UNI-T ut60c, ut60b, ut60e Operating Manual

Temperature: Operating : 0oC~40oC (32oF ~104 oF). Storage : -10oC~50oC (14oF~122oF). :
75% @ 0oC - 30oC; 50% @ 31oC - 40oC;
UL & CUL
General Specifications
l Maximum Voltage between
any Terminals and Grounding: 1000V.
l
Fused Protection for Input Terminal: Model
UT60B:
Glass fuse, 0.5A, 250V, fast type, 5x20mm.
l
Fused Protection for Input Terminal: Model
UT60C/E:
l
Fused Protection for 10AInput Terminal:
UT60B/C/E: Glass fuse, 10A, 250V, fast type, 5x20mm.
Glass fuse, 0.5A, 250V, fast type, 5x20mm.
l Maximum Display: Digital: 3999 l Measurement Speed: Updates 3 times/second. l
l Relative Humidity: l Altitude: Operating: 2000 m.
Storage : 10000 m.
l Battery Type: One piece of 9V (NEDA1604 or 6F22 or
006P).
l Battery Deficiency: Display l Dimensions (HxWxL): 177 x 85 x 40 mm. l Weight: Approximate 300g (battery included). l Safety/Compliances: IEC61010 CAT.III 1000V,
CAT.IV 600V overvoltage and double insulation standard.
l Certifications: ,
26
Model UT60B/C/E: OPERATING MANUAL
Accuracy: (a% reading + b digits), guarantee for 1 year. Operating temperature: 23oC 5oC. Relative humidity: 75%. Temperature coefficient: 0.1 x (specified accuracy)/1oC
A. DC Voltage
Range Resolution Accuracy
Overload
Protection
400mV
0.1mV
(0.8%+3)
4V
40V
400V
(0.8%+1)
1000V
1V
(1%+3)
1000V DC
750V AC rms
continuous.
1mV
10mV
100mV
l Frequency response 40Hz~400Hz.
400
0.1
1
4k
40k
10
400k 100
4M
1k
Accuracy Specifications (1)
Remarks: Input impedance
C. Resistance
Range Resolution Accuracy
Overload
Protection
(1%+2)
(1.2%+2) (1.5%+2)
1000Vp
Remarks: Open circuit voltage approximate 0.45V.
40M
10k
Measure at
REL mode
(1.2%+2)
B. AC Voltage
Range Resolution Accuracy
Overload
Protection
4V
40V
400V
(1%+5)
750V
1V
(1.2%+5)
1000V DC
750V AC rms
continuous.
Remarks:
l Input impedance 10M . l Model UT60B/UT60C: displays effective value of sine
wave (mean value response). Model UT60E: displays true rms value.
100mV
10mV
1mV
27