All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Analytical Instruments, the manufacturer of this instrument, cannot accept
responsibility for conditions beyond its knowledge and control. No statement expressed
or implied by this document or any information disseminated by the manufacturer or its
agents, is to be construed as a warranty of adequate safety control under the user’s process
conditions.
ii
Oxygen Analyzer Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer Oxygen Analyzer
Table of Contents
Specific Model Information..................................iv
The instrument for which this manual was supplied may incorporate
one or more options not supplied in the standard instrument. Commonly
available options are listed below, with check boxes. Any that are incorporated in the instrument for which this manual was supplied are indicated by a
check mark in the box.
Instrument Serial Number: __________________________
The instrument with the above serial number has the following
Options:
o 3010PA-CThree gas inputs, for sample, zero and span gases, with
three solenoid-actuated gas-flow control valves built in.
Valves are automatically synchronized to the analyzer's
electronic control sequences.
o 3010PA–FBuilt-in flame arresters for Groups C and D service.
o 3010PA–GBuilt-in flame arresters for Groups C and D service, plus
gas-control valves as in –C option, above.
o 3010PA–HBuilt-in flame arresters for Group B (hydrogen) service.
o 3010PA–IBuilt-in flame arresters for Group B (hydrogen) service,
plus gas-control valves as in –C option, above.
o 3010PA–M4-20 mA current signal outputs for Percent of Full Scale
and Range ID, in addition to voltage outputs.
o 3010PA–SEntire sample system including cell block and all wetted
parts fabricated from stainless steel.
o 19" Rack Mount
The 19" Relay Rack Mount units are available with either
one or two series 3000 analyzer Control Units installed in
a standard 19" panel and ready to mount in a standard
rack. See Appendix for details.
o Cell Class*____________________ (B-1 standard).
Enter Class Designation
*See Part II, Chapter 2 and/or any addendum that may be attached
to this manual for cell specifications.
iv
Oxygen Analyzer Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer Oxygen Analyzer
Preface
Overview
The Analytical Instruments Model 3010PA Percent Oxygen Analyzer is
a versatile microprocessor-based instrument for detecting oxygen in a
variety of background gases. It is a “split architecture” instrument. This
means that a general purpose Control Unit, designed for nonhazardous areas
only, remotely controls a specially designed Analysis Unit, or remote probe,
that can operate in a hazardous area.
Part I of this manual covers the Model 3010PA General Purpose flushpanel and/or rack-mount Control Unit only. This Control Unit is for indoor
use in a nonhazardous environment. The Analysis Units (or Remote Probes)
they control, can be designed for a variety of hazardous environments. Part II
of this manual covers the 3010P Analysis Unit.
Typical Applications
A few typical applications of the Model 3010PA are:
•Monitoring inert gas blanketing
•Air separation and liquefaction
•Chemical reaction monitoring
•Semiconductor manufacturing
•Petrochemical process control
•Quality assurance
•Gas analysis certification.
Model and Part Number Designations
The part numbers are the most specific identification. When using this
manual for operation, maintenance, or ordering parts, check the part numbers
v
Model 3010Model 3010
Model 3010
Model 3010Model 3010
on your Instruments to be sure of a match. Where an underscore (_) appears
in a model number, the unit has more than one application. For example,
3010P_C means that the same unit is part of the 3010PAC and the 3010PBC
models.
3010TA:NEC Type Trace Oxygen Analyzer with flush mount Control
3010PA:NEC Type Percent Oxygen Analyzer with flush mount
3010TB:NEC type Trace Oxygen Analyzer with bulkhead mount
3010PB:NEC type Percent Oxygen Analyzer with bulkhead mount
PAPA
PA
PAPA
Unit. Consists of 3010TA Control Unit, PN D-64596A and a
3010T Analysis Unit, PN D-65478.
Control Unit. Consists of 3010PA Control Unit, PND-64596B and a 3010P Analysis Unit, PND-65479.
Control Unit. Consists of 3010TB/PB Control Unit, PND-66190A, and a 3010T Analysis Unit, PN D-65478.
Control Unit. Consists of 3010TB Control Unit, PN D-66190
B or C, and a 3010T Analysis Unit, PN D-65479.
3010TAC:CENELEC type Trace Oxygen Analyzer with flush mount
Control Unit. Consists of 3010TA Control Unit, PN
D-66192A, and a 3010T_C Analysis Unit, PN D-66193.
