The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
How to Power On 2-13 How to Use Help 2-14 How to Use the Status Menu 2-14 How to Set Functions 2-15 How to Set Complex Functions 2-18
Theory of Operation
Circuit Description 3-1
Logic Conventions 3-1 Module Overview 3-1
ii Contents
Performance Verification
Brief Procedures 4-1
Performance Tests 4-13
Table of Contents
General Instructions 4-1 Conventions 4-2 Self Tests 4-4
Verify Internal Adjustment, Self Compensation,
and Diagnostics 4-4
Functional Tests 4-6
Verify All Input Channels 4-7 Verify the Time Base 4-8 Verify the Main and Delayed Trigger Systems 4-9 Verify the File System (Optional on TDS 620A and 640A) ... 4-11
Prerequisites 4-13 Equipment Required 4-14
Test Record 4-17
Signal Acquisition System Checks 4-21
Check Accuracy of Offset (Zero Setting) 4-21 Check DC Gain and Voltage Measurement Accuracy 4-23 Check Analog Bandwidth 4-29 Check Delay Between Channels 4-33
Time Base System Checks 4-37
Check Accuracy for Long-Term Sample Rate, Delay Time,
and Delta Time Measurements 4-37
Trigger System Checks 4-40
Check Accuracy (Time) for Pulse-Glitch or Pulse-Width
Triggering 4-40 Check Accuracy, Trigger-Level or Threshold, DC Coupled .. 4-43 Sensitivity, Edge Trigger, DC Coupled 4-46
Output Signal Checks 4-51
Check Outputs—AUX1 (for TDS 620A) or CH 3 (for
TDS 640A or 644A), Main and Delayed Trigger 4-51 Check Probe Compensator Output 4-54
Option 05 Video Trigger Checks 4-59
Check Video Trigger 4-59
Adjustment Procedures
Adjustment Procedures 5-1
TDS 620A, 640A, & 644A Service Manual
Requirements for Performance 5-1
Personnel 5-1 Warm-Up Period 5-1 Access 5-2 System 5-2
iii
Table of Contents
Optional Peripherals 5-2 Test Equipment 5-2
Usage 5-2
Performing the Adjustments 5-3 Complete Adjustment 5-3 Individual Adjustments 5-3 Partial Adjustment 5-4 Adjustment After Repair 5-4 Adjustment Dependencies 5-4
List of Modules 6-10 General Instructions 6-10 Summary of Procedures 6-10
iv Contents
Table of Contents
Access Procedure 6-16 Procedures for External Modules 6-17
Front-Panel Knobs 6-17
Line Fuse and Line Cord 6-18
EMI Gaskets 6-19
Rear Cover and Cabinet 6-20
Front Cover, Trim Ring, Menu Buttons, and Attenuator Panel 6-23 A12 Front-Panel Assembly 6-25
Display-Frame Assembly 6-28 Cabinet Modules 6-29
Procedures for Outer-Chassis Modules 6-31
A15 Attenuator Assembly 6-31
Fan 6-32 A14 D1 Bus and Analog-Power and Digital-Power Cables .. 6-33 A23 SerPar Board — RS232/Centronics Hardcopy Interface 6-35 A29 Video Trigger Board 6-37 A11 Processor/Display Board 6-42 Top Cover and Board Brackets 6-44
Rear-Panel Cables 6-45 A10 Acquisition Board 6-46
Floppy Disk 6-48
Rear Chassis 6-50
Procedures for Inner-Chassis Modules 6-52
A16 Low Voltage Power Supply 6-52 A20 or A30 Display Assembly and Supply Fuse 6-53
Front Subpanel 6-58
Main Chassis 6-59
Disassembly for Cleaning 6-60
Troubleshooting 6-63
Diagnostics 6-63 Firmware Updates 6-64
TDS 620A, 640A, & 644A Service Manual ix
Table of Contents
Options
Electrical Parts List
Diagrams
Mechanical Parts List
vi Contents
List of Figures
Figure 2-1: Power-Cord Plug Identification 2-8 Figure 2-2: Map of Display Functions 2-14 Figure 4-1: Map of Display Functions 4-3 Figure 4-2: Verifying Adjustments and Signal-Path Compensation .. 4-5 Figure 4-3: Universal Test Hookup for Functional Tests 4-7 Figure 4-4: Measurement of DC Offset Accuracy at Zero Setting ... 4-23 Figure 4-5: Initial Test Hookup 4-24 Figure 4-6: Measurement of the DC Accuracy for Delta
