The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
that in all previously published material. Specifications and price change privileges reserved.
Printed in the U.S.A.
T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000
TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.
WARRANTY
T ektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship
for a period of three (3) years from the date of shipment. If a product proves defective during this warranty period,
T ektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a
replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the
warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid.
T ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the
T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any
other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair
damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction
caused by the use of non-T ektronix supplies; or d) to service a product that has been modified or integrated with other
products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO
REP AIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO
THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE
LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE
OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
Service Assurance
If you have not already purchased Service Assurance for this product, you may do so at any time during the product’s
warranty period. Service Assurance provides Repair Protection and Calibration Services to meet your needs.
Repair Protection extends priority repair services beyond the product’s warranty period; you may purchase up to three
years of Repair Protection.
Calibration Services provide annual calibration of your product, standards compliance and required audit documentation,
recall assurance, and reminder notification of scheduled calibration. Coverage begins upon registration; you may purchase
up to five years of Calibration Services.
Service Assurance Advantages
H Priced well below the cost of a single repair or calibration
H Avoid delays for service by eliminating the need for separate purchase authorizations from your company
H Eliminates unexpected service expenses
For Information and Ordering
For more information or to order Service Assurance, contact your T ektronix representative and provide the information
below . Service Assurance may not be available in locations outside the United States of America.
NameVISA or Master Card number and expiration
Companydate or purchase order number
AddressRepair Protection (1,2, or 3 years)
City , State, Postal codeCalibration Services (1,2,3,4, or 5 years)
CountryInstrument model and serial number
PhoneInstrument purchase date
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
Injury Precautions
Use Proper Power Cord. To avoid fire hazard, use only the power cord specified
for this product.
Avoid Electric Overload. To avoid electric shock or fire hazard, do not apply a
voltage to a terminal that is outside the range specified for that terminal.
Avoid Electric Shock. To avoid injury or loss of life, do not connect or disconnect
probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Do Not Operate Without Covers. To avoid electric shock or fire hazard, do not
operate this product with covers or panels removed.
Use Proper Fuse. To avoid fire hazard, use only the fuse type and rating specified
for this product.
Do Not Operate in Wet/Damp Conditions. To avoid electric shock, do not operate
this product in wet or damp conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fire hazard, do not
operate this product in an explosive atmosphere.
Product Damage
Precautions
TDS 510A Service Manual
Use Proper Power Source. Do not operate this product from a power source that
applies more than the voltage specified.
Provide Proper Ventilation. To prevent product overheating, provide proper
ventilation.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
ix
General Safety Summary
Symbols and Terms
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
T erms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
DANGER
High Voltage
Protective Ground
(Earth) T erminal
ATTENTION
Refer to Manual
Double
Insulated
x
TDS 510A Service Manual
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, disconnect the main power by means
of the power cord or, if provided, the power switch.
Use Caution When Servicing the CRT. To avoid electric shock or injury, use
extreme caution when handling the CRT. Only qualified personnel familiar with
CRT servicing procedures and precautions should remove or install the CRT.
CRTs retain hazardous voltages for long periods of time after power is turned off.
Before attempting any servicing, discharge the CRT by shorting the anode to
chassis ground. When discharging the CRT, connect the discharge path to ground
and then the anode. Rough handling may cause the CRT to implode. Do not nick
or scratch the glass or subject it to undue pressure when removing or installing it.
When handling the CRT, wear safety goggles and heavy gloves for protection.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
X-Radiation. To avoid x-radiation exposure, do not modify or otherwise alter the
high-voltage circuitry or the CRT enclosure. X-ray emissions generated within
this product have been sufficiently shielded.
TDS 510A Service Manual
xi
Service Safety Summary
xii
TDS 510A Service Manual
Preface
Manual Structure
This preface contains information needed to properly use this manual to service
the TDS 510A Digitizing Oscilloscope, as well as general information critical to
safe and effective servicing of this oscilloscope.
This manual is divided into sections, such as Specification and Theory ofOperation. Further, it is divided into subsections, such as Product Description
and Removal and Installation Procedures.
Sections containing procedures also contain introductions to those procedures.
Be sure to read these introductions because they provide information needed to
do the service correctly and efficiently. The following is a brief description of
each manual section.
HSpecification contains a product description of the TDS 510A Digitizing
Oscilloscope and tables of the characteristics and descriptions that apply to
it.
HOperating Information includes general information and operating instruc-
tions at the level needed to safely power on and service this oscilloscope. A
statement of the service strategy that this manual supports, and instructions
for shipment of the oscilloscope are found in this section.
HTheory of Operation contains circuit descriptions that support general service
and fault isolation.
HPerformance Verification contains a collection of procedures for confirming
that this oscilloscope functions properly and meets warranted limits.
HAdjustment Procedures contains a collection of procedures for adjusting this
oscilloscope to meet warranted limits.
HMaintenance contains information and procedures for doing preventive and
corrective maintenance of this oscilloscope. Instructions for cleaning, for
module removal and installation, and for fault isolation to a module are
found here.
HOptions contains information on servicing any of the factory-installed
options that may be present in your oscilloscope.
HElectrical Parts List contains a statement referring you to Mechanical
Replaceable Parts, where both electrical and mechanical modules are listed.
TDS 510A Service Manual
xiii
Preface
Manual Conventions
HDiagrams contains a block diagram and an interconnection diagram useful
for isolating failed modules.
HMechanical Parts List includes a table of all replaceable modules, their
descriptions, and their Tektronix part numbers.
This manual uses certain conventions which you should become familiar with
before doing service.
Modules
Safety
Symbols
Throughout this manual, any replaceable component, assembly, or part of these
oscilloscope is referred to generically as a module. In general, a module is an
assembly, like a circuit board, rather than a component, like a resistor or an
integrated circuit. Sometimes a single component is a module; for example, each
chassis part of the oscilloscope is a module.
Symbols and terms related to safety appear in the Safety Summary found at the
beginning of this manual.
Besides the symbols related to safety, this manual uses the following symbols:
STOP. The stop labels information which must be read in order to correctly do
service and to avoid incorrectly using or applying service procedures.
The clock icon labels procedure steps which require a pause to wait for
the oscilloscope to complete some operation before you can continue.
Various icons such as the example icon at the left are used in procedures
to help identify certain readouts and menu functions on screen.
Related Manuals
xiv
The TDS 510A Digitizing Oscilloscope comes with the following manuals:
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual contains a tutorial
to quickly show you how to operate the TDS 510A Digitizing Oscilloscope and
an in depth discussion of how to more completely use their features. Applications are also discussed.
TDS 420A, TDS 430A, TDS 460A & TDS 510A Reference contains a brief
overview of oscilloscope operation.
TDS 510A Service Manual
Preface
TDS Family Programmer Manual contains information for programmed
operation via the GPIB interface. Included are the complete command set, setup
information, and programming examples.
TDS 510A Technical Reference contains performance verification procedures and
specifications.
TDS 510A Service Manual
xv
Preface
xvi
TDS 510A Service Manual
Introduction
Service Strategy
STOP. Throughout this manual, any field-replaceable component, assembly, or
part of this oscilloscope is referred to generically as a module.
This manual contains all the information needed for periodic maintenance of the
TDS 510A Digitizing Oscilloscope. (Examples of such information are
procedures for checking performance and for readjustment.) Further, it contains
all information for corrective maintenance down to the module level. This means
that the procedures, diagrams, and other troubleshooting aids help isolate failures
to a specific module, rather than to components of that module. Once a failure is
isolated, replace the module with a fresh unit obtained from the factory.
All modules are listed in Mechanical Parts List. To isolate a failure to a module,
use the fault isolation procedures found in Section 6, Maintenance Information.
To remove and replace any failed module, follow the instructions in Removaland Installation Procedures, also found in Section 6.
Service Offerings
Warranty Repair Service
Repair or Calibration
Service
Tektronix provides service to cover repair under warranty as well as other
services that may provide a cost-effective answer to your service needs.
Whether providing warranty repair service or any of the other services listed
below, Tektronix service technicians, trained on Tektronix products, are best
equipped to service your TDS 510A Digitizing Oscilloscope. Tektronix
technicians are appraised of the latest information on improvements to the
product as well as the latest new options to the product.
Tektronix warrants this product for three years from date of purchase, excluding
probes for which the warranty is one year. (The warranty appears after the title
page and copyright page in this manual.) Tektronix technicians provide warranty
service at most Tektronix service locations worldwide. Your Tektronix product
catalog lists all service locations worldwide.
The following services may be purchased to tailor repair and/or calibration of
your TDS 510A Digitizing Oscilloscope to fit your requirements.
TDS 510A Service Manual
xvii
Introduction
Option 95. The oscilloscope, when ordered with option 95, is shipped with a Test
Data Report.
At-Depot Service. Tektronix offers several standard-priced adjustment (calibration) and repair services:
HA single repair and/or adjustment.
HCalibrations using equipment and procedures that meet the traceability
standards specific to the local area.
HAnnual maintenance agreements that provide for either calibration and repair
or calibration only of the oscilloscope.
Of these services, the annual maintenance agreement offers a particularly
cost-effective approach to service for many owners of the TDS 510A Digitizing
Oscilloscope. Such agreements can be purchased to span several years.
On-Site Service. The annual maintenance agreement can be purchased with
on-site service, with repair and calibration done at your facility. This service
reduces the time your oscilloscope is out of service when calibration or repair is
required.
Self Service
Tektronix supports repair to the module level by providing Module Exchange
and Module Repair and Return.
Module Exchange. This service reduces down time for repair by allowing you to
exchange most modules for remanufactured ones. Tektronix ships you an
updated and tested exchange module from the Beaverton, Oregon service center,
typically within 24 hours. Each module comes with a 90-day service warranty.
Module Repair and Return. This service returns to you within 10 days the same
module that you shipped to Tektronix. The module shipped is repaired, tested,
and returned to you from the Beaverton, Oregon service center. It is not updated
to match current modules of the same type. Again, each module comes with a
90-day service warranty.
For More Information. Contact your local Tektronix service center or sales
engineer for more information on any of the repair or adjustment services just
described.
xviii
TDS 510A Service Manual
Before You Begin
Introduction
This manual is for servicing the TDS 510A Digitizing Oscilloscope. To prevent
injury to yourself or damage to the oscilloscope, do the following tasks before
you attempt service:
HBe sure you are a qualified service person.
HRead the Safety Summary found at the beginning of this manual.
HRead Service Strategy in this section and Supplying Operating Power in
section 2.
When using this manual for servicing, be sure to heed all warnings, cautions, and
notes.
TDS 510A Service Manual
xix
Introduction
xx
TDS 510A Service Manual
Specifications
This chapter begins with a general description of the traits of the TDS 510A
Digitizing Oscilloscope. Three sections follow, one for each of three classes of
traits: nominal traits, warranted characteristics, and typical characteristics.
General Product Description
The TDS 510A is a portable, four-channel oscilloscope suitable for use in a
variety of test and measurement applications and systems. Key features include:
H500 MHz maximum analog bandwidth
H500 Megasamples/second maximum digitizing rate
HFour-channel acquisition
HExtensive triggering capabilities: including edge, logic, and glitch; optional
NTSC, PAL, SECAM, HDTV, and FlexFormat
video trigger modes
HWaveform Math — Invert a single waveform and add, subtract, and multiply
two waveforms, and optionally integrate or differentiate a single waveform
or perform an FFT (fast fourier transform) on a waveform to display its
magnitude or phase versus its frequency
HEight-bit digitizers
HUp to 50,000-point record length per channel
HFull GPIB software programmability; hardcopy output using GPIB, RS-232,
or Centronics ports (RS-232 and Centronics are option 13)
HComplete measurement and documentation capability
HIntuitive graphic icon operation blended with the familiarity of traditional
horizontal and vertical knobs
HOn-line help at the touch of a button
HA full complement of advanced functions, like continuously updated
measurement results and local pass/fail decision making
TM
(user definable format)
TDS 510A Service Manual
HSpecialized display modes, including variable persistence, dot or vector
mode, sin(x)/x or linear display filters, and Fit-to-Screen compression of the
waveform record to fit on the screen
1–1
Specifications
User Interface
HA complement of advanced acquisition modes such as peak-detect, high-res-
olution, sample, envelope, and average
HA unique graphical user interface (GUI), an on-board help mode, and a
logical front-panel layout
HVGA output for driving remote monitors
HNVRAM and an optional 1.44 Mbyte, DOS 3.3 or later, floppy disk drive for
saving waveforms and oscilloscope setups
Use a combination of front-panel buttons, knobs, and on-screen menus to control
the many functions of this oscilloscope. The front-panel controls are grouped
according to function: vertical, horizontal, trigger, and special. Set a function you
adjust often, such as vertical positioning or the time base setting, directly by its
own front-panel knob. Set functions which you change less often, such as
vertical coupling and horizontal mode, indirectly using selected menus.
