Teledyne 3010TA User Manual

Oxygen Analyzer Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer Oxygen Analyzer
OPERA TING INSTRUCTIONS
Model 3010TA
Trace Oxygen Analyzer
Flush Mount Control Unit, PN D-64596A*
NEC Type Analysis Unit, PN D-65478*
DANGER
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM. HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PER-
SIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED. ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING.
BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
P/N M65693
11/22/99
ECO:#99-0459
i
Model 3010Model 3010
Model 3010
Model 3010Model 3010
TT
AA
T
A
TT
AA
Copyright © 1999 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an autho­rized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumen­tation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.
Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process
conditions.
ii
Oxygen Analyzer Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer Oxygen Analyzer
Table of Contents
Specific Model Information..................................iv
Preface ................................................................v
Part I: Control Unit, Model TA................. Part I: 1-1
Part II: Analysis Unit, Model T ............... Part II: 1-1
Appendix ......................................................... A-1
iii
Model 3010Model 3010
Model 3010
Model 3010Model 3010
TT
AA
T
A
TT
AA
Specific Model Information
The instrument for which this manual was supplied may incorporate one or more options not supplied in the standard instrument. Commonly available options are listed below, with check boxes. Any that are incorpo­rated in the instrument for which this manual was supplied are indicated by a check mark in the box.
Instrument Serial Number: __________________________
The instrument with the above serial number has the following Options:
o 3010TA-C Three gas inputs, for sample, zero and span gases, with
three solenoid-actuated gas-flow control valves built in. Valves are automatically synchronized to the analyzer's electronic control sequences.
o 3010TA–F Built-in flame arresters for Groups C and D service. o 3010TA–G Built-in flame arresters for Groups C and D service, plus
gas-control valves as in –C option, above.
o 3010TA–H Built-in flame arresters for Group B (hydrogen) service. o 3010TA–I Built-in flame arresters for Group B (hydrogen) service,
plus gas-control valves as in –C option, above.
o 19" Rack Mount
The 19" Relay Rack Mount units are available with either one or two series 3000 analyzer Control Units installed in a standard 19" panel and ready to mount in a standard rack. See Appendix for details.
o Cell Class* ____________________ (L-2C standard).
Enter Class Designation
* See Part II, Chapter 2 and/or any addendum that may be
attached to this manual for cell specifications.
iv
Oxygen Analyzer Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer Oxygen Analyzer
Preface
Overview
The Analytical Instruments Model 3010TA Trace Oxygen Analyzer is a versatile microprocessor-based instrument for detecting parts-per-million (ppm) levels of oxygen in a variety of background gases. It is a “split archi­tecture” instrument. This means that a general purpose Control Unit, de­signed for nonhazardous areas only, remotely controls a specially designed Analysis Unit, or remote probe, that can operate in a hazardous area.
Part I of this manual covers the Model 3010TA General Purpose flush­panel and/or rack-mount Control Unit only. This Control Unit is for indoor use in a nonhazardous environment. The Analysis Units (or Remote Probes) they control, can be designed for a variety of hazardous environments. Part II of this manual covers the 3010T Analysis Unit.
Typical Applications
A few typical applications of the Model 3010TA are:
Monitoring inert gas blanketing
Air separation and liquefaction
Chemical reaction monitoring
Semiconductor manufacturing
Petrochemical process control
Quality assurance
Gas analysis certification.
Model and Part Number Designations
The part numbers are the most specific identification. When using this manual for operation, maintenance, or ordering parts, check the part numbers
v
Model 3010Model 3010
Model 3010
Model 3010Model 3010
on your Instruments to be sure of a match. Where an underscore (_) appears in a model number, the unit has more than one application. For example, 3010T_C means that the same unit is part of the 3010TAC and the 3010TBC models.
