Thurlby Thandar Instruments User Manual [en, de, fr, it]
AC1000
Low Distortion Power Source
INSTRUCTION MANUAL
Introduction
The AC1000 is an innovative, low cost, pure power source designed specifically for use with a harmonics
analyser such as the TTi HA1600, permitting compliance quality measurements to EN61000–3–2.
The AC supply available at a standard wall socket is usually distorted. This comes about because of
nonlinear loads (non–resistive loads) on the AC supply such as transformers, fluorescent lights,
switched–mode power supplies etc. In a typical load there are rectifiers and capacitors that cause current
to be drawn from the supply only near the peak of the waveform. The effect of this is to flatten off the top
of the sine wave causing significant distortion.
In a typical factory environment this distortion can be so significant that it is easily visible using an
oscilloscope.
EMC legislation (EN61000–3–2) has made mandatory the measurement of harmonic current produced
by electronic equipment, and requires that the source supply provides a pure voltage waveform to the
equipment under test. The harmonic currents can differ significantly when a pure source is used so it is
important that any testing is performed with a device such as the AC1000.
Compact and portable the AC 1000 is rated up to 1000 V·A for 230 V operation and will supply 1000 V·A
at 230 V input at up to 35°C ambient. In contrast to other techniques which use a passive filtering
approach, the AC1000 uses an innovative technique which actively corrects the AC supply to produce a
harmonically pure output.
The AC1000 is also very useful for stress testing power supplies in electronic equipment. The use of a
harmonically pure, distortion free device – such as the AC1000 – is essential to fully stress and evaluate
the power supply under test.
Table of Contents
Specification 2
Safety 4
EMC 5
Installation 6
Operation 7
Maintenance 8
Instructions en Francais 9
Bedienungsanleitung auf Deutsch 13
Istruzioni in Italiano 17
Instrucciones en Español 21
1
Specification
Input Voltage: Factory set to 230V, 115V or 100V; 50Hz or 60Hz.
Installation Category II.
Supply Tolerances: 230V setting : 198V – 253V
115V setting: 99V – 127V
100V setting: 90V – 110V
50Hz setting: ± 0·5Hz
60Hz setting: ± 0·6Hz
Output Voltage: Tracks the amplitude of the fundamental of the input voltage. A variable
voltage input may be used to adjust the output voltage to within the limits
specified by EN61000–3–2.
Output Distortion: Dependent on the purity of the input but will generally meet the requirements
of EN61000–3–2.
Output Current: Maximum continuous output current is 4·4A.
Output Power: Maximum output power is input voltage x 4·4 VA
Input Connection: IEC connector; front panel switch.
Output Connection: U.K., Schuko, or other national outlet sockets. Load power switch can be set
to DIRECT or CORRECTED for ‘A–B’ comparisons.
Protection: Thermal trip automatically diverts load to a DIRECT connection in the event of
thermal overload.
Operating Range:
Storage Range:
Environmental: Indoor use at altitudes to 2000m, Pollution Degree 2.
Safety: Complies with EN61010–1
EMC: Complies with EN50081-1 and EN50082-1.
Size: 307 x 105 x 225 mm (W x H x D)
Weight: 5 kg.
+5°C to +35°C at full rated output; 20% to 80% RH (non-condensing).
–40°C to +70°C
2
f) EN61000-4-6 (2009) Conducted RF
EC Declaration of Conformi ty
We Thurlby Thandar Instruments Ltd
Glebe Road
Huntingdon
Cambridgeshire PE29 7DR
England
declare that the
AC1000 Low Distortion Power Source
meets the intent of the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
Compliance was demonstrated by conformance to the following specifications which have been
listed in the Official Journal of the European Communities.
EMC
Emissions: a) EN61326-1 (2006) Radiated, Class B
b) EN61326-1 (2006) Conducted, Class B
c) EN61326-1 (2006) Har monic s, referring to EN61000-3-2 (2006)
Immunity: EN61326-1 (2006) Immunity T able 1, Performance B, referring to:
a) EN61000-4-2 (2009) Electrostatic Discharge
b) EN61000-4-3 (2006) Electromagnetic Field
c) EN61000-4-11 (2004) Voltage Interrupt
d) EN61000-4-4 (2004) Fast Transient
e) EN61000-4-5 (2006) Surge
Safety
EN61010-1 (1993) Installation Category II, Pollution Degree 2.