3010PAC:CENELEC type Percent Oxygen Analyzer with flush mount
Control Unit. Consists of 3010PA Control Unit, PN D-66192
B or C, and a 3010P_C Analysis Unit, PN D-66191.
3010TBC:CENELEC type Trace Oxygen Analyzer with bulkhead mount
Control Unit. Consists of 3010TB Control Unit, PN
D-66194A, and a 3010T_C Analysis Unit, PN D-66193.
3010PBC:CENELEC type Percent Oxygen Analyzer with bulkhead
mount Control Unit. Consists of 3010PB Control Unit, PN
D-66194 B or C, and a 3010P_C Analysis Unit, PN
D-66191.
Options:See Specific Model Information sheet, on page iv for details.
Main Features of the Analyzer
The Model 3010PA series Oxygen Analyzers are sophisticated yet
simple to use. The main features of these analyzers include:
vi
•A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•High resolution, accurate readings of oxygen content: from low
0-1 % levels through 0-100 %. Large, bright, meter readout.
Oxygen Analyzer Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer Oxygen Analyzer
•Optional stainless steel cell block available.
•Advance design Micro-Fuel Cell sensor with a one year
warranty and an expected lifetime of two years.
•Versatile analysis over a wide range of applications.
•Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•Three user definable output ranges allow best match to users
process and equipment: 0-1 % through 0-100 %.
•Air-calibration range for convenient spanning at 20.9 %.
•Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•Two adjustable concentration alarms and a system failure alarm.
•Self-diagnostic testing, at startup and on demand, with continuous
power-supply monitoring.
•Two way RFI protection.
•RS-232 serial digital port for use with a computer or other
digital communications device.
•Analog outputs for Concentration and Analysis Range: 0–1 V dc
standard. Additional isolated 4–20 mA dc optional.
•Compact and versatile design: flush-panel, rack-mountable, or
bulkhead mounted Control Units available.
vii
Model 3010Model 3010
Model 3010
Model 3010Model 3010
Commonwealth of Europe (CE) for Radio Frequency Interference,
Electromagnetic Interference (RFI/EMI), and Low Voltage Directive
(LVD).
The following International Symbols are used throughout the Instruction Manual for your visual and immediate warnings and when you
have to attend CAUTION while operating the instrument:
PAPA
PA
PAPA
Model 3010PA complies with all of the requirements of the
STAND-BY, Instrument is on Stand-by,
but circuit is active
GROUND
Protective Earth
CA UTION, The operator needs to refer to the manual
for further information. Failure to do so may
compromise the safe operation of the equipment.
The Analytical Instruments Model 3010PA Analyzer Control Unit,
together with a 3010P Analysis Unit, is a versatile microprocessor-based
instrument for detecting percent amounts of oxygen in a variety of gases.
Part I, this part, of this manual covers the Model 3010PA series
General Purpose flush-panel and/or rack-mount Control Units. (The Analysis Unit is covered in Part II of this manual.) The Control Unit is for indoor
use in a nonhazardous environment only. The Analysis Units (or Remote
Probes) it controls can be designed for a variety of hazardous environments.
1.2Control Unit Front Panel
The standard 3010PA Control Unit is housed in a rugged metal case
with all remote controls and displays accessible from the front panel. See
Figure 1-1. The front panel has a digital meter, an alphanumeric display,
and thirteen buttons for operating the analyzer.
Part I: 1-1
1 Introduction Model 3010PA
Figure 1-1: Front of Unmounted Control Unit
Function Keys: Six touch-sensitive membrane switches are used to
change the specific function performed by the analyzer:
•AnalyzePerform analysis for oxygen content of a sample gas.
•SystemPerform system-related tasks (described in detail in
chapter 4, Operation.).
•SpanSpan calibrate the analyzer.
•ZeroZero calibrate the analyzer.
•AlarmsSet the alarm setpoints and attributes.
•RangeSet up the 3 user definable ranges for the instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD display:
•Left & Right ArrowsSelect between functions currently
displayed on the VFD screen.
•Up & Down ArrowsIncrement or decrement values of
functions currently displayed.
1-2: Part I
Oxygen Anal yzerPart I: Control Unit
•EnterMoves VFD display on to the next screen in a series.
