Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
SAFETY MESSAGES
Your safety and the safety of others is very important. We have provided many important safety
messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety
message is associated with a safety alert symbol. These symbols are found in the manual and
inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for details on the
specific danger.
CAUTION: Hot Surface Warning
CAUTION: Electrical Shock Hazard
Technician Symbol: All operations marked with this symbol are to be performed
by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central safety
grounding point for the instrument.
CAUTION
The analyzer should only be used for the purpose
and in the manner described in this manual.
If you use the analyzer in a manner other than that for which
it was intended, unpredictable behavior could ensue with
possible hazardous consequences.
ii
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
10 M200AU SPARE PARTS LIST ................................................................ 10-1
APPENDIX A ELECTRICAL SCHEMATICS ..................................................A-1
vi
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
LIST OF FIGURES
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
IGURE
F
2-1: R
2-2: R
2-3: I
2-4: F
2-5: A
4-1: B
5-1: S
5-2: S
7-1: C
7-2: D
8-1: R
8-2: S
8-3: R
8-4: M
8-5: P
8-6: P
8-7: P
9-1: S
9-2: CPU B
EMOVAL OF SHIPPING SCREWS
EAR PANEL
NLET AND EXHAUST VENTING
RONT PANEL
SSEMBLY LAYOUT
LOCK DIAGRAM
AMPLE MENU TREE
ETUP MENU TREE
ALIBRATION SETUP
IAGRAM OF
EPLACING THE PARTICULATE FILTER
AMPLE PUMP ASSEMBLY
EACTION CELL ASSEMBLY
OLYBDENUM CONVERTER ASSEMBLY
NEUMATIC DIAGRAM
NEUMATIC DIAGRAM WITH ZERO/SPAN VALVE OPTION
NEUMATIC DIAGRAM WITH EXTERNAL CONVERTER OPTION
PAN CALIBRATION VOLTAGE
OARD JUMPER SETTINGS
9-3: RS-232 PIN A
9-4: V/F B
9-5: P
9-6: E
9-7: O
9-8: F
9-9: NO
9-10: NO
9-11: PMT C
9-12: H
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
LIST OF TABLES
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
ABLE
T
2-1: F
2-1: F
4-1: F
4-2: O
5-1: S
5-2: S
5-3: S
5-3: S
5-4: S
5-5: S
5-6: DAS D
5-7: C
INAL TEST AND CALIBRATION VALUES
INAL TEST AND CALIBRATION VALUES (CONTINUED
RONT PANEL STATUS
ZONE GENERATOR START-UP TIMING
AMPLE MENU
AMPLE MENU
ETUP MENU
ETUP MENU (CONTINUED
ETUP MENU
ETUP MENU
ATA CHANNEL EDITING
ALIBRATE
, S
5-8: M200AU O
5-9: S
TATUS OUTPUT PIN ASSIGNMENTS
5-10: RS-232 P
5-11: RS-232 S
5-12: RS-232 T
5-13: RS-232 C
5-14: RS-232 C
5-15: RS-232 I
NTERFACE COMMAND TYPES
5-16: RS-232 T
5-17: RS-232 W
5-18: RS-232 C
5-19: RS-232 C
5-20: RS-232 C
5-21: RS-232 D
6-1: Z
6-2: E
7-1: T
7-2: M
7-3: E
7-4: M
7-5: M
7-6: M
7-7: E
7-8: Z/S V
7-9: A
7-10: A
7-11: Z
7-12: E
ERO/SPAN VALVE OPERATION
XAMPLE OF AUTOCAL SETUP
YPES OF ZERO/SPAN CHECK AND CALIBRATION
7-18: C
8-1: P
8-2: P
9-1: T
9-1: T
9-1: T
9-1: T
9-2: F
9-2: F
9-3: S
9-4: D
9-4: D
9-4: D
9-4: D
9-5: M
9-6: T
9-6: T
9-7: M
9-8: V/F B
9-9: P
9-10: P
9-11: O
10-1: TELEDYNE API M200AU S
10-1: TELEDYNE API M200AU S
ALIBRATION QUALITY CHECK
REVENTATIVE MAINTENANCE SCHEDULE
REVENTATIVE MAINTENANCE CALENDAR
EST FUNCTIONS
EST FUNCTIONS (CONTINUED
EST FUNCTIONS (CONTINUED
EST FUNCTIONS (CONTINUED
RONT PANEL WARNING MESSAGES
RONT PANEL WARNING MESSAGES (CONTINUED
UMMARY OF DIAGNOSTIC MODES
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
INTENTIONALLY BLANK
x
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
1 HOW TO USE THIS MANUAL
The Model 200AU has been designed to provide serviceability, reliability and ease of operation.
