Tektronix TDS3014B, TDS3032B, TDS3034B, TDS3052B, TDS3054B Read This First

...
TDS3000B Series Read This First
The following specification, performance verification, and information changes and additions have been made to the following TDS3000B Series instrument manuals.
TDS3000B User Manual Changes
Make the following changes to the TDS3000B series user manual:
1. In the General Safety Summary section, page viii, add the
following new heading and text at the bottom of the page:
This oscilloscope uses an LCD backlight lamp that contains mercury. Disposal may be regulated due to environmental considerations. For disposal or recycling information, please contact your local authorities or the Electronic Industries Alliance (www.eiae.org).
2. In Appendix A (Specifications), page A-6, Edge Trigger
Sensitivity, replace the specification with the following:
0.6 div from DC to 50 MHz, increasing to 1 div at oscillo­scope bandwidth.
3. In Appendix A (Specifications), page A-12, EMC Certifications
and Compliances, add the following new text below EN 61000-3-2 AC power line harmonic emissions:
EN 61000-3-3
Voltage changes, fluctuations and flicker
*P071171000*
071171000
TDS3000B Series Read This First
4. In Appendix A (Specifications), page A-13, EMC Certifications
and Compliances, footnote 3, replace the text with the following:
The increase in trace noise while subjected to a test field (3 V/m over the frequency range 80 MHz to 1 GHz, with 80% amplitude modulation at 1 kHz) is not to exceed 8 major divisions peak- to­peak. Ambient conducted fields may induce triggering when the trigger threshold is offset less than 4 major divisions from ground reference.
5. In Appendix E (Performance Verification), Check Channel
Edge-Trigger Sensitivity, page E-15, step 20, replace step 20 with the following steps:
20a. Set the output amplitude of the leveled sine wave generator so the peak--to--peak measurement is approximately
2.5 V. Note the generator output amplitude setting.
20b. Set the generator output amplitude to one fifth of the output amplitude that was set in step 20a.
TDS3000B Service Manual Changes
Make the following changes to the TDS3000B Series Service Manual (Tektronix part number 071-0972-03)
1. In the General Safety Summary section, page x, replace the
Components That Contain Mercury heading and text with the following:
This oscilloscope uses an LCD backlight lamp that contains mercury. Disposal may be regulated due to environmental considerations. For disposal or recycling information, please contact your local authorities or the Electronic Industries Alliance (www.eiae.org).
2. In Section 1 (Specifications), page 1 --5, Edge Trigger Sensitivity
specification, replace the specification with the following:
0.6 div from DC to 50 MHz, increasing to 1 div at oscillo­scope bandwidth.
www.tektronix.com
Copyright E Tektronix, Inc.