If none remains, returns to the
•Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the
Digital Meter Display: The meter display is a LED device that
produces large, bright, 7-segment numbers that are legible in any lighting.
It is accurate across all analysis ranges from 0-1 % through 0-100 %
Alphanumeric Interface Screen: The VFD screen is an easy-to-use
interface between operator and analyzer. It displays values, options, and
messages that give the operator immediate feedback.
I/O Power Button: The red I/O button switches the instrument power
between I (ON) and O (a Keep-Alive state). In the O state, the instrument’s
circuitry is operating, but there are no displays or outputs.
Analyze
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When
chassis is exposed or when access door is open
and power cable is connected, use extra care to
avoid contact with live electrical circuits .
screen.
Analyze
screen.
Access Door: For access to the front panel electronics, the front panel
swings open when the latch in the upper right corner of the panel is pressed
all the way in with a narrow gauge tool. Accessing the main circuit board
and other electronics requires unfastening the rear panel screws and sliding
the unit out of the case.
1.3 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.
1.4Control Unit Rear Panel
The Control Unit rear panel, shown in Figure 1-2, contains the
electrical connectors for external inputs and outputs. The input/output
functions are described briefly here and in detail in the Installation chapter
of this manual.
Part I: 1-3
1 Introduction Model 3010PA
Figure 1-2: Model 3010PA Rear Panel
•Power ConnectionUniversal AC power source.
•Analog Outputs0-1 V dc concentration and 0-1 V dc
range ID. Optional isolated 4-20 mA dc
and 4-20 mA dc range ID.
•Alarm Connections2 concentration alarms and 1 system
alarm.
•RS-232 PortSerial digital concentration signal
output and control input.
•Remote ProbeProvides all electrical interconnect to
the Analysis Unit or Remote Probe.
•Remote Span/ZeroDigital inputs allow external control of
analyzer calibration.
•Calibration ContactTo notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•Range ID ContactsFour separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
1-4: Part I
Oxygen Anal yzerPart I: Control Unit
•Remote ProbeInterfaces with an Analysis Unit or
Remote Probe (external sensor/sample
system).
•Network I/OSerial digital communications for local
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3010PA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
Part I: 1-5
1 Introduction Model 3010PA
1-6: Part I
Oxygen Anal yzerPart I: Control Unit
Operational Theory
2.1Introduction
The Model 3010PA Oxygen Analyzer Control Unit uses an 8031
microcontroller with 32 kB of RAM and 128 kB of ROM to control all
signal processing, input/output, and display functions for the Model
3010PA analyzer. (The sample system and Micro-Fuel Cell sensor are
covered in Part II, Analysis Unit, in this manual.) System power is supplied
from a universal power supply module designed to be compatible with any
international power source.
2.2Electronics and Signal Processing
All of the Analyzer electronics are located on Printed Circuit Board
(PCB) assemblies inside the Control Unit chassis. The PCB locations are
illustrated in section 5, Maintenance.
Refer to Figure 2-1, Block Diagram of the 3010PA CU Electronics:
In the presence of oxygen, the sensor (in the Analysis Unit) generates
a current. A current to voltage amplifier (in the Control Unit) converts this
current to a voltage.
The second stage amplifier amplifies the voltage. It also uses a signal
from the thermistor (which is physically located in the Analysis Unit cell
block) to provide temperature compensation for the sensor signal. The
thermistor is a temperature dependent resistance that changes the gain of
the amplifier in proportion to the temperature changes in the block. This
thermistor signal compensates for the change in the cell output due to the
temperature changes. The result is a signal that is temperature independent.
The output from the second stage amplifier is sent to an 18-bit analog to
digital converter controlled by the microprocessor.
Part I: 2-1
2 Operational TheoryModel 3010PA
Figure 2-1: Block Diagram of the 3010PA CU Electronics
2-2: Part I
Oxygen Anal yzerPart I: Control Unit
The digital concentration signal—along with input from the control
panel—is processed by the microprocessor, and appropriate control signals
are directed to the display, alarms and communications port as well as to
the optional gas control valves in the Analysis Unit.
The same digital information is also sent to a 12 bit digital to analog
converter that produces the 0-1 V dc and the optional 4-20 mA dc analog
concentration signal outputs, and the analog range ID outputs.
The microprocessor monitors the power supply, and activates the
system failure alarm if a malfunction is detected.