The M200AU’s microprocessor continually checks operating parameters such as temperature,
flow, and critical voltages. The instrument’s modular design uses captive screws to facilitate
repair and ease of access. If you encounter any difficulty refer to Section 9 General
Troubleshooting Hints.
We recognize that the need for information in this manual changes as time passes. When the
instrument first arrives, it is necessary to get it up and running quickly and verify its correct
operation. As time passes, more detailed information is often required on special configurations,
calibration alternatives and other operational details. Finally there is the need for periodic
maintenance and to quickly troubleshoot problems to assure maximum reliability and data
integrity.
To address these needs, we have created three indexes to the information inside. They are:
Table of Contents:
Outlines the contents of the manual in the order the information is presented. This is a good
overview of the topics covered in the manual. There is also a list of Tables and a list of Figures.
Index to M200AU Front Panel Menus:
The Menu Index (Figure 5-1, Figure 5-2, Table 5-1 and Table 5-2) briefly describes the front
panel menus and refers you to other sections of the manual that have a detailed explanation of
each menu selection.
Troubleshooting Section 9:
The Troubleshooting Section, outlined in the Table of Contents, allows you to diagnose and
repair the instrument based on variables in the TEST menu, the results of DIAGNOSTIC tests,
and performance faults such as excessive noise or drift. The troubleshooting section also
explains the operation, adjustment, diagnosis and testing of each instrument subsystem.
If you are unpacking the instrument for the first time, please refer to Getting Started in
Section 2.
1-1
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
INTENTIONALLY BLANK
1-2
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
2 GETTING STARTED
2.1 Unpacking
1. Verify that there is no apparent shipping damage. If damage has occurred please advise
shipper first, then Teledyne API.
CAUTION
To avoid personal injury, always use two persons to
lift and carry the Model 200AU.
2. Before operation, it is necessary to remove the shipping hold-down screws. Remove the
instrument cover, then refer to Figure 2-1 for screw location.
3. Also check for internal shipping damage, and generally inspect the interior of the
instrument to make sure all circuit boards and other components are in good shape.
4. Please check the voltage and frequency label on the rear panel of the instrument for
compatibility with the local power before plugging in the M200AU.
2.2 Electrical and Pneumatic Connections
Refer to Figure 2-2 to locate the rear panel electrical and pneumatic connections.
1. Attach the pump to the Exhaust Out port on the instrument rear panel.
2. If you are connecting to a calibrator, attach a vented sample inlet line to the sample inlet
port. The pressure of the sample gas at the inlet port should be at ambient pressure. The
exhaust from the pump should be vented to atmospheric pressure. See Figure 2-3 for inlet
and exhaust line venting recommendations during calibration.
3. If desired, attach the analog output connections to a strip chart recorder and/or
datalogger. Refer to Figure 9-4 for the jumper settings for the desired analog output voltage
range. Factory default setting is 0-5VDC.
4. Connect the power cord to the correct voltage line, then turn to Section 2.3 Initial
Operation.
2-1
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
WARNING
Analyzer Exhaust –Sample Pump
For brief periods after power-up, analyzer
exhaust may contain ozone.
Make sure pump exhaust is routed to well ventilated
area at atmospheric pressure.
WARNING
Lethal voltages present inside case.
Do not operate with cover off during normal operation.
Before operation check for correct input voltage and frequency for
both the analyzer and the sample pump.
Do not operate without proper chassis grounding.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting
electrical subassemblies.
2-2
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Figure 2-1: Removal of Shipping Screws & Check for Correct Power
2-3
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Figure 2-2: Rear Panel
2-4
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Figure 2-3: Inlet and Exhaust Venting
2-5
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
2.3 Initial Operation
1. After confirming proper supply voltage, turn on the instrument power.
2. The display should immediately light, displaying the instrument type (M200AU) and the
computer's memory configuration. If you are unfamiliar with the M200AU, we recommend
that you read the overview Section 4 before proceeding. A diagram of the software menu
trees is in Figure 5-1 and Figure 5-2.
