Tektronix TDS1002, TDS1012, TDS2002, TDS2012, TDS2014 User manual

...
Supplement
TDS1000 and TDS2000 Series User Manual Specifications Update
071-1858-01
www.tektronix.com
*P071185801*
071185801
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Informat ion in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Dri ve P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and techni cal support:
H In North America, call 1-800-833-9200. H Worldwide, visit www.tektronix.com to find contacts in your area.
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life
Handling
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. Production of this equipment required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. In order to avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
The symbol shown to the left indicates that this product complies with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
Mercury Notification. This product uses an LCD backlight lamp that contains mercury. Disposal may be regulated due to environmental considerations. Please contact your local authorities or, within the United States, the Electronics Industries Alliance (www.eiae.org) for disposal or recycling information.
Restriction of Hazardous
Substances
TDS1000 and TDS2000 Series User Manual Specifications Update
This product has been classified as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS Directive. This product is known to contain lead, cadmium, mercury, and hexavalent chromium.
1
Specifications Update
Specifications Update
The next table shows specifications for the new TDS1001 and TDS2004 models, and includes three corrections. These specifications supersede the corresponding specifications in the TDS1000 and TDS2000 series oscilloscope user manual.
The corrections are for the following specifications:
H Channel-to-Channel Common Mode Rejection, typical
H Trigger Sensitivity, Edge Trigger Type
H Not Equal Guardband
NOTE.AllVertical specifications are relative to a 1X probe. To modify displayed values, change the Probe option attenuation (default is 10X) to match the probe.
Updates for the TDS1000 and TDS2000 series oscilloscope specifications
Inputs
Channel-to-Channel TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014, TDS2022, TDS2024 Common Mode Rejection, typical
Channel-to-Channel TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014 TDS2022 ,TDS2024 Crosstalk
100:1 at 60 Hz 20:1at20MHz
Measured on MATH Ch1 -- Ch2 waveform, with test signal applied between signal and common of both channels, and with the same VOLTS/DIV and coupling settings on each channel
Measured on MATH Ch3 -- Ch4 waveform for 4-channel models
100:1 at 20 MHz
Measured on one channel, with test si gnal applied between signal and common of the other channel, and with the same VOLTS/DIV and coupling setti ngs on each channel
100:1 at 60 Hz 20:1at30MHz
100:1 at 30 MHz 100:1 at 50 MHz 100:1 at 100 MHz
100:1 at 60 Hz 10:1at50MHz
2
TDS1000 and TDS2000 Series User Manual Specifications Update
Specifications Update
g
gPea
(50s/divt
o
,
Updates for the TDS1000 and TDS2000 series oscilloscope specifications (cont.)
Vertical
Analog Bandwidth in TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014 TDS2022 ,TDS2024 Sample and Average modes at BNC or with P2220 probe, DC Coupled
Analog Bandwidth in
kDetect mode
5 s/div}), typical
Rise Time at BNC, TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014 TDS2022 ,TDS2024 typical
Horizontal
Sample Rate Range TDS2022, TDS2024 TDS1001, TDS1002, TDS1012, TDS2002, TDS2004, TDS2012,
SEC/DIV Range TDS2022, TDS2024 TDS1001, TDS1002, TDS1012, TDS2002, TDS2004, TDS2012,
* Bandwidth reduced to 6 MHz with a 1X probe.
40 MHz*{
20 MHz* (when vertical scale is set to < 5 mV)
TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014, TDS2022, TDS2024
30 MHz*{ 50 MHz*{ 75 MHz*{
20 MHz* (when vertical scale is set to < 5 mV)
<8.4ns <5.8ns <3.5ns <2.1ns
5S/sto2GS/s 5S/sto1GS/s
2.5ns/divto50s/div,ina1,2.5,5 sequence
60 MHz*{ 100 MHz*{ 200 MHz*{
0 °Cto+40°C(32°Fto104°F)
160 MHz*{ 0 °Cto+50°C(32°Fto122°F)
TDS2014
TDS2014
5 ns/div to 50 s/div, in a 1, 2.5, 5 sequence
{ When vertical scale is set to 5mV.
} The oscilloscope reverts to Sample mode when the SEC/DIV (horizontal scale) is set from 2.5 s/div to 5 ns/div on 1 GS/s
models, or from 2.5 s/div to 2.5 ns/div on 2 GS/s models. The Sample mode can still capture 10 ns glitches.
TDS1000 and TDS2000 Series User Manual Specifications Update
3
Specifications Update
Updates for the TDS1000 and TDS2000 series oscilloscope specifications (cont.)
Trigger
Trigger Sensitivity, Edge Trigger Type
Not Equal Guardband t > 330 ns: ±5% guardband < ±(5.1% + 16.5 ns)
* Bandwidth reduced to 6 MHz with a 1X probe.
