This document applies to TDS2000C and TDS1000C-EDU
models with serial numbers of C030000 and above.
This document a pplies to firmw are version 1.00 and above.
Warning
The servicing instructions are for use by qualified personnel
only. To avoid personal injury, do not perform any servicing
unless you are qualified to do so. Refer to all safety summaries
prior to performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
OpenChoice is a registered trademark of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O . B ox 5 00
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to find contacts in your area.
Table of Contents
General safety summary ........................ .................................. ................................vi
Service safety summary..........................................................................................viii
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manualv
General safety summary
General safet
To avoid fir
e or personal
injury
ysummary
Review the fo
this product or any products connected to it.
To avoid pot
Only qualified personnel should perform service procedures.
Use proper
certified for the country of use.
Connect a
instrument before connecting the probe to the circuit under test. Connect the
probe reference lead to the circuit under test before connecting the probe input.
Disconnect the probe input and the probe reference lead from the circuit under test
before disconnecting the probe from the measurement instrument.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
nals of the product, ensure that the product is properly grounded.
termi
Observe all terminal ratings. To av o id fire or shock hazard, observe all ratings
arkings on the product. Consult the product manual for further ratings
and m
information before making connections to the product.
llowing safety precautions to avoid injury a nd prevent damage to
ential hazards, use this product only as s peci fied.
power cord. Use only the power cord specified for this product and
nd disconnect properly. Connect the probe output to the measurement
nect the probe reference lead to earth ground only.
Con
Do not apply a potential to any terminal, including the common terminal, that
eeds the maximum ratingofthatterminal.
exc
Power disconnect. The power switch disconnects the product from the power
urce. See instructions for the location. Do not block the pow er switch; it must
so
remain accessible to the user at all times.
o not operate without covers. Do not operate this product with covers or panels
D
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
viTDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
General safety summary
Terms in this manual
Symbols and terms on the
product
Do not operate i
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide prop
on installing the product so it has proper ventilation.
These terms may appear in this manual:
WAR N ING.
in injury or loss of life.
CAUTION
damage to this product or other property.
These t
erms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read
the m a
n wet/damp conditions.
er ventilation. Refer to the manual's installation instructions for details
Warning statements identify conditions or practices that could result
. Caution statements identify conditions or practices that could result in
rking.
WARNING indicates an injury hazard not immediately accessible as you
the marking.
read
CAUTION indicates a hazard to property including the product.
The following symbol(s) m ay appear on the product:
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manualvii
Service s afety summary
Service safet
ysummary
Only qualifiesafety summary and the General safety summary before performing any service
procedures.
Do not service alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering fi rst aid and resuscitation is
present.
Disconnect power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
Use care when servicing with power on. Dangerous voltages or currents may exist
in this p
test leads before removing protective panels, soldering, or replacing components.
To avoi
d personnel should perform service procedures. Read this Service
roduct. Disconnect power, remove battery (if applicable), and disconnect
d electric shock, do not touch exposed connections.
viiiTDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Preface
Manual Conventions
This service manual provides information to verify performance of,
calibrate, troubleshoot, disassemble, and replace parts on the TDS2000C and
TDS1000C-ED
Unless noted otherwise, the term "oscilloscope" refers to all models in the
TDS2000C an
U Series Digital Storage Oscilloscopes.
d TDS1000C-EDU series.
Modules
Replaceable Parts
Safety
Related Documentation
This manu
before attempting service.
Throughout this manual, any replaceable component, assembly, o r part is referred
to by the
A module is composed of electrical and mechanical assemblies, circuit cards,
interc
This manual refers to any field-replaceable assembly or mechanical part
specifically by its name or generically as a replaceable part. In general, a
aceable part is any circuit board or assembly, such as th e hard disk drive, or a
repl
mechanical part, such as the I/O port connectors, that is listed in the replaceable
parts list.
Sym
at the beginning of this manual.
al uses certain conventions that you should become familiar with
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manualix
he oscilloscope user manual, available on the Tektronix Web
T
site.Go to www.tektronix.com/manuals and search for the user
manual for your product (available in 11 languages).
The oscilloscope programmer manual, available on the Tektronix
Web site. Go to www.tektronix.com/manuals and search for the
programmer manual for your product.
Preface
xTDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Specifications
These specifications apply to all TDS2000C and TDS1000C-EDU series
oscilloscopes. To verify that an oscilloscope meets specifications, it must first
meet the foll
owing conditions:
The oscilloscope must have been operating continuously for twenty minutes
within the s
pecified operating temperature.
You must perform the Do Self Cal operation, accessible through the Utility
menu, if th
e operating temperature has changed by more than 5 ° C (9 °F)
since the last time the Do Self Cal operation was performed.