Menus
Indicators
General Purpose Knob
GUI
Pressing one (sometimes two) front-panel button(s), such as vertical menu,
displays a main menu of related functions, such as coupling and bandwidth at the
bottom of the screen. Pressing a main-menu button, such as coupling, displays a
side menu of settings for that function, such as AC, DC, or GND (ground)
coupling, at the right side of the screen. Pressing a side-menu button selects a
setting such as DC.
On-screen readouts help you keep track of the settings for various functions,
such as vertical and horizontal scale and trigger level. Some readouts use the
cursors or the automatic parameter extraction feature (called measure) to display
the results of measurements made or the status of the instrument.
Assign the general purpose knob to adjust a selected parameter function. Change
parameters more quickly by toggling the SHIFT button. Use the same method as
for selecting a function, except the final side-menu selection assigns the general
purpose knob to adjust some function, such as the position of measurement
cursors on screen, or the setting for fine gain.
The user interface also makes use of a GUI, or Graphical User Interface, to make
setting functions and interpreting the display more intuitive. Some menus and
status are displayed using iconic representations of function settings such as
those shown here for Full, 100 MHz, and 20 MHz bandwidth. Such icons allow
you to more readily determine status or the available settings.
1–2
TDS 510A Service Manual
Signal Acquisition System
The signal acquisition system provides four vertical channels with calibrated
vertical scale factors from 1 mV to 10 V per division. All four channels can be
acquired simultaneously.
Each of the four channels can be displayed, vertically positioned, and offset, can
have their bandwidth limited (100 MHz or 20 MHz) and their vertical coupling
specified. Fine gain can also be adjusted.
Besides the four channels, up to three math waveforms and four reference
waveforms are available for display. (A math waveform results when dual
waveform operations, such as add, are specified on any two channels. A
reference waveform results when you save a live waveform in a reference
memory.)
Horizontal System
Specifications
Trigger System
There are three horizontal display modes: main only, main intensified, and
delayed only. You can select among various horizontal record length settings.
A feature called “Fit to Screen” allows the user to view entire waveform records
within a 10 to 15 division screen area. In other words, waveforms are compressed to fit on the screen.
You can delay by time with respect to the main trigger both the delayed display
and the intensified zone on the main intensified display. You can set them to
display immediately after the delay (delayed runs after main mode). The delayed
display can also be set to display at the first valid trigger after the delay (delayed
triggerable mode).
You can also delay by a selected number of events the delayed display (or the
intensified zone). In this case, the events source is the delayed trigger source. For
any events signal, the delayed-trigger system conditions the signal by determining the source, coupling, and so on, of that signal.
The triggering system supports a varied set of features for triggering the
signal-acquisition system. Trigger signals recognized include:
TDS 510A Service Manual
HEdge (main- and delayed-trigger systems): This familiar type of triggering is
fully configurable for source, slope, coupling, mode (auto or normal), and
holdoff.
HLogic (main-trigger system): This type of triggering can be based on pattern
(asynchronous) or state (synchronous). In either case, logic triggering is
1–3
Specifications
configurable for sources, for boolean operators to apply to those sources, for
logic pattern or state on which to trigger, for mode (auto or normal), and for
holdoff. Time-qualified logic triggering can also be specified.
HPulse (main-trigger system): Pulse triggering is configurable for triggering
on runt or glitch pulses, or on pulse widths or periods inside or outside limits
that you specify. It is also configurable for source, polarity, mode, and
holdoff.
HVideo (with option 05: Video Trigger): Video triggering is compatible with
standard NTSC, PAL, SECAM, and HDTV formats. An additional feature
called FlexFormat
format on which to trigger.
You can choose where the trigger point is located within the acquired waveform
record by selecting the amount of pretrigger data displayed. Select presets of
10%, 50%, and 90% of pretrigger data in the horizontal menu, or assign the
general purpose knob to set pretrigger data to any value within the limits of
trigger position resolution.
TM
(flexible format) allows the user to define the video
Acquisition Control
You can specify a mode and manner to acquire and process signals:
HSelect equivalent-time sampling on repetitive signals or interpolation of
HUse peak-detect, high-resolution, sample, envelope, and average modes to
HSet the acquisition to stop after a single acquisition (or sequence of
HSelect channel sources for compliance with limit tests. You can direct the
On-Board User Assistance
Help and autoset can assist you in setting up the oscilloscope to take your
measurements.
points sampled on non-repetitive signals. Both can increase apparent sample
rate on the waveform when maximum real-time rates are reached.
acquire signals.
acquisitions if acquiring in average or envelope modes).
oscilloscope to signal you or generate hard copy output based on the results.
Also, you can create templates for use in limit tests.
1–4
Help
Help displays operational information about any front-panel control. When help
mode is in effect, manipulating any front-panel control causes the oscilloscope to
TDS 510A Service Manual
Specifications
display information about that control. When help is first invoked, an introduction to help is displayed on screen.
Autoset
Autoset automatically sets up the oscilloscope for a viewable display based on
the input signal.
Measurement Assistance
Once you have set up to make your measurements, the cursor and measure
features can help you quickly make those measurements.
Cursor
Three types of cursors are provided for making parametric measurements on the
displayed waveforms. Horizontal bar cursors (H Bar) measure vertical parameters (typically volts). Vertical bar cursors (V Bar) measure horizontal parameters
(typically time or frequency) and now extend to the top and bottom of the screen.
Paired cursors measure both amplitude and time simultaneously. These are delta
measurements; that is, measurements based on the difference between two
cursors.
Both H Bar and V Bar cursors can also be used to make absolute measurements;
that is measurements relative to a defined level or event. For the H Bars, either
cursor can be selected to read out its voltage with respect to any channel’s
ground reference level. V Bars measure time with respect to the trigger point
(event) of the acquisition. The cursors can control the portion of the waveform
on which automatic measurements are made.
Measure
Storage and I/O
TDS 510A Service Manual
When the video trigger option installed (Option 05), the video line number can
be selected using the vertical cursors. IRE amplitude (NTSC) can be measured
using the horizontal cursors with or without the video trigger option installed.
Measure can automatically extract parameters from the signal input to the
oscilloscope. Any four out of the more than 20 parameters available can be
displayed to the screen. The waveform parameters are measured continuously
with the results updated on screen as the oscilloscope continues to acquire
waveforms.
Acquired waveforms may be saved in any of four nonvolatile REF (reference)
memories or, if available, on a 3.5 inch, DOS 3.3-or-later compatible disk. The
disk is available as option 1F. Any or all of the saved waveforms may be
displayed for comparison with the waveforms currently being acquired.
1–5
Specifications
Display and Zoom
The oscilloscope is fully controllable and capable of sending and receiving
waveforms over the GPIB interface. This feature makes the instrument ideal for
making automated measurements in a production or research and development
environment that calls for repetitive data taking. Self-compensation and
self-diagnostic features built into the oscilloscope to aid in fault detection and
servicing are also accessible using commands sent from a GPIB controller.
The hardcopy feature allows you to output waveforms and other on-screen
information to a variety of graphic printers and plotters from the oscilloscope
front panel, providing hard copies without requiring you to put the oscilloscope
into a system-controller environment. You can make hardcopies in a variety of
popular output formats. You can also save hardcopies in a disk file. The
hardcopies obtained are based on what is displayed on-screen at the time
hardcopy is invoked. The hardcopies can be stamped with date and time and
spooled to a queue for printing at a later time. You can output screen information
via GPIB, RS-232C, or Centronics interfaces.
The oscilloscope offers flexible display options. You can customize the
following attributes of your display:
HIntensity: waveforms, readouts, and graticule
HStyle of waveform display(s): vectors or dots, intensified or non-intensified
samples, infinite persistence, and variable persistence with color coding
HInterpolation method: Sin(x)/x or Linear
HDisplay format: xy or yt with various graticule selections including NTSC
and PAL to be used with video trigger
The oscilloscope also provides an easy way to focus in on those waveform
features you wish to examine up close. By invoking zoom, you can magnify the
waveform parameter using the vertical and horizontal controls to expand (or
contract) and position it for viewing.
1–6
TDS 510A Service Manual
Nominal Traits
This section contains tables that list the electrical and mechanical nominal traits
that describe the TDS 510A Digitizing Oscilloscope.
Nominal traits are described using simple statements of fact, such as “Four, all
identical” for the trait “Input Channels, Number of,” rather than in terms of
limits that are performance requirements.
T able 1–1: Nominal Traits — Signal Acquisition System
NameDescription
Bandwidth Selections20 MHz, 100 MHz, and FULL (500 MHz)
Digitizers, Number ofFour, all identical
Digitized Bits, Number of8 bits
Input Channels, Number ofFour, all identical, called CH1 – CH4
Input CouplingDC, AC, or GND
Input Impedance Selections
Ranges, OffsetVolts/Div Setting
Range, Position±5 divisions
Range, Sensitivity1 mV/div to 10 V/div
1
Displayed vertically with 25 digitization levels (DLs) per division and 10.24 divisions dynamic range with zoom off. A DL
is the smallest voltage level change that can be resolved by the 8-bit A-D Converter, with the input scaled to the
volts/division setting of the channel used. Expressed as a voltage, a DL is equal to 1/25 of a division times the volts/division setting.
2
GND input coupling disconnects the input connector from the attenuator and connects a ground reference to the input of
the attenuator.
3
The sensitivity ranges from 1 mV/div to 10 V/div in a 1–2–5 sequence of coarse settings. Between a pair of adjacent
coarse settings, the sensitivity can be finely adjusted. The resolution of such a fine adjustment is 1% of the more
sensitive coarse setting. For example, between 50 mV/div and 100 mV/div , the volts/division can be set with 0.5 mV
resolution.
Range, Seconds/Division500 ps/div to 10 s/div
Record Length
4
500 samples, 1000 samples, 2500 samples, 5000 samples, 15,000 samples, and 50,000
samples. Up to four 50 K waveform records may be saved in NVRAM.
Range, Time Base Delay Time16 ns to 250 seconds
1
The range of real-time rates, expressed in samples/second, at which a digitizer samples signals at its inputs and stores
the samples in memory to produce a record of time-sequential samples.
2
The range of waveform rates for equivalent time or interpolated waveform records.
3
The Waveform Rate (WR) is the equivalent sample rate of a waveform record. For a waveform record acquired by
real-time sampling of a single acquisition, the waveform rate is the same as the real-time sample rate; for a waveform
created by interpolation of real-time samples from a single acquisition or by equivalent-time sampling of multiple
acquisitions, the waveform rate is faster than the real time sample rate. For all three cases, the waveform rate is
1/(Waveform Interval) for the waveform record, where the waveform interval (WI) is the time between the samples in the
waveform record.
4
The maximum record length of 50,000 samples is selectable with all acquisition modes except Hi Res. In Hi Res, the
maximum record length is 15,000 samples.
1–8
TDS 510A Service Manual
T able 1–3: Nominal Traits — Triggering System
NameDescription
Range, Delayed Trigger Time Delay16 ns to 250 s
Range, Events Delay2 to 10,000,000
Range (Time) for Pulse-Glitch or
Pulse-Width Triggering
Ranges,Trigger Level or ThresholdSource
2 ns to 1 s
Nominal Traits
Range
Any Channel
Auxiliary
Line
Video Trigger Modes of Operation
(Option 05 Video Trigger)
Supports the following video standards:
HNTSC (525/60) – 2 field mono or 4 field
HPAL (625/50) – 2 field mono, 8 field
HSECAM
HHDTV –
HFlexFormat
T able 1–4: Nominal Traits — Display System
±12 divisions from center of screen
±4 V
±300 V
(787.5/60)
(1050.60)
(1125/60)
(1250/60)
TM
(user definable standards)
User can specify: field rate, number of lines, sync pulse width and polarity, line rate,
and vertical interval timing.
