Frequency and Amplitude Control with Arbitrary Waveforms 50
Sync Out Settings with Arbitrary Waveforms 51
Waveform Hold in Arbitrary Mode 52
Output Filter Setting 52
Pulse and Pulse-trains 53
Pulse Set-up 53
Pulse-train Setup 55
Waveform Hold in Pulse and Pulse-Train Modes 58
1
Modulation 59
Introduction 59
External Modulation 59
Internal Modulation 60
Sum 61
Inter-Channel Synchronisation 63
Synchronising Two Generators 66
Memory Card and Other System Operations from the Utility Menu 68
Memory Card – General Information 68
System Operations from the Utility Menu 70
Calibration 74
Equipment Required 74
Calibration Procedure 74
Calibration Routine 75
Remote Calibration 78
Remote Operation 79
Power on Settings 86
Remote Commands 87
Channel Selection 88
Frequency and Period 88
Amplitude and DC Offset 88
Waveform Selection 89
Arbitrary Waveform Create and Delete 89
Arbitrary Waveform Editing 90
Waveform Sequence Control 92
Mode Commands 92
Input/Output control 92
Modulation Commands 93
Synchronising Commands 93
Status Commands 93
Miscellaneous Commands 94
Remote Command Summary 95
Maintenance 99
Appendix 1. Warning and Error Messages 100
Appendix 2. SYNC OUT Automatic Settings 104
Appendix 3. Factory System Defaults 105
Appendix 4. Waveform Manager Plus Arbitrary Waveform Creation and Management Software 106
Block Diagrams 107
2
Introduction
This range of synthesised programmable arbitrary waveform generators has the following
features:
• 1, 2 or 4 independent arb channels
• Additional DC to 50MHz fixed amplitude sine and squarewave outputs on 2- and
4-channel instruments
• Up to 100MHz sampling frequency
• Sinewaves up to 40MHz, squarewaves up to 50MHz
• Output level 2.5mV to 10Vp−p into 50Ω with 12 bit vertical resolution
• 1M points horizontal resolution per channel
• Compact Flash card for non−volatile waveform memory
• Waveform linking, looping and sequencing
• Interchannel triggering, summing, modulation and phase control
• GPIB, RS232 and USB interfaces
The instrument uses a combination of direct digital synthesis and variable clock techniques to
provide high performance and extensive facilities in a compact instrument. It can generate a wide
variety of waveforms between 0·1mHz and 50MHz with high resolution and accuracy.
Arbitrary waveforms may be defined with 12 bit vertical resolution and from 8 to 1048576
horizontal points. In addition a number of standard waveforms are available including sine,
square, triangle, ramp and pulse.
Arbitrary waveforms may be replayed at a user specified waveform frequency or period, or the
sample rate may be defined in terms of period or frequency. Alternatively, an external Arb clock
may be used at frequencies up to 50MHz.
Extensive waveform editing features between defined start and end points are incorporated,
including waveform insert, point edit, line draw, amplitude adjust and invert. More comprehensive
features are available using the arbitrary waveform creation software supplied. This is a powerful
Windows−based design tool that enables the user to create waveforms from mathematical
expressions, from combinations of other waveforms, freehand, or using a combination of all three
techniques. Waveforms created in this way can be downloaded via the RS232, GPIB or USB
interfaces or transferred directly to the generator on a removable memory card written to by the
PC using the USB-connected card reader/writer provided.
Up to 500 different waveforms may be stored with the length and name specified by the user; the
total size of all the waveforms stored is limited only by the size of the memory card. Waveforms
may be linked together to form a sequence of up to 1024 steps. Each waveform may have a user
defined repeat count from 1 to 32768.
All waveforms can be swept over their full frequency range at a rate variable between 1
millisecond and 15 minutes. Sweep can be linear or logarithmic, single or continuous. Single
sweeps can be triggered from the front panel, the trigger input, or the digital interfaces. A sweep
marker is provided.
Amplitude Modulation is available for all waveforms and is controlled from the previous channel or
from an external generator via the MODULATION input socket.
Signal Summing is available for all waveforms and is controlled from the previous channel or from
an external generator via the SUM input socket.
3
All waveforms are available as a Triggered Burst whereby each active edge of the trigger signal
will produce one burst of the carrier. The number of cycles in the burst can be set between 1 and
1048575. The Gated mode turns the output signal On when the gating signal is true and Off
when it is false. Both Triggered and Gated modes can be operated from the previous or next
channel, from the internal Trigger Generator (0.005Hz to 100kHz), from an external source (dc to
1MHz) or by a key press or remote command.
Any number of channels can be synchronised with user defined phase angle between channels.
This can be used to generate multi−phase waveforms or synchronised waveforms of different
frequencies.
The signals from the REF IN/OUT socket and the SYNC OUT socket can be used to synchronise
two instruments where more than 4 channels are required.
The generator parameters are clearly displayed on a backlit LCD with 4 rows of 20 characters.
Soft−keys and sub menus are used to guide the user through even the most complex functions.
All parameters can be entered directly from the numeric keypad. Alternatively most parameters
can be incremented or decremented using the rotary control. This system combines quick and
easy numeric data entry with quasi−analogue adjustment when required.
The generator has RS232, GPIB and USB interfaces as standard which can be used for remote
control of all of the instrument functions or for the down−loading of arbitrary waveforms. As well
as operating in conventional RS232 mode the serial interface can also be used in addressable
mode whereby up to 32 instruments can be linked to a single PC serial port.
4
Specifications apply at 18−28ºC after 30 minutes warm−up, at maximum output into 50Ω
Range: 0·1mHz to 40MHz
Resolution: 0·1mHz or 10 digits
Accuracy: 10 ppm for 1 year
Temperature Stability: Typically <1 ppm/ºC.
Output Level:
Harmonic Distortion: <0.15% THD to 100kHz; <–60dBc to 20kHz
<–50dBc to 1MHz,
Non−harmonic Spurii:
2.5mV to 10Vp−p into 50Ω
<−40dBc to 10MHz
<−30dBc to 40MHz
<–60dBc to 1MHz, <–60dBc + 6dB/octave 1MHz to 40MHz
Specifications
Square
Range: 1mHz to 50MHz
Resolution: 1mHz or 8 digits
Accuracy: 10ppm for 1 year
Output Level:
Rise and Fall Times: <8ns
Triangle
Range: 0.1mHz to 500kHz
Resolution: 0.1mHz or 10 digits
Accuracy: 10 ppm for 1 year
Output Level:
Linearity Error: <0.1% to 30 kHz
Ramps and Sin(x)/x
Range: 0.1mHz to 500kHz
Resolution: 0.1mHz (10 digits)
Accuracy: 10 ppm for 1 year
Output Level:
Linearity Error: <0.1% to 30 kHz
2.5mV to 10Vp−p into 50Ω
2.5mV to 10Vp−p into 50Ω
2.5mV to 10Vp−p into 50Ω
5
Pulse and Pulse Train
Output Level:
Rise and Fall Times: <8ns
Period:
Range: 40ns to 100s
Resolution:
Accuracy: 10ppm for 1 year
Delay:
Range:
Resolution:
Width:
Range:
Resolution:
Note that the pulse width and absolute value of the delay may not exceed the pulse period at any
time.
