Tektronix TDS3054C, TDS3052C, TDS3014C, TDS3034C, TDS3012C DATASHEET

...
Digital Phosphor Oscilloscopes
TDS3000C Series Data Sheet
Ease-of-Use Features
Portable Design
Front-panel Measurement Data
25 Automatic Me asurements
FFT Standard
Multiple Language User Interface
WaveAlert®Automatic Waveform Anomaly Detection
TekProbe®Interface Supports Active, Differential, and Current Probes for Automatic Scaling and Units
Lightweight Design (only 7 lb./3.2 kg) for Easy Transport
Optional Internal Battery Operation Provides up to Three Hours Witho ut Line Power
USB Ho st Port for Easy Storage and Transfer of
Features & Benets
Key Performance Specifications
100 MHz,
2 or 4 Channels
Sample Rates up to 5 GS/s Real Time on All Channels
3,600 wfms/s Continuous Waveform Capture Rate
Suite of Adva nced Triggers
300 MHz, and 500 MHz Bandwidth Models
Applicati
on Modules for Specialized Analysis
Advanced Analysis M odu le
Limit Test
Telecommunications Mask Testing Module
Extended
601 Serial Digital Video Module
ing Module
Video Module
Applications
Digital Design and Debug
Video Installation and Service
Power Supply Design
Education and Training
Telecommunications Mask Testing
Manufacturing Test
General Bench Testing
Data Sheet
Performance You Need at a Price You Can Afford
Performance i
The TDS3000C Series Digital Phosphor Oscilloscopes (DPO) provide you with the perfo from 100 MHz to 500 MHz, with up to 5 GS/s sample rates for accurate representation of your signal.
DPOs Provide a
rmance you need at a price you can afford. Bandwidths range
n an Affordable Package
Greater Level of Insight into Complex
Signals
To solve a problem, rst you need to see it. The TDS3000C Series combines 3,600 wfms/s continuous waveform capture rate an d real-time intensity grading so you can see the problem and solve it.
Fast waveform capture rates on a continuous basis save time by quickly revealing the nature of faults so advanced triggers can be ap p lied to isolate them.
Real-time intensity grading highlights the details about the history of a signal’s activity, making it easier to understand the characteristics of the waveforms you’ve captured. Unlike other comparable oscilloscopes, the history rem
ains even after the acquisition is stopped.
Quickly Debug and Characterize Signals with DRT Sampling Te
The TDS3000C Series combines unique digital real-time (DRT) sampling technolog a wide range of signal types on all channels simultaneously. With the TDS3000C Series there is n o change in sampling rate when additional channels are turned on, unlike other comparable oscilloscopes. This sampling technology makes it possible to capture high-frequency information, such as glitches and edge anomalies, that elude other oscillosc reconstruction of each waveform.
chnology and sin(x)/x Interpolation
y with sin(x)/x interpolation to allow you to accurately characterize
opes in its class, while sin(x)/x interpolation ensures precise
Look for unintentional circuit noise with the TDS3000C Series’ FFT capability.
Custom video triggering allows the TDS3000C Series to trigger on standards such as RS-343 (26.2 kHz scan rate).
2 www.tektronix.com
The TDS3000C Series provides breakthrough test speeds for telecommunications line card testing. The telecom QUICKMENU puts all the commonly used telecom test functions on a single menu.
Digital Phosphor Oscilloscopes — TDS3000C Series
Easy Setup and Use
When working under tight deadlines, you need your oscilloscope to be intuitive; you want to minimize time spent learning and relearning how to use it. The TDS3000C Series oscilloscopes help reduce your learning curve. Simple navigat want to be quickly, so that you spend less time learning and more time on the task at hand.
ion and dedicated front-panel controls get you to where you
The TDS300 applicati
0C Series with the TDS3LIM module is ideal for manufacturing test
ons where fast Go/No-Go decisions are required.
Simple Docume
ntation and Analysis
The TDS3000C Series comes equipped with a USB host port so you can easily store a
OpenChoice images and wa
nd transfer measurement information to your PC.
®
PC Communication Software allows you to simply pull screen
veform data into a standalone desktop application or directly
into Microsoft Word and Excel.
To compl e m en
t OpenChoice, National Instruments LabVIEW SignalExpress™ Tektronix Edition Software provides you with extended capabilities including advanced analysis, data logging, remote instrument control, and live waveform analysis.
