absolute.
+ Operation with silencer at the outlet: Use the silencer only in case of low gas throughput.
Operating the pump at a high inlet pressure or pump­ing dusty gases for a long time may cause clogging of the silencer. Due to clogging of the silencer an internal overpressure may build up, which may damage bear­ings, diaphragms and valves.
Check the silencer regularly for permeability and re-
place if necessary, or install a hose nozzle instead (see “Accessories / Spare parts”, pg. 84). Preferably use the hose nozzle as outlet and make sure that the ex­haust pipeline cannot become blocked.
+ Keep the electrical power cord away from heated sur-
faces.
• Provide a rm, level platform for the equipment. Check
that the system which you are going to evacuate is
mechanically stable. Check that all ttings are secure.
Ensure a stable position of the pump without any me­chanical contact other than the pump feet.
• Comply with maximum permissible pressures at in- let and outlet and with maximum permissible pres- sure differences between inlet and outlet. See section “Technical data”, pg. 26. Do not operate the pump with overpressure at the inlet.
page 16 of 126
• Comply with maximum permissible pressures at the pressure transducer. See section “Technical data”, pg.
26.
• Avoid overpressure of more than 17.5 psi absolute (1.2
bar absolute) in the event that inert gas is connected to the pump, or to a venting valve.
Note: Flexible elements will shrink when evacuated.
• Connect hoses gas tight at inlet and outlet of the pump.
• Connect hoses gas tight at the pressure transducer.
• Ensure that no foreign objects can be drawn into the
NOTICE
pump.
• Check the power source and the pump’s rating plate
to be sure that the power source and the equipment match in voltage, phase, and frequency.
Make sure ventilation is adequate to maintain recommend­ed operating temperature. Keep a minimum distance of 2 in (5 cm) between the cooling fan and surrounding items (e.g., housing, walls, etc.), or else install an external au­tomatic ventilation system. Do not place the pump on soft surfaces (e.g., rubber foam) during operation. This may reduce the air circulation in the pump housing and affect the fans’ cooling effect. Check fans regularly for dust/dirt. Clean fan guard grills if necessary to avoid a reduction of ventilation.
Position the pressure transducer and its vacuum line in such a way that condensate cannot ow towards the pres­sure transducer.
Use only hoses at the inlet and outlet of the pump with an inner diameter at least as large as the diameter of the pump’s tubing (to avoid overpressure at the outlet, and reduction of pumping speed at the inlet).
page 17 of 126
Allow the equipment to equilibrate to ambient temperature if you bring it from cold environment into a room prior to operation. Notice if there is water condensation on cold surfaces.
Comply with all applicable and relevant safety require-
ments (regulations and guidelines). Implement the re­quired actions and adopt suitable safety measures.

Ambient conditions

Do not reach for this product if it has fallen into liquid.
NOTICE
There is a risk of deadly electrical shock. Unplug the system immediately.
+ Do not use this product in an area where it can fall or
be pulled into water or other liquids.
• Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment outdoors, installation in higher altitudes, conductive pollution or external condensation on the pump.
• Do not operate this product near ames.
To the best of our knowledge the equipment is in com­pliance with the requirements of the applicable EC-direc­tives and harmonized standards (see ”Declaration of Con­formity”) with regard to design, type and model. Directive EN 61010-1 gives in detail the conditions under which the equipment can be operated safely (see also IP degree of protection, “Technical data”, pg. 26).

