Tektronix TDS420A, TDS430A, TDS460A, TDS510A User Manual

User Manual
TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes
070-9701-03
www.tektronix.com
Copyright © T ektronix, Inc. All rights reserved. T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved. T ektronix, Inc., P.O. Box 500, Beaverton, OR 97077 TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.

WARRANTY

T ektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If a product proves defective during this warranty period, T ektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid. T ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-T ektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REP AIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY . TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

Getting Started
Operating Basics
Reference
General Safety Summary vii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preface ix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Product Description 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Start Up 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
At a Glance 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tutorial 2–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessing Help 3–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Acquisition Modes 3–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Delayed Triggering 3–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Determining Status 3–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display Modes 3–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Edge Triggering 3–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fast Fourier Transforms (Optional) 3–27. . . . . . . . . . . . . . . . . . . . . . . . . . .
File System 3–43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hardcopy 3–47. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Horizontal Control 3–55. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Limit Testing 3–63. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Logic Triggering (TDS 510A Only) 3–67. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring Waveforms 3–73. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Cal (TDS 510A Only) 3–83. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Compensation 3–89. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pulse Triggering (TDS 510A Only) 3–91. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Remote Communication 3–97. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Roll Mode (TDS 400A Only) 3–99. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Saving and Recalling Setups 3–103. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Saving and Recalling Waveforms 3–105. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Selecting Channels 3–111. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting Up Automatically: Autoset and Reset 3–113. . . . . . . . . . . . . . . . . . .
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
i
Table of Contents
Appendices
Glossary
Signal Path Compensation 3–115. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Taking Cursor Measurements 3–117. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Vertical Control 3–123. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Video Triggering (TDS 400A) 3–127. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Waveform Differentiation (Optional) 3–131. . . . . . . . . . . . . . . . . . . . . . . . . .
Waveform Integration (Optional) 3–135. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Waveform Math 3–139. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Zoom 3–143. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix A: Options and Accessories A–1. . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix B: Packaging for Shipment B–1. . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix C: Remote Display C–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix D: Programmer Disk D–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Index
ii
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

List of Figures

Table of Contents
Figure 1–1: Rear Panel Controls Used in Start Up 1–4. . . . . . . . . . . . . . .
Figure 1–2: ON/STBY Button 1–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–1: Connecting a Probe for the Examples 2–11. . . . . . . . . . . . . . . .
Figure 2–2: SETUP Button Location 2–12. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–3: The Displayed Setup Menu 2–12. . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–4: SET LEVEL TO 50% Button 2–13. . . . . . . . . . . . . . . . . . . . . .
Figure 2–5: The Display After Factory Initialization 2–14. . . . . . . . . . . . .
Figure 2–6: The VERTICAL and HORIZONTAL Controls 2–15. . . . . . .
Figure 2–7: TRIGGER Controls 2–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–8: AUTOSET Button Location 2–16. . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–9: The Display After Pressing Autoset 2–17. . . . . . . . . . . . . . . . .
Figure 2–10: Display Signals Requiring Probe Compensation 2–17. . . . . .
Figure 2–11: The Channel Buttons and Lights 2–18. . . . . . . . . . . . . . . . . . .
Figure 2–12: The Vertical Main Menu and Coupling Side Menu 2–20. . . .
Figure 2–13: The Menus After Changing Channels 2–21. . . . . . . . . . . . . .
Figure 2–14: Measure Main Menu and Select Measurement Side
Menu 2–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–15: Four Simultaneous Measurement Readouts 2–23. . . . . . . . .
Figure 2–16: General Purpose Knob Indicators 2–24. . . . . . . . . . . . . . . . .
Figure 2–17: Snapshot of Channel 1 2–25. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2–18: Save/Recall Setup Menu 2–27. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–1: Initial Help Screen 3–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–2: Acquisition Menu and Readout 3–8. . . . . . . . . . . . . . . . . . . . .
Figure 3–3: Acquire Menu Stop After 3–9. . . . . . . . . . . . . . . . . . . . . . .
Figure 3–4: Delayed Runs After Main 3–11. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–5: Delayed Triggerable 3–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–6: Delayed Trigger Menu (TDS 400A shown) 3–14. . . . . . . . . . . .
Figure 3–7: Status Menu System (TDS 400A shown) 3–15. . . . . . . . . . .
Figure 3–8: Banner Display 3–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–9: Display Menu Style 3–17. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–10: Trigger Point and Level Indicators 3–19. . . . . . . . . . . . . . . . .
Figure 3–11: Edge Trigger Readouts 3–21. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–12: Record View, Trigger Position, and Trigger Level
Bar Readouts 3–22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
iii
Table of Contents
Figure 3–13: Main Trigger Menu Edge Type 3–23. . . . . . . . . . . . . . . . .
Figure 3–14: TRIGGER Controls and Status Lights 3–26. . . . . . . . . . . . . .
Figure 3–15: System Response to an Impulse 3–28. . . . . . . . . . . . . . . . . . . .
Figure 3–16: Define FFT Waveform Menu 3–29. . . . . . . . . . . . . . . . . . . . . .
Figure 3–17: FFT Math Waveform in Math1 3–31. . . . . . . . . . . . . . . . . . . .
Figure 3–18: Cursor Measurement of an FFT Waveform 3–32. . . . . . . . . .
Figure 3–19: Waveform Record vs. FFT Time Domain Record 3–34. . . . .
Figure 3–20: FFT Time Domain Record vs. FFT Frequency
Domain Record 3–34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–21: How Aliased Frequencies Appear in an FFT 3–37. . . . . . . . .
Figure 3–22: Windowing the FFT Time Domain Record 3–40. . . . . . . . . .
Figure 3–23: FFT Windows and Bandpass Characteristics 3–41. . . . . . . .
Figure 3–24: File Utilities 3–43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–25: File System Labelling Menu 3–44. . . . . . . . . . . . . . . . . . . .
Figure 3–26: Utility Menu System I/O 3–48. . . . . . . . . . . . . . . . . . . . . . .
Figure 3–27: Date and Time Display 3–49. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–28: Connecting the Digitizing Oscilloscope Directly to
the Hardcopy Device 3–50. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–29: Connecting the Digitizing Oscilloscope and Hardcopy
Device Via a PC 3–52. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–30: Record View and Time Base Readouts 3–55. . . . . . . . . . . . . .
Figure 3–31: Horizontal Controls 3–56. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–32: Aliasing 3–58. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–33: Comparing a Waveform to a Limit Template 3–63. . . . . . . .
Figure 3–34: Acquire Menu Create Limit Test Template 3–64. . . . . . . .
Figure 3–35: Logic Trigger Readouts 3–68. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–36: Logic Trigger Menu 3–70. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–37: Logic Trigger Menu Time Qualified TRUE 3–72. . . . . . .
Figure 3–38: Measurement Readouts 3–75. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–39: Measure Menu 3–76. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–40: Measure Menu Gating 3–78. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–41: Measure Menu Reference Levels 3–79. . . . . . . . . . . . . . . .
Figure 3–42: Measure Delay Menu Delay To 3–80. . . . . . . . . . . . . . . . . .
Figure 3–43: Snapshot Menu and Readout 3–81. . . . . . . . . . . . . . . . . . . . . .
Figure 3–44: Probe Cal Menu and Gain Compensation Display 3–85. . . .
Figure 3–45: Re-use Probe Calibration Data Menu 3–87. . . . . . . . . . . . . . .
Figure 3–46: How Probe Compensation Affects Signals 3–89. . . . . . . . . . .
Figure 3–47: Probe Adjustment 3–90. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–48: Pulse Trigger Readouts 3–91. . . . . . . . . . . . . . . . . . . . . . . . . .
iv
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Table of Contents
Figure 3–49: Main Trigger Menu Glitch Class 3–92. . . . . . . . . . . . . . . .
Figure 3–50: Main Trigger MenuRunt Class 3–95. . . . . . . . . . . . . . . . . .
Figure 3–51: Connecting the Digitizing Oscilloscope to a Controller 3–97
Figure 3–52: Utility Menu 3–98. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–53: Roll Mode (500 Point Record Length) 3–100. . . . . . . . . . . . . .
Figure 3–54: Trigger Mode Menu 3–102. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–55: Save/Recall Setup Menu 3–104. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–56: Save Waveform Menu (TDS 400A shown) 3–106. . . . . . . . . . .
Figure 3–57: More Menu 3–107. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–58: Save Format Menu 3–108. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–59: The Channel Readout 3–111. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–60: Waveform Selection Priority 3–112. . . . . . . . . . . . . . . . . . . . . .
Figure 3–61: Performing a Signal Path Compensation 3–116. . . . . . . . . . . .
Figure 3–62: Cursor Types 3–117. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–63: Cursor Modes 3–118. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–64: H Bars Cursor Menu and Readouts 3–119. . . . . . . . . . . . . . . .
Figure 3–65: Paired Cursor Menu and Readouts 3–119. . . . . . . . . . . . . . . .
Figure 3–66: Video Line and IRE Units (TDS 400A shown) 3–121. . . . . . . .
Figure 3–67: Vertical Readouts and Channel Menu 3–124. . . . . . . . . . . . . .
Figure 3–68: Main Trigger Menu Video Type 3–127. . . . . . . . . . . . . . . . .
Figure 3–69: Video Trigger Menu — Class 3–128. . . . . . . . . . . . . . . . . . . . . .
Figure 3–70: Video Trigger Menu TV Delay Mode 3–129. . . . . . . . . . . . .
Figure 3–71: Video Trigger Scan Parameter 3–130. . . . . . . . . . . . . . . . . .
Figure 3–72: Video Trigger Scan Rate & Interlace 3–130. . . . . . . . . . . . .
Figure 3–73: Derivative Math Waveform 3–132. . . . . . . . . . . . . . . . . . . . . . .
Figure 3–74: Peak-Peak Amplitude Measurement of a Derivative
Waveform 3–133. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–75: Integral Math Waveform 3–136. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–76: H Bars Cursors Measure an Integral Math Waveform 3–137
Figure 3–77: More Menu 3–139. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3–78: Dual Waveform Math Menus 3–141. . . . . . . . . . . . . . . . . . . . .
Figure 3–79: Zoom Mode with Horizontal Lock Set to None 3–145. . . . . . .
Figure 3–80: Zoom Preview Mode 3–146. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure D–1: Equipment Needed to Run the Example Programs D–1. . . .
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
v
Table of Contents

List of Tables

Table 1–1: Fuse and Fuse Cap Part Numbers 1–4. . . . . . . . . . . . . . . . . . .
Table 3–1: TDS 460A, TDS 430A, and TDS 420A Resolution Bits 3–6. .
Table 3–2: TDS 510A Resolution Bits 3–6. . . . . . . . . . . . . . . . . . . . . . . . . .
Table 3–3: XY Format Pairs 3–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 3–4: Logic Triggers 3–68. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 3–5: Measurement Definitions 3–73. . . . . . . . . . . . . . . . . . . . . . . . . .
Table 3–6: Probe Cal Status 3–86. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 3–7: Pulse Trigger Definitions 3–91. . . . . . . . . . . . . . . . . . . . . . . . . . .
Table A–1: International Power Cords A–2. . . . . . . . . . . . . . . . . . . . . . . .
Table A–2: Standard Accessories A–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table A–3: Optional Accessories A–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table A–4: Compatible Probes A–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table A–5: Accessory Software A–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vi
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
Only qualified personnel should perform service procedures.

Injury Precautions

Use Proper Power Cord. To avoid fire hazard, use only the power cord specified
for this product. Avoid Electric Overload. To avoid electric shock or fire hazard, do not apply a
voltage to a terminal that is outside the range specified for that terminal. Avoid Overvoltage. To avoid electric shock or fire hazard, do not apply potential
to any terminal, including the common terminal, that varies from ground by more than the maximum rating for that terminal.
Avoid Electric Shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Do Not Operate Without Covers. To avoid electric shock or fire hazard, do not operate this product with covers or panels removed.
Use Proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.
Do Not Operate in Wet/Damp Conditions. To avoid electric shock, do not operate this product in wet or damp conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fire hazard, do not operate this product in an explosive atmosphere.
Product Damage
Precautions
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Use Proper Power Source. Do not operate this product from a power source that applies more than the voltage specified.
Provide Proper Ventilation. To prevent product overheating, provide proper ventilation.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
vii
General Safety Summary

Symbols and Terms

T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
T erms on the Product. These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the
marking. WARNING indicates an injury hazard not immediately accessible as you read the
marking. CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
Certifications and
Compliances
DANGER
High Voltage
Protective Ground
(Earth) T erminal
ATTENTION
Refer to Manual
Refer to the specifications chapter of the performance verification and specifica­tions manual for a listing of certifications and compliances that apply to this product.
viii
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Preface

Related Manuals

This is the User Manual for the TDS 420A, TDS 430A, TDS 460A, and TDS 510A Digitizing Oscilloscopes.
The Getting Started chapter briefly describes the digitizing oscilloscope, prepares you to install it, and tells you how to put it into service.
The Operating Basics chapter covers basic principles of the operation of the oscilloscope. These articles help you understand why your oscilloscope works the way it does.
The Reference chapter teaches you how to perform specific tasks. See page 3–1 for a complete list of tasks covered in that chapter.
The Appendices provide an option and accessories listing and other useful information.
The following documents are related to the use or service of the digitizing oscilloscope:
H The TDS Family Programmer Manual describes using a computer to control
the digitizing oscilloscope through the GPIB interface. H The TDS 420A, TDS 430A, TDS 460A & TDS 510A Reference gives you a
quick overview of how to operate your digitizing oscilloscope. H The TDS 420A, TDS 430A & TDS 460A Performance Verification and
TDS 510A Performance Verification manuals tell how to verify the
performance of the digitizing oscilloscope. H The TDS Family Option 13 Instruction Manual describes using the optional
Centronicsr and RS-232 interfaces for obtaining hardcopy (only for TDS
oscilloscopes equipped with that option).
H The TDS 420A, TDS 430A & TDS 460A Service Manual and the TDS 510A
Service Manual provide information for maintaining and servicing your
digitizing oscilloscope to the module level.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
ix
Preface

Conventions

In the Getting Started and Reference chapters, you will find various procedures which contain steps of instructions for you to perform. To keep those instructions clear and consistent, this manual uses the following conventions:
H Names of front panel controls and menu labels appear in boldface print. H Names also appear in the same case (initial capitals, all uppercase, etc.) in
the manual as is used on the oscilloscope front panel and menus. Front panel names are all upper case letters, for example, VERTICAL MENU, CH 1, and SETUP.
H Instruction steps are numbered. The number is omitted if there is only
one step.
H When steps require that you make a sequence of selections using front panel
controls and menu buttons, an arrow ( front panel button and a menu, or between menus. Also, whether a name is a main menu or side menu item is clearly indicated: Press VERTICAL
MENU 100 MHz
Coupling (main) DC (side) Bandwidth (main)
(side).
) marks each transition between a
Using the convention just described results in instructions that are graphical­ly intuitive and simplifies procedures. For example, the instruction just given replaces these five steps:
1. Press the front panel button VERTICAL MENU.
2. Press the main menu button Coupling.
3. Press the side-menu button DC.
4. Press the main menu button Bandwidth,
5. Press the side menu button 100 MHz.
H Sometimes you may have to make a selection from a pop-up menu: Press
TRIGGER MENU
repeatedly press the main menu button Type until Edge is highlighted in the pop-up menu.
Type (main) Edge (pop-up). In this example, you
x
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Product Description

Your Tektronix digitizing oscilloscope is a superb tool for acquiring, displaying, and measuring waveforms. Its performance addresses the needs of both lab and portable applications with the following features:
H 200 MHz maximum analog bandwidth on the TDS 420A
400 MHz maximum analog bandwidth on the TDS 430A, TDS 460A
500 MHz maximum analog bandwidth on the TDS 510A H 500 Megasamples/second maximum digitizing rate on the TDS 510A
100 Megasamples/second maximum digitizing rate on the TDS 420A,
TDS 430A, and TDS 460A H Roll mode and triggered roll mode for display of slower waveforms on the
TDS 420A, TDS 430A, and TDS 460A H Waveform Math Invert a single waveform and add, subtract, and multiply
two waveforms. On instruments equipped with option 2F, integrate or
differentiate a single waveform or perform an FFT (fast fourier transform) on
a waveform to display its magnitude or phase versus its frequency. H Up to 30,000-point record length per channel (120,000-point optional) on the
TDS 420A, TDS 430A, and TDS 460A. Up to 50,000-point record length
per channel on the TDS 510A H Full GPIB software programmability. GPIB hardcopy output. On instru-
ments equipped with option 13, hardcopy output using the RS-232 or
Centronics ports.
H Complete measurement and documentation ability H Intuitive graphical icon operation blended with the familiarity of traditional
horizontal and vertical knobs
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
1–1
Product Description
H Four channels and four eight-bit digitizers on the TDS 420A, TDS 460A,
and TDS 510A. Two channels and two eight-bit digitizers on the TDS 430A H On-line help at the touch of a button Appendix A lists the options and accessories.
The product specification is in the performance verification manual that is shipped as a standard accessory with the digitizing oscilloscope.
1–2
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Start Up

Operation

Before you use the digitizing oscilloscope, ensure that it is properly installed and powered on.
To properly install and power on the digitizing oscilloscope, do the following steps:

Installation

1. Be sure you have the appropriate operating environment. Specifications for
temperature, relative humidity, altitude, vibrations, and emissions are included in performance verification and specification manuals (Tektronix part numbers 070-9705-xx and 070-9706-xx).
2. Leave space for cooling. Do this by verifying that the air intake and exhaust
holes on the sides of the cabinet (where the fan operates) are free of any airflow obstructions. Leave at least 2 inches (5.1 cm) free on each side.
WARNING. To avoid electrical shock, be sure that the power cord is disconnected before checking the fuse.
3. Check the fuse to be sure it is the proper type and rating (see Figure 1-1 for
the fuse location). You can use either of two fuses (see Table 1–1 for the fuse data).
4. Check that you have the proper electrical connections:
H For TDS 400A serial number below B080000: 90 to 132 V for 48 Hz
through 62 Hz, 100 to 132 V or 180 to 250 V for 48 through 440 Hz, and may require up to 240 W.
H For TDS 400A serial number B080000 - Up: 100 to 240 V ±10%,
50/60 Hz nominal, or 115 V ±10% for 400 Hz, and may require up to 240 W.
H For TDS 510A all serial numbers: 90 to 250 V for 45 Hz to 440 Hz, and
may require up to 300 W.
5. Connect the proper power cord from the rear-panel power connector (see
Figure 1-1 for the connector location) to the power system.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
1–3
Start Up
TDS 400A (Below B080000)
Power
Connector
Fuse Principal
Power Switch
TDS 400A ( B080000 - Up)
Figure 1-1: Rear Panel Controls Used in Start Up
Fuse
Power
Connector
TDS 510A
Power
Connector
FusePrincipal
Power Switch
T able 1–1: Fuse and Fuse Cap Part Numbers
Oscilloscope Fuse Fuse
Part Number
TDS 420A, TDS 430A, and TDS 460A
TDS 510A
Below B080000: 5 A FAST, 250 V, 3AG B080000 - Up: 8 A, 250V , 3AG
Below B080000: 4 A (T), 250 V. B080000 - Up:
6.3 A FAST, 250 V .25 inch y 1.25 inch (UL
198.6, 3AG): 6 A FAST, 250 V
5 mm y 20 mm (IEC 127): 5 A (T), 250 V
159-0014-00 159-0046-00
159-0255-00 159-0381-00
159-0013-00 200-2264-00
159-0210-00 200-2265-00
Fuse Cap Part Number
200-2264-00 200-2264-00
200-2265-00 200-2265-00
1–4
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Start Up

Front Cover Removal

Power On

ON/STBY Button
Remove the front cover by grasping its left and right edges and snapping it off of the front subpanel. (When reinstalling, align and snap back on.)
1. Check that the rear-panel principal power switch is on (see Figure 1-1 for the
location of the switch ). The principal power switch controls all AC power to the instrument.
NOTE. TDS400A instruments with serial number B080000 or above do not have a principal power switch.
2. If the oscilloscope is not powered on (the screen is blank), push the front-panel ON/STBY button to toggle it on (Figure 1-2).
Once the digitizing oscilloscope is installed, you can leave the principal power switch on (TDS400A instruments below B080000 and all TDS510A instruments) and use the ON/STBY button .

Self Test

Power Off

Figure 1-2: ON/STBY Button
The digitizing oscilloscope automatically performs power-on tests each time it is turned on. It comes up with a display screen that states whether or not it passed self test. If the self test does not detect any problems, the status display screen disappears a few seconds after the self test is complete.
Check the self test results. If the self test fails, call your local Tektronix Service Center. Depending on the
type of failure, you may still be able to use the oscilloscope before it is serviced.
Press the ON/STBY switch to turn off the oscilloscope.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
1–5
Start Up

Before You Begin

Signal Path Compensation (SPC) lets you compensate your oscilloscope for the current ambient temperature. SPC helps ensure maximum possible accuracy for your most critical measurements. See Signal Path Compensation in Section 3 for information on this feature.
1–6
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Overview

This chapter describes the basic concepts of operating the digitizing oscilloscope. Understanding the basic concepts of your digitizing oscilloscope helps you use it much more effectively.
At a Glance quickly shows you how the oscilloscope is organized and gives some very general operating instructions. It also contains an overview of the following maps:
H Front Panel Map H Rear Panel Map H Display Map H Basic Menu Operation
The Tutorial contains tutorial examples and explains basic system concepts: H Setting Up for the Examples explains how to set up the digitizing oscillo-
scope to use the examples.
H Example 1: Displaying a Waveform teaches you how to reset the digitizing
oscilloscope, display and adjust waveforms, and use the autoset function.
H Example 2: Displaying Multiple Waveforms explains how to add, control,
and delete multiple waveforms.
H Example 3: Taking Automated Measurements introduces you to the
automated measurement system.
H Example 4: Saving Setups discusses saving and recalling the digitizing
oscilloscope setups.
H Triggering explains how to set the triggers to convert unstable displays or
blank screens into meaningful waveforms.
H Scaling and Positioning Waveforms explains how to change the position and
displayed size of waveforms.
H Measurements explains using automated, cursor, and graticule measurements
to display numeric information on the displayed waveforms.
To explore these topics in more depth and to read about topics not covered in this chapter, see Reference. Page 3–1 lists the topics covered.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–1
Overview
2–2
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

At a Glance

The At a Glance section contains illustrations of the display and the front and rear panels. These illustrations help you understand and operate the digitizing oscilloscope. This section also contains a visual guide to using the menu system.

Front Panel Map Left Side

File System,
page 3–43
Side Menu Buttons, page 2–8
ON/STBY Switch,
page 1-4
Main Menu Buttons,
page 2–8
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
CLEAR MENU Removes Menus from the Display
2–3
At a Glance

Front Panel Map Right Side (TDS 400A)

Saving and Recalling Setups, page 3–103
Reset the Oscilloscope, page 2–12
Help, page 3–3
Status, page 3–15
Selecting Channels,
page 3–111
Waveform Math,
page 3–139
Removing Waveforms,
page 3–111
Ground
Autoset,
page 3–113
Saving and Recalling Waveforms, page 3–105
File System,
page 3–43
Measurement
System,
page 3–73
Hardcopy,
page 3–47
Acquisition Modes, page 3–5
Cursor Measurements, page 3–117
Shift, when lit, selects alternate menus (printed in blue) and coarse knob speed.
Display Modes, page 3–17 Remote Communication, page 3–97
Probe Compensation,
page 3–89
2–4
Vertical Control,
page 3–123
Zoom,
page 3–143
Horizontal Control,
page 3–55
Triggering, page 3–25 Delay Triggering, page 3–11 Edge Triggering, page 3–21
Video Triggering, page 3–127
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Front Panel Map Right Side (TDS 510A)

At a Glance
Saving and Recalling
Waveforms, page 3–105
File System, page 3–43
Autoset, page 3–113
Help, page 3–3
Status, page 3–15
Saving and Recalling
Setups, page 3–103
Reset the Oscilloscope,
page 2–12
Selecting Channels,
page 3–111
Measurement System,
Cursor Measurements,
page 3–117
page 3–73
Hardcopy, page 3–47
File System, page 3–43
Display Modes, page 3–17
Remote Communication, page 3–97
Acquisition Modes,
page 3–5
Cursor Measurements,
page 3–117
Waveform Math,
page 3–139
Vertical Control,
page 3–123
Removing Waveforms,
page 3–111
Zoom,
page 3–143
Ground
Horizontal Control,
page 3–55
Triggering, page 3–25 Delay Triggering, page 3–11 Edge Triggering, page 3–21 Logic Triggering, page 3–67 Pulse Triggering, page 3–91
Probe Compensation,
page 3–89
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–5
At a Glance

Rear Panel Map

GPIB
Connector
TDS 400A (Below B080000)
Connector, page C–1
Centronics Connector
(Optional)
VIDEO VGA
Compatible
RS-232 Connector (Optional)
Power
Fuse,
page 1–3
AUX TRIGGER/EXT CLOCK
(Provides Auxiliary Trigger
and External Clock Input)
Connector,
page 1–3
Principal
Power Switch,
page 1–5
Serial
Number
Fuse,
page 1–3
TDS 400A (B080000 - Up)
Power
Connector,
page 1–3
Principal
Power Switch,
page 1–3
Fuse,
page 1–3
TDS 510A
Centronics Connector
(Optional)
Serial
Number
RS-232 Connector
(Optional)
Power Connector,
page 1–3
GPIB Connector
Security Bracket
VGA Output (Monochrome)
Rear Panel Connectors SIGNAL OUTPUT –
(Provides CH3 analog signal output)
AUX TRIGGER INPUT – (Provides auxiliary trigger signal input)
MAIN TRIGGER OUTPUT – (Provides main trigger (TTL) output)
DELAYED TRIGGER OUTPUT – (Provides delayed trigger (TTL) output)
2–6
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Display Map

At a Glance
Acquisition Status,
page 3–7
Shows what part of the waveform record
Trigger Position (T),
page 3–25
Indicates position of
vertical bar cursors in
the waveform record
is displayed, page 3–55.
The value entered with the
general purpose knob.
Waveform Record Icon
When the general purpose
knob is activated, the knob
icon appears here.
When present, the general purpose knob makes coarse adjustments;
when absent, fine adjustments.
Trigger level on
waveform (may be an
arrow at right side of
screen instead of a bar)
Channel Level and
Waveform Source
Vertical Scale of Each Channel, page 3–123
The main menu with choices
of major actions
Cursor Measurements, page 3–117
The side menu with choices of specific actions
Trigger Parameters, page 3–21
Horizontal Scale and Time
Base Type, page 3–55
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–7
At a Glance

To Operate a Menu (TDS 400A)

1
Press front-panel menu button.
2
Press one of these buttons to
select from main menu.
3
Press one of these buttons to select from
side menu (if displayed).
4
If side menu item has an adjustable value (shown in reverse
video), adjust it with the general purpose knob.
2–8
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

To Operate a Menu (TDS 510A)

Press front-panel menu button. (Press
1
SHIFT first if button label is blue.)
At a Glance
2
Press one of these buttons to select
from main menu.
3
Press one of these buttons to select
from side menu (if displayed).
4
If side menu item has an adjustable
value (shown in reverse video),
adjust it with the general purpose
knob or keypad.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–9
At a Glance

To Operate a Pop-Up Menu

Press
to display pop-ups.
Press again
to make selection.
Alternatively, press SHIFT
first to make selection in the
opposite direction.
A pop-up selection changes the other
main menu titles.
Press to remove
menus from screen.
2–10
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Tutorial

This section quickly acquaints you with some of the fundamental operations required to use your digitizing oscilloscope to take measurements. Start this tutorial by doing Setting Up for the Examples.

Setting Up for the Examples

Perform the following tasks to connect input signals to the digitizing oscillo­scope, to reset it, and to become acquainted with its display screen. Once completed, these tasks ready the digitizing oscilloscope for use in the examples that follow.

Connect the Input Signal

Remove all probes and signal inputs from the input BNC connectors along the lower right of the front panel. Then, using one of the probes supplied with the digitizing oscilloscope, connect from the CH 1 connector of the digitizing oscilloscope to the Probe Compensation connector (Figure 2-1).
TDS 400A
TDS 510A
Figure 2-1: Connecting a Probe for the Examples
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–11
Tutorial

Reset the Oscilloscope

SETUP Button
Do the following steps to reset the digitizing oscilloscope to a known factory default state. (Reset the oscilloscope anytime you begin a new task and need to start fresh with known default settings.)
1. Press the save/recall SETUP button to display the Setup menu (see
Figure 2-2).
Figure 2-2: SETUP Button Location
The digitizing oscilloscope displays main menus along the bottom of the screen. Figure 2-3 shows the Setup main menu.
OK Confirm Factory Init Menu Item and Button
Recall Factory Setup
Menu Item and Button
Figure 2-3: The Displayed Setup Menu
2. Press the button directly below the Recall Factory Setup menu item.
The display shows side menus along the right side of the screen. The buttons to select these side menu items are to the right of the side menu.
Because an accidental instrument reset could destroy a setup that took a long time to create, the digitizing oscilloscope asks you to verify the Recall Factory Setup selection (see Figure 2-3).
3. Press the button to the right of the OK Confirm Factory Init side menu
item.
2–12
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Tutorial
NOTE. This manual uses the following notation to represent the sequence of selections you made in steps 1, 2 and 3: Press save/recall SETUP
Factory Setup
(main) OK Confirm Factory Init (side).
Recall
Note that a clock icon appears on screen. The oscilloscope displays this icon when performing operations that take longer than several seconds.
4. Press SET LEVEL TO 50% (see Figure 2-4) to be sure the oscilloscope
triggers on the input signal.
Examine the Display
Elements
SET LEVEL TO 50% Button
Figure 2-4: SET LEVEL TO 50% Button
Read the following information to become familiar with the oscilloscope display. Figure 2-5 shows the display that results from the oscilloscope reset. There are several important points to observe:
H The trigger level bar shows that the waveform is triggered at a level near
50% of its amplitude (from step 4).
H The trigger position indicator shows that the trigger position of the
waveform is located at the horizontal center of the graticule.
H The channel reference indicator shows the vertical position of channel 1
with no input signal. This indicator points to the ground level for the channel when its vertical offset is set to 0 V in the vertical menu; when vertical offset is not set to 0 V, the indicator points to the vertical offset level.
H The trigger readout shows that the digitizing oscilloscope is triggering on
channel 1 (Ch1) on a rising edge and that the trigger level is about 200-300 mV.
H The time base readout shows that the main time base is set to a horizontal
scale of 500 ms/div.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–13
Tutorial
H The channel readout indicates that channel 1 (Ch1) is displayed with DC
coupling. (In AC coupling, ~ appears after the volts/div readout.) The digitizing oscilloscope always displays channel 1 at reset.
Right now, the channel, time base, and trigger readouts appear in the graticule area because a menu is displayed. You can press the CLEAR MENU button at any time to remove any menus and to move the readouts below the graticule.
Trigger Position
Indicator
Channel Ground
Reference Indicator
2–14
Channel Readout
Time Base
Readout
Trigger
Readout
Figure 2-5: The Display After Factory Initialization
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Example 1: Displaying a Waveform

The digitizing oscilloscope provides front panel knobs for you to adjust a waveform, or it can automatically set up its controls to display a waveform. Do the following tasks to learn how to adjust a waveform and how to autoset the digitizing oscilloscope.
Tutorial
Adjusting the Waveform
Display
The display shows the probe compensation signal. It is a 1 kHz square wave of approximately 0.5 V amplitude. Do the following steps to adjust the size and placement of the waveform using the front-panel knobs.
Figure 2-6 shows the main VERTICAL and HORIZONTAL sections of the front panel. Each has SCALE and POSITION knobs.
1. Turn the vertical SCALE knob clockwise. Observe the change in the
displayed waveform and the channel readout at the bottom of the display.
Figure 2-6: The VERTICAL and HORIZONTAL Controls
2. Turn the vertical POSITION knob first one direction, then the other.
Observe the change in the displayed waveform. Then return the waveform to the center of the graticule.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–15
Tutorial
3. Turn the horizontal SCALE knob one click clockwise. Observe the time
base readout at the bottom of the display. The time base should be set to 200 ms/div now, and you should see two complete waveform cycles on the display.

Autoset the Oscilloscope

When you first connect a signal to a channel and display it, the signal displayed may not be scaled and triggered correctly. Use the autoset function and you should quickly get a meaningful display.
You should have a stable display of the probe compensation waveform from the last step. Do the following steps to first create an unstable display and then to automatically obtain a stable display:
1. To create an unstable display, slowly turn the trigger MAIN LEVEL knob
(see Figure 2-7) first one direction, then the other. Observe what happens when you move the trigger level above the highest part of the displayed waveform. Leave the trigger level in that untriggered state.
2. Press AUTOSET (see Figure 2-8) and observe the stable waveform display.
MAIN LEVEL Knob
2–16
Figure 2-7: TRIGGER Controls
AUTOSET Button
Figure 2-8: AUTOSET Button Location
Figure 2-9 shows the display after pressing AUTOSET. If necessary, you can adjust the waveform using the knobs discussed earlier in this example.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Tutorial
Figure 2-9: The Display After Pressing Autoset
NOTE. If the corners on your displayed signal look rounded or pointed (see Figure 2-10), then you may need to compensate your probe. See pages 3–83 and 3–89 for probe calibration and compensation procedures.
Figure 2-10: Display Signals Requiring Probe Compensation
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–17
Tutorial

Example 2: Displaying Multiple Waveforms

In this example you learn how to display and control more than one waveform at a time.

Adding a Waveform

The VERTICAL section of the front panel contains the channel selection buttons. On the TDS 420A, TDS 460A, and TDS 510A Digitizing Oscillo­scopes, they are CH 1, CH 2, CH 3, CH 4, and MORE (Figure 2-11). On the TDS 430A, they are CH 1, CH 2, and MORE.
Figure 2-11: The Channel Buttons and Lights
2–18
Each of the channel (CH) buttons has a light above or beside its label. Do the following steps to add a waveform to the display:
1. If you are not continuing from the previous example, follow the instructions
on page 2–11 under the heading Setting Up for the Examples.
2. Press SETUP Init
(side).
3. Press AUTOSET.
4. Press CH 2.
The display shows a second waveform, which represents the signal on channel 2.
Recall Factory Setup (main) OK Confirm Factory
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
There are several other things to observe: H The channel readout on the display now shows the settings for both Ch1
and Ch2.
H There are two channel indicators at the left edge of the graticule. Right
now, they overlap.
H The light by the CH 2 button is now on, and the vertical controls are
now set to adjust channel 2.
H The trigger source is not changed by adding a channel. (You can change
the trigger source by using the TRIGGER MENU.)
5. Turn the vertical POSITION knob clockwise to move the channel 2
waveform up on the graticule. Notice that the channel reference indicator for channel 2 moves with the waveform.
Tutorial
6. Press VERTICAL MENU
Coupling (main).
The vertical menu gives you control over many vertical channel parameters (Figure 2-12). Although there can be more than one channel displayed, the vertical menu and buttons only adjust the selected channel.
Each menu item in the Vertical menu displays a side menu. Right now, the Coupling item in the main menu is highlighted, which means that the side menu shows the coupling choices.
7. Press W
(side) to toggle the selection to 50 W; this changes the input
coupling of channel 2 from 1 MW to 50 W. The channel readout for channel 2 (near the bottom of the graticule) now shows an W indicator.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–19
Tutorial
Ch2 Reference Indicator
Side Menu Title
Assign Controls to
Another Channel
Figure 2-12: The Vertical Main Menu and Coupling Side Menu
Pressing a channel (CH) button sets the vertical controls to that channel. It also adds the channel to the display if that waveform is not already displayed. To explore assigning controls to different channels, do the following steps:
1. Press CH 1. Observe that the side menu title shows Ch1 (see Figure 2-13) and that the
indicator next to CH 1 is on. Note the highlighted menu item in the side menu also changes from the 50 W channel 2 setting to the 1 MW impedance setting of channel 1.
2. Press CH 2
W (side) to toggle the selection to 1MW. This returns the
coupling impedance of channel 2 to its initial state.
2–20
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Side Menu Title
Tutorial

Remove a Waveform

Figure 2-13: The Menus After Changing Channels
Pressing the WAVEFORM OFF button removes the waveform for the currently selected channel. If the waveform you want to remove is not already selected, select that channel using the channel (CH) button. To remove a waveform from the display, do the following steps:
1. Press WAVEFORM OFF (under the vertical SCALE knob).
Since the CH 2 light was on when you pressed the WAVEFORM OFF button, the channel 2 waveform was removed.
The channel (CH) lights now indicate channel 1. Channel 1 has become the selected channel. When you remove the last waveform, all the CH lights are turned off.
2. Press WAVEFORM OFF again to remove the channel 1 waveform.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–21
Tutorial

Example 3: Taking Automated Measurements

The digitizing oscilloscope can measure many waveform parameters automati­cally and read out the results on screen. Do the following tasks to discover how to set up the oscilloscope to measure waveforms automatically.
Display Measurements
Automatically
To take automated measurements, do the following steps:
1. If you are not continuing from the previous example, follow the instructions on page 2–11 under the heading Setting Up for the Examples.
2. Press SETUP Init
(side).
3. Press AUTOSET.
4. Press MEASURE to display the Measure main menu (see Figure 2-14).
➞ Recall Factory Setup (main) OK Confirm Factory
2–22
Figure 2-14: Measure Main Menu and Select Measurement Side Menu
5. If it is not already selected, press Select Measrmnt that menu item indicates which channel the measurement will be taken from. All automated measurements are made on the selected channel.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
(main). The readout for
Tutorial
The Select Measurement side menu lists the measurements that can be taken. Up to four can be taken and displayed at any one time. Pressing the button next to the –more– menu item displays the other measurement selections.
6. Press Frequency
more
(side) repeatedly until the Frequency item appears, then press
Frequency
(side). If the Frequency menu item is not visible, press
(side).
Observe that the frequency measurement appears within the right side of the graticule area. The measurement readout includes the notation Ch1, meaning that the measurement is taken on the channel 1 waveform. (To take a measurement on another channel, select that channel, and then select the measurement.)
7. Press Positive Width Positive Duty Cycle
(side) more(side) Rise Time (side)
(side).
All four measurements are displayed.
8. To move the measurement readouts outside the graticule area, press CLEAR MENU (see Figure 2-15).
Figure 2-15: Four Simultaneous Measurement Readouts
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–23
Tutorial
Remove Measurement
Readouts
Examine the Measurement
Reference Levels
General Purpose Knob
Setting and Readout
General Purpose
Knob Icon
To remove measurements you no longer want displayed: Press MEASURE
ment 2, and Measurement 4
Remove Measrmnt (main) Measurement 1, Measure-
(side) to remove those measurements. Leave the
rise time measurement displayed.
To examine the current values:
Press Reference Levels
(main) High Ref (side).
Highlighted Menu Item with Boxed
Readout Value
Change the Measurement
Reference Levels
Figure 2-16: General Purpose Knob Indicators
By default, the measurement system uses the 10% and 90% levels of the waveform for taking the rise time measurement. You can change these values to other percentages or change them to absolute voltage levels.
To examine the current values, press Reference Levels
(main) High Ref
(side). The general purpose knob is now set to adjust the high reference level (Fig-
ure 2-16).
2–24
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Tutorial
There are several important things to observe on the screen: H The knob icon appears at the top of the screen. The knob icon indicates that
the general purpose knob is set to adjust a parameter.
H The upper right corner of the screen shows the readout High Ref: 90%. H The High Ref side menu item is highlighted, and a box appears around the
90% readout in the High Ref menu item. The box indicates that the general purpose knob is currently set to adjust that parameter.
To adjust the high level to 80%, turn the general purpose knob.
Display a Snapshot of
Automated Measurements
You can pop up a display of almost all of the automated measurements. To display a snapshot of automated measurements of the selected channel, do the following steps:
1. Press Snapshot waveform measurements. (See Figure 2-17).
(main) to pop up a snapshot of all available single
Figure 2-17: Snapshot of Channel 1
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–25
Tutorial

Example 4: Saving Setups

The digitizing oscilloscope can save its control settings and recall them later to quickly re-establish the previously saved state. Do the following tasks to learn how to save, and then recall, a setup.

Save a Setup

First, you need to create an instrument setup you want to save. Perform the following steps to create and save a setup that is complex enough that you might prefer not to go through all these steps each time you want that display:
1. If you are not continuing from the previous example, follow the instructions
on page 2–11 under the heading Setting Up for the Examples.
2. Press SETUP Init
(side).
3. Press
4. Press MEASURE
5. Press CH 2
To save the setup, do the following steps:
6. Press SETUP
AUTOSET.
the –more– side menu item if the Frequency selection does not appear in the side menu.)
menu (see Figure 2-18).
Recall Factory Setup (main) OK Confirm Factory
Select Measrmnt (main) Frequency (side). (Press
CLEAR MENU.
Save Current Setup (main) to display the Setup main
2–26
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Tutorial

Recall a Setup

Figure 2-18: Save/Recall Setup Menu
7. Press one of the To Setup side menu buttons to store the current instrument
settings into that setup location. Remember which setup location you selected for use later.
Once you have saved a particular setup, you can change the settings as you wish, knowing that you can come back to that setup at any time.
8. Press MEASURE
Positive Width (side) to add that measurement to the
display.
To recall the setup, press SETUP
(side) for the setup location you used in the last exercise.
Setup
Recall Saved Setup (main) Recall
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
2–27
Tutorial
2–28
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Overview

This chapter describes the details of operating the digitizing oscilloscope. It contains an alphabetical list of tasks you can perform with the digitizing oscilloscope. Use this chapter to answer specific questions about instrument operation. The following tasks are included:
H Accessing Help H Remote Communication H Acquisition Modes H Roll Mode H Delayed Triggering H Saving and Recalling Setups H Determining Status H Saving and Recalling Waveforms H Display Modes H Selecting Channels H Edge Triggering H Setting up Automatically H Fast Fourier Transforms H Signal Path Compensation H File System H Taking Cursor Measurements H Hardcopy H Vertical Control H Horizontal Control H Video Triggering H Limit Testing H Waveform Differentiation H Measuring Waveforms H Waveform Integration H Probe Cal H Waveform Math H Probe Compensation H Zoom H Pulse Triggering
Many of these tasks list steps you perform to accomplish the task. You should read Conventions on page x of Preface before reading about these tasks.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–1
Overview
3–2
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Accessing Help

The on-line help system provides brief information about each of the digitizing oscilloscope controls.
To use the on-line help system: Press HELP to display information on any front panel button, knob, or menu
item (see Figure 3-1). Press HELP again to return to the regular operating mode. Whenever the
oscilloscope is in help mode, pressing any button (except HELP or SHIFT), turning any knob, or pressing any menu item displays help text on the screen that discusses that control.
On-line help is available for each menu selection displayed at the time the HELP button is first pressed. If you are in help mode and want to see help on selections from non-displayed menus, you first exit help mode, display the menu you want information on, and press HELP again to re-enter help mode.
Figure 3-1: Initial Help Screen
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–3
Accessing Help
3–4
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Acquisition Modes

The acquisition system has several options for converting analog data into digital form. The Acquisition menu lets you determine the acquisition mode, whether or not to permit equivalent time sampling, and how to start and stop acquisitions.

Description of Modes

The digitizing oscilloscope supports five acquisition modes:
H Sample H Peak Detect H Hi Res H Envelope H Average
The Sample, Peak Detect, and Hi Res modes operate in real-time on a single trigger event, provided the digitizing oscilloscope can acquire enough samples for each trigger event. Envelope and Average modes operate on multiple acquisitions. The digitizing oscilloscope averages or envelopes several wave­forms on a point-by-point basis.

Sample Mode

Peak Detect Mode

Hi Res Mode

TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
In Sample mode, the oscilloscope creates a record point by saving the first sample (of perhaps many) during each acquisition interval. (An acquisition interval is the time covered by the waveform record divided by the record length.) This is the default mode.
Peak Detect mode alternates between saving the highest sample in one acquisi­tion interval and lowest sample in the next acquisition interval. This mode only works with real-time, non-interpolated sampling.
If you set the time base so fast that it requires real-time interpolation or equivalent-time sampling, the mode automatically changes from Peak Detect to Sample, although the menu selection does not change.
In Hi Res mode, the digitizing oscilloscope averages all samples taken during an acquisition interval to create a record point. That average results in a higher-reso­lution, lower-bandwidth waveform.
3–5
Acquisition Modes
NOTE. In Hi Res mode the frequency of the external clock signal (TDS 460A, TDS 430A, and TDS 420A only) must be less than or equal to the frequency set in the external clock menu. If the frequency of the external clock signal is greater than the frequency in the menu, the displayed waveform will have the wrong amplitude.
This mode only works with real-time, non-interpolated sampling. If you set the time base so fast that it requires real-time interpolation or equivalent-time sampling, the mode automatically becomes Sample, although the menu selection does not change.
A key advantage of Hi Res is its potential for increasing resolution regardless of the input signal. Tables 3–1 and 3–2 illustrate how you can obtain up to 15 significant bits with Hi Res mode. Note that resolutions above 15 bits are not allowed. The bits of resolution shown in the tables are theoretically achievable. Actual resolution may vary as a function of the correlated noise sources in the test environment.
T able 3–1: TDS 460A, TDS 430A, and TDS 420A Resolution Bits
Time Base Speed Bits of Resolution
1 ms and faster 8 bits 2 ms to 5 ms 9 bits 10 ms to 20 ms 10 bits 50 ms to 100 ms 11 bits 200 ms to 500 ms 12 bits
T able 3–2: TDS 510A Resolution Bits
Time Base Speed Bits of Resolution
400 ns and faster 8 bits 1 ms to 2 ms 9 bits 5 ms to 10 ms 10 bits 20 ms to 50 ms 11 bits 50 ms to 100 ms 11 bits 100 ms to 200 ms 12 bits 500 ms 13 bits 1 ms to 2 ms 14 bits 5 ms and slower 15 bits
3–6
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Acquisition Modes

Envelope Mode

Average Mode

Envelope mode lets you acquire and display a waveform record that shows the extremes in variation over several acquisitions. You specify the number of acquisitions over which to accumulate the data. The oscilloscope saves the highest and lowest values in two adjacent intervals similar to the Peak Detect mode. But Envelope mode, unlike Peak Detect, gathers peaks over many trigger events.
The final display shows the most extreme values for all the acquisitions for each point in the waveform record.
NOTE. Envelope and Average acquisition modes disable Roll mode. See Roll Mode beginning on page 399.
Average mode lets you acquire and display a waveform record that is the averaged result of several acquisitions. This mode reduces random noise. The oscilloscope acquires data after each trigger event using Sample mode.

Checking the Acquisition Readout

To determine the acquisition sampling rate, the acquisition state (running or stopped), and the acquisition mode, check the acquisition readout at the top of the display (see Figure 3-2). The “running” state shows the sample rate (or External Clock when external clock is enabled) and acquisition mode. The stopped state shows the number of acquisitions acquired since the last stop or major change.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–7
Acquisition Modes
Acquisition Readout
Figure 3-2: Acquisition Menu and Readout

Selecting an Acquisition Mode

The oscilloscope provides several acquisition modes. To bring up the acquisition menu (Figure 3-2) and choose how the digitizing oscilloscope creates points in the waveform record:
Press SHIFT ACQUIRE MENU Res, Envelope, or Average
NOTE. With some longer record lengths, the digitizing oscilloscope will not allow selecting Hi Res mode or will reduce the record length setting.
When you select Envelope or Average, you can enter the number of waveform records to be enveloped or averaged using the general purpose knob.

Selecting Repetitive Sampling

To limit the digitizing oscilloscope to real-time sampling or let it choose between real-time or equivalent-time sampling:
Mode (main) Sample, Peak Detect, Hi
(side).
3–8
Press SHIFT ACQUIRE MENU (side).
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Repetitive Signal (main) ON or OFF

To Stop After

Acquisition Modes
ON (Enable ET) uses both the real time and the equivalent time features of the digitizing oscilloscope.
OFF (Real Time Only) limits the digitizing oscilloscope to real time sampling. If the digitizing oscilloscope cannot accurately get enough samples for a complete waveform, the oscilloscope uses the interpolation method selected in the display menu to fill in the missing record points.
To choose the event that signals the oscilloscope to stop acquiring waveforms, do the following step:
Press SHIFT ACQUIRE MENU only, Single Acquisition Sequence, or Limit Test Condition Met Figure 3-3).
Stop After (main) RUN/STOP button
(side) (see
Figure 3-3: Acquire Menu Stop After
RUN/STOP button only
RUN/STOP button. Pressing the RUN/STOP button once stops the acquisitions. The upper left hand corner in the display indicates Stopped and shows the number of acquisitions. If you press the button again, the digitizing oscilloscope resumes taking acquisitions.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
(side) lets you start or stop acquisitions by toggling the
3–9
Acquisition Modes
Single Acquisition Sequence (side) lets you run a single sequence of acquisitions by pressing the RUN/STOP button.
In Envelope or Average mode, the digitizing oscilloscope makes the specified number of acquisitions to complete the averaging or enveloping task.
If the oscilloscope is in equivalent-time mode and you press Single Acquisition Sequence
(side), it continues to recognize trigger events and acquire samples
until the waveform record is filled.
NOTE. To quickly select Single Acquisition Sequence without displaying the Acquire and Stop After menus, press SHIFT FORCE TRIG. (You still must display the Acquire menu and then the Stop After menu to leave Single Acquisi­tion Sequence operation.)
Limit Test Condition Met (side) lets you acquire waveforms until waveform data
exceeds the limits specified in the limit test. Then acquisition stops. At that point, you can also specify other actions for the oscilloscope to take using the selections available in the Limit Test Setup main menu.
NOTE. In order for the digitizing oscilloscope to stop acquisition when limit test conditions are met, limit testing must be turned ON, using the Limit Test Setup main menu.
Setting up limit testing requires several more steps. See Limit Testing, on page 3–63.
3–10
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Delayed Triggering

The digitizing oscilloscope provides a main time base and a delayed time base. The delayed time base, like the main time base, requires a trigger signal and an input source dedicated to that signal. You can only use delay with the edge trigger and certain classes of main pulse triggers. This section describes how to delay the acquisition of waveforms.
There are two different ways to delay the acquisition of waveforms: delayed runs after main and delayed triggerable. Only delayed triggerable uses the delayed trigger system.
Delayed runs after main looks for a main trigger, then waits a user-defined time, and then starts acquiring (see Figure 3-4).
Wait for
Main
Trigger
Wait User-specified
Time
Acquire
Data
Figure 3-4: Delayed Runs After Main
Delayed triggerable looks for a main trigger and then, depending on the type of delayed trigger selected, makes one of the types of delayed triggerable mode acquisitions listed below (see Figure 3-5).
Wait for
Main
Trigger
Wait User-specified Time,
Number of Delayed
Trigger Events or Number
of External Clocks
Wait for Delay Trigger Event
Acquire
Data
Figure 3-5: Delayed Triggerable
After Time waits the user-specified time, then waits for the next delayed trigger event, and then acquires.
After Events waits for the specified number of delayed trigger events and then acquires.
After Events/Time (TDS 510A only) waits for the specified number of delayed trigger events, then waits the user-specified time, and then acquires.
External clks (TDS 400A) waits for the specified number of external clocks and then acquires.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–11
Delayed Triggering

To Run After Delay

NOTE. When using the delayed triggerable mode, the digitizing oscilloscope provides a conventional edge trigger for the delayed time base. The delayed time base will not trigger if the main trigger type (as defined in the Main Trigger menu) is logic, or if the main trigger type is pulse with the runt trigger class selected.
You use the Horizontal menu to select and define either delayed runs after main or delayed triggerable. Delayed triggerable, however, requires further selections in the Delayed Trigger menu. Do the following steps to set the delayed time base to run immediately after delay:

To Trigger After Delay

1. Press HORIZONTAL MENU (side)
Delayed Runs After Main (side).
2. Use the general purpose knob to set the delay time.
If you press Intensified timebase record that shows where the delayed timebase record occurs relative to the main trigger. For Delayed Runs After Main mode, the start of the intensified zone corresponds to the start of the delayed timebase record. The end of the zone corresponds to the end of the delayed record.
To learn how to set the intensity level, see Display Modes on page 3–17.
The Main Trigger menu settings must be compatible with Delayed Triggerable. To select Delayed Triggerable mode, do the following steps:
1. Press TRIGGER MENU.
2. Press Type
3. Press HORIZONTAL MENU
(side)
(main) and either Edge or Pulse as fits your application.
Delayed Triggerable (side).
(side), you display an intensified zone on the main
Time Base (main) Delayed Only
Time Base (main) Delayed Only
3–12
NOTE. The Delayed Triggerable menu item is not selectable unless incompatible Main Trigger menu settings are eliminated. If such is the case, the Delayed Triggerable menu item is dimmer than other items in the menu.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Delayed Triggering
By pressing Intensified (side), you can display an intensified zone that shows where the delayed timebase record may occur (a valid delay trigger event must be received) relative to the main trigger on the main timebase. For Delayed Triggerable After mode, the start of the intensified zone corresponds to the possible start point of the delayed timebase record. The end of the zone continues to the end of main timebase, since a delayed time base record may be triggered at any point after the delay time elapses.
To learn how to define the intensity level of the normal and intensified waveform, see Display Modes on page 3–17.
Now you need to bring up the Delayed Trigger menu so you can define the delayed trigger event.
4. On a TDS 400A, press SHIFT DELAYED TRIG Triggerable After Time, Events, or Ext clks
5. On the TDS 510A, press SHIFT DELAYED TRIG Triggerable After Time, Events, or Events/Time
Delay by (main)
(side) (Figure 3-6).
Delay by (main)
(side).
6. Enter the delay time or events using the general purpose knob or the keypad. Hint: You can go directly to the Delayed Trigger menu (see steps 4 and 5).
By selecting either Triggerable After Time, Events, or Events/Time, the oscilloscope automatically switches to Delayed Triggerable in the Horizontal menu. If you wish to leave Delayed Triggerable, you still need to display the Horizontal menu.
The Source menu lets you select which input is the delayed trigger source.
7. Press Source
(main) Ch1, Ch2, Ch3, Ch4, DC Aux, or Auxiliary (side).
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–13
Delayed Triggering
Figure 3-6: Delayed Trigger Menu (TDS 400A shown)
8. To define how the input signal is coupled to the delayed trigger, press Coupling
(main) DC, AC, HF Rej, LF Rej, or Noise Rej (side). For
descriptions of these coupling types, see To Specify Coupling on page 3–23.
9. To select the slope that the delayed trigger occurs on, press Slope (main). Choose between the rising edge and falling edge slopes.
When using Delayed Triggerable mode to acquire waveforms, two trigger bars are displayed. One trigger bar indicates the level set by the main trigger system; the other indicates the level set by the delayed trigger system.
10. Press Level
(main) Level, Set to TTL, Set to ECL, or Set to 50% (side).
For a description of these level settings, see To Set Level on page 3–24.
3–14
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Determining Status

The Status menu lets you see information about the oscilloscope state.

To Display the Status

To display the Status menu (see Figure 3-7):
Firmware Date
Press SHIFT STATUS I/O
(side).
System displays information about the firmware, Horizontal, Zoom, Acquisition, Limit Test, Measure, and Hardcopy (TDS 510A) systems.
Display displays information about the display system. Trigger displays information about the triggers. Waveforms displays information about the various waveforms, including live,
math, and reference. I/O displays information about the I/O port(s).
Status (main) System, Trigger, Waveforms, or
Figure 3-7: Status Menu System (TDS 400A shown)
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–15
Determining Status

To Display the Banner

To display the banner (firmware version, options, and copyright): Press SHIFT STATUS
Banner (main). (See Figure 3-8.)
Figure 3-8: Banner Display
3–16
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Display Modes

This section describes how to adjust the oscilloscope display style, intensity level, graticule, and format.

Change Display Settings

To bring up the Display menu, press DISPLAY.

Select the Display Style

Press DISPLAY Infinite Persistence, or Variable Persistence
Style (main) Vectors, Intensified Samples, Dots,
(side) (Figure 3-9).
Figure 3-9: Display Menu Style
Vectors style displays draw vectors (lines) between the record points. Dots style displays waveform record points as dots. Intensified Samples style also displays waveform record points as dots. However,
the points actually sampled are displayed intensified relative to the interpolated points.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–17
Display Modes
In addition to choosing Intensified Samples in the side menu, the oscilloscope must be interpolating (equivalent time must be off) or Zoom must be on with its horizontal expansion greater that 1X. See Zoom beginning on page 3–143.
Variable Persistence accumulates the record points on screen over many acquisi-
tions and displays them only for a specific time interval. In that mode, the display behaves like that of an analog oscilloscope.
Infinite Persistence accumulates the record points until you change some control (such as scale factor) causing the display to be erased.

Adjust Intensity

Set Display Readout
Options
Intensity lets you set overall, text/graticule, and waveform intensity (brightness) levels. To set the intensity:
Press DISPLAY
Contrast
(side). Enter the intensity percentage values with the general purpose
Intensity (main) Overall, Text/Grat, Waveform, or
knob or keypad.
NOTE. The Intensified setting for Timebase in the horizontal menu causes a zone on the waveform to be intensified relative to the rest of the waveform. If the contrast is set to 100%, you can not distinguish the intensified portion from the rest of the waveform because both are the same brightness.
To control whether the trigger indicator, trigger level bar, and current date and time appear on the display and what style trigger level bar, long or short, is displayed:
1. Press DISPLAY
Readout (main).
2. To select whether or not to display a T indicating the trigger point, Toggle Display T @ Trigger Point
(side). You can select ON or OFF. (The
trigger point indicates the position of the trigger in the waveform record.)
3–18
3. To select either the short or the long trigger bar or to turn the trigger bar off, toggle Trigger Bar Style
(side). (See Figure 3-10. Note that both styles are
shown for illustrating purposes, but you can only display one style at a time.) The trigger bar is only displayed if the trigger source is an active, displayed
waveform. Also, two trigger bars are displayed when delay triggerable acquisitions are displayed . The trigger bar is a visual indicator of the trigger level.
4. Press Display Date/Time
(side) to turn it on or off. Push Clear Menu to see
the current date and time. See To Date/Time Stamp Your Hardcopy on page 3–49 for more information.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Trigger Point Indicator
Trigger BarLong Style
-or-
Trigger BarShort Style
Display Modes

Select Interpolation Filter

Select the Graticule Type

Figure 3-10: Trigger Point and Level Indicators
To select the display filter type (sin(x)/x interpolation and linear interpolation): Press DISPLAY
polation
(side).
Filter (main) Sin(x)/x Interpolation or Linear Inter-
To select the graticule type: Press DISPLAY
or PAL
(side).
Graticule (main) Full, Grid, Cross Hair, Frame, NTSC
Full provides a grid, cross hairs, and a frame. Grid displays a frame and a grid. Cross Hair provides cross hairs and a frame. Frame displays just a frame. NTSC provides a grid useful for measuring NTSC-class waveforms. PAL provides a grid useful for measuring PAL-class waveforms.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–19
Display Modes
NOTE. Selecting either NTSC or PAL graticule automatically changes the vertical scale, position settings, coupling, and sets to zero any vertical offset of any channel displayed. These settings are not restored after switching to other graticule types.

Select the Display Format

There are two display formats: YT and XY. YT is the conventional oscilloscope display format. It shows a signal voltage (the
vertical axis) as it varies over time (the horizontal axis). XY format compares the voltage levels of two waveform records point by point.
That is, the digitizing oscilloscope displays a graph of the voltage of one waveform record against the voltage of another waveform record. This mode is particularly useful for studying phase relationships.
To set the display axis format: Press DISPLAY
Format (main) XY or YT (side).
When you choose the XY mode, the input you have selected is assigned to the X-axis and the digitizing oscilloscope automatically chooses the Y-axis input (see Table 3–3).
T able 3–3: XY Format Pairs
X-Axis Channel (User Selectable) Y -Axis Channel (Fixed)
Ch 1 Ch 2 Ch 3 (TDS 420A, TDS 460A, and TDS 510A
only)
Ch 4 (TDS 420A, TDS 460A, and TDS 510A only)
3–20
If you press the WAVEFORM OFF button for an XY channel, the digitizing oscilloscope deletes the XY graph of of the channels.
Since selecting YT or XY affects only the display, the horizontal and vertical scale and position knobs and menus control the same parameters regardless of the mode selected. Specifically, in XY mode, the horizontal scale continues to control the time base and the horizontal position continues to control which portion of the waveforms are displayed.
XY format is a dot-only display, although it can have persistence. The Vector style selection has no effect when you select XY format.
You cannot display Math waveforms in XY format. They disappear from the display when you select XY.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Edge Triggering

An edge trigger event occurs when the trigger source passes through a specified voltage level (the trigger level) in a specified direction (the trigger slope). You will likely use edge triggering for most of your measurements. This section describes how to use edge triggering. It also explains how to select the trigger mode, auto or normal, for all trigger types.

To Check Edge Trigger Status

To quickly check if edge triggers are selected, check the Trigger readout. When edge triggers are selected, the trigger readout displays the trigger source, as well as the trigger slope and level (see Figure 3-11).
Main Time Base Time/Div
Main Time Base
Figure 3-11: Edge Trigger Readouts
To determine the trigger status, check the TRIGD, READY, and ARM status lights in the Trigger control area (see Figure 3–26):
H When TRIG’D is lighted, the digitizing oscilloscope has recognized a valid
trigger and is filling the posttrigger portion of the waveform.
H When READY is lighted, the digitizing oscilloscope can accept a valid
trigger event and it is waiting for that event to occur.
Main Trigger
Source = Ch 1
Main Trigger Slope = Rising Edge
Main Trigger Level
H When ARM is lighted, the trigger circuitry is filling the pretrigger portion of
the waveform record.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–21
Edge Triggering
Trigger Position Relative to the
Display and Waveform Record.
H When both TRIG’D and READY are lighted, the digitizing oscilloscope has
recognized a valid main trigger and is waiting for a delayed trigger. When it recognizes a delayed trigger it will fill in the posttrigger portion of the delayed waveform.
H When ARM, TRIG’D, and READY are all off, the digitizer is stopped. To determine where the trigger point is located in the waveform record and with
respect to the display, check the record view at the top of the display (see Figure 3-12).
To see the trigger point and level on the waveform display, check the graphic indicators of trigger position and level, which you can optionally display. Figure 3-12 shows the trigger position indicator and the short-style trigger level bar. See Set Display Readout Options on page 3–18 for more information.
Trigger Point Indicator
Indicating the Trigger Position
on the Waveform Record.
Trigger Bar Indicating the Trigger
Level on the Waveform Record.
Figure 3-12: Record View, Trigger Position, and Trigger Level Bar Readouts

To Select Edge Triggering

Use the Edge Trigger menu to select edge triggering and to perform the procedures for selecting the source, coupling, slope, trigger level, mode, and holdoff that follow.
3–22
To bring up the Edge Trigger menu, press TRIGGER MENU Edge
(pop-up) (see Figure 3-13).
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Type (main)
Edge Triggering

To Select a Source

To select which source you want for the trigger: Press TRIGGER MENU
Ch1, Ch2, Ch3, Ch4, AC Line, or Auxiliary
Type (main) Edge (pop-up) Source (main)
(side).

To Specify Coupling

Figure 3-13: Main Trigger Menu Edge Type
To select the coupling you want, press TRIGGER MENU Edge
(pop-up) Coupling (main) DC, AC, HF Rej, LF Rej, or Noise Rej
Type (main)
(side).
NOTE. If you select the Auxiliary trigger source, select DC or HF Rej coupling.
DC passes all of the input signal. In other words, it passes both AC and DC
components to the trigger circuit. AC passes only the alternating components of an input signal (above 30 Hz). It
removes the DC component from the trigger signal. HF Rej removes the high frequency portion of the triggering signal. That allows
only the low frequency components to pass on to the triggering system. High frequency rejection attenuates signals above 30 kHz.
LF Rej does the opposite of high frequency rejection. Low frequency rejection attenuates signals below 80 kHz.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–23
Edge Triggering
Noise Rej provides lower sensitivity. It requires additional signal amplitude for stable triggering, reducing the chance of falsely triggering on noise.

To Set Slope

To Set Level

To select the slope that the edge trigger occurs on:
1. Press the TRIGGER MENU
(main).
Slope
➞ Type (main) Edge (pop-up)
2. Select the rising or falling edge from the side menu.
Press the TRIGGER MENU el
(main) Level, Set to TTL, Set to ECL, or Set to 50% (side).
Type (main) Edge (pop-up) Lev-
Level lets you enter the trigger level using the general purpose knob or the keypad. (You may also change the level using the MAIN LEVEL knob.)
Set to TTL fixes the trigger level at +1.4 V. Set to ECL fixes the trigger level at 1.3 V.
NOTE. When you set the volts/div smaller than 200 mV, the oscilloscope reduces the Set to TTL or Set to ECL trigger levels below standard TTL and ECL levels. This happens because the trigger level range is fixed at ±12 divisions from the center. At 100 mV (the next smaller setting after 200 mV) the trigger range is ±1.2 V, which is smaller than the typical TTL (+1.4 V) or ECL (–1.3 V) level.

To Set Mode & Holdoff

3–24
Set to 50% fixes the trigger level to approximately 50% of the peak-to-peak value of the trigger source signal.
To select the trigger mode and change the holdoff time:
1. Press the TRIGGER MENU Normal
(side). The trigger modes operate as follows:
Mode & Holdoff (main) Auto or
H In Auto mode the oscilloscope acquires a waveform after a specific time
has elapsed even if a trigger does not occur. The amount of time the oscilloscope waits depends on the time base setting.
H In Normal mode the oscilloscope acquires a waveform only if there is a
valid trigger.
2. To change the holdoff time, press Holdoff
(side). Enter the value in % using
the general purpose knob or keypad.
If you want to enter a large number using the general purpose knob, press the SHIFT button before turning the knob. When the light next to the SHIFT button
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Edge Triggering
is on and the display says Coarse Knobs in the upper right corner, the general purpose knob speeds up significantly.
Holdoff is automatically reset to 0% when you change the main time base time/division setting. However, it is not reset if you change the delayed time base time/division (that is, Intensified or Delayed Only).

To Force a Trigger

To Single Trigger

Triggering From the Front
Panel
To force the oscilloscope to immediately start acquiring a waveform record even without a trigger event, press the FORCE TRIG front panel button.
To trigger on the next valid trigger event and then stop, press SHIFT FORCE TRIG. Now press the RUN/STOP button each time you want to initiate a single sequence of acquisitions.
The trigger buttons and knob let you quickly adjust the trigger level or force a trigger (see Figure 3-14). You use the following trigger controls for all trigger types, except were noted.
T o Set the MAIN LEVEL. To manually change the trigger level, turn the MAIN LEVEL knob. It adjusts the trigger level (or threshold level) no matter what
menu, if any, is displayed.
T o Set to 50%. To quickly obtain an edge trigger, press SET LEVEL TO 50%. The oscilloscope sets the trigger level to the halfway point between the peaks of the trigger signal. Set Level to 50% has no effect when the trigger type is video.
Note that the MAIN LEVEL knob and menu items apply only to the main trigger level. To modify the delayed trigger level, use the Level item in the Delayed Trigger menu.
T o Force a Trigger. To force the oscilloscope to immediately start acquiring a waveform record even without a trigger event, press the FORCE TRIG front panel button.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–25
Edge Triggering
Trigger Status Lights
Figure 3-14: TRIGGER Controls and Status Lights
T o Single Trigger. To trigger on the next valid trigger event and then stop, press
SHIFT FORCE TRIG. Now press the RUN/STOP button each time you want to initiate a single sequence of acquisitions.
3–26
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Fast Fourier Transforms (Optional)

Advanced DSP Math, provides the Fast Fourier Transform (FFT). (For more information see Waveform Differentiation, on page 3–131, Waveform Integra- tion, on page 3–135, and Waveform Math, on page 3–139.) The FFT allows you to transform a waveform from a display of its amplitude against time to one that plots the amplitudes of the various discrete frequencies the waveform contains. You can also display the phase shifts of those frequencies. Use FFT math waveforms in the following applications:
H Testing impulse response of filters and systems H Measuring harmonic content and distortion in systems H Characterizing the frequency content of DC power supplies H Analyzing vibration H Analyzing harmonics in 50 and 60 cycle lines H Identifying noise sources in digital logic circuits

Description

The FFT computes and displays the frequency content of a waveform you acquire as an FFT math waveform. The resulting waveform is a display of the magnitude or phase angle of the various frequencies the waveform contains with respect to those frequencies. For example, Figure 3-15 shows the non-trans­formed impulse response of a system in channel 2 at the top of the screen. The FFT-transformed magnitude and phase appear in the two math waveforms below the impulse.
The horizontal scale for FFT math waveforms is always expressed in frequency per division with the beginning (left-most point) of the waveform representing zero frequency (DC).
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–27
Fast Fourier Transforms
Normal Waveform of an
Impulse Response
FFT Waveform of the
Magnitude Response
FFT Waveform of the
Phase Response

Operation

Figure 3-15: System Response to an Impulse
To obtain an FFT of your waveform:
1. Connect the waveform to the desired channel input and select that channel.
2. Adjust the vertical and horizontal scales and trigger the display (or press AUTOSET).
The topic Offset, Position, and Scale, on page 3–34, provides information about optimizing your setup for FFT displays.
3. Press MORE to access the menu for turning on math waveforms.
4. Select a math waveform. Your choices are Math1, Math2, and Math3
(main).
5. If the selected math waveform is not FFT, press Change Math Definition
FFT (main). See Figure 3-16.
(side)
3–28
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Fast Fourier Transforms
Figure 3-16: Define FFT Waveform Menu
6. Press Set FFT Source to
(side) repeatedly until the channel source selected
in step 1 appears in the menu label.
7. Press Set FFT Vert Scale to
(side) repeatedly to choose from the following
vertical scale types: H dBV RMS Magnitude is displayed using log scale, expressed in dB
relative to 1 V
where 0 dB =1 V
RMS
RMS
.
H Linear RMS Magnitude is displayed using voltage as the scale. H Phase (deg) Phase is displayed using degrees as the scale, where
degrees wrap from 180_ to +180_.
H Phase (rad) Phase is displayed using radians as the scale, where
radians wrap from –p to +p.
The topic Considerations for Phase Displays, on page 3–37, provides in depth information on setup for phase displays.
8. Press Set FFT Window to
(side) repeatedly to choose from the following
window types: H Rectangular Best type for resolving frequencies that are very close to
the same value but worst for accurately measuring the amplitude of those
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–29
Fast Fourier Transforms
frequencies. Best type for measuring the frequency spectrum of non-repetitive signals and measuring frequency components near DC.
H Hamming Very good for resolving frequencies that are very close to
the same value with somewhat improved amplitude accuracy over the rectangular window.
H Hanning — Very good for measuring amplitude accuracy but degraded
for resolving frequencies.
H Blackman-Harris Best window for measuring the amplitude of
frequencies but worst at resolving frequencies.
See Selecting a Window, on page 3–39, provides in depth information on choosing the right window for your application.
9. If you did not select Phase (deg) or Phase (rad) in step 7, skip to step 12. Phase suppression is only used to reduce noise in phase FFTs.
10. If you need to reduce the effect of noise in your phase FFT, press Suppress phase at amplitudes <
(side).
11. Use the general purpose knob or the keypad to adjust the phase suppression level. FFT magnitudes below this level will have their phase set to zero.
See Adjust Phase Suppression, on page 3–38, provides additional informa­tion on phase suppression.
12. Press OK Create Math Wfm
(side) to display the FFT of the waveform you
input in step 1 (see Figure 3-17).
3–30
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Fast Fourier Transforms
Cursor Measurements of
an FFT
Figure 3-17: FFT Math Waveform in Math1
Once you have displayed an FFT math waveform, use cursors to measure its frequency amplitude or phase angle.
1. Be sure MORE is selected in the channel selection buttons and that the FFT math waveform is selected in the More main menu.
2. Press CURSOR tion
(main) H Bars (side).
Mode (main) Independent (side) Func-
3. Use the general purpose knob to align the selected cursor (solid line) to the top (or to any amplitude on the waveform you choose).
4. Press SELECT to select the other cursor. Use the general purpose knob to align the selected cursor to the bottom (or to any amplitude on the waveform you choose).
5. Read the amplitude between the two cursors from the D: readout. Read the amplitude of the selected cursor relative to either 1 V volts), or the zero phase level
(0 degrees or 0 radians) from the @: readout.
(0 dB), ground (0
RMS
(The waveform reference indicator at the left side of the graticule indicates the level where phase is zero for phase FFTs.)
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–31
Fast Fourier Transforms
Figure 3-18 shows the cursor measurement of a frequency magnitude on an FFT. The @: readout reads 0 dB because it is aligned with the 1 V
RMS
level.
The D: readout reads 24.4 dB indicating the magnitude of the frequency it is measuring is –24.4 dB relative to 1 V
. The source waveform is turned
RMS
off in the display. The cursor units will be in dB or volts for FFTs measuring magnitude and in
degrees or radians for those FFTs measuring phase. The cursor unit depends on the selection made for Set FFT Vert Scale to
(side). See step 7 on
page 3–29 for more information.
6. Press V Bars
(side). Use the general purpose knob to align one of the two
vertical cursors to a point of interest along the horizontal axis of the waveform.
7. Press SELECT to select the alternate cursor.
8. Align the selected cursor to another point of interest on the math waveform.
9. Read the frequency difference between the cursors from the D: readout. Read
the frequency of the selected cursor relative to the zero frequency point from the @: readout.
The cursor units will always be in Hz, regardless of the setting in the Time Units side menu. The first point of the FFT record is the zero frequency point for the @: readout.
3–32
Figure 3-18: Cursor Measurement of an FFT Waveform
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Fast Fourier Transforms
10. Press Function (main) Paired (side).
11. Place the vertical bar of each paired cursor to the points along the horizontal
axis you are interested in.
12. Read the amplitude between the X of the two paired cursors from the top-most D: readout. Read the amplitude of the short horizontal bar of the selected (solid) cursor relative to either 1 V zero phase level frequency between the long horizontal bars of both paired cursors from the bottom D: readout.
(0 degrees or 0 radians) from the @: readout. Read the
(0 dB), ground (0 volts), or
RMS
Automated Measurements
of an FFT
You can also use automated measurements to measure FFT math waveforms. Use the same procedure as is found under Waveform Differentiation on page 3–132.

