Teledyne 205, 201, 207, 203 User Manual

TELEDYNE HASTINGS
INSTRUMENTS
INSTRUCTION MANUAL
201/203/205/207 SERIES
FLOW METERS/CONTROLLERS
CERTIFIED
Manual Print History
The print history shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new revision is created, all addenda associated with the previous revision of the manual are incorporated into the new revision of the manual. Each new revision includes a revised copy of this print history page.
Revision B (Document Number 141-1199) ............................................................................ November 1999
Revision C (Document Number 141-102002) ........................................................................... October 2002
Revision D (Document Number 141-082005) ............................................................................. August 2005
Revision E (Document Number 141-062008) .................................................................................. June 2008
Revision F (Document Number 141-092008) ........................................................................ September 2008
Revision G (Document Number 141-022009) ......................................................................... February 2009
Revision H (Document Number 141-102009) ........................................................................... October 2009
Revision J (Document Number 141-102009A) .......................................................................... October 2009
Revision K (Document Number 141-082010) ............................................................................. August 2010
Revision L (Document Number 141-072011) ................................................................................... July 2011
Visit www.teledyne-hi.com for WEEE disposal guidance.
CAUTION:
CAUTION:
CAUTION:
The ins truments des cribed in this manual are available with multiple pin-outs .
Ensure that all electrical connections are correct.
The ins truments des cribed in this manual are des igned for INDOOR us e only .
The ins truments described in this manual are des ig ned for Clas s 2 installations
in accordanc e with IPC s tandards
Hastings Instruments reserves the right to change or modify the design of its equipment without
any obligation to provide notification of change or intent to change.
141-072011 - 201/203 Series Page 2 of 23
Table of Contents
1. GENERAL INFORMATION ............................................................................................................................................ 4
1.1. FEATURES .................................................................................................................................................................... 4
1.2. SPECIFICATIONS HFM-201/205 ................................................................................................................................... 5
1.3. SPECIFICATIONS HFC-203/207 .................................................................................................................................... 5
1.4. OPTIONAL 4 - 20 MA CURRENT OUTPUT ...................................................................................................................... 6
1.5. OTHER ACCESSORIES ................................................................................................................................................... 6
2. INSTALLATION AND OPERATION ............................................................................................................................. 8
2.1. RECEIVING INSPECTION ............................................................................................................................................... 8
2.2. POWER REQUIREMENTS ............................................................................................................................................... 8
2.3. OUTPUT SIGNAL........................................................................................................................................................... 8
2.4. MECHANICAL CONNECTIONS ....................................................................................................................................... 9
2.5. ELECTRICAL CONNECTIONS ......................................................................................................................................... 9
2.6. OPERATION ................................................................................................................................................................ 10
2.7. RANGE CHANGING: .................................................................................................................................................... 14
2.8. VALVE-OVERRIDE CONTROL ..................................................................................................................................... 14
3. THEORY OF OPERATION ........................................................................................................................................... 15
3.1. OVERALL FUNCTIONAL DESCRIPTION: ...................................................................................................................... 15
3.2. SENSOR: ..................................................................................................................................................................... 15
3.3. ELECTRONICS: ........................................................................................................................................................... 15
3.4. SHUNT: ...................................................................................................................................................................... 16
3.5. VALVE: ...................................................................................................................................................................... 17
4. MAINTENANCE .............................................................................................................................................................. 18
4.1. AUTHORIZED MAINTENANCE ..................................................................................................................................... 18
4.2. TROUBLESHOOTING ................................................................................................................................................... 18
4.3. ADJUSTMENTS ........................................................................................................................................................... 19
4.4. END CAP REMOVAL: .................................................................................................................................................. 20
4.5. PRINTED CIRCUIT BOARD REPLACEMENT .................................................................................................................. 20
4.6. SENSOR REPLACEMENT: ............................................................................................................................................ 20
4.7. ORIFICE CHANGES: .................................................................................................................................................... 20
5. WARRANTY .................................................................................................................................................................... 23
5.1. WARRANTY REPAIR POLICY ...................................................................................................................................... 23
5.2. NON-WARRANTY REPAIR POLICY ............................................................................................................................. 23
141-072011 - 201/203 Series Page 3 of 23
1. General Information
The Hastings HFM-201/HFM-205/HFC-203/HFC-207 series Mass flow meters (HFM-201/HFM-
205) and controllers (HFC-203/HFC-207) are designed to accurately measure and control mass flow over the range of 30 slm to 2500 slm, without corrections or compensations for gas pressure and temperature with an accuracy of better than ±1% from the mean (±2% FS for 201/203 units in the ½” fitting size base with flows >500 slm). Hastings mass flow instruments do not require any periodic maintenance under normal operating conditions with clean gases. No damage will occur from the use of moderate overpressures (~500 psi/3.45 MPa) or overflows. Instruments are normally calibrated with the appropriate standard calibration gas (nitrogen) then a correction factor is used to adjust the output for the intended gas. Special calibrations for other gases, such as oxygen, helium and argon, are available upon special order.