Pulse trains of up to 10 pulses may be specified, each pulse having independently defined width,
delay and level. The baseline voltage is separately defined and the sequence repetition rate is
set by the pulse train period.
Arbitrary
Up to 500 user defined waveforms may be stored on the removable memory card. Waveforms
can be defined by front panel editing controls, by downloading of waveform data via RS232, GPIB
or USB, or by writing directly to the removable memory card using the USB card reader/writer
connected to a PC.
Waveform Memory Size: 1M points per channel – minimum waveform size is 8 points
2.5mV to 10Vp−p into 50Ω
8−digit
−99·99s to + 99·99s
0·001% of period or 10ns, whichever is greater (8 digits)
10ns to 99·99s
0·001% of period or 10ns, whichever is greater (8 digits)
Vertical Resolution: 12 bits
Sample Clock Range: 100mHz to 100MHz
Resolution: 8 digits
Accuracy: 10ppm for 1 year
Sequence
Up to 1024 waveforms may be linked. Each waveform can have a loop count of up to 32,768.
A sequence of waveforms can be looped up to 1,048,575 times or run continuously.
Output Filter
Selectable between 40MHz Elliptic, 20MHz Bessel or none.
Noise
Digital noise generated by a 35-bit linear feedback register clocked at 100MHz. User’s external
filter defines bandwidth and response.
6
OPERATING MODES
Triggered Burst
Each active edge of the trigger signal will produce one burst of the waveform.
Carrier Waveforms: All standard and arbitrary
Maximum Carrier Frequency: The smaller of 2.5MHz or the maximum for the selected
Number of Cycles: 1 to 1,048,575
Trigger Repetition Rate: 0.005Hz to 100kHz internal
Trigger Signal Source: Internal from keyboard, previous channel, next channel or trigger
Trigger Start/Stop Phase:
Gated
Waveform will run while the Gate signal is true and stop while false.
Carrier Waveforms: All standard and arbitrary.
Maximum Carrier Frequency: The smaller of 2.5MHz or the maximum for the selected
Trigger Repetition Rate: 0.005Hz to 100kHz internal
Gate Signal Source: Internal from keyboard, previous channel, next channel or trigger
Gate Start/Stop Phase:
waveform. 100Msamples/s for ARB or Sequence.
dc to 1MHz external.
generator.
External from TRIG IN or remote interface.
± 360° settable with 0.1° resolution, subject to waveform
frequency and type.
waveform. 100Msamples/s for ARB or Sequence.
dc to 1MHz external.
generator.
External from TRIG IN or remote interface.
± 360° settable with 0.1° resolution, subject to waveform
frequency and type.
Sweep
Frequency sweep capability is provided for both standard and arbitrary waveforms. Arbitrary
waveforms are expanded or condensed to exactly 4096 points and DDS techniques are used to
perform the sweep.
Carrier Waveforms: All standard and arbitrary except pulse, pulse train and
sequence.
Sweep Mode: Linear or logarithmic, triggered or continuous.
Sweep Direction: Up, down, up/down or down/up.
Sweep Range: From 1mHz to 40MHz in one range. Phase continuous.
Independent setting of the start and stop frequency.
Sweep Time: 1ms to 999s (3 digit resolution).
Marker: Variable during sweep.
Sweep Trigger Source: The sweep may be free run or triggered from the following
sources: Manually from keyboard. Externally from TRIG IN input
or remote interface.
Sweep Hold: Sweep can be held and restarted by the HOLD key.
Multi channel sweep: Any number of channels may be swept simultaneously with
independent sweep parameters for each channel. Amplitude,
Offset and Waveform can be set independently for each channel.
7
Tone Switching
Capability provided for both standard and arbitrary waveforms. Arbitrary waveforms are
expanded or condensed to exactly 4096 points and DDS techniques are used to allow
instantaneous frequency switching.
Carrier Waveforms: All waveforms except pulse, pulse train and sequence.
Frequency List: Up to 16 frequencies from 1mHz to 40MHz.
Trigger Repetition Rate: 0.005Hz to 100kHz internal
Source: Internal from keyboard, previous channel, next channel or trigger
Tone Switching Modes:
Gated:
Triggered:
FSK: The tone is output when the trigger signal goes true and the next
Using 2 channels with their outputs summed together it is possible to generate DTMF test
signals.
dc to 1MHz external.
Usable repetition rate and waveform frequency depend on the
tone switching mode.
generator.
External from TRIG IN or remote interface.
The tone is output while the trigger signal is true and stopped, at
the end of the current waveform cycle, while the trigger signal is
false. The next tone is output when the trigger signal is true
again.
The tone is output when the trigger signal goes true and the next
tone is output, at the end of the current waveform cycle, when the
trigger signal goes true again.
tone is output, immediately, when the trigger signal goes true
again.
Trigger Generator
Internal source 0.005 Hz to 100kHz square wave adjustable in 10us steps. 3 digit resolution.
Available for external use from any SYNC OUT socket.
OUTPUTS
Main Output - One for each channel
Output Impedance:
Amplitude:
Amplitude Accuracy:
Amplitude Flatness: ±0.2dB to 1MHz; ±0.4dB to 40MHz
DC Offset Range:
DC Offset Accuracy: Typically 3% ±10mV, unattenuated.
Resolution: 3 digits or 1mV for both Amplitude and DC Offset.
50Ω
5mV to 20Vp−p open circuit (2.5mV to 10Vp−p into 50Ω).
Amplitude can be specified open circuit (hi Z) or into an assumed
load of 50Ω or 600Ω in Vpk−pk, Vrms or dBm.
2% ±1mV at 1kHz into 50Ω.
±10V. DC offset plus signal peak limited to ±10V from 50Ω.
8
Sync Out - One for each channel
Multifunction output user definable or automatically selected to be any of the following:
Waveform Sync:
(all waveforms)
Position Markers:
(Arbitrary only)
Burst Done: Produces a pulse coincident with the last cycle of a burst.
Sequence Sync: Produces a pulse coincident with the end of a waveform sequence.
Trigger: Selects the current trigger signal. Useful for synchronizing burst or
Sweep Sync: Outputs a trigger signal at the start of sweep to synchronize an
Phase Lock Out: Used to synchronise two generators. Produces a positive edge at the
Output Signal Level: Logic level of <0.8V to >3V for all outputs except Sweep Sync. Sweep
A square wave with 50% duty cycle at the main waveform frequency, or
a pulse coincident with the first few points of an arbitrary waveform.
Any point(s) on the waveform may have associated marker bit(s) set
high or low.
gated signals.
oscilloscope or recorder. Can additionally output a sweep marker.
0° phase point.
Sync is a 3-level waveform, logic level as above at start of sweep, with
narrow 1V pulses at each marker point.
