VACUUBRAND Liquid level sensor User Manual

Page 1
page 1 of Seite 4
Liquid level sensor (699908 / 699909)
Safety information!
Use the level sensors 699908 and 699909 only in combination with VACUUBRAND vacuum con-
troller CVC 3000 or VNC 2 (connection only possible to VACUUBRAND VACUU•BUS system).
Remove all packing material, remove the product from its packing-box and inspect the equipment. Do
not use the equipment if it is damaged.
- Level sensor 699908 for catchpot 500 ml (cat. no.: 638497)
- Level sensor 699909 for condensate ask 1000 ml (cat. no.: 638877) Do not immerse the level sensor into liquid.
Comply with all relevant safety requirements (regulations and guidelines) and adopt suitable safety
measures.
Comply with all applicable national safety regulations. Pay attention to the maximal permitted ambient temperature. Use only genuine spare parts and accessories. Otherwise safety and performance of the equipment might be reduced. Ensure that maintenance is done only by suitable trained and supervised techni­cians.
environmental protection), equipment returned to the manufacturer can be repaired only under certain conditions.
Use and operation
The level sensor is intended to be assembled at the bottleneck of a VACUUBRAND catchpot 500 ml (cat. no.: 638497) or at a condensate ask 1000 ml (cat. no.: 638877) recommended at the catchpot of an exhaust waste vapour condenser.
Important: Assemble the sensor as close as possible to the catchpot.
Assembling the level sensor at the catchpot 500 ml
Assemble the level sensor to the bottleneck.Loosen the adjustment screw (1) at the sensor.Adjust the sensor as close as possible to the condensate ask.Screw adjustment screw.
liquid level to detect nonpolar liquids (e. g. toluene)
liquid level to detect polar liquids (e. g. water)
Disassembling the level sensor from the mount­ing plate
(e. g. to empty the condensate ask)
1
Remove clip with sensor from the bottleneck.
When reassembling ensure that the sensor is posi-
tioned as close as possible to the condensate ask.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check care­fully the validity of this document with respect to his product. Manual-no.: 999189 / 12/03/2013
Page 2
page 2 of Seite 4
Assembling the level sensor at the condensate ask 1000 ml
opening of the strap
closing of the strap
Attach both straps on the condensate ask. Position mounting plate under the straps. Assemble the liquid level sensor in a way that the sensor
becomes activated at the desired maximum liquid level.
Close straps.Loosen the adjustment screw at the sensor.Adjust the sensor as close as possible to the condensate ask.Screw adjustment screw.
liquid level to detect nonpolar liquids (e. g. toluene)
liquid level to detect polar liquids (e. g.
water)
level sensor
straps
Disassembling the level sensor from the mount­ing plate
(e. g. to empty the condensate ask)
Unscrew adjustment screw at the level sensor and
remove sensor. Reassemble the sensor in the original position. When reassembling ensure that the sensor is posi-
tioned as close as possible to the condensate ask.
adjustment screw
mounting plate
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check care­fully the validity of this document with respect to his product. Manual-no.: 999189 / 12/03/2013
Page 3
page 3 of Seite 4
Connecting to a controller
Plug in the connection cable of the level sensor on a switched-off VACUUBRAND vacuum controller CVC 3000 or VNC 2 (use Y adapter VACUU•BUS (VACUUBRAND cat. no: 636656) if necessary). When switching on the controller the level sensor is detected automatically. The display shows a symbol
for the sensor only if the sensor alarms or in case of an error. When coming near hands or objects the release of an unintentional message is possible.
The sensitivity of the sensor is adjusted at the factory in a way to detect also nonpolar liquids. Disconnecting the sensor plug from a switched-on controller leads to an error message.
Individual positioning of the level sensor at the condensate ask 1000 ml
- Fill solvent into the condensate ask until the desired liquid level is achieved.
- Connect level sensor to controller and switch on controller.
- Assemble the liquid level sensor loosely at the top of the condensate ask.
- Bring the sensor close to the liquid surface until the sensor becomes activated.
- Fix level sensor by closing the straps.
Level sensor alarm
The level sensor becomes activated if the liquid level in the catchpot reaches the position of the sen­sor.
An acoustic signal is given only if ”Sound On” is selected at the controller. Controller CVC 3000: Display of a ashing catchpot and a warning triangle and ashing display (10
blips).
Controller VNC 2: Display ”Ext. F.” and ashing display (9 blips). Note: Strong electromagnetic elds may inuence the alarm level of the capacitive measurement ele-
ment.
Display CVC 3000
Vac control
Display VNC 2
1013mbar Ext. Err.
1013 . 2
mbar
If a process stop may lead to a dangerous situation take appropriate safety measures.
Attention: The error message ”full catchpot” has to be quit by pressing key on/off at the controller CVC 3000 (obey reset time), at the controller VNC 2 the error message disappears automatically when the fault is eliminated. A reset of the error message if the sensor has been removed is only possible by choosing the factory set conguration. After draining the catchpot (obey regulations for disposal) the process has to be restarted by pressing key Start/Stop (CVC 3000) or pressing key ”Enter” (VNC 2). We recommend to check the function of the level sensor after rst assembly and than regularly (depend­ing on operation conditions): Fill the catchpot with liquid, the level sensor must alarm. The release time
is 10 seconds.
In case of false alarm (empty catchpot) it might be necessary to adjust the sensor using the individual catchpot.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check care­fully the validity of this document with respect to his product. Manual-no.: 999189 / 12/03/2013
Page 4
page 4 of Seite 4
Adjustment
The level sensor is adjusted at the factory on a coated VACUUBRAND catchpot 500 ml. Readjustment
of the level sensor might be necessary using an individual catchpot.
Fix the sensor on an empty (VACUUBRAND) catchpot as close as possible. Controller CVC 3000: Choose ”Defaults On”. Controller VNC 2: Press key ”Esc” while switching on and choose a factory set conguration. The readjustment is than made automatically for both controllers. Attention: When choosing a factory set conguration all individual settings are reset to factory set values.
Technical data
Type 699908 / 699909
Measuring principle capacitive sensor with digital evaluation
Adjustment
Sensitivity
at the factory on a coated VACUUBRAND catchpot
500 ml (cat.-no.: 638497)
adjusted to liquids with a dielectric coefcient
εr > 1.8
Integration time until release 10 s
Reset time 2.5 s
Power supply 8V DC to 30V DC, typical 24V DC
Current draw < 5 mA
Output signal VACUU•BUS
Max. permitted ambient temperature at operation / storage
Max. permitted relative atmospheric moisture
during operation
+10°C to +40°C / -10°C to +60°C
30 to 85%, no condensation
Degree of protection according to IEC 529 IP 65
Connection cable approx. 2m, VACUU•BUS
Dimensions housing L x W x H (without cable, clip, mounting plate)
50 mm x 50 mm x 30 mm
Weight approx. 0.1 kg
Materials PA / PP / TPE / PE coated spring steel
We reserve the right for technical modication without prior note!
VACUUBRAND GMBH + CO KG
-Technology for vacuum systems-
© 2013 VACUUBRAND GMBH + CO KG Printed in Germany 999189
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check care­fully the validity of this document with respect to his product. Manual-no.: 999189 / 12/03/2013
Alfred-Zippe-Str. 4 - 97877 Wertheim T +49 9342 808-0 - F +49 9342 808-5555 E-Mail: info@vacuubrand.com Web: www.vacuubrand.com