All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free
from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgments provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s
process conditions.
Teledyne Analytical Instrumentsii
Thermal Conductivity Analyzer
Table of Contents
Specific Model Information................................. iv
Part I: Control Unit, Model 2010B .........Part I: 1-1
Part II: Analysis Unit, Model 2010B......Part II: 1-1
The instrument for which this manual was supplied may incorporate one or more
options not supplied in the standard instrument. Commonly available options are listed
below, with check boxes. Any that are incorporated in the instrument for which this
manual is supplied are indicated by a check mark in the box.
Instrument Serial Number: _______________________
Options Included in the Instrument with the Above Serial Number:
❑❑
❑ C:Auto Calibration valves (zero/span/sample) built-in gas control
❑❑
valves are electronically controlled to provide synchronization
with the analyzer’s operations.
❑❑
❑ G:Stainless steel cell block with nickel filaments and Stainless
❑❑
Steel fittings and tubing.
❑❑
❑ H:Stainless steel cell block with gold filaments for corrosive gas
❑❑
streams and Stainless Steel fittings and tubing.
❑❑
❑ K:19" Rack Mount option with one or two analyzer Control
❑❑
Units installed and ready to mount in a standard rack.
❑❑
❑ K2:19" Rack Mount option with two Control Units mounted.
❑❑
❑❑
❑ K3:19" Rack Mount option with one Control Unit mounted and a
❑❑
blank cover installed in the second Control Unit location.
❑❑
❑ L:Gas selector panel consisting of sample/ref flow meters and
❑❑
control valves for metering input of sample/calibrations support
gases.
❑❑
❑ F:Flame Arrestors for Class 1, Div. 1, Groups C/D service.
❑❑
❑❑
❑ P:Flame Arrestors for Class 1, Div. 1, Groups C/D service, and
❑❑
Auto Cal valves option (Ref. C above) and GP use.
❑ ❑
❑ Q:Flame Arrestors for Group B (hydrogen) service, and Auto Cal
A-2 Recommended 2-Year Spare Parts List......................... A-3
A-3 Drawing List ................................................................... A-4
Teledyne Analytical Instruments
vii
Model 2010B
D ANGER
COMBUSTIBLE GAS USAGE WARNING
The customer should ensure that the principles of operating of
this equipment are well understood by the user. Misuse of this
product in any manner, tampering with its components, or
unauthorized substitution of any component may adversely
affect the safety of this instrument.
Since the use of this instrument is beyond the control of
Teledyne, no responsibility by Teledyne, its affiliates, and
agents for damage or injury from misuse or neglect of this
equipment is implied or assumed.
Teledyne Analytical Instrumentsviii
Thermal Conductivity AnalyzerPart I: Control Unit
Introduction
1.1Overview
The Model 2010 is a family of split configuration conductivity analyzers.
Each analyzer consist of a Control Unit suitable for installation in a general
purpose area-enclosure rated NEMA-4, and a Analysis Unit which is housed in
an explosion-proof enclosure. The Analysis Unit enclosure is rated for NEMA
4/7 Class I, Div. 1, Groups B,C,D and is approved by U/L and CSA.
The Analytical Instruments Model 2010B Thermal Conductivity Analyzer is
a versatile microprocessor-based instrument for measuring a component gas in a
background gas, or in a specific mixture of background gases. It compares the
thermal conductivity of a sample stream with that of a reference gas of known
composition. The 2010B can—
•measure the concentration of one gas in a mixture of two gases.
•measure the concentration of a gas in a specific mixture of background gases.
•measure the purity of a sample stream containing a single impurity or a mixture of impurities.
The standard 2010B is pre-programmed with automatic linearization
algorithms for a large number of gases and gas mixtures. The factory can add to
this data base for custom applications, and the sophisticated user can add his
own unique applications.
This manual section covers the Model 2010B General Purpose flush-panel
and rack-mount control units only.
Many of the Model 2010B features covered in this manual are optional,
selected according to the customers specific application. Refer to the specific
model information sheet (page IV) for the options incorporated in the instrument.