Auxiliary Sine Out
Frequency Range: DC to 50MHz, set by System Clock
Output Signal Level:
System Clock
Frequency Range: DC to 50MHz, 0.1Hz resolution
INPUTS
Trig In
Frequency Range:
Signal Range: Threshold level adjustable ±5V; maximum input ±10V.
Minimum Pulse Width: 50ns, for Trigger and Gate modes; 50us for Sweep mode.
Polarity: Selectable as high/rising edge or low/falling edge.
Input Impedance:
Modulation In
Frequency Range: DC – 100kHz.
Signal Range:
Input Impedance:
1Vp−p into 50Ω
DC − 1MHz.
10kΩ
VCA: Approximately 1V pk−pk for 100% level change at maximum
output; maximum input ±10V.
SCM: Approximately ± 1Vpk for maximum output.
Typically 1 kΩ.
Sum In
Frequency Range:
Signal Range:
Input Impedance:
9
DC − 30 MHz (25MHz on 2- and 4-channel instruments).
Approximately 2 Vpk−pk input for 20Vpk−pk output; maximum input
±10V.
Typically 1kΩ.
Hold
Holds an arbitrary waveform at its current position. A TTL low level or switch closure causes the
waveform to stop at the current position and wait until a TTL high level or switch opening which
allows the waveform to continue. The front panel MAN HOLD key or remote command may also
be used to control the Hold function. The Hold input may be enabled independently for each
channel.
Input Impedance:
Maximum Input Voltage: ±10V
Ref Clock In/Out
Set to Input: Input for an external 10MHz reference clock. TTL/CMOS threshold
Set to Output: Buffered version of the internal 10MHz clock. Output levels nominally
Set to Phase Lock: Used together with SYNC OUT on a master and TRIG IN on a slave
Maximum Input Voltage: +5V, –1V.
Arb Clock In/Out
10kΩ
level.
1V and 4V from 50Ω.
to synchronise (phase lock) two separate generators.
Set to Input: Input for External Arb Clock. TTL/CMOS threshold level,
Set to Output: Outputs System Clock, logic levels <0.8V and >3V.
Frequency Range: DC to 50MHz
Maximum Input Voltage: +5V, –1V.
INTER-CHANNEL OPERATION
Inter-channel Modulation:
The waveform from any channel may be used to Amplitude Modulate (AM) or Suppressed Carrier
Modulate (SCM) the next channel. Alternatively any number of channels may be Modulated (AM
or SCM) with the signal at the MODULATION input socket.
Carrier frequency: Entire range for selected waveform.
Carrier waveforms: All standard and arbitrary waveforms.
Modulation Types:
AM:
SCM:
Modulation source: Internal from the previous channel.
Frequency Range: DC to >100 kHz.
Internal AM:
Depth:
Resolution:
Carrier Suppression (SCM):
External Modulation Signal
Range:
Double sideband with carrier.
Double sideband suppressed carrier.
External from Modulation input socket.
The external modulation signal may be applied to any number of
channels simultaneously.
0% to 105%
1%.
> −40dB.
VCA: Approximately 1V pk−pk for 100% level change at maximum
output.
SCM: Approximately ± 1Vpk for maximum output.
10
Inter-channel Analog Summing:
Waveform Summing sums the waveform from any channel into the next channel.
Alternatively any number of channels may be summed with the signal at the SUM input socket.
Carrier frequency: Entire range for selected waveform.
Carrier waveforms: All standard and arbitrary waveforms.
Sum source: Internal from the previous channel.
External from SUM IN socket.
Frequency Range: DC to >25MHz.
External Signal Range:
Approximately 5Vpk−pk input for 20Vpk−pk output.
Inter-channel Synchronisation:
Two or more channels may be synchronised together. Each synchronised channel may be
assigned a phase angle relative to the other synchronised channels. Arbitrary waveforms and
waveform sequences may be synchronised but certain constraints apply to waveform lengths and
clock frequency ratios. With one channel assigned as the Master and other channels as Slaves a
frequency change on the master will be repeated on each slave thus allowing multi−phase
waveforms at the same frequency to be easily generated.
Channels may be clocked using the master channel, the System Clock or an External Arb Clock.
The signals from the REF IN/OUT socket and the SYNC OUT socket can be used to synchronise
two instruments where more than 4 channels are required.
Inter-channel Triggering:
Any channel can be triggered by the previous or next channel.
The previous/next connections can be used to ’daisy chain’ a trigger signal from a ‘start’ channel,
through a number of channels in the ‘chain’ to an ‘end’ channel. Each channel receives the
trigger out signal from the previous (or next) channel, and drives its selected trigger out to the
next (or previous) channel. The ‘end’ channel trigger out can be set up to drive the ‘start’
channel, closing the loop.
In this way, complex and versatile inter−channel trigger schemes may be set up. Each channel
can have its trigger out and its output waveform set up independently. Trigger out may be
selected from Waveform End, Position Markers, Sequence Sync or Burst Done.
INTERFACES
0.1 degree
0.1 degree or 360 degrees/number of points whichever is the greater
<± 6ns (internal clock)
<± 2ns (External Arb or System Clock)
Full remote control facilities are available through the RS232, USB or GPIB interfaces.
RS232:
IEEE−488:
USB 1.1
Variable Baud rate, 38400 Baud maximum. 9−pin D−connector.
Conforms with IEEE488.1 and IEEE488.2
11
GENERAL
Display: 20 character x 4 row alphanumeric LCD.
Data Entry: Keyboard selection of mode, waveform etc.; value entry direct by numeric
Memory Card: Removable memory card conforming to the Compact Flash memory card
Stored Settings:
Size: 3U (130mm) height; 350mm width (2 and 4 channels),
Weight: 7.2 kg. (16 lb), 2 and 4 channels; 4.1kg (9lb) 1 channel.
Power: 220-240V, nominal 50/60Hz; 110-120V or 100V nominal 50/60/400Hz;
Operating Range:
Storage Range:
Environmental: Indoor use at altitudes up to 2000m, Pollution Degree 2.
Options: 19 inch rack mounting kit.
Safety:
EMC: Complies with EN61326
keys or by rotary control.
standard. Sizes from 32MB to 1GB can be used.
Up to 500 complete instrument set−ups may be stored and recalled from
the memory card. Up to 500 arbitrary waveforms can also be stored
independent of the instrument settings.
212mm (½−rack) single channel; 335mm long.
nominal voltage adjustable internally; operating range ±10% of nominal;
150VA max. for 4 channels, 100VA max. for 2 channel, 60VA max. for 1
channel. Installation Category II.
+5°C to 40°C, 20−80% RH.
−20°C to + 60°C.
Complies with EN61010−1.
12
Safety
This generator is a Safety Class I instrument according to IEC classification and has been
designed to meet the requirements of EN61010−1 (Safety Requirements for Electrical Equipment
for Measurement, Control and Laboratory Use). It is an Installation Category II instrument
intended for operation from a normal single phase supply.
