Introduction 2
Specification 3
EMC 7
Safety 8
Installation 9
Operation 10
Calibration and Maintenance 13
Instructions en Francais
Sécurité 14
Installation 15
Fonctionnement 16
Calibrage et maintenance 20
Bedienungsanleitung auf Deutsch
Sicherheit 21
Installation 22
Betrieb 23
Kalibrierung und Wartung 27
Istruzioni in Italiano
Sicurezza 28
Installazione 29
Funzionamento 30
Calibrazione e Manutenzione 34
Instrucciones en Español
Seguridad 35
Instalación 36
Funcionamiento 37
Calibrado y Mantenimiento 41
1
Introduction
The TG550 is a precision 5MHz function generator with waveforms of sine, t r iangle, ramp. pulse
and DC from a variable amplitude 50Ω or 600Ω output. A TTL/CMOS auxiliary output is also
provided.
The frequency range is 0· 005Hz to 5MHz selected by a seven decade range multiplier and
calibrated vernier. Both the frequency vernier and the sweep input can give 1000:1 frequency
change within a selected range. Frequency can be locked t o a c r ystal reference to give superior
frequency stability. A 10:1 symmetry range per mits ramp and pulse waveforms to be produced.
The generator output level is 20Vpk-pk maximum from a 50/600Ω source. The level is set via a
switched attenuator plus vernier with a total range of >80dB. DC offs et is vernier adjustable over
a ±10V range with centre detent for 0V. Waveform quality is good at all frequenc ies and sig nal
levels.
The TG550 also has internal LIN/LOG sweep with sweep output. Amplitude modulation of up to
100% is possible using the internal 400Hz oscillator or an external source.
The large liquid crystal display shows frequency and amplitude simultaneously. Good resolution
and a fast display update rate are maintained at all frequencies. The display can also be used as
an external 20MHz frequency meter with up to 7-digit resolution.
Additional display annunciators indicate generator status when sweep, AM, etc. are in use.
2
coarse and fine verniers.
Vernier Range:
1000:1 on each range.
of range full scale (unspecified with Symmetry on).
displayed value ± 0·01%.
5MHz range.
Amplitude Flatness:
± 0·2dB to 200kHz; ± 2dB to 5MHz.
Linearity:
Rise and Fall Times:
<45ns.
Mark-Space Ratio:
1:1 ±1% to 100kHz.
Range:
±10V (unterminated).
frequency divided by 10 when Symmetry is selected.
External: linear.
Sweep Width:
Up to 1000:1 within each range.
Sweep Rate:
Internal sweep rate adjustable from typically 20ms to 20s.
External Sweep:
By signal applied by SWEEP IN/OUT socket.
Specification
Specifications apply at 18º - 28º after one hour warm-up, at maximum output int o 50Ω.
FREQUENCY
Frequency Range: 0·005Hz to 5MHz in 7 overlapping decade ranges with adjustment by
Frequency Display: Auto-ranging reciprocal measurement giving 4-digit resolution down to
10Hz (to 1Hz on the 5Hz range); resolution is fixed at 0·01Hz below 10Hz
(0·001Hz below 1Hz on the 5Hz range).
Accuracy is ± 1 digit 5MHz to 2·1Hz (to 0·21Hz, Symmetry off, or to
0·42Hz, Symmetry on, on the 5Hz range); below this the accuracy is ± 1%
Frequency Locking: With vernier setting between 10% and 100% of range maximum,
frequency can be locked to the cr ystal reference to maintain it at the
WAVEFORMS
Waveform specifications apply f or t he top decade of each frequency rang e.
SINE
Distortion: Less than 0·5% on 500Hz, 5kHz and 50kHz ranges; less than 1% on 5Hz,
50Hz and 500kHz ranges; all harmonics >25dB below fundamental on
TRIANGLE
Better than 99% to 200kHz.
SQUAREWAVE
DC
SYMMETRY
Symmetry Range: Variable typically between 1:9 and 9:1 (on top decade of each range),
MODULATION MODES
Modulation can be used with all waveforms and at any frequency.
Sweep
Sweep Mode: Internal: linear or logarithmic; s weep Start and Stop frequencies
displayed at a press of a button.