Measurements 4-25
Figure 4-7: Measurement of DC Accuracy at Maximum Offset
and Position 4-28 Figure 4-8: Initial Test Hookup 4-29 Figure 4-9: Measurement of Analog Bandwidth 4-32 Figure 4-10: Initial Test Hookup 4-34 Figure 4-11: Measurement of Channel Delay 4-35 Figure 4-12: Initial Test Hookup 4-37 Figure 4-13: Measurement of Accuracy — Long-Term and
Delay Time 4-38 Figure 4-14: Initial Test Hookup 4-40 Figure 4-15: Measurement of Time Accuracy for Pulse and
Glitch Triggering 4-42 Figure 4-16: Initial Test Hookup 4-43 Figure 4-17: Measurement of Trigger-Level Accuracy 4-45 Figure 4-18: Initial Test Hookup 4-47 Figure 4-19: Measurement of Trigger Sensitivity 4-48 Figure 4-20: Initial Test Hookup 4-51 Figure 4-21: Measurement of Main Trigger Out Limits 4-52 Figure 4-22: Initial Test Hookup 4-54 Figure 4-23: Measurement of Probe Compensator Frequency 4-56 Figure 4-24: Subsequent Test Hookup 4-57 Figure 4-25: Measurement of Probe Compensator Amplitude 4-58 Figure 4-26: Jitter Test Hookup 4-60 Figure 4-27: Jitter Test Displayed Waveform 4-60 Figure 4-28: Jitter Test When Completed 4-62 Figure 4-29: Triggered Signal Range Test — 300 mV 4-63 Figure 4-30: Triggered Signal Range Test — 75 mV 4-64 Figure 4-31: 60 Hz Rejection Test Hookup 4-65 Figure 4-32: 60 Hz Rejection Test Setup Signal 4-65 Figure 4-33: Subsequent 60 Hz Rejection Test Hookup 4-66
TDS 620A, 640A, & 644A Service Manual vii
List of Figures
Figure 4-34: 60 Hz Rejection Test Result 4-67 Figure 4-35: Line Count Accuracy Test Hookup 4-68 Figure 4-36: Line Count Accuracy Test Setup Waveform 4-68 Figure 4-37: Line Count Accuracy Correct Result Waveform 4-69 Figure 4-38: PG502 Setup for Sync Duty Cycle Test 4-71 Figure 4-39: Sync Duty Cycle Test: One-Div Neg Pulse Waveform .. 4-72 Figure 4-40: Sync Duty Cycle Test: Critically Adjusted Pulse 4-73 Figure 5-1: Accessing the Protection Switch 5-8 Figure 5-2: Hookup for Probe Compensation 5-10 Figure 5-3: Performing Probe Compensation 5-11 Figure 5-4: Proper and Improper Probe Compensation 5-11 Figure 5-5: Exposing the Inner Probe Tip 5-12 Figure 5-6: Initial Test Hookup 5-12 Figure 5-7: Exposing the Probe Body 5-14 Figure 5-8: Initial Test Hookup 5-15 Figure 5-9: Locations of P6139A Probe Adjustments 5-17 Figure 5-10: Adjustments versus Front-Corner Response 5-17 Figure 5-11: TDS 620A/640A Five and Ten Percent Luminance
Patches 5-19
Figure 5-12: TDS 644A Five and Ten Percent Luminance Patches .. 5-22 Figure 6-1: External Modules 6-12 Figure 6-2: Outer-Chassis Modules 6-13 Figure 6-3: Inner-Chassis Modules - TDS 620A or 640A with
A20 Display Assembly 6-14
Figure 6-4: Inner-Chassis Modules - TDS 644A with A30
Display Assembly 6-15
Figure 6-5: Knob Removal 6-18 Figure 6-6: Line Fuse and Line Cord Removal 6-19 Figure 6-7: Rear Cover and Cabinet Removal 6-22 Figure 6-8: Front Cover, Trim Ring, Menu Buttons, and Attenuator
Troubleshooting Procedure 6-68 Figure 6-33: Power Supply Voltage Measurement Locations 6-70 Figure 6-34: Color (TDS 644A) Display Troubleshooting Procedure . 6-71 Figure 6-35: Horizontal and Vertical Sync Signals - Color Display . 6-72 Figure 6-36: A Video Signal with White, Black, and Blanking
Please take a moment to review these safety precautions. They are provided for your protection and to prevent damage to the Digitizing Oscilloscopes. This safety information applies to all operators and service personnel.