Coupling Sensitivity TDS1001, TDS1002, TDS1012,
TDS2002, TDS2004, TDS2012, TDS2014
DC EXT 200 mV from DC to 100 MHz* 200 mV from DC to 100 MHz*
EXT/5 1 V from DC to 100 MHz* 1 V from DC to 100 MHz*
CH1, CH2, CH3, CH4
165 ns < t 330 ns: guardband = --16.5 ns/+33 ns
t 165 ns: guardband = ±16.5 ns
1 div from DC to 10 MHz*
1.5 div from 10 MHz to Full
Certifications and Compliances
TDS2022, TDS2024
350 mV from 100 MHz to 200 MHz*
1.75 V from 100 MHz to 200 MHz*
1 div from DC to 10 MHz*
1.5 div from 10 MHz to 100 MHz
2 div from 100 MHz to full
EN 61326. EMC requirements for Class A electrical equipment for measurement, control, and laboratory use. Annex D.
H IEC 61000--4--2. Electrostatic discharge immunity
H IEC 61000--4--3. RF electromagnetic field immunity
H IEC 61000--4--4. Electrical fast transient / burst immunity
H IEC 61000--4--5. Power line surge immunity
H IEC 61000--4--6. Conducted RF Immunity
H IEC 61000--4--11. Voltage dips and interruptions immunity
EN 61000- 3- 2. AC power line harmonic emissions
1
EN 61000- 3- 3. Voltage changes, fluctuations, and flicker
1
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
4
TDS1000 and TDS2000 Series User Manual Specifications Update
Specifications Update
Australia / New Zealand
Declaration of Conformity
-EMC
EMC Compliance
FCC Compliance
Russian Federation
Peoples Republic of China
EC Declaration of
Conformity - Low Voltage
Complies with EMC provision of Radiocommunications Act per these stan­dard(s):
H AS/NZS 2064.1/2. Industrial, Scientific, and Medical Equipment: 1992
Meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility when it is used with the product(s) stated in the specifications table. Refer to the EMC specification published for the stated products. May not meet the intent of the directive if used with other products.
Emissions comply with FCC 47 CFR, Part 15, Subpart B for Class A equipment.
This product was certified by the GOST ministry of Russia to be in compliance with all applicable EMC regulations.
This product has received the Chinese Metrology Certification. (CMC).
Compliance was demonstrated to the following specification as listed in the Official Journal of the European C ommunities:
U.S. Nationally
Recognized Testing
Laboratory Listing
Canadian Certification
Additional Compliance
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC.
H EN 61010-1:2001. Safety requirements for electrical equipment for
measurement control and laboratory use.
H EN 61010-2-031:2002. Particular requirements for handheld probe assem-
blies for electrical measurement and test equipment.
H UL 61010B--1:2004, 2nd Edition. Standard for electrical measuring and test
equipment.
H UL 61010B--2--031:2003. Particular requirements for handheld probe
assemblies for electrical measurement and test equipment.
H CAN/CSA C22.2 No. 1010.1:1997. Particular requirements for electrical
equipment for measurement, control, and laboratory use. Part 1.
H CAN/CSA C22.2 No. 61010--2--031:1994. Particular requirements for
handheld probe assemblies for electrical measurement and test equipment.
H IEC 61010--1:2001. Safety requirements for electrical equipment for
measurement, control, and laboratory use.
TDS1000 and TDS2000 Series User Manual Specifications Update
5
Specifications Update
H IEC 61010--031:2002. Particular requirements for handheld probe assemblies
for electrical measurement and test equipment.
Equipment Type
Pollution Degree
Descriptions
Test and measuring equipment.
A measure of the contaminates that could occur in the environment around and within a product. Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
H Pollution Degree 1. No pollution or only dry, nonconductive pollution
occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
H Pollution Degree 2. Normally only dry, nonconductive pollution occurs.
Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical office/home environment. Temporary condensation occurs only when the product is out of service.
H Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution
that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
H Pollution Degree 4. Pollution that generates persistent conductivity through
conductive dust, rain, or snow. Typical outdoor locations.
Pollution Degree
Installation (Overvoltage)
Category Descriptions
Overvoltage Category
6
Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor use only.
Terminals on this product may have different installation (overvoltage) category designations. The installation categories are:
H Measurement Category IV. For measurements performed at the source of
low-voltage installation.
H Measurement Category III. For measurements performed in the building
installation.
H Measurement Category II. For measurements performed on circuits directly
connected to the low-voltage installation.
H Measurement Category I. For measurements performed on circuits not
directly connected to MAINS.
Overvoltage Category II (as defined in IEC 61010-1).
TDS1000 and TDS2000 Series User Manual Specifications Update