This instrument has been tested in accordance with EN61010−1 and has been supplied in a safe
condition. This instruction manual contains some information and warnings which have to be
followed by the user to ensure safe operation and to retain the instrument in a safe condition.
This instrument has been designed for indoor use in a Pollution Degree 2 environment in the
temperature range 5°C to 40°C, 20% − 80% RH (non−condensing). It may occasionally be
subjected to temperatures between +5° and −10°C without degradation of its safety. Do not
operate while condensation is present.
Use of this instrument in a manner not specified by these instructions may impair the safety
protection provided. Do not operate the instrument outside its rated supply voltages or
environmental range.
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
When the instrument is connected to its supply, terminals may be live and opening the covers or
removal of parts (except those to which access can be gained by hand) is likely to expose live
parts. The apparatus shall be disconnected from all voltage sources before it is opened for any
adjustment, replacement, maintenance or repair.
Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided
as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of
the hazard involved.
If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture
or chemical corrosion the safety protection may be impaired and the apparatus should be
withdrawn from use and returned for checking and repair.
Make sure that only fuses with the required rated current and of the specified type are used for
replacement. The use of makeshift fuses and the short−circuiting of fuse holders is prohibited.
This instrument uses a Lithium button cell for non−volatile memory battery back−up; typical life is
5 years. In the event of replacement becoming necessary, replace only with a cell of the correct
type, i.e. 3V Li/Mn0
in accordance with local regulations; do not cut open, incinerate, expose to temperatures above
60°C or attempt to recharge.
Do not wet the instrument when cleaning it and in particular use only a soft dry cloth to clean the
LCD window. The following symbols are used on the instrument and in this manual:−
20mm button cell type 2032. Exhausted cells must be disposed of carefully
2
Caution−refer to the accompanying documentation,
incorrect operation may damage the instrument.
terminal connected to chassis ground.
l
mains supply OFF.
mains supply ON.
alternating current.
13
EC Declaration of Conformity
We Thurlby Thandar Instruments Ltd
Glebe Road
Huntingdon
Cambridgeshire PE29 7DR
England
meet the intent of the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
Compliance was demonstrated by conformance to the following specifications which have been
listed in the Official Journal of the European Communities.
EMC
Emissions: a) EN61326-1 (2006) Radiated, Class A
b) EN61326-1 (2006) Conducted, Class B
c) EN61326-1 (2006) Harmonics, referring to EN61000-3-2 (2006)
Immunity: EN61326-1 (2006) Immunity Table 1, referring to:
a) EN61000-4-2 (1995) Electrostatic Discharge
b) EN61000-4-3 (2006) Electromagnetic Field
c) EN61000-4-11 (2004) Voltage Interrupt
d) EN61000-4-4 (2004) Fast Transient
e) EN61000-4-5 (2006) Surge
f) EN61000-4-6 (2007) Conducted RF
Performance levels achieved are detailed in the user manual.
Safety
EN61010-1 Installation Category II, Pollution Degree 2.
14
CHRIS WILDING
TECHNICAL DIRECTOR
1 May 2009
This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use. Test limits used were:
a) Radiated: Class A*
b) Conducted: Class B
c) Harmonics: EN61000-3-2 (2006) Class A; the instrument is Class A by product category.
* Note: Typically, radiated emissions will meet Class B limits but some operating configurations of multi-channel
instruments may have emissions marginally exceeding Class B but within Class A.
Immunity
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use.
Test methods, limits and performance achieved are shown below (requirement shown in
brackets):
EMC
a) EN61000-4-2 (1995) Electrostatic Discharge : 4kV air, 4kV contact, Performance A (B).
b) EN61000-4-3 (2006) Electromagnetic Field:
c) EN61000-4-11 (2004) Voltage Interrupt: ½ cycle and 1 cycle, 0%: Performance A (B);
d) EN61000-4-4 (2004) Fast Transient, 1kV peak (AC line), 0·5kV peak (signal connections),
Performance A (B).
e) EN61000-4-5 (2006) Surge, 0·5kV (line to line), 1kV (line to ground), Performance A (B).
f) EN61000-4-6 (2007) Conducted RF, 3V, 80% AM at 1kHz (AC line only; signal
connections <3m, therefore not tested), Performance A (A).
According to EN61326-1 the definitions of performance criteria are:
Performance criterion A: ‘During test normal performance within the specification limits.’
Performance criterion B: ‘During test, temporary degradation, or loss of function or
performance which is self-recovering’.
Performance criterion C: ‘During test, temporary degradation, or loss of function or
performance which requires operator intervention or system reset occurs.’
Cautions
To ensure continued compliance with the EMC directive observe the following precautions:
3V/m, 80% AM at 1kHz, 80MHz – 1GHz: Performance A (A) and 1.4GHz to 2GHz:
Performance A (A); 1V/m, 2.0GHz to 2.7GHz: Performance A (A).
25 cycles, 70% and 250 cycles, 0%: Performance B (C).
a) Connect the generator to other equipment using only high quality, double-screened cables.
b) After opening the case for any reason ensure that all signal and ground connections are
remade correctly and that case screws are correctly refitted and tightened.
c) In the event of part replacement becoming necessary, only use components of an identical
type, see the Service Manual.
15
Mains Operating Voltage – Single Channel Instrument
Check that the instrument operating voltage marked on the rear panel is suitable for the local
supply. Should it be necessary to change the operating voltage, proceed as follows:
1) Disconnect the instrument from all voltage sources.
2) Remove the screws which retain the top cover and lift off the cover.
3) Change the transformer connections following the diagram below. To change the connection,
cut the brown wire from the switch at the switch end of the crimped butt connector; make safe
the exposed end of the connector. Strip 6mm of insulation from the end of the cut brown wire;
fit into the connector for the new supply voltage and crimp. Check that the connection is
mechanically secure and that there are no loose strands.
Installation
for 230V operation connect the brown transformer wire to the brown wire from the switch
for 115V operation connect the red transformer wire to the brown wire from the switch
for 100V operation connect the black transformer wire to the brown wire from the switch
4) Refit the cover and the secure with the same screws.
5) To comply with safety standard requirements the operating voltage marked on the rear panel
must be changed to clearly show the new voltage setting.
6) Change the fuse to one of the correct rating, see below.
Mains Operating Voltage – 2- and 4-Channel Instruments
These instruments have a universal input range and will operate from a nominal 100V, 115V or
230V mains supply without adjustment. Check that the local supply meets the AC Input
requirement given in the Specification.
Fuse
Ensure that the correct mains fuse is fitted for the set operating voltage. The correct mains fuse
types are:
2- and 4-channel instruments 2A(T)250V HRC 2A(T)250V HRC
To replace the fuse, disconnect the mains lead from the inlet socket and withdraw the fuse drawer
below the socket pins. Change the fuse and replace the drawer.
The use of makeshift fuses or the short−circuiting of the fuse holder is prohibited.
Mains Lead
When a three core mains lead with bare ends is provided it should be connected as follows:−
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
Mounting
This instrument is suitable both for bench use and rack mounting. It is delivered with feet for
bench mounting. The front feet include a tilt mechanism for optimal panel angle.
