Tenma 72-6859 User Manual

Page 1
72-6859
20MHz FUNCTION GENERATOR
INSTRUCTION MANUAL
Page 2
Tenma 72-6859 Instruction manual
Table of Contents
Introduction 2
Specification 2
EMC 5
Safety 4
Installation 5
Operation 7
Maintenance 8
www.tenma.com 1
Page 3
Tenma 72-6859 Instruction manual
This instrument is a 20MHz function generator with waveforms of sine, square, triangle, ramp, pulse and DC from a variable amplitude 50 output. A TTL output is also provided.
The frequency range is 0.2Hz to 20MHz selected by an eight decade range multiplier and calibrated vernier. The sweep input can be used to give a >20:1 frequency change within a selected range.
The generator output level is 20Vpk-pk maximum from a 50 source. The level is set via switched attenuators plus vernier with a total range of 66dB. DC offset is vernier adjustable over a ±10V range with 0V defined by a centre detent.
A switchable symmetry control with a 6:1 range permits ramp and pulse waveforms to be produced.
FREQUENCY
Frequency Range: 0.2Hz to 20MHz in 8 overlapping decade ranges with fine adjustment
by a vernier. Vernier Range: >10:1 on each range. Vernier Accuracy: Typically ± 5% of full scale.
Introduction
Specification
SWEEP MODE (EXTERNAL)
Sweep Range: Typically >20:1. Input Impedance: Input Sensitivity: Typically 0 to 2V for 10:1 sweep. Maximum Allowable
Input Voltage: Maximum Slew Rate of
sweep voltage:
OPERATING MODES
(Specifications apply for the top decade of each frequency range and output 10V peak-to-peak into 50Ω termination).
Sine
Distortion: Typically 2% on 200, 2k and 20k ranges. Amplitude Flatness: ± 0.2dB to 200kHz; ± 2dB to 20MHz.
Triangle
Linearity: Typically 99% on kHz ranges.
Squarewave
82 k
± 10V
0.1V/us
Rise and Fall Times: <22ns
DC
Range:
±10V from 50
Symmetry
Symmetry Range: Typically variable from 1:6 to 6:1 up to 500kHz.
2 www.tenma.com
Page 4
Tenma 72-6859 Instruction manual
OUTPUTS
50
Three switch-selectable ranges with 26dB vernier control within each attenuator range.
0dB Range:
1V to 20V peak-to peak from 50 (0.5V to 10V into 50).
–20dB Range:
–40dB Range:
DC Offset Control Range:
TTL
Capable of driving 4 standard TTL loads.
GENERAL
Power: 230V or 115V nominal 50/60Hz, adjustable internally; operating
Operating Range: + 5ºC to +40ºC, 20% to 80% RH.
Storage Range:
Environmental: Indoor use at altitudes up to 2000m, Pollution Degree 2.
Electrical Safety: Complies with IEC 1010-1
EMC: Complies with EN50081-1 and EN50082-1
Size: 220(W) x 82(H) x 230(D) mm, (10.3 x 3.4 x 9.2”) excluding feet.
100mV to 2V peak-to-peak (50mV to 1V into 50Ω)
10mV to 0.2V peak-to-peak (5mV to 0.1V into 50Ω)
±10V from 50. DC offset plus signal peak limited to ±10V (± 5V into 50). DC offset plus waveform attenuated proportionally in
20dB and 40dB position.
range ±14% of nominal; 30VA max. Installation Category II.
40ºC to +70ºC
Weight: 1.5 kg (3.3lb).
www.tenma.com 3
Page 5
Tenma 72-6859 Instruction manual
This instrument is Safety Class I according to IEC classification and has been designed to meet the requirements of IEC 1010-1 (Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for operation from a normal single phase supply.
This instrument has been tested in accordance with IEC 1010-1 and has been supplied in a safe condition. This instruction manual contains some information and warnings which have to be followed by the user to ensure safe operation and to retain the instrument in a safe condition.
This instrument has been designed for indoor use in a Pollution Degree 2 environment in the temperature range 5°C to 40°C, 20% - 80% RH (non-condensing). It may occasionally be subjected to temperatures between +5°C and –10°C without degradation of its safety. Do not operate while condensation is present.
Use of this instrument in a manner not specified by these instructions may impair the safety protection provided. Do not operate the instrument outside its rated supply voltages or environmental range.
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conductor inside or outside the instrument will make the instrument dangerous. Intentional interruption is prohibited. The protective action must not be negated by the use of an extension cord without a protective conductor.
Safety
When the instrument is connected to its supply, terminals may be live and opening the covers or removal of parts (except those to which access can be gained by hand) is likely to expose live parts. The apparatus shall be disconnected from all voltage sources before it is opened for any adjustment, replacement, maintenance or repair.
Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of the hazard involved.
If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture or chemical corrosion the safety protection may be impaired and the apparatus should be withdrawn from use and returned for checking and repair.
Make sure that only fuses with the required rated current and of the specified type are used for replacement. The use of makeshift fuses and the short-circuiting of fuse holders is prohibited.
Do not wet the instrument when cleaning it.
The following symbols are used on the instrument and in this manual:-
Caution -refer to the accompanying documentation, incorrect operation may damage the instrument.
terminal connected to chassis ground.
mains supply OFF.
l
mains supply ON.
alternating current.
