Tektronix TDS 400 Series, TDS 500 Series, TDS 600 Series, TDS 800 Series Instruction Manual

Instruction Manual
TDS Family Option 13 RS-232/Centronics Hardcopy Interface
070-8567-01
First Printing: December 1992
www.tektronix.com
*P070856701*
070856701
Instrument Serial Numbers
Each instrument manufactured by Tektronix has a serial number on a panel insert or tag, or stamped on the chassis. The first letter in the serial number designates the country of manufacture. The last five digits of the serial number are assigned sequentially and are unique to each instrument. Those manufactured in the United States have six unique digits. The country of manufacture is identified as follows:
B010000 Tektronix, Inc., Beaverton, Oregon, USA E200000 Tektronix United Kingdom, Ltd., London J300000 Sony/Tektronix, Japan H700000 Tektronix Holland, NV, Heerenveen, The Netherlands
Instruments manufactured for Tektronix by external vendors outside the United States are assigned a two digit alpha code to identify the country of manufacture (e.g., JP for Japan, HK for Hong Kong, IL for Israel, etc.).
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
Printed in U.S.A.
Copyright E Tektronix, Inc., 1992. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. The following are registered trademarks: TEKTRONIX, TEK, TEKPROBE, and SCOPEĆMOBILE.
WARRANTY
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Welcome
This instruction manual provides operating and service information for the TDS Family Option 13, RSĆ232/Centronics Hardcopy Interface. The informaĆ tion contained here applies to the TDS 400, TDS 500, TDS 600, and TDS 800 series digitizing oscilloscopes. This manual includes the following sections:
H Related Manuals (follows this introduction).
H Operating Information provides a product description and operating
procedures for using the RSĆ232/Centronics Hardcopy Interface capabilĆ ities.
H Service contains moduleĆlevel service information including a troubleĆ
shooting tree, removal and replacement instructions, and a replaceable parts list.
Related Manuals
Other documentation for the TDS Family of digitizing oscilloscopes includes the following:
H The User and Reference manuals give more detailed operating informaĆ
tion for the oscilloscope.
H The Service manual provides extended moduleĆlevel service information.
TDS Family Option 13 Instruction Manual
v
Welcome
vi
Welcome
Operating Information
Contents
Safety xi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating Information 1Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Product Description 1Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Before You Begin 1Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hardcopy Setup 1Ć2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Communication, Hardware, and Software Setup 1Ć4. . . . . . . . . . . .
RSĆ232 Communication Setup 1Ć4. . . . . . . . . . . . . . . . . . . . . . . . . .
RSĆ232 Hardware Setup 1Ć5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RSĆ232 Software Setup 1Ć6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connections 1Ć7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Printing the Hardcopy 1Ć7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Information
Service Information 2Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Before Servicing 2Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Strategy for Servicing 2Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preventing Electrostatic Discharge (ESD) 2Ć2. . . . . . . . . . . . . . . . . .
Troubleshooting 2Ć3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting 2Ć3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Diagnostics 2Ć4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Executing Extended Diagnostics Routine 2Ć4. . . . . . . . . . . . . . . . .
Removal and Replacement Procedures 2Ć5. . . . . . . . . . . . . . . . . . . . . . .
Before You Begin 2Ć5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Rear Cover and Cabinet Removal 2Ć6. . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500, 600, and 800 Oscilloscopes 2Ć6. . . . . . . . . . . . . . . . . . .
TDS 400 Oscilloscopes 2Ć7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Circuit Board Assembly Removal 2Ć9. . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500, 600, and 800 Oscilloscopes 2Ć9. . . . . . . . . . . . . . . . . . .
TDS 400 Oscilloscopes 2Ć10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Circuit Board Removal From Assembly 2Ć11. . . . . . . . . . . . . . . . . . . .
TDS 500, 600, and 800 Oscilloscopes 2Ć11. . . . . . . . . . . . . . . . . . .
TDS 400 Oscilloscopes 2Ć11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS Family Option 13 Instruction Manual
vii
Contents
Replacement of Circuit Board 2Ć12. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500, 600, and 800 Oscilloscopes 2Ć12. . . . . . . . . . . . . . . . . . .
TDS 400 Oscilloscopes 2Ć12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replace Cabinet and Rear Cover 2Ć13. . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500, 600, and 800 Oscilloscopes 2Ć13. . . . . . . . . . . . . . . . . . .
