UNI-T UT70C Operating Manual

Page
3 4 5 6 8 8
9 10 12 14 14 14 16 16 17 18
19 23 24 26 28 30 31 32
33 34 34 35 35 36
Table of Contents
Title
Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons Display Symbols Measurement Ranges
A. Selecting a Measurement Range B. Manual Ranging and Autoranging
Measurement Operation
A. AC Voltage Measurement B. DC Voltage Measurement C. DC Milivolt Measurement D. Measuring Continuity, Resistance,
Conductance & Capacitance E. Testing Diodes F. DC or AC Current Measurement G. Frequency & Duty Cycle Measurement
MAX MIN Recording Mode PEAK MAX MIN Mode Operation of Hold Mode Operation of Auto Hold Mode The Use of Relative Value and Relative Percentage Value Mode Turning on the Display Backlight Analogue Bar Graph Display Full Icons Display Sleep Mode General Specifications
Model UT70C: OPERATING MANUAL
1
Page
37 37 37
38 38 38 39 39 40 41 41 42 43 44 45 45 45
46
Accuracy Specifications
A. AC Voltage B. DC Voltage C. Continuity, Resistance & Conductance Test D. Capacitance
E.
Diodes Test
F.
DC Current
G.
AC Current
H. Frequency & Duty Cycle
Maintenance
A. General Service B. Testing the Fuses C. Replacing the Battery D. Replacing the Fuses
RS232C Serial Port
A. RS232C Port Cable B. Setting of RS232C Serial Ports C. System Requirements for Installing the
UT70C Interface Program
Model UT70C: OPERATING MANUAL
2
Overview
This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly.
Warning To avoid electric shock or personal injury, read the "Safety Information" and "Rules for Safe Operation" carefully before using the Meter.
Your multimeter is an intelligent digital one, a precise instrument with a resolution of 8,000 counts and up-to­date automatic computer calibrating function. High resolution A/D converter and micro-controller data processing technique is adopted in the Meter, featured with intelligence, high precision and multi-functions. The Meter can be widely used in laboratory, field service, domestic and other applications. All the functions and ranges have overloaded protection.
Model UT70C: OPERATING MANUAL
3
Unpacking Inspection
Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part:
In the event you find any missing or damage, please contact your dealer immediately.
Item
1 2 3 4
5 6
7 8
Description
Operating Manual Test Lead Test Clip 9V Battery (NEDA1604, 6F22 or 006P) (installed) RS232C Interface Cable CD-ROM (Installation Guide & Computer Interface Software) Holster Tilt Stand
Qty
1 piece
1 pair 1 pair
1 piece 1 piece
1 piece 1 piece
1 piece
Model UT70C: OPERATING MANUAL
4
Safety Information
CA T. II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient voltage overvoltages than CAT. III
CA T. III: Distribution level, fixed installation, with smaller transient overvoltages than CAT. IV
CAT IV: Primary supply level, overhead lines, cablesystems etc.
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test.
A Note identifies the information that user should pay attention on.
International electrical symbols used on the Meter and in this Operating Manual are explained on page 8.
Model UT70C: OPERATING MANUAL
5
CE Version: The Meter complies with the standards IEC61010-1:in pollution degree 2, overvoltage category CAT III 1000V, CAT IV 600V and double insulation . UL V ersion: The Meter complies with the standards UL61010B-1, in pollution degree 2, overvoltage category CAT II 1000V and double insulation.
Rules For Safe Operation (1)
Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules:
l Before using the Meter inspect the case. Do
not use the Meter if it is damaged or the case
(or part of the case) is removed. Look for cracks
or missing plastic. Pay attention to the insulation
around the connectors. l Inspect the test leads for damaged insulation
or exposed metal. Check the test leads for
continuity. Replace damaged test leads with
identical model number or electrical
specifications before using the Meter. l Do not apply more than the rated voltage, as
marked on the Meter, between the terminals or
between any terminal and grounding. l The rotary switch should be placed in the right
position and no any changeover of range shall
be made during measurement is conducted to
prevent damage of the Meter. l Never working at an effective voltage over 60V
in DC or 30V rms in AC for there is danger of
electric shock. l Use the proper terminals, function, and range
for your measurements. l Do not use or store the Meter in an environment
of high temperature, humidity, explosive,
inflammable and strong magnetic field. The
performance of the Meter may deteriorate after
dampened. l When using the test leads, keep your fingers
behind the finger guards.