3. The M200AU requires about 30 minutes for all internal components to come up to
temperature. During this time the ozone generator power is OFF until the membrane dryer
has time to purge itself, therefore there will be no response from the instrument, even if span
gas is coming in the sample port. During this time temperatures and other conditions are out
of specification. Because many warning conditions could be displayed the software
suppresses warning conditions for 30 minutes after power up. After 30 minutes, warning
messages will be displayed until the respective warning conditions are within specifications.
Use the CLR key on the front panel to clear warning messages.
4. While waiting for instrument temperatures to come up, you can check for correct
operation by using some of the M200AU's diagnostic and test features.
5. Examine the TEST functions by comparing the values listed in Table 2-1 to those in the
display. Remember that as the instrument warms up the values may not have reached their
final values yet. If you would like to know more about the meaning and utility of each TEST
function, refer to Table 9-1. Also, now is a good time to verify that the instrument was
shipped with the options you ordered. Table 2-1 also contains the list of options. Section 6
covers setting up the options.
6. When the instrument is warmed up, re-check the TEST functions against Table 2-1. All
of the readings should compare closely with those in the Table. If they do not, see Section
9.1.1. The next task is to calibrate the analyzer. There are several ways to do a calibration,
they are summarized in Table 7-1. For instruments not equipped with the external converter
option, we recommend calibration with zero air and span gas coming in through the sample
port. The procedure is:
2-6
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Step 1 - Enter the expected NOx and NO span gas concentrations:
Step Number Action Comment
1. Press
CAL-CONC-NOX
This key sequence causes the M200AU to prompt for the
expected NO
concentration. Enter the NOx span
x
concentration value by pressing the key under each digit until
the expected value is set.
2. Press ENTR ENTR stores the expected NOx span value. This value will be
used in the internal formulas to compute subsequent NO
x
concentration values.
3. Press
CAL-CONC-NO
In the same CAL-CONC sub menu press the NO button and
enter the expected NO span value, then ENTR. As before this
value will be used in the internal formulas to compute the
subsequent NO concentration values.
4. Press EXIT-EXIT Returns instrument to SAMPLE mode.
5. Press
SETUP-RNGEMODE-SING-ENTR
6. Press
SETUP-RNGE-SET
If necessary, you may want to change ranges. Normally the
instrument is shipped in single range mode set at 500 ppb. We
recommend doing the initial checkout on the 500 ppb range.
After SETUP-RNGE-SET, enter 500 and press ENTR. The
instrument will now be in the 500 ppb range.
7. Press EXIT Returns instrument to SAMPLE mode.
2-7
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Step 2 - Calibrate the instrument:
Zero/Span Calibration Procedure
Step Number Action Comment
1. Input Zero gas Allow Zero gas to enter the sample port on the rear of the
instrument.
2. Press CAL The M200AU enters the calibrate mode from sample mode.
When the CAL button is pressed, the adaptive filter is
activated. This allows the instrument to respond rapidly to
concentration changes regardless of their magnitude.
3. Wait 10 min Wait for reading to stabilize at the zero value. If you wait less
than 10 minutes the final zero value may drift.
4. Press ZERO The ZERO button will be displayed.
5. Press ENTR Pressing ENTR actually changes the calculation equations and
zeroes the instrument.
6. Input Span Gas Switch gas streams to span gas.
7. Wait 10 min Wait for reading to stabilize at the span value. If you wait less
than 10 minutes the final span value may drift.
8. Press SPAN The SPAN button should be displayed. If it is not, check the
Troubleshooting Section 9.2.8 for instructions on how to
proceed. In certain circumstances at low span gas
concentrations (<100ppb), both the ZERO and SPAN buttons
will appear.
9. Press ENTR Pressing ENTR actually changes the calculation equations so
that the concentration displayed is the same as the expected
span concentration you entered above, thus spanning the
instrument.
10. Press EXIT Pressing EXIT returns the instrument to SAMPLE mode.
2-8
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Step 3 - Review the quality of the calibration:
Calibration Quality Check Procedure
Step Number Action Comment
1. Scroll the TEST
function menu until
the NO
SLOPE is
x
displayed.
2. Scroll the TEST
function menu until
the NO SLOPE is
displayed.
The SLOPE value for NO
not in this range, check Section 7.8 or 9. If the SLOPE value
is in the acceptable range, the instrument will perform
optimally.