4 www.tenma.com
Page 6
Tenma 72-6859 Instruction manual
This function generator has been designed to meet the requirements of the European EMC Directive 89/336/EEC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN50081-1 (1992) Generic emission standard for residential commercial and light industry. Test methods and limits used were:
a) EN55022 Conducted, Class B
b) EN55022 Radiated, Class B
Immunity
EN50082-1 (1992) Generic immunity standard for residential, commercial and light industry. Test methods and limits used were:
a) EN60801-2 (1993) Electrostatic Discharge, 8 kV air discharge.
b) IEC801-3 (1984) RF Field, 3 V/m.
c) IEC801-4 (1988) Fast Transient, 1 kV peak (AC line), 0.5kV peak (signal lines).
EMC
Cautions
To ensure continued compliance with the EMC directive the following precautions should be observed:
a) connect the generator to other equipment using only high quality, double-screened
cables.
b) after opening the case for any reason ensure that all signal and ground connections
are remade correctly before replacing the covers. Always ensure all case screws are correctly refitted and tightened.
c) In the event of part replacement becoming necessary, only use components of an
identical type, see the Service Manual.
www.tenma.com 5
Page 7
Tenma 72-6859 Instruction manual
Mains Operating Voltage
The operating voltage of the instrument is shown on the rear panel. Should it be necessary to change the operating voltage from 230V to 115V or vice-versa, proceed as follows:
1. Disconnect the instrument from all voltage sources.
2. Remove the 4 screws which hold the upper and lower case halves together and lift off the
case upper.
3. Change the transformer connections following the diagrams below.
230V Operation - Primaries in series
Installation
115V Operation - Primaries in parallel
4. Reassemble in the reverse order.
5. To comply with safety standard requirements the operating voltage marked on the rear
panel must be changed to clearly show the new voltage setting.
Mains Lead
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conductor inside or outside the instrument will make the instrument dangerous. Intentional interruption is prohibited.
6 www.tenma.com
Page 8
Tenma 72-6859 Instruction manual
Power
The ON/OFF switch is on the rear panel of the instrument.
Frequency Selection
Frequency range is selected by an eight position multiplier switch with fine adjustment by calibrated vernier.
Symmetry
A slide switch selects the symmetry control which varies the duty cycle 1:6 to 6:1 to produce sawtooth and variable pulse-width waveforms.
Note: The symmetry of the Aux Output does not vary when Sine or Triangle are selected because the Aux Output is generated by a zero crossing comparator.
Function Selection
The output waveform shape is selected by a four position switch to give sine, square, triangle and DC. A DC level only is useful as it permits input threshold testing of a circuit without having to connect up an external supply.
Operation
Main Output
The amplitude of the 50Ω MAIN output is controlled by the 3-position ATTENUATOR switches and AMPLITUDE control. Maximum output is 20 volts peak-to-peak from 50 and 10 volts peak­to-peak when terminated with a 50 load.
The AMPLITUDE control has greater than 26dB range and used in conjunction with the ATTENTUATORS a range of 0dB to 66dB can be achieved. This provides a range of 20V peak­to-peak down to 10mV peak-to-peak, or 10V peak-to-peak down to 5mV peak-to-peak into 50Ω.
Greater attenuation can be achieved by using standard 50 BNC attenuators. To maintain waveform integrity only 50 cable should be used and the receiving end should be terminated with a 50 load. The 50 MAIN output will withstand a short circuit for a period of 2 minutes at maximum output and greater periods at lower output levels.
Do not apply external voltages to this output.
DC Offset
The DC OFFSET control has a range of ±10 volts from 50 in all output modes; the control has a centre detent for 0 volts. DC offset plus signal peak is limited to ± 10V (± 5V into 50). DC offset plus waveform is attenuated proportionally in the 20dB and 40dB positions.
Aux Output
The AUX output provides a fixed TTL pulse output at the same frequency and phase as the 50 MAIN output and is capable of driving 4 standard TTL loads.
Sweep Input
The generator frequency can be swept, DC programmed or modulated by a suitable control voltage applied to the SWEEP IN socket. The instrument sums the SWEEP IN voltage with the internal control voltage derived from the FREQUENCY vernier to determine the operating frequency. A positive voltage increases the frequency; for frequency control with positive-going DC inputs the vernier should therefore be set to the lower frequency limit of the range to be swept. For example, a 0V to +2V signal will sweep the generator 1 decade up from range minimum, set by the vernier, to range maximum.
www.tenma.com 7
Page 9
Tenma 72-6859 Instruction manual
Similarly, a negative voltage decreases the frequency and for negative-going DC inputs the vernier should be set to the upper frequency limit of the range to be swept. For example, a 0V to
2V signal will sweep the generator 1 decade down from range maximum, set by the vernier, to range minimum.
To use a sweep signal which is symmetrical about ground, the vernier should be set to give a frequency at approximately the centre of the band to be swept.
Do not apply external voltages exceeding ±10V.
For service repair and calibration please contact your local Tenma distributor or go to
www.tenma.com
Cleaning
If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild detergent.
Maintenance
WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER ALLOW WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER CLEAN WITH SOLVENTS.
Mains Fuse
The transformer primary is protected by an integral thermal fuse which is not user replaceable. Should a fault develop which causes the thermal fuse to blow, the generator must be returned to the manufacturer or agent for service.
8 www.tenma.com
Page 10
TEST EQUIPMENT
405 Pioneer Blvd.
Springboro, Ohio 45066
Book Part No. 48591-0730 Issue 1
Loading...