UNI-T UT803 Operating Manual

Model UT803
OPERATING MANUAL
TABLE OF CONTENTS TITLE PAGE
3 4 5 6 8 9 10 11 12 16 16 18 20 22 24 26 28 30 32 33 33 33 34 34 35 35
Model UT803: OPERATING MANUAL
1
Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons Display Symbols Measurement Operation A. DC or AC Voltage Measurement B. DC or AC Current Measurement C. Measuring Resistance D. Testing for Continuity E. Testing Diodes F. Capacitance Measurement G. Frequency Measurement H. Temperature Measurement I . Measuring Transistor Operation of Hold Mode The POWER Button The SELECT Button Turning on the Display Backlight The RANGE Button The MAX MIN Button AC/AC+DC Button
2
TITLE PAGE
35 36 36 37 38 38 39 41 42 43 44 44 45 46 47 47 48 48 49 51 52 52 53 53 54 54 55 55 55
Model UT803: OPERATING MANUAL
POWER INPUT Switch Sleep Mode RS232 Button General Specifications Accuracy Specifications A. DC Voltage B. AC Voltage C. DC Current D. AC Current E. Resistance F. Continuity Test G. Diode Test H. Capacitance I . Frequency J. Temperature K. Transistor Maintenance A. General Service B. Replacing the Fuses C. Replacing the Battery RS232C and USB Serial Port System Requirements for Installing the UT803 Interface Program RS232C Serial Port A. Connecting between the Meter and computer B. RS232C Port Cable C. Setting of RS232C Serial Ports USB Serial Port A. Connecting between the Meter and computer B. Setting of USB Serial Ports
3
Overview
This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly .
Warning To avoid electric shock or personal injury, read the “Safety Information” and “Rules for Safety Operation” carefully before using the Meter.
Digital Bench-Type T rue RMS Multimeter Model UT803 (hereafter referred to as “the Meter”) has autorange and manual range options with maximum reading 5999 and 3 5/6 digits which has a unique outlook design.
In addition to all the conventional features including DC/AC voltage, current, resistance, diode, continuity test, capacitance, temperature 0C and 0F, transistor, max/min, there is a data hold, low battery display, sleep mode, RS232C and USB standard serial port for easy connection with computer to realize macro recording
and monitoring and capture of transient dynamic data, displaying change of waveform during the measurement, providing data and evidence to engineering technicians for scientific research.
This is also a highly applied digital multimeter of good performance with full overload protection and display backlight function.
Model UT803: OPERATING MANUAL
4
Unpacking Inspection
Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part:
Item Description Qty
In the event you find any missing or damage, please contact your dealer immediately.
Model UT803: OPERATING MANUAL
Operating Manual Test Lead Alligator Clip Multi-Purpose Socket Point Contact Temperature Probe (to be used under 2300C temperature measurement) Power Cord (AC 250V) RS232C Interface Cable USB Interface Cable Installation Guide & Computer Interface Software (CD-ROM)
1 2 3 4 5 6 7 8 9
1 piece 1 pair 1 pair 1 piece 1 piece 1 piece 1 piece 1 piece 1 piece
5
Safety Information
This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT. I 1000V, CAT.II 600V) and double insulation.
CAT III: Distribution level, fixed installation, with smaller transient overvoltages than CAT. IV CAT IV: Primary supply level, overhead lines, cable systems etc.
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired. In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test.
A Note identifies the information that user should pay attention on.
International electrical symbols used on the Meter and in this Operating Manual are explained on page 8.
Model UT803: OPERATING MANUAL
6
Rules For Safe Operation
Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules:
Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads with identical model number or electrical specifications before using the Meter. Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and grounding. The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter.
When the Meter working at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken for there is danger of electric shock. Use the proper terminals, function, and range for your measurements. If the value to be measured is unknown, use the maximum measurement position and reduce the range step by step until a satisfactory reading is obtained. Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened. When using the test leads, keep your fingers behind the finger guards. Disconnect circuit power and discharge all high­voltage capacitors before testing resistance, continuity, diodes, current, or capacitance. Before measuring current, check the Meter’s fuses and turn off power to the circuit before connecting the Meter to the circuit.
Model UT803: OPERATING MANUAL
l
l
l
l
l
l
l
l
l
l
l
7
Replace the battery as soon as the battery indicator
appears. With a low battery, the Meter might
produce false readings that can lead to electric shock and personal injury. Remove test leads, temperature probe, RS232C interface cable, USB interface cable and alligator clip from the Meter and turn the Meter power off before opening the Meter case. When servicing the Meter, use only the same model number or identical electrical specifications replacement parts. The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident. Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident. The Meter is suitable for indoor use.
Turn the Meter power off when it is not in use. Take out the battery when not using for a long time if using battery to power on the Meter. When using battery to power on the Meter, constantly check the battery as it may lead when it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter.