NameDescription
CRT Type7 inch diagonal monochrome, magnetic deflection CRT with horizontal raster orientation.
The phosphor is P4
Video Display Resolution640 pixels horizontally by 480 pixels vertically in a display area of 5.04 inches horizontally
by 3.78 inches vertically
Waveform Display GraticuleSingle Graticule: 401 × 501 pixels, 8 × 10 divisions, where divisions are 1 cm by 1 cm
Waveform Display Grey ScaleSixteen levels in infinite-persistence and variable persistence display styles.
TDS 510A Service Manual
1–9
Nominal Traits
T able 1–5: Nominal Traits — Interfaces, Output Ports, and Power Fuse
NameDescription
Interface, GPIBGPIB interface complies with IEEE Std 488–1987
Interface, RS-232 (Option 13)RS-232 interface complies with EIA/TIA 574 (talk only)
Interface, Centronics (Option 13)Centronics interface complies with Centronics interface standard C332-44 Feb 1977,
REV A
Interface, VideoVGA video output1, non interlaced, with levels that comply with ANSI R5343A standard.
DB-15 connector.
Logic Polarity for Main- and Delayed-Trigger Outputs
Fuse RatingEither of two fuses2 may be used: a .25I × 1.25I (UL 198.6, 3AG): 6 A FAST, 250 V, or
1
VGA compatible at 30.6 kHz line sync rate.
2
Each fuse type requires its own fuse cap.
Negative TRUE. High to low transition indicates the trigger occurred.
a 5 mm × 20 mm, (IEC 127): 5 A (T), 250V.
T able 1–6: Nominal Traits — Mechanical
NameDescription
Cooling MethodForced-air circulation with no air filter
Construction MaterialChassis parts constructed of aluminum alloy; front panel constructed of plastic laminate;
circuit boards constructed of glass-laminate. Cabinet is aluminum and is clad in Tektronix
Blue vinyl material.
Finish TypeTektronix Blue vinyl-clad aluminum cabinet
WeightStandard oscilloscope
12.3 kg (27 lbs), with front cover.
20.0 kg (44 lbs), when packaged for domestic shipment.
Rackmount oscilloscope
12.3 kg (27 lbs) plus weight of rackmount parts, for the rackmounted oscilloscope
(Option 1R).
25.6 kg (56 lbs), when the rackmounted oscilloscope is packaged for domestic shipment.
Rackmount conversion kit
2.3 kg (5 lbs), parts only; 3.6 kg (8 lbs), parts plus package for domestic shipping.
1–10
TDS 510A Service Manual
T able 1–6: Nominal Traits — Mechanical (Cont.)
NameDescription
Overall DimensionsStandard oscilloscope
Height: 193 mm (7.6 in), with the feet installed.
Width: 445 mm (17.5 in), with handle.
Depth: 434 mm (17.1 in), with front cover installed.
Rackmount oscilloscope
Height: 178 mm (7.0 in).
Width: 483 mm (19.0 in).
Depth: 558.8 mm (22.0 in).
Nominal Traits
TDS 510A Service Manual
1–11
Nominal Traits
1–12
TDS 510A Service Manual
W arranted Characteristics
This section lists the electrical and environmental warranted characteristics that
describe the TDS 510A Digitizing Oscilloscope.
Warranted characteristics are described in terms of quantifiable performance
limits which are warranted.
NOTE. In these tables, those warranted characteristics that are checked in the
procedure Performance Verification, found in Section 4, appear in boldface type
under the column Name.
As stated above, this subsection lists onlywarranted characteristics. A list of
typical characteristics starts on page 1–19.
Performance Conditions
The electrical characteristics found in these tables of warranted characteristics
apply when the scope has been adjusted at an ambient temperature between
+20_ C and +30_ C, has warmed-up for at least 20 minutes, and is operating at
an ambient temperature between +4_ C and +50_ C (unless otherwise noted).
T able 1–7: Warranted Characteristics — Signal Acquisition System
NameDescription
Accuracy, DC Gain±1%
(For all sensitivities from 1 mV/div to 10 V/div with offsets from 0 V to ±100 V.)
T able 1–7: Warranted Characteristics — Signal Acquisition System (Cont.)
NameDescription
Analog Bandwidth, DC-50 W Coupled or
DC-1 MW Coupled
Volts/Div
5 mV/div – 10 V/div
Bandwidth
DC – 500 MHz
4
2 mV/div – 4.98 mV/div
1 mV/div – 1.99 mV/div
DC – 350 MHz
DC – 250 MHz
Cross T alk (Channel Isolation)w100:1 at 100 MHz and w30:1 at the derated bandwidth for any two channels having
equal volts/division settings
Input Impedance, DC-1 MW Coupled1 MW ± 0.5% in parallel with 10 pF ±2 pF
Input Impedance, DC-50 W Coupled
Input Voltage, Maximum, DC-1MW,
50 W ± 1% with VSWR v1.3:1 from DC – 500 MHz
300 V CAT II; derate at 20 dB/decade above 1 MHz
AC-1 MW, or GND Coupled
Input Voltage, Maximum, DC-50W or
5 V
, with peaks v ±30 V
RMS
AC-50 W Coupled
Lower Frequency Limit, AC Coupledv10 Hz when AC–1 MW Coupled; v200 kHz when AC – 50 W Coupled
Delay Between Channels, Full Band-
v250 ps for any two channels with equal volts/division and coupling settings
5
width, Equivalent Time
1
Net Offset = Offset – (Position Volts/Div). Net Offset is the nominal voltage level at the center of the A-D converter
dynamic range. Offset Accuracy is the accuracy of this Voltage level.
2
The samples must be acquired under the same setup and ambient conditions.
3
To ensure the most accurate measurements possible, run an SPC calibration first. When using the oscilloscope at a
Volts/Div setting v5 mV/div, an SPC calibration should be run once per week to ensure that instrument performance
levels meet specifications.
4
The limits given are for the ambient temperature range of 0_ C to +30_ C. Reduce the upper bandwidth frequencies by
4.0 MHz for each _ C above +30_ C.
5
The AC Coupled Lower Frequency Limits are reduced by a factor of 10 when 10X, passive probes are used.
T able 1–8: Warranted Characteristics — Time Base System
NameDescription
Accuracy, Long Term Sample Rate and
±25 ppm over any w1 ms interval
Delay Time
1–14
TDS 510A Service Manual
T able 1–9: Warranted Characteristics — Triggering System
Sensitivity, Edge-Type Trigger, D
NameDescription
Warranted Characteristics
Accuracy (Time) for Pulse-Glitch or
Pulse-Width Triggering
C
Coupled
1
Time Range
2 ns to 1 ms
1.02 ms to 1 s
Trigger Source
CH1 – CH4
Accuracy
±(20% of setting + 0.5 ns)
±(100 ns + 0.0025% of setting)
Sensitivity
0.35 division from DC to 50 MHz, increasing to 1 division at
500 MHz
0.55 division from DC to 50 MHz, increasing to 1.5 division at
500 MHz
Auxiliary0.25 volts from DC to 50 MHz
Width, Minimum Pulse and Rearm,
for Pulse Triggering
Pulse Class
Glitch
Runt
Width
Minimum Pulse Width
2 ns
2.5 ns
2 ns
Minimum Rearm Width
2 ns + 5% of Glitch Width Setting
2.5 ns
2 ns + 5% of Width Upper Limit
Setting
Accuracy , Trigger Level or Threshold,
DC Coupled
Trigger Source
Any Channel
Accuracy
±(2% of |(Setting – Net Offset)| + 0.3 div volts/div setting +
2
Offset Accuracy)
Auxiliary
Jitter (Option 05 Video Trigger)17 ns
Input Signal Sync Amplitude for Stable
0.6 division to 4 division (1 division to 4 divisions in Numerical Field)
on HDTV signal; 60 ns
p-p
±(6% of |Setting| + 8% of p-p signal + 100 mV)
on NTSC or PAL signal
p-p
Triggering (Option 05 Video Trigger)
1
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on
the selected slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not
“roll” across the screen on successive acquisitions. The TRIG’D LED stays constantly lighted when the SEC/DIV setting
is 2 ms or faster but may flash when the SEC/DIV setting is 10 ms or slower.
2
The waveform interval (WI) is the time between the samples in the waveform record. Also, see the footnote for the
characteristics Sample Rate Range and Equivalent Time or Interpolated Waveform Rate on page 1–8 and net offset on
page 1–14.
TDS 510A Service Manual
1–15
Warranted Characteristics
T able 1–10: Warranted Characteristics — Interfaces, Output Ports and Power Requirements
NameDescription
Logic Levels, Main- and Delayed-Trigger
Outputs
Characteristic
V
(HI)
out
V
(LO)
out
Limits
w2.5 V open circuit; w1.0 V into a 50 W load to ground
v0.7 V into a load of v4 mA;
v0.25 V into a 50 W load to ground
Output Voltage and Frequency,
Probe Compensator
Characteristic
Output Voltage
Frequency
Limits
0.5 V (base-top) ±1% into a w50 W load
1 kHz ±5%Output Voltage, Channel 3 Signal Out20 mV/division ±10% into a 1 MW load; 10 mV/division ±10% into a 50W load
Source Voltage90 to 250 VAC
CAT II, continuous range
RMS
Source Frequency45 Hz to 440 Hz
Power Consumptionv300 W (450 VA)
T able 1–11: Warranted Characteristics — Environmental, Safety , and Reliability
NameDescription
AtmosphericsTemperature:
Operating: +0_ C to +50_ C (disk drive operation limited to +4_ C minimum);
Non-operating: –22_ C to +60_ C
Relative humidity with floppy disk (optional):
Operating: To 80%, at or below +29_ C;
Operating: To 20%, at or below +50_ C
Non-operating: To 90%, at or below +40_ C;
Non-operating: To 5%, at or below +50_ C
Relative humidity without floppy disk:
Operating: To 95%, at or below +40_ C;
Operating: To 75%, from +41_ C to +55_ C
Altitude:
To 4570 m (15,000 ft.), operating;
To 12190 m (40,000 ft.), non-operating
DynamicsRandom vibration without floppy disk installed:
0.31 g
3.07 g
, from 5 to 500 Hz, 10 minutes each axis, operating;
RMS
, from 5 to 500 Hz, 10 minutes each axis, non-operating
RMS
User-Misuse SimulationElectrostatic Discharge Susceptibility: Up to 8 kV with no change to control settings or
impairment of normal operation; up to 15 kV with no damage that prevents recovery of
normal operation by the user
1–16
TDS 510A Service Manual
Specifications
T able 1–12: Certifications and compliances
EC Declaration of ConformityMeets intent of Directive 89/336/EEC for Electromagnetic Compatibility and Low Voltage Directive
73/23/ECC for Product Safety. Compliance was demonstrated to the following specifications as
listed in the Official Journal of the European Communities:
EN 50081-1 Emissions:
EN 55022Class B Radiated and Conducted Emissions
EN 60555-2AC Power Line Harmonic Emissions
EN 50082-1 Immunity:
IEC 801-2Electrostatic Discharge Immunity
IEC 801-3RF Electromagnetic Field Immunity
IEC 801-4Electrical Fast T ransient/Burst Immunity
IEC 801-5Power Line Surge Immunity
Low Voltage Directive 73/23/EEC:
EN 61010-1 Safety requirements for electrical equipment for measurement,
control, and laboratory use
FCC ComplianceEmissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits
CertificationsUnderwriters Laboratories certified to Standard UL311 1-1 for Electrical Measuring and Test
Equipment.
Canadian Standards Association certified to Standard CAN/CSA-C22.2 No. 1010.1-92, Safety
Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use.