3010TA: NEC Type Trace Oxygen Analyzer with flush mount Control
3010PA: NEC Type Percent Oxygen Analyzer with flush mount
3010TB: NEC type Trace Oxygen Analyzer with bulkhead mount
3010PB: NEC type Percent Oxygen Analyzer with bulkhead mount
TT
AA
T
A
TT
AA
Unit. Consists of 3010TA Control Unit, PN D-64596A and a 3010T Analysis Unit, PN D-65478.
Control Unit. Consists of 3010PA Control Unit, PN D-64596B and a 3010P Analysis Unit, PN D-65479.
Control Unit. Consists of 3010TB/PB Control Unit, PN D-66190A, and a 3010T Analysis Unit, PN D-65478.
Control Unit. Consists of 3010TB Control Unit, PN D-66190
B or C, and a 3010T Analysis Unit, PN D-65479.
3010TAC: CENELEC type Trace Oxygen Analyzer with flush mount
Control Unit. Consists of 3010TA Control Unit, PN D-66192A, and a 3010T_C Analysis Unit, PN D-66193.
3010PAC: CENELEC type Percent Oxygen Analyzer with flush mount
Control Unit. Consists of 3010PA Control Unit, PN D-66192 B or C, and a 3010P_C Analysis Unit, PN D-66191.
3010TBC: CENELEC type Trace Oxygen Analyzer with bulkhead mount
Control Unit. Consists of 3010TB Control Unit, PN D-66194A, and a 3010T_C Analysis Unit, PN D-66193.
3010PBC: CENELEC type Percent Oxygen Analyzer with bulkhead
mount Control Unit. Consists of 3010PB Control Unit, PN D-66194 Bor C, and a 3010P_C Analysis Unit, PN D-66191.
Options: See Specific Model Information sheet, on page iv for details.
Main Features of the Analyzer
The Model 3010TA series Oxygen Analyzers are sophisticated yet
simple to use. The main features of these analyzers include:
A 2-line alphanumeric display screen, driven by microprocessor electronics, that continuously prompts and informs the operator.
vi
High resolution, accurate readings of oxygen content: from low ppm levels through 25%. Large, bright, meter readout.
Oxygen Analyzer Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer Oxygen Analyzer
Stainless steel cell block is standard equipment.
Advance design Micro-Fuel Cell sensor, Specially designed model for trace analysis, has a one year warranty and an expected lifetime of two years.
Versatile analysis over a wide range of applications.
Microprocessor based electronics: 8-bit CMOS microprocessor with 32 kB RAM and 128 kB ROM.
Three user definable output ranges allow best match to users process and equipment: 0-10 ppm through 0-250,000 ppm for trace analysis.
Air-calibration range for convenient spanning at 20.9 %.
Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest.
Two adjustable concentration alarms and a system failure alarm.
Self-diagnostic testing, at startup and on demand, with continuous power-supply monitoring.
Two way RFI protection.
RS-232 serial digital port for use with a computer or other digital communications device.
Analog outputs for Concentration and Analysis Range: 0–1 V dc standard. Isolated 4–20 mA dc standard.
Compact and versatile design: flush-panel, rack-mountable, or bulkhead mounted Control Units available.