If you prefer not to use a PC for analysis, the TDS3000C Series comes standard with 25 automatic measurements, waveform add, subtract, divide, and multiply math functions, and Fast Fou rier Transform (FFT). Unlike other comparable oscilloscopes, the TDS3000C Series math and m easurement allows you t
o use the full acquisition record length or isolate a specific
occurrence within an acquisition.
Trace and identify ITU-R BT.601 video signals with the TDS3SDI 601 Serial Digital Video Module.
WaveAlert waveform anomaly detection alerts you to any waveform that deviates from the “normal” input such as the glitch on channel 2.
Easily transfer, document, and analyze data on your PC.
Instrument Control
Utilizing the built-in Ethernet port, e*Scope web-based remote control allows you to a control TDS3000C Series oscillosco pe from anywhere, using the Internet and your PC.
www.tektronix.com 3
Data Sheet
Work Where you Need to
The TDS3000C Series packs the power of a DPO in a compact design that is only 5.9 in. (149 mm) deep, freeing up valuable benchtop space. And when you need to move your oscilloscope to another lab, its portable 7 lb. (3.2 kg) design
makes for easy transport.
If your work demands even more mobility, then the optional battery pack will give you up to t
hree hours of operation without line power.
Characteristics
TDS3BATC provides you with up to three hours of portable battery operation.
TDS3000C Serie
Characteristic TDS3012C TDS3014C TDS3032C TDS3034C TDS3052C TDS3054C
Bandwidth 100 MHz 100 MHz 300 MHz 300 MHz 500 MHz 500 MHz Calculated Rise Time
(typical) Input Channels External Trigger Input Included on all models Sample Rate on each
channel Record Length 10 kpoints
Vertical Resolution 9 bits Vertical Sensitivity,
1M Vertical Sensitivity,
50 Input Cou
Input Impedance DC Gain A Maximum Input
Voltage, 1 M Maximum Input
Voltage, 50 Position Range ±5 div Bandwidth Limit 20 MHz 20 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz Time Base Range 4 ns to 10 s 4 ns to 10 s 2 ns to 10 s 2 ns to 10 s 1 ns to 10 s 1 ns to 10 s Time Base Accuracy ±20 ppm over any 1 ms time interval
Input/Output Interfaces
Ethernet Port RJ-45 connector, supports 10Base-T LAN USB Port Front-panel USB 2.0 host port
GPIB Port Full talk/listen modes, setting and measurements
RS-232-C Port DB-9 male connector, full talk/listen modes; control of all modes, settings and measurements
VGA Video Port DB-15 female connector, monitor output for direct display on large VGA-equipped monitors
External Trigger Input
pling
ccuracy
s Electrical Characteristics
3.5 ns 3.5 ns 1.2 ns 1.2 ns 0.7 ns 0.7 ns
242424
1.25 GS/s 1.25 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s 5 GS/s
BNC connector, input impedance >1 Min parallel with 17 pF; max input voltage is 150 V
1mV/divto10V/div
1mV/divto1V/div
AC, DC, GN
1MΩ in pa
(Optional with TDS3GV Communications Module)
(Optional with TDS3GV Communications Module)
(Optional with TDS3GV Communications Module)
rallel with 13 pF or 50
150 V
with peaks at 400 V
RMS
5V
with peaks at 30 V
RMS
Supports USB ash drive
Baudratesupto38,400
D
±2%
RMS
4 www.tektronix.com
Digital Phosphor Oscilloscopes — TDS3000C Series
Acquisition Modes
Mode Description
DPO Captures and displays complex waveforms, random
events and subtle patterns in actual signal behavior. DPOs provide 3 dimensions of signal information in real time: Amplitude, time, and the distribution of amplitude over time
Peak Detect
High-frequency and random glitch capture. Captures glitches as narrow as 1 ns (typical) using acquisition hardware at all time base settings
WaveAlert
®
Monitors the incoming signals on all channels and alerts the user to any waveform that deviates from the normal
waveform being acquired Sample Sample data only Average
Waveform averaged, selectable from 2 to 512 Envelope Min-max values acquired over one or more acquisitions Single Sequence Use the Single Sequence button to capture a single
triggered acquisition sequence at a time
Trigger System
Characteristic
Main Trigger Modes
Description
Auto (supports Roll Mode for 40 ms/div and slower),
Normal, Single Sequence B Trigger Trigger After Time Range Trigger After Events
Trigger after time or events
13.2nsto50s
1 to 9,999,999 events Range
Trigger Types
Trigger Description
Edge
Conventional level-driven trigger. Positive or negative
slope on any channel. Coupling selections: AC, DC,