T emperature (operating)0_ C to +50_ C
Altitude (maximum operating)2000 meters
Safety ClassTest and Measuring
CSA Certified Power CordsCSA Certification includes the products and power cords appropriate for use in the North America
power network. All other power cords supplied are approved for the country of use.
Overvoltage CategoryCategory:Examples of Products in this Category:
CA T IIIDistribution-level mains, fixed installation
CA T IILocal-level mains, appliances, portable equipment
CA T I
Pollution Degree 2Do not operate in environments where conductive pollutants may be present.
Signal levels in special equipment or parts of equipment, telecommunications, electronics
TDS 510A Service Manual
1–17
Specifications
1–18
TDS 510A Service Manual
Typical Characteristics
ettli
rrors
The tables in this section list the typical characteristics that describe the
TDS 510A Digitizing Oscilloscope.
Typical characteristics are described in terms of typical or average performance.
Typical characteristics are not warranted.
T able 1–13: Typical Characteristics — Signal Acquisition System
NameDescription
Frequency Limit, Upper, 100 MHz Bandwidth Limited
Frequency Limit, Upper , 20 MHz Bandwidth Limited
The numbers given are valid 0_C to +30_C and will increase as the temperature increases due to the degradation in
bandwidth. Rise time is calculated from the bandwidth. It is defined by the following formula:
Rise Time (ns) +
400
BW (MHz)
Note that if you measure rise time, you must take into account the rise time of the test equipment (signal source, etc.) that
you use to provide the test signal. That is, the measured rise time (RTm) is determined by the instrument rise time (RTi)
and the rise time of the test signal source (RTgen) according to the following formula:
2
RT
+ RT
m
2
The values given are the maximum absolute difference between the value at the end of a specified time interval after the
i
2
) RT
gen
2
mid-level crossing of the step, and the value one second after the mid-level crossing of the step, expressed as a
percentage of the step amplitude.
3Net Offset = Offset – (Position Volts/Div). Net Offset is the nominal voltage level at the center of the A-D converter
dynamic range. Offset Accuracy is the accuracy of this Voltage level.
T able 1–14: Typical Characteristics — Time Base System
NameDescription
Aperture Uncertaintyv5 ps
Accuracy, Delta Time MeasurementFor single-shot acquisitions using single-shot sample acquisition modes.
1
Time Measurement Error ≤ .15* Si + (25 ppm |Reading|) + (Time Per Div B 1000)
Add 50 ps (typical) Channel Skew for 2 channel measurements.
Sample calculation:
To measure the width of a 65.5 ns pulse at 1 GS/sec sampling,
Time Measurement Error
The limits are given for signals having pulse height ≥ 5 div, pulse duration v 10 divisions, reference level = 50%
mid-point, filter set to Sin (x) /x acquired at ≥ 5 mV/div, 1.4 vTr/Si v 5, where Si is the sample interval and tr is the signal
rise time.
1–20
TDS 510A Service Manual
T able 1–15: Typical Characteristics — T riggering System
NameDescription
Trigger Source
Any Channel
Auxiliary
Input, Auxiliary Trigger
Trigger Position Error , Edge T riggeringAcquire Mode
The input resistance is w1.5 kW; the maximum safe input voltage is ±20 V (DC + peak
AC).
Trigger-Position Error
Typical Characteristics
1,2
Sample, Hi-Res, Average
Peak Detect, Envelope
Holdoff, V ariable, Main Trigger
Minimum: For any horizontal scale setting, the minimum holdoff is 10 times that setting,
but is never less than 1 ms or longer than 5 s.
Maximum: For any horizontal scale setting, the maximum holdoff is at least 2 times the
minimum holdoff for that setting, but is never more than 10 times the minimum holdoff for
that setting.
Width, Minimum Pulse and Rearm, for
Logic Triggering or Events Delay
3
Lowest Frequency for Successful Operation
5 ns
30 Hz
of “Set Level to 50%” Function
Sensitivity, Edge Trigger, Not DC Coupled4Trigger Coupling
AC
Noise Reject
High Frequency Reject
Low Frequency Reject
Sensitivities, Logic-Type Trigger/Events
Delay , DC Coupled
6
Sensitivities, Pulse-T ype Runt Trigger
6
1.0 division, from DC to 100 MHz with a minimum slew rate of 25 divisions/ms at the
trigger level or the threshold crossing.
1.0 division, from DC to 200 MHz with a minimum slew rate of 25 divisions/ms at the
trigger level or the threshold crossing.
Sensitivities, Pulse-T ype Trigger Width and
6
Glitch
1.0 division with a minimum slew rate of 25 div/ms at the trigger level or the threshold
crossing. For <5 ms pulse width or rearm time, 2 divisions are required.
Sync Width (Option 05 Video Trigger)min. 400 ns for HDTV signals
Sync Duty Cycle
min. 50 to 1
(Option 05 Video Trigger)
±(1 WI+ 1 ns)
±(2 WI + 1 ns)
Typical Signal Level for Stable Triggering
Same as DC-coupled limits4 for frequencies above
60 Hz. Attenuates signals below 60 Hz.
Three and one half times the DC-coupled limits.
5
One and one half times the DC-coupled limits5 from
DC to 30 kHz. Attenuates signals above 30 kHz.
One and one half times the DC-coupled limits5 for
frequencies above 80 kHz. Attenuates signals below
80 kHz.
TDS 510A Service Manual
1–21
Typical Characteristics
T able 1–15: Typical Characteristics — T riggering System (Cont.)
NameDescription
Hum Rejection7
(Option 05 Video Trigger)
1
The trigger position errors are typically less than the values given here. These values are for triggering signals having a
NTSC and PAL: –20 dB without any trigger spec deterioration. Triggering will continue
down to 0 dB with some performance deterioration.
slew rate at the trigger point of ±0.5 division/ns.
2
The waveform interval (WI) is the time between the samples in the waveform record. Also, see the footnote for the
characteristics Sample Rate Range and Equivalent Time or Interpolated Waveform Rate on page 1–8 and Net Offset on
page 1–14.
3
The minimum pulse width and rearm width required for logic-type triggering or events delaying to occur.
4
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on
the selected slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not
“roll” across the screen on successive acquisitions. The TRIG’D LED stays constantly lighted when the SEC/DIV setting
is 2 ms or faster but may flash when the SEC/DIV setting is 10 ms or slower.
5
See the characteristic Sensitivity, Edge-T ype T rigger, DC Coupled in Table 1–9, which begins on page 1–15.
6
The minimum signal levels required for stable logic or pulse triggering of an acquisition or for stable counting of a
DC-coupled events delay signal. Also, see the footnote for Sensitivity, Edge-T ype T rigger, DC Coupled in this table.
(Stable counting of events is counting that misses no events or produces no extra events.)
7
The input signal is clamped at the bottom of the sync for the trigger path only. To remove the hum from the display , use a
Tektronix V ideo Display Clamp.
T able 1–16: Typical Characteristics — Data Handling
NameDescription
Time, Data-Retention, Nonvolatile
1,2
Memory
Battery life is w 5 years
Floppy disk3.5 inch, 720 K or 1.44 Mbyte, DOS 3.3-or-later compatible
1
The time that reference waveforms, stored setups, and calibration constants are retained when there is no power to the
oscilloscope.
2
Data is maintained by small lithium-thionyl-chloride batteries internal to the memory ICs. The amount of lithium is so
small in these ICs that they can typically be safely disposed of with ordinary garbage in a sanitary landfill.
1–22
TDS 510A Service Manual
Installation
This chapter contains information about supplying operating power, the
operating environment, applying and interrupting power, repackaging for
shipment, and installed options.
Supplying Operating Power
STOP. Read all information and heed all warnings in this section before
connecting the oscilloscope to a power source.
WARNING. AC POWER SOURCE AND CONNECTION. The oscilloscope
operates from a single-phase power source. It has a three-wire power cord and
two-pole, three-terminal grounding type plug. The voltage to ground (earth) from
either pole of the power source must not exceed the maximum rated operating
voltage, 250 volts.
Before making connection to the power source, be sure the oscilloscope has a
suitable two-pole, three-terminal grounding-type plug.
GROUNDING. This oscilloscope is safety Class 1 equipment (IEC designation).
All accessible conductive parts are directly connected through the grounding
conductor of the power cord to the grounded (earthing) contact of the power
plug.
WARNING. The power input plug must be inserted only in a mating receptacle
with a grounding contact where earth ground has been verified by a qualified
service person. Do not defeat the grounding connection. Any interruption of the
grounding connection can create an electric shock hazard.
For electric shock protection, the grounding connection must be made before
making connection to the oscilloscope input or output terminals.
TDS 510A Service Manual
2–1
Installation
Power Cord Information
Operating Voltage
Memory Backup Power
A power cord with appropriate plug configuration is supplied with each
oscilloscope. Table 2–1 gives the color-coding of the conductors in the power
cord. If you require a power cord other than the one supplied, refer to Table 2–2
Power-Cord and Plug Identification.
This oscilloscope operates with any line voltage from 90–250 VAC
RMS
with any
line frequency from 45–440 Hz. There are two fuses, either of which may be
used throughout the line voltage and frequency ranges. (The two fuses are not
totally interchangeable as each requires a different fuse cap.)
Memory modules with on-board batteries allow the oscilloscope to retain some
types of data upon loss of the AC power source. The stored adjustment
constants, saved front-panel settings, current front-panel settings (oscilloscope
status), and waveforms saved in memory are retained.
The on-board batteries of the memory modules have a shelf life of about five
years. Partial or total loss of stored settings upon power on may indicate that the
memory modules need to be replaced.
2–2
TDS 510A Service Manual
T able 2–2: Power Cord Identification
Plug ConfigurationNormal UsageOption Number
Installation
North America
115 V
Europe
230 V
United Kingdom
230 V
Australia
230 V
Standard
A1
A2
A3
TDS 510A Service Manual
North America
230 V
Switzerland
230 V
A4
A5
2–3
Installation
Operating Environment
The following environmental requirements are provided to ensure proper
operation and long oscilloscope life.
Operating Temperature
Ventilation Requirements
The oscilloscope can be operated where the ambient air temperature is between
+0_ and +50_ C and can be stored in ambient temperatures from –22_ to
+60_ C. Operation of the disk drive is limited to +4_ C minimum. After storage
at temperatures outside the operating limits, allow the chassis to stabilize at a
safe operating temperature before applying power.
The oscilloscope is cooled by air drawn in and exhausted through its cabinet side
panels by an internal fan. To ensure proper cooling of the oscilloscope, allow at
least two inches clearance on both sides and 3/4 inch on the bottom of the
oscilloscope. (The feet on the bottom of the oscilloscope provide the required
clearance when set on flat surfaces.) The top of the oscilloscope does not require
ventilation clearance.
CAUTION. If air flow is restricted, the oscilloscope power supply may
temporarily shut down.
2–4
TDS 510A Service Manual
Applying and Interrupting Power
Consider the following information when you power up or power down the
oscilloscope, or when power is interrupted due to an external power failure.
Installation
Power On
Power Off
Upon power on, the oscilloscope runs its power-on self check. If it passes, the
oscilloscope displays a “passed” status message and a prompt to press CLEAR
MENU to continue. If it fails, the oscilloscope displays a diagnostic log that
identifies the area(s) that failed and a prompt to press CLEAR MENU to
continue. See Chapter 6, Maintenance, for information on diagnostics and fault
isolation.
CAUTION. DO NOT power off the oscilloscope when either running a signal path
compensation or when doing any of the adjustments described in Chapter 5,
Adjustment Procedures. To do so might result in the loss of internally stored
adjustment constants.
In general, do not power off the oscilloscope when doing operations that affect
the data types listed in Table 2–3. Wait for the oscilloscope to finish the
operation when doing adjustments, saving waveforms, or saving setups.
Improper power off or unexpected loss of power to the oscilloscope can result in
the following corruptions of nonvolatile RAM (NVRAM). The following table
describes the messages displayed when power is restored after an abnormal
power off.
TDS 510A Service Manual
T able 2–3: Effects of Corrupted Data
Corrupted Data TypeResults
Adjustment Constants:
Signal Path CompensationA signal path compensation is required.
Voltage ReferenceA voltage reference adjustment is required
(Chapter 5).