Teledyne Analytical Instruments
Part I 1-1
1 Introduction Model 2010B
1.2Typical Applications
A few typical applications of the Model 2010B are:
•Power generation
•Air liquefaction
•Chemical reaction monitoring
•Steel manufacturing and heat treating
•Petrochemical process control
•Quality assurance
•Refrigeration and storage
•Gas proportioning control.
1.3Main Features of the Analyzer
The main features of the Model 2010B Thermal Conductivity Analyzer
include:
•Three independent, user definable, analysis ranges allow up to
three different gas applications with one concentration range
each, or up to three concentration ranges for a single gas application, or any combination.
•Special recalibration range for multiple applications. Recalibrating one, recalibrates all.
•Automatic, independent linearization for each range.
•Auto Ranging allows analyzer to automatically select the proper
preset range for a given single application. Manual override
allows the user to lock onto a specific range of interest.
•RS-232 serial digital port for use with a computer or other digital
communications device.
•Six adjustable concentration set points with two alarms and a system
failure alarm relay.
•Extensive self-diagnostic testing, at startup and on demand.
•A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•High resolution, accurate indication of target or impurity gas
concentration from large, bright, meter readout. (0-9999 ppm
through 0-100 % depending on types of gas involved.)
•Standard, proven sensor cell design.
Teledyne Analytical Instruments1-2 Part I
Thermal Conductivity AnalyzerPart I: Control Unit
•Wide range of custom applications, ranges, and linearization.
•Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•Auto and remote calibration capabilities.
•Four analog outputs: two for measurement (0–1 V dc and Isolated 4–20 mA dc) and two for range identification.
•Compact and versatile design: Small footprint, yet internal components are accessible.
1.4Model Designations
The Model 2010B is ordinarily custom programmed at the factory to fit the
customer’s application. Many parameters, including the number of channels, the
gas application, the materials specification of the sampling system, and others,
are options. The most common options, are covered in this manual. See the
Specific Model Information checklist in the front matter of this manual for those
that apply to your Model 2010B analyzer. Some standard models that are not
covered in this manual are listed here.
Models 2000A:Both analysis section and control unit are in a single
general purpose enclosure.
Models 2020:Both the analysis section and control unit are in a single
explosion proof enclosure.
1.5Operator Interface (Front Panel)
The standard 2010B is housed in a rugged metal case with all controls and
displays accessible from the front panel. See Figure 1-1. The front panel has
thirteen buttons for operating the analyzer, a digital meter, and an alphanumeric
display. They are described briefly here and in detail in the Operations chapter
of this manual.
Function Keys: Six touch-sensitive membrane switches are used to
change the specific function performed by the analyzer:
•AnalyzePerform analysis for target-gas content of a sample
gas.
•SystemPerform system-related tasks (described in detail in
chapter 4, Operation.).
Teledyne Analytical Instruments
Part I 1-3
1 Introduction Model 2010B
Outer Door
(O p e n)
Teledy n e A nalytic al In s tr u ments
Viewing
Window
Outer Door
Latch
Thermal Conductivity Analyzer
0.0 % Anlz
AL-1
2010B
Figure 1-1: Model 2010B Front Panel
Digital
Meter
LCD
Screen
Inn e r D o o r
Latch
(Pressing the latch button
will open the inner Door)
Control
Panel
•SpanSpan calibrate the analyzer.
•ZeroZero calibrate the analyzer.
•AlarmsSet the alarm setpoints and attributes.
•RangeSet up the user definable ranges for the instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD (Vacuum Fluorescent
Display) display:
•Left & Right ArrowsSelect between functions currently
displayed on the VFD screen.
•Up & Down ArrowsIncrement or decrement values of
functions currently displayed.
•EnterMoves VFD on to the next screen in a series. If none
remains, returns to the
Analyze
screen.
•Escape Moves VFD back to the previous screen in a series. If
none remains, returns to the
Analyze
screen.
Teledyne Analytical Instruments1-4 Part I
Thermal Conductivity AnalyzerPart I: Control Unit
Digital Meter Display: The meter display is a VFD device that
produces large, bright, 7-segment numbers that are legible in any lighting. It
produces a continuous trace readout from 0-9999 ppm or a continuous percent
readout from 1-100 %. It is accurate across all analysis ranges.