Symbols and Terms These two terms appear in manuals:
caut
• L
• ij««»™^ statements identify conditions or practices that could result in
These two terms appear on equipment:
• CAUTION indicates a personal injury hazard not immediately accessible
!?!L1 statements identify conditions or practices that could result in
damage to the equipment or other property.
personal injury or loss of life.
as one reads the marking, or a hazard to property including the equip-ment itself.
Safety Summary
• DANGER indicates a personal injury hazard immediately accessible as one reads the marking.
This symbol appears in manuals:
®
Static-Sensitive Devices
These symbols appear on equipment:
A
DANGER
High Voltage
Protective
ground (earth)
terminal
ATTENTION
Refer to manual
TDS 620A, 640A, & 644A Service Manual
xiii
Safety Summary
Specific Precautions Observe all of the following precautions to ensure your personal safety and
to prevent damage to either the TDS 620A, 640A, & 644A or equipment connected to it.
Do Not Perform Service While Alone
Do not perform internal service or adjustment of this product unless another person capable of rendering first aid and resuscitation is present.
Use Care When Servicing With Power On
Dangerous voltages exist at several points in this product. To avoid personal injury, do not touch exposed connections or components while power is on. Disconnect power before removing protective panels, soldering, or replacing
components.
Power Source
The TDS 620A, 640A, & 644A is intended to operate from a power source that will not apply more than 250 VRMs between the supply conductors or
between either supply conductor and ground. A protective ground connec-tion, through the grounding conductor in the power cord, is essential for safe system operation.
Grounding the Digitizing Oscilloscopes
The TDS 620A, 640A, & 644A are each grounded through the power cord. To avoid electric shock, plug the power cord into a properly wired receptacle where earth ground has been verified by a qualified service person. Do this
before making connections to the input or output terminals of the TDS 620A, 640A, & 644A.
Without the protective ground connection, all parts of the TDS 620A, 640A, & 644A are potential shock hazards. This includes knobs and controls that may appear to be insulators.
Use the Proper Power Cord
Use only the power cord and connector specified for your product. Use only a power cord that is in good condition.
Use the Proper Fuse
To avoid fire hazard, use only the fuse specified in the parts list for your
product, and which is identical in type, voltage rating, and current rating.
xiv
Safety
Safety Summary
Do Not Remove Covers or Panels
To avoid personal injury, do not operate the TDS 620A, 640A, & 644A without the panels or covers.
Do Not Operate in Explosive Atmospheres
The TDS 620A, 640A, & 644A provides no explosion protection from static discharges or arcing components. Do not operate the TDS 620A, 640A, & 644A in an atmosphere of explosive gasses.
Electric Overload
Never apply a voltage to a connector on the TDS 620A, 640A, & 644A that is outside the range specified for that connector.
TDS 620A, 640A, & 644A Service Manual
XV
Safety Summary
xvi Safety
Specifications
This subsection begins with a general description of the traits of the TDS 620A, 640A, & 644A Digitizing Oscilloscopes. Three subsections follow, one for each of three classes of traits: nominal traits, warranted characteris-tics, and typical characteristics.
General Product The TDS 620A, 640A, & 644A are portable, four-channel digitizing oscillo-Description scopes suitable for use in a variety of test and measurement applications
p
and systems. Key features include:
• A maximum digitizing rate of 2 GS/s on each of the full-featured chan-nels (four on the TDS 640A and 644A, two on the TDS 620A) simulta-neously with an analog bandwidth of 500 MHz.
• Four input channels, each with 8-bit vertical resolution, and each with a record length of 2,000 samples and 8-bit vertical resolution.
• Extensive triggering capabilities: such as edge, logic, and glitch. Video
trigger (Option 05) is also available. The video trigger modes are NTSC, SECAM, PAL, HDTV, and FlexFormat™ (user definable format).
• Acquisition modes such as sample, envelope, and average.
• A full complement of advanced functions, like continuously-updated measurements, results and local pass/fail decision making.
• Specialized display modes, such as variable persistence with color
grading, dot or vector mode, sin(x)/x or linear display filters, and user selectable color pallettes. The "Fit to Screen" features compresses the entire waveform record to fit on the screen.