3
Modulation Depth:
Variable 0 to 100% typical.
Modulation Frequency:
400Hz internal. DC to 100kHz external.
External Modulation:
By signal applied to COUNT/AM IN socket.
Output Impedance:
50Ω.
range.
Attenuator:
0, –20, – 40 or – 60dB.
plus waveform attenuated proportionally by the attenuator.
Output Characteristics:
Frequency, symmetry and phase the same as main outputs.
loads.
Input Impedance:
10kΩ.
Input Sensitivity:
0 to 3V for 1000:1 sweep.
Voltage:
±10V.
Sweep Linearity:
Better than 1%
Sweep Voltage:
0·1V/µs.
Input Impedance:
40kΩ.
Input Sensitivity:
Approximately 2V peak-peak for 100% modulat ion.
Voltage:
Amplitude Modulation
OUTPUTS
50Ω Main Out
Amplitude:
DC Offset Range:
2mV to 20V peak-peak open circuit (1m V t o 10V peak-peak into 50Ω)
in four switch-selectable ranges with 20dB vernier control within each
±10V. DC offs et plus signal peak limited to ±10V (±5V into 50Ω); CLIP
shows in display when offset + signal peak exceeds ±10V. DC offset
600Ω Main Out
Alternative output socket providing the same signal as the 50Ω Main Out but from 600Ω. The
50Ω and 600Ω sockets ar e not independent .
Aux Out
Output Level: 0 to 5V TTL/CMOS logic levels capable of driving 2 standard TTL
Sweep Out
When internal sweep is selected the sweep signal is available as a 0 to 3V ramp from 600Ω.
INPUTS
Sweep In
The SWEEP IN/OUT socket is set to Sweep In when EXT SWEEP is selected.
Maximum Allowable Input
Maximum Slew Rate of
AM In
The AM/COUNT IN socket is set to AM input when EXT AM is selected.
Maximum Allowable Input
4
±10V.
Input Impedance:
1MΩ//25pF.
Input Sensitivity:
50mVrms (sinewave) typical; 80mVrms max.
1Vrms above 1MHz.
± 1% of range f ull scale (unspec ified with Symmetry on).
±5% of range f ull scale.
Display corrected for attenuator setting.
Frequency Range:
5Hz to 20MHz, fully autoranging.
Input Sensitivity:
50mVrms (sinewave).
Measurement Time:
Selectable 0·5s or 5s.
Resolution:
6 digits in 0·5s; 7 digits in 5s.
Accuracy:
±1 digit ± timebase accuracy.
ppm/ºC.
adjustment. Installation Category II.
Power Consumption:
25VA max.
Operating Range:
+5°C to +40°C, 20% to 80% RH.
Storage Range:
-10°C to +65°C
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
feet.
Weight:
1.9kg (4.2lb).
Safety:
Complies with EN61010-1.
EMC:
Complies with EN61326.
Count In
The AM/COUNT IN socket is set to external freq uenc y measur em ent when EXT COUNT is
selected.
Maximum Allowable Input
Voltage:
50Vdc/50Vrms to 400Hz with respect to ground
, reducing to
DISPLAY FUNCTIONS
The LCD shows generator frequenc y simultaneously with output amplitude/offset, together with
various status annunciators. Alternatively, it is the external freq uency measur em ent display.
Internal Measurement Accur acy
Frequency: Auto-ranging reciprocal measurement giving 4-digit resolution down
to 10Hz (to 1Hz on the 5Hz range); resolution is fixed at 0·01Hz
below 10Hz (0·001Hz below 1Hz on the 5Hz range).
Accuracy is ± 1 digit 5MHz to 2·1Hz (to 0·21Hz, Symmetry off, or to
0·42Hz, Symmetry on, on the 5Hz range); below this the accuracy is
Amplitude: Display shows peak-to-peak amplitude or rms value. Display
corrected for attenuator setting. 3-digit res olution, ac curacy typically
DC Offset: 3-digit resolution; accuracy typically ±2% setting ± 1 dig it .