Alphanumeric Interface Screen: The VDF screen is an easy-to-use
interface between operator and analyzer. It displays values, options, and
messages that give the operator immediate feedback.
Standby Button: The Standby turns off the display and outputs,
but circuitry is still operating.
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When
chassis is exposed or when access door is open
and power cable is connected, use extra care to
avoid contact with live electrical circuits.
Access Door: For access to the thermal conductivity sensor or the front
panel electronics, the front panel swings open when the latch in the upper right
corner of the panel is pressed all the way in with a narrow gauge tool. Accessing
the main electronics circuit board requires unfastening rear panel screws and
sliding the electronics drawer out of the case. (See chapter 5.)
1.6 Recognizing Difference Between LCD &
VFD
LCD (Liquid Crystal Display) has GREEN background with BLACK
characters. VFD has DARK background with GREEN characters. In the case
of VFD (Vacuum Fluorescent Display) - NO CONTRAST ADJUSTMENT IS
NEEDED.
1.7Equipment Interface (Rear Panel)
The rear panel, shown in Figure 1-2, contains the electrical connectors for
external input and output. The connectors are described briefly here and in detail
in chapter 3, Installation.
Teledyne Analytical Instruments
Part I 1-5
1 Introduction Model 2010B
Figure 1-2: Model 2010B Rear Panel
•Power Connection85-250 V AC power source.
•Analog Outputs0-1 V dc concentration plus 0-1 V dc
range ID, and isolated 4-20 mA dc plus
4-20 mA dc range ID.
Teledyne Analytical Instruments1-6 Part I
Thermal Conductivity AnalyzerPart I: Control Unit
•Alarm Connections2 concentration alarms and 1 system
alarm.
•RS-232 PortSerial digital concentration signal output
and control input.
•Remote ProbeUsed in the 2010B to interface the
external Analysis Unit.
•Remote Span/ZeroDigital inputs allow external control of
analyzer calibration.
•Calibration ContactTo notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•Network I/OSerial digital communications for local
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 2010B is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
Teledyne Analytical Instruments
Part I 1-7
1 Introduction Model 2010B
Teledyne Analytical Instruments1-8 Part I
Thermal Conductivity AnalyzerPart I: Control Unit
Operational Theory
2.1Introduction
The analyzer is composed of two subsystems:
1. Thermal Conductivity Sensor
2. Electronic Signal Processing, Display and Control.
The sensor is a thermal conductivity comparator that continuously
compares the thermal conductivity of the sample gas with that of a reference
gas having a known conductivity.
The electronic signal processing, display and control subsystem simplifies operation of the analyzer and accurately processes the sampled data. A
microprocessor controls all signal processing, input/output, and display
functions for the analyzer.
2.2Sensor Theory
For greater clarity, Figure 2-1 presents two different illustrations, (a)
and (b), of the operating principle of the thermal conductivity cell.
2.2.1Sensor Configuration
The thermal conductivity sensor contains two chambers, one for the
reference gas of known conductivity and one for the sample gas. Each
chamber contains a pair of heated filaments. Depending on its thermal
conductivity, each of the gases conducts a quantity of heat away from the
filaments in its chamber. See Figure 2-1(a).
The resistance of the filaments depends on their temperature. These
filaments are parts of the two legs of a Wheatstone bridge circuit that unbalances if the resistances of its two legs do not match. See Figure 2-1(b).
If the thermal conductivities of the gases in the two chambers are
different, the Wheatstone bridge circuit unbalances, causing a current to flow
in its detector circuit. The amount of this current can be an indication of the
amount of impurity in the sample gas, or even an indication of the type of
gas, depending on the known properties of the reference and sample gases.
The temperature of the measuring cell is regulated to within 0.1 °C by a
sophisticated control circuit. Temperature control is precise enough to compensate for diurnal effects in the output over the operating ranges of the
analyzer. (See Specifications in the Appendix for details.)
2.2.2Calibration
Because analysis by thermal conductivity is not an absolute measurement, calibration gases of known composition are required to fix the upper
and lower parameters (“zero” and “span”) of the range, or ranges, of analysis. These gases must be used periodically, to check the accuracy of the
analyzer.