A rack kit for mounting in a 19” rack is available from the Manufacturers or their overseas agents.
Ventilation
The generator uses a fan fitted to the rear panel. Take care not to restrict the rear air inlet or the
vents at the front (sides and underneath). In rack-mounted situations allow adequate space
around the instrument and/or use a fan tray for forced cooling.
Brown
Blue
Green / Yellow
WARNING! THIS INSTRUMENT MUST BE EARTHED
Mains Live
−
Mains Neutral
−
Mains Earth
−
17
Front Panel Connections
MAIN OUT (1 per channel)
This is the 50Ω output from the channel’s main generator. It will provide up to 20V peak−to−peak
e.m.f. which will yield 10V peak−to−peak into a matched 50Ω load. It can tolerate a short circuit
for 60 seconds.
Do not apply an external voltage to this output.
SYNC OUT (1 per channel)
This is a TTL/CMOS level output which may be set to any of the following signals from the
SYNC OUT screen.
Connections
waveform sync
position marker
Burst done
Sequence sync
Trigger
Sweep sync
Phase lock
A sync marker phase coincident with the MAIN OUT waveform. For
standard waveforms, (sine, cosine, haversines, square, triangle, sinx/x
and ramp), the sync marker is a squarewave with a 1:1 duty cycle with
the rising edge at the 0º phase point and the falling edge at the 180º
phase point. For arbitrary waveforms the sync marker is a positive pulse
coincident with the first few points (addresses) of the waveform.
When position (pos’n) marker is selected, the instrument generates a
pulse marker pattern for arbitrary waveforms. The pulse pattern is
programmable from the
screen.
Provides a signal during Gate or Trigger modes which is low while the
waveform is active at the main output and high at all other times.
Provides a signal which is low during the last cycle of the last waveform
in a sequence and high at all other times.
Provides a positive going version of the actual trigger signal; internal,
external, manual and remote all produce a trigger sync.
Goes high at the start of sweep and remains high for the duration of the
first frequency step. In addition, a half-amplitude marker pulse can be
set to be output at any of the frequency steps.
Produces a positive edge coincident with the start of the current
waveform; this is used for synchronising instruments. This waveform
may not appear coherent.
edit waveform menu on the MODIFY
SYNC OUT logic levels are nominally 0V and 5V from typically 50Ω. SYNC OUT will withstand a
short circuit.
TRIG IN
This is the external input for Trigger, Gate, Sweep and Sequence operations. It is also the input
used to synchronise the generator (as a slave) to another (which is the master).
SUM IN
This is the input socket for external signal summing. The channel(s) with which this signal is to
be summed are selected on the
18
Do not apply an external voltage to this output.
Do not apply an external voltage exceeding ±10V.
SUM screen.
Do not apply an external voltage exceeding ±10V.
MODULATION IN
This is the input socket for external modulation. Any number of channels may be AM or SCM
modulated with this signal; the target channels are selected on the
Do not apply an external voltage exceeding ±10V.
Rear Panel Connections
REF CLOCK IN/OUT
The function of the CLOCK IN/OUT socket is set from the REF/SYS CLOCKmenu on the
UTILITY screen, see System Operations section.
MODULATION screen.
input
output
master/slave
As an output the logic levels are nominally 1V and 4V from typically 50Ω. CLOCK OUT will
withstand a short−circuit. As an input the threshold is TTL/CMOS compatible.
Do not apply external voltages exceeding + 5V or –1V to this signal connection.
HOLD IN
Controls the waveform hold function. The input impedance is nominally 10kΩ.
Do not apply an external voltage exceeding ±10V.
ARB CLOCK IN/OUT
Set to an Input, this is the input for a user-supplied ARB clock in the frequency range DC to
50MHz.
This is the default setting. The socket becomes an input for an external
10MHz reference clock. The system automatically switches over from the
internal clock when the external reference is applied.
The internal 10MHz clock is made available at the socket.
When two or more generators are synchronised the slaves are set to slave
and the master is set to master.
Set to an Output, it outputs the System Clock at TTL/CMOS compatible logic levels.
Do not apply an external voltage exceeding + 5V or –1V.
MAIN OUT (1 per channel)
These plugged panel positions are provided for the user to fit a 50Ω BNC as an alternative to
each front panel MAIN OUT socket where rear panel connections are required in a rack-mounted
system. The front panel MAIN OUT connection must be carefully disconnected from the pcb and
the pcb then rewired, using high quality 50Ω coax, to the new rear panel connector.
Do not apply an external voltage to these outputs.
19
RS232
9−pin D−connector compatible with addressable RS232 use. The pin connections are shown
below:
Pin NameDescription
Pin 2, 3 and 5 may be used as a conventional RS232 interface with XON/XOFF handshaking.
Pins 7, 8 and 9 are additionally used when the instrument is used in addressable RS232 mode.
Signal grounds are connected to instrument ground. The RS232 address is set from the
remote menu on the UTILITY screen, see System Operations section.
GPIB (IEEE−488)
The GPIB interface is not isolated; the GPIB signal grounds are connected to the instrument
ground.
1
−
2 TXD Transmitted data from instrument
3 RXD Received data to instrument
4
−
5 GND Signal ground
6
−
7 RXD2 Secondary received data
8 TXD2 Secondary transmitted data
9 GND Signal ground
No internal Connection
No internal connection
No internal connection
The implemented subsets are:
The GPIB address is set from the remote menu on the UTILITY screen, see System
Operations section.
USB
The USB port is connected to instrument ground. It accepts a standard USB cable. If USB has
been selected as the current interface and the driver has been installed from the CD, the
Windows plug-and-play function should automatically recognise that the instrument has been
connected. See the USB folder on the CD for information on installing the driver on a PC.
MEMORY CARD
The MEMORY CARD slot accepts a standard Compact Flash Card, size 32MB to 1GB. The
‘Memory Card Active’ lamp on the front panel is lit during memory card reads and writes.
SH1 AH1 T6 TE0 L4 LE0 SR1 RL1 PP1 DC1 DT1 C0 E2
20
Initial Operation
This section is a general introduction to the organisation of the instrument and is intended to be
read before using the generator for the first time. Detailed operation is covered in later sections
starting with Standard Waveform Operation.
In this manual front panel keys and sockets are shown in capitals, e.g. CREATE, SYNC OUT; all
soft−key labels, entry fields and messages displayed on the LCD are shown in a different
type−font, e.g.
Switching On
The power switch is located at the bottom left of the front panel.
At power up the generator displays the installed software revision whilst loading its waveform
RAM; if an error is encountered the message
firmware updated
Loading takes a few seconds, after which the status screen is displayed, showing the generator
parameters set to their default values, with the MAIN OUT outputs set off. Refer to the System
Operations section for how to change the power up settings to either those at power down or to
any one of the stored settings. Recall the status screen at any time with the STATUS key; a
second press returns the display to the previous screen.
General
STANDARD WAVEFORMS, sine.
system ram error, battery fault or
will be displayed, see the Warnings and Error Messages section.