TDS 400 Oscilloscopes 2Ć13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Execute Diagnostics 2Ć13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replaceable Parts List 2Ć15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Parts Ordering Information 2Ć15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Part Replacement 2Ć15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Replaceable Parts List 2Ć15. . . . . . . . . . . . . . . . . . . . . . . . . .
Item Names 2Ć15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Indentation System 2Ć16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Abbreviations 2Ć16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Option 13 - TDS 500, 600, and 800 Oscilloscopes 2Ć17. . . . . . . .
Option 13 - TDS 400 Oscilloscopes 2Ć18. . . . . . . . . . . . . . . . . . . . .
viii
Contents
List of Figures
Figure 1Ć1:ăHardcopy Format 1Ć2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 1Ć2:ăHardcopy Layout Formats 1Ć3. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 1Ć3:ăRSĆ232 Communication Setup 1Ć4. . . . . . . . . . . . . . . . . . . . . .
Figure 1Ć4:ăRSĆ232 Hardware Setup 1Ć5. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 1Ć5:ăRSĆ232 Software Setup 1Ć6. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 1Ć6:ăConnecting the Oscilloscope to the Printer or Plotter 1Ć7. . .
Figure 2Ć1:ăTroubleshooting Tree 2Ć3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2Ć2:ăRear Cover and Cabinet Removal
(TDS 500, 600, and 800) 2Ć6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2Ć3:ăRear Cover and Cabinet Removal (TDS 400) 2Ć8. . . . . . . . . .
Figure 2Ć4:ăCircuit Board Assembly Removal
(TDS 500, 600, and 800) 2Ć9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2Ć5:ăA23 Option 13 Assembly Removal (TDS 400) 2Ć10. . . . . . . . .
Figure 2Ć6:ăRemoving Circuit Board From Assembly
(TDS 500, 600, and 800) 2Ć11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2Ć7:ăRemoving Circuit Board From Assembly (TDS 400) 2Ć12. . . .
Figure 2Ć8:ăRSĆ232/Centronics Hardcopy Interface Replaceable
Parts (TDS 500, 600, and 800) 2Ć17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2Ć9:ăRSĆ232/Centronics Hardcopy Interface Replaceable
Parts (TDS 400) 2Ć19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS Family Option 13 Instruction Manual
ix
List of Figures
x
Contents
Safety
Please take a moment to review these safety precautions. They are for your protection and to prevent damage to the digitizing oscilloscope. This safety information applies to all operators and service personnel.
Symbols and Terms
These two terms appear in manuals:
H
statements identify conditions or practices that could result in
damage to the equipment or other property.
H
statements identify conditions or practices that could result in
personal injury or loss of life.
These two terms appear on equipment:
H CAUTION indicates a personal injury hazard not immediately accessible
as one reads the marking, or a hazard to property including the equipĆ ment itself.
H DANGER indicates a personal injury hazard immediately accessible as
one reads the marking.
This symbol appears in manuals:
TDS Family Option 13 Instruction Manual
StaticĆSensitive Devices
These symbols appear on equipment:
DANGER
High Voltage
Protective
ground (earth)
terminal
ATTENTION
Refer to
manual
xi
Safety
Specific Precautions
Observe all of these precautions to ensure your personal safety and to prevent damage to either the digitizing oscilloscope or other connected equipment.
Power Source
The digitizing oscilloscope operates from a power source that will not apply more than 250 V supply conductor and ground. A protective ground connection, through the grounding conductor in the power cord, is essential for safe system operaĆ tion.
between the supply conductors or between either
RMS
Grounding the Digitizing Oscilloscope
The digitizing oscilloscope is grounded through the power cord. To avoid electric shock, plug the power cord into a properly wired receptacle where earth ground has been verified by a qualified service person. Do this before making connections to the input or output terminals of the digitizing oscilloĆ scope.
Without the protective ground connection, all parts of the digitizing oscilloĆ scope are potential shock hazards. This includes knobs and controls that may appear to be insulators.
Use the Proper Power Cord
Use only the power cord and connector specified for your product. Use only a power cord that is in good condition.
Use the Proper Fuse
To avoid fire hazard, use only the fuse specified in the parts list for your product, matched by type, voltage rating, and current rating.