vii
Model 3010Model 3010
Model 3010
Model 3010Model 3010
TT
AA
T
A
TT
AA
viii
Part I: Control Unit
OPERATING INSTRUCTIONS
Model 3010TA
Oxygen Analyzer
Part I: Control Unit
Flush Mount
Part Numbers: D-64596A
Part I: i
Model 3010TA Oxygen Analyzer
Table of Contents
1 Introduction
1.1 Overview........................................................................ 1-1
1.2 Control Unit Front Panel................................................. 1-1
1.3 Recognizing Difference Between LCD & VFD............... 1-3
1.4 Control Unit Rear Panel................................................. 1-3
2 Operational Theory
2.1 Introduction .................................................................... 2-1
2.2 Electronics and Signal Processing ................................ 2-1
2.3 Temperature Control ...................................................... 2-3
3 Installation
3.1 Unpacking the Control Unit............................................ 3-1
3.2 Mounting the Control Unit .............................................. 3-1
3.3 Electrical Connections ................................................... 3-3
3.4 Testing the System......................................................... 3-9
4 Operation
4.1 Introduction .................................................................... 4-1
4.2 Using the Data Entry and Function Buttons ................... 4-2
4.3 The
4.3.1 Tracking the O2 Readings during Calibration ......... 4-4
4.3.2 Setting up an Auto-Cal........................................... 4-5
4.3.3 Password Protection.............................................. 4-5
4.3.4 Logout.................................................................... 4-9
4.3.5 System Self-Diagnostic Test .................................. 4-9
4.3.6 Version Screen ...................................................... 4-10
4.3.7 Showing Negative Oxygen readings ..................... 4-10
4.4 The
System
4.3.3.1 Entering the Password................................... 4-6
4.3.3.2 Installing or Changing the Password ............. 4-7
Zero
Function ..................................................... 4-3
and
Span
Functions ....................................... 4-11
ii: Part I
Part I: Control Unit
4.4.1 Zero Cal................................................................. 4-12
4.4.1.1 Auto Mode Zeroing ........................................ 4-12
4.4.1.2 Manual Mode Zeroing.................................... 4-12
4.4.1.3 Cell Failure .................................................... 4-13
4.4.2 Span Cal................................................................ 4-14
4.4.2.1 Auto Mode Spanning ..................................... 4-14
4.4.2.2 Manual Mode Spanning................................. 4-15
4.4.3 Span Failure .......................................................... 4-16
4.5 The
4.6 The
4.6.1 Setting the Analog Output Ranges......................... 4-19
4.6.2 Fixed Range Analysis ............................................ 4-19
4.7 The
4.8 Signal Output ................................................................. 4-21
Alarms Range
Analyze
Function...................................................... 4-16
Function ...................................................... 4-18
Function.................................................... 4-20
5 Maintenance
5.1 Fuse Replacement ......................................................... 5-1
5.2 System Self Diagnostic Test........................................... 5-2
5.3 Major Internal Components............................................ 5-3
5.4 Cleaning ........................................................................ 5-4
Part I: iii
Model 3010TA Oxygen Analyzer
iv: Part I
Oxygen Anal yzer Part I: Control Unit
Introduction
1.1 Overview
The Analytical Instruments Model 3010TA Analyzer Control Unit, together with a 3010T Analysis Unit, is a versatile microprocessor-based instrument for detecting trace amounts of oxygen in a variety of gases.
Part I, this part, of this manual covers the Model 3010TA series General Purpose flush-panel and/or rack-mount Control Units. (The Analysis Unit is covered in Part II of this manual.) The Control Unit is for indoor use in a nonhazardous environment only. The Analysis Units (or Remote Probes) it controls can be designed for a variety of hazardous environments.
1.2 Control Unit Front Panel
The standard 3010TA Control Unit is housed in a rugged metal case with all remote controls and displays accessible from the front panel. See Figure 1-1. The front panel has a digital meter, an alphanumeric display, and thirteen buttons for operating the analyzer.
Part I: 1-1
1 Introduction Model 3010TA
T
e
l
e
d
y
n
e
A
n
a
l
y
t
i
c
a
l
I
n
s
t
ru
m
e
n
t
Digital Meter
s
Alphanumeric Display
301
0T
O
X
Y
G
E
N
A
N
A
L
Y
Z
E
R
Data Entry Buttons
F uncti o n B u tto n s
Figure 1-1: Front of Unmounted Control Unit
Function Keys: Six touch-sensitive membrane switches are used to
change the specific function performed by the analyzer:
Door Latch
Standby Sw itch
Analyze Perform analysis for oxygen content of a sample gas.
System Perform system-related tasks (described in detail in
chapter 4, Operation.).
Span Span calibrate the analyzer.
Zero Zero calibrate the analyzer.
Alarms Set the alarm setpoints and attributes.
Range Set up the 3 user definable ranges for the instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD display:
Left & Right Arrows Select between functions currently
displayed on the VFD screen.
Up & Down Arrows Increment or decrement values of
functions currently displayed.