On multi-channel instruments the status shown is that of the channel selected by the SETUP
keys; this is the channel currently enabled for editing and is always the last channel selected
whether power has been switched off or not. Change the basic generator parameters for the
selected channel as described in the Standard Waveform Operation section and switch the output
on with the MAIN OUT key; the ON lamp will light to show that the output is on.
Display Contrast
All parameter settings are displayed on the 20 character x 4 row backlit liquid crystal display
(LCD). The contrast may vary a little with changes of ambient temperature or viewing angle but
can be optimised for a particular environment by using the front panel contrast control. Insert a
small screwdriver or trimmer tool through the adjustment aperture marked LCD and rotate the
control for optimum contrast.
Keyboard
Pressing the front panel keys displays screens which list parameters or choices relative to the key
pressed. Selections are then made using the display soft−keys and numeric values are changed
using the numeric keys or rotary control, see the Principles of Editing section.
The keys are grouped as follows:
• WAVE SELECT keys call screens from which all standard or already defined arbitrary
waveforms can be selected.
• WAVE EDIT keys call screens from which arbitrary waveforms can be created and modified.
• FREQuency, AMPLitude, OFFSET and MODE keys display screens which permit their
respective parameters to be edited either from the numeric keypad or using the rotary
control/cursor keys.
• Numeric keys permit direct entry of a value for the parameter currently selected. Values are
accepted in three formats: integer (20), floating point (20·0) and exponential (2 EXP 1). For
example, to set a new frequency of 50kHz press FREQ followed by 50000 ENTER or
5 EXP 4 ENTER. ENTER confirms the numeric entry and changes the generator setting to
the new value.
CE (Clear Entry) undoes a numeric entry digit by digit. ESCAPE returns a setting being edited
to its last value.
21
• MODULATION, SUM, TRIG IN and SYNC OUT call screens from which the parameters of
those input/outputs can be set, including whether the port is on or off.
• SWEEP and SEQUENCE similarly call screens from which all parameters can be set and the
functions run.
• Each channel has a key which directly switches the MAIN OUT of that channel on and off.
• MAN TRIG is used for manual triggering (when TRIG IN is appropriately set) and for
synchronising two or more generators when suitably connected together. MAN HOLD is used
to manually pause arbitrary waveform output and sweep; the output is held at the level it was
at when MAN HOLD was pressed.
• UTILITY gives access to menus for a variety of functions such as remote control interface
set−up, power−up parameters, error message settings and store/recall settings to/from
memory card; the STORE and RECALL keys can also be used to directly access the memory
card settings files.
• The INTER CHannel and COPY CHannel keys (multi-channel instruments only) directly call
screens from which channel-to-channel synchronisation and set-up copying can be set.
• The SETUP keys (multi-channel instruments only) select the channel to be edited; the lamp
lights beside the channel currently enabled for editing.
• Eight soft−keys around the display are used to directly set or select parameters from the
currently displayed menu; their operation is described in more detail in the next section.
• The STATUS key always returns the display to the default start−up screen which gives an
overview of the generators status. Pressing STATUS again returns the display to the previous
screen.
Further explanations will be found in the detailed descriptions of the generator’s operation.
Principles of Editing
Each screen called up by pressing a front panel key shows parameter value(s) and/or a list of
choices. Parameter values can be edited by using the ROTARY CONTROL in combination with
the left and right arrowed CURSOR keys, or by direct numeric keyboard entry; choices are made
using the soft−key associated with the screen item to be selected. The examples which follow
assume factory default settings.
The channel to be edited must first be selected by pressing the appropriate SETUP key;
the lamp lights beside the SETUP key of the channel currently enabled for editing.
A diamond beside a screen item indicates that it is selectable; hollow diamonds identify
deselected items and filled diamonds denote selected items. For example, press MODE to get
the screen shown below:
MODE:
♦continuous
◊gated setup…◊
◊triggered setup…◊
22
The filled diamond indicates that the selected mode is continuous. Gated or
Triggered modes are selected by pressing the associated soft−key which will make the
diamond beside that item filled and the diamond beside
illustrates how an ellipsis (three dots following the screen text) indicates that a further screen
follows when that item is selected. In the case of the MODE screen illustrated, pressing the
continuous hollow. This screen also
setup... soft−key on the bottom line brings up the TRIGGER SETUP menu; note that
selecting this item does not change the
continuous/gated/triggered selection.
Some screen items are marked with a double−headed arrow (a split diamond) when selected to
indicate that the item’s setting can be changed by further presses of the soft−key, by pressing
either cursor key or by using the rotary control. For example, pressing FILTER brings up the
screen shown below.
FILTER
mode: auto
◊type: 40MHz eliptic
Repeated presses of the mode soft−key will toggle the mode between its two possible settings
auto and manual. Similarly, when type is selected, repeated presses of the type
of
soft−key (or cursor keys or use of the rotary control) will step the selection through all possible
settings of the filter type.
In addition to their use in editing items identified by a double−headed arrow as described above,
the CURSOR keys and ROTARY CONTROL operate in two other modes.
In screens with lists of items that can be selected (i.e. items marked with a diamond) the cursor
keys and rotary control are used to scroll all items through the display if the list has more than
three items; look, for example at the STD (standard waveform) and UTILITY screens.
In screens where a parameter with a numeric value is displayed the cursor keys move the edit
cursor (a flashing underline) through the numeric field and the rotary control will increment or
decrement the value; the step size is determined by the position of the edit cursor within the
numeric field.
Thus for
STANDARD FREQUENCY set to 1.000000000 MHz rotating the control will
change the frequency in 1kHz steps. The display will auto−range up or down as the frequency is
changed, provided that autoranging permits the increment size to be maintained; this will in turn
determine the lowest or highest setting that can be achieved by turning the control. In the
example above, the lowest frequency that can be set by rotating the control is 1 kHz, shown on
the display as
1.000000000 kHz.
This is the limit because to show a lower frequency the display would need to autorange below
1kHz to
xxx.xxxxxxx Hz in which the most significant digit represents 100Hz, i.e. the
1kHz increment would be lost. If, however, the starting frequency had been set to
1.000000000 MHz, i.e. a 100 Hz increment, the display would have autoranged at 1kHz to
900.0000000 Hz and could then be decremented down to 100.0000000 Hz without
losing the 100 Hz increment.
Turning the control quickly will step numeric values in multiple increments.
Principles of Operation
The instrument operates in one of two different modes depending on the waveform selected.
DDS mode is used for sine, cosine, haversine, triangle, sinx/x and ramp waveforms. Clock
Synthesis mode is used for square, pulse, pulse train, arbitrary and sequence.
In both modes the waveform data is stored in RAM. As the RAM address is incremented the
values are output sequentially to a Digital−to−Analogue Converter (DAC) which reconstructs the
waveform as a series of voltages steps which are subsequently filtered before being passed to
the main output connector.
The main differences between DDS and Clock Synthesis modes are the way in which the
addresses are generated for the RAM and the length of the waveform data.