Do Not Remove Covers or Panels
To avoid personal injury, do not operate the equipment without the panels or covers.
Electric Overload
Never apply to a connector a voltage that is outside the range specified for that connector.
Do Not Operate in Explosive Atmospheres
This equipment provides no explosion protection from static discharges or arcing components. Do not operate the digitizing oscilloscope in an atmoĆ sphere of explosive gases.
xii
Safety
Safety
Do Not Service Alone
Do not perform internal service or adjustment of this product unless another person capable of rendering first aid and resuscitation is present.
Use Care When Servicing With Power On
Dangerous voltages exist at several points in this product. To avoid personal injury, do not touch exposed connections or components while power is on.
Disconnect power before removing protective panels, soldering, or replacing components.
TDS Family Option 13 Instruction Manual
xiii
Safety
xiv
Safety
Operating Information
This section provides operating information for the TDS Family Option 13, RSĆ232/Centronics Hardcopy Interface. The following subsections are included:
H Product Description gives a general description of the RSĆ232/CentronĆ
ics Hardcopy Interface and the basic hardware and firmware requireĆ ments.
H Before You Begin tells how to run the internal diagnostic routine.
H Hardcopy Setup tells how to set RSĆ232 and Centronics hardcopy paĆ
rameters from the oscilloscope.
H Communication, Hardware, and Software Setup contains procedures for
setting RSĆ232 communication, hardware, and software configurations from the oscilloscope. There are no optional settings for Centronics.
H Connections contains basic steps for connecting the oscilloscope to the
printer or plotter.
H Printing the Hardcopy tells how to send the onĆscreen information to the
printer or plotter.
Product Description
Before You Begin
The TDS Family Option 13 adds Centronics and RSĆ232 hardcopy (talk only) capability to the TDS Family of digitizing oscilloscopes. The option consists of a single circuit board, its attached connectors, and rear panel labels.
The RSĆ232/Centronics Hardcopy Interface allows you to make hard copies of waveforms (and other onĆscreen information) to a variety of graphic printers and plotters from the scope front panel.
To make this option work, you must have both the RSĆ232/Centronics HardĆ copy Interface hardware and Version 2.0 firmware (or future versions) installed on the oscilloscope.
To ensure the hardcopy feature is working properly, execute the internal diagnostic routine by doing the following steps:
1. Power on the digitizing oscilloscope.
The oscilloscope automatically executes a short internal diagnostic routine that checks oscilloscope functions. When finished, the oscilloĆ scope displays an onĆscreen report of any failed modules or features.
TDS Family Option 13 Instruction Manual
1Ć1
Operating Information
2. If the hardcopy feature fails, refer to the Troubleshooting procedures starting on page 2Ć3.
3. If no failures are reported, go to Hardcopy Setup.
Hardcopy Setup
Hardcopy parameters must be set up before you can make a hardcopy through either the RSĆ232 or Centronics interface.
Setting Hardcopy Parameters
The Format menu lets you select the way the oscilloscope formats the hardcopy data.
1. Press SHIFT HARDCOPY MENU Format(main).
2. Select Thinkjet, Deskjet, Laserjet, Epson, Interleaf, TIFF, PCX, BMP,
EPS Image, EPS Mono, and EPS Color or HPGL (side). Press
-more- (side) to see all choices. (See Figure 1Ć1.)
EPS Mono and Color (EPS stands for Encapsulated Postscript) formats are compatible with the Tektronix Phaser Color Printer, HPGL is compatĆ ible with the Tektronix HC100 Plotter (or equivalent), and Epson is comĆ patible with the Tektronix HC200 Printer (or equivalent dot matrix printer).
1Ć2
Figure 1Ć1:ăHardcopy Format
Operating Information
Operating Information
The Layout menu lets you specify landscape or portrait format for the hardĆ copy.
Press SHIFT HARDCOPY MENU
Layout (main) Landscape or
Portrait (side) (see Figure 1Ć2).
Landscape Format Portrait Format
Figure 1Ć2:ăHardcopy Layout Formats
The Port menu lets you choose the hardcopy output channel.
Press SHIFT HARDCOPY MENU Port (main) RSĆ232 or Centronics (side).
If you select RSĆ232 as the hardcopy port, you must also set the commuĆ nication, hardware, and software parameters starting on page 1Ć5. If you select Centronics as the hardcopy port, no further setup is required. ProĆ ceed to Connections on page 1Ć7.
TDS Family Option 13 Instruction Manual
1Ć3
Operating Information
Communication, Hardware, and Software Setup
The following communication, hardware, and software settings are only for the RSĆ232 interface. No optional settings are offered for the Centronics interface.
RSĆ232 Communication Setup
Press SHIFT UTILITY System (main)  I/O (popĆup) Configure (main)
Hardcopy (side) (see Figure 1Ć3).
1Ć4
Figure 1Ć3:ăRSĆ232 Communication Setup
Operating Information
Operating Information
RSĆ232 Hardware Setup
The oscilloscope hardware settings must match the rearĆpanel parameter switch settings on the printer or plotter (see Connections on page 1Ć7).
1. Press SHIFT UTILITY Hardware Setup (main) (see Figure 1Ć4).
2. Press Baud Rate (side) 4800, 9600, or 19200. You can also use the general purpose knob (or the keypad, if available) to enter a selection.
3. Press Stop Bits (side) the general purpose knob or the keypad to enter a selection.
4. Press Parity (side) and toggle to select Even, Odd or None.
5. Press Hard Flagging (side) and toggle to select ON or OFF.
Port (main) RSĆ232 (popĆup)
and toggle to select 300, 600, 1200, 2400,
 and toggle to select 1 or 2. You can also use
TDS Family Option 13 Instruction Manual
Figure 1Ć4:ăRSĆ232 Hardware Setup
1Ć5
Operating Information
RSĆ232 Software Setup
Press Software Setup (main)  Soft Flagging (side) and toggle to select ON or OFF (see Figure 1Ć5).
Figure 1Ć5:ăRSĆ232 Software Setup
1Ć6
Operating Information
Operating Information
Connections
To connect the oscilloscope directly to an RSĆ232 or CentronicsĆbased printer or plotter (see Figure 1Ć6) do the following steps.
Oscilloscope
RSĆ232 or Centronics DoubleĆshielded Cable
Printer or Plotter
(e.g., Tek HC100)
Figure 1Ć6:ăConnecting the Oscilloscope to the Printer or Plotter
1. Leave power on to the oscilloscope.
2. Power off the printer or plotter.
3. Set the parameter switches on the rear panel of the RSĆ232 or CentronĆ ics printer or plotter to match the oscilloscope hardware settings. ConĆ sult the printer or plotter instruction manual for help with setting the switches.
Printing the Hardcopy
4. Power on the printer or plotter.
5. Connect a doubleĆshielded RSĆ232 or CentronicsĆcompatible cable to the corresponding ports on the printer or plotter and the oscilloscope. Below is a list of recommended cables for use with RSĆ232 and CentronĆ icsĆbased printers.
Tektronix
Printer
RSĆ232
Part Number Printer Cable Description
012Ć1298Ć00 Oscilloscope to 25Ćpin male RSĆ232
printer
Centronics 012Ć1250Ć00 Oscilloscope to Centronics printer
Press HARDCOPY.
Refer to your oscilloscope user manual for more information on how to use the hardcopy command.
TDS Family Option 13 Instruction Manual
1Ć7
Operating Information
1Ć8
Operating Information
WARNING
The following servicing instructions are for use only by qualified personnel. To avoid injury, do not perform any servicing other than that stated in the operating instructions unless you are qualified to do so. Refer to all safety summaries before performing any service.
Service Information
This section contains information you need to troubleshoot and service the RSĆ232/Centronics Hardcopy Interface. It includes the following subsections:
H Troubleshooting contains information for determining whether the
RSĆ232/Centronics Hardcopy Interface has failed. A troubleshooting tree and procedures for executing the oscilloscope short and extended internal diagnostic routines are included.
H Removal and Replacement Procedures gives stepĆbyĆstep procedures
for removal and replacement of the RSĆ232/Centronics Hardcopy InterĆ face circuit board.
H Replaceable Parts List includes replacement part numbers with a correĆ
sponding exploded view and ordering information.
Before Servicing
Strategy for Servicing
To prevent injury to yourself or damage to the oscilloscope, do the following before you attempt service:
H Be sure you are a qualified service person.
H Read the section entitled Safety found at the beginning of this manual.
H Read the following Strategy for Servicing and Preventing Electrostatic
Discharge (ESD) Precautions on page 2Ć2.
The Troubleshooting subsection starting on page 2Ć3, contains procedures to determine whether the RSĆ232/Centronics Hardcopy Interface feature has failed. If the hardcopy feature fails, replace the circuit board with a replaceĆ ment part obtained from the factory. See the Replaceable Parts List starting on page 2Ć11, for replacement part numbers and ordering information.
TDS Family Option 13 Instruction Manual
2Ć1
Service Information
Preventing Electrostatic Discharge (ESD)
CAUTION
Electrostatic discharge can damage any semiconductor component on this oscilloscope option circuit board.
Precautions
When performing any service that requires internal access to the oscilloĆ scope, follow the following precautions to avoid damaging internal modules and their components due to ESD.
1. Minimize handling of staticĆsensitive modules.
2. Transport and store staticĆsensitive modules in their static protected containers or on a metal rail. Label any package that contains staticĆsenĆ sitive modules.
3. Discharge the static voltage from your body by wearing a grounded antistatic wrist strap while handling these modules. Perform service on staticĆsensitive modules only at a staticĆfree work station.
4. Remove anything capable of generating or holding a static charge on the work station surface.
5. When possible, handle circuit boards by the edges.
6. Do not slide the modules over any surface.
7. Avoid handling modules in areas that have a floor or workĆsurface coverĆ ing capable of generating a static charge.
2Ć2
Service Information
Troubleshooting
This subsection contains information and diagnostic procedures to deterĆ mine if the RSĆ232/Centronics Hardcopy Interface is faulty. The diagnostic routine checks oscilloscope functions but does not check printer function.
Troubleshooting
Execute extended
START
diagnostic routine
(see Page 2Ć4).
Use the troubleshooting tree in Figure 2Ć1 to determine if the RSĆ232/CenĆ tronics Hardcopy Interface is working properly.
No
Yes
Yes
Hardcopy feature passes.
No
Verify Cable.
Attempt to make a hard copy.
Pass?
Verify instrument configuration.
Attempt to make a hard copy.
Pass?
DONE
TDS Family Option 13 Instruction Manual
Order a replacement circuit board for the
RSĆ232/Centronics Hardcopy Interface (see the
Replaceable Parts List starting on page 2Ć11).
Replace the RSĆ232/Centronics Hardcopy Interface circuit board (follow the Removal and Replacement
Procedures starting on page 2Ć5).
Figure 2Ć1:ăTroubleshooting Tree
Pass?
No
Yes
DONE
2Ć3
Troubleshooting
Diagnostics
The oscilloscope has two levels of internal diagnostics: short confidence and extended. The oscilloscope automatically executes the short set at power on. If the hardcopy feature fails the short diagnostic routine, use the followĆ ing procedure to execute the extended diagnostics:
Prerequisites: Power on the oscilloscope and allow a 20 minute warmĆup before doing this procedure.
Executing Extended Diagnostics Routine
1. Press SHIFT UTILITY System (main) Diag/Err (popĆup) Execute (main) OK Confirm Run Test (side).
The internal diagnostics routine checks oscilloscope functions, but does not check printer function. When finished, the oscilloscope displays an onĆscreen report of any failed modules, features, or interfaces.
2. If the hardcopy feature fails, order a replacement RSĆ232/Centronics circuit board (see the Replaceable Parts List starting on page 2Ć11) then follow the Removal and Replacement Procedures starting on page 2Ć5.
2Ć4
Service Information
Removal and Replacement Procedures
Before You Begin
Use the procedures in this subsection to remove and replace a faulty RSĆ232/Centronics Hardcopy Interface circuit board.
WARNING
To prevent possible injury to service personnel or damage to the product components, read the safety precautions found at the beginning of this manual, and read Before Servicing, Strategy for Servicing, and Preventing ESD found at the beginning of this section.
WARNING
Disconnect the power cord from the line voltage source before doing any procedure in this subsection. Failure to do so could cause serious injury or death.
TDS Family Option 13 Instruction Manual
2Ć5
Removal and Replacement Procedures
Rear Cover and Cabinet Removal
TDS 500, 600, and 800 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ20 TorxR tip.
1. Install the front cover. Set the oscilloscope face down with the front cover on the work surface and the bottom facing up (see Figure 2Ć2).
2. Unplug the power cord from its receptacle at the rear panel.
3. Using the screwdriver with size TĆ20 TorxR (number 2 in Figure 2Ć2) securing the rear cover (number 1) to the oscilloscope. Lift off the rear cover.
4. Grasp the left and right edges of the cabinet (number 3) at the rear. Pull upward to slide the cabinet off the oscilloscope. Do not bind or snag the cabinet on the oscilloscope internal cabling as you remove it.
1
tip, remove the four screws
2
3
Figure 2Ć2:ăRear Cover and Cabinet Removal (TDS 500, 600, and 800)
2Ć6
Service Information
Removal and Replacement Procedures
TDS 400 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ15 TorxR tip.
1. Install the front cover. Set the oscilloscope face down with the front cover on the work surface and the bottom facing up (see Figure 2Ć3).
2. Unplug the power cord from its receptacle at the rear panel.
3. Using the screwdriver with size TĆ15 TorxR TorxR screw at the left side of the oscilloscope (number 4).
4. Remove the four screws (number 2 in Figure 2Ć3) securing the rear cover (number 1) to the oscilloscope. Lift off the rear cover.
5. Lift the cabinet (number 3) upwards to slide it off the oscilloscope. Take care not to bind or snag the cabinet on the oscilloscope's internal cabĆ ling as you remove it.
tip, remove the single TĆ15
TDS Family Option 13 Instruction Manual
2Ć7
Removal and Replacement Procedures
1
2
3
Figure 2Ć3:ăRear Cover and Cabinet Removal (TDS 400)
2Ć8
4
Service Information
Removal and Replacement Procedures
Circuit Board Assembly Removal
TDS 500, 600, and 800 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ20 TorxR tip.
1. Using the screwdriver with size TĆ20 TorxR (number 4 in Figure 2Ć4) securing the rear plate of the circuit board assembly.
2. Disconnect the cable connector (number 2) from the connector on the processorĆdisplay circuit board (number 1).
3. Pull out the RSĆ232/Centronics Hardcopy Interface circuit board assemĆ bly (number 3).
3
2
1
tip, remove the two screws
4
Figure 2Ć4:ăCircuit Board Assembly Removal (TDS 500, 600, and 800)
TDS Family Option 13 Instruction Manual
2Ć9
Removal and Replacement Procedures
TDS 400 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ15 TorxR tip.
1. Remove the three board supports (number 4 in Figure 2Ć5) by depressĆ ing the snap lock at the bottom end of the board support and tilting it outwards. The tabbed top end slips out of a notch in the top rail.
2. Using the screwdriver with size TĆ15 TorxR
tip, remove the two screws
on the back and the two screws on the side (numbers 1 and 2) securing the rear plate of the circuit board assembly.
3. Grasp the board (number 3) by its edge and pull outward to unplug it from J1 of the backplane assembly to complete the removal.
1
2
3
Figure 2Ć5:ăA23 Option 13 Assembly Removal (TDS 400)
2Ć10
4
Service Information
Removal and Replacement Procedures
Circuit Board Removal From Assembly
TDS 500, 600, and 800 Oscilloscopes
Equipment Required: One 3/16Ćinch nutdriver.
1. Using the nutdriver, remove the four lock connectors (number 3 in Figure 2Ć6) securing the circuit board to the rear plate.
2. Separate the circuit board with attached cable (number 1), from the rear plate with attached brackets (number 2).
2
3
1
Figure 2Ć6:ăRemoving Circuit Board From Assembly (TDS 500, 600, and 800)
TDS 400 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ15 TorxR tip.
1. Unplug the two cable housings from the circuit board (numbers 3 and 5 in Figure 2Ć7).
2. Using the screwdriver, remove the two TĆ15 TorxR screws (number 4) and separate the rear plate (number 1) from the circuit board.
TDS Family Option 13 Instruction Manual
2Ć11
Removal and Replacement Procedures
1
5
3
2
4
Replacement of Circuit Board
Figure 2Ć7:ăRemoving Circuit Board From Assembly (TDS 400)
TDS 500, 600, and 800 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ20 TorxR tip and one
3/16Ćinch nutdriver.