CHRIS WILDING
TECHNICAL DIRECTOR
2 January 2013
3
l
incorrect operation may damage the instrument.
Safety
This power supply is a Safety Class I instrument according to IEC classification and has been
designed to meet the requirements of EN61010-1 (Safety Requirements for Electrical Equipment
for Measurement, Control and Laboratory Use). It is an Installation Category II instrument
intended for operation from a normal single phase supply.
This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe
condition. This instruction manual contains some information and warnings which have to be
followed by the user to ensure safe operation and to retain the instrument in a safe condition.
This instrument has been designed for indoor use in a Pollution Degree 2 environment in the
temperature range 5°C to 35°C, 20% - 80% RH (non-condensing). It may occasionally be
subjected to temperatures between +5° and -10°C without degradation of its safety. Do not
operate while condensation is present.
Use of this instrument in a manner not specified by these instructions may impair the safety
protection provided. Do not operate the instrument outside its rated supply voltages or
environmental range. In particular excessive moisture may impair safety.
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
When the instrument is connected to its supply, terminals may be live and opening the covers or
removal of parts (except those to which access can be gained by hand) is likely to expose live
parts. The apparatus shall be disconnected from all voltage sources before it is opened for any
adjustment, replacement, maintenance or repair. Capacitors inside the power supply may still be
charged even if the power supply has been disconnected from all voltage sources but will be
safely discharged about 1 minute after switching off power.
Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided
as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of
the hazard involved.
If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture
or chemical corrosion the safety protection may be impaired and the apparatus should be
withdrawn from use and returned for checking and repair.
Make sure that only fuses with the required rated current and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuse holders is prohibited.
Do not wet the instrument when cleaning it. The following symbols are used on the instrument
and in this manual:-
mains supply OFF.
mains supply ON.
alternating current (ac)
Caution – refer to the accompanying documentation;
4
This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use. Test limits used were:
a) Radiated : Class B
b) Conducted : Class B
c) Harm onics: EN61000-3-2 (2006) Class A; the instrument is Class A by product category.
Note that emissions from the AC1000 are within specification when fully loaded with a compliant
load, e.g. a pure resistance. If the load is non-compliant, emissions from the AC1000 will similarly
be non-compliant.
Immunity
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use.
EMC
Test methods, limits and performance achieved were:
a) EN61000-4-2 (2009) Electrostatic Discharge : 4kV air, 4kV contact, Performance A.
b) EN61000-4-3 (2006) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance B.
c) EN61000-4-11 (2004) Voltage Interrupt, 1 cycle, 100%, Performance B.
d) EN61000-4-4 (2004) Fast Transient, 1kV peak (AC line), 0.5kV peak (DC Outputs),
e) EN61000-4-5 (2006) Surge, 0.5kV (line to line), 1kV (line to ground), Performance A.
f) EN61000-4-6 (2009) Conducted RF, 3V, 80% AM at 1kHz (AC line only; DC Output
According to EN61326 the definitions of performance criteria are:
Performance criterion A: ‘During test normal performance within the specification limits.’
Performance criterion B: ‘During test, temporary degradation, or loss of function or
performance which is self-recovering’.
Performance criterion C: ‘During test, temporary degradation, or loss of function or
performance which requires operator intervention or system reset occurs.’
Cautions
Performance B.
connections <3m not tested), Performance A.
To ensure continued compliance with the EMC directive the following precautions
should be observed:
a) after opening the case for any reason ensure that all signal and ground
connections are remade correctly before replacing the cover. Always ensure all
case screws are correctly refitted and tightened.
b) In the event of part replacement becoming necessary, only use components of an
identical type, see the Service Manual.
5
Mains Operating Voltage and Frequency
Check that the instrument operating voltage and frequency marked on the rear panel are correct
for the local supply; these are factory set and cannot be changed by the user.
There are 6 combinations of supply voltage and frequency: 230V/50Hz, 230V/60Hz, 115V/50Hz,
115V/60Hz, 100V/50Hz and 100V/60Hz; the Specification section gives the operating range for
each version. To operate correctly the AC1000 requires an accurate and stable input frequency.
It is not designed to operate from portable generators or battery inverters and it will not function
when powered from a source that has a high impedance at 1kHz such as an isolating transformer
with a split bobbin.
The AC1000 does not function as an uninterruptable power supply (UPS), a transient suppressor
or a voltage stablizer.