Noise Reject, HF Reject, LF Reject Video
Trigger on all lines or individual lines, odd/even or all
elds on NTSC, PAL, SECAM Extended Video
Trigger on specic lines in broadcast and
non-broadcast (custom) standards and on analog
HDTV formats (1080i, 1080p, 720p, 480p). Requires
TDS3VID or TDS3SDI application module Pulse Width
(or Glitch) Runt
Trigger on a pulse width <, >, =, to a selectable time
limit ranging from 39.6 ns to 50 s
Trigger on a pulse that crosses one threshold but fails to
cross a second threshold before crossing the rst again Slew Rate Trigger on pulse edge rates that are either faster or
slower than a set rate. Edges can be rising, falling, or
either Pattern
Species AND, OR, NAND, NOR when true or false for
aspecifictime State Any logic state. Triggerable on rising or falling edge of
a clock. Logic triggers can be used on combinations of
2 inputs (not 4) Comm Provides isolated pulse triggering required to perform
DS1/DS3 telecommunications mask testing per ANSI
T1.102 standard. Requires TDS3TMT application
module Alternate
Sequentially uses each active channel as a trigger
source
Waveform Measurements
Characteristic
Cursors Automatic Measurements
Description
Amplitude, Time Display any four measurements from any combination
of waveforms. Or display all measurements with measurement snapshot feature. Measurements include Period, Frequency, +Width, -Width, Rise time, Fall time, +Duty cycle, -Duty cycle, +Overshoot, High, Low, Max, Min, Peak-to-peak, Amplitude, Mean, Cycle mean, RMS, Cycle RMS, Burst width, Delay, Phase, Area* Cycle Area*
1
Measurement Statistics Mean, Min, Max, Standard deviation. Requires
TDS3AAM application module
Thresholds
User-denable thresholds for automatic measurements; settable in percent or voltage
Gating Isolate a specic occurrence within an acquisition to
take measurements, using either the screen or cursors
*1Requires TDS3AAM application module.
Waveform Math
Characteristic
Arithmetic FFT
Description
Add, subtract, multiply, and divide waveforms Spectral magnitude. Set FFT vertical scale to Linear
RMS or dBV RMS, and FFT window to Rectangular, Hamming, Hanning, or Blackman-Harris
nced Math*
Adva
*1Requires TDS3AAM application module.
1
grate, Differentiate, Dene extensive algebraic
Inte
essions including analog waveforms, math
expr
tions, scalars, up to two user-adjustable variables
func
esults of parametric measurements. For example:
and r
g (Ch1-Mean(Ch1)) × 1.414 × VAR1)
(Int
: Requires TDS3AAM application module.
Note
Waveform Processing
cription
Characteristic
oset
Aut
Des
Single-button, automatic setup of all channels for vertical, horizontal and trigger systems, with undo autoset
Deskew
Channel-to-channel deskew ±10 ns may be manually entered for better timing measurements and more accurate math waveforms
Display Characteristics
Characteristic
Display Type Display Resolution Interpolation
Description
6.5 in. (165.1 mm) liquid-crystal TFT color display 640 horizontal × 480 vertical pixels (VGA)
Sin(x)/x Waveform styles Dots, vectors, variable persistence, innite persistence Graticules Full, grid, crosshair, and frame. NTSC, PAL, SECAM,
and vectorscope (100% and 75% color bars) with
optional TDS3VID or TDS3SDI application modules Format
YT, XY, and Gated XYZ (XY with Z-axis blanking
available on 4-channel instruments only)
Power Source
Characteristic
Description
AC line power
Source voltage 100 V
RMS
to 240 V
RMS
±10%
Source frequency 45 Hz to 440 Hz from 100 V to 120 V
45 Hz to 66 Hz from 120 V to 240 V
Power consumption 75 W maximum
Battery power
Requires TDS3BATC, rechargeable lithium ion battery
pack
Operating time, typical
3hours
1
,
www.tektronix.com 5
Data Sheet
Environmental and Safety
Characteristic Operating
Temperature Humidity
Altitude To 3,000 m 15,000 m Electromagnetic
Compatibility
Safety UL61010B-1, CSA1010.1, IEC61010-1, EN61010-1
0ºCto+50ºC -40ºCto+71ºC
Operating and Nonoperating:
Up to 95% RH at or below +30 ºC
Operating and Nonoperating:
Up to 45% RH +30 ºC up to +50 ºC
Meets or exceeds EN61326 Class A, Annex D
radiated and conducted emissions and immunity;
EN6100-3-2 AC Powerline Harmonic Emissions;
EN6100-3-3 Voltage Changes, Fluctuation, and
Flicker; FCC 47 CFR, Part 15, Subpart B, Class
A; Australian EMC framework
Non-operating
Physical Characteristics
Instrument
Dimensions
Width 375.0 14.8 Height 176.0 6.9 Depth 149.0 5.9
Weight kg lb.