Frequency ResponseA frequency response adjustment is
required (Chapter 5).
Error LogErrors logged are lost.
Reference WaveformsWaveform Lost
Saved SetupsSetup Lost
2–5
Installation
Repackaging Instructions
Use a corrugated cardboard shipping carton having a test strength of at least
126 g (275 pounds) and with an inside dimension at least 15 cm (six inches)
greater than the oscilloscope dimensions.
If the oscilloscope is being shipped to a Tektronix Service Center, enclose the
following information: the owners address, name and phone number of a contact
person, type and serial number of the oscilloscope, reason for returning, and a
complete description of the service required.
Seal the shipping carton with an industrial stapler or strapping tape.
Mark the address of the Tektronix Service Center and also your own return
address on the shipping carton in two prominent locations.
Installed Options
Your oscilloscope may be equipped with one or more options. Except for the
line-cord options described by Table 2–2 (on page 2–3 of this section), all
options and optional accessories are listed and described in Chapter 7, Options.
For further information and prices of options, see your Tektronix Products
catalog or contact your Tektronix Field Office.
2–6
TDS 510A Service Manual
Operating Instructions
This section contains general operating instructions for the TDS 510A Digitizing
Oscilloscope. The complete user instructions are found in the User Manual.
Additional instructions are integrated into the service procedures found in later
sections of this manual. For instance, the procedures found in Chapter 4
Performance Verification contain instructions for making the front-panel settings
required to check each instrument characteristic included there. Also, the general
instructions for operating internal diagnostic routines are found in Chapter 6,
Maintenance. You may also find the Product Description in Chapter 1 useful for
understanding how the oscilloscope functions.
Screen Layout
The screen layout is illustrated in Figure 2–1 on page 2–8. Note that the figure
illustrates a full graticule; you may also select a grid, crosshair, or frame
graticule from the display menu.
Basic Procedures
How to Power On
Before doing service, read the following operating instructions. These instructions are at the level appropriate for servicing this oscilloscope.
Push the principal power switch found on the rear panel of the oscilloscope, then
push the ON/STBY (standby) switch to toggle the oscilloscope into operation.
The switch at the rear panel is the true power disconnect switch. The ON/STBY(standby) switch simply toggles operation on and off.
WARNING. The principal power switch at the rear panel is the true power
disconnect switch. The ON/STBY (standby) switch simply toggles operation on
and off. When connected to a power source and when the principal power switch
is on, the internal power supplies and other circuitry of this oscilloscope are
energized regardless of the setting of the ON/STBY switch.
TDS 510A Service Manual
2–7
Operating Instructions
Brief status information
Graticule and waveforms
Waveform reference
symbols show ground levels
and waveform sources
Vertical scale, horizontal scale,
and trigger level readouts
Pop-up menu
Position of waveform
record relative to
the screen and display
General purpose
knob readout
Side menu area.
Readouts for
measurements
move here when
CLEAR MENU
is pressed.
How to Use Help
How to Use the Status
Menu
Main menu display area. Readouts in lower graticule
area move here when CLEAR MENU is pressed.
Figure 2–1: Map of Display Functions
Push the HELP front-panel button to enter help mode. Front-panel knobs and
buttons now display information about their function when turned or pushed.
Push HELP again to exit help mode.
To get help information on a menu item, display the menu desired (if you are in
help mode, exit help first). Push HELP. Now the menu buttons display
information about their function when pushed.
Push the SHIFT, then the STATUS front-panel buttons to display the status
menu. You will find messages reflecting the state of the acquisition system,
whether it is running or stopped (and if it is stopped, why), as well as setup-related information.
2–8
TDS 510A Service Manual
Operating Instructions
ertical
ctio
trols: Access
orizontal
ctio
trols
rigger
ctio
trols
How to Set Functions
V
all vertical functions and inputs from
Special Function Controls:
Access all functions except
vertical, horizontal, and trigger
functions from this front-panel
control block.
Most functions can be set either by using one or two front-panel buttons or
knobs, or by pushing a front-panel button to use a main menu, and then a side
menu to set the function. The following steps illustrate both procedures.
1. Locate the block that contains the function to be set.
Fun
n Con
this front-panel control block.
Fun
H
Access all horizontal functions
from this front-panel block.
n Con
:
Access all trigger functions from
Fun
T
this front-panel block.
n Con
:
TDS 510A Service Manual
2–9
Operating Instructions
2. Select the waveform source(s). Position, scale, and set trigger level for
waveform source(s) directly from the front-panel. (Examples of possible
control selections are given in the steps that follow.)
4231
8
7
56
a. Input waveforms into these channels (7). Example: CH 1.
b. Push any channel button (8) to display its waveform. The last channel
turned on determines which waveform is positioned and scaled. The
indicator above the channel last selected is lighted. Example: Push
CH 1; then CH 2.
c. Vertically (1) and horizontally (2) scale and position the waveform(s)
selected. Example: Set the scale to 100 mV/div and center the waveform
on screen.
2–10
d. Stop and start acquiring waveforms (3). Example: Push RUN/STOP if
not acquiring.
TDS 510A Service Manual
Operating Instructions
e. Adjust trigger level (6) to trigger the waveform(s) selected or use these
buttons (5) to either set a trigger level at the mid-amplitude level of the
selected waveform or to force a single trigger. Example: Push SET
LEVEL TO 50%.
3. Set all other functions using menus.
a. Choose the waveform source (8) first if setting a vertical function;
otherwise skip to step b. Example: Push CH 2.
b. Push SHIFT (4) if the function to be set is highlighted in blue;
otherwise skip to step c.
9
14
13
12
c. Push the front-panel button that corresponds to the menu containing the
function. A main menu (14) for selecting among related functions
appears. Example: Push VERTICAL MENU.
d. Select a function from the main menu using the main menu buttons (12).
A side menu for selecting among that the available settings for that
function appears. Example: Push Coupling (13).
2–11
Operating Instructions
e. Select the setting for the function from the side menu (9) using the side
menu buttons (11). Example: Push AC (10).
How to Set Complex
Functions
A few functions require more than just two levels (main and side) of menus to
completely specify their settings. In such cases, either the main menu, the side
menu, or both are modified to supply additional choices. The procedures that
follow show both schemes.
1. Set up a function using pop-up menus:
15
18
2–12
17
a. For some selections, pushing a main-menu button pops up a menu (18)
b. Pushing the button that popped up the menu (17) toggles through the
pop-up menu choices. Example: Repeatedly push Type to toggle through
the pop-up menu. Notice the other main-menu button labels change
accordingly. Toggle back to Edge.
c. Complete the setting of the desired mode by selecting from the main
menu and the side menu that results. Example: Push Mode & Holdoff
(16), and then push Normal (15).
TDS 510A Service Manual
16
Operating Instructions
2. Set up a function using the general purpose knob (20). (The examples of
possible menu selections in substeps that follow assume you’ve pushed
TRIGGER MENU.)
1920
21
a. Pushing some main menu buttons displays a side menu with labels
containing readouts that can be varied. Example: Push Level (21).
b. Pushing the side-menu button assigns the knob to control the readout
appearing in the button label. It also copies the readout to the general
purpose knob readout area in the right corner of the screen. Example:
Push Level (19).
c. Use the general purpose knob (20) to adjust the trigger level to the
setting desired. Example: Turn the knob to –20 mV.
More About the General Purpose Knob. As you just saw, the general purpose knob
can be used to extend the number of choices available to a side-menu button. The
general purpose knob can also be assigned to control the following functions:
TDS 510A Service Manual
HCursor positioning
HDisplay intensities
HDelay time
2–13
Operating Instructions
HNumber of events
HTrigger position
HHoldoff
HOffset
HVariable persistence
In all cases, the menus are used to select the function to which the general
purpose knob is assigned. The following attributes apply to this knob:
HDepending on the function it is assigned to control, the general purpose knob
may vary numerical readouts, position objects (cursors) on screen, or select
between icon-label settings that show up in side-menu labels. Once assigned,
the general purpose knob can change parameters by toggling the SHIFT
button.
HThe general purpose knob has a readout area at the upper right corner of the
screen. (See Figure 2–1.) This readout always reflects the name and value of
the function that the general purpose knob is currently controlling.
HWhenever the general purpose knob assignment is changed, a knob icon
appears immediately to the left of the general purpose knob readout to notify
you of the assignment change. The icon is removed as soon as you use the
general purpose knob to change the value of the function it is assigned to.
HTo assign the general purpose knob to control a function, display the menu
containing the function; then select the function. (Note that not all functions
can be controlled by the general purpose knob.)
HWhenever the menu is removed, the general purpose knob is not assigned
and does not control any function. (An exception is the cursor function. If
cursors are turned on, removing the menu leaves the knob assigned to
control the cursors until reassigned by selecting another menu and function
that uses the knob.)
HThe general purpose knob also has a SELECT button. This button is used to
toggle the knob between the control of either of the two cursors displayed
when cursors are turned on in the cursor menu.
Display and Utility Menus. Using the techniques described for using menus, you
can access and change functions in the display menu and utilities menu. In the
Display menu, you can set the following functions:
HIntensity: waveforms, readouts, graticule.
2–14
HStyle of waveform display(s): vectors or dots, intensified or non-intensified
samples, and infinite or variable persistence.
TDS 510A Service Manual
Operating Instructions
HDisplay format: XY or YT.
HGraticule format: type.
HWaveform interpolation filter and readout options.
From the Utility menu, you can configure the GPIB port (talk/listen, address,
etc.) and access internal routines for self diagnostics and self compensation.
Instructions for setting up communication over the GPIB are found in Chapter 5,
Adjustment Procedures.
TDS 510A Service Manual
2–15
Operating Instructions
2–16
TDS 510A Service Manual
Circuit Description
This chapter describes the electrical operation of the TDS 510A Digitizing
Oscilloscope using the major circuit blocks or modules.
This chapter has two main parts:
HLogic Conventions describes how logic functions are discussed and
HModule Overview describes circuit operation from a functional-circuit block
Logic Conventions
The oscilloscope contains many digital logic circuits. This manual refers to these
circuits with standard logic symbols and terms. Unless otherwise stated, all logic
functions are described using the positive-logic convention: the more positive of
the two logic levels is the high (1) state, and the more negative level is the low
(0) state. Signal states may also be described as “true” meaning their active state
or “false” meaning their non-active state. The specific voltages that constitute a
high or low state vary among the electronic devices.
represented in this manual.
perspective.
Module Overview
General
Input Signal Path
Active-low signals are indicated by a tilde prefixed to the signal name (~RESET). Signal names are considered to be either active-high, active-low, or to
have both active-high and active-low states.
This module overview describes the basic operation of each functional circuit
block as shown in Figure 9–2 on page 9–2.
The Tektronix TDS 510A Digitizing Oscilloscope is a portable, four-channel
instrument. Each channel provides a calibrated vertical scale factor. All of the
four channels can be acquired simultaneously.
A signal enters the oscilloscope through a probe connected to a BNC on the A15
Attenuator board.
Attenuators. Circuitry in the attenuator selects the input coupling, termination,
and the attenuation factor. The processor system, by way of the acquisition
system, controls the attenuators. For example, if 50 W input termination is
TDS 510A Service Manual
3–1
Circuit Description
selected and the input is overloaded, the processor system switches the input to
the 1 MW position.
Probe Coding Interface. Probe coding interface signals pass through the A15
Attenuator to the acquisition system, where they are sensed and controlled.
Acquisition System. The acquisition system amplifies the input signals, converts
them to digital signals, and controls the acquisition process under direction of the
processor system. The acquisition system includes the trigger, acquisition timing,
and acquisition mode generation and control circuitry.
D1 Bus. The acquisition system passes the digital values representing the
acquired waveform through the A14 D1 Bus to the A11 DRAM Processor/Display board. This happens after a waveform acquisition is complete if the digital
signal processor in the processor system requests the waveform.
Voltage Controlled
Oscillator (VCO)
Processor System. The processor system contains a 68020 microprocessor that
controls the entire instrument. It includes the firmware. It also includes a GPIB
interface. You can reprogram the firmware from a remote controller using the
GPIB and an external software package.