External Frequency Measurement
Timebase Accuracy: ±10ppm initial error; ± 5 ppm/year ag eing r ate; typically less than 0·5
GENERAL
POWER REQUIREMENTS
AC Input Voltage: 220V AC-240V AC or 110V AC-120V AC ±10%, 50/60Hz, by internal
Size: 260 (W) x 88 (H) x 235 (D) mm (10.2 x 3.4 x 9.2") excluding handle and
5
Performance levels achieved are detailed in the user manual.
EC Declaration of Conformity
We Thurlby Thandar Instruments Ltd
Glebe Road
Huntingdon
Cambridgeshire PE29 7DR
England
declare that the
TG550 Precision 5MHz Function Generator
meets the intent of the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
Compliance was demonstrated by conformance to the following specifications which have been
listed in the Official Journal of the European Comm unit ies.
EMC
Emissions: a) EN61326-1 (2006) Radiated, Class B
b) EN61326-1 (2006) Conducted, Class B
c) EN61326-1 (2006) Harmonics, referring to EN61000-3-2 (2006)
Immunity: EN61326-1 (2006) Immunity Table 1, referring to:
a) EN61000-4-2 (2009) Electrostatic Discharge
b) EN61000-4-3 (2006) Electromagnetic Field
c) EN61000-4-11 (2004) Voltage Interr upt
d) EN61000-4-4 (2004) Fast Transient
e) EN61000-4-5 (2006) Surge
f) EN61000-4-6 (2009) Conducted RF
Safety
EN61010-1 Installation Category II, Pollution Degree 2.
CHRIS W ILDING
TECHNICAL DIRECTOR
2 January 2013
6
This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC.
Compliance was demonstrated by meeting the test limits of t he following standards:
Emissions
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use. Test limits used were:
a) Radiated: Class B
b) Conducted: Class B
c) Harmonics: EN61000-3-2 (2006) Class A; the instrument is Class A by product category.
Immunity
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use.
Test methods, limits and per formance achieved are shown below (requirement shown in
brackets):
a) EN61000-4-2 (2009) Electros tatic Dischar ge : 4kV air, 4kV contact, Performance A (B) .
EMC
b) EN61000-4-3 (2006) Electr omagnetic Field:
3V/m, 80% AM at 1kHz, 80MHz – 1GHz: Performance B
Performance A (A) ; 1V/ m , 2.0GHz to 2.7GHz: Performance A (A).
c) EN61000-4-11 (2004) Voltage Interrupt: ½ cycle and 1 cycle, 0%: Per formance A (B);
25 cycles, 70% and 250 cycles, 0%: Performance B (C).
d) EN61000-4-4 (2004) Fast Transient, 1kV peak ( AC line), 0·5kV peak (signal connections),
Performance A (B).
e) EN61000-4-5 (2006) Surg e, 0·5kV (line to line), 1kV (line to gr ound), Performance A ( B) .
f) EN61000-4-6 (2009) Conducted RF, 3V, 80% AM at 1kHz (AC line only; signal
connections <3m, therefor e not tested), Performance B
Note: If the generat or is subjected to sufficiently large m odulated RF fields (radiated or
conducted) some minor modulation of t he out put waveform may be visible at certain frequencies
up to about 100MHz. However, possible degradation will be small and infrequent and not
deemed to be a problem in practice. In all other res pect s t he inst rument will operate normally
(Performance A) in fields up t o 3V/ m .
According to EN61326-1 the definitions of performance criteria are:
Performance criterion A: ‘During test normal performance within the specification limits.’
Performance criterion B: ‘During t est , temporary degradation, or loss of function or
performance which is self-recovering’.
(A) and 1.4GHz to 2GHz:
(A).
Performance criterion C: ‘During t est, temporary degradation, or loss of function or
performance which requires operator int er vention or s ystem r eset occurs.’
Cautions
To ensure continued compliance with the EMC directive observe the following precautions:
a) Connect t he generator to other equipment using only high quality, double−screened cables.
b) Aft er opening the case for any reason ensure that all signal and g round connections are
remade correctly.
c) In the event of part replacement becoming nec essary, only use components of an ident ical
type, see the Service Manual.
7
may damage the instrument.
terminal connected to chassis ground.
alternating current.
Safety
This instrument is Safety Class I accor ding to IEC classification and has been designed to meet
the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use). It is an Installation Category II instrument intended for oper ation
from a normal single phase supply.