2.3 Temperature Control
For accurate analysis this instrument is temperature controlled not to
fall beneath a certain temperature. This temperature is 22oF. This is to
prevent the sensor from freezing in cold environments.
Part I: 2-3
2 Operational TheoryModel 3010PA
2-4: Part I
Oxygen Anal yzerPart I: Control Unit
Installation
Installation of Model 3010 Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control Unit and
the Analysis Unit
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Unit. (Installation of the
Analysis Unit is covered in Part II of this manual.)
3.1Unpacking the Control Unit
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the Control Unit and
inspect it for damage. Immediately report any damage to the shipping agent.
3.2Mounting the Control Unit
The Model 3010PA Control Unit is for indoor use in a general purpose
area. It is NOT for hazardous environments of any type.
The standard model is designed for flush panel mounting. Figure 3-1 is
an illustration of a Model 3010 standard Control Unit front panel and mounting bezel. There are four mounting holes—one in each corner of the rigid
frame. Drawing number D-64596, at the back of this manual, contains a
panel cutout diagram.
On special order, a 19" rack-mounting can be provided. Per order, one
or two 3010 series Control Units are flush-panel mounted on the 19" rack
panel. See Figure 3-2.
Figure 3-1: Front Panel of the Model 3010 Control Unit
Part I: 3-1
3 InstallationModel 3010PA
Mounting
Holes (4)
Latch
Hinge
Figure 3-1: Front Panel of the Model 3010 Control Unit
Figure 3-2: Single and Dual 19" Rack Mounts
All operator controls are mounted on the control panel, which is hinged
on the left edge and doubles as a door to provide access to the internal
components of the instrument. The door is spring loaded and will swing
open when the button in the center of the latch (upper right corner) is pressed
3-2: Part I
Oxygen Anal yzerPart I: Control Unit
all the way in with a narrow gauge tool (less than 0.18 inch wide), such as a
small hex wrench or screwdriver Allow clearance for the door to open in a
90-degree arc of radius 7.625 inches. See Figure 3-3.
Figure 3-3: Required Front Door Clearance
3.3Rear Panel Connections
Figure 3-4 shows the Control Unit rear panel. Connections for power,
communications, and both digital and analog signal outputs are described in
the following paragraphs. Wire size and maximum length data appear in the
Drawings in the back of this manual.
Figure 3-4: Rear Panel of the Model 3010 Control Unit
For safe connections, ensure that no uninsulated wire extends outside of
the connectors they are attached to. Stripped wire ends must insert completely into terminal blocks. No uninsulated wiring should be able to come in
contact with fingers, tools or clothing during normal operation.
Part I: 3-3
3 InstallationModel 3010PA
Primary Input Power: The universal power supply requires a 85–250
V ac, 47-63 Hz power source. The power cord receptacle and fuse block are
located in the same assembly. Insert the female plug end of the power cord
into the power cord receptacle.
CAUTION: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
I/O
source. The red
switching power on or off to the displays and outputs only.
Fuse Installation: The fuse block, at the right of the power cord
receptacle, accepts US or European size fuses. A jumper replaces the fuse in
whichever fuse receptacle is not used. Fuses are not installed at the factory.
Be sure to install the proper fuse as part of installation. (See Fuse Replace-ment in chapter 5, maintenance.)
Analog Outputs: There are four DC output signal connectors with
spring terminals on the panel. There are two wires per output with the
polarity noted. See Figure 3-5. The outputs are:
switch on the front panel is for
0–1 V dc % of Range:Voltage rises linearly with increasing oxygen, from
0 V at 0 % to 1 V at full scale. (Full scale = 100%
of programmed range.)
0–1 V dc Range ID:0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.
4–20 mA dc % Range: (Optional) Current increases linearly with increasing
oxygen, from 4 mA at 0 % to 20 mA at full scale.
(Full scale = 100% of programmed range.)
4–20 mA dc Range ID: (Optional) 8 mA = Low Range, 12 mA = Medium
Range, 16 mA = High Range, 20 mA = Air Cal.