Model UT70C: OPERATING MANUAL
6
Rules For Safe Operation (2)
l Disconnect circuit power and discharge all high-
voltage capacitors before testing resistance,
conductance, continuity, diodes, current, or
capacitance. l Before measuring current, check the Meter's
fuses and turn off power to the circuit before
connecting the Meter to the circuit. l Replace the battery as soon as the battery
indicator
appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury.
l Remove test leads, RS232C interface cable and
test clip from the Meter and turn the Meter power off before opening the Meter case.
l When servicing the Meter, use only the same
model number or identical electrical specifications replacement parts.
l The internal circuit of the Meter shall not be
altered at will to avoid damage of the Meter and any accident.
l Soft cloth and mild detergent should be used
to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident.
l The Meter is suitable for indoor use. l Turn off the Meter when it is not in use and
take out the battery when not using for a long time.
l Constantly check the battery as it may leak when
it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter.
Model UT70C: OPERATING MANUAL
7
International Electrical Symbols
The Meter Structure (see figure 1)
(figure 1)
1 Front Case 2 LCD Display 3 Functional Buttons 4 Rotary Switch 5 Input Terminals
AC (Alternating Current) DC (Direct Current) AC or DC Grounding Double Insulated Deficiency of Built-In Battery Continuity Test Diode Capacitance Test Fuse Warning. Refer to the Operating Manual Conforms to Standards of European Union
Model UT70C: OPERATING MANUAL
8
mV
V
OFF
V
Rotary Switch
Below table indicated for information about the rotary switch positions.
Rotary Switch
Position
Function
Power is turned off. AC voltage measurement.
DC voltage measurement. DC millivolt measurement.
: Continuity test. : Resistance measurement. : Capacitance test.
Diode Test. AC or DC Current Measurement
from 0.001A to 10.00A . AC or DC Current Measurement from 0.1mA to 800.0mA.
A
mA
Model UT70C: OPERATING MANUAL
9
Functional Buttons (1)
Below table indicated for information about the functional button operations.
Button Operation PerformedMeasuring
Function
Continuity Test
Turn the continuity buzzer on and off in resistance measurement mode.
At OFF position
At REC mode or REC mode plus Hold mode
Press while turning on the Meter to disable the beeper at all functions except short circuit and diode testing.
To toggle between 100ms response time and 1ms response time under REC mode except at resistance, capacitance and continuity range.
MAX
MIN
Any rotary switch position except at conductance mode.
Starts recording of maximum and minimum values. Steps the display through the sequence of present, high (MAX), low (MIN) and average (AVG) readings at any mode. In 1ms response time, it steps through only MAX and MIN reading.
At OFF position
To select 1 second high accuracy response time, press and hold down MAX MIN while turning the rotary switch to any function setting.
RANGEAny rotary
switch position
1. Press RANGE to enter the manual ranging mode; the Meter beeps.
2. Press RANGE to step through the ranges available for the selected function; the Meter beeps.
3. Press and hold RANGE for over 1 second to return to autoranging; the Meter beeps.
Model UT70C: OPERATING MANUAL
10
HOLD Any rotary
switch position
Press HOLD to enter and exit the Hold mode in any mode; the Meter beeps.
Functional Buttons (2)
Conductance mode
Press HOLD to toggle between conductance and resistance value. Present resistance value is kept.
Press to select capacitance test.
A , mA Press to select DC or AC
current. DC is default.
At OFF position
To disable the Sleep Mode, press
while turning on the Meter.
REL
Any rotary switch position
Press REL to enter and exit the
and
% mode in any measuring
mode; the Meter beeps.
