Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
SAFETY MESSAGES
Your safety and the safety of others is very important. We have provided many important safety
messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety
message is associated with a safety alert symbol. These symbols are found in the manual and
inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for details on
the specific danger.
CAUTION: Hot Surface Warning
CAUTION: Electrical Shock Hazard
Technician Symbol: All operations marked with this symbol are to be
performed by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central
safety grounding point for the instrument.
CAUTION
The analyzer should only be used for the purpose
and in the manner described in this manual.
If you use the analyzer in a manner other than that for which
it was intended, unpredictable behavior could ensue with
possible hazardous consequences.
1
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
TABLE 3.2–M501NHCONVERTER REBUILD PARTS LIST......................................................................... 19
4
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
5
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
1.0 M201A AMMONIA ANALYZER
This manual is to be used in conjunction with the M200A NOx Analyzer manual p/n
02246 supplied with this instrument.
The M201A Ammonia Analyzer consists of a modified M200A Nitrogen Oxides Analyzer
and M501NH Ammonia Converter:
A functional block diagram of the system is shown in Figure 1.
The instrument comes in three pneumatic configurations:
• The basic system is shown in Figure 3;
• The M201A with Z/S valves is shown in Figure 4;
• The M201A with External Calibrator Option (ECO) is shown in Figure 5.
Figure 2 can be used to connect the M201A system for installation. Please note that for a
M201A with either a Zero/Span or ECO option, you must connect the umbilical cord,
02255, between the analyzer and the converter assembly.
1.1 Principle of Operation
The Teledyne API Model 201A Analyzer measures ammonia by converting NH3 to nitric
oxide by the following reaction:
4NH
The resulting NO is then measured by the chemiluminescent reaction of NO with ozone.
Consult Section 4.1 in the M200A manual (provided with this instrument) for more
details on NO measurement.
A molybdenum converter is continuously in the pneumatic pathway, providing the TN
channel signal. The
converters are in series, creating the TNx signal.
The ammonia concentration is computed by the difference between the TNx and TN
channels, as shown in the following equation:
NH
See the pneumatic diagram in Figure 1.
+ 5O2 → 4NO + 6H2O
3
M501NH ammonia converter is periodically switched so both
→ TNx - TN
3
The actual formula for computation of ammonia concentration is more complicated than
the above equation, as it includes converter efficiencies, plus sample temperature and
pressure. The difficulty in ammonia measurement is that the molybdenum converter
6
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
converts some of the ammonia to NO, thus the TN channel data contains a small
ammonia signal. Secondly, the ammonia converter does not convert all of the ammonia
or ambient NO2 to NO. Finally, the high temperature in the ammonia converter causes
the oxygen and nitrogen in ambient air to react, creating a background of NOx which
needs to be factored into computing the result. Fortunately, all of the above
shortcomings have been overcome in the M201A by careful calibration, and calculation
of the result. .
Periodically, the AutoZero valve switches allowing the analyzer to read zero
background. The AutoZero readings are subtracted from all of the other readings. This
improves zero baseline stability.
As with many chemical reactions the conversion of ammonia in the presence of other
oxides of nitrogen is complicated.
In the molybdenum converter operating at 315
o
C, the following significant reactions are
taking place:
Mo + NO
+ Mo + O2 → MoO3 + NO <10% Eff
NH
3
→ MoO3 + NO ~100% Eff
2
Notice in the second reaction some of the ammonia is converted to NO by the
molybdenum metal.
The M501NH NH3 Converter operates at 825
o
C. At this high temperature, several
reactions occur:
NO → NO Efficiency = ~100 %
NOx → NO Efficiency = ~85%
NH3 → NO Efficiency = ~85%
N2 + O2 → NO Efficiency = <1%
As can be seen from the above reactions, the calculation of the ammonia concentration
and overall calibration of the instrument must be done carefully if accurate ammonia
concentrations are to be measured. The calibration procedure is covered in Section 2.
1.1.1 Special Considerations for Ammonia Measurement
Ammonia is a difficult gas to measure due to its chemical characteristics. Because of
its large dipole moment, the gas tends to adsorb on surfaces and diffuse into many
materials. The following precautions should be observed when designing an ammonia
sampling system and connecting the M201A analyzer to the system:
Materials:
Do NOT use TEFLON tubing or fittings in contact with NH3 sample or span gas.
7
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
Do NOT use copper tubing or fittings designed for household plumbing.
Use ONLY Chromatography grade (cleaned, passivated) stainless steel tubing
Use ONLY Glass tubing for sample inlet manifold.
These rules apply also to your calibrator’s internal plumbing. Verify that the calibrator
pneumatic pathway that contacts ammonia only uses stainless steel or glass tubing.
The M702SS calibrator supplied with the External Calibration Option is built with
stainless steel fittings and tubing for all internal plumbing wetted by NH3.
Typical ambient NH3 levels are < 15 ppb, unless there is some large source of
ammonia nearby, so the sampling and analysis system must be carefully designed and
well maintained.
1.1.2 Sample Filtration
The M201A does not come with a sample filter installed. Depending on the options
ordered, a sample filter may be separately packaged with the instrument. If the filter is
installed, the instrument response time is significantly increased, due to its large surface
area. Also, the condition of the filter (particulate loading and type) can adversely affect
the response time.
For minimum response time operation, the instrument should be supplied with clean
sample gas at ambient pressure.
If the sample filter is not included, the only additional filter in the M201A pneumatic
pathway is a ¼” dia stainless steel sintered filter used to protect the flow control orifice
from plugging. This filter will plug rapidly if the sample gas is not filtered before entering
the analyzer. Some means of filtering the sample gas, such as an external stainless
steel filter should be added upstream of the analyzer.
8
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
1.2 M201A Analyzer Specifications
Ranges 0-50 to 0-2000 ppb in 1 ppb increments
Range Modes Single, Independent, AutoRange
Noise at Zero
Noise at Span
Lower Detectable Limit 1 ppb RMS
Zero Drift
Span Drift < 1.0% FS Range / 7 days
Lag Time 40 sec
Rise Time 90% 120 sec
Fall Time 90% 120 sec
Sample Flow Rate 1000 cc/min
Linearity 2% of full scale
Precision 1.0% of reading
Temp Range 10-30oC
Dimensions HxWxD 7” x 17” x 23.6” (18cm x 43cm x 61cm)