The SLOPE value for NO should be 1.0 ± 0.3. If the value is
not in this range, check Section 7.8 or 9. If the SLOPE is in
the acceptable range, the instrument will perform optimally.
should be 1.0 ± 0.3. If the value is
x
NOTE:
slopes should be equal within ± 0.1.
x
x
4. Scroll the TEST
function menu until
the NO
OFFSET is
x
displayed.
The NO and NO
The M200AU will display the OFFSET parameter for the NO
equation. This number should be near zero. A value of 0.0 mV
–10 +150 mV indicates calibration in the optimal range. If the
OFFSET value is outside this range, check Section 7 or 9 for
procedures to correct the OFFSET value to near zero.
5. Scroll the TEST
function menu until
The instrument will now display the NO OFFSET value. It
should also have a value near zero (0.0 mV –10 +150 mV).
the NO OFFSET is
displayed.
Step 4 - The M200AU is now ready to measure sample gas.
2-9
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Figure 2-4: Front Panel
2-10
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Figure 2-5: Assembly Layout
2-11
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Table 2-1: Final Test and Calibration Values
Test Values
Observed
Value
Units Nominal Range Reference Section
RANGE ppb 5-2000 5.3.4
STABILITY ppb 0.001-2000 9.1.1, 9.2.5,
Table 9-1
SAMP FLW cc/min 1000 ± 100 9.3.7, Table 9-1
OZONE FL cc/min 80 ± 15 9.3.6, 9.3.7
PMT mV 0-5000 9.3.8.1, Table 9-1
PREREACT mV 0-1000 Table 9-1
HVPS V 450 - 900 constant 9.3.8.5
DCPS mV 2500 ± 200 9.3.5, 9.3.4
RCELL TEMP
BOX TEMP
PMT TEMP
MOLY TEMP
o
C 40 ± 1 9.3.8.2
o
C 8-48 9.3.4.1
o
C -5 ± 1 9.3.8.4
o
C 315 ± 5 9.3.4.1
RCEL PRES IN-Hg-A 1 - 4 constant 9.3.7
SAMP PRES IN-Hg-A 25 - 30 constant 9.3.7
Electric Test & Optic Test
Electric Test
PMT Volts mV 2000 ± 500 9.1.3.2
NO Conc PPB 1000 ± 250 9.1.3.2
NOx Conc PPB 1000 ± 250 9.1.3.2
Optic Test
PMT Volts mV 2000 ± 1000 9.1.3.3
NO Conc PPB 1000 ± 500 9.1.3.3
NOx Conc PPB 1000 ± 500 9.1.3.3
(table continued)
2-12
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
Table 2-1: Final Test and Calibration Values (Continued)
Parameter
NO Span Conc PPB 1 - 2000 Table 7-3
NOx Span Conc PPB 1 - 2000 Table 7-3
NO Slope -
NOx Slope -
NO Offset mV -10 to +150 Table 7-18
NOx Offset mV -10 to +150 Table 7-18
Moly Efficiency % 0.96 - 1.02 7.6.6, 5.2.2.6
Stability at Zero PPB 0.001 - 2000 Table 9-1
Stability at Span PPB < 1 @ 400 ppb Table 9-1
Sample Flow cc/min 1000 ± 100 9.3.7, Figure 9-8
Ozone Flow cc/min 80 ± 15 9.3.7, Figure 9-8
Factory Installed Options Option Installed
Power Voltage/Frequency
Observed
Value
Units Nominal Range Reference Section
1.0 ± 0.3
1.0 ± 0.3
Measured Flows
Table 7-18
Table 7-18
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Rack Mount, External Pump w/ Slides
Rack Mount, External Pump w/o Slides
Stainless Zero/Span Valves
Current Loop - NOx Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - NO Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - NO2 Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - TST Chan 4-20 mA 0-20 mA Isolated Non-Isolated
PROM # Serial #
Date Technician
2-13
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
INTENTIONALLY BLANK
2-14
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
3 SPECIFICATIONS, AGENCY APPROVALS,
WARRANTY
3.1 Specifications
Ranges In 1ppb increments from 5ppb to 2000ppb
independent ranges or autoranging
Noise at Zero
Noise at Span
Lower Detectable Limit
Zero Drift
Span Drift
Lag Time 20 sec
Rise Time
Fall Time
Sample Flow Rate 1000 cc/min. ± 10%
Linearity
Precision 0.5% of reading above 50 ppb
Hydrocarbon Interference Propylene rejection ratio > 20,000:1
Temperature Range 20-30
Temp Coefficient < 0.1% per
Humidity 0-95% RH non-condensing
Voltage Coefficient < 0.1% per V
Dimensions HxWxD 7" H x 17" W x 23.6" D (18cm x 43cm x 61cm)
Weight, Analyzer 43 lbs (20 kg)
Weight, Ext Pump Pack 21 lbs (9.5 kg)
Power, Analyzer
Power, Analyzer
Power, Ext Pump 110V~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 150 watts
Power, Ext Pump
Environmental
Recorder Output 100 mV, 1, 5, 10V, isolated or non-isolated current loop
Analog Resolution 1 part in 1024 of selected voltage or current range