Operating conditions

These pumps are not approved for operation in po-
tentially explosive atmospheres. Do not operate the pumps in potentially explosive atmospheres.
page 18 of 126
Pumps without the
not approved for the pumping of potentially explo­sive atmospheres. Do not pump potentially explo­sive atmospheres with those pumps
Pumps bearing the
are approved for the pumping of potentially explo-
sive atmospheres according to their classication II 3G IIC T3 X according to ATEX, but they are not approved for operation in potentially explosive at­mospheres (see section “ Equipment marking (ATEX)”, pg. 24).
The pumps are not suitable to pump any of the sub-
stances listed below.
Do not pump:
- unstable substances
mark on the rating plate are
`
mark on their rating plates
`
Important information:
`
- substances which react explosively under impact
(mechanical stress) without air
- substances which react explosively when being ex-
posed to elevated temperatures without air,
- substances subject to auto-ignition,
- substances which are inammable without air
- explosive substances.
The pumps are not approved for operation below
ground. Do not operate the pump below ground.
Do not pump substances which may form depos- its inside the pump. The pumps are not suitable for
pumping substances which may form deposits inside the pump. Deposits and condensate in the pump may lead to increased temperatures even to the point of exceeding the maximum permitted temperatures.
• Check the inlet and outlet of the pump, if there is a
danger of forming deposits inside the pump, e.g., in the pump chambers (the pump chamber is the part between diaphragm and head cover. See section “Re­placing diaphragms and valves”, pg. 91). Inspect the pump chambers regularly and clean if necessary.
page 19 of 126
Consider interactions and chemical reactions of the pumped media. Ensure that the materials of the
pump’s wetted parts are compatible with the pumped substances, see section “Technical data”, pg. 26.
When changing the substances pumped, we recom-
mend purging the pump with air or inert gas prior to changing the pumped media. Purging the pump will pump out residues and it will reduce the possibility of reactions of the pumped substances with each other and with the pump’s materials.

Safety during operation

Adopt suitable measures to prevent the release of dan-
gerous, toxic, explosive, corrosive, noxious or pollut-
ing uids, vapors and gases. To prevent any emission
of such substances from the pump outlet, install an appropriate collecting and disposal system and take protective action for pump and environment.
You must take suitable precautions to prevent any for-
mation of explosive mixtures in the pump chamber or at the outlet of the pump. In case, e.g., of a diaphragm failure, mechanically generated sparks, hot surfaces or static electricity may ignite these mixtures. Use inert gas for venting, if necessary.
Drain appropriately or otherwise remove any poten-
tially explosive mixtures at the outlet of the pump, or dilute them to non-explosive concentrations.
Never operate this pump if it has a damaged cord or
plug.
+ If the pump is not working properly, has been dropped
or has fallen into water, contact your pump service pro­vider.
+ Prevent any part of the human body from coming into
contact with vacuum.
page 20 of 126
+ Always provide a free and pressureless exhaust outlet
to avoid damage to pump valves and risk of bursting.
+ Attention: Due to clogging of the silencer an internal
overpressure may build up, which may damage bear­ings, diaphragms and valves. Preferably use the hose nozzle as outlet and make sure that the exhaust pipe­line cannot become blocked.
+ Attention: At pressures above approximately 795 Torr
(1060 mbar) the pressure reading becomes incorrect due to saturation of the pressure transducer. The dis-
play ashes. Release pressure immediately! Risk of
bursting!
+ Comply with applicable regulations when disposing of
chemicals. Take into consideration that chemicals may be contaminated. Take adequate precautions to pro­tect people from the effects of dangerous substances (chemicals, thermal decomposition products of uoro­elastomers). Use appropriate protective clothing and safety goggles.
+ Interruption of the pump (e.g., due to power failure),
failure of connected components or of parts of the sup­ply, or change in parameters must not be allowed to lead to dangerous conditions. In case of a diaphragm failure or in case of a leak in the manifold, pumped substances might be released into the environment or into the pump housing or motor.
Comply with all notes regarding proper use of the
pumps, as well as operation and maintenance guid­ance.
+ The residual leak rate of the equipment might render
possible an exchange of gas, albeit extremely slight, between the environment and the vacuum system.
Adopt suitable measures to prevent contamination of
the pumped substances or the environment.
page 21 of 126
• Ensure that no parts of your clothing, hair or ngers
can be caught or drawn in at the inlet of the pump.
Never insert ngers or drop any other object into the
inlet or outlet.
• You must take suitable precautions to prevent any
dangerous situation from arising if the controller starts a NT VARIO pump, switches a coolant valve or opens a venting valve.
Attention: If the controller is set to Autostart, the pro- cess will start immediately after a power failure with­out pressing any further key. It is your responsibility to ensure that automatic start-up of the system will not lead to any dangerous condition. Provide appropriate
NOTICE
safety measures. Check prior to starting the process whether the option ”Autostart” (menu: conguration) is enabled.
Pay attention to the safety symbol ”hot surfaces” on the equipment. Hot parts may cause burns if touched. Adopt suitable measures to prevent any danger aris­ing from hot surfaces or electric sparks. Ensure that hot surfaces of the pump do not cause burns. Provide a suitable contact guard if necessary.
Prevent the backpressure of gases and the backow of
condensates at the outlet.
Provide appropriate protective measures to allow for the possibility of failure and malfunction. The protective mea­sures must also allow for the requirements of the respec­tive application.
In case of overload, the motor is shut down by a self-hold
thermal cutout in the winding. Note: Only manual reset is possible. Press the START/
STOP key at the controller to conrm the error message.
Switch off the pump and disconnect from the power source. Identify and eliminate the cause of failure. Allow the pump
to cool down sufciently before restart.
page 22 of 126