Instrument only 3.2 7.0
with accessories 4.5 9.8
ment Shipping
Instru
Package Dimensions mm in.
Width 502.0 19.8 Height 375.0 14.8 Depth 369.0 14.5
mount (RM3000)
Rack
Dimension mm in.
Width 484.0 19.0 Height 178.0 7.0 Depth 152.0 6.0
mm
in.
Ordering Information
TDS3000C Series Digital Phosphor O scilloscopes
Model Description
TDS3012C 100 MHz, 2 Channel, 1.25 GS/s TDS3014C 100 MHz, 4 Channel, 1.25 GS/s TDS3032C 300 MHz, 2 Channel, 2.5 GS/s TDS3034C 300 MHz, 4 Channel, 2.5 GS/s TDS3052C 500 MHz, 2 Channel, 5 GS/s TDS3054C 500 MHz, 4 Channel, 5 GS/s
Standard Accessories
Accessory Description
One P6139B, 500 MHz, 10x passive probe per channel
User Manual and Translated Front-panel Overlay
Power Cord Please specify plug option - see chart below Accessory T Protective Front Cover OpenChoice®PC
Communication Software
NI LabVIEW SignalExpress™ Tektronix Edition LE
Traceable Certicate of Calibration
Documentation CD 3 Year Warranty
ray
Please specify preferred language option - see chart
below
Enables fast and easy communication between
Windows PC and the TDS3000C Series through
LAN, GPIB, or RS-232. Transfer and save settings,
waveforms, measurements, and screen images
A fully interactive measurement software environment
optimized for the TDS3000C Series. Enables you
to acquire, generate, analyze, compare, import,
and save measurement data and signals using an
intuitive drag-and-drop user interface that does not
require any programming. Standard TDS3000C
Series support for acquiring, controlling, viewing, and
exporting your live signal data is permanently available
through the software. A 30-day trial period of the full
version provides additional signal processing, advance
analysis, mixed signal, sweeping, limit testing, and
user-dened step capabilities. Order SIGEXPTE for
permanent full version capability
NIM/NIST
Covering all labor and parts excluding probes and
accessories
Options
Option
Opt. A0 Opt. A1 Opt. A2 Opt. A3 Opt. A5 Switzerland Opt. A6 Opt. A10 China Opt. A11 Opt. A99
Description
North America
Universal Euro
United Kingdom
Australia
Japan
India
No power cord
Language Options
Option
Opt. L0 Opt. L1 Opt. L2 Opt. L3 German Opt. L4 Spanish Opt. L5 Opt. L6 Opt. L7 Simplied Chinese Opt. L8 Traditional Chinese Opt. L9 Opt. L10 Opt. L99
Description
English
French
Italian
Japanese
Portuguese
Korean
Russian
No manual
6 www.tektronix.com
Digital Phosphor Oscilloscopes — TDS3000C Series
Recommended Accessories
Accessory Description
TDS3GV GPIB, VGA, RS-232 interface TDS3AAM
Advanced Analysis Module. Adds extended math capability, arbitrary math expressions, measurement statistics, and additional automated measurements
TDS3LIM Limit Testing Module. Adds custom waveform limit
testing capabilities
TDS3TMT Telecom Mask Testing Module. Adds pass/fail
compliance of ITU-T G.703 and ANSI T1.102 standards, custom mask testing, and more
TDS3VID
Extended Video Analysis Module. Adds video quickmenu, autoset, hold, line count trigger, video picture mode, vectorscope*
2
mode, HDTV format
trigger graticules, and more
TDS3SDI Serial/
Digital Video Module. Adds 601 serial digital
analog video conversion, video picture,
video to vectors
2
cope*
, and analog HDTV triggering capabilities,
and more
TDS3BATC Lithium-ion battery pack for up to 3 hours continuous
operation without line power TDS3CHG Fast charger for battery pack AC3000 Soft case for carrying instrument HCTEK4321 Hard plastic case for carrying instrument
(requires AC3000) RM3000 Rackmount kit SIGEXPTE NI LabVIEW SignalExpress™ Tektronix Edition
Software full version Service Manual English only (071-2507-xx)
TNGTDS01
Extensive instructions and step-by-step lab exercises
provide education about the operation of TDS3000C
Series Oscilloscopes. Kit includes self-paced CD-ROM
based manual and signal source board. Optional
hardcopy manual available for order separately
*2Vectorscope does not support composite video.