This instrument has been tested in accordanc e with EN61010-1 and has been supplied in a safe
condition. This instruction manual contains some information and warnings which have to be
followed by the user to ensure saf e oper at ion and t o r etain the inst rument in a safe condition.
This instrument has been designed f or indoor use in a Pollution Degree 2 environment in the
temperature range 5°C to 40°C, 20% - 80% RH (non-condensing). It may occasionally be
subjected to temperatures between +5°C and –10°C without degradation of its safety. Do not
operate while condensation is present.
Use of this instrument in a manner not spec ified by these instructions may impair the safety
protection provided. Do not operate the instrum ent outside its rat ed supply voltages or
environmental range.
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conduct or inside or outside t he inst r um ent will make the
instrument dangerous. Int entional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
When the instrument is connected to its supply, terminals may be live and opening the covers or
removal of parts (except those to which access can be gained by hand) is likely to expose live
parts. The apparatus shall be disconnected from all voltage sources before it is opened for any
adjustment, replacement, m aint enance or repair.
Any adjustment, maintenance and repair of the opened instrument under voltage shall be
avoided as far as possible and, if inevitable, shall be carried out only by a skilled person who is
aware of the hazard involved.
If the instrument is clearly def ec t ive, has been subject to mechanical damage, excessive
moisture or chemical corrosion the safety protection may be impaired and the apparatus should
be withdrawn from use and returned for checking and repair.
Make sure that only fuses with the required rated current and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuse holders is prohibited.
Do not wet the instrument when cleaning it.
The following symbols are used on the instrument and in this m anual:-
Caution - refer to the accompanying documentation, incorrect operation
8
Mains Operating Voltage
The operating voltage of the instr um ent is shown on the rear panel. Should it be necessary to
change the operating voltage from 230V t o 115V or vice-versa, proceed as follows:
1. Disconnect the instrument fr om all voltage sources.
2. Remove the 4 screws which hold the upper and lower case halves together and lift off the
case lower.
3. Remove the 6 screws securing the main pcb to the case upper and lift the pcb free,
complete with front and rear panels.
4. Change the appropriate zero-ohm links beside the transformer on the pcb:
Link LK2 only for 230V operation
Link LK1 and LK3 only for 115V operation
5. Refit the pcb to the case upper, ensuring all connections (especially safety earth) ar e
remade as before, and refit the case lower.
6. To comply with safety standard requirem ents t he oper ating voltage marked on the rear
panel must be changed to clearly show the new voltage setting.
Installation
7. Change the fuse to suit the new operating voltage, see below.
Fuse
The correct time-lag fuse must be fitted for the selected operating voltage.
For 230V operation use 125mA (T) 250V HBC.
For 115V oper at ion use 250mA (T) 250V HBC.
Make sure that only fuses with the required rated cur rent and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuse holders are prohibited.
Mains Lead
When a three c or e m ains lead with bare ends is pr ovided it should be connected as follows:
Brown - Mains live
Blue - Mains Neutral
Green/Yellow - Earth
Any interruption of the mains earth conduct or inside or outside t he inst r um ent will make the
instrument dangerous. Int entional interruption is prohibited.
WARNING! THIS INSTRUMENT MUST BE EARTHED
9
Switching On
Switch on the generator using the ON/O FF switch on the r ear panel. To fully disconnect from the
AC supply unplug the mains cord from the back of the instrument or switch off at the AC supply
outlet; make sure that the means of disconnection is readily accessible. Disconnect from the AC
supply when not in use.
Ensure that the push-button switches in the SWEEP and AM/EX TERNAL COUNTER sections of
the control panel are all set to the position, i.e. set t he generator to its normal,
unmodulated, running mode. The display will show the frequency and peak-to-peak out put
amplitude with appropriate range annunciators.
Frequency
Frequency Setting
Frequency is set by pressing the appropriate FREQUENCY RANGE push button and rotating the
calibrated FREQUENCY vernier; precise adjustments can be made using the FINE control. The
vernier controls provide a 1000:1 frequency change within each rang e. The auto-ranging meter
makes reciprocal freq uenc y measur em ents down to 2·1Hz (to 0· 21Hz on the 5Hz range) ; the
display update rate in this mode is 130ms or 1 waveform period, whichever is the longer. Below
these frequencies a f ast measurement update rate is maintained by changing the measurement
method but accuracy is reduced, see Specification.