Specifica
tions are provided in the following tables. All specifications are
guaranteed unless noted "typical." Specifications that are m arked with the symbol
are checked in the Performance Verification section. (See page 27, PerformanceVer ification.)
Signal Acquisition System Characteristics
NOTE.
attenuation factors unless otherwise specified. This is due to the way the displayed
sensitivity works. This does not affect actual methods of attachment. It only alters
the relationship between displayed scale factors and the specifications.
Table 1: Signal acquisition system characteristics
Characteristic
Number of Input
annels
Ch
Input CouplingDC, AC, or GND
Input Impedance, DC
Coupled
Description
TDS1001C-EDU
S1002C-EDU
TD
TDS1012C-EDU
TDS2001C
S2002C
TD
TDS2012C
TDS2022C
TDS2004C
DS2014C
T
TDS2024C
1MΩ ±2%inparallelwith20pF±3pF
All amplitude-related or modified specifications require 1X probe
Two nominally identical channels, sampled
simultaneously.
Four nominally identical channels, sampled
simultaneously.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual1
Specifications
Table 1: Signal acquisition system characteristics (cont.)
Characteristic
Maximum Input Voltage
Description
At the front panel connector, 300 V
, Installation Category II; derate at 20 dB/decade above 100 kHz to
RMS
13 V peak AC at 3 MHz and above.
Based on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50 V offset
±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level up to 300 V. For
nonsinusoidal waveforms, peak value must be less than 450 V. Excursions above 300 V should be less
than 100 ms in duration, and the duty factor is limited to < 44%. The RMS signal level must be limited to
300 V. If these values are exceeded, damage to the instrument may result.
8 bits except at 2 mV/divNumber of Digitized Bits
Displayed vertically with 25 digitization levels per division, 10 divisions dynamic range.
2 mV/div setting is generated by digital m ultiplication and the resolution is reduced. Given 100 levels
available, the resolution is >6.5 bits.
Sensitivity Range2 mV/div to 5 V/div in 1-2-5 sequence with the probe attenuation set to 1X.
Display Gain Variable
The Display Gain Variable function allows the user to vary the vertical display gain continuously over the
full range.
Resolution is that of the coarse gain from which these data are constructed. Fine gain values (for
instance 120 mV/div) are acquired at the next higher coarse gain setting (in this case 200 mV/div).
The Display Gain Variable is achieved by digital mult iplication of the data to obtain the settings between
the 1-2-5 gain settings.
Hard copy outputs to printer will be exactly the same as seen on screen even when the Display Gain
Variable is used.
However, WAVEFORM DATA obtained through the I/O interface is limited to the 1-2-5 gain settings.
Probe Scale Factors
1X, 10X, 20X, 50X, 100X, 200X, 500X, 1000X voltage attenuation.
5, 1, 500 m, 200 m, 20 m, 10 m, 1 m V/A current scale factor.
This adjusts the d isplay scale factor of the instrument to accommodate various probe types.
Accuracy of the probe used must be added to the accuracy specifications of instrument.
No automatic probe interface is provided, so you must verify that the settings match the probe
characteristics. The probe check function allows setting of the proper attenuation for voltage probes.
Acquisition Modes
Sample, Peak Detect, Average
Envelope mode not provided
Retained Front Panel
Settings
Front panel settings are retained when the instrument power is turned off and on with the power switch.
The settings are retained when the line power is turned off and on.
The instrument periodically saves front panel settings after settings are changed. There is a delay of
three seconds after the last change and before the storage of the settings in memory.
2TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 1: S ignal acquisition system characteristics (cont.)
Specifications
Characteristic
Math Modes
Voltage Measurement
Functions
√ DC Gain Accuracy,
Sample or Average
Acquisition
DC Voltage
Measurement Accuracy,
Average Acquisition
Mode
Delta Volts Measurement
Accuracy, Average
Acquisition Mode
This is the difference between the measured DC gai n and the nominal DC gain, divided by the nominal
DC gain and expressed as a percent.
±3%, 5 V/div through 10 mV/div
±4%, 5 mV/div and 2 mV/div
Vertical position = 0: ±(3% of |reading| + 0.1 div + 1 mV)
Vertical position ≠ 0 and vertical scale = 2 mV/div to 200 mV/div: ±[3% of |reading + vertical position|
+ 1% of |vertical position| + 0. 2 div + 7 mV]
Vertical position ≠ 0 and vertical scale > 200 mV/div: ±[3% of |reading + vertical position| + 1% of
|vertical position| + 0.2 div + 175 mV]
Delta volts between any two averages of 16 waveforms acquired under the same setup and ambient
conditions.