Hz V , V ,
mV
, A ,
mA
1. Press Hz to enter frequency measurement mode; the Meter beeps.
2. Press Hz again to enter duty cycle measurement mode; the Meter beeps.
3. Press Hz again to return to the last measurement mode; the Meter beeps.
Display Backlight
Press once to turn the Display Backlight on and it shall shut off automatically in around 60 seconds later.
Model UT70C: OPERATING MANUAL
11
No. MeaningSymbol
Display Symbols (1) (see figure 2)
(figure 2)
1
The Meter is in the auto range mode in which the Meter automatically selects the range with the best resolution.
2
The battery is low.
Warning: To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.
3
Indicates negative reading.
4
Test of diode.
5
The continuity buzzer is on.
6
The relative value mode is on, which display the present value minus the stored value.
%
7
The relative percentage value mode is on. It displays the percentage value of the relative value and stored value.
DATA
HOLD
8
Data hold is active.
DC AC Indicator for DC/AC voltage or current.
The displayed value is the mean value.
9
Display the present chosen range.8000mV
10
Percent, used for duty cycle measurements.
%
11
Model UT70C: OPERATING MANUAL
12
V, mV
Display Symbols (2)
12
,
k , M
W: Ohm. The unit of resistance. kW: kilohm. 1 x 103 or 1000 ohms. MW: Megohm. 1 x 106 or 1,000,000
ohms.
Hz, kHz,
MHz
Hz: Hertz. The unit of frequency. kHz: Kilohertz. 1 x 103 or 1000 hertz. MHz: Meghertz. 1 x 106 or1,000,000
hertz.
V: Volts. The unit of voltage. mV: Millivolt. 1 x 10-3 or 0.001 volts.
A: Amperes (amps). The unit of
current.
mA: Milliamp. 1 x 10-3 or 0.001
amperes.
F: Farad. The unit of capacitance. mF: Microfarad. 1 x 10
-6
or 0.000001
farads.
nF: Nanofarad. 1 x 10-9 or
0.000000001 farads.
S: Siemens. The unit of
conductance.
nS: Nanosiemens. 1 x 10-9 or
0.000000001 siemens.
Maximum, minimum and average recording mode enabled. Present reading displayed.
13
100ms response time. Maximum reading displayed. Minimum reading displayed. Average reading displayed.
14
Provides an analogue indication of the present input, quick response.
15
Polarity indicator for the analogue bar graph display, no display for positive.
The input value is too large for the selected range.
16
Computer and the Meter is connected properly. Data output is in progress.
17
A, mA
F , F, nF
nS
100ms
MAX
MIN
AVG
Analogue
Bar
Graph
RS232C
K : M :
:
Model UT70C: OPERATING MANUAL
13
Measurement Ranges (1)
A measurement range determines the highest value the Meter can measure. Most Meter functions have more than one range. See "Accuracy Specifications."
A. Selecting a Measurement Range
Being in the right measurement range is important: l If the range is too low for the input, the Meter displays
to indicate an overload.
l If the range is too high, the Meter will not display
the most accurate measurement.
B. Manual Ranging and Autoranging
The Meter has both manual range and autorange options: l In the autorange mode, the Meter selects the best
range for the input detected. This allows you to switch test points without having to reset the range.
l In the manual range mode, you select the range.
This allows you to override autorange and lock the Meter in a specific range.
The Meter defaults to the autorange mode in measurement functions that have more than one range. When the Meter is in the autorange mode, is displayed.
To enter and exit the manual range mode:
Model UT70C: OPERATING MANUAL
14
1. Press RANGE. The Meter enters the manual range mode and turns off. Each presses of RANGE increments the range. When the highest range is reached, the Meter wraps to the lowest range.
2. To exit the manual range mode, press and hold RANGE for over one second. The Meter returns to the autorange mode and is displayed.
Note
l In Hold and MAX MIN mode, press RANGE exits
these measurement modes and enters manual range mode.
l In Hz measurement mode, press RANGE to attenuate
10 times of input signal amplitude which can increase the Meter sensitivity and stabilize the correct reading.
Measurement Ranges (2)
Model UT70C: OPERATING MANUAL
15