Operation
When SYMMETRY is switched ON the SYMMETRY control can be used to vary the duty cycle
from 1:10 to 10:1 to pr oduce s awtooth and variable pulse-width waveforms. Optimum
performance of the SYMMETRY cont r ol is achieved on the top decade of each frequency range.
When SYMMETRY is selected SYM shows in the display and the generator frequency is divided
by 10; the display shows the correct frequency. The changeover in frequency measurement
method now happens at 0·21Hz on the 50Hz range and 0·42Hz on the 5Hz range.
Frequency Locking
The set frequency can be maintained precisely by using the FREQUENCY LOCK feature. Press
the LOCK button once to light the READY lamp. With READY lit set the required frequency in the
usual way, ensuring that the FREQUENCY RANGE has been selected such that the
FREQUENCY vernier control is within its top decade, i.e. between 0·5 and 5.
Press LOCK a second time to activate frequency locking; the LOCKED lamp will light to show
that frequency locking is oper at ional. Fr equency locking is disabled by pressing LOCK again (the
LOCKED lamp goes off) or by changing the frequency using the vernier or fr equency range
switches.
Frequency Locking employs a ‘measure-and-cor r ect’ technique; at low frequencies the time
between measurements is long (2s at 0·5Hz) and frequency stability may therefore be poorer.
Warning is given if fr equency lock is about to be lost by flashing the LOCKED lamp; t his m ay
occur with extreme variations of time and temperat ur e which may cause the natural frequency of
the generator to drift outside the frequency locking control range.
Waveform
The output waveform shape is selected by depressing one of the three function buttons t o give
sine, square or triangle. With all three switches out (acc om plished by half -depressing any one)
the output will be a DC level only; this is useful as it permits input threshold testing of a cir cuit
without having to connect up an external DC supply.
10
Amplitude
The amplitude of the MAIN OUT outputs is set by the AMPLITUDE control and the
ATTENUATOR switches. Maximum output is 20 volts peak-to-peak open-cir cuit , 10 volts peak-topeak when terminated with the outputs characteristic impedance.
The AMPLITUDE control has greater t han 20dB range; the ATTENUATOR switches permit
selection of up to -60dB attenuation (bot h s witches press ed t ogether). Used together t hes e
controls provide a range of 20V peak-to-peak down to 2mV peak-to-peak, or 10V peak-to-peak
down to 1mV peak-to-peak into the output’s characteristic impedanc e.
Still greater attenuation of the 50Ω MAIN OUT can be achieved by using standard 50Ω BNC
attenuators. To maintain waveform integrity only 50Ω cable should be used and the receiving end
should be terminated with a 50Ω load.
The MAIN OUT outputs will withstand a short circuit for a period of 10 minutes at maximum
output and greater periods at lower output levels. However, damage will occur if an external
voltage is connected to the output sockets.
The 50Ω and 600Ω outputs ar e not independent ; terminating one will affect the amplitude of the
other.
The amplitude is displayed on the right-hand side of t he LCD; pressing the PK-PK/RMS button
beside the display will alternate the reading between peak-to-peak volts (Vp-p) and r.m.s. volts
(Vrms). The readout is correctly adjusted for att enuator setting and waveform shape.
The display shows open circuit voltage; the actual voltage at the socket will be half the displayed
value when the output is terminated with its characteristic impedance.
DC Offset
The DC OFFSET control has a range of ±10 volts fr om 50Ω/600Ω in all output modes; the c ont r ol
has a centre detent for 0 volts. DC offset plus signal peak is limited to ±10V (±5V into the
characteristic output impedance); CLI P shows in the display when this limit is exceeded.
DC offset plus waveform is attenuated proportionally by the ATTENUATOR switches.
The DC offset can be displayed in place of the output amplitude by pressing the OFFSET button
beside the display. The display shows open circuit voltage offset; the actual voltage at the socket
will be half the displayed value when the output is terminated with its characteristic impedance.