(3% of |reading| + 0.05 div)
Defined in Section 4.6 of IEEE std 1057. The difference between the upper and lower frequencies, at
which the amplitude response, as seen in the data record, is 0.707 (-3 dB) of the response seen in the
data record at the specified reference frequency. Specifies only the -3 dB point. It does not include
the in-band response.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual3
Specifications
Table 1: Signal acquisition system characteristics (cont.)
Characteristic
√ Analog Bandwidth,
DC Coupled, Sample or
Average
Analog Bandwidth, DC
Coupled, Peak Detect,
typical
Analog Bandwidth
Selections
Description
This is analog bandwidth when the instrument is DC coupled in sample or average mode. V/div values
are accurate for probe attenuation settings of 1X. No probe should be installed for these measurements.
System bandwidth is type tested to be equivalent to this specification with the provided probe in 10X
mode. Use Section 4.6.1 of IEEE 1057, with the reference frequency of 1 kHz at an amplitude of
5 divisions, driven from a 50 Ω source with external termination at the input BNC (25 Ω effective source).
TDS2022C, TDS2024C:
to ≥ 200 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full with temperature between
0°C and 35 °C. DC to > 160 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full for
temperatures between 0 °C and 50 °C. < 5 mV/div settings are limited to 20 MHz bandwidth.
TDS2012C, TDS2014C:
DC to ≥ 100 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS2002C, TDS2004C:
DC to ≥ 70 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 MV/div settings
are limited to 20 MHz bandwidth.
TDS2001C:
DC to ≥ 50 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS1012C-EDU:
DC to ≥ 100 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS1002C-EDU:
DC to ≥ 60 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 MV/div settings
are limited to 20 MHz bandwidth.
TDS1001C-EDU:
DC to ≥ 40 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
This is the analog bandwidth when the instrument is DC coupled. V/div values are accurate for probe
attenuation settings of 1X. No probe should be installed for these measurements.
This is the upper frequency for Analog Bandwidth when the instrument has 20 MHz bandwidth limiting
turned on.
20 MHz
Bandwidth of all trigger paths are similarly limited, except the External Trigger, which is not affected
by BW Limit function. Each channel is separately limited, allowing different bandwidths on different
channels of the same instrument.
This is the lower frequency for Analog Bandwidth when the instrument is AC-coupled
≤ 10 Hz.
<1 Hz when 10X, passive probes are used.
Model
TDS2022C, TDS2024C
TDS2012C, TDS2014C
TDS2002C, TDS2004C
TDS2001C
TDS1012C-EDU
TDS1002C-EDU
TDS1001C-EDU
Rise time is generally calculated from the followi ng formula: Rise time in ns = 350 / Bandwidth in MHz
This is the capability of the instrument to capture single event pulses using the Peak Detect Acquisition
Mode.
The minimum single pulse widths for guaranteed 50% or greater amplitude capture are as follows:
Section 4.11.1 of IEEE std. 1057. It is the r atio of the level of a signal input into one channel to that of
the same signal present in another channel due to stray coupling.
Model
TDS2022C
TDS2024C
TDS1012C-EDU
TDS2012C
TDS2014C
TDS1002C-EDU
TDS2002C
TDS2004C
TDS1001C-EDU
TDS2001C
Common Mode Rejection Ratio
100:1 at 60 Hz, reducing to 10:1 with 50 MHz sine wave, with equal
Volts/Div and Coupling settings on each channel.
100:1 at 60 Hz, reducing to 20:1 with a sine wave with frequency equal to
½ the -3 dB bandwidth and with equal Volts//Div and Coupling settings
on each channel.
Crosstalk
≥ 100:1 with a 100 MHz sine wave and with equal V/div settings on each
channel
≥ 100:1 with a 50 MHz sine wave and with equal V/div settings on each
channel
≥ 100:1 with a 30 MHz sine wave and with equal V/div settings on each
channel
≥ 100:1 with a 20 MHz sine wave and with equal V/div settings on each
channel
Position Accuracy
range ±1.8 V
range ±45 V
6TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Specifications
Time Base Syst
em
Table 2: Time base system
Characteris
Sample-Rate
Recor
√ Long-Term Sample
Rate and Horizontal
sition Time Accuracy
Po
Horizontal Position Time
nge
Ra
Horizontal Window
Display
tic
Range
d Length
Description
This is the ra
its inputs and stores the samples in memory to produce a record of time-sequential samples. (IEEE
1057, 2.2.1)
(Sin x)/x interpolationWaveform Interpolation
Waveform interpolation is activated for sweep speeds of 100 ns/div and faster.
This is the total number of samples contained in a single acquired waveform record (Memory Length in
IEEE 1057.2.2.1).
2,500 samples per record.
5 n s/div to 50 s/div in 1–2.5–5 sequenceSeconds/Division Range
Sec/Div Variable function is not available for this product.
s is the maximum, total, long-term error in sample-rate or horizontal position time accuracy, expressed
Thi
in parts per million.
0 ppm over any ≥1 ms interval.
±5
Horizontal scale settingHorizontal position time range
5ns/divto10ns/div
5 ns/div to 100 μs/div
2
250 μs/divto10s/div
25 s/div to 50 s/div
The user controls the time from the trigger to the center graticule on the display with the Horizontal
Position knob.
The resolution of the Horizontal Position time is 1/25 of a horizontal division.
The Window Zone enables a user to select a part of the display to be magnified. The user controls the
location with the Horizontal Position knob and the width with the Sec/Div knob. Vertical cursors show the
selected zone. To magnify the zone, the user selects the window from the menu.
nge of real-time rates, expressed in samples/second, at which a digitizer samples signals at
Sample-rate range
5 S/s to 2000 MS/s
5 S/s to 1000 MS/s
-EDU
C
-4 div * s/div to 20 ms
4 div * s/div to 50 ms
-
-4 div * s/div to 50 s
-4 div * s/div to 250 s
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual7
Specifications
Table2: Timebasesystem(cont.)
Characteristic
Delta Time Measurement
Accuracy
Time Measurement
Functions
Description
This is the accuracy of delta time measurements made on any single waveform.
The limits are given in the following table for signals having an amplitude ≥ 5 divisions, a slew rate at the
measurement points of ≥ 2.0 divisions/ns, and acquired ≥ 10 mV/div.
Condition
Single shot, sample mode,
full bandwidth selected
> 16 averages,
full bandwidth selected
The Sample Interval is the time between the samples in the waveform record.
Frequency, Period, Rise Time, Fall Time, Positive Width, negative Width, Positive Duty Cycle, Phase,
External/5 selection attenuates the external signal by 5.
When Bandwidth Limit is selected for a channel, the bandwidth of that channel’s trigger path will also be
d. The bandwidth of the External Trigger path is not affected by the bandwidth limit.
limite
The trigger position is set by the Horizontal Position knob.
500 ns minimum to 10 s maximum
The ability to set large values of Holdoff is limited by the difficulty in adjusting the Holdoff at Sec/Div
ings less than 100 ms/Div. This is because Holdoff c annot be set in Scan Mode, which begins at
sett
100 ms/div when Trigger Mode is AUTO. By adjusting Trigger Mode to NORMAL, the Scan Mode
operation is turned off, and Holdoff can be adjusted at larger seconds/Div settings.
1M±2%inparallelwith20pF±3pF
300 V
an
Ba
±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level up to 300 V. For
nonsinusoidal waveforms, peak value must be less than 450 V. Excursions above 300 V s hould be less
han 100 ms duration and the duty factor is limited to < 44%. R MS signal level must be limited to 300 V. If
t
these values are exceeded, damage to the instrument may result.
Line Trigger mode provides a source to synchronize the trigger with the AC line input.
Input Amplitude requirements: 85 V
Input Frequency requirements: 45 Hz - 440 Hz.
, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz
RMS
d above
sed on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50 V offset
EDU
Channel 1
Channel 2
External
External/5
AC Line
Channel 1
Channel 2
Channel 3
Channel 4
External
l/5
Externa
AC Line
- 265 VAC.
AC
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual9
Specifications
Table 3: Triggering system (cont.)
CharacteristicDescription
AC, DC, Noise Reject, High Frequency Reject, Low Frequency RejectTrigger Coupling
The External Trigger path does not have a DC blocking capacitor ahead of the trigger input circuit. The
rolloff associated with AC coupling happens after the input circuit. When attempting to trigger on an AC
signal that has a DC offset, care must be used to avoid overloading the input of the External Trigger
circuit. For signals that have a large DC offset, using Channel 1 or Channel 2 (and Channel 3 or Channel
4 for the TDS20X4B) with AC coupling is preferred.
Trigger Slope
√ Sensitivity,
Edge-TypeTrigger,
DC Coupled
Rising Edge, Falling Edge
Measurement Style A) The minimum signal levels for achieving stable frequency indication on the Trigger
Frequency C ounter within 1% of correct indication.
Measurement Style B) Section 4 10.2 in IEEE Std. #1057. The minimum signal levels required for stable
edge triggering of an acquisition when the trigger Source is DC coupled.
Trigger SourceSensitivity (Measurement
Channel Inputs
Ext
Ext/5
Trigger Frequency Readout typically stabilizes at 50% more signal than generates a stable visual display.
All products
TDS1001C-EDU
TDS1002C-EDU
TDS2001C
TDS2002C
TDS2004C
TDS1012C-EDU
TDS2012C
TDS2014C
TDS2022C
TDS2024C
All products
TDS2022C
TDS2024C
All products
TDS2022C
TDS2024C
Sensitivity (Measurement
style A ), typical
1.5 div from DC to
10 MHz ( > 2 mV/div)
4 div from DC to 10 MHz
(2 mV/Div)
3 div between 10 MHz
and 40 MHz
3 div between 10 MHz
and 60 MHz
3 div between 10 MHz
and 50 MHz
3 div between 10 MHz
and 70 MHz
3 div between 10 MHz
and 100 MHz
3 div between 10 MHz
and 200 MHz
300 mV from DC to
100 MHz
500 mV from 100 MHz to
200 MHz
1.5 Vfrom DC to100 MHz1 V from DC to 100 MHz
2.5 V from 100 MHz to
200 MHz
style B)
1 div from DC to 10 MHz >
2mV/div)
2.5 div from DC to 10 MHz
(2 mV/Div)
1.5 div between 10 MHz and
40 MHz
1.5 div between 10 MHz and
60 MHz
1.5 div between 10 MHz and
50 MHz
1.5 div between 10 MHz and
70 MHz
1.5 div between 10 MHz and
100 MHz
1.5 div from 10 MHz to
100 MHz
2.0 div above 100 MHz to
200 MHz
200 mV from DC to 100 MHz
350 mV from 100 MHz to
200 MHz
1.75 V from 100 MHz to
200 MHz
10TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
for Successful
Operation of “Set
Level to 50%”
Function, typical
Trigger Level
Ranges, typical
Trigger Level
Accuracy, DC
Coupled, typical
Video Trigger
Default Settings
for Video Trigger
Video Trigger
Source Selection
Video Trigger
Polarity Selection
Video Sync
Selection
Formats and Field
Rates
Trigger SourceSensitivity
ACSame as DC Coupled limits for frequencies 50 Hz and
Noise Rej
HF Rej
LF RefSame as DC Coupled limits for frequencies above 300 kHz.
Since AC coupling is not done in the front end, use of a 10M probe does not affect the low frequency
corner.
50 Hz.Lowest Frequency
Using a 10M probe will not affect the operation of this function.
Input Channel±8 divisions from center screen
Ext±1.6 V
Ext/5:
The settable resolution for Trigger Level is 0.02 division for an input channel source, 4 mV for Ext source,
and 20 mV for Ext/5 source.
This is the amount of deviation allowed between the level on the waveform at which triggering occurs and
the level selected for DC-coupled triggering signals
±(0.2 d iv + 5 mV) for signals within ±4 divisions from the center screen, having rise and fall times
of ≥ 20 ns.
Ext: ±(6% of setting + 40 mV) for signals less than ±800 mV
Ext/5: ±(6% of setting + 200 mV) for signals less than ±4 V
Trigger Mode: Auto
Trigger Coupling: AC
Same as Trigger Source Selections listed above except Line Trigger. Line Trigger source is meaningless
in this mode.
Normal (Negative going Sync Signal), Invert (Positive going Sync Signal)
Line, Line #, Odd Field, Even F ield, Field: PAL/SECAM, NTSC formats
Field rates: 50 Hz to 60 Hz.Video Trigger
Line rates: 15 kHz to 20 kHz (NTSC, PAL, SECAM)
Specifications
above
Effective in Sample or Average Mode, > 10 mV/div to
5 V/div. Reduces DC Coupled trigger sensitivity by 2X.
Same as DC Coupled limits from DC to 7 kHz.
±8 V
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual11
Falling edge for positive polarity pulse. Rising edge for negative polarity pulse.
16.5 ns or 1 part per thousand, whichever is larger
t > 330 ns: ±5% < guardband < ±(5.1% + 16.5 ns)
t ≤ 330 ns: guardband = ±16.5 ns.
All pulses, even from the most stable sources, have some amount of jitter. To avoid disqualifying pulses
that are intended to qualify but are not absolutely correct values, we provide an arbitrary guardband. Any
measured pulse width within the guardband will qualify. If you are looking for pulse width differences that
are smaller than the guardband width, offsetting the center should allow discriminating differences down
to the guardband accuracy.
330 ns < 1: ±5% ≤ guardband < ±(5.1% + 16.5 ns)
165 ns < 1 < 330 ns: guardband = -16.5 ns/+33 ns
t ≤ 165 ns: guardband = ±16.5 ns
All pulses, even from the most stable sources, have some amount of jitter. To avoid disqualifying pulses
that are intended to qualify but are not absolutely correct values, we provide an arbitrary guardband. Any
measured pulse width outside the guardband will qualify. If you are looking for pulse width differences
that are smaller than the guardband width, offsetting the center should allow discriminating differences
down to the guardband accuracy. Not equal has slightly better ability to deal with small pulse widths
than equal. The accuracy is not better.
Equal: The oscilloscope triggers when the trailing edge of the pulse crosses the trigger level.
Not Equal: If the pulse is narrower than the specified width, the trigger point is the trailing edge. Otherwise,
the oscilloscope triggers when a pulse continues longer than the time specified as the Pulse Width.
Less than: The trigger point is the trailing edge.
Greater than (also called the time out trigger): The oscilloscope triggers when a pulse continues longer
than the time specified as the Pulse Width.
This provides a highly accurate means of identifying the frequency of trigger signals. Since averaging
takes place over a longer time span, the number of stable digits is improved over the automatic
measurement of the same type.
Typical sensitivity
400 mV of composite video
12TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 3: Triggering system (cont.)
Specifications
Frequency
Counter
Resolution
Frequency
Counter Accuracy,
typical
Frequency
Counter
Frequency Range,
typical
Counter Signal
Source
6 digits
±51 ppm including all reference errors and ±1 count errors.
AC coupled. 10 Hz minimum to rated bandwidth.
Pulse width or edge sele cted trigger source.Frequency
Frequency counter measures selected trigger source at all times in pulse width and edge mode, including
when oscilloscope acquisition is halted due to changes in run status, or acquisition of a single shot
event has completed.
Frequency counter does NOT measure pulses that do not qualify as legitimate trigger events.
Pulse Width mode: Counts pulses of sufficient magnitude inside the 250 ms measurement window
that qualify as triggerable events (for example, all narrow pulses in a PWM pulse train if set to < mode
and the limit is set to a relatively small number).
Edge Trigger mode: Counts all pulses of sufficient magnitude.
Display Specifications
Table 4: Display specifications
CharacteristicDescription
Display Type
Display Resolution
Brightness, typical
Contrast Ratio and
Control, typical
11.5 cm (width) * 8.64 cm (height), 14.3 cm diagonal (5.7”), QVGA, active TFT color liquid crystal display
(LCD) with color characters/waveforms on a black background. Surface antiglare (3H) treatment.
This is the number of individually addressable pixels
320 horizontal by 240 vertical pixels
The video display contains both the character and waveform displays.
This is the light output of the backlight.
400 cd/m
Available black room contrast ratio, full black to full white. 400 minimum, 500 typical.
2
, typical. 320 cd/m2min.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual13
Specifications
Interfaces an
d Output Ports Specifications
Table 5: Interfaces and output ports specifications
Characteris
USB DeviceUSB 2.0 Full S
USB H ostUSB 2.0 Full Speed host. 12 Mb/sec maximum. Supports USB Mass Storage Class. Bulk Only Subclass
USB Host CurrentProvides full 0.5 A of 5 V.
GPIB InterfaceGPIB access via TEK-USB-488 accessory.
Output V
Frequency, typical
Data Ha
tic
oltage and
Description
USB-TMC communications with Tektronix extensions.
Standard
only. Prov
Standard
Standard
Output voltage: 5.0 V ±10% into 1 MΩ load.Probe Compensator,
Frequency: 1 kHz
ides full 0.5 A of 5 V.
ndling Characteristics
peed device. 12 Mb/second maximum. Supports PICTBRIDGE compatibility and provides
Table 6: Data handling characteristics
cteristic
Chara
ntion of Front
Rete
Panel Settings
Stored Waveforms and
Multiple Front Panel
tings
Set
iption
Descr
Front panel settings are stored periodically in memory. The settings are not lost when the instrument is
ed off or if there is a power failure.
turn
Two Channel 1, Channel 2, or Math waveforms can be stored in nonvolatile waveform memory A or
B. One, both, or neither of A or B waveform memories can be displayed. For the TDS20X4C, four
eforms, A, B, C, D, or Math can be stored. Ten user setups of the current instrument settings can be
wav
saved and restored from nonvolatile memory. Additional storage is available when an appropriate mass
storage device is connected via USB.
Power Distribution System
ble 7: Power distribution system
Ta
CharacteristicDescription
ess than 30 W a t 85 to 275 V
Power Consumption
Source VoltageFull Range: 100 to 240 V
Source Frequency360 Hz to 440 Hz from 100 VACto 120 VAC.
Fuse Rating
L
ACRMS
45 Hz to 66 Hz from 100 V
3.15 Amps, T rating, 250 V; IEC and UL approved.
input.
AC
± 10%, Installation Category II (Covers range of 90 to 264 V
to 240 VAC.
AC
AC
14TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Specifications
Mechanical Ch
aracteristics
Table 8: Mechanical characteristics
Characteris
Weight
Size
Cooling
tic
Method
Description
Requirement
2.0 k g (4.4 lbs), stand-alone instrument
2.2 kg (4.9 lbs), with accessories
3.6kg(8lbs
Height
Width
Depth
Convect
s that follow are nominal:
), when packaged for domestic shipment
ion cooled
Environmental Performance
Table 9: Environmental performance
CharacteristicDescription
Temperature
Humidity
Altitude
Operating0° C to +50° C (32 °F to 122 °F), with 5° C/minute maximum gradient,
Nonoperating
ating
Oper
Nonoperating
erating
Op
Nonoperating
158 mm (6.2
326.3 mm (1
124.1 mm (
nonco
-40°
5% to
5% to 45% RH above +40° C up to +50° C, noncondensing, and as limited
by a Maximum Wet-Bulb Temperature of +37° C (derates relative humidity
5 % RH at +50° C)
to 4
o 85% relative humidity (% RH) at up to +40° C
5% t
5% to 45% RH above +40° C up to + 50° C, noncondensing.
Above +50° C limited by a Maximum Wet-Bulb Temperature of +37° C
erates relative humidity to 12% RH at +71° C)
(d
to 3000 meters (10,000 feet)
Up
p to 3000 meters (10,000 feet).
U
Altitude is limited by possible damage to the L CD at higher altitudes. This
damage is independent of operation
2in)
2.85 in)
4.88 in)
ndensing, up to 3000 m altitude
C to + 71° C (-40 °F to 159.8 °F), with 5° C/minute maximum gradient
85% relative humidity (% R H) at up to +40° C
Data Logging System Characteristics (Not Available on T DS1000C-EDU M odels)
NOTE. This software feature directs the oscilloscope to autom atically collect
data over a period of time. After you configure the trigger conditions to use,
you can use the data logging menu to set up the oscilloscope so that it will save
all of the triggered waveform to a USB memory device, within a time duration
that you have set.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual15
Specifications
Table 10: Data l
CharacteristicDescription
DurationThe time perio
SourceThe signal source which you want to save the waveform.
Select Fold
ogging system characteristics (not available on TDS1000C-EDU models)
The file fold
You can create the new folder or change the existing folder as the folder where you want to save the
waveform data.
hannel 2, Channel 3, Channel 4, Math
er where you save the waveform data.
Limit Testing System Characteristics (Not Available on TDS1000C-EDU Models)
NOTE. This software feature directs the oscilloscope to monitor an active input
signal against a template and to output pass or fail results by judging whether
the input signal is within the bounds of the template.
Table 11: Limit testing system chara cteristics (not available on TDS1000C-EDU models)
teristic
Charac
e
Sourc
re Ref Channel
Compa
Run/StopTo enable or disable the limit testing function.
Template SetupYou can use this menu item to set up a limit test waveform template. The template is the mask signal that
SourceThe location of the signal source that is used to create the limit test template.
Vertical LimitThe vertical limit in vertical divisions.
Horizontal LimitThe horizontal limit in horizontal divisions.
Destination Ref ChannelThe location of the reference memory location that is used to store the limit test template.
Display TemplateDisplays or does not display a stored test template.
ption
Descri
The signal source which you want to do the limit testing.
el 1, Channel 2, Channel 3, Channel 4, Math
Chann
ference channel # where the template is saved.
The re
RefA, RefB, RefC, RefD. The limit testing system will compare the source signal with this template.
Run, Stop.
define as the boundary to compare with the input source signal. You can create the template from
you
internal or external waveforms with specific horizontal and vertical tolerances.
Channel 1, Channel 2, Channel 3, Channel 4, Math
0~1000 mdiv
0~500 mdiv.
RefA, RefB, RefC, RefD.
On, Off.
16TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Table 11: Limit testing system characteristics (not available on TDS1000C-EDU models) (cont.)
CharacteristicDescription
Action on Violation
Stop AfterDefines the conditions that will cause the oscilloscope to end limit testing.
Defines the actions the o scilloscope will take after a violation is detected.
Save Image: The oscilloscope will automatically save a screen image when a violation is detected.
Save Waveform: The oscilloscope will automatically save a digital copy of the source waveform when
a violation is detected.
Manual: Lets you stop the test by toggling the “Run/stop” choice.
Waveforms: Lets you set the numbers of waveforms to test before stopping limit testing.
Violations: Lets you set the numbers of violations to detect before stopping limit testing.
Elapsed time: Lets you set the elapsed test time in seconds to pass before stopping limit testing.
Specifications
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual17
Specifications
18TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
WheretoFindOperatingInformation
For information on installing and operating your TDS2000C or TDS1000C-EDU
Series Digital Storage Oscilloscope, refer to the user manual for your product.
The user manu
www.tektronix.com/manuals.
als are available in eleven languages and are on the Web at
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual19
Where to Find Operating Information
20TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Theory of Operation
This section covers the electrical operation of the TDS2000C and TDS1000C-EDU
series oscilloscopes to the module level by describing the basic operation of e ach
functional c
NOTE. The descriptions for the 2-channel and 4-channel models are slightly
different.
ircuit block.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual21
Theory of Operation
System Level B
lock Diagrams
Two high-level block diagrams are provided, one for 2-channel models and one
for 4-channel models. (See Figure 1 on page 22.) (See Figure 2 on page 23.)
Figure 1: Module-level block diagram (2-channel)
22TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Theory of Operation
Figure 2: Module-level block diagram (4-channel)
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual23
Theory of Operation
Main Board
The Main board is also called the acquisition board. The Main board of the
4-channel oscilloscopes is essentially two 2-channel oscilloscopes tied together
through a common microprocessor, and some special interconnects to support
combining the display and trigger systems. The focus of the Main board
discussion
noted as necessary.
is the 2-channel system, with differences for the 4-channel models
Acquisi
tion System
At a minimu
digitizer/trigger system ASIC, a signal processing/display/system services
RAM, flash PROM, a system microprocessor, USB controller, USB RAM, system
communication RAM, and special power supplies. For a 4-channel oscilloscope,
the attenuators a re duplicated. Most of the other a spects of the circuitry remain
unchanged.
Signals from the channel 1 and channel 2 and other input connectors pass through
attenuators and an AC-coupling switch to the amplifier ASIC. The Ext Trig input
has an abbreviated version of this path, lacking some of the attenuator settings
and th
The amplifier ASIC contains buffers and variable gain amplifiers, as well as
s that provide 20 MHz bandwidth limiting. The task of the amplifier ASIC
filter
is to convert from a 1 MΩ single-ended environment in the front end to a much
lower impedance differential (and thus less noise-sensitive) environment for the
acquisition process. The amplifierASICassuresthattheinputsignalisamplified
to a level that will allow the fullest possible use of the digitizer.
The a cquisition ASIC contains samplers and peak detectors for each input
channel, a common amplifier, an A/D converter, and the trigger logic. The
digitized waveform samples are transferred to the processing and display
n 4-channel systems, the two acquisition ASICs are interconnected so that a
I
trigger on one ASIC can cause a trigger on the other.
m, the Main board contains attenuators, an amplifier ASIC, a
FPGA,
e AC coupling switch.
FPGA.
he processor system adds the m icroprocessor and flash PROM to the processing
T
and display system. The processor system interprets the front-panel control
changes detected by the disp lay ASIC, provides control parameters based on
user setting requests, computes waveform measurements, and manages the USB
interfaces via the d edicated USB controller. Saved setups, waveforms, and
calibration constants are stored in nonvolatile memory sections within the flash
PROM. The processor system shares DRAM with the display system.
24TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Theory of Operation
Processing and Display
System
Input Signal Interface
Probe Compensation
The processing
system oscillator. Digitized acquisition samples are received by the display
ASIC and stored in DRAM. Once data are received by the display ASIC, various
corrections are applied, display rasterization is performed, and the waveform is
placed in a display buffer. At the same time, the waveform is read from the display
buffers and written to the LCD. Additional circuitry in the display ASIC supports
scanning th
clock, and p erforming various memory mapping tasks required by all elementary
microprocessor based system s. In a 4-channel system, the two display ASICs are
interconnected so that one ASIC can provide display information for the second.
The processing and display system handles some of the computational tasks.
Other tasks are performed by the processor system. Since all array processing
is performed in the processing and display system, no computations can be
performed that involve data from two different channel sets. Thus, subtracting
channel 3 data from channel 2 is prohibited. Channel 1 and Channel 2 data may
be combined in all of the supported ways.
BNC connectors are mounted on the Main board for all signal inputs. The signal
inputs are compatible with the supplied TPP0101 and TPP0201 probes.
The Probe Comp and ground terminals are provided for probe adjustment.
and display system consists of the display ASIC, DRAM, and
e front panel, handling DRAM refresh, providing the processor
External Trigger
Main Board Power
Power Supply
The Ext Trig channel is proces se d on the chain containing the highest numbered
normal input channel.
To support various functions on the Main board, several secondary power sup
are generated. For the amplifier and acquisition ASICs, the Main board creates a
+2.5 V and -2.5 V supply. The +2.5 V supply is derived from the +3.3 V logic
supply. The -2.5 V supply is derived from the -4 V supply. One three-terminal
regulator provides +5 V for internal uses. A second three-terminal regulator
provides USB power to preclude USB faults from seriously disrupting operation
of the