Considerations for Using FFTs

There are several characteristics of FFTs that affect how they are displayed and should be interpreted. Read the following topics to learn how to optimize the oscilloscope setup for good display of your FFT waveforms.
The FFT Frequency
Domain Record
The following topics discuss the relation of the source waveform to the record length, frequency resolution, and frequency range of the FFT frequency domain record. (The FFT frequency domain waveform is the FFT math waveform that you display.)
FFTs May Not Use All of the Waveform Record. The FFT math waveform is a display of the magnitude or phase data from the FFT frequency domain record. This frequency domain record is derived from the FFT time domain record, which is derived from the waveform record. All three records are described below.
Waveform Record the complete waveform record acquired from an input channel. The waveform record is not a DSP Math waveform.
FFT Time Domain Record that part of the waveform record that is input to the FFT. This time domain record waveform becomes the FFT math waveform. Its record length depends on the length of the waveform record defined above.
FFT Frequency Domain Record the FFT math waveform after digital signal processing converts the FFT time domain record into a frequency domain record.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–33
Fast Fourier Transforms
Figure 3-19 compares the waveform record to the FFT time domain record.
FFT Time Domain Record =
Waveform Record
Waveform Record 10 K
Zero Phase
Reference
FFT Time Domain Record = 10k
Waveform Record > 10 K
Zero Phase
Reference

Offset, Position, and Scale

Figure 3-19: Waveform Record vs. FFT Time Domain Record
FFTs Transform Time Records to Frequency Records. The FFT time domain record is input for the FFT. The resulting record is one half the length of the FFT input (see Figure 3-20) because the FFT computes both positive and negative frequencies. Since the negative values mirror the positive values, only the positive values are displayed.
FFT Time Domain Record
FFT
FFT Frequency Domain Record
Figure 3-20: FFT Time Domain Record vs. FFT Frequency Domain Record
The following topics contain information to help you display your FFT properly.
3–34
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Fast Fourier Transforms
Adjust for a Non-Clipped Display . To properly display your FFT waveform, scale the source waveform so it is not clipped:
H Scale and position the source waveform so it is contained on screen. (Off
screen waveforms may be clipped, resulting in errors in the FFT waveform. If you use off screen waveforms, turn on Pk-Pk in the measurement menu and monitor the source waveform for clipping.)
H Use vertical position and vertical offset to position your source waveform.
As long as the source waveform is not clipped, its vertical position and vertical offset will not affect your FFT waveform except at DC. (DC correction is discussed below.)
Adjust Offset and Position to Zero for DC Correction. Normally, the output of a standard FFT computation yields a DC value that is twice as large as it should be with respect to the other frequencies. Also, the selection of window type introduces errors in the DC value of an FFT.
The displayed output of the FFT on TDS oscilloscopes is corrected for these errors to show the true value for the DC component of the input signal. The Position and Offset must be set to zero for the source waveform in the Vertical menu. When measuring the amplitude at DC, remember that 1 VDC equals 1 V
and the display is in dB.
RMS

Record Length

Acquisition Mode

Most often, you will want to use a short record length because more of the FFT waveform can be seen on screen and long record lengths can slow oscilloscope response. However, long record lengths lower the noise relative to the signal and increase the frequency resolution for the FFT. More important, they might be needed to capture the waveform feature you want to include in the FFT.
To speed up oscilloscope response when using long record lengths, you can save your source waveform in a reference memory and perform an FFT on the saved waveform. That way the DSP will compute the FFT based on saved, static data and will only update if you save a new waveform.
Selecting the right acquisition mode can produce less noisy FFTs.
Set up in Sample or Normal Mode. Use sample mode until you have set up and turned on your FFT. Sample mode can acquire repetitive and nonrepetitive waveforms and does not affect the frequency response of the source waveform.
Hi Res and Average Reduce Noise. After the FFT is set up and displayed, it might be useful to turn on Hi Res mode, to reduce the effect of noise in the signal. However, Hi Res does affect the frequency response of the source waveform.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–35
Fast Fourier Transforms
If the pulse is repetitive, Average mode may be used to reduce noise in the signal at a cost of slower display response. However, averaging does affect the frequency response of the source waveform.
Do not use Peak Detect and Envelope mode. They can add significant distortion to the FFT results.

Zoom and Interpolation

Undersampling (Aliasing)

Once you have your waveform displayed optimally, you may magnify (or reduce) it vertically and horizontally to inspect any feature you desire. Just be sure the FFT waveform is the selected waveform. Depending upon the record length, FFT math waveforms may be zoomed or not.
Sin(x)/x interpolation may distort the magnitude and phase displays of the FFT depending on which window was used. You can easily check the effects of the interpolation by switching between sin(x)/x and linear interpolation and observing the difference in measurement results on the display. If significant differences occur, use linear interpolation.
Aliasing occurs when the oscilloscope acquires a source waveform with frequency components outside of the frequency range for the current sample rate. In the FFT waveform, the actual higher frequency components are under­sampled, and therefore, they appear as lower frequency aliases that fold back around the Nyquist point (see Figure 3-21).
The greatest frequency that can be input into any sampler without aliasing is ½ the sample frequency. Since source waveforms often have a fundamental frequency that does not alias but have harmonic frequencies that do, use the following methods to recognize and deal with aliases:
H A source waveform with fast edge transition times creates many high
frequency harmonics.
3–36
H Sample the source signal at rates that are at least 2X that of the highest
frequency component having significant amplitude.
H Filter the input to bandwidth limit it to frequencies below that of the Nyquist
frequency. H Recognize and ignore the aliased frequencies. If you think you have aliased frequencies in your FFT, select the source channel
and adjust the horizontal scale to increase the sample rate. Since you increase the Nyquist frequency as you increase the sample rate, the alias signals should appear at their proper frequency.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Nyquist Frequency
Point
Amplitude
Frequency
Aliased Frequencies Actual Frequencies
Figure 3-21: How Aliased Frequencies Appear in an FFT
Fast Fourier Transforms
Considerations for Phase
Displays
When you set up an FFT math waveform to display the phase angle of the frequencies contained in a waveform, take into account the reference point the phase is measured against. You may also need to use phase suppression to reduce noise in your FFTs.
Establish a Zero Phase Reference Point. The phase of each frequency is measured with respect to the zero phase reference point. The zero reference point is the point at the center of the FFT math waveform but corresponds to various points on the source (time domain) record. (See Figure 3-19 on page 3–34.)
To measure the phase relative to most source waveforms, you need only to center the positive peak around the zero phase point. (For instance, center the positive half cycle for a sine or square wave around the zero phase point.) Use the following method:
H First be sure the FFT math waveform is selected in the More menu, then set
horizontal position to 50% in the Horizontal menu. This positions the zero phase reference point to the horizontal center of the screen.
H In the Horizontal menu, vary the trigger position to center the positive peak
of the source waveform at the horizontal center of screen. Alternately, you can adjust the trigger level (knob) to bring the positive peak to center screen if the phase reference waveform has slow enough edges.
When impulse testing and measuring phase, align the impulse input into the system to the zero reference point of the FFT time domain waveform:
H Set the trigger position to 50% and horizontal position to 50% for all record
lengths less than 15 K.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–37
Fast Fourier Transforms
H For records with a 15 K length, set the trigger position to 33%. Use the
horizontal position knob to move the trigger T on screen to the center
horizontal graticule line. H For records with 30 K, 50 K, or 60 K lengths, set the trigger position to
16.6%, 10%, or 8.3%, respectively. Use the horizontal position knob to
move the trigger T on screen and to the center horizontal graticule line. H Trigger on the input impulse.
Adjust Phase Suppression. Your source waveform record may have a noise component with phase angles that randomly vary from pi to pi. This noise could make the phase display unusable. In such a case, use phase suppression to control the noise.
You specify the phase suppression level in dB with respect to 1 V
RMS
. If the magnitude of the frequency is greater than this threshold, then its phase angle will be displayed. However, if it is less than this threshold, then the phase angle will be set to zero and be displayed as zero degrees or radians. (The waveform reference indicator at the left side of the graticule indicates the level where phase is zero for phase FFTs.)
It is easier to determine the level of phase suppression you need if you first create a frequency FFT math waveform of the source and then create a phase FFT waveform of the same source. Do the following steps to use a cursor measurement to determine the suppression level:
1. Do steps 1 through 7 of Operation that begins on page 328. Select dBV
RMS
(side) for the Set FFT Vert Scale to (side).
2. Press CURSOR
tion
(main) H Bars (side). Use the general purpose knob to align the
Mode (main) Independent (side) Func-
selected cursor to a level that places the tops of the magnitudes of frequen­cies of interest above the cursor but places other magnitudes completely below the cursor.
3. Read the level in dB from the @: readout. Note the level for use in step
4. Press MORE
Set FFT Vert Scale to
(main) Change Waveform Definition menu (side). Press
(side) repeatedly to choose either Phase (rad) or
5.
Phase (deg).
3–38
5. Press Suppress Phase at Amplitudes (side). Use the general purpose knob
or keypad to set phase suppression to the value obtained using the H Bar cursor. Do not change the window selection or you will invalidate the results obtained using the cursor.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Fast Fourier Transforms

FFT Windows

To learn how to optimize your display of FFT data, read about how the FFT windows data before computing the FFT math waveform.
Windowing Process. The oscilloscope multiplies the FFT time domain record by one of four FFT windows before it inputs the record to the FFT function. Figure 3-22 shows how the time domain record is processed.
The FFT windowing acts like a bandpass filter between the FFT time domain record and the FFT frequency domain record. The shape of the window controls the ability of the FFT to resolve (separate) the frequencies and to accurately measure the amplitude of those frequencies.
Selecting a Window. You can select your window to provide better frequency resolution at the expense of better amplitude measurement accuracy in your FFT, better amplitude accuracy over frequency resolution, or to provide a compromise between both. You can choose from these four windows: Rectangular, Hamming, Hanning, and Blackman-Harris.
In step 8 (page 3–29) in Operation, the four windows are listed. In general, choose a window that can just resolve between the frequencies you want to measure. That way, you will have the best amplitude accuracy and leakage elimination while still separating the frequencies.
You can often determine the best window empirically by first using the window with the most frequency resolution (rectangular), and then proceeding toward that window with the least (Blackman-Harris) until the frequencies merge. Use the window just before the window that lets the frequencies merge for best compromise between resolution and amplitude accuracy.
NOTE. If the Hanning window merges the frequencies, try the Hamming window before settling on the rectangular window. Depending on the distance of the frequencies you are trying to measure from the fundamental, the Hamming window sometimes resolves frequencies better than the Hanning.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–39
Fast Fourier Transforms
FFT Time Domain Record
Xs
FFT Window
FFT Time Domain Record
After Windowing
FFT
FFT Frequency Domain Record
Figure 3-22: Windowing the FFT Time Domain Record
Window Characteristics. When evaluating a window for use, you may want to
examine how it modifies the FFT time domain data. Figure 3-23 shows each window, its bandpass characteristic, bandwidth, and highest side lobe. Consider the following characteristics:
H The narrower the central lobe for a given window, the better it can resolve a
frequency.
H The lower the lobes on the side of each central lobe are, the better the
amplitude accuracy. The Hanning and Blackman/Harris windows taper the data at the end of the record to zero.
H Narrow lobes increase frequency resolution because they are more selective.
Lower side lobe amplitudes increases accuracy because they reduce leakage.
3–40
When using bell shaped widows, be sure that the most interesting parts of the signal in the time domain record are positioned in the center region of the window so that the tapering does not cause severe errors.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Fast Fourier Transforms
FFT Window Type Bandpass Filter
Rectangular Window
Hamming Window
Hanning Window
Blackman-Harris Window
0 dB
-20
-40
-50
0 dB
-20
-40
-60
0 dB
-20
-40
-60
-80
0 dB
-20
-40
-60
-80
-100
-101
–3 dB
Bandwidth
0.89
1.28
1.28
1.28
Highest
Side Lobe
13 dB
43 dB
32 dB
94 dB
Figure 3-23: FFT Windows and Bandpass Characteristics
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–41
Fast Fourier Transforms
3–42
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

File System

The file system provides file utilities and a floppy disk drive for saving hardcopies, setups, and waveforms. This section describes how to manage (delete, rename, and so forth) these files using the file system.