During calibration, the bridge circuit is balanced, with zero gas against
the reference gas, at one end of the measurement range; and it is sensitized
with span gas against the reference gas at the other end of the measurement
range. The resulting electrical signals are processed by the analyzer electronics to produce a standard 0-1V, or an isolated 4–20 mA dc, output signal, as
described in the next section.
2-2 Part I
Teledyne Analytical Instruments
Thermal Conductivity AnalyzerPart I: Control Unit
2.2.3Effects of Flowrate and Gas Density
Because the flowrate of the gases in the chambers affects their cooling
of the heated filaments, the flowrate in the chambers must be kept as equal,
constant, and low as possible.
When setting the sample and reference flowrate, note that gases lighter
than air will have an actual flowrate higher than indicated on the flowmeter,
while gases heavier than air will have an actual flowrate lower than indicated. Due to the wide range of gases that are measured with the Thermal
Conductivity Analyzer, the densities of the gases being handled may vary
considerably.
Then, there are limited applications where the reference gas is in a
sealed chamber and does not flow at all. These effects must be taken in
consideration by the user when setting up an analysis.
2.2.4Measurement Results
Thermal conductivity measurements are nonspecific by nature. This fact
imposes certain limitations and requirements. If the user intends to employ
the analyzer to detect a specific component in a sample stream, the sample
must be composed of the component of interest and one other gas (or specific, and constant, mixture of gases) in order for the measured heat-transfer
differences to be nonambiguous.
If, on the other hand, the user is primarily interested in the purity of a
process stream, and does not require specific identification of the impurity,
the analyzer can be used on more complex mixtures.
2.3Electronics and Signal Processing
The Model 2010B Thermal Conductivity Analyzer uses an 8031
microcontroller, Central Processing Unit—(CPU) with 32 kB of RAM and
128 kB of ROM to control all signal processing, input/output, and display
functions for the analyzer. System power is supplied from a universal power
supply module designed to be compatible with any international power
source. (See Major Internal Components in chapter 5 Maintenance for the
location of the power supply and the main electronic PC boards.)
The signal processing electronics including the microprocessor, analog
to digital, and digital to analog converters are located on the Motherboard at
the bottom of the case. The Preamplifier board is mounted on top of the
Motherboard as shown in the figure 5.4. These boards are accessible after
Teledyne Analytical Instruments
Part I 2-3
2 Operational Theory Model 2010B
removing the back panel. Figure 2-2 is a block diagram of the Analyzer
electronics.
Thermistor
Temperature
Control
Temperature
Control
Sensor
Heater
Heater
Differential
Amplifier
Fine
Adjustment
Selt-Test Signal to MUX
Variable
Gain
Amplifier
Auto-
Range
Coarse
Adjustment
M
U
X
A to D
Converter
Digitial to
Analog
Converter
(DAC)
To CPU
0-1 V dc
Concentration
and Range
4-20 m A dc
Concentration
and Range
Alarm 1
Alarm 2
System
Failure
Alarm
2-4 Part I
Keyboard
Central
Processing
Unit
(CPU)
Displays
Power
Supply
Processing
A to D Conv
Figure 2-2: Block Diagram of the Model 2010B Electronics
Teledyne Analytical Instruments
RS-232
Range
Contacts (4)
External
Valve
Control
Remote Span
Control
Remote Zero
Control
Cal
Contact
Thermal Conductivity AnalyzerPart I: Control Unit
In the presence of dissimilar gases the sensor generates a differential
voltage across its output terminals. A differential amplifier converts this
signal to a unipolar signal, which is amplified in the second stage, variable
gain amplifier, which provides automatic range switching under control of
the CPU. The output from the variable gain amplifier is sent to an 18 bit
analog to digital converter.
The digital concentration signal along with input from the Gas Selector
Panel is processed by the CPU and passed on to the 12-bit DAC, which
outputs 0-1 V dc Concentration and Range ID signals. An voltage-to-current
converter provides 4-20 mA dc concentration signal and range ID outputs.