Status Option 12 Status Outputs from opto-isolator
Measurement Units ppb, ug/m
1
<25 ppt RMS
1
<0.25% RMS of reading above 50 ppb
2
<0.1 ppb / 24 hours
2
<0.5% FS or 50 ppt RMS whichever is greater / 7 days
3
95% in <50 sec6
3
95% in <50 sec6
1
50 ppt RMS
<0.2 ppb / 7 days
1% of full scale or ± 0.1 ppb whichever is greater
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
3.2 EPA Equivalency Designation
Teledyne Instruments Advanced Pollution Instrumentation Division, Model 200AU High
Sensitivity Nitrogen Oxides Analyzer is designated as Reference Method Number RFNA-1194099 as defined in 40 CFR Part 53, when operated under the following conditions:
1. Range: Any range from 50 parts per billion (ppb) to 1 ppm.
2. Ambient temperature range of 20 to 30oC.
3. Line voltage range of 105-125 VAC, 60Hz; 220-240 VAC, 50Hz.
4. With 5-micron TFE filter element installed in the internal filter assembly.
5. Sample flow of 1000 ± 100 cc/min.
6. Vacuum pump capable of 6"Hg Abs pressure @ 2 slpm or better.
7. Software settings:
A. Dynamic span OFF
B. Dynamic zero OFF
C. Cal-on-NO2 OFF
D. Dilution factor OFF
E. AutoCal ON or OFF
F. Independent range ON or OFF
G. Autorange ON or OFF
H. Temp/Pres compensation ON
I. Converter Efficency 0.96 to 1.02
Under the designation, the Analyzer may be operated with or without the following options:
1. Rack mount with slides.
2. Rack mount without slides, ears only.
3. Rack mount for external pump w/o tray.
4. Stainless steel zero/span valves.
5. 4-20mA, isolated outputs.
6. Status outputs.
7. RS-232 output.
8. 1 micron sample filter.
3-2
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
3.3 Warranty
WARRANTY POLICY
Prior to shipment, Teledyne API equipment is thoroughly inspected and tested. Should
equipment failure occur, Teledyne API assures its customers that prompt service and support
will be available.
COVERAGE
After the warranty period and throughout the equipment lifetime, Teledyne API stands ready to
provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in
the industry. All maintenance and the first level of field troubleshooting is to be performed by
the customer.
NON-TELEDYNE API MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by Teledyne API is warranted and will be repaired to
the extent and according to the current terms and conditions of the respective equipment
manufacturers warranty.
GENERAL
TELEDYNE API warrants each Product manufactured by Teledyne API to be free from defects
in material and workmanship under normal use and service for a period of one year from the date
of delivery. All replacement parts and repairs are warranted for 90 days after the purchase.
If a Product fails to conform to its specifications within the warranty period, Teledyne API shall
correct such defect by, in Teledyne API's discretion, repairing or replacing such defective
Product or refunding the purchase price of such Product.
The warranties set forth in this section shall be of no force or effect with respect to any Product:
(i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used
in any manner other than in accordance with the instruction provided by Teledyne API or (iii)
not properly maintained.
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES
THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF
MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER
WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE
REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR
BREACH OF ANY WARRANTY CONTAINED HEREIN. TELEDYNE API SHALL
NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES
ARISING OUT OF OR RELATED TO THIS AGREEMENT OF TELEDYNE API'S
PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR
OTHERWISE.
3-3
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F
TERMS AND CONDITIONS
All units or components returned to Teledyne API should be properly packed for handling and
returned freight prepaid to the nearest designated Service Center. After the repair, the equipment
will be returned, freight prepaid.
3-4
Loading...
+ 178 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.