➨ DANGER indicates a hazardous situation which, if not avoided, will re-
☞ WARNING indicates a hazardous situation which, if not avoided, could
• CAUTION indicates a hazardous situation which, if not avoided, could
page 2 of 22
sult in death or serious injury.
result in death or serious injury.
result in minor or moderate injury.
NOTICE
After sales service:
Contact your local dealer or call +49 9342 808-5500.
NOTICE is used to address practices not related to personal injury.
Danger for persons with cardiac pacemaker.
Warning! Magnetic eld.
Trademark index:
VACUU•LAN
TURBO•MODETM, VARIO® (US-Reg.No 3,833,788), VARIO-SPTM, VACUUBRAND® (US-Reg.No 3,733,388) and
also the shown company logos are trademarks of VACUUBRAND GMBH + CO KG in Germany and/or
other countries.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999226 // 05/05/2014
Cleaning the MPT 100 / MPT 200 .............................................................................16
Important notes concerning bake-out ..............................................................................................16
Calibration in the factory .........................................................................................17
Readjustment of the vacuum gauge head .............................................................18
Notes on return to the factory .................................................................................19
Health and safety clearance form ...........................................................................20
EC Declaration of Conformity..................................................................................21
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999226 // 05/05/2014
page 4 of 22
Safety information
General information
To operate the vacuum gauge head MPT 100 / MPT 200, a vacuum
gauge DCP 3000 is required!
If using a different controller, order the instructions for use of the
MPT 100 / MPT 200 from VACUUBRAND.
Read this manual and the manual of the vacuum gauge DCP 3000 carefully be-
NOTICE
fore installing or operating the equipment. Observe the instructions contained
in these manuals.
Remove all packing material from the packing box. Remove the product from its
packing box and retain all packaging until the equipment is inspected and tested.
Inspect the equipment promptly and carefully.
If the equipment is damaged, notify the supplier and the carrier in writing within
three days. Retain all packing material for inspection. State the item number of the
product together with the order number and the supplier’s invoice number. Failure
to check and give notice of damage will void any and all warranty claims for those
deciencies.
Store the equipment in dry and non-corrosive conditions if the equipment is not used
immediately.
☞Do not use any damaged equipment.
Intended use
☞ Ensure that the individual components are only connected, combined and oper-
ated according to their design and as indicated in the instructions for use. Use
only original manufacturer’s spare parts and accessories. Otherwise the safety
and performance of the equipment, as well as the electromagnetic compatibility
of the equipment might be reduced.
The CE mark may be voided if not using original manufacturer’s spare parts.
☞ Comply with all notes on correct vacuum and electrical connections, see section
”Use and operation”.
☞ If installing the device into a cabinet or a housing check maximum temperature.
Install an external automatic ventilation system if necessary. Ensure that the
maximum permitted gas temperature at the sensor head (see ”Technical data”) is
not exceeded.
MPT 100:
The device is designed for ambient and gas temperatures at the vacuum gauge
head connection of +10°C to +40°C at continuous operation.
MPT 200:
The device is designed for ambient and gas temperatures at the vacuum gauge
head connection of +5°C to +60°C at continuous operation.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999226 // 05/05/2014
NOTICE
NOTICE
page 5 of 22
Use the equipment only as intended, that is, for measurement of vacuum in vessels
designed for that purpose. Any other use will automatically invalidate all warranty
and liability claims.
Remain aware of safety and risks.
Connecting the device
☞ Do not permit any uncontrolled pressurizing (e.g., make sure that pipelines
cannot become blocked) to avoid a risk of bursting!
• Comply with maximum permissible pressures at the gauge head, see section
”Technical data”.
• Connect the vacuum gauge head gas tight to the vacuum application.
• Consider possible reactions of the process media due to the heat generated by
the product.
• When working with residues, aggressive or condensable media, install a gas
washing bottle if necessary.
Remove the protective cover and keep. Connect the vacuum gauge head via small
ange to the vacuum application.
The vacuum gauge head may be mounted in any orientation, though an orientation
is to be preferred in which condensate cannot ow towards the measuring chamber.
NOTICE
Switch off the DCP 3000 before connecting the MPT 100 / MPT 200 via VACUU•BUS
cable to the vacuum gauge.
Allow the equipment to equilibrate to ambient temperature if you bring it from cold
environment into a room prior to operation. Notice if there is water condensation on
cold surfaces.
Soiling and damage will affect the function of the vacuum gauge head.
Comply with all applicable and relevant safety requirements (regulations and
guidelines). Implement the required actions and adopt suitable safety meas-
ures.