Recommended Probes
Probe Description
P6243 P5205 P5210
1GHz,1 pF input C 10x active probe
1.3 kV, 100 MHz high-voltage differential probe
5.6 kV, 50 MHz high-voltage differential probe P5100 2.5 kV, 100x high-voltage passive probe TCP202 50 MHz, 15 A AC/DC current probe TCP303* TCP305* TCP312*
3
3
3
15 MHz, 150 A current probe 50 MHz, 50 A current probe
100 MHz, 30 A current probe TCPA300 100 MHz probe amplier TCP404XL*
4
2 MHz, 500 A current probe TCPA400 50 MHz probe amplier ADA400A
*3Requires TCPA300 probe amplier.
4
Requires TCPA400 probe amplier.
*
100x, 10x, 1x, 0.1x high-gain differential amplier
Service Options
Option
Available at time of purchase
Description
Opt. CA1 Provides a single calibration event or coverage for the
designated calibration interval, whichever comes rst Opt. C3 Calibration Service - 3 years Opt. C5 Calibration Service - 5 years Opt. D1 Calibration Data Report Opt. D3 Calibration Data Report - 3 years (with Opt. C3) Opt. D5 Calibration Data Report - 5 years (with Opt. C5) Opt. R5 Repair Service - 5 year
Available after purchase
TDS30xxC-CA1 Provides a single calibration event or coverage for the
designated calibration interval, whichever comes rst TDS30xxC-R1PW TDS30xxC-R2PW
Repair service coverage 1 year post warranty
Repair service coverage 2 years post warranty TDS30xxC-R5DW Repair service coverage 5 years (includes product
warranty period); 5-year period starts at time of
customer instrument purchase
Product(s) are manufactured in ISO registered facilities.
Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats.
www.tektronix.com 7
Data Sheet
Contact Tektronix:
ASEAN / Australa
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Central East Eu
Mexico, Central/South America & Caribbean (52) 56 04 50 90
* European toll-free number. If not accessible, call: +41 52 675 3777
rope, Ukraine, and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Asia, and North Africa +41 52 675 3777
Middle Eas t,
The Netherlands 00800 2255 4835*
People’s Rep
Republic of
United Kingdom & Ireland 00800 2255 4835*
sia (65) 6356 3900
Austria 00800 2255 4835*
Belgium 00800 22
Brazil +55(11)37597600
Canada 1 800 833 9200
Denmark +4580881401
Finland +41526
France 008 00 2255 4835*
Germany 00800 2255 4835*
Hong Kong 400 8
India 000 800 650 1835
Italy 00800 2255 4835*
Japan 81 (3) 67
Luxembourg +41526753777
ublic of China 400 820 5835
Poland +41 52 675 3777
Korea 001 800 8255 2835
Russia & CIS +7 (495) 7484900
South Africa +41526753777
Spain 00800
Sweden 008 00 2255 4835*
Switzerland 00800 2255 4835*
Tai wa n 886 (
55 4835*
75 3777
20 5835
14 3010
Norway 800 16098
Portugal 80 08 12370
2255 4835*
2) 2722 9622
USA 1 800 833 9200
Updated 25 May 2010
www.tektronix.com
For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com
t © Tektronix, Inc. All rights r eserved. Tektronix products are covered by U.S. and foreign patents,
Copyrigh
d pending. Information in this publication supersedes that in all previously published material.
issued an
tion and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of
Specica
x, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks
Tek tro ni
espective companies.
of their r
29 Nov 2010 41W-12482-18
Loading...