The processor also includes a digital signal processor. This signal processor
processes each waveform as directed by the system processor. Waveforms and
any text to be displayed are passed on to the display system. The A11 DRAM
Processor/Display board contains both the processor and display systems.
Display System. Text and waveforms are processed by different parts of the
display circuitry. The display system sends the text and waveform information to
the tube assembly as a video signal. The display system also generates and sends
vertical (VSYNC) and horizontal (HSYNC) sync signals to the tube assembly. A
VGA-compatible video output is at the rear of the oscilloscope.
Master clocks for the acquisition system are generated by a voltage controlled
oscillator on the A21 VCO Board and circuitry on the A10 Acquisition board.
The circuitry makes up a phased locked loop. The master clock is divided down
by the A10 Acquisition circuitry under control of the processor system.
3–2
Tube Assembly
All information (waveforms, text, graticules, and pictographs) is displayed by
the A20 Display system. The Display system generates the high voltages
necessary to drive the display tube. It also contains the video amplifier,
horizontal oscillator, and the vertical and horizontal yoke driver circuitry.
TDS 510A Service Manual
Circuit Description
Front Panel
Rear Panel
Low V oltage Power Supply
The processor system sends instructions to and receives information from the
Front-Panel Processor on the A12 Front Panel board. The Front-Panel Processor
reads the front-panel switches and potentiometers. Any changes in their settings
are reported to the processor system. The Front-Panel Processor also turns the
LEDs on and off and generates the bell signal.
Front-panel menu switches are also read by the Front-Panel Processor. The
processor sends any changes in menu selections to the processor system. The
ON/STBY switch is one of the menu switches. However, it is not read by the
Front-Panel Processor, but passes through the A12 Front Panel board and the
A11 DRAM Processor/Display board to the low voltage power supply.
The front panel also generates the probe compensation signals SIGNAL
and GND.
The GPIB connector provides access to stored waveforms, and allows external
control of the oscilloscope.
You can make hardcopies on the GPIB, RS-232 and Centronics ports.
The low voltage power supply is a switching power converter with active power
factor control. It supplies power to all of the circuitry in the oscilloscope.
Fan
The principal POWER switch, located on the rear panel, controls all power to
the oscilloscope including the Low Voltage Power Supply. The ON/STBY
switch, located on the front panel, also controls all of the power to the oscilloscope except for part of the circuitry in the Low Voltage Power Supply.
If the power is going off, the power supply sends a power fail (~PF) warning to
the processor system.
The fan provides forced air cooling for the oscilloscope. It connects to +25 V
from the Low Voltage Power Supply by way of the A11 DRAM Processor/Display module.
TDS 510A Service Manual
3–3
Circuit Description
3–4
TDS 510A Service Manual
Performance Verification Procedures
Two types of Performance Verification procedures can be performed on this
product; Brief Procedures and Performance Tests. You may not need to perform
all of these procedures, depending on what you want to accomplish.
HTo rapidly confirm that the oscilloscope functions and was adjusted properly,
just do the brief procedures under Self Tests, which begin on page 4–5.Advantages: These procedures are quick to do, require no external
equipment or signal sources, and perform extensive functional and accuracy
testing to provide high confidence that the oscilloscope will perform
properly. They can be used as a quick check before making a series of
important measurements.
HTo further check functionality, first do the Self Tests just mentioned; then do
the brief procedures under Functional Tests that begin on page 4–7.Advantages: These procedures require minimal additional time to perform,
require no additional equipment other than a standard-accessory probe, and
more completely test the internal hardware of the oscilloscope. They can be
used to quickly determine if the oscilloscope is suitable for putting into
service, such as when it is first received.
HIf more extensive confirmation of performance is desired, do the Perfor-
mance Tests, beginning on page 4–15, after doing the Functional and Self
Tests just referenced.
Advantages: These procedures add direct checking of warranted specifications. They require more time to perform and suitable test equipment is
required. (See Equipment Required beginning on page 4–16.)
If you are not familiar with operating this oscilloscope, read the user manual.
These contain instructions that will acquaint you with the use of the front-panel
controls and the menu system.
TDS 510A Service Manual
4–1
Performance Verification Procedures
Conventions
Throughout these procedures the following conventions apply:
HEach test procedure uses the following general format:
HEach procedure consists of as many steps, substeps, and subparts as required
Title of Test
Equipment Required
Prerequisites
Procedure
to do the test. Steps, substeps, and subparts are sequenced as follows:
1. First Step
a. First Substep
HFirst Subpart
HSecond Subpart
b. Second Substep
2. Second Step
HIn steps and substeps, the lead-in statement in italics instructs you what to
do, while the instructions that follow tell you how to do it, as in the example
step below, “Initialize the oscilloscope” by doing “Press save/recall SETUP.
Now, press the main-menu button...”.
Initialize the oscilloscope: Press save/recall SETUP. Now, press the
main-menu button Recall Factory Setup; then the side-menu button OKConfirm Factory Init.
HWhere instructed to use a front-panel button or knob, or select from a main
or side menu, or verify a readout or status message, the name of the button or
knob appears in boldface type: “press SHIFT; then UTILITY, press the
main-menu button System until Cal is highlighted in the pop-up menu.
Verify that the status message is Pass in the main menu under the VoltageReference label.”
STOP. The symbol at the left is accompanied by information you must read to do
the procedure properly.
4–2
HRefer to Figure 4–1: “Main menu” refers to the menu that labels the seven
menu buttons under the display; “side menu” refers to the menu that labels
TDS 510A Service Manual
Performance Verification Procedures
the five buttons to the right of the display. “Pop-up menu” refers to a menu
that pops up when a main-menu button is pressed.
Brief status information
Graticule and waveforms
Waveform reference
symbols show ground levels
and waveform sources
Vertical scale, horizontal scale,
and trigger level readouts
Pop-up menu
Position of waveform
record relative to
the screen and display
General purpose
knob readout
Side menu area.
Readouts for
measurements
move here when
CLEAR MENU
is pressed.
Figure 4–1: Map of Display Functions
Main menu display area. Readouts in lower graticule
area move here when CLEAR MENU is pressed.
TDS 510A Service Manual
4–3
Performance Verification Procedures
4–4
TDS 510A Service Manual
Brief Procedures
Self Tests
The Self Tests use internal routines to confirm basic functionality and proper
adjustment. No test equipment is required to do these test procedures.
The Functional Tests utilize the probe-compensation output at the front panel as
a test-signal source for further verifying that the oscilloscope functions properly.
A probe, such as the P6139A, is required to do these test procedures.
This procedure uses internal routines to verify that the oscilloscope functions and
was adjusted properly. No test equipment or hookups are required.
Verify Internal Adjustment,
Self Compensation, and
Diagnostics
Equipment
Required
PrerequisitesSee page 4–15.
1. Verify that internal diagnostics pass: Do the following substeps to verify
passing of internal diagnostics.
a. Display the System diagnostics menu:
HPress SHIFT; then press UTILITY.
HRepeatedly press the main-menu button System until Diag/Err is
highlighted in the pop-up menu.
b. Run the System Diagnostics:
HFirst disconnect any input signals from all four channels.
HPress the main-menu button Execute; then press the side-menu
button OK Confirm Run Test.
c. Wait: The internal diagnostics do an exhaustive verification of proper
oscilloscope function. This verification will take up to three and a half
minutes. At some time during the wait, a “clock” icon (shown at left)
will appear on-screen. When the verification is finished, the resulting
status will appear on the screen.
None
TDS 510A Service Manual
d. Confirm no failures are found: Verify that no failures are found and
reported on-screen.
e. Confirm the three adjustment sections have passed status:
4–5
Brief Procedures
HPress SHIFT; then press UTILITY.
HHighlight Cal in the pop-up menu by repeatedly pressing the
main-menu button System. See Figure 4–2.
HVerify that the word Pass appears in the main menu under the
following menu labels: Voltage Reference, Frequency Response,
and Pulse Trigger. See Figure 4–2.
f.Run the signal-path compensation: Press the main-menu button Signal
Path; then press the side-menu button OK Compensate Signal Paths.
g. Wait: Signal-path compensation may take five minutes to run. While it
progresses, a “clock” icon (shown at left) is displayed on-screen. When
compensation completes, the status message will be updated to Pass or
Fail in the main menu. See step h.
h. Confirm signal-path compensation returns passed status: Verify that the
word Pass appears under Signal Path in the main menu. See Figure 4–2.
4–6
1
Highlight Cal
3
Verify Pass
2
Verify Pass
Figure 4–2: Verifying Adjustments and Signal-Path Compensation
2. Return to regular service: Press CLEAR MENU to exit the system menus.
TDS 510A Service Manual
Functional Tests
Brief Procedures
The purpose of these procedures is to confirm that the oscilloscope functions
properly. The only equipment required is one of the standard-accessory probes
and, to check the file system, a 3.5 inch, 720 K or 1.44 Mbyte floppy disk.
STOP. These procedures verify functions; that is, they verify that the oscilloscope
features operate. They do not verify that they operate within limits.
Therefore, when the instructions in the functional tests that follow call for you to
verify that a signal appears on-screen “that is about five divisions in amplitude”
or “has a period of about six horizontal divisions,” etc., do NOT interpret the
quantities given as limits. Operation within limits is checked in Performance
Tests, which begin on page 4–15.
STOP. DO NOT make changes to the front-panel settings that are not called out
in the procedures. Each verification procedure will require you to set the
oscilloscope to certain default settings before verifying functions. If you make
changes to these settings, other than those called out in the procedure, you may
obtain invalid results. In this case, just redo the procedure from step 1.
Verify All Input Channels
When you are instructed to press a menu button, the button may already be
selected (its label will be highlighted). If this is the case, it is not necessary to
press the button.
Equipment
Required
PrerequisitesNone
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground (typically black) to PROBE COMPENSA-TION GND. If using a P6243 or P6245 probe, you may want to attach a
Y-lead connector and two SMD KlipChips as shown in Figure 4–3.
One probe such as the P6243, P6245 or P6139A
TDS 510A Service Manual
4–7
Brief Procedures
SignalGnd
Figure 4–3: Universal T est Hookup for Functional Tests
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
2. Verify that all channels operate: Do the following substeps — test CH 1
first, skipping substepa and b since CH 1 is already set up for verification
and as the trigger source from step 1.
a. Select an unverified channel:
HPress WAVEFORM OFF to remove the channel just verified from
display.
HPress the front-panel button that corresponds to the channel you are
to verify.
HMove the probe to the channel you selected.
b. Match the trigger source to the channel selected:
4–8
HPress TRIGGER MENU.
TDS 510A Service Manual
Brief Procedures
HPress the main-menu button Source.
HPress the side-menu button that corresponds to the channel selected
(Ch2, Ch3, or Ch4).
c. Set up the selected channel:
HSet the vertical SCALE to 200 mV.
HSet the horizontal SCALE to 200 ms. Press CLEAR MENU to
remove any menu that may be on the screen.
HPress SET LEVEL TO 50%.
d. Verify that the channel is operational: Confirm that the following
statements are true.
HThe vertical scale readout for the channel under test shows a setting
of 200 mV, and a square-wave probe-compensation signal about
2.5 divisions in amplitude is on-screen. See Figure 4–1 on page 4–3
to locate the readout.
HThe vertical POSITION knob moves the signal up and down the
screen when rotated.
HTurning the vertical SCALE knob counterclockwise decreases the
amplitude of the waveform on-screen, turning the knob clockwise
increases the amplitude, and returning the knob to 200 mV returns
the amplitude to about 2.5 divisions.
e. Verify that the channel acquires in all acquisition modes: Press SHIFT;
then press ACQUIRE MENU. Use the side menu to select, in turn, each
of the three hardware acquire modes and confirm that the following
statements are true. Refer to the icons at the left of each statement as you
confirm those statements.
HSample mode displays an actively acquiring waveform on-screen.
(Note that there is noise present on the peaks of the square wave.)
HPeak Detect mode displays an actively acquiring waveform
on-screen with the noise present in Sample mode “peak detected.”
HHi Res mode displays an actively acquiring waveform on-screen
with the noise that was present in Sample mode reduced.