Ambient conditions
• Adopt suitable measures in case of differences from recommended conditions,
e.g., using the equipment outdoors, or in an atmosphere of conductive pollution
or external condensation.
To the best of our knowledge the equipment is in compliance with the requirements
of the applicable EC-directives and harmonized standards (see ”Declaration of Conformity”) with regard to design, type and model. Directive EN 61010-1 gives in detail
the conditions under which the equipment can be operated safely (see also IP degree of protection).
Operating conditions
➨The devices are not suitable for applications which involve dangerous or ex-
plosive gases or explosive or ammable mixtures.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999226 // 05/05/2014
page 6 of 22
• Ensure that the materials of the vacuum gauge head’s wetted parts are compatible with the substances in the vacuum system, see section “Technical data”.
Safety during operation
➨ Adopt suitable measures to prevent the release of dangerous, toxic, explosive,
corrosive, noxious or polluting uids, vapours and gases.
☞ Prevent any part of the human body from coming in contact with the vacuum.
☞ Attention: Maximum pressure reading: 1*103 mbar. Pressure values above
1000 mbar can not be displayed. Danger of unnoticed overpressure! Risk of
bursting!
☞ Attention! Magnetic eld.
☞ Persons with active body aids (e.g. cardiac pacemakers) must keep a minimum
safety distance of approx. 10 cm (4in) to the magnetic unit of the MPT 100 /
MPT 200 (see also safety symbols on the vacuum gauge head).
NOTICE
NOTICE
Electronic equipment is never 100% fail-safe. This may lead to an ill-dened status of the equipment or of other connected devices. Provide appropriate protective
measures to allow for the possibility of failure and malfunction. The protective measures must also allow for the requirements of the respective application.
Maintenance and repair
The vacuum gauge head itself is maintenance-free.
A damp cloth may be used to clean the vacuum gauge head at the outside. Ensure
that the MPT 100 / MPT 200 is rst disconnected from the vacuum gauge DCP 3000.
➨Note: The equipment may be contaminated with harmful or dangerous substan-
ces, which have been processed during operation. Ensure that the equipment
is completely decontaminated before maintenance commences. Take adequate
precautions to protect people from the effects of dangerous substances if contamination has occurred. Use appropriate protective clothing, safety goggles and
protective gloves.
The vacuum gauge head is not prepared for customer repair.
Return the vacuum gauge head to the factory for repair.
Vacuum gauge head failures due to contamination are not covered by the warranty.
Returned products will not be repaired until the completed “Health and safety clearance form“ has been received.
In order to comply with laws (occupational, health and safety regulations, safety at
work law and regulations for environmental protection) components and measuring
instruments can only be returned when certain procedures (see section ”Notes on return to the factory”) are followed.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999226 // 05/05/2014
Technical data of vacuum gauge heads MPT 100 / MPT 200
TypeMPT 100
combined vacuum gauge head comprising two
Measuring principle
Measuring range (absolute)
Repeatability
(in percent of displayed value)
Measurement uncertainty
(in percent of displayed value
at constant temperature)
Maximum permissible pressure
at vacuum gauge head (absolute)
Leak rate1*10
components: thermal conductivity according to Pi-
rani and cold cathode tube; dependent on gas type
3
1*10
mbar - 5*10-9 mbar
(7.5*102 Torr - 3.7*10-9 Torr)
0
1*10
mbar - 1*10-7 mbar:
(7.5*10-1 Torr - 7.5*10-8 Torr):
2
1*10
mbar - 1*10-2 mbar:
(7.5*101 Torr - 7.5*10-3 Torr):
1*10-2 mbar - 1*10-7 mbar:
(7.5*10-3 Torr - 7.5*10-8 Torr):
1.5 bar (1125 Torr)
-9
mbar l/s
±5%
±15%
±25%
Measuring cycle40 ms
Maximum permissible temperature of gaseous
media at measurement connection
Maximum permissible temperature during
bake-out
for short periods (less than 5 minutes) up to 80°C
continuous operation: 40°C,
max. 180°C
(without electronics and magnetic unit)
Ambient temperature range (operation)+10°C to +40°C
Ambient temperature range (storage)-10°C to +60°C
Permissible relative atmospheric moisture
during operation (no condensation)
Maximum permissible installation altitude above
mean sea level
Range of supply voltage
(via VACUU•BUS cable)
30% to 85%
2000 m
24 VDC
Max. power consumption (supply)2.5 W
Degree of protection according to IEC 529IP 40
Weight580 g
Dimensions L x W x H144 mm x 85 mm x 53 mm
Vacuum connectionsmall ange KF DN 25
Internal volume of measurement chamber9.5 cm
3
Cable lengthapprox. 2.0 m
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully