Tektronix TDS1001C EDU, TDS1002C EDU, TDS1012C EDU, TDS2001C, TDS2002C Service Manual

...
xx
TDS2000C and TDS1000C-EDU Series
ZZZ
Digital Storage Oscilloscopes
Service Manual
Revision A
This document applies to TDS2000C and TDS1000C-EDU models with serial numbers of C030000 and above. This document a pplies to rmw are version 1.00 and above.
Warning
The servicing instructions are for use by qualied personnel only. To avoid personal injury, do not perform any servicing unless you are qualied to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
077-0446-02
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
OpenChoice is a registered trademark of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O . B ox 5 00 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tektronix.com to nd contacts in your area.
Table of Contents
General safety summary ........................ .................................. ................................ vi
Service safety summary.......................................................................................... viii
Preface .............................................................................................................. ix
Manual Conventions................................ ................................ .......................... ix
Related Documentation ...................................................................................... ix
Specications ........ .................................. ................................ ............................. 1
Signal Acquisition System Characteristics.. . .. .. . . . . . . . . . . . . . . . . . . . . . . .... .. .. .. . . . . . . . . . . . . .... .. .. .. . . . . . 1
Time Base System .. ................................ .................................. ......................... 7
Triggering System ...... .................................. ................................ ..................... 9
Display Specications............................ .................................. .......................... 13
Interfaces and Output Ports Specications................................................................. 14
Data Handling Characteristics............................... ................................ ................ 14
Power Distribution System .................................................................................. 14
Mechanical Characteristics .................................................................................. 15
Environmental Performance................................................................................. 15
Data Logging System Characteristics (Not Available on TDS1000C-EDU Models) ................. 15
Limit Testing System Characteristics (Not Available on TDS1000C-EDU Models) ... . . . .. . . .. . . . . . 16
Where to Find Operating Information........... .................................. .............................. 19
Theory of Operation................... .................................. ................................ .......... 21
System Level Block Diagrams ........................ .................................. .................... 22
Main Board .................................................................................................... 24
Acquisition System .. . .. .. .. . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . .. . . . 2 4
Processing and Display System .......... ................................ .............................. 25
Input Signal Interface................................ ................................ .................... 25
Probe Compensation......... ................................ .................................. .......... 25
External Trigger.......................................................................................... 25
Main Board Power....................................................................................... 25
Power Supply.................................................................................................. 25
Display Module ... .................................. .................................. ........................ 26
Front Panel .................. .................................. ................................ ................ 26
Two-Channel Oscilloscopes . .. .. . . .... .. .. .. . . . . . . .... .. .. .. . . . . .... .. .. .. . . . . .... .. .. .. . . . . .... .. .. .. . 26
Four-Channel Oscilloscopes .. .. .. . .. .. .. .. .. .. .. .. .. .. . .. .. .. .. .. .. .. .. .. .. .. .. .. . .. .. .. .. .. .. .. .. .. . .. 26
Performance Verication ....... ................................ .................................. ................ 27
Required Equipment............................ ................................ .............................. 27
Test Record.... .................................. ................................ .............................. 28
Performance Verication Procedures....... .................................. .............................. 29
Self Test................................................................................................... 29
Self Calibration........... .................................. .................................. ............ 29
Check DC Gain Accuracy............................................................................... 29
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual i
Table of Contents
Check Bandwidt
Check Sample Rate Accuracy and Delay Time Accuracy......................... .................. 32
Check Edge Trigger Sensitivity ........ ................................ ................................ 33
Check External Edge Trigger Sensitivity.......................... .................................. .. 35
Check Vertical Position Accuracy... .. .. .. .. .. . . .. . . . . .. . . . . . . .. . . . . .. . . . . .. . . . . .. . . . . .. . . . . .. . . . . . . .. . 36
Adjustment Procedures ........................................................................................... 41
Required Equipment.......................................................................................... 41
Adjustment Procedure........................................................................................ 41
Enable the Service Menu........................ .................................. ...................... 42
Adjustment Procedure ................... ................................ ................................ 44
Maintenance........................................................................................................ 49
Preventing ESD ............................................................................................... 49
Inspection and Cleaning................ ................................ .................................. .... 49
General Care.............................................................................................. 50
Interior Cleaning ......................................................................................... 50
Exterior Cleaning .......... .................................. ................................ ............ 50
Exterior Inspection......................................... .................................. ............ 51
Interior Inspection ..... ................................ .................................. ................ 51
Cleaning Procedure, Interior........................................ ................................ .... 52
Removal and Installation Procedures................. ................................ ...................... 54
Summary of Procedures............... .................................. ................................ 54
Required Tools ........................... .................................. .............................. 56
Rear Feet............ .................................. ................................ .................... 56
Flip Feet................................................................................................... 57
Front-Panel Knobs ....................................................................................... 57
Power Button ......................... ................................ .................................. .. 57
Rear Case ................................................................................................. 58
Front Feet ............................... .................................. ................................ 59
Power Supply Module............. ................................ .................................. .... 60
Internal Assembly.. .................................. .................................. .................. 61
Front-Panel Cable...................................... ................................ .................. 62
Main Board Module (Serial number C030000 or later) ............................................. 63
Display Module .......................................................................................... 66
Front-Panel Module...................................................................................... 67
Keypad .............. .................................. ................................ .................... 68
Troubleshooting......................... .................................. ................................ .... 69
Adjustment After Repair ................................................................................ 69
Required Tools and Equipment................. ................................ ........................ 69
Troubleshooting Tree.................................................................................... 69
Probe Comp Output...................................................................................... 74
Troubleshooting the Power Supply .................................................................... 75
h ........................................................................................ 31
ii TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table of Contents
Troubleshooti
Troubleshooting the Front Panel ....... ................................ ................................ 77
Troubleshooting the Main Board. ................................ .................................. .... 80
Repackaging Instructions ................ ................................ .................................. .. 83
Diagrams ........................................................................................................... 85
Replaceable Parts .................................................................................................. 87
Parts Order
Using the Replaceable Parts List............................................................................ 88
Parts Lists and Exploded Views............................................................................. 89
Appendix A: Example of a Vertical Position Accuracy Test Spreadsheet.. . .. . . . . . . . . . . . . . . . . . . .... .. .. .. . . 99
Sample Filled-In Vertical Position Accuracy Test Spreadsheet . .... .. .. .. . . . . .... .. .. .. . . . . .... .. .. .. . . . 99
ng the Display.................... .................................. ...................... 76
ing Information .................................................................................. 87
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual iii
Table of Contents
List of Figure
Figure 1: Module-level block diagram (2-channel)............................................................ 22
Figure 2: Mo
Figure 3: Example of a line graph for the Vertical Position Accuracy test . . . . . .. . . . . . . .. . . . . . . .. . . . . . . .. . . . 39
Figure 4: Adjustment setups ..................................................................................... 43
Figure 5: Locator for trim and cabinet removal (4-channel model shown) ................................. 55
Figure 6: Locator for internal modules (4-channel model shown)............................... ............ 55
Figure 7: Instrument cable locations (Main Board SN C030000 or later) .................................. 64
Figure 8:
Figure 9: Oscilloscope troubleshooting tree (page 2 of 4) . . . . . . . . . . .. . . . . . . . . .. . . . . . . . . .. . . . . . . . . . . .. . . . . . . . . 71
Figure 10: Oscilloscope troubleshooting tree (page 3 of 4)... . . . . . . . .... .. .. .. . . .... .. .. .. .... .. .. .. . . .... .. . 72
Figure 11: TDS2000C and TDS1000C-EDU series block diagram ......................................... 86
Figure 12: Exploded diagram, 2-channel models.............................................................. 89
Figure 13: Exploded diagram, 4-channel models.............................................................. 92
e 14: Power supply module, cables, and wires........................................................... 94
Figur
Figure 15: Exploded diagram, back case and trim............................................................. 96
dule-level block diagram (4-channel)............................................................ 23
Oscilloscope troubleshooting tree (page 1 of 4) .. .. .. . . .. . . . . . . . . .. . . . . . . . . .. . . . . . . . . .. . . . . . . . . . . .. 70
s
iv TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
List of Tables
Table 1: Signal acquisition system characteristics. .. .. .. .. .. . . . . .... .. .. .. . . . . .... .. .. .. . . . . . . .... .. .. .. . . . . ... 1
Table 2: Tim
Table 3: Triggering system ....................... ................................ ................................. 9
Table 4: Display spec ications................................................................................... 13
Table 5: Interfaces and output ports specications .. .................................. ........................ 14
Table 6: Data handling characteristics .......................................................................... 14
Table 7: Power distribution system.............................................................................. 14
Table 8: M
Table 9: Environmental performance ........................................................................... 15
Table 10: Data logging system characteristics (not available on TDS1000C-EDU models).............. 16
Table 11: Limit testing system characteristics (not available on TDS1000C-EDU models).. .. .. .. .. .. .. 16
Table 12: Performance verication.............................................................................. 27
Table 13: Test record.......... .................................. .................................. ................ 28
14: Required equipment ................................................................................... 41
Table
Table 15: Adjustment steps..................................... .................................. ................ 45
Table 16: External inspection checklist......................................................................... 51
Table 17: Internal inspection checklist.......................................................................... 52
Table 18: Tools and equipment .................................................................................. 69
Table 19: Probe Comp output.................. ................................ .................................. 74
ble 20: Troubleshooting the power supply................................................................... 75
Ta
Table 21: Troubleshooting the front panel – service........................... ................................ 77
Table 22: Troubleshooting the front panel – Service Diag.................................................... 77
Table 23: Troubleshooting the front panel – Mfg. test ...... .................................. ................ 77
Table 24: Expected signals at J202............ .................................. ................................ 78
Table 25: Troubleshooting the Main board ............... .................................. .................... 80
Table 26: List of error codes ..................................................................................... 81
Table 27: Replaceable parts list.................................................................................. 88
Table 28: Replaceable parts list, 2-channel models............................................................ 90
Table 29: Replaceable parts list, 4-channel models............................................................ 93
Table 30: Replaceable parts list; power supply module, instrument cables, and wires .................... 94
Table 31: Replaceable parts list, back case and trim .......................................................... 97
Table 32: Vertical position accuracy test spreadsheet . .. .. .... .. .. .. .... .. .. . . . . . . .... .. .. .. .... .. .. .. .... .. .. 99
e base system ................. .................................. .................................. ... 7
echanical characteristics ........... .................................. ................................ 15
Table of Contents
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual v
General safety summary
General safet
To avoid fir
e or personal
injury
ysummary
Review the fo this product or any products connected to it.
To avoid pot
Only qualied personnel should perform service procedures.
Use proper
certied for the country of use.
Connect a
instrument before connecting the probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output
nals of the product, ensure that the product is properly grounded.
termi
Observe all terminal ratings. To av o id re or shock hazard, observe all ratings
arkings on the product. Consult the product manual for further ratings
and m information before making connections to the product.
llowing safety precautions to avoid injury a nd prevent damage to
ential hazards, use this product only as s peci ed.
power cord. Use only the power cord specied for this product and
nd disconnect properly. Connect the probe output to the measurement
nect the probe reference lead to earth ground only.
Con
Do not apply a potential to any terminal, including the common terminal, that
eeds the maximum ratingofthatterminal.
exc
Power disconnect. The power switch disconnects the product from the power
urce. See instructions for the location. Do not block the pow er switch; it must
so remain accessible to the user at all times.
o not operate without covers. Do not operate this product with covers or panels
D
removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
vi TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
General safety summary
Terms in this manual
Symbols and terms on the
product
Do not operate i
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide prop
on installing the product so it has proper ventilation.
These terms may appear in this manual:
WAR N ING.
in injury or loss of life.
CAUTION
damage to this product or other property.
These t
erms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the m a
n wet/damp conditions.
er ventilation. Refer to the manual's installation instructions for details
Warning statements identify conditions or practices that could result
. Caution statements identify conditions or practices that could result in
rking.
WARNING indicates an injury hazard not immediately accessible as you
the marking.
read
CAUTION indicates a hazard to property including the product.
The following symbol(s) m ay appear on the product:
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual vii
Service s afety summary
Service safet
ysummary
Only qualifie safety summary and the General safety summary before performing any service procedures.
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
Use care when servicing with power on. Dangerous voltages or currents may exist in this p test leads before removing protective panels, soldering, or replacing components.
To avoi
d personnel should perform service procedures. Read this Service
roduct. Disconnect power, remove battery (if applicable), and disconnect
d electric shock, do not touch exposed connections.
viii TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Preface
Manual Conventions
This service manual provides information to verify performance of, calibrate, troubleshoot, disassemble, and replace parts on the TDS2000C and TDS1000C-ED
Unless noted otherwise, the term "oscilloscope" refers to all models in the TDS2000C an
U Series Digital Storage Oscilloscopes.
d TDS1000C-EDU series.
Modules
Replaceable Parts
Safety
Related Documentation
This manu before attempting service.
Throughout this manual, any replaceable component, assembly, o r part is referred to by the
A module is composed of electrical and mechanical assemblies, circuit cards, interc
This manual refers to any eld-replaceable assembly or mechanical part specically by its name or generically as a replaceable part. In general, a
aceable part is any circuit board or assembly, such as th e hard disk drive, or a
repl mechanical part, such as the I/O port connectors, that is listed in the replaceable parts list.
Sym at the beginning of this manual.
al uses certain conventions that you should become familiar with
term module.
onnecting cables, and user-accessible controls.
bolsandtermsrelatedtosafetyappearintheServiceSafetySummaryfound
To read about Use these documents
nstallation and
I Operation
Programmer Commands
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual ix
he oscilloscope user manual, available on the Tektronix Web
T site.Go to www.tektronix.com/manuals and search for the user
manual for your product (available in 11 languages).
The oscilloscope programmer manual, available on the Tektronix Web site. Go to www.tektronix.com/manuals and search for the
programmer manual for your product.
Preface
x TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Specications
These specications apply to all TDS2000C and TDS1000C-EDU series oscilloscopes. To verify that an oscilloscope meets specications, it must rst meet the foll
owing conditions:
The oscilloscope must have been operating continuously for twenty minutes within the s
pecied operating temperature.
You must perform the Do Self Cal operation, accessible through the Utility menu, if th
e operating temperature has changed by more than 5 ° C (9 °F)
since the last time the Do Self Cal operation was performed.
Specica
tions are provided in the following tables. All specications are guaranteed unless noted "typical." Specications that are m arked with the symbol are checked in the Performance Verication section. (See page 27, Performance Ver ication.)
Signal Acquisition System Characteristics
NOTE.
attenuation factors unless otherwise specied. This is due to the way the displayed sensitivity works. This does not affect actual methods of attachment. It only alters the relationship between displayed scale factors and the specications.
Table 1: Signal acquisition system characteristics
Characteristic
Number of Input
annels
Ch
Input Coupling DC, AC, or GND Input Impedance, DC
Coupled
Description
TDS1001C-EDU
S1002C-EDU
TD TDS1012C-EDU TDS2001C
S2002C
TD TDS2012C TDS2022C
TDS2004C
DS2014C
T TDS2024C
1MΩ ±2%inparallelwith20pF±3pF
All amplitude-related or modied specications require 1X probe
Two nominally identical channels, sampled simultaneously.
Four nominally identical channels, sampled simultaneously.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 1
Specications
Table 1: Signal acquisition system characteristics (cont.)
Characteristic
Maximum Input Voltage
Description
At the front panel connector, 300 V
, Installation Category II; derate at 20 dB/decade above 100 kHz to
RMS
13 V peak AC at 3 MHz and above. Based on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50 V offset
±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level up to 300 V. For nonsinusoidal waveforms, peak value must be less than 450 V. Excursions above 300 V should be less than 100 ms in duration, and the duty factor is limited to < 44%. The RMS signal level must be limited to 300 V. If these values are exceeded, damage to the instrument may result.
8 bits except at 2 mV/divNumber of Digitized Bits Displayed vertically with 25 digitization levels per division, 10 divisions dynamic range. 2 mV/div setting is generated by digital m ultiplication and the resolution is reduced. Given 100 levels
available, the resolution is >6.5 bits. Sensitivity Range 2 mV/div to 5 V/div in 1-2-5 sequence with the probe attenuation set to 1X. Display Gain Variable
The Display Gain Variable function allows the user to vary the vertical display gain continuously over the
full range.
Resolution is that of the coarse gain from which these data are constructed. Fine gain values (for
instance 120 mV/div) are acquired at the next higher coarse gain setting (in this case 200 mV/div).
The Display Gain Variable is achieved by digital mult iplication of the data to obtain the settings between
the 1-2-5 gain settings.
Hard copy outputs to printer will be exactly the same as seen on screen even when the Display Gain
Variable is used.
However, WAVEFORM DATA obtained through the I/O interface is limited to the 1-2-5 gain settings. Probe Scale Factors
1X, 10X, 20X, 50X, 100X, 200X, 500X, 1000X voltage attenuation.
5, 1, 500 m, 200 m, 20 m, 10 m, 1 m V/A current scale factor.
This adjusts the d isplay scale factor of the instrument to accommodate various probe types.
Accuracy of the probe used must be added to the accuracy specications of instrument.
No automatic probe interface is provided, so you must verify that the settings match the probe
characteristics. The probe check function allows setting of the proper attenuation for voltage probes. Acquisition Modes
Sample, Peak Detect, Average
Envelope mode not provided Retained Front Panel
Settings
Front panel settings are retained when the instrument power is turned off and on with the power switch.
The settings are retained when the line power is turned off and on.
The instrument periodically saves front panel settings after settings are changed. There is a delay of
three seconds after the last change and before the storage of the settings in memory.
2 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 1: S ignal acquisition system characteristics (cont.)
Specications
Characteristic
Math Modes
Voltage Measurement Functions
DC Gain Accuracy, Sample or Average Acquisition
DC Voltage Measurement Accuracy, Average Acquisition Mode
Delta Volts Measurement Accuracy, Average Acquisition Mode
Analog Bandwidth
Description
All Units: Channel 1 - Channel 2 Channel 2 - Channel 1 Channel 1 + Channel 2 Channel 1 * Channel 2 FFT
4 channel units (TDS20X4C): Channel 3 - Channel 4 Channel 4 - Channel 3 Channel 3 + Channel 4 Channel 3 * Channel 4
Mean, Cyc RMS, Peak-to-Peak, Max, Min, RMS, CursorRMS
This is the difference between the measured DC gai n and the nominal DC gain, divided by the nominal DC gain and expressed as a percent.
±3%, 5 V/div through 10 mV/div ±4%, 5 mV/div and 2 mV/div
Vertical position = 0: ±(3% of |reading| + 0.1 div + 1 mV) Vertical position 0 and vertical scale = 2 mV/div to 200 mV/div: ±[3% of |reading + vertical position|
+ 1% of |vertical position| + 0. 2 div + 7 mV] Vertical position 0 and vertical scale > 200 mV/div: ±[3% of |reading + vertical position| + 1% of
|vertical position| + 0.2 div + 175 mV]
Delta volts between any two averages of 16 waveforms acquired under the same setup and ambient conditions.
(3% of |reading| + 0.05 div)
Dened in Section 4.6 of IEEE std 1057. The difference between the upper and lower frequencies, at which the amplitude response, as seen in the data record, is 0.707 (-3 dB) of the response seen in the data record at the specied reference frequency. Species only the -3 dB point. It does not include the in-band response.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 3
Specications
Table 1: Signal acquisition system characteristics (cont.)
Characteristic
Analog Bandwidth, DC Coupled, Sample or Average
Analog Bandwidth, DC Coupled, Peak Detect, typical
Analog Bandwidth Selections
Description
This is analog bandwidth when the instrument is DC coupled in sample or average mode. V/div values
are accurate for probe attenuation settings of 1X. No probe should be installed for these measurements.
System bandwidth is type tested to be equivalent to this specication with the provided probe in 10X
mode. Use Section 4.6.1 of IEEE 1057, with the reference frequency of 1 kHz at an amplitude of
5 divisions, driven from a 50 source with external termination at the input BNC (25 effective source).
TDS2022C, TDS2024C:
to 200 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full with temperature between
0°C and 35 °C. DC to > 160 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full for
temperatures between 0 °C and 50 °C. < 5 mV/div settings are limited to 20 MHz bandwidth.
TDS2012C, TDS2014C:
DC to 100 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS2002C, TDS2004C:
DC to 70 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 MV/div settings
are limited to 20 MHz bandwidth.
TDS2001C:
DC to 50 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS1012C-EDU:
DC to 100 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS1002C-EDU:
DC to 60 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 MV/div settings
are limited to 20 MHz bandwidth.
TDS1001C-EDU:
DC to 40 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
This is the analog bandwidth when the instrument is DC coupled. V/div values are accurate for probe
attenuation settings of 1X. No probe should be installed for these measurements.
TDS1012C-EDU, TDS2012C, TDS2014C, TDS2022C, TDS2024C:
DC to 75 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. Settings less
than 5 mV /div are limited to 20 MHz bandwidth.
TDS1002C-EDU, TDS2002C, TDS2004C:
DC to 50 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. Settings less
than 5 mV /div are limited to 20 MHz bandwidth.
TDS1001C-EDU, TDS2001C:
DC to 30 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. Settings less
than 5 mV /div are limited to 20 MHz bandwidth.
20 MHz bandwidth limit ON/OFF
4 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 1: S ignal acquisition system characteristics (cont.)
Specications
Characteristic
Upper-Frequency Limit, 20 MHz Bandwidth Limited, typical
Lower- Frequency Limit, AC Coupled
Rise Time, typical
Peak Detect Mode Pulse Response
Vertical Position Ranges
Description
This is the upper frequency for Analog Bandwidth when the instrument has 20 MHz bandwidth limiting turned on.
20 MHz Bandwidth of all trigger paths are similarly limited, except the External Trigger, which is not affected
by BW Limit function. Each channel is separately limited, allowing different bandwidths on different channels of the same instrument.
This is the lower frequency for Analog Bandwidth when the instrument is AC-coupled
10 Hz. <1 Hz when 10X, passive probes are used.
Model
TDS2022C, TDS2024C TDS2012C, TDS2014C TDS2002C, TDS2004C TDS2001C TDS1012C-EDU TDS1002C-EDU TDS1001C-EDU Rise time is generally calculated from the followi ng formula: Rise time in ns = 350 / Bandwidth in MHz
This is the capability of the instrument to capture single event pulses using the Peak Detect Acquisition Mode.
The minimum single pulse widths for guaranteed 50% or greater amplitude capture are as follows:
Model
TDS1002C-EDU TDS1012C-EDU TDS2002C TDS2004C TDS2012C TDS2014C TDS2022C TDS2024C
TDS1001C-EDU TDS2001C
These are the ranges of the user-settable input offset voltage.
Volts/Div Setting
2 mV/div to 200 mV/div > 200 mV/div to 5 V/div
Expected full bandwidth rise time
2.1 ns
3.5 ns
5.0 ns
7.0 ns
3.5 ns
5.8 ns
8.8 ns
Sec/Div Setting
50 sec/div to 5 μS/div
50 sec/div to 5 μS/div
Position Range
±1.8V ±45V
Minimum Pulse Width
12 ns
13 ns
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 5
Specications
Table 1: Signal acquisition system characteristics (cont.)
Vertical Position Accuracy
Common Mode Rejection Ratio (CMRR), typical
Crosstalk (Channel Isolation)
This is the accuracy of the nominal voltage level represented by the code at the vendor of the A-D
converter's dynamic range.
Volts/Div Setting
2 mV/div to 200 mV/div ±(1% of |selected value| + 0.1 div + 5 mV) within the
> 200 mV/div to 5 V/div ±(1% of |selected value| + 0.1 div + 125 mV) within the
With the same signal applied to each channel, CMRR is the ratio of the acquired signal amplitude to the
amplitude of the M ATH difference waveform, either (Channel 1 - Channel 2), (Channel 2 - Channel 1),
(Channel 3 - Channel 4) or (Channel 4 - Channel 3)
Model
TDS1012C-EDU
TDS2012C
TDS2014C
TDS2022C
TDS2024C
TDS1001C-EDU
TDS1002C-EDU
TDS2001C
TDS2002C
TDS2004C
Section 4.11.1 of IEEE std. 1057. It is the r atio of the level of a signal input into one channel to that of
the same signal present in another channel due to stray coupling.
Model
TDS2022C
TDS2024C
TDS1012C-EDU
TDS2012C
TDS2014C
TDS1002C-EDU
TDS2002C
TDS2004C
TDS1001C-EDU
TDS2001C
Common Mode Rejection Ratio
100:1 at 60 Hz, reducing to 10:1 with 50 MHz sine wave, with equal Volts/Div and Coupling settings on each channel.
100:1 at 60 Hz, reducing to 20:1 with a sine wave with frequency equal to ½ the -3 dB bandwidth and with equal Volts//Div and Coupling settings on each channel.
Crosstalk
100:1 with a 100 MHz sine wave and with equal V/div settings on each channel
100:1 with a 50 MHz sine wave and with equal V/div settings on each channel
100:1 with a 30 MHz sine wave and with equal V/div settings on each channel
100:1 with a 20 MHz sine wave and with equal V/div settings on each channel
Position Accuracy
range ±1.8 V
range ±45 V
6 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Specications
Time Base Syst
em
Table 2: Time base system
Characteris
Sample-Rate
Recor
Long-Term Sample Rate and Horizontal
sition Time Accuracy
Po
Horizontal Position Time
nge
Ra
Horizontal Window Display
tic
Range
d Length
Description
This is the ra its inputs and stores the samples in memory to produce a record of time-sequential samples. (IEEE 1057, 2.2.1)
Model
TDS2012C TDS2014C TDS2022C TDS2024C
TDS1001C-EDU TDS1002C TDS1012C-EDU TDS2001C TDS2002 TDS2004C
(Sin x)/x interpolationWaveform Interpolation Waveform interpolation is activated for sweep speeds of 100 ns/div and faster.
This is the total number of samples contained in a single acquired waveform record (Memory Length in IEEE 1057.2.2.1).
2,500 samples per record. 5 n s/div to 50 s/div in 1–2.5–5 sequenceSeconds/Division Range Sec/Div Variable function is not available for this product.
s is the maximum, total, long-term error in sample-rate or horizontal position time accuracy, expressed
Thi in parts per million.
0 ppm over any 1 ms interval.
±5
Horizontal scale setting Horizontal position time range
5ns/divto10ns/div
5 ns/div to 100 μs/div
2 250 μs/divto10s/div 25 s/div to 50 s/div The user controls the time from the trigger to the center graticule on the display with the Horizontal
Position knob. The resolution of the Horizontal Position time is 1/25 of a horizontal division. The Window Zone enables a user to select a part of the display to be magnied. The user controls the
location with the Horizontal Position knob and the width with the Sec/Div knob. Vertical cursors show the selected zone. To magnify the zone, the user selects the window from the menu.
nge of real-time rates, expressed in samples/second, at which a digitizer samples signals at
Sample-rate range
5 S/s to 2000 MS/s
5 S/s to 1000 MS/s
-EDU
C
-4 div * s/div to 20 ms 4 div * s/div to 50 ms
-
-4 div * s/div to 50 s
-4 div * s/div to 250 s
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 7
Specications
Table2: Timebasesystem(cont.)
Characteristic
Delta Time Measurement Accuracy
Time Measurement Functions
Description
This is the accuracy of delta time measurements made on any single waveform.
The limits are given in the following table for signals having an amplitude 5 divisions, a slew rate at the measurement points of 2.0 divisions/ns, and acquired 10 mV/div.
Condition
Single shot, sample mode, full bandwidth selected
> 16 averages, full bandwidth selected
The Sample Interval is the time between the samples in the waveform record. Frequency, Period, Rise Time, Fall Time, Positive Width, negative Width, Positive Duty Cycle, Phase,
Delay
Time Measurement Accuracy
±(1 Sample Internal + 100 ppm * |reading| + 0.6 ns)
±(1 Sample Internal + 100 ppm * |reading| + 0.4 ns)
8 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Specications
Triggering Sy
stem
Table 3: Triggering system
Characteris
Trigger Types Edge, Video, Pulse Width Trigger Source
Selection
Horizontal Trigger Posit
Trigg
External Trigger Input Impedance
External Trigger Maximum Input Voltage
ine Trigger
L Characteristics
Edge Trigger
tic
ion er Holdoff Range
Trigger Modes Auto, Normal
Description
Models Trigger source selections TDS1001C-EDU
TDS1002C-EDU TDS1012C­TDS2001C TDS2002C TDS2012C TDS2022C
TDS2004C TDS2014C TDS2024C
External/5 selection attenuates the external signal by 5. When Bandwidth Limit is selected for a channel, the bandwidth of that channel’s trigger path will also be
d. The bandwidth of the External Trigger path is not affected by the bandwidth limit.
limite The trigger position is set by the Horizontal Position knob.
500 ns minimum to 10 s maximum The ability to set large values of Holdoff is limited by the difculty in adjusting the Holdoff at Sec/Div
ings less than 100 ms/Div. This is because Holdoff c annot be set in Scan Mode, which begins at
sett 100 ms/div when Trigger Mode is AUTO. By adjusting Trigger Mode to NORMAL, the Scan Mode operation is turned off, and Holdoff can be adjusted at larger seconds/Div settings.
1M±2%inparallelwith20pF±3pF
300 V an
Ba ±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level up to 300 V. For nonsinusoidal waveforms, peak value must be less than 450 V. Excursions above 300 V s hould be less
han 100 ms duration and the duty factor is limited to < 44%. R MS signal level must be limited to 300 V. If
t these values are exceeded, damage to the instrument may result.
Line Trigger mode provides a source to synchronize the trigger with the AC line input. Input Amplitude requirements: 85 V Input Frequency requirements: 45 Hz - 440 Hz.
, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz
RMS
d above sed on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50 V offset
EDU
Channel 1 Channel 2 External External/5 AC Line
Channel 1 Channel 2 Channel 3 Channel 4 External
l/5
Externa AC Line
- 265 VAC.
AC
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 9
Specications
Table 3: Triggering system (cont.)
Characteristic Description
AC, DC, Noise Reject, High Frequency Reject, Low Frequency RejectTrigger Coupling The External Trigger path does not have a DC blocking capacitor ahead of the trigger input circuit. The
rolloff associated with AC coupling happens after the input circuit. When attempting to trigger on an AC signal that has a DC offset, care must be used to avoid overloading the input of the External Trigger circuit. For signals that have a large DC offset, using Channel 1 or Channel 2 (and Channel 3 or Channel 4 for the TDS20X4B) with AC coupling is preferred.
Trigger Slope Sensitivity,
Edge-TypeTrigger, DC Coupled
Rising Edge, Falling Edge
Measurement Style A) The minimum signal levels for achieving stable frequency indication on the Trigger Frequency C ounter within 1% of correct indication.
Measurement Style B) Section 4 10.2 in IEEE Std. #1057. The minimum signal levels required for stable edge triggering of an acquisition when the trigger Source is DC coupled.
Trigger Source Sensitivity (Measurement
Channel Inputs
Ext
Ext/5
Trigger Frequency Readout typically stabilizes at 50% more signal than generates a stable visual display.
All products
TDS1001C-EDU
TDS1002C-EDU
TDS2001C
TDS2002C TDS2004C
TDS1012C-EDU TDS2012C TDS2014C
TDS2022C TDS2024C
All products
TDS2022C TDS2024C
All products TDS2022C
TDS2024C
Sensitivity (Measurement
style A ), typical
1.5 div from DC to 10 MHz ( > 2 mV/div)
4 div from DC to 10 MHz (2 mV/Div)
3 div between 10 MHz and 40 MHz
3 div between 10 MHz and 60 MHz
3 div between 10 MHz and 50 MHz
3 div between 10 MHz and 70 MHz
3 div between 10 MHz and 100 MHz
3 div between 10 MHz and 200 MHz
300 mV from DC to 100 MHz
500 mV from 100 MHz to 200 MHz
1.5 Vfrom DC to100 MHz 1 V from DC to 100 MHz
2.5 V from 100 MHz to 200 MHz
style B)
1 div from DC to 10 MHz > 2mV/div)
2.5 div from DC to 10 MHz (2 mV/Div)
1.5 div between 10 MHz and 40 MHz
1.5 div between 10 MHz and 60 MHz
1.5 div between 10 MHz and 50 MHz
1.5 div between 10 MHz and 70 MHz
1.5 div between 10 MHz and 100 MHz
1.5 div from 10 MHz to 100 MHz
2.0 div above 100 MHz to 200 MHz
200 mV from DC to 100 MHz
350 mV from 100 MHz to 200 MHz
1.75 V from 100 MHz to 200 MHz
10 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 3: Triggering system (cont.)
Characteristic Description
Sensitivity, Edge-Type Trigger, non-DC Coupled, typical
for Successful Operation of “Set Level to 50%” Function, typical
Trigger Level Ranges, typical
Trigger Level Accuracy, DC Coupled, typical
Video Trigger
Default Settings for Video Trigger
Video Trigger Source Selection
Video Trigger Polarity Selection
Video Sync Selection
Formats and Field Rates
Trigger Source Sensitivity
AC Same as DC Coupled limits for frequencies 50 Hz and
Noise Rej
HF Rej LF Ref Same as DC Coupled limits for frequencies above 300 kHz. Since AC coupling is not done in the front end, use of a 10M probe does not affect the low frequency
corner. 50 Hz.Lowest Frequency Using a 10M probe will not affect the operation of this function.
Input Channel ±8 divisions from center screen
Ext ±1.6 V Ext/5: The settable resolution for Trigger Level is 0.02 division for an input channel source, 4 mV for Ext source,
and 20 mV for Ext/5 source.
This is the amount of deviation allowed between the level on the waveform at which triggering occurs and the level selected for DC-coupled triggering signals
±(0.2 d iv + 5 mV) for signals within ±4 divisions from the center screen, having rise and fall times of 20 ns.
Ext: ±(6% of setting + 40 mV) for signals less than ±800 mV Ext/5: ±(6% of setting + 200 mV) for signals less than ±4 V
Trigger Mode: Auto Trigger Coupling: AC Same as Trigger Source Selections listed above except Line Trigger. Line Trigger source is meaningless
in this mode. Normal (Negative going Sync Signal), Invert (Positive going Sync Signal)
Line, Line #, Odd Field, Even F ield, Field: PAL/SECAM, NTSC formats
Field rates: 50 Hz to 60 Hz.Video Trigger Line rates: 15 kHz to 20 kHz (NTSC, PAL, SECAM)
Specications
above Effective in Sample or Average Mode, > 10 mV/div to
5 V/div. Reduces DC Coupled trigger sensitivity by 2X. Same as DC Coupled limits from DC to 7 kHz.
±8 V
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 11
Specications
Table 3: Triggering system (cont.)
Characteristic Description
Video Trigger Sensitivity, typical
Pulse-Width Trigger
Pulse-Width Trigger Modes
Pulse Width Trigger Edge
Pulse Width Range 33 ns width 10 seconds Pulse Width
Resolution Equal Guardband
Not Equal Guardband
Pulse-Width Trigger Point
Trigger Frequency Counter
Trigger Frequency Counter
This is the minimum peak–to–peak video signal required for stable Video-Type triggering.
Source
Input Channels 2 divisions of composite video Ext Ext/5 2 V of composite video
< (Less than), > (Greater than), = (Equal), (Not equal)
Falling edge for positive polarity pulse. Rising edge for negative polarity pulse.
16.5 ns or 1 part per thousand, whichever is larger
t > 330 ns: ±5% < guardband < ±(5.1% + 16.5 ns) t 330 ns: guardband = ±16.5 ns. All pulses, even from the most stable sources, have some amount of jitter. To avoid disqualifying pulses
that are intended to qualify but are not absolutely correct values, we provide an arbitrary guardband. Any measured pulse width within the guardband will qualify. If you are looking for pulse width differences that are smaller than the guardband width, offsetting the center should allow discriminating differences down to the guardband accuracy.
330 ns < 1: ±5% guardband < ±(5.1% + 16.5 ns) 165 ns < 1 < 330 ns: guardband = -16.5 ns/+33 ns t 165 ns: guardband = ±16.5 ns All pulses, even from the most stable sources, have some amount of jitter. To avoid disqualifying pulses
that are intended to qualify but are not absolutely correct values, we provide an arbitrary guardband. Any measured pulse width outside the guardband will qualify. If you are looking for pulse width differences that are smaller than the guardband width, offsetting the center should allow discriminating differences down to the guardband accuracy. Not equal has slightly better ability to deal with small pulse widths than equal. The accuracy is not better.
Equal: The oscilloscope triggers when the trailing edge of the pulse crosses the trigger level. Not Equal: If the pulse is narrower than the specied width, the trigger point is the trailing edge. Otherwise,
the oscilloscope triggers when a pulse continues longer than the time specied as the Pulse Width. Less than: The trigger point is the trailing edge. Greater than (also called the time out trigger): The oscilloscope triggers when a pulse continues longer
than the time specied as the Pulse Width.
This provides a highly accurate means of identifying the frequency of trigger signals. Since averaging takes place over a longer time span, the number of stable digits is improved over the automatic measurement of the same type.
Typical sensitivity
400 mV of composite video
12 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 3: Triggering system (cont.)
Specications
Frequency Counter Resolution
Frequency Counter Accuracy, typical
Frequency Counter Frequency Range, typical
Counter Signal Source
6 digits
±51 ppm including all reference errors and ±1 count errors.
AC coupled. 10 Hz minimum to rated bandwidth.
Pulse width or edge sele cted trigger source.Frequency Frequency counter measures selected trigger source at all times in pulse width and edge mode, including
when oscilloscope acquisition is halted due to changes in run status, or acquisition of a single shot event has completed.
Frequency counter does NOT measure pulses that do not qualify as legitimate trigger events. Pulse Width mode: Counts pulses of sufcient magnitude inside the 250 ms measurement window
that qualify as triggerable events (for example, all narrow pulses in a PWM pulse train if set to < mode and the limit is set to a relatively small number).
Edge Trigger mode: Counts all pulses of sufcient magnitude.
Display Specications
Table 4: Display specications
Characteristic Description
Display Type
Display Resolution
Brightness, typical
Contrast Ratio and Control, typical
11.5 cm (width) * 8.64 cm (height), 14.3 cm diagonal (5.7”), QVGA, active TFT color liquid crystal display (LCD) with color characters/waveforms on a black background. Surface antiglare (3H) treatment.
This is the number of individually addressable pixels
320 horizontal by 240 vertical pixels The video display contains both the character and waveform displays.
This is the light output of the backlight.
400 cd/m Available black room contrast ratio, full black to full white. 400 minimum, 500 typical.
2
, typical. 320 cd/m2min.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 13
Specications
Interfaces an
d Output Ports Specications
Table 5: Interfaces and output ports specications
Characteris
USB Device USB 2.0 Full S
USB H ost USB 2.0 Full Speed host. 12 Mb/sec maximum. Supports USB Mass Storage Class. Bulk Only Subclass
USB Host Current Provides full 0.5 A of 5 V.
GPIB Interface GPIB access via TEK-USB-488 accessory.
Output V Frequency, typical
Data Ha
tic
oltage and
Description
USB-TMC communications with Tektronix extensions. Standard
only. Prov Standard
Standard
Output voltage: 5.0 V ±10% into 1 Mload.Probe Compensator, Frequency: 1 kHz
ides full 0.5 A of 5 V.
ndling Characteristics
peed device. 12 Mb/second maximum. Supports PICTBRIDGE compatibility and provides
Table 6: Data handling characteristics
cteristic
Chara
ntion of Front
Rete Panel Settings
Stored Waveforms and Multiple Front Panel
tings
Set
iption
Descr
Front panel settings are stored periodically in memory. The settings are not lost when the instrument is
ed off or if there is a power failure.
turn Two Channel 1, Channel 2, or Math waveforms can be stored in nonvolatile waveform memory A or
B. One, both, or neither of A or B waveform memories can be displayed. For the TDS20X4C, four
eforms, A, B, C, D, or Math can be stored. Ten user setups of the current instrument settings can be
wav saved and restored from nonvolatile memory. Additional storage is available when an appropriate mass storage device is connected via USB.
Power Distribution System
ble 7: Power distribution system
Ta
Characteristic Description
ess than 30 W a t 85 to 275 V
Power Consumption Source Voltage Full Range: 100 to 240 V Source Frequency 360 Hz to 440 Hz from 100 VACto 120 VAC.
Fuse Rating
L
ACRMS
45 Hz to 66 Hz from 100 V
3.15 Amps, T rating, 250 V; IEC and UL approved.
input.
AC
± 10%, Installation Category II (Covers range of 90 to 264 V
to 240 VAC.
AC
AC
14 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Specications
Mechanical Ch
aracteristics
Table 8: Mechanical characteristics
Characteris
Weight
Size
Cooling
tic
Method
Description
Requirement
2.0 k g (4.4 lbs), stand-alone instrument
2.2 kg (4.9 lbs), with accessories
3.6kg(8lbs Height Width Depth Convect
s that follow are nominal:
), when packaged for domestic shipment
ion cooled
Environmental Performance
Table 9: Environmental performance
Characteristic Description
Temperature
Humidity
Altitude
Operating 0° C to +50° C (32 °F to 122 °F), with 5° C/minute maximum gradient,
Nonoperating
ating
Oper
Nonoperating
erating
Op Nonoperating
158 mm (6.2
326.3 mm (1
124.1 mm (
nonco
-40° 5% to 5% to 45% RH above +40° C up to +50° C, noncondensing, and as limited
by a Maximum Wet-Bulb Temperature of +37° C (derates relative humidity
5 % RH at +50° C)
to 4
o 85% relative humidity (% RH) at up to +40° C
5% t 5% to 45% RH above +40° C up to + 50° C, noncondensing. Above +50° C limited by a Maximum Wet-Bulb Temperature of +37° C
erates relative humidity to 12% RH at +71° C)
(d
to 3000 meters (10,000 feet)
Up
p to 3000 meters (10,000 feet).
U Altitude is limited by possible damage to the L CD at higher altitudes. This
damage is independent of operation
2in)
2.85 in)
4.88 in)
ndensing, up to 3000 m altitude
C to + 71° C (-40 °F to 159.8 °F), with 5° C/minute maximum gradient
85% relative humidity (% R H) at up to +40° C
Data Logging System Characteristics (Not Available on T DS1000C-EDU M odels)
NOTE. This software feature directs the oscilloscope to autom atically collect
data over a period of time. After you congure the trigger conditions to use, you can use the data logging menu to set up the oscilloscope so that it will save all of the triggered waveform to a USB memory device, within a time duration that you have set.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 15
Specications
Table 10: Data l
Characteristic Description
Duration The time perio
Source The signal source which you want to save the waveform.
Select Fold
ogging system characteristics (not available on TDS1000C-EDU models)
d.
30 min, 1 hour, 1.5 hour, 2 hour, 2.5 hour, 3 hour, 3.5 hour, 4 hour, 4.5 hour, 5 hour, 5.5 hour, 6 hour, 6.5 hour, 7 hour, 7.5 hour, 8 hour.
Channel 1, C
er
The le fold You can create the new folder or change the existing folder as the folder where you want to save the
waveform data.
hannel 2, Channel 3, Channel 4, Math
er where you save the waveform data.
Limit Testing System Characteristics (Not Available on TDS1000C-EDU Models)
NOTE. This software feature directs the oscilloscope to monitor an active input
signal against a template and to output pass or fail results by judging whether the input signal is within the bounds of the template.
Table 11: Limit testing system chara cteristics (not available on TDS1000C-EDU models)
teristic
Charac
e
Sourc
re Ref Channel
Compa
Run/Stop To enable or disable the limit testing function.
Template Setup You can use this menu item to set up a limit test waveform template. The template is the mask signal that
Source The location of the signal source that is used to create the limit test template.
Vertical Limit The vertical limit in vertical divisions.
Horizontal Limit The horizontal limit in horizontal divisions.
Destination Ref Channel The location of the reference memory location that is used to store the limit test template.
Display Template Displays or does not display a stored test template.
ption
Descri
The signal source which you want to do the limit testing.
el 1, Channel 2, Channel 3, Channel 4, Math
Chann
ference channel # where the template is saved.
The re
RefA, RefB, RefC, RefD. The limit testing system will compare the source signal with this template.
Run, Stop.
dene as the boundary to compare with the input source signal. You can create the template from
you internal or external waveforms with specic horizontal and vertical tolerances.
Channel 1, Channel 2, Channel 3, Channel 4, Math
0~1000 mdiv
0~500 mdiv.
RefA, RefB, RefC, RefD.
On, Off.
16 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 11: Limit testing system characteristics (not available on TDS1000C-EDU models) (cont.)
Characteristic Description
Action on Violation
Stop After Denes the conditions that will cause the oscilloscope to end limit testing.
Denes the actions the o scilloscope will take after a violation is detected.
Save Image: The oscilloscope will automatically save a screen image when a violation is detected. Save Waveform: The oscilloscope will automatically save a digital copy of the source waveform when
a violation is detected.
Manual: Lets you stop the test by toggling the “Run/stop” choice. Waveforms: Lets you set the numbers of waveforms to test before stopping limit testing. Violations: Lets you set the numbers of violations to detect before stopping limit testing. Elapsed time: Lets you set the elapsed test time in seconds to pass before stopping limit testing.
Specications
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 17
Specications
18 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Loading...
+ 84 hidden pages