HEnvelope mode displays an actively acquiring waveform on-screen
with the noise displayed.
TDS 510A Service Manual
HAverage mode displays an actively acquiring waveform on-screen
with the noise reduced.
f.Test all channels: Repeat substeps a through e until all four input
channels are verified.
4–9
Brief Procedures
3. Remove the test hookup: Disconnect the probe from the channel input and
the probe-compensation terminals.
Verify the Time Base
Equipment
Required
PrerequisitesNone
One probe such as the P6243, P6245 or P6139A
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. See
Figure 4–3 on page 4–8.
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup; then press the
side-menu button OK Confirm Factory Init.
c. Modify default settings:
HSet the vertical SCALE to 200 mV.
HSet the horizontal SCALE to 200 ms.
HPress SET LEVEL TO 50%.
HPress CLEAR MENU to remove the menus from the screen.
2. Verify that the time base operates: Confirm the following statements.
a. One period of the square-wave probe-compensation signal is about five
horizontal divisions on screen for the 200 ms horizontal scale setting (set
in step 1c).
b. Rotating the horizontal SCALE knob clockwise expands the waveform
on-screen (more horizontal divisions per waveform period), counterclockwise rotation contracts it, and returning the horizontal scale to
200 ms returns the period to about five divisions.
c. The horizontal POSITION knob positions the signal left and right on
screen when rotated.
3. Remove the test hookup: Disconnect the probe from the channel input and
the probe-compensation terminals.
4–10
TDS 510A Service Manual
Brief Procedures
Verify the Main and
Delayed Trigger Systems
Equipment
Required
PrerequisitesNone
One probe such as the P6243, P6245, or P6139A
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. See
Figure 4–3 on page 4–8.
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
c. Modify default settings:
HSet the vertical SCALE to 200 mV.
HSet the horizontal SCALE for the M (main) time base to 200 ms.
HPress SET LEVEL TO 50%.
HPress TRIGGER MENU.
HPress the main-menu button Mode & Holdoff.
HPress the side-menu button Normal.
HPress CLEAR MENU to remove the menus from the screen.
2. Verify that the main trigger system operates: Confirm that the following
statements are true:
HThe trigger level readout for the main trigger system changes with the
trigger-LEVEL knob.
HThe trigger-LEVEL knob can trigger and untrigger the square-wave
signal as you rotate it. (Leave the signal untriggered, which is indicated
by the display not updating.)
HPressing SET LEVEL TO 50% triggers the signal that you just left
untriggered. (Leave the signal triggered.)
TDS 510A Service Manual
4–11
Brief Procedures
3. Verify that the delayed trigger system operates:
a. Select the delayed time base:
HPress HORIZONTAL MENU.
HPress the main-menu button Time Base.
HPress the side-menu button Delayed Triggerable; then press the
side-menu button Delayed Only.
HSet the horizontal SCALE for the D (delayed) time base to 200 ms.
b. Select the delayed trigger level menu:
HPress SHIFT; then press DELAYED TRIG.
HPress the main-menu button Level; then press the side-menu button
Level.
c. Confirm that the following statements are true:
HThe trigger-level readout for the delayed trigger system changes as
you turn the general purpose knob.
HAs you rotate the general purpose knob, the square-wave probe-com-
pensation signal can become triggered and untriggered. (Leave the
signal untriggered, which is indicated by the display not updating.)
HPressing the side-menu button Set to 50% triggers the probe-com-
pensation signal that you just left untriggered. (Leave the signal
triggered.)
d. Verify the delayed trigger counter:
HPress the main-menu button Delay by Time.
HUse the keypad to enter a delay time of 1 second. Press 1, then press
ENTER.
HVerify that the trigger READY indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
4. Remove the test hookup: Disconnect the probe from the channel input and
the probe-compensation terminals.
4–12
TDS 510A Service Manual
Brief Procedures
Verify the File System
Equipment
Required
PrerequisitesNone
One probe such as the P6243, P6245 or P6139A
One 720 K or 1.44 Mbyte, 3.5 inch DOS-compatible disk.
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. See
Figure 4–3 on page 4–8.
b. Insert the test disk: Insert the disk in the disk drive to the left of the
monitor.
HPosition the disk so the metal shutter faces the drive.
HPosition the disk so the stamped arrow is on the top right side. In
other words, place the angled corner in the front bottom location.
HPush the disk into the drive until it goes all the way in and clicks
into place.
c. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
d. Modify default settings:
HSet the vertical SCALE to 200 mV.
HSet the horizontal SCALE for the M (main) time base to 200 ms.
Notice the waveform on the display now shows two cycles instead of
five.
HPress SET LEVEL TO 50%.
HPress CLEAR MENU to remove the menus from the screen.
e. Save the settings:
HPress SETUP.
HPress the main-menu button Save Current Setup; then press the
side-menu button To File.
HTurn the general purpose knob to select the file to save. Choose
(or ). With this choice, you will save a file
starting with , then containing 5-numbers, and a extension.
TDS 510A Service Manual
4–13
Brief Procedures
For example, the first time you run this on a blank, formatted disk or
on the Example Programs Disk, the oscilloscope will assign the
name to your file. If you ran the procedure again, the
oscilloscope would increment the name and call the file
.
HPress the side-menu button Save To Selected File.
2. Verify the file system works:
HPress the main-menu button Recall Factory Setup and the side-menu
button OK Confirm Factory Init to restore the 500 ms time base and
the five cycle waveform.
HPress the main-menu button Recall Saved Setup; then press the
side-menu button From File.
HTurn the general purpose knob to select the file to recall. For example, if
you followed the instructions above and used a blank disk, you had the
oscilloscope assign the name TEK00000.SET to your file.
HPress the side-menu button Recall From Selected File.
HVerify that oscilloscope retrieved the saved setup from the disk. Do this
by noticing the horizontal SCALE for the M (main) time base is again
200 ms and the waveform shows only two cycles just as it was when you
saved the setup.
3. Remove the test hookup:
HDisconnect the probe from the channel input and the probe-compensa-
tion terminals.
HRemove the disk from the disk drive. Do this by pushing in the tab of
the disk drive.
4–14
TDS 510A Service Manual
Performance Tests
This section contains a collection of procedures for checking that the TDS 510A
Digitizing Oscilloscope performs as warranted.
The procedures are arranged in four logical groupings: Signal Acquisition System
Checks, Time Base System Checks, Triggering System Checks, and Output Ports
Checks. They check all the characteristics that are designated as checked in
Specifications. (The characteristics that are checked appear in boldface typeunder Warranted Characteristics in Specifications.)
STOP. These procedures extend the confidence level provided by the basic
procedures described on page 4–5. The basic procedures should be done first,
then these procedures performed if desired.
Prerequisites
The tests in this section comprise an extensive, valid confirmation of performance and functionality when the following requirements are met:
HThe cabinet must be installed on the oscilloscope.
HYou must have performed and passed the procedures under Self Tests, found
on page 4–5, and those under Functional Tests, found on page 4–7.
HA signal-path compensation must have been done within the recommended
calibration interval and at a temperature within ±5_ C of the present
operating temperature. (If at the time you did the prerequisite Self Tests, the
temperature was within the limits just stated, consider this prerequisite met.)
HThe oscilloscope must have been last adjusted at an ambient temperature
between +20_ C and +30_ C, must have been operating for a warm-up
period of at least 20 minutes, and must be operating at an ambient temperature between +5_ C and +50_ C. (The warm-up requirement is usually met
in the course of meeting the Self Tests and Functional Tests prerequisites
listed above.)
TDS 510A Service Manual
4–15
Performance Tests
Equipment Required
T able 4–1: Test Equipment
Item Number and
Description
These procedures use external, traceable signal sources to directly check
warranted characteristics. The required equipment list follows this introduction.
19. Generator, Sine Wave100 kHz to at least 400 MHz. Variable
75 W, 36 in, male to male BNC
connectors
Impedance 75 W; connectors: female
BNC input, male BNC output
amplitude from 12 mV to 2 V
p-p
.
Tektronix part number
012-1338-00
Tektronix part number
01 1-0102-01
Rohde & Schwarz SMY
Frequency accuracy >2.0%
3
Used to Test Video Option 05
Equipped Instruments Only
Used to Test Video Option 05
Equipped Instruments Only
Used to Test Video Option 05
Equipped Instruments Only
Checking Analog Bandwidth,
Trigger Sensitivity , Samplerate, External Clock, and
Delay-Time Accuracy
20. Meter, Level and Power
Sensor
21. Splitter, PowerFrequency range: DC to 1 GHz.
Frequency range:10 MHz to 400MHz.
Amplitude range: 6 mVp-p to 2 V
p-p
Rohde & Schwarz URV 35,
with NRV-Z8 power sensor
Rohde & Schwarz RVZ
3
Checking Analog Bandwidth
3
and Trigger Sensitivity
Checking Analog Bandwidth
Tracking: >2.0%
22. Generator, FunctionFrequency range 5 MHz to 10 MHz.
T ektronix CFG280Checking External Clock
Square wave transition time v25 ns.
Amplitude range: 0 to 10 V
p-p
into 50 W
23. Adapter (four required)Male N to female BNCT ektronix 103–0045–00Checking Analog Bandwidth
24. AdapterFemale N to male BNCT ektronix 103–0058–00Checking Analog Bandwidth
25. Generator,
Leveled Sine Wave,
200 kHz to 250 MHz; Variable amplitude from 5 mV to 4 V
into 50 W
p-p
Tektronix SG 503 Leveled
Sine Wave Generator
1, 3
Checking Trigger Sensitivity
at low frequencies
Medium-Frequency
(optional)
26. Generator, Leveled Sine
Wave, High-Frequency
(optional)
1
Requires a TM 500 or TM 5000 Series Power Module Mainframe.
2
Warning: The optional P6243 and P6245 probes that may be used with this oscilloscope provide an extremely low loading
250 MHz to 1 GHz; Variable amplitude from 500 mV to 4 V
into 50 W;
p-p
6 MHz reference
Tektronix SG 504 Leveled
Sine Wave Generator
SG 504 Output Head
1
3
with
Checking Analog Bandwidth
and Trigger Sensitivity at high
frequencies
capacitance (<1 pF) to ensure the best possible signal reproduction. These probes should not be used to measure
signals exceeding ±8 V , or errors in signal measurement will be observed. Above 40 V, damage to the probe may result.
To make measurements beyond ±8 V, use either the P6139A probe (good to 500 V), or refer to the catalog for a recommended probe.
3
You can replace items 19, 20, or 21 with a Tektronix SG503 (item 25) or SG504 (item 26) – if available.
TDS 510A Service Manual
4–17
Performance Tests
4–18
TDS 510A Service Manual
Performance Tests
TDS 510A Test Record
Photocopy this and the next three pages and use them to record the performance
test results for your TDS 510A Digitizing Oscilloscope.
T est Record
Instrument Serial Number:Certificate Number:
Temperature:RH %:
Date of Calibration:Technician:
Performance TestMinimumIncomingOutgoingMaximum
Offset Accuracy
CH1 Offset+1 mV
+101 mV
+1.01 V
CH2 Offset+1 mV
+101 mV
+1.01 V
CH3 Offset+1 mV
+101 mV
+1.01 V
CH4 Offset+1 mV
+101 mV
+1.01 V
DC Voltage Measurement Accuracy (Averaged)
CH1 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
CH1 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
CH1200 mV Vert scale setting,
–5 Div position setting, +10 V offset
CH1200 mV Vert scale setting,
+5 Div position setting, –10 V offset
CH11 V Vert scale setting,
–5 Div position setting, +10 V offset
CH11 V Vert scale setting,
+5 Div position setting, –10 V offset
CH2 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
CH2 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
CH2200 mV Vert scale setting,
–5 Div position setting, +10 V offset
– 1.6 mV
– 25.1 mV
– 251 mV
– 1.6 mV
– 25.1 mV
– 251 mV
– 1.6 mV
– 25.1 mV
– 251 mV
– 1.6 mV
– 25.1 mV
– 251 mV
+ 1.0355 V____________________+ 1.0445 V
– 1.0445 V____________________– 1.0355 V
+ 11.5085 V____________________+ 11.6915 V
– 11.6915 V____________________– 11.5085 V
+ 17.6075 V____________________+ 18.3925 V
– 18.3925 V____________________– 17.6075 V
+ 1.0355 V____________________+ 1.0445V
– 1.0445 V____________________– 1.0355 V
+ 11.5085 V____________________+ 11.6915 V
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
+ 1.6 mV
+ 25.1 mV
+ 251 mV
+ 1.6 mV
+ 25.1 mV
+ 251 mV
+ 1.6 mV
+ 25.1 mV
+ 251 mV
+ 1.6 mV
+ 25.1 mV
+ 251 mV
TDS 510A Service Manual
4–19
Performance Tests
Test Record (Cont.)