• A unique graphical user interface (GUI), an on-board help mode, and a logical front-panel layout which combine to deliver a new standard in usability.
• Full GPIB software programmability. Hardcopy output using GPIB, RS-232, or Centronics ports. RS-232 and Centronics are standard on
the TDS 644A and optional, as option 13, on the TDS 620A and 640A.
• VGA output for driving remote monitors.
• A 1.44 Mbyte, DOS 3.3 or later-compatible, floppy disk drive (option 1F
on the TDS 620A and 640A) and NVRAM mass storage for saving waveforms, hardcopies, and oscilloscope setups.
• On the TDS 644A, a color display for distinguishing among waveforms,
their measurements, and associated text.
TDS 620A, 640A, & 644A Service Manual
1-1
Specifications
User Interface Use a combination of front-panel buttons, knobs, and on-screen menus to
control the oscilloscope's many functions. The front-panel controls are grouped according to function: vertical, horizontal, trigger, and special. Set a function you adjust often, such as vertical positioning or the time base setting, directly by its own front-panel knob. Set functions which you change
less often, such as vertical coupling and horizontal mode, indirectly using
selected menus.
Menus
Pressing one (sometimes two) front-panel button(s), such as vertical menu, displays a main menu of related functions, such as coupling, bandwidth, etc., at the bottom of the screen. Pressing a main-menu button, such as coupling, displays a side menu of settings for that function, such as AC, DC, or GND (ground) coupling, at the right side of the screen. Pressing a side-
menu button selects a setting such as DC.
Indicators
On-screen readouts help you keep track of the settings for various functions, such as vertical and horizontal scale and trigger level. Some readouts use the cursors or the automatic parameter extraction feature (called measure) to display the results of measurements made or the status of the instrument.
General Purpose Knob
Assign the general purpose knob to adjust a selected parameter function.
More quickly change parameters by toggling the SHIFT button. Use the same method as for selecting a function, except the final side-menu selec-tion assigns the general purpose knob to adjust some function, such as the
position of measurement cursors on screen, or the setting for a channel's fine gain.
GUI
I ^ The user interface also makes use of a GUI, or Graphical User Interface, to
make setting functions and interpreting the display more intuitive. Some
J
< menus and status are displayed using iconic representations of function
settings such as those shown here for full, 100 MHz, and 20 MHz band-
J \ width. Such icons allow you to more readily determine status or the available
settings.
7-2
Specifications
Specifications
Signal Acquisition Svstem
The signal acquisition system provides full-featured vertical channels (four on the TDS 640A and 644A, two on the TDS 620A) with calibrated vertical scale factors from 1 mV to 10 V per division. All four channels can be ac-quired simultaneously.
Each of the four channels can be displayed, vertically positioned, and offset, can have their bandwidth limited (100 MHz or 20 MHz) and their vertical coupling specified. Fine gain can also be adjusted.
Besides these channels, up to three math waveforms and four reference waveforms are available for display. (A math waveform results when you specify dual waveform operations, such as add, on any two channels. A
reference waveform results when you save a live waveform in a reference
memory.)
Horizontal Svstem There are three horizontal display modes: main only, main intensified, and
delayed only. You can select among various horizontal record length set-tings.
A feature called "Fit to Screen" allows you to view entire waveform records within the 10 division screen area. In other words, waveforms are com-
pressd to fit on the screen. (see Table 1-1)
Table 1-1: Record Length versus Divisions per Record, Samples
per Division, and Sec/Div Sequence
Divisions per Record
Record Length
Record Length
Fit to Screen OFF Fit to Screen ON 50 (1-2.5-5) 50 (1-2.5-5)
500 10 divs 10 divs
1000 20 divs 10 divs
2000 40divs 10divs
Both the delayed only display and the intensified zone on the main intensi-fied display may be delayed by time with respect to the main trigger. Both can be set to display immediately after the delay (delayed runs after main
mode). The delayed display can also be set to display at the first valid trigger after the delay (delayed-triggerable mode).
The delayed display (or the intensified zone) may also be delayed by a selected number of events. In this case, the events source is the delayed-trigger source. The delayed trigger can also be set to occur after a number of events plus an amount of time.