1-2: Part I
Oxygen Anal yzer Part I: Control Unit
Enter Moves VFD display on to the next screen in a series. If
none remains, returns to the
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the
Digital Meter Display: The meter display is a LED device that produces large, bright, 7-segment numbers that are legible in any lighting. It is accurate across all analysis ranges. The –TA models produce continuous readout from 0-10,000 ppm and then switch to continuous percent readout from 1-25 %.
Alphanumeric Interface Screen: The backlit VFD screen is an easy­to-use interface between operator and analyzer. It displays values, options, and messages that give the operator immediate feedback.
I/O Power Button: The red I/O button switches the instrument power between I (ON) and O (a Keep-Alive state). In the O state, the instrument’s circuitry is operating, but there are no displays or outputs.
Analyze
screen.
Analyze
screen.
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When chassis is exposed or when access door is open and power cable is connected, use extra care to avoid contact with live electrical circuits .
Access Door: For access to the front panel electronics, the front panel swings open when the latch in the upper right corner of the panel is pressed all the way in with a narrow gauge tool. Accessing the main circuit board and other electronics requires unfastening the rear panel screws and sliding the unit out of the case.
1.3 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has DARK background with GREEN characters. In the case of VFD - NO CONTRAST ADJUSTMENT IS NEEDED.
1. 4 Control Unit Rear Panel
The Control Unit rear panel, shown in Figure 1-2, contains the electri­cal connectors for external inputs and outputs. The input/output functions are described briefly here and in detail in the Installation chapter of this manual.
Part I: 1-3
1 Introduction Model 3010TA
Figure 1-2: Model 3010TA Rear Panel
Power Connection Universal AC power source.
Analog Outputs 0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20 mA dc range ID.
Alarm Connections 2 concentration alarms and 1 system
alarm.
RS-232 Port Serial digital concentration signal output
and control input.
Remote Probe Provides all electrical interconnect to the
Analysis Unit or Remote Probe.
Remote Span/Zero Digital inputs allow external control of
analyzer calibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and readings are not monitoring sample.
Range ID Contacts Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
1-4: Part I
Oxygen Anal yzer Part I: Control Unit
Remote Probe Interfaces with an Analysis Unit or
Remote Probe (external sensor/sample system).
Network I/O Serial digital communications for local
network access. For future expansion. Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the Model 3010TA is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
Part I: 1-5
1 Introduction Model 3010TA
1-6: Part I
Oxygen Anal yzer Part I: Control Unit
Operational Theory
2.1 Introduction
The Model 3010TA Oxygen Analyzer Control Unit uses an 8031 microcontroller with 32 kB of RAM and 128 kB of ROM to control all signal processing, input/output, and display functions for the Model 3010TA analyzer. (The sample system and Micro-Fuel Cell sensor are covered in Part II, Analysis Unit, in this manual.) System power is supplied from a universal power supply module designed to be compatible with any international power source.
2.2 Electronics and Signal Processing
All of the Analyzer electronics are located on Printed Circuit Board (PCB) assemblies inside the Control Unit chassis. The PCB locations are illustrated in section 5, Maintenance.
Refer to Figure 2-1, Block Diagram of the 3010TA CU Electronics:
In the presence of oxygen, the sensor (in the Analysis Unit) generates a current. A current to voltage amplifier (in the Control Unit) converts this current to a voltage.
The second stage amplifier amplifies the voltage. It also uses a signal from the thermistor (which is physically located in the Analysis Unit cell block) to provide temperature compensation for the sensor signal. The thermistor is a temperature dependent resistance that changes the gain of the amplifier in proportion to the temperature changes in the block. This thermistor signal compensates for the change in the cell output due to the temperature changes. The result is a signal that is temperature independent. The output from the second stage amplifier is sent to an 18-bit analog to digital converter controlled by the microprocessor.