Auxiliary Output
The AUX OUT output provides a fixed 0 to 5V TTL/CMOS pulse output at the same frequency,
symmetry and phase as the MAIN OUT outputs and is capable of driving 2 standard TTL loads.
Internal Sweep
With INT SWEEP selected the output of the internal sweep generat or is summed with the
FREQUENCY vernier. Select LIN or LO G s weep mode; t he annunc iators LIN-SWP or SWP-LOG
will show in the display accordingly.
Frequency Selection
Using the FREQUENCY RANGE switches select the lowest range within which the upper sweep
limit can be set; in this way the best setting resolution and the widest sweep range can be
realised.
Sweep Limits
Turn the STOP control to minimum (fully-anticlockwise). Hold the SET START button and set the
START (lowest) freq uency on the display using the FREQ UENCY vernier; the START frequency
must be set first. Now release the SET START button, hold in the SET STOP button and set the
STOP (highest) frequency on the display using the STOP control; release the SET STOP button.
Note that both the FREQUENCY vernier and the STOP control have a log characteristic when
LOG mode is selected.
The sweep limits can be checked at any time by depressing the SET START or SET STOP.
11
Sweep Rate
The generator sweeps both up and down at the speed set by the RATE control, i.e. it does not
reset rapidly after the upward sweep. This has the advantage that the suitability of the selected
sweep time can be continuously monitored.
Sweep Out
When INT SWEEP is selected the SWEEP IN/OUT socket becomes an out put providing a 0V to
3V ramp from 600Ω im pedance to drive the X-input of an oscilloscope or chart r ecor der.
Terminating this output with a resistor will reduce the output amplitude. Terminating with 600Ω will
halve the output giving 0V to 3V.
Holding in the SET STOP button holds the ramp at the maximum and can therefore be used t o
set full scale on the oscilloscope or chart recor der.
Note that when SWEEP IN/OUT is used as an output to drive the X -input of an oscilloscope, the
oscilloscope should be set to DC coupling and not AC coupling which will cause a double image
at slow sweep speeds.
External Sweep
When EXT SWEEP is selected the SWEEP IN/OUT socket becomes an input. The generator
frequency can be swept, DC programmed or modulated by a suitable control voltage applied to
this input. The instrument sums the SWEEP IN voltage with the internal control voltage derived
from the FREQUENCY vernier to determine the operating frequency; the display shows the
resultant frequency.
A positive voltage increases the frequency; f or frequency control with positive-going DC inputs
the vernier should therefore be set to t he lower frequency limit of the range t o be swept. For
example, a 0V to +3V signal will sweep the generator 3 decades up from range minimum, set by
the vernier, to range maximum.
Similarly, a negative voltage decreases the frequency and for negative-going DC inputs the
vernier should be set to the upper freq uency limit of the range to be swept. For example, a 0V to
-3V signal will sweep the generator 3 decades down from range maximum, s et by t he vernier, to
range minimum.
To use a sweep signal which is symmetrical about ground, t he vernier should be set to give a
frequency at approximately the centre of the band to be swept.
In external sweep mode the LIN/LOG selection is inoperative, i.e. t he FRE QUENCY vernier is
always operating linearly, as marked.
Note: Non-linear operation may result when the sweep input voltage is excessive; that is, when
the attempted generator frequency exceeds the range limits.
Amplitude Modulation
Depressing the AM ON/OFF button selects AM. The depth of m odulat ion can be adj usted over a
0% to 100% range using the modulation DEPTH cont rol. When AM is selected the output
amplitude will drop to 50% at 0% modulation.
With the AM INT/EXT button in the INT (internal) position, the modulation sour c e is an inter nal
400Hz sinewave oscillator. The INT-AM annunciator is displayed.
12
With the AM INT/EXT button in the EXT (external) position, and the EXT COUNT/EX T AM button
in the EXT AM position, exter nal AM modulation is possible via the AM/COUNT IN socket; the
AM-EXT annunciator is displayed. The modulating signal applied should have no DC offset or
should be AC coupled. A 2V peak-to-peak signal gives approximately 100% modulation with the
modulation control at maximum. Modulating the generator with a squarewave gives step changes
in the output amplitude which are suitable for testing signal compressors and automatic gain
controlled circuits.
Loading...
+ 30 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.