23
Clock Synthesis Mode
In Clock Synthesis mode the addresses are always sequential (an increment of one) and the
clock rate is adjusted by the user in the range 100MHz to 0·1Hz. The frequency of the waveform
is clock frequency ÷ waveform length, thus allowing short waveforms to be played out at higher
repetition rates than long waveforms, e.g. the maximum frequency of an 8 point waveform is
100e 6÷8 or 12·5 MHz but a 1000 point waveform has a maximum frequency of 100e6÷1000
100kHz.
Arbitrary waveforms have a user defined length of 8 to 1048576 points. Squarewaves use a fixed
length of 2 points and pulse and pulse train have their length defined by the user selected period
value.
DDS Mode
In DDS mode (Direct Digital Synthesis) all waveforms are stored in RAM as 4096 points. The
frequency of the output waveform is determined by the rate at which the RAM addresses are
changed. The address changes are generated as follows:
or
The RAM contains the amplitude values of all the individual points of one cycle (360º) of the
waveform; each sequential address change corresponds to a phase increment of the waveform of
360º/4096. Instead of using a counter to generate sequential RAM addresses, a phase
accumulator is used to increment the phase.
On each clock cycle the phase increment, which has been loaded into the phase increment
register by the CPU, is added to the current result in the phase accumulator; the 12 most
significant bits of the phase accumulator drive the lower 12 RAM address lines, the upper 8 RAM
address lines are held low. The output waveform frequency is now determined by the size of the
phase increment at each clock. If each increment is the same size then the output frequency is
constant; if it changes, the output frequency changes as in sweep mode.
24
The generator uses a 44 bit accumulator and a 100 MHz clock frequency; the frequency setting
resolution is 0·1 mHz.
Only the 12 most significant bits of the phase accumulator are used to address the RAM. At a
waveform frequency of F
CLK/4096 (~24·4kHz), the natural frequency, the RAM address
increments at every clock. At all frequencies below this (i.e. at smaller phase increments) one or
more addresses are output for more than one clock period because the phase increment is not
big enough to step the address at every clock. Similarly at frequencies above the natural
frequency the larger phase increment causes some addresses to be skipped, giving the effect of
the stored waveform being sampled; different points will be sampled on successive cycles of the
waveform.
Standard Waveform Operation
This section deals with the use of the instrument as a standard function generator, i.e. generating
sine, square, triangle, dc, ramp, haversine, cosine, havercosine and sinx/x waveforms. All but
squarewave are generated by DDS which gives 10−digit frequency resolution; squarewave is
generated by Clock Synthesis which results in only 8−digit frequency resolution. Refer to
Principles of Operation in the previous section for a fuller explanation of the differences involved.
The
STANDARD WAVEFORMS screen also includes arbitrary and sequence for simplicity of
switching between these and standard waveforms; they do, however, have their own screens
(accessed by pressing ARB and SEQUENCE respectively) and are described in detail in their
appropriate sections. Pulse and pulse−train are also accessed from the ‘standard waveforms’
screen but are sufficiently different to justify their own section in the manual.
Much of the following descriptions of amplitude and offset control, as well as of Mode, Sweep,
etc., in following sections, apply to arbitrary and sequence as well as standard waveforms; for
clarity, any differences of operation with arbitrary, sequence, pulse and pulse−train are described
only in those sections.
Setting Generator Parameters
Waveform Selection
Pressing the STD key gives the STANDARD WAVEFORMS screen which lists all the
waveforms available; the rotary control or cursor keys can be used to scroll the full list back and
forward through the display. The currently selected waveform (sine with the factory defaults
setting) is indicated by the filled diamond; the selection is changed by pressing the soft−key
beside the required waveform.
Frequency
Pressing the FREQ key gives the STANDARD FREQUENCY screen. With freq selected as
shown above, the frequency can be entered directly from the keyboard in integer, floating point or
exponential format, e.g. 12·34 kHz can be entered as 12340, 12340·00, or 1·234 exp 4 etc.
However, the display will always show the entry in the most appropriate engineering units, in this
case 12·34000000 kHz.
STANDARD WAVEFORMS
♦sine
◊square
◊triangle
STANDARD FREQUENCY
10·00000000 kHz
♦freq period◊
With
period selected instead of freq the frequency can be set in terms of a period, e.g.
123·4µs can be entered as ·0001234 or 123·4e−6; again the display will always show the entry in
the most appropriate engineering units.
Squarewave, generated by Clock Synthesis has 8−digit resolution for both frequency and period
entry.
Turning the rotary control will increment or decrement the numeric value in steps determined by
the position of the edit cursor (flashing underline); the cursor is moved with the left and right
arrowed cursor keys.
Note that the upper frequency limits vary for the different waveform types; refer to the
Specifications section for details.
25
Amplitude
Pressing the AMPL key gives the AMPLITUDE screen.
The waveform amplitude can be set in terms of peak−to−peak Volts (Vpp), r.m.s. Volts (Vrms) or
dBm (referenced to a 50Ω or 600Ω load). For Vpp and Vrms the level can be set assuming that
the output is open−circuit (
is selected termination is always assumed and the
changed to
displayed amplitude values for 600Ω termination take this into account.
With the appropriate form of the amplitude selected (indicated by the filled diamond) the
amplitude can be entered directly from the keyboard in integer, floating point or exponential
format, e.g. 250mV can be entered as ·250 or 250 exp −3, etc., However, the display will always
show the entry in the most appropriate engineering units, in this case 250mV.
Turning the rotary control will increment or decrement the numeric value in steps determined by
the position of the edit cursor (flashing underline); the cursor is moved with the left and right
arrowed cursor keys.
AMPLITUDE:
+20·0 Vpp
♦Vpp Vrms◊
◊dBm load:hiZ◊
load:hiZ) or terminated (load:50Ω or load:600Ω); when dBm
load:hiZsetting is automatically
load:50Ω. Note that the actual generator output impedance is always 50Ω; the
Alternate presses of the ± key will invert the MAIN OUT output; if DC OFFSET is non−zero, the
signal is inverted about the same offset. The exception to this is if the amplitude is specified in
dBm; since low level signals are specified in −dBm (0dBm = 1mW into 50Ω = 224mVrms) the
− sign is interpreted as part of a new amplitude entry and not as a command to invert the signal.
Note that for DC, sinx/x, pulse, pulse train, arbitrary and sequence amplitude can only be
displayed and entered in the Vpp form; further limitations on pulse, pulse−train, arbitrary and
sequence amplitude are discussed in the appropriate sections.
DC Offset
Pressing the OFFSET key gives the DC OFFSET screen. The offset can be entered directly
from the keyboard in integer, floating point or exponential format, e.g. 100mV can be entered as
·1 or 100 exp −3 etc. However, the display will always show the entry in the most appropriate
engineering units, in this case 100mV. During a new offset entry the ± key can be used at any
time to set the offset negative; alternate presses toggle the sign between + and −.