Part I: 2-1
2 Operational Theory Model 3010TA
Figure 2-1: Block Diagram of the 3010TA CU Electronics
2-2: Part I
Oxygen Anal yzer Part I: Control Unit
The digital concentration signal—along with input from the control panel—is processed by the microprocessor, and appropriate control signals are directed to the display, alarms and communications port as well as to the gas control valves in the Analysis Unit.
The same digital information is also sent to a 12 bit digital to analog converter that produces the 4-20 mA dc and the 0-1 V dc analog concentra­tion signal outputs, and the analog range ID outputs.
The microprocessor monitors the power supply, and activates the system failure alarm if a malfunction is detected.
2.3 Temperature Control
For accurate analysis this instrument is temperature controlled not to fall beneath a certain temperature. This temperature is 22oF. This is to prevent the sensor from freezing in cold environments.
Part I: 2-3
2 Operational Theory Model 3010TA
2-4: Part I
Oxygen Anal yzer Part I: Control Unit
Installation
Installation of Model 3010 Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control Unit and the Analysis Unit
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Unit. (Installation of the
Analysis Unit is covered in Part II of this manual.)
3.1 Unpacking the Control Unit
The analyzer is shipped with all the materials you need to install and prepare the system for operation. Carefully unpack the Control Unit and inspect it for damage. Immediately report any damage to the shipping agent.
3.2 Mounting the Control Unit
The Model 3010TA Control Unit is for indoor use in a general purpose area. It is NOT for hazardous environments of any type.
The standard model is designed for flush panel mounting. Figure 3-1 is an illustration of a Model 3010 standard Control Unit front panel and mount­ing bezel. There are four mounting holes—one in each corner of the rigid frame. Drawing number D-64596, at the back of this manual, contains a panel cutout diagram.
On special order, a 19" rack-mounting can be provided. Per order, one or two 3010 series Control Units are flush-panel mounted on the 19" rack panel. See Figure 3-2.
Part I: 3-1
3 Installation Model 3010TA
Mounting Holes (4)
Latch
Hinge
Figure 3-1: Front Panel of the Model 3010 Control Unit
Figure 3-2: Single and Dual 19" Rack Mounts
All operator controls are mounted on the control panel, which is hinged on the left edge and doubles as a door to provide access to the internal components of the instrument. The door is spring loaded and will swing
3-2: Part I
Oxygen Anal yzer Part I: Control Unit
open when the button in the center of the latch (upper right corner) is pressed all the way in with a narrow gauge tool (less than 0.18 inch wide), such as a small hex wrench or screwdriver Allow clearance for the door to open in a 90-degree arc of radius 7.125 inches. See Figure 3-3.
Figure 3-3: Required Front Door Clearance
3.3 Electrical Connections
Figure 3-4 shows the Control Unit rear panel. Connections for power, communications, and both digital and analog signal outputs are described in the following paragraphs. Wire size and maximum length data appear in the Drawings at the back of this manual.
Figure 3-4: Rear Panel of the Model 3010 Control Unit
For safe connections, ensure that no uninsulated wire extends outside of the connectors they are attached to. Stripped wire ends must insert com­pletely into terminal blocks. No uninsulated wiring should be able to come in contact with fingers, tools or clothing during normal operation.
Part I: 3-3
3 Installation Model 3010TA
Primary Input Power: The universal power supply requires a 85–250
V ac, 47-63 Hz power source. The power cord receptacle and fuse block are located in the same assembly. Insert the female plug end of the power cord into the power cord receptacle.
CAUTION: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power source. The red switching power on or off to the displays and out­puts only.
Fuse Installation: The fuse block, at the right of the power cord
receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation. (See Fuse Replace- ment in chapter 5, maintenance.)
Analog Outputs: There are four DC output signal connectors with
spring terminals on the panel. There are two wires per output with the polarity noted. See Figure 3-5. The outputs are:
I/O
switch on the front panel is for
0–1 V dc % of Range: Voltage rises linearly with increasing oxygen, from
0 V at 0 ppm to 1 V at full scale ppm. (Full scale = 100% of programmed range.)
0–1 V dc Range ID: 0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.