3-4: Part I
Figure 3-5: Analog Output Connections
Oxygen Anal yzerPart I: Control Unit
Alarm Relays: The three alarm-circuit connectors are spring terminals
for making connections to internal alarm relay contacts. Each provides a set
of Form C contacts for each type of alarm. Each has both normally open and
normally closed contact connections. The contact connections are indicated
by diagrams on the rear panel. They are capable of switching up to 3 amperes at 250 V ac into a resistive load. See Figure 3-6. The connectors are:
Threshold Alarm 1:• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as failsafe or nonfailsafe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2:• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as failsafe or nonfailsafe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm:Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be defeated. Actuates if self test fails.
(Reset by pressing
press
I/O
again and any other button EXCEPT
System
Further detail can be found in chapter 4, section 4-5.
to resume.
I/O
button to remove power. Then
Figure 3-6: Types of Relay Contacts
Part I: 3-5
3 InstallationModel 3010PA
Digital Remote Cal Inputs: Accept 0 V (off) or 24 V dc (on) inputs
for remote control of calibration. (See Remote Calibration Protocol below.)
Zero:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the
grounded at the source of the signal. Signal must be removed
before zeroing is complete, or the zeroing will repeat. The
Analysis Unit internal valves operate synchronously to
supply the zero gas. See Remote Probe Connector at end of
section 3.3.
Span:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the
grounded at the source of the signal. Signal must be removed
before spanning is complete, or the spanning will repeat. The
Analysis Unit internal valves operate synchronously to
supply the span gas. See Remote Probe Connector at end of
section 3.3.
Zero
mode. Either side may be
Span
mode. Either side may be
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 3010 Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
3-6: Part I
Oxygen Anal yzerPart I: Control Unit
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Probe connector (paragraph 3.3) provides signals
to the Analysis Unit to ensure that the zero and span gas
valves will be controlled synchronously.
Range ID Relays: Four dedicated Range ID relay contacts. The first
three ranges are assigned to relays in ascending order—Low range is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and High
range is assigned to Range 3 ID. The fourth range is reserved for the Air Cal
Range (25%).
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future
versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or
other digital device. It requires a standard 9-pin D connector.
The data is status information, in digital form, updated every two
seconds. Status is reported in the following order:
•The concentration in percent
•The range in use (HI, MED, LO)
•The span of the range (0-10 %, etc)
•Which alarms—if any—are disabled (AL–x DISABLED)
•Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 3-1.
Table 3-1: Commands via RS-232 Input
CommandDescription
as<enter>Immediately starts an autospan.
az<enter>Immediately starts an autozero.
co<enter>Reports "Raw Cell Output" (current output of the sensor
itself) in µA. For example—
Cell Output: 99 µA
st<enter>Toggling input. Stops/Starts any status message output from
the RS-232, until st<enter> is sent again.
Part I: 3-7
3 InstallationModel 3010PA
The RS-232 protocol allows some flexibility in its implementation.
Table 3-2 lists certain RS-232 values that are required by the 3010PA.
Table 3-2: Required RS-232 Options
ParameterSetting
Baud2400
Byte8 bits
Paritynone
Stop Bits1
Message Interval2 seconds
Remote Probe Connector: The Model 3010PA is a split architecture
(dual-chassis) instrument, which has a Remote Probe, or Analysis Unit. The
Remote Probe connector is used for controlling the Analysis Unit internal
sample, zero, and span gas valves (which are optional), and for receiving the
oxygen sensor and thermistor signals. See Figure 3-7. The connections at
the Analysis Unit are covered in detail in Part II, section 3.4, of this manual.
Figure 3-7: Remote Probe Connector Pinouts
If you use your own gas control valves, use the interconnect diagram in
Figure 3-8 for the valves. (See drawing D-64950 for wire recommendations.)
3-8: Part I
Oxygen Anal yzerPart I: Control Unit
Figure 3-8: Remote Probe Connector Pinouts
The voltage from the solenoid outputs is nominally 0 V for the OFF
and 15 V dc for the ON conditions. The maximum combined current that
can be pulled from these output lines is 100 mA. (If two lines are ON at the
same time, each must be limited to 50 mA, etc.)
If more current and/or a different voltage is required, use relays, power
amplifiers, or other matching circuitry to provide the actual driving current.
Note that each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This can limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-9.
Figure 3-9: FET Series Resistance
3.4Testing the System
After The Control Unit and the Analysis Unit are both installed and
interconnected, and the system gas and electrical connections are complete,
the system is ready to test. Before plugging either of the units into their
respective power sources:
Part I: 3-9
Loading...
+ 75 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.