Turning the rotary control will increment or decrement the numeric value in steps determined by
the position of the edit cursor (flashing underline); the cursor is moved by the left and right
arrowed cursor keys. Because DC offset can have negative values, the rotary control can take the
value below zero; although the display may autorange to a higher resolution if a step takes the
value close to zero, the increment size is maintained correctly as the offset is stepped negative.
For example, if the display shows
DC OFFSET:
program +0·00 mVdc
(actual +0·00 mVdc)
load:hiZ◊
26
program = +2
with the cursor in the most significant digit, the rotary control will decrement the offset in 100mV
steps as follows:
program = +2
program = +1
05· mVdc
05· mVdc
05· mVdc
program = +5·00 mVdc
program = −9
program = −1
The actual DC offset at the MAIN OUT socket is attenuated by the fixed−step output attenuator
when this is in use. Since it is not obvious when the signal is being attenuated the actual offset is
shown in brackets as a non−editable field below the programmed value.
For example, if the amplitude is set to 2·5Vpp the output is not attenuated by the fixed attenuator
and the actual DC offset (in brackets) is the same as that set. The
DC OFFSET:
program +1.50 Vdc
(actual +1.50 Vdc)
load: hiZ◊
If the amplitude is now reduced to 250mVpp which introduces the attenuator, the actual DC offset
changes by the appropriate factor:
DC OFFSET:
program +1.50 Vdc
(actual +151. mVdc)
load: hiZ◊
The above display shows that the set DC offset is +1·50V but the actual offset is +151mV. Note
that the actual offset value also takes into account the true attenuation provided by the fixed
attenuator, using the values determined during the calibration procedure. In the example
displayed the output signal is 250mVpp exactly and takes account of the small error in the
fixed attenuator; the offset is 151.mV exactly, taking account of the effect of the known
attenuation (slightly less than the nominal) on the set offset of 1·50V.
5·0 mVdc
95· mVdc
DC OFFSET display shows:
Whenever the set DC offset is modified by a change in output level in this way a warning
message that this has happened will be displayed. Similarly, because the DC offset plus signal
peak is limited to ± 10V to avoid waveform clipping, a warning message will be displayed if this
condition is set. This is explained more fully in the Warnings and Error Messages section.
The output attenuation is controlled intelligently to minimise the difference between the
programmed and actual offset when the combination of programmed amplitude and offset allows
this. Thus when the offset is set to 150mV, for example, the amplitude can be reduced to
nominally 50mVpp before the fixed attenuator causes the actual offset to be different from the
programmed value.
Warnings and Error Messages
Two classes of message are displayed on the screen when an illegal combination of parameters
is attempted.
WARNING messages are shown when the entered setting causes some change which the user
might not necessarily expect. Examples are:
1. Changing the amplitude from, for example, 2·5 Volts pk−pk to 25mV pk−pk brings in the
step attenuator; if a non−zero offset has been set then this will now be attenuated too. The
message
the screen but the setting will be accepted; in this case the actual, attenuated, offset will
be shown in brackets below the set value.
DC offset changed by amplitude will be shown temporarily on
27
2. With the output level set to 10V pk−pk, increasing the DC offset beyond ± 5V will cause
the message
change will be accepted (producing a clipped waveform) and the user may then choose to
change the output level or the offset to produce a signal which is not clipped.
Offset + Sum + level may cause clipping. The offset
(clip?) will show in the display beside AMPLITUDEorDC OFFSET while the
clipped condition exists.
ERROR messages are shown when an illegal setting is attempted, most generally a number
outside the range of values permitted. In this case the entry is rejected and the parameter setting
is left unchanged. Examples are:
1. Entering a frequency of 1MHz for a triangle waveform. The error message:
Frequency out of range for the selected waveform is shown.
2. Entering an amplitude of 25Vpp. The error message:
Maximum output level exceeded is shown.
3. Entering a DC offset of 20V. The error message:
Maximum DC offset value exceeded is shown.
The messages are shown on the display for approximately two seconds. The last two messages
can be viewed again by pressing the
Memory Card and System Operations section.
last error... soft−key on the UTILITY screen, see
Each message has a number and the full list appears in Appendix 1.
The default set−up is for all warning and error messages to be displayed and for a beep to sound
with each message. This set−up can be changed on the
screen. The
Each feature can be turned ON and OFF with alternate presses of the associated soft−key; the
factory default is for all features to be ON.
SYNC Output
SYNC OUT is a multifunction CMOS/TTL level output that can be automatically or manually set to
be any of the following:
• waveform sync : A square wave with 50% duty cycle at the main waveform
frequency, or a pulse coincident with the first few points of an
arbitrary waveform. Can be selected for all waveforms.
• position marker : If an arbitrary waveform is selected, any point(s) on that
waveform may have associated marker bit(s) set high or low.
• burst done : Produces a pulse coincident with the last cycle of the burst.
errormenu is shown below:
error beep: ON
◊
◊ error message: ON
warn beep: ON
◊ warn message: ON
These will then show as pulses when
selected.
error... menu on the UTILITY
position markeris
28
• sequence sync : Produces a pulse coincident with the end of a waveform sequence.
• trigger : Selects the current trigger signal (internal, external or manual).
Useful for synchronising burst or gated signals.
• sweep sync : Outputs the sweep trigger and sweep marker signals.
• phase lock : Used to synchronise two or more generators. Produces a positive
edge at the 0º phase point.
The setting up of the signals themselves is discussed in the relevant sections later in this manual,
e.g.
trigger is described in the Triggered Burst/Gate section and position marker
under the Arbitrary Waveform Generation.
Pressing the SYNC OUT key calls the
SYNC OUT:
output: on
◊ mode: auto
src: waveform sync
SYNC OUT is turned on and off by alternate presses of the output soft−key.
The selection of the signal to be output from the SYNC OUT socket is made using the
(source) soft−key; repeated presses of src cycle the selection through all the choices
(
waveform sync, position marker, etc.) listed above. Alternatively, with the src
selected (double−headed arrow) the rotary control or cursor keys can be used to step backwards
and forwards through the choices.
The source selection of the SYNC OUT waveform can be made automatic (
user−defined (
SYNC OUT waveform most appropriate for the current main waveform is selected.
For example,
arbitrary waveforms, but
automatic selection will be mentioned in each of the appropriate main waveform mode sections
and a full table is given in Appendix 2.
The automatic selection can still be changed manually by the
mode has been selected but the selection will immediately revert to the automatic choice as soon
as any relevant parameter (e.g. main waveform frequency or amplitude) is adjusted.
must be selected by the mode soft−key for a source other than the automatic choice to remain
set. The
sync
requirements, e.g. position markers on arbitrary waveforms.
for all continuous main waveforms, but manual will need to be used for special
manual) with alternate presses of the mode soft−key. In automatic mode the
waveform sync is automatically selected for all continuous standard and
trigger is selected in trigger or gated waveform modes. The
auto selection will generally set the most frequently used signal, e.g. waveform
SYNC OUT setup screen.
src
auto) or
src soft−key even when auto
Manual
29
Loading...
+ 80 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.