4–20 mA dc % Range: Current increases linearly with increasing oxygen,
from 4 mA at 0 ppm to 20 mA at full scale ppm. (Full scale = 100% of programmed range.)
4–20 mA dc Range ID: 8 mA = Low Range, 12 mA = Medium Range, 16
mA = High Range, 20 mA = Air Cal Range.
3-4: Part I
Figure 3-5: Analog Output Connections
Oxygen Anal yzer Part I: Control Unit
Alarm Relays: The three alarm-circuit connectors are spring terminals for making connections to internal alarm relay contacts. Each provides a set of Form C contacts for each type of alarm. Each has both normally open and normally closed contact connections. The contact connections are indicated by diagrams on the rear panel. They are capable of switching up to 3 am­peres at 250 V ac into a resistive load. See Figure 3-6. The connectors are:
Threshold Alarm 1: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm: Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently configured as fail-safe and latching. Cannot be de­feated. Actuates if self test fails.
(Reset by pressing press
I/O
again and any other button EXCEPT
System
Further detail can be found in chapter 4, section 4-5.
to resume.
I/O
button to remove power. Then
Figure 3-6: Types of Relay Contacts
Part I: 3-5
3 Installation Model 3010TA
Digital Remote Cal Inputs: Accept 0 V (off) or 24 V dc (on) inputs
for remote control of calibration. (See Remote Calibration Protocol below.) Zero: Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the grounded at the source of the signal. Signal must be removed before zeroing is complete, or the zeroing will repeat. The Analysis Unit internal valves operate synchronously to supply the zero gas. See Remote Probe Connector at end of section 3.3.
Span: Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the grounded at the source of the signal. Signal must be removed before spanning is complete, or the spanning will repeat. The Analysis Unit internal valves operate synchronously to supply the span gas. See Remote Probe Connector at end of section 3.3.
Zero
mode. Either side may be
Span
mode. Either side may be
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 3010 Analyzer, the customer's controller must monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
3-6: Part I
Oxygen Anal yzer Part I: Control Unit
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Probe connector (paragraph 3.3) provides signals
to the Analysis Unit to ensure that the zero and span gas valves will be controlled synchronously.
Range ID Relays: Four dedicated Range ID relay contacts. The first
three ranges are assigned to relays in ascending order—Low range is as­signed to Range 1 ID, Medium range is assigned to Range 2 ID, and High range is assigned to Range 3 ID. The fourth range is reserved for the Air Cal Range (25%).
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or other digital device. It requires a standard 9-pin D connector.
The data is status information, in digital form, updated every two
seconds. Status is reported in the following order:
The concentration in ppm or percent
The range in use (HI, MED, LO)
The span of the range (0-100 ppm, etc.)
Which alarms—if any—are disabled (AL–x DISABLED)
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 3-1.
Table 3-1: Commands via RS-232 Input
Command Description as<enter> Immediately starts an autospan. az<enter> Immediately starts an autozero. st<enter> Toggling input. Stops/Starts any status message output from
the RS-232, until st<enter> is sent again.
Part I: 3-7
3 Installation Model 3010TA
The RS-232 protocol allows some flexibility in its implementation.
Table 3-2 lists certain RS-232 values that are required by the 3010TA.
Table 3-2: Required RS-232 Options
Parameter Setting
Baud 2400
Byte 8 bits
Parity none
Stop Bits 1
Message Interval 2 seconds
Remote Probe Connector: The Models 3010 are split architecture
(dual-chassis) instruments, which have a Remote Probe, or Analysis Unit. The Remote Probe connector is used for controlling the Analysis Unit internal sample, zero, and span gas valves, and for receiving the oxygen sensor and thermistor signals. See Figure 3-7. The connections at the Analy­sis Unit are covered in detail in Part II, section 3.4, of this manual.
Figure 3-7: Remote Probe Connector Pinouts
If you use your own gas control valves, use the interconnect diagram in
Figure 3-8. (See drawing D-64950 for wire recommendations.)
3-8: Part I