Instrument Serial Number:Certificate Number:
Temperature:RH %:
Date of Calibration:Technician:
Performance TestMaximumOutgoingIncomingMinimum
CH2200 mV Vert scale setting,
+5 Div position setting, –10 V offset
CH21 V Vert scale setting,
–5 Div position setting, +10 V offset
CH21 V Vert scale setting,
+5 Div position setting, –10 V offset
CH3 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
CH3 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
CH3200 mV Vert scale setting,
–5 Div position setting, +10 V offset
CH3200 mV Vert scale setting,
+5 Div position setting, –10 V offset
CH31 V Vert scale setting,
–5 Div position setting, +10 V offset
CH31 V Vert scale setting,
+5 Div position setting, –10 V offset
CH4 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
CH4 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
CH4200 mV Vert scale setting,
–5 Div position setting, +10 V offset
CH4200 mV Vert scale setting,
+5 Div position setting, –10 V offset
CH41 V Vert scale setting,
–5 Div position setting, +10 V offset
CH41 V Vert scale setting,
+5 Div position setting, –10 V offset
Analog Bandwidth
CH1 100 mV424 mV____________________N/A
CH2100 mV424 mV____________________N/A
CH3100 mV424 mV____________________N/A
CH4 100 mV424 mV____________________N/A
– 11.6915 V____________________– 11.5085 V
+ 17.6075 V____________________+ 18.3925 V
– 18.3925 V____________________– 17.6075 V
+ 1.0355 V____________________+ 1.0445 V
– 1.0445 V____________________– 1.0355 V
+ 11.5085 V____________________+ 11.6915 V
– 11.6915 V____________________– 11.5085 V
+ 17.6075 V____________________+ 18.3925 V
– 18.3925 V____________________– 17.6075 V
+ 1.0355 V____________________+ 1.0445 V
– 1.0445 V____________________– 1.0355 V
+ 11.5085 V____________________+ 11.6915 V
– 11.6915 V____________________– 11.5085 V
+ 17.6075 V____________________+ 18.3925 V
– 18.3925 V____________________– 17.6075 V
4–20
TDS 510A Service Manual
Performance Tests
Test Record (Cont.)
Instrument Serial Number:Certificate Number:
Temperature:RH %:
Date of Calibration:Technician:
Performance TestMaximumOutgoingIncomingMinimum
Delay Between Channels
Delay Between ChannelsN/A____________________250 ps
Time Base System
Long Term Sample Rate/
Delay Time @ 100 ns/10.0 ms
Trigger System Accuracy
Pulse-Glitch or Pulse-Width,
Hor. scale ≤ 1 ms
Lower Limit
Upper Limit
Pulse-Glitch or Pulse-Width,
Hor. scale > 1 ms
Lower Limit
Upper Limit
Main Trigger, DC Coupled, Positive Slope9.940 V____________________10.060 V
Main Trigger, DC Coupled, Negative Slope9.940 V____________________10.060 V
Delayed Trigger, DC Coupled, Positive Slope9.940 V____________________10.060 V
Delayed Trigger, DC Coupled, Negative Slope9.940 V____________________10.060 V
CH1 Sensitivity, 50 MHz, MainPass/Fail____________________Pass/Fail
CH1 Sensitivity , 50 MHz, DelayedPass/Fail____________________Pass/Fail
CH1 AUX Trigger InputPass/Fail____________________Pass/Fail
CH1 Sensitivity , full bandwidth, MainPass/Fail____________________Pass/Fail
CH1 Sensitivity , full bandwidth, DelayedPass/Fail____________________Pass/Fail
Output Signal Checks
MAIN TRIGGER OUTPUT, 1 MW
High
Low
MAIN TRIGGER OUTPUT, 50 W
High
Low
DELA YED TRIGGER OUTPUT, 50 W
High
Low
DELA YED TRIGGER OUTPUT, 1 MW
High
Low
–2.5 Div____________________+2.5 Div
2.5 ns
2.5 ns
1.0 ms
1.0 ms
High ≥ 2.5 V__________
High ≥ 1.0 V__________
High ≥ 1.0 V__________
High ≥ 2.5 V__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________Low ≤ 0.25 V
__________
__________Low ≤ 0.25 V
__________
__________Low ≤ 0.7 V
7.5 ns
7.5 ns
3.0 ms
3.0 ms
Low ≤ 0.7 V
TDS 510A Service Manual
4–21
Performance Tests
Test Record (Cont.)
Instrument Serial Number:Certificate Number:
Temperature:RH %:
Date of Calibration:Technician:
Performance TestMaximumOutgoingIncomingMinimum
CH 3 or AX 1 SIGNAL OUTPUT, 1 MWPk-Pk ≥ 90 mV____________________Pk-Pk ≤ 110 mV
CH 3 or AX 1 SIGNAL OUTPUT, 50 WPk-Pk ≥ 45 mV____________________Pk-Pk ≤ 55 mV
Probe Compensator Output Signal
Frequency (CH1 Freq.)950 Hz____________________1.050 kHz
Voltage (difference)495 mV____________________505 mV
4–22
TDS 510A Service Manual
Signal Acquisition System Checks
These procedures check those characteristics that relate to the signal-acquisition
system and are listed as checked under Warranted Characteristics in Specifica-
tions.
Performance Tests
Check Offset Accuracy
(Zero Setting)
Equipment
Required
PrerequisitesSee page 4–15.
1. Preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
HPress CLEAR MENU to remove the menus from the screen.
b. Modify the default settings:
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Hi Res.
HPress CURSOR.
None
TDS 510A Service Manual
HPress the main-menu button Function; then press the side-menu
button H Bars.
HPress CLEAR MENU.
HBe sure to disconnect any input signals from all four channels.
2. Confirm input channels are within limits for offset accuracy at zero offset:
Do the following substeps — test CH 1 first, skipping substep a. since CH 1
is already set up to be checked from step 1.
a. Select an unchecked channel: Press WAVEFORM OFF to remove the
channel just confirmed from the display. Then, press the front-panel
button that corresponds to the channel you are to confirm.
b. Set the vertical scale: Set the vertical SCALE to one of the settings
listed in Table 4–2 that is not yet checked. (Start with the first setting
listed.)
4–23
Performance Tests
HPress VERTICAL MENU. Press the main-menu button Fine Scale.
HUse the keypad to enter the vertical scale. For the 1 mV setting,
press 1, SHIFT, m, then ENTER. For the 101 mV setting, press
101, SHIFT, m, and then ENTER. For the 1.01 V setting, press
1.01 and then ENTER.
HPress CLEAR MENU.
T able 4–2: DC Offset Accuracy (Zero Setting)
Vertical
Vertical Scale
Setting
1 mV0±1.6 mV
101 mV0±25.1 mV
1.01 V0±251 mV
1
Vertical position is set to 0 divisions and vertical
offset to
step 1.
Position and
Offset Setting
0 V when the oscilloscope is initialized in
Offset Accuracy
1
Limits
c. Display the test signal: The waveform position and offset were
initialized for all channels in step 1 and are displayed as you select each
channel and its vertical scale.
d. Measure the test signal: Align the active cursor over the waveform by
rotating the general purpose knob. Ignore the other cursor. See
Figure 4–4.
e. Read the measurement results at the absolute (@:) cursor readout, not
the delta (D:) readout on screen. That is, read the offset relative to the
ground reference. See Figure 4–4.
f.Check against limits: Do the following subparts in the order listed.
HCHECK that the measurement results are within the limits listed for
the current vertical scale setting.
HEnter voltage on test record.
HRepeat substeps b through f until all vertical scale settings listed in
Table 4–2, are checked for the channel under test.
4–24
TDS 510A Service Manual
Read the measurement
2
results.
Align the active cursor
1
over the waveform.
Performance Tests
Check DC Voltage
Measurement Accuracy
Figure 4–4: Measurement of DC Offset Accuracy at Zero Setting
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
WARNING. The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure.
Equipment
Required
PrerequisitesThe oscilloscope must meet the prerequisites listed on page 4–15
Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision coaxial cables (Item 5)
TDS 510A Service Manual
4–25
Performance Tests
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
HSet the output of a DC calibration generator to 0 volts.
HConnect the output of a DC calibration generator through a
dual-banana connector followed by a 50 W precision coaxial cable to
one side of a BNC T connector. See Figure 4–5.
HConnect the Sense output of the generator through a second
dual-banana connector followed by a 50 W precision coaxial cable to
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 4–5.
Output Sense
DC Calibrator
Dual Banana to
BNC Adapters
Figure 4–5: Initial Test Hookup
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
c. Modify the default settings:
50 W Coaxial Cables
Digitizing Oscilloscope
BNC T
Connector
4–26
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Average 16.
2. Confirm input channels are within limits for DC accuracy at maximum offset
and position: Do the following substeps — test CH 1 first, skipping substep
2a since CH 1 is already selected from step 1.
TDS 510A Service Manual
Performance Tests
a. Select an unchecked channel:
HPress WAVEFORM OFF to remove the channel just confirmed
from the display.
HPress the front-panel button that corresponds to the channel you are
to confirm.
HSet the generator output to 0 V.
HMove the test hookup to the channel you selected.
b. Turn on the measurement Mean for the channel:
HPress MEASURE, then press the main-menu button Select
Measrmnt for CHx.
HPress the side-menu button more until the menu label Mean appears
in the side menu (its icon is shown at the left). Press the side-menu
button Mean.
HPress CLEAR MENU.
c. Set the vertical scale: Set the vertical SCALE to one of the settings
listed in Table 4–3 that is not yet checked. (Start with the first setting
listed.)
T able 4–3: DC Voltage Measurement Accuracy
Position
Scale
Setting
5 mV–5+1 V+1.040 V+1.0355 V to +1.0445 V
200 mV–5+10 V+11.6 V+11.5085 V to +1 1.6915 V
1 V–5+10 V+18V+17.6075 V to +18.3925 V
Setting
(Divs)
+5 V–1 V–1.040 V–1.0445 V to –1.0355 V
+5–10 V–1 1.6 V–11.6915 V to –1 1.5085 V
+5–10 V–18 V–18.3925 V to –17.6075 V
Offset
Setting
Generator
Setting
Accuracy Limits
d. Display the test signal:
TDS 510A Service Manual
HPress VERTICAL MENU. Press the main-menu button Position.
HUse the keypad to set vertical position to –5 divisions (press –5, then
ENTER, on the keypad). The baseline level will move off screen.
HPress the main-menu button Offset.
4–27
Performance Tests
measurement called
mean and read the
HUse the keypad to set vertical offset to the positive-polarity setting
listed in the table for the current vertical scale setting. The baseline
level will remain off screen.
HSet the generator to the level and polarity indicated in the table for
the vertical scale, position, and offset settings you have made. The
DC test level should appear on screen. (If it doesn’t return, the DC
accuracy check is failed for the current vertical scale setting of the
current channel.)
e. Measure the test signal: Press CLEAR MENU. Read the measurement
results at the Mean measurement readout. See Figure 4–6.
Turn on the
results here.
4–28
Figure 4–6: Measurement of DC Accuracy at Maximum Offset and Position
f.Check against limits:
HCHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and
position/offset/generator settings. Enter value on test record.
HRepeat substep d, reversing the polarity of the position, offset, and
generator settings as is listed in the table.
HCHECK that the Mean measurement readout on screen is within the
limits listed for the current vertical scale setting and position/offset/
generator settings. Enter the value on test record.
TDS 510A Service Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.