Maintenance and repair

NOTICE
In order to comply with laws (occupational, health and safety regulations, safety at work law and regulations for environmental protection) vacuum pumps, components and measuring instruments can only be returned when certain procedures (see section “Notes on return to the factory”, pg. 120) are followed.
Take advantage of our service seminars, which put special focus on the maintenance and repair of vacuum pumps. For details and for the online ”Instructions for repair” man­ual see www.vacuubrand.com.
In normal use, the lifetime of the diaphragms and valves is typically 15,000 operating hours. Bearings have a typi­cal durability of 40000 h.
Ensure that the pump cannot be operated acciden-
tally. Never operate the pump if covers or other parts of the pump are disassembled.
Switch off the pump. Disconnect the electrical
power cord and wait two minutes before starting
maintenance to allow the capacitors to discharge.
Note: The pump may be contaminated with process
chemicals, which have been pumped during operation. Ensure that the pump is completely decontaminated before maintenance commences.
+ Take adequate precautions to protect people from the
effects of dangerous substances if contamination has occurred. Use appropriate protective clothing, safety goggles and protective gloves.
+ Wear parts have to be replaced regularly. + Never operate a defective or damaged pump.
page 23 of 126
+ Vent the pump before starting maintenance. Isolate the
pump and other components from the vacuum system.
Allow sufcient cooling of the pump. Drain condensate,
if applicable.
NOTICE
Ensure that maintenance is done only by suitably trained and supervised technicians. Ensure that the maintenance technician is familiar with the safety procedures which re­late to the products processed by the pumping system. Only dismantle the pump as far as necessary.
page 24 of 126
Important information: Equipment marking (ATEX)
`
VACUUBRAND equipment bearing mark (see rating plate)
II 3G IIC T3 X
`
Internal Atm. only
Tech. File Ref.: VAC-EX01
and
VACUUBRAND equipment bearing mark (see rating plate)
X see manual
`
For equipment labelled with „ tion according to Directive 94/9/EC (ATEX) is valid: Internal Atm. only, Tech. File Ref.: VAC-EX01.
The classication II 3G IIC T3 X according to ATEX is only valid for the in­ner part (wetted part, pumped gas or vapor) of the equipment. The equip­ment is not suitable for use in external, potentially explosive atmospheres (environment).
The overall category of the equipment depends on the connected com-
ponents. If the connected components do not comply with the classi­cation of the VACUUBRAND equipment, the specied category of the
VACUUBRAND equipment is no longer valid.
X see manual“ the following classica-
`
II 3G IIC T3 X,
`
Vacuum pumps and vacuum gauges in category 3 are intended for con­nection to equipment in which during normal operation explosive atmo­spheres caused by gases, vapors or mists normally don’t occur; or, if they do occur, are likely to do so only infrequently and for a short period only. Equipment in this category ensures the requisite level of protection dur­ing normal operation. The use of gas ballast or the operation of venting valves is only permit­ted if thereby explosive atmospheres normally don’t occur in the interior of the equipment or, if they do occur, are likely to do so only infrequently and for a short period.
page 25 of 126
The pumps are marked with ”X” (according to EN 13463-1), i.e., restric­tions of the operation conditions:
• The equipment is designated for a low degree of mechanical stress
and has to be installed in a way so that it cannot be damaged from outside.
Pumping units have to be installed so that they are protected against
shocks from the outside and against glass splinters in the event of breakage (implosion).
• The equipment is designated for an ambient and gas inlet temperature
during operation of +10 to +40°C. Never exceed these ambient and gas inlet temperatures. If pumping / measuring gases which are not potentially explosive, extended gas inlet temperatures are permissible. See instructions for use, section “Gas inlet temperatures” or “Techni­cal data”.
After any intervention at the equipment (e.g., repair / maintenance) the ul­timate vacuum of the pump has to be checked. Only if the pump achieves its specied ultimate vacuum is the pump’s leak rate low enough to en­sure that no explosive atmospheres will occur in the interior of the equip­ment. After any intervention at the vacuum sensor, the leak rate of the equip­ment has to be checked.
Attention: This manual is not available in all languages of the EU. The user must not op­erate the device if he does not understand this manual. In this case a technically cor­rect translation of the complete manual has to be available. The manual must be com­pletely read and understood before opera-
tion of the device. All specied measures
must be applied, or else must be replaced by equivalent measures at the user’s own risk.
page 26 of 126