Sample/Division (Sec/Div Sequence)
TDS 620A, 640A, & 644A Service Manual 1-3
Specifications
Trigger System The triggering system supports a varied set of features for triggering the
• Edge (main- and delayed-trigger systems): This familiar type of trigger-
ing is fully configurable for source, slope, coupling, mode (auto or normal), and holdoff.
• Logic (main-trigger system): This type of triggering can be based on
pattern (asynchronous) or state (synchronous). In either case, logic triggering is configurable for sources, for boolean operators to apply to those sources, for logic pattern or state on which to trigger, for mode
(auto or normal), and for holdoff. Time qualification may be selected in
pattern mode.
• Pulse (main-trigger system): Pulse triggering is configurable for trigger-
ing on runt or glitch pulses, or on pulse widths or periods inside or outside limits that you specify. It is also configurable for source, polarity,
mode, and holdoff.
• Video (with option 05: Video Trigger): Video triggering is compatible with
standard NTSC, PAL, SECAM, and HDTV formats. An additional feature called FlexFormat™ (flexible format) allows the user to define the video format on which to trigger.
You can choose where the trigger point is located within the acquired wave-form record by selecting the amount of pretrigger data displayed. Presets of 20%, 50%, and 80% of pretrigger data can be selected in the horizontal
menu, or the general purpose knob can be assigned to set pretrigger data
to any value within the 20% to 80% limits.
Acquisition Control You can specify a mode and manner to acquire and process signals that
matches your measurement requirements.
• Select the mode for interpolation of points sampled on non-repetitive signals (linear or sin (x)/x. This can increase the apparent sample rate on the waveform when maximum real-time rates are reached.
• Use sample, envelope, and average modes to acquire signals.
• Set the acquisition to stop after a single acquisition (or sequence of acquisitions if acquiring in average or envelope modes), or after a limit condition has been met.
• Select channel sources for compliance with limit tests. You can direct the TDS to signal you or generate hard copy output either to a printer or to a floppy disk (with option 1F) based on the results. Also, you can create templates for use in limit tests.
7-4 Specifications
Specifications
On-Board User Help and autoset
can
assist you in setting up the Digitizing Oscilloscope to
Assistance make your measurements.
Help
Help displays operational information about any front-panel control. When help mode is in effect, manipulating any front-panel control causes the Digitizing Oscilloscope to display information about that control. When help is first invoked, an introduction to help is displayed on screen.
Autoset
Autoset automatically sets up the Digitizing Oscilloscope for a viewable display based on the input signal.
Measurement Once you have set up to make your measurements, the cursor and measure
Assistance features can help you quickly make those measurements.
Cursor
Three types of cursors are provided for making parametric measurements on the displayed waveforms. Horizontal bar cursors (H Bar) measure vertical
parameters (typically volts). Vertical bar cursors (V Bar) measure horizontal parameters (typically time or frequency) and now extend to the top and bottom of the screen. Paired cursors measure both amplitude and time
simultaneously. These are delta measurements; that is, measurements
based on the difference between two cursors.
Both H Bar and V Bar cursors can also be used to make absolute measure-ments; that is measurements relative to a defined level or event. For the
H Bars, either cursor can be selected to read out its voltage with respect to any channels ground reference level. For the V Bars, it's time with respect to the trigger point (event) of the acquisition and the cursors can control the
portion of the waveform on which automatic measurements are made.
For time measurements, units can be either seconds or Hertz (for 1/time).
When the video trigger option installed (Option 05), you can measure the video line number using the vertical cursors. You can measure IRE ampli-tude (NTSC) using the horizontal cursors with or without the video trigger option installed.
Measure
Measure can automatically extract parameters from the signal input to the
Digitizing Oscilloscope. Any four out of the more than 25 parameters avail-able can be displayed to the screen. The waveform parameters are mea-sured continuously with the results updated on-screen as the Digitizing Oscilloscope continues to acquire waveforms.
TDS 620A, 640A, & 644A Service Manual 1-5
Specifications
Digital Signal Processing (DSP)
An important component of the multiprocessor architecture of this Digitizing Oscilloscope is Tektronix's proprietary digital signal processor, the DSF! This dedicated processor supports advanced analysis of your waveforms when doing such compute-intensive tasks as interpolation, waveform math, and signal averaging. It also teams with a custom display system to deliver specialized display modes (See Display, later in this description.)
and I/O Acquired waveforms may be saved in any of four nonvolatile REF (reference)
memories or on a 3.5 inch, DOS 3.3-or-later compatible disk. Any or all of the saved waveforms may be displayed for comparison with the waveforms
being currently acquired.