Model UT803: OPERATING MANUAL
l
l
l
l
l
l
l l
l
8
Model UT803: OPERATING MANUAL
International Electrical Symbols
AC or DC Grounding Double Insulated Warning. Refer to the Operating Manual Deficiency of Built-In Battery Continuity Test Diode Capacitance Test Fuse Conforms to Standards of European Union
9
The Meter Structure (see figure 1)
1. LCD Display 2. Rotary Switch
3. Input Terminals 4. Functional Buttons
Model UT803: OPERATING MANUAL
(Figure 1)
10
Rotary Switch
Below table indicated for information about the rotary switch positions:
AC and DC voltage measurement :Continuity test. :Diode test. :Resistance measurement. Capacitance test :Frequency measurement. :Temperature in Fahrenheit Temperature in celsius Transistor test AC or DC current measurement range from 0.1uA to 5999uA. AC or DC current measurement range from 0.01mA to 599.9mA. AC or DC current measurement range from 0.01A to 10.00A.
V
mA
A
A
Model UT803: OPERATING MANUAL
Rotary Switch Position
Function
Hz
0
F
0
C
hFE
Hz
0
F
11
Functional Buttons
Below table indicated for information about the functional button operations.
Turn the power on and off. Turn the display backlight on and off. Switches between AC and DC measurement. Switches between continuity and diode and resistance measurements Switches between frequency and Fahrenheit temperature. Press HOLD to enter and exit the Hold mode in any mode, the Meter beeps. Press RANGE to switch between manual and auto ranging. Turn on or off the serial port interface without changing the original setting. Starts recording of maximum and minimum values. Steps the display through high (MAX) and low (MIN). To select AC or AC+DC measurement.
Model UT803: OPERATING MANUAL
Button
POWER
LIGHT
SELECT
HOLD
RANGE
RS232C
MAX MIN
AC AC+DC
Operation Performed
l l l
12
HOLD
hFE MIN
MAX
RS232
AC+DC Trus RMS
Auto Range Manual
Trms
0
C umVA
DC munF 0F
~mV
Mk Hz
12 10 1 11
5.6.7
14
2
16
13 3 4
8
9
15
Model UT803: OPERATING MANUAL
Display Symbols (see figure 2)
(Figure 2)
13
Symbol
Number
Model UT803: OPERATING MANUAL
True RMS
HOLD
AC DC
AC+DC
OL
,K ,M
A, mA, µA
Meaning
Indicator for true rms value. Data hold is active. Sleep Mode feature is enabled. Indicates negative reading. Indicator for AC voltage or current. Indicator for DC voltage or current Indicator for AC+DC measurement The input value is too large for the selected range. :Ohm. The unit of resistance. :kilohm. 1 x 103 or 1000 ohms. :Megaohm.1x106 or 1,000,000 ohms. A :Amperes (amps).The unit of current. mA:Milliamp. 1 x 10-3 or 0.001 amperes. µA : Microamp. 1x 10-6 or 0.000001 amperes.
1 2 3 4 5 6 7 8 9
K M
14
Symbol
Number
Model UT803: OPERATING MANUAL
Meaning
V : Volts. The unit of voltage. mV: Millivolt.1 x 10-3 or 0.001 volts. F : Farad.The unit of capacitance. mF: Millifarad.1 x 10-3 or 0.001 farads µF : Microfarad.1x10-6 or 0.000001 farads. nF : Nanofarad.1 x 10-9 or 0.000000001 farads. Centigrade temperature Temperature in Fahrenheit Hz : Hertz. The unit of frequency in cycles/second. kHz : Kilohertz. 1 x 103 or 1,000 hertz. MHz: Megahertz. 1 x 106 or 1,000,000 hertz. Unit of transistor. Test of diode The continuity buzzer is on.
9
10 11
V, mV
F, mF, µF,nF
0
C, 0F
Hz, kHz,
MHz
β
15
Symbol
Number
Model UT803: OPERATING MANUAL
AutoRange
Manual
Meaning
12
13 14 15
16
Indicator of Auto or manual range
Display of maximum or minimum value. Data output is in progress. The battery is low.
Warning: To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears. Transistor test is on
MAX
RS232
MIN
hFE
16
Measurement Operation
A. DC or AC Voltage Measurement (See figure 3)
Warning To avoid harms to you or damages to the Meter from electric shock, please do not attempt to measure voltages higher than 1000V although readings may be obtained.
To measure DC/AC voltage, connect the Meter as follows:
Model UT803: OPERATING MANUAL
1.
2.
3.
Insert the red test lead into the V terminal and the black test lead into the COM terminal. If the measured value is less than 600mV, insert the red test lead into mV terminal instead and press RANGE button to select manual range 600.0mV mode, the LCD displays “MANUAL” and “mV”. Set the rotary switch to V ; press SELECT button to select DC or AC measurement mode. Connect the test leads across with the object being measured. The measured value shows on the display . AC measurement displays True RMS value.
(Figure 3)