Technical data

General technical data valid for all pumps

Maximum permissible inlet pressure (absolute)
Maximum permissible outlet pressure (absolute)
Maximum pressure difference between inlet and outlet
Permissible ambient temperature storage / operation
Permissible relative atmospheric mois­ture during operation (no condensation)
Maximum permissible installation alti­tude above mean sea level
Rated motor power hp (kW) 0.71 (0.530)
No-load speed rpm 30 - 2400
Maximum permissible range of supply voltage (±10%)
Attention: Observe specications of rating plate!
psi
(bar)
psi
(bar)
psi
(bar)
°F
(°C)
% 30 to 85
ft
(m)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
(2000)
100-120 V~ 50/60 Hz,
200-230 V~ 50/60 Hz
16
(1.1)
16
(1.1)
16
(1.1)
6500
Maximum rated current at: 100-120 V~ 50/60 Hz 200-230 V~ 50/60 Hz
Device fuse 2 slow blow fuses 8A
Motor protection thermal cutout, manual reset
Degree of protection IEC 529 IP 40
A-weighted emission sound pressure level* (uncertainty KpA : 3 dB(A))
Inlet small ange KF DN 25
Outlet silencer / thread G 1/2”
Dimensions L x W x H
approx.
Weight approx. lbs. (kg) 68.8 (31.2)
* Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 1500rpm and ulti-
mate vacuum with standard silencer at outlet.
A A
dB(A) 50
in
(mm)
21.8 x 10.2 x 16.5 (554 x 260 x 420)
8.0
3.5
page 27 of 126
Type
ME 16 NT
VARIO
Maximum pumping speed (ISO 21360)
Ultimate vacuum (absolute) **
** Ultimate vacuum at setting ”Pump down” at speed ”HI”
cfm
(m3/h)
Torr
(mbar)
approx. 11.2 (approx. 19)
75
(100)