1.1. Features
LINEAR BY DESIGN. The Hastings 200 series flow instrument is inherently linear (no linearization
circuitry is employed). Should recalibration in the field be desired (a calibration standard is required), the customer needs to simply set the zero and span points. There will be no appreciable linearity change of the instrument when the flowing gas is changed.
MODULAR SENSOR. The Hastings 200 series flow instrument incorporates a removable/replaceable
sensor module. Field repairs to units can be achieved with a minimum of production line downtime.
METER SETTLING TIME. Changes in flow rate for the HFM-201/HFM-205 are detected in less than 2
seconds when using the speed up circuitry.
LOW TEMPERATURE DRIFT. The temperature coefficient of span for the Hastings 200 series flow
instrument is typically less than 0.05% of full scale/° C from 15° - 45° C. The temperature coefficient of zero is typically less than 0.1% of reading/° C from 0° - 50° C.
FIELD RANGEABLE. The HFM-201/HFC-203 series is available in ranges from 30 slm to 1100 slm.
The HFM-205/HFC-207 series is available in ranges from 1000 slm to 3000 slm. For HFC-203 & HFC­207 controllers, an orifice change is required as well. Calibration is required after all changes.
CURRENT LOOP. The 4 - 20 mA option gives the user the advantages of a current loop output to
minimize environmental noise pickup.
141-072011 - 201/203 Series Page 4 of 23
1.2. Specifications HFM-201/205
Accuracy1 and Linearity ....................................................................................................... ±1% F.S.
Repeatability .................................................................................................................. ±0.05% F.S.
Standard Pressure Rating ...................................................................................................... 500 psig
Pressure Coefficient ........................................................... +0.0067%/psi (0 - 1000 psig N
High-Pressure Option ................................................................................ Proof tested to 1500 psig
Leak Integrity ................................................................................................................. < 1x10
Temperature Coefficient ........................................................... Zero: ±0.035% FS/° C (0° - 60° C)
.................................................................................................... Span: ±0.05% FS/° C (0° - 60° C)
STP ..................................................................................................................... 0° C and 760 Torr
Power (±15 Volt flow meter) ............................................. ±(14 - 16) VDC @ ±30 mA (< 1 Watt)
Power (24 Volt flow meter) ................................................................... (14 - 32) VDC < 1.9 Watts
Flow Signal ............................................. (inherently linear) 0 - 5.00 VDC, 0 - 20 mA or 4 - 20 mA
Wetted Material
2
............ 302 SS, 316 SS, Viton®, 82/18 Au/Ni Braze, Ni 200, Trace Silver Solder
Connector ....................................................................... 15-pin subminiature D (9 pin for 24 Volt)
®
Fittings½-in. Swagelok
- 201/203 small, 3/4-in. Swagelok® - 201/203 large, 1” Swagelok® - 205/207
Weight (approx.) ................... 201L: 3.55 lb (1610 g), 201H: 3.25 lb (1470g), 205: 8.1 lb (3650 g)
1.3. Specifications HFC-203/207
) typical
2
-9
sccs
Accuracy1 and Linearity ...................................................................................................... ±1% F.S.
Repeatability .................................................................................................................. ±0.05% F.S.