Fuse
The AC fuse is located in the fuse drawer in the lower part of the IEC inlet connector. To change
the fuse remove the line cord and open the fuse drawer with a suitable tool.
The correct fuse type for all operating voltages is 20 x 5mm 5A(T) 250V HBC time-lag.
Make sure that only fuses with the required current rating and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuseholders are prohibited.
Installation
Mains Lead
When a three core mains lead with bare ends is provided this should be connected as follows:
As the colours of the wires in the mains lead of this apparatus may not correspond with the
coloured markings identifying the terminals in your plug proceed as follows:
The wire which is coloured green-and-yellow must be connected to the terminal in the plug which
is marked by the letter E or by the safety earth symbol shown above or coloured green or greenand-yellow.
The wire which is coloured blue must be connected to the terminal which is marked with the letter
N or coloured black.
The wire which is coloured brown must be connected to the terminal which is marked with the
letter L or coloured red.
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited.
BROWN - MAINS LIVE
BLUE - MAINS NEUTRAL
GREEN/YELLOW - EARTH Safety Earth Symbol
WARNING! THIS INSTRUMENT MUST BE EARTHED.
6
Connections
With the POWER switch off ( ), connect the AC1000 to the AC supply using the mains lead
provided.
With the LOAD POWER switch set to DIRECT, connect the load to the AC socket on the front
panel; the unit should have been supplied with a socket suitable for local appliance connectors.
If the 'load' is a HA1600 Power and Harmonics Analyser, connect the fixed LOAD POWER lead of
the HA1600 to the front panel socket of the AC1000.
Switching On Supply Power
It is recommended that the load is turned on by its own power switch and that the LOAD POWER
switch is set to DIRECT before the AC1000 is turned on by setting the POWER switch to
the LOAD POWER switch set to DIRECT the correction circuit is bypassed and the load is
connected directly to the incoming AC line; this avoids the possibility of any surge current
associated with switching on the load causing damage to the power output amplifier. Note that if a
HA1600 is connected to the AC1000, the load connected to the HA1600 should be switched on
too.
When the power is first switched on, an internal relay bypasses the correction circuit so that any
surge current to the load bypasses the power output stage should the LOAD POWER switch
have been left set to CORRECTED. The buzzer sounds a pulsed tone while the correction circuit
is being bypassed, see Audible Warnings section.
Operation
l. With
Switching On CORRECTED Load Power
With the POWER switch on ( l ) and the LOAD POWER switch in the DIRECT position the load is
connected directly to the incoming AC supply.
A few seconds after the POWER switch has been turned on ( l ) the buzzer tone will cease,
indicating that the correction circuit has stabilised. The LOAD POWER switch may then be set to
CORRECTED, providing the load with a corrected supply which is harmonically pure. There is a
brief interruption of the AC supply to the load (<2ms) at switch-over; this, however, is well within
the minimum interruption period (10ms) that equipment must withstand to comply with other
aspects of EMC immunity, so should present no problem.
Voltage, Frequency, Power and Harmonic Purity in CORRECTED Mode
The AC1000 does not change the frequency of the incoming AC supply. Note that not all national
supplies guarantee to meet the ±0·5% frequency specification required by EN61000-3-2, although
in practice this will generally be the case.
Similarly, local supply voltage variations mean that the AC1000 will not generally meet the ±2%
voltage tolerance required by EN61000-3-2 unless it is supplied from a variable voltage of
suitable power rating such that the output voltage can be adjusted to the required limits. In
practice this is easily done because the supply voltage changes quite slowly, giving a reasonable
time within specification before further adjustment is needed.
The AC1000 will supply up to 1000VA at 230V, up to 500VA at 115V or up to 440VA at 100V; the
maximum load current in all cases is 4·4Arms.
Because the input distortion of the incoming AC supply is undefined it is not possible to guarantee
the harmonic purity of the output. However, in practice the AC1000 will produce an output which
meets the stringent source requirements of EN61000-3-2 even when the AC supply is severely
distorted.
If the AC1000 is being used as a power source for a HA1600, the HA1600 can be used to monitor
the quality of the corrected supply voltage. Voltage, frequency and harmonic distortion can be
continuously monitored and a 'PASS' or 'FAIL' message shown in the Meter display to indicate
whether the corrected source is compliant with the requirements of EN 61000-3-2 or not.
7
Loading...
+ 18 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.