Gas inlet temperatures

Operating condition Inlet pressure
Continuous operation
> 75 Torr (100 mbar) (high gas load)
MD 12 NT
VARIO
7.9
(13.4)
1.1
(1.5)
Permitted range of gas temperatures at inlet
50 °F to 104 °F (+10°C to +40°C)
MV 10 NT
VARIO
7.1
(12.1)
0.2
(0.3)
Continuous operation
Short-time (< 5 minutes)
* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
< 75 Torr (100 mbar) (low gas load)
< 75 Torr (100 mbar) (low gas load)
32 °F to 140 °F* (0°C to +60°C*)
14 °F to 176 °F* (-10°C to +80°C*)
Controller CVC 3000
external gauge head VSK 3000,
Pressure transducer
Display LCD graphic display, illuminated
Pressure units / scale (selectable) Torr, mbar or hPa
ceramic diaphragm (alumina), capacitive, absolute pressure,
gas type independent
Measuring range (absolute) 795 - 0.1 Torr (1060 - 0.1 mbar)
Maximum control range (absolute)* 795 - 0.1 Torr (1060 - 0.1 mbar)
Resolution (display) 0.1 Torr (mbar)
Maximum permissible pressure at pressure transducer (absolute)
* The actual vacuum control range in your application might be reduced due to ultimate vacuum
of the pump, volume of gas present, etc.
1125 Torr (1.5 bar)
page 28 of 126
Controller CVC 3000
Measurement uncertainty (absolute) after careful adjustment and at constant temperature
Temperature coefcient <± 0.05 Torr/K (0.07 mbar/K)
<± 0.75 Torr (1 mbar)
Maximum permissible temperature of gaseous media**
continuous operation: 104°F (40°C),
for short periods (less than 5 minutes)
up to 176°F (80°C)
Ambient temperature range (operation) 50°F to 104°F (10°C to +40°C)
Ambient temperature range (storage) 14°F to 158°F (-10°C to +70°C)
Permissible relative atmospheric mois­ture during operation (no condensation)
Maximum permitted current of connect­ed valves (connected components)
Degree of protection IEC 529 (front side)
Degree of protection IEC 529 (gauge head VSK 3000)
Measurement connection of pressure transducer
hose nozzle for tubing
I.D. 1/4” / 3/8” (6 / 10 mm)
30% to 85%
4A
IP 42
IP 54
small ange DN 16 or
Cable length of pressure transducer approx. 5’6” (2m)
Supply voltage VSK 3000
6-30 V= / 5 mA
(via VACUU • BUS)
Communication VSK 3000 VACUU • BUS
Interface RS-232 C
Dimensions CVC 3000 L x W x H
(with foot) Dimensions VSK 3000
(138 mm x 124 mm x 115 mm)
5.4” x 4.9” x 4.5”
2.4” (60 mm)
housing diameter length with measurement connection
Weight CVC 3000
VSK 3000 (with small ange)
** if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
2.4” (60 mm) with small ange
0.97 lb (440 g)
0.41 lb (185 g)
We reserve the right for technical modication without prior notice!
page 29 of 126

Wetted parts

Components Wetted materials
Pump
Housing cover Aluminum alloy
Head cover Aluminum alloy (AlSi12)
Diaphragm clamping disc Aluminum alloy (AlSi12)
Diaphragm FPM
Valves FPM
O-rings FPM
Small ange Stainless steel
Tubing PTFE
Fittings ETFE / ECTFE
Inlet / distributor Aluminum alloy
Outlet PP
Silencer
VSK 3000
Sensor Aluminum oxide ceramic
Seal at sensor Chemically resistant uoroelastomer
Sensor housing PPS glass ber reinforced
Small ange Stainless steel
Hose nozzle PP
Aluminum, anodized / PTFE / PTFE carbon reinforced / spring steel
We reserve the right for technical modication without prior notice!

Abbreviations

ECTFE: Ethylene/Chlorotriuoroethylene ETFE: Ethylene/Tetrauoroethylene FPM: Fluoroelastomer PP: Polypropylene
PPS: Polyphenylene sulde PTFE: Polytetrauoroethylene

Pump parts

Position Component
page 30 of 126
1 Inlet
2 Outlet (silencer or thread 1/2”)
3 ON/OFF switch
4 Mains connection
5 Fuse holder
6 Pump rating plate
7 Fan
8 Handle
9 CVC 3000 controller
10 Pressure transducer VSK 3000
11 VACUU•BUS cable to CVC 3000 controller
ME 16 NT VARIO, MD 12 NT VARIO, MV 10 NT VARIO
8
9
8
1
2
11
10
3
7
6
4
7
5
Loading...