The CPU also provides appropriate control signals to the Displays,
Alarms, and External Valve Controls, and accepts digital inputs for external
Remote Zero and Remote Span commands. It monitors the power supply
through an analog to digital converter as part of the data for the system
failure alarm.
The RS-232 port provides two-way serial digital communications to
and from the CPU. These, and all of the above electrical interface signals are
described in detail in chapter 3 Installation.
2.4. Temperature Control
For accurate analysis the sensor of this instrument is temperature controlled to 50oC.
The Temperature Control keeps the temperature of the measuring cell
regulated to within 0.1 degree C. A thermistor is used to measure the temperature, and a zero-crossing switch regulates the power in a cartridge-type
heater. The result is a sensor output signal that is temperature independent.
A second temperature control system is used to maintain the Analysis
Unit internal temperature at 220C minimum.
Teledyne Analytical Instruments
Part I 2-5
2 Operational Theory Model 2010B
2-6 Part I
Teledyne Analytical Instruments
Thermal Conductivity AnalyzerPart I: Control Unit
Installation
Installation of the Model 2010B Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Testing the system.
3.1Unpacking the Analyzer
The analyzer is shipped ready to install and prepared for operation.
Carefully unpack the analyzer and inspect it for damage. Immediately report
any damage to the shipping agent.
3.2Mounting the Control Unit
The Model 2010B Control Unit is designed for bulkhead mounting in
general purpose area, NOT for hazardous environments of any type. The
Control Unit is for indoor/outdoor use.
Figure 3-1 is an illustration of the 2010B standard front panel and
mounting bezel. There are four mounting holes—one in each corner of the
rigid frame. See the outline drawing, at the back of this manual for overall
dimensions.
All operator controls are mounted on the inner control panel, which is
hinged on the left edge and doubles as a door to provide access to the internal
components of the instrument. The door will swing open when the button of
the latch is pressed all the way in with a narrow gauge tool (less than 0.18
inch wide), such as a small hex wrench or screwdriver Allow clearance for
the door to open in a 90-degree arc of radius 11.75 inches. See Figure 3-2.
Teledyne Analytical Instruments
Part I 3-1
3 Installation Model 2010B
NPT Fittings
supplied by
customer
0.0 % An lz
AL-1
Viewing
Window
HHinge
Ou ter Door
Latch
Figure 3-1: Front Panel of the Model 2010B Control Unit
3-2 Part I
1
1
.
7
5
i
n
Figure 3-2: Required Front Door Clearance
Teledyne Analytical Instruments
H
Thermal Conductivity AnalyzerPart I: Control Unit
3.3Electrical Connections (Rear Panel)
Figure 3-3 shows the Model 2010B Electrical Connector Panel. There
are terminal blocks for connecting power, communications, and both digital
and analog concentration outputs.
For safe connections, ensure that no uninsulated wire extends outside of
the connectors they are attached to. Stripped wire ends must insert completely into terminal blocks. No uninsulated wiring should be able to come in
contact with fingers, tools or clothing during normal operation.
Figure 3-3: Rear Panel of the Model 2010B
3.3.1Primary Input Power
The power terminal block and fuse receptacles are located in the same
rear panel assembly.
Teledyne Analytical Instruments
Part I 3-3
3 Installation Model 2010B
DANGER:POWER IS APPLIED TO THE INSTRUMENT'S CIR-
CUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. THE RED I/O
SWITCH ON THE FRONT PANEL IS FOR SWITCHING POWER ON OR OFF TO THE DISPLAYS AND
OUTPUTS ONLY.
NOTE:AC POWER MAY BE PRESENT ON THE RELAY
CONTACTS WHEN THE POWER CORD IS REMOVED!
The Control Unit is universal power 100-240V, 50-60 Hz. The Analysis Unit requires 110 or 220 VAC and is selectable via switch located inside
the explosion-proof enclosure.
3.3.2Fuse Installation
The fuse receptacles accepts European size fuses only (5x20mm). Both
sides of the line are fused. If the European size fuses are selected, both sides
at the line will be fused. Be sure to install the proper fuse as part of installation. (See Fuse Replacement in chapter 5, maintenance.)