To Access the File Utilities

To bring up the File Utilities menu:
1. Press the SETUP button to bring up the Save/Recall Setup menu, or press
2. Press File Utilities in the main menu to bring up the File Utilities side menu.
the WAVEFORM button to bring up the Save/Recall Waveform menu, or press the Shift HARDCOPY button to bring up the Hardcopy menu.
(See Figure 3-24.)
Free disk space
Figure 3-24: File Utilities
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–43
File System

To Delete a File or Directory

To delete a file or directory, turn the general purpose knob until it scrolls the cursor over the name of the file or directory to delete. Then, press the side menu Delete button.
To delete all files in the file list, set the cursor to the *.* selection. The digitizing oscilloscope deletes a directory and all its contents.

To Rename a File or Directory

To rename a file or directory, turn the general purpose knob until it scrolls the cursor over the name of the file or directory to delete. For example, to rename the target file whose default name is TEK????? set the cursor over its name. Then, press the side menu Rename button.
The labelling menu should appear. Turn the general purpose knob or use the main-menu arrow keys to select each letter. Press Enter Char from the main menu to enter each letter. When you have entered the name, press the side menu OK Accept item. (See Figure 3-25).
3–44
Figure 3-25: File System Labelling Menu
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

To Copy a File or Directory

To copy a file or directory, turn the general purpose knob until it scrolls the cursor over the name of the file to copy. Then, press the side menu Copy button. The file menu reappears with the names of directories to copy to. Select a directory and press the side-menu button labelled Copy <name> to Selected Directory.
To copy all files, select the *.* entry. The digitizing oscilloscope copies a directory and all its contents.

To Print a File or Directory

To print a file, turn the general purpose knob until it scrolls the cursor over the name of the file to print. Then, press the side-menu Print button.
The Print-To side menu should appear. Select the port to print to from GPIB, RS-232, or Centronics. (See Figure 3-25.) Then the digitizing oscilloscope sends the file in its raw form out the port.
File System

To Create a Directory

To Set Confirm Delete

To Set Overwrite Lock

To create a new directory, press the side menu Create Directory button. The labelling menu appears. Turn the general purpose knob or use the main-
menu arrow keys to select each letter. Press Enter Char from the main menu to enter each letter. When you have entered the name, press the side menu OK Accept item. (See Figure 3-25.)
To turn on or off the confirm delete message, toggle the side menu Confirm Delete button.
When the confirm delete option is OFF, the digitizing oscilloscope immediately deletes files or directories. When the confirm option is ON, the digitizing oscilloscope warns you before it deletes files and gives you a chance to reconsider.
To turn on or off the file overwrite lock, toggle the side menu Overwrite Lock button.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–45
File System

To Format a Disk

When overwrite lock is on, the digitizing oscilloscope will not permit you to write over an existing file of the same name. An important reason to allow overwriting is to let you write files using a target file name that contains wild card characters (“?”). This means the digitizing oscilloscope creates sequential files whose names are similar except for the sequential numbers that go in the real name in the place of the question marks.
To format a 720 Kbyte or 1.44 Mbyte disk, turn the general purpose knob until it scrolls the cursor over the name of the drive to format (fd0:). Then, press the side menu Format button.
3–46
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual

Hardcopy

Supported Formats

You can get a copy of the digitizing oscilloscope display by using the hardcopy feature. This section lists hardcopy formats and describes how to setup and execute a hardcopy.
Different hardcopy devices use different formats. The digitizing oscilloscope supports the following formats:
H HP Thinkjet inkjet printer H HP Deskjet inkjet printer H HP Laserjet laser printer H Seiko DPU 411/412 portable thermal printer H HPGL Color Plot H EpsonR H InterleafR H Tag Image File Format (TIFFR) H PC PaintbrushR (PCXR) H Microsoft WindowsR file format (BMPR) H Encapsulated PostscriptR (Image, Mono Plot, and Color Plot)
Some formats, particularly Interleaf, Postscript, TIFF, PCX, BMP, and HPGL, are compatible with various desktop publishing packages. You can paste files created from the oscilloscope directly into a document when using any of those desktop publishing systems.
EPS Mono and Color formats are compatible with the Tektronix Phaser Color Printer, HPGL is compatible with the Tektronix HC100 Plotter, and Epson is compatible with the Tektronix HC200 Printer.

To Set Up for Making Hardcopies

Before you make a hardcopy, you need to set up communications and hardcopy parameters. This discussion assumes that the hardcopy device is already connected to the GPIB port on the rear panel.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–47
Hardcopy
To Set Communication
Parameters
To set up the communication parameters to talk to a printer attached directly to the oscilloscope GPIB port:
1. Press SHIFT UTILITY System (main) I/O (pop-up) Config-
(main) (see Figure 3-26).
ure
2. Press Hardcopy (Talk Only)
(side).
3–48
To Set Hardcopy
Parameters
Figure 3-26: Utility Menu System I/O
To specify the hardcopy format, layout, and type of hardcopy port:
1. Press SHIFT HARDCOPY MENU.
2. Press Format
(main) Thinkjet, Deskjet, Laserjet, Epson, DPU-411,
DPU-412, PCX, TIFF, BMP, EPS Image, EPS Mono, EPS Color (EPS stands for Encapsulated Postscript), Interleaf, or HPGL
–more–
3. Press SHIFT HARDCOPY MENU
Portrait
4. Press SHIFT HARDCOPY MENU
(side) to see all of these format choices.)
Layout (main) Landscape or
(side).
Port (main) to specify the output
(side). (Press
channel to send your hardcopy through. Select one of the optional hardcopy ports: GPIB, RS-232, Centronics, or File.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Hardcopy
To Date/Time Stamp Your
Hardcopy

To Set the Date and Time

To date and time stamp your hardcopy, do the following steps:
1. Press DISPLAY (side) to toggle the setting to On.
2. If the date and time are not set, a message instructing you to do so is displayed. If that is the case, skip steps 3 and 4 and continue with step 1 of To Set the Date and Time below. Then redo this procedure.
3. Press Clear Menu to remove the menu from the display so the date and time can be displayed. See Figure 3-27. (The date and time is removed from the display when menus are displayed.)
4. Press HARDCOPY to print your date/time stamped hardcopy.
If you need to set the date and time of the oscilloscope, do the following steps:
1. Press SHIFT UTILITY Year, Day Month, Hour, or Minute.
Readout Options (main) Display Date and Time
Config (pop-up) Set Date & Time (main)
Date and Time Display
To Print Directly to a
Hardcopy Device
Figure 3-27: Date and Time Display
2. Use the general purpose knob to set the parameter you have chosen to the value desired.
3. Repeat steps 1 and 2 to set other parameters as desired.
4. Press OK Enter Date/Time
sets the seconds to zero.
5. Press CLEAR MENU to see the date/time displayed with the new settings.
To make your hardcopies, use the procedures that follow.
(side) to put the new settings into effect. This
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–49
Hardcopy
Connect to a Hardcopy Device. To connect the digitizing oscilloscope directly to a hardcopy device, determine which interface and cable the device uses, and connect accordingly. (See Figure 3-28.)
Digitizing Oscilloscope
GPIB, RS-232,
or Centronics Cable
Hardcopy Device
Figure 3-28: Connecting the Digitizing Oscilloscope Directly to the Hardcopy Device
Print. To print a single hardcopy or send additional hardcopies to the digitizing
oscilloscope spool (queue) while waiting for earlier hardcopies to finish printing, press HARDCOPY.
Abort. To stop and discard the hardcopy being sent, press HARDCOPY again while the hardcopy in process message is still on screen.
Add to the Spool. To add additional hardcopies to the printer spool, press HARDCOPY again after the hardcopy in process message is removed from the screen.
3–50
You can add hardcopies to the spool until it is full. When the spool is filled by adding a hardcopy, the message Hardcopy in Process Press HARDCOPY to abort remains displayed. When the printer empties enough of the spool to finish adding the last hardcopy it does so and then removes the message.
Clear the Spool. To remove all hardcopies from the spool, press SHIFT HARDCOPY MENU Spool
(side).
Clear Spool (main) OK Confirm Clear
The oscilloscope takes advantage of unused RAM when spooling hardcopies. The size of the spool is, therefore, variable. The number of hardcopies that can be spooled depends on three variables:
H The amount of unused RAM H The hardcopy format chosen H The complexity of the display
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Hardcopy

To Save to a Floppy Disk

To send hardcopies to a floppy disk (see File System on page 3–43), do the following steps:
1. Set up the digitizing oscilloscope communication and hardware parameters as outlined in To Set Up for Making Hardcopies on page 3–47.
2. Insert a formatted 720 Kbyte or 1.44 Mbyte floppy disk into the disk slot of the digitizing oscilloscope.
NOTE. See page 3–43 for information on formatting disks, deleting and renaming files, setting overwrite lock, and using wild card characters.
3. Press SHIFT specify that any hardcopy will be output to a file on the floppy drive. The file list and its scrollbar will appear.
4. Turn the general purpose knob to place the scroll bar over the file in which to store the hardcopy.
NOTE. Upon power on, the digitizing oscilloscope creates the wild card file
TEK?????.FMT for storing hardcopies, where “.FMT is replaced by the hardcopy format you select. Selecting this file and pressing Hardcopy stores a hardcopy in a uniquely named, sequentially numbered file. For instance, the digitizing oscilloscope saves the first hardcopy you save to the file
TEK00001.FMT, the second to TEK00002.FMT, and so on.
HARDCOPY MENU Port (main) File (side) to
5. Display what you want on your hardcopy.
6. Press HARDCOPY to print your hardcopy to the selected file.

To Print Using a Controller

TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
To make your hardcopies, use the procedures that follow.
Connect to a Hardcopy Device. To connect to a controller with two ports between the digitizing oscilloscope and the hardcopy device, connect from the digitizing oscilloscope GPIB connector (rear panel) to the controller GPIB port and from the controller RS-232 or Centronics port to the hardcopy device (see Fig­ure 3-29). Use the GPIB port to remotely request and receive a hardcopy from the digitizing oscilloscope. Use the RS-232 or a Centronics port on the controller to print output.
3–51
Hardcopy
Digitizing Oscilloscope
PC Compatible
GPIB Cable
Hardcopy Device
Centronics or RS-232 Cable
Figure 3-29: Connecting the Digitizing Oscilloscope and Hardcopy Device Via a PC
Print. If your controller is PC-compatible and it uses the Tektronix GURU or
S3FG210 (National Instruments GPIB-PCII/IIA) GPIB package, do the following steps to print a hardcopy:
1. Use the MS-DOS cd command to move to the directory that holds the
software that came with your GPIB board. For example, if you installed the software in the GPIB-PC directory, type: cd GPIB-PC
2. Run the IBIC program that came with your GPIB board. Type: IBIC
3. Type: IBFIND DEV1 where DEV1 is the name for the digitizing
oscilloscope you defined using the IBCONF.EXE program that came with the GPIB board.
NOTE. If you defined another name, use it instead of “DEV1. Also, remember that the device address of the CONF.EXE program should match the address set in the
digitizing oscilloscope as set with the IB-
digitizing oscilloscope
Utility menu (typically, use “1”). Making hardcopies using some hardcopy formats may generate a time-out on
your controller. If a time-out occurs, increase the time-out setting of your controller software.
4. Type: IBWRT “HARDCOPY START”.
3–52
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
Hardcopy
NOTE. Be sure the digitizing oscilloscope Utility menu is set to Talk/Listen and not Hardcopy (Talk Only) or you will get an error message at this step. Setting the digitizing oscilloscope Utility menu was described in To Set Up for Making Hardcopies on page 3–47.
5. Type: IBRDF <Filename>where <Filename> is a valid DOS file name you want to call your hardcopy information. It should be 8 characters long with up to a 3 character extension. For example, you could type ibrdf screen1”.
6. Exit the IBIC program by typing: EXIT
7. Copy the data from your file to your hardcopy device. Type: COPY <Filename> <Output port> </B> where:
<Filename> is the name you defined in step 5. <Output port> is the PC output port your hardcopy device is connected to
(such as LPT1 or LPT2). For example, to copy (print) a file called screen1 to a printer attached to the
lpt1 parallel port, type copy screen1 lpt1: /B”.
Your hardcopy device should now print a picture of the digitizing oscilloscope screen.
NOTE. If you transmit hardcopy files across a computer network, use a binary (8-bit) data path.
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–53
Loading...