1. Replace the failed circuit board with a replacement RSĆ232/Centronics Hardcopy Interface circuit board ordered from the factory (see ReplaceĆ able Parts List starting on page 2Ć11 for ordering information). Do in reverse order steps 1 and 2 of the Circuit Board Removal From AssemĆ bly procedure on page 2Ć11.
2. Reinstall the RSĆ232/Centronics Hardcopy Interface circuit board assemĆ bly in to the oscilloscope. Do in reverse order steps 1, 2, and 3 of the Circuit Board Assembly Removal procedure on page 2Ć9.
TDS 400 Oscilloscopes
2Ć12
Equipment Required: One screwdriver with a size TĆ15 TorxR tip and one 3/16Ćinch nutdriver.
Service Information
Removal and Replacement Procedures
1. Replace the failed circuit board with a replacement ordered from the factory (see Replaceable Parts List starting on page 2Ć11 for ordering information). Do in reverse order steps 1 and 2 of the Circuit Board Removal From Assembly procedure on page 2Ć11.
2. Reinstall the RSĆ232/Centronics Hardcopy Interface circuit board assemĆ bly in to the oscilloscope. Do in reverse order steps 1, 2, and 3 of the Circuit Board Assembly Removal procedure on page 2Ć10
Replace Cabinet and Rear Cover
TDS 500, 600, and 800 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ20 TorxR tip and one
torque driver.
1. Do step 1 of the Rear Cover and Cabinet Removal procedure on page 2Ć6.
2. Do in reverse order steps 2, 3, and 4 of the Rear Cover and Cabinet Removal procedure on page 2Ć6.
NOTE
When reinstalling the four screws at the rear panel, use the torque driver to tighten the screws to 16 inchĆlbs (18.4 cmĆkg) torque.
TDS 400 Oscilloscopes
Equipment Required: One screwdriver with a size TĆ15 TorxR tip and one
torque driver.
1. Do step 1 of the Rear Cover and Cabinet Removal procedure on page 2Ć7.
Execute Diagnostics
TDS Family Option 13 Instruction Manual
2. Do in reverse order steps 2 through 5 of the Rear Cover and Cabinet Removal procedure on page 2Ć7.
NOTE
When reinstalling the four screws at the rear panel, use the torque driver to tighten the screws to 8 inchĆlbs (9.2 cmĆkg) torque.
To ensure the RSĆ232/Centronics Hardcopy Interface is working correctly, perform the Diagnostics procedure on page 2Ć4.
2Ć13
Removal and Replacement Procedures
2Ć14
Service Information
Replaceable Parts List
This subsection contains a list of the components that are replaceable for the RSĆ232/Centronics Hardcopy Interface. As described below, use this list to identify and order replacement parts.
Parts Ordering Information
Replacement parts are available from or through your local Tektronix, Inc., service center or representative.
Changes to Tektronix instruments are sometimes made to accommodate improved components as they become available and to give you the benefit of the latest circuit improvements. Therefore, when ordering parts, it is important to include the following information in your order:
H Part number
H Instrument type or model number
H Instrument serial number
H Instrument modification number, if applicable
If a part you order has been replaced with a different or improved part, your local Tektronix service center or representative will contact you concerning any change in the part number.
Part Replacement
The RSĆ232/Centronics Hardcopy Interface is serviced by replacing the failed part. You may purchase these replacement parts in the same way as other replacement parts.
Using the Replaceable Parts List
TDS Family Option 13 Instruction Manual
The tabular information in the Replaceable Parts List is arranged for quick retrieval. Understanding the structure and features of the list will help you find all the information you need for ordering replacement parts.
Item Names
In the Replaceable Parts List, an Item Name is separated from the descripĆ tion by a colon (:). Because of space limitations, an Item Name may someĆ times appear as incomplete. For further Item Name identification, U.S. Federal Cataloging Handbook H6Ć1 can be used where possible.
2Ć15
Replaceable Parts List
Indentation System
This parts list is indented to show the relationship between items. The following example is of the indentation system used in the Description column:
1ą 2ą 3ą 4ą 5 Name & Description Assembly and/or Component
Attaching parts for Assembly and/or Component
(END ATTACHING PARTS)
Detail Part of Assembly and/or Component Attaching parts for Detail Part
(END ATTACHING PARTS)
Parts of Detail Part Attaching parts for Parts of Detail Part
(END ATTACHING PARTS)
Attaching parts always appear at the same indentation as the item it mounts, while the detail parts are indented to the right. Indented items are part of, and included with, the next higher indentation. Attaching parts must be purchased separately, unless otherwise specified.