3.3.3Analog Outputs
There are four DC output signal connectors with spring terminals on the
panel. There are two wires per output with the polarity noted. See Figure 3-
4. The outputs are:
0–1 V dc % of Range:Voltage rises linearly with increasing concentration,
from 0 V at 0 concentration to 1 V at full scale.
(Full scale = 100% of programmable range.)
0–1 V dc Range ID:0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3, 1 V = Cal Range.
4–20 mA dc % Range: Current rises linearly with concentration, from 4
mA at 0 concentration to 20 mA at full scale. (Full
scale = 100% of programmable range.)
4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =
Range 3, 20 mA = Range 4.
3-4 Part I
Teledyne Analytical Instruments
Thermal Conductivity AnalyzerPart I: Control Unit
Figure 3-4: Analog Output Connections
Examples:
The analog output signal has a voltage which depends on gas concen-
tration relative to the full scale of the range. To relate the signal output to the
actual concentration, it is necessary to know what range the instrument is
currently on, especially when the analyzer is in the autoranging mode.
The signal output for concentration is linear over the currently selected
analysis range. For example, if the analyzer is set on a range that was defined
as 0–10 % hydrogen, then the output would be as shown in Table 3-1.
Table 3-1: Analog Concentration Output—Example
PercentVoltage SignalCurrent Signal
HydrogenOutput (V dc)Output (mA dc)
To provide an indication of the range, the Range ID analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
RangeVoltage (V)Current (mA)Application
Range 10.2580-1 % H
Range 20.50120-10 % H
in N2
2
in N2
2
Range 30.75160-1 % H2 in Air
Range 4 (Cal)1.00200-1 % H2 in N2
3.3.4Alarm Relays
The three alarm-circuit connectors are spring terminals for making
connections to internal alarm relay contacts. Each provides a set of Form C
contacts for each type of alarm. Each has both normally open and normally
closed contact connections. The contact connections are indicated by diagrams on the rear panel. They are capable of switching up to 3 amperes at
250 V ac into a resistive load. See Figure 3-5. The connectors are:
Threshold Alarm 1:• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2:• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm:Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be defeated.
3-6 Part I
Actuates when cell can not balance during zero
calibration.
Actuates when span parameter out off its limited
parameter.
Teledyne Analytical Instruments
Thermal Conductivity AnalyzerPart I: Control Unit
Actuates when self test fails.
(Reset by pressing
press
I/O
again and any other button EXCEPT
System
Further detail can be found in chapter 4, section 4-5.
Figure 3-5: Types of Relay Contacts
3.3.5Digital Remote Cal Inputs
to resume.
I/O
button to remove power. Then
Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra-
tion. (See Remote Calibration Protocol below.)
Zero:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appropriately. See 3.3.9 Remote Probe Connector. (With the –C
option, the internal valves operate automatically.)
Span:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appropriately. See 3.3.9 Remote Probe Connector. (With the –C
option, the internal valves operate automatically.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 2010B Analyzer, the customer's controller must
monitor the Cal Relay Contact.
Zero
mode. Either side may be
Span
mode. Either side may be
Teledyne Analytical Instruments
Part I 3-7
3 Installation Model 2010B
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.
When the contact is closed, the display would display the last reading of
the gas concentration value and output signal would output the last reading
from the sample gas (SAMPLE and HOLD).
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will close quickly.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will close quickly.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Probe connector (paragraph 3.3.9) provides
signals to operate the zero and span gas valves synchronously. However, if you have the –C, -P, or -Q Internal valve
option, which includes zero and span gas inputs, the 2010B
automatically selects the zero, span and sample gas flow.
3.3.6Range ID Relays
Four dedicated Range ID relay contacts. For any single application they
are assigned to relays in ascending order. For example: if all ranges have the
same application, then the lowest range is assigned to the Range 1 ID relay,
and the highest range is assigned to the Range 3 ID relay. Range 4 is the Cal
Range ID relay.
3.3.7Network I/O
A serial digital input/output for local network protocol. At this printing,
this port is not yet functional. It is to be used in future versions of the instrument.
3-8 Part I
Teledyne Analytical Instruments
Loading...
+ 68 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.