The source and destination of waveforms to be saved may be chosen. You can save any of the four channels to any REF memory or to move a stored
reference from one REF memory to another. Reference waveforms may also
be written into a REF memory location via the GFIB interface.
The Digitizing Oscilloscope is fully controllable and capable of sending and
receiving waveforms over the GFIB interface (IEEE Std 488.1-1987/IEEE Std 488.2-1987 standard). This feature makes the instrument ideal for
making automated measurements in a production or research and develop-
ment environment that calls for repetitive data taking. Self-compensation and self-diagnostic features built into the Digitizing Oscilloscope to aid in fault detection and servicing are also accessible using commands sent from a GFIB controller.
Another standard feature is hardcopy. This feature allows you to output waveforms and other on-screen information to a variety of graphic printers and plotters from the TDS front panel, providing hard copies without requir-
ing you to put the TDS into a system-controller environment. You can make
hardcopies in a variety of popular output formats, such as FCX, TIFF, BMF!
RLE, EFS, Interleaf, and EFS mono or color. You can also save hardcopies
in a disk file in any of the formats above. The hardcopies obtained are based on what is displayed on-screen at the time hardcopy is invoked. The hardco-
pies can be stamped with date and time and spooled to a queue for printing at a later time. You can output screen information via GFIB, RS-232C, or Centronics interfaces.
7-6 Specifications
Specifications
Display
The TDS 620A, 640A, & 644A Digitizing Oscilloscopes offers flexible display options. You can customize the following attributes of your display:
Color (on the TDS 644A): waveforms, readouts, graticule, and variable
persistence with color coding;
Intensity: waveforms, readouts, and graticule;
Style of waveform display(s): vectors or dots, intensified or non-intensi-fied samples, infinite persistence, and variable persistence with color coding;
Interpolation method: Sin(x)/x or Linear;
Display format: xy or yt with various graticule selections including NTSC
and PAL to be used with video trigger (option 05).
Zoom
This oscilloscope also provides an easy way to focus in on those waveform features you wish to examine up close. By invoking zoom, you can magnify the waveform parameter using the vertical and horizontal controls to expand
(or contract) and position it for viewing.
TDS 620A, 640A, & 644A Service Manual
1-7
Specifications
7-8 Specifications
Nominal Traits
This subsection contains a collection of tables that list the various nominal traits that describe the TDS 620A, 640A, & 644A Digitizing Oscilloscope.
Electrical and mechanical traits are included.
Nominal traits are described using simple statements of fact such as "Four, all identical" for the trait "Input Channels, Number of," rather than in terms of
limits that are performance requirements.
Table 1-2: Nominal Traits — Signal Acquisition System
Name
Bandwidth Selections
Samplers, Number of Four, simultaneous
Digitized Bits, Number of
Input Channels, Number of Four, all identical (TDS 640A/644A)
Input Coupling
Input Impedance Selections
Ranges, Offset
Range, Position ±5 divisions
Range, Sensitivity, CH 1-CH 4 on the TDS 640A, 644A; CH1, CH2, AUX1, and AUX2 on the TDS 620A
Description
20 MHz, 100 MHz, and FULL (500 MHz)
1
8 bits
Two, all identical (TDS 620A)
DC, AC, or GND
1 MO or 50 O
Volts/Div Setting Offset Range
1 mV/div-99.5 mV/div ±1 V
100 mV/div-995 mV/div
1 V/div-10 V/div ±100 V
1 mV/div to 10 V/div
2
±10 V
1 Displayed vertically with 25 digitization levels (DLs) per division and 10.24 divisions dynamic range with zoom off. A DL is the
smallest voltage level change that can be resolved by the 8-bit A-D Converter, with the input scaled to the volts/division setting of the channel used. Expressed as a voltage, a DL is equal to 1/25 of a division times the volts/division setting.
2 The sensitivity ranges from 1 mV/div to 10 V/div in a 1-2-5 sequence of coarse settings. Between a pair of adjacent coarse set-
tings, the sensitivity can be finely adjusted. The resolution of such a fine adjustment is 1% of the more sensitive of the pair. For example, between 5o mV/div and 1o0 mV/div, the volts/division can be set with 0.5 mV resolution.
TDS 620A, 640A, & 644A Service Manual 1-9
Loading...
+ 258 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.