Abbreviations
Abbreviations conform to American National Standards Institute (ANSI) standard Y1.1
2Ć16
Service Information
Mfr. Code
Replaceable Parts List
Option 13 - TDS 500, 600, and 800 Oscilloscopes
CROSS INDEX - MFR. CODE NUMBER TO MANUFACTURER
Manufacturer Address City, State, Zip Code
TK1163
00779 AMP INC 2800 FULLING MILL
80009 TEKTRONIX INC 14150 SW KARL BRAUN DR
Fig. & Index
2Ć8-1
2Ć8-2 407-3825-00 2 BRACKET,CKT BD:PLASTIC,REAR TK1163 ORDER BY DESC
2Ć8-3 386-6182-00 1 PLATE,REAR:ALUMINUM,RS232/CENTRONIC 80009 386618200
2Ć8-4 131-0890-01 4 LOCK,CONNECTOR:4-40 X 0.312 L,HEX HD,STL 00779 205818-2
2Ć8-5 211-0730-00 2 SCREW,MACHINE: 80009 211073000
POLYCAST INC 9898 SW TIGARD ST TIGARD OR 97223
HARRISBURG PA 17105
BEAVERTON OR 97077-0001
No.
PO BOX 3608
PO BOX 500
Tektronix
Part No.
671-2437-00 1 CIRCUIT BD ASSY:RS232/CENTRONIC 80009 671243700
070-8567-01 1 MANUAL,TECH:INSTRUCTION,TDS FAMILY OPT13 80009 070856701
Effective Dscont Qty 12345 Name & Description
Serial No.
ACCESSORIES
3
Mfr.
Code
Mfr. Part No.
5
2
1
4
Figure 2Ć8:ăRSĆ232/Centronics Hardcopy Interface Replaceable Parts (TDS 500, 600, and 800)
TDS Family Option 13 Instruction Manual
2Ć17
Replaceable Parts List
Option 13 - TDS 400 Oscilloscopes
CROSS INDEX - MFR. CODE NUMBER TO MANUFACTURER
Mfr. Code
0KB01
80009 TEKTRONIX INC 14150 SW KARL BRAUN DR
Manufacturer Address City, State, Zip Code
STAUFFER SUPPLY 810 SE SHERMAN PORTLAND OR 97214
PO BOX 500
BEAVERTON OR 97077-0001
Fig. & Index
No.
2-9-1
-2 174-2612-00 1 CA ASSY,SP,ELEC:LOW PROFILE,RS232 80009 174261200
-3 407-4238-00 1 BRACKET,CKT BD: 80009 407423800
-4 348-1279-00 1 GASKET,EMI:VIDEO (OUT),SST 80009 348127900
-5 348-1350-00 1 GASKET,EMI:25 PIN D,CENTRONIX 80009 348135000
-6 214-3903-00 4 SCREW,JACK:4-40 X 0.312 L,HEX HD 80009 214390300
-7 174-3013-00 1 CA ASSY,SP,ELEC:CENTRONIX 80009 174301300
-8 213-0882-00 2 SCREW,TPG,TR:6-32 X 0.437 TAPTITE,PNH,STL 0KB01 ORDER BY DESC
-9 214-3796-00 1 HEATSINK,XSTR:ALUMINUM,VERT MOUNT 80009 214379600
Tektronix
Part No.
Effective Dscont
671-2756-00 1 CIRCUIT BD ASSY:RS232/CENTRONIX 80009 671275600
070-8567-01 1 MANUAL,TECH:INSTRUCTION 80009 070856701
Serial No.
12345 Name & Description
Qty
STANDARD ACCESSORIES
Mfr.
Code
Mfr.
Part No.
2Ć18
Service Information
Replaceable Parts List
4
3
2
1
7
8
9
5
6
Figure 2Ć9:ăRSĆ232/Centronics Hardcopy Interface Replaceable Parts (TDS 400)
TDS Family Option 13 Instruction Manual
2Ć19
Replaceable Parts List
2Ć20
Service Information