Std. Pressure Rating ............................................................................................................. 500 psig
High Pressure Option ................................................................................ Proof tested to 1500 psig
Pressure Coefficient .......................................................... +0.0067%/psi (0 - 1000 psig N2) typical
Control Valve DP* ............................................................................................. Per customer order
-9
Leak Integrity ................................................................................................................ < 1x10
sccs
Temperature Coefficient ........................................................... Zero: ±0.035% FS/° C (0° - 60° C)
..................................................................................................... Span: ±0.05% FS/° C (0° - 60° C)
STP ..................................................................................................................... 0° C and 760 Torr
Power (±15 Volt flow meter) ............................................. ±(14 - 16) VDC @ ±30 mA (< 1 Watt)
Power (24 Volt flow meter) ................................................................... (14 - 32) VDC < 1.9 Watts
Flow Signal ............................................................... (inherently linear) 0 - 5.00 VDC or 4 - 20 mA
Command Signal ................................................................................... 0 - 5.00 VDC or 4 - 20 mA
2
Wetted Material
........................................................... 316 SS, Nickel, Viton, 82/18 Au/Ni Braze,
.............................................................................................................. Trace Silver Solder, Kalrez®
Connector ....................................................................... 15-pin subminiature D (9 pin for 24 Volt)
®
Fittings½-in. Swagelok
- 201/203 small, 3/4-in. Swagelok® - 201/203 large, 1” Swagelok® - 205/207
Weight (approx.) ................... 203L: 5.2 lb (2360 g), 203K: 5.15 lb (2340g), 207: 14.9 lb (6760 g)
1 Stated accuracy is for nitrogen or other gas specific calibration and using this gas only. 2 Other materials are available. Viton is the standard O-ring option. * Consult factory for pressure ranges.
141-072011 - 201/203 Series Page 5 of 23
1.4. Optional 4 - 20 mA Current Output
An option to the standard 0 - 5 VDC output is the 4 - 20 mA current output that is proportional to flow. The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flow meter. The current loop output is useful for remote applications where pickup noise could substantially affect the stability of the voltage output or long cable runs where cable resistance would cause a voltage signal to decay.
The current loop signal replaces the voltage output on pin 6 of the DA-15 connector (pin 2 of DE-9 24 Volt). The current loop is sourced internally from the +15 VDC supply and must be returned to the signal supply common after (pin 5 [±15VDC] or pin 8 [24 VDC]) passing through the monitoring circuitry to complete the loop. The load must be between 0 and 600 Ohm.
The 4 - 20 mA I/O option can accept a current input for flow controllers. The 0 - 5 VDC command signal on pin 14 (pin 3 for 24 Volts) can be replaced by a 4 - 20mA command signal. The loop presents an impedance of 50 Ohms and is returned to the signal common. On the 24 Volt units, the signal common will need to be tied to the power supply common external to the flow controller to complete the loop since the signal common is galvanically isolated internally from the supply common. This line must be current limited (<50 mA) to prevent short circuit currents from damaging the receiving circuit in the flow controller.
24 Volt Supply
±15 Volt Supply
1.5. Other Accessories
1.5.1. Hastings Power Supplies
Hastings power supply/display units are available in one and four channel versions. They convert 100, 115 or 230VAC to the ±15 VDC required to operate the flow meter and provide a digital indication of the flow rate. Interface terminals for the retransmission of the flow meter analog output signal are located on the rear of the panel.
The THCD-100 and the THPS-100/400 models are built with controllers in mind but will work with meters as well. The Model 40 is for flow meters only. Throughout this manual, when reference is made to a power supply, it is assumed the customer is using a Hastings power supply. Hastings THCD and THPS power supplies are CE marked, but the Model 40 does not meet CE standards at this time. The Model 40, The THCD and THPS-100 are not compatible with 4 – 20 mA analog signals. With the THPS-400, individual channels’ input signals, as well as their commands, become 4 – 20 mA compatible when selected. The THPS-400 also sports a Totalizer
141-072011 - 201/203 Series Page 6 of 23
feature. More information about Hastings Instruments’ power supplies can be found on the Hastings web site. http://www.teledyne-hi.com/products/powerpod-series.htm
1.5.2. Interconnecting Cables
Cables are available from Hastings, in various lengths, to connect from the 15 pin "D" connector on the back of Hastings power supplies directly to any of the 200 series and 300 series flow instruments (including digital versions). More information about the available cables can be found in the THPS 400 bulletin on the Hastings web site. http://www.teledyne-hi.com/pdfs/bulletins.htm
141-072011 - 201/203 Series Page 7 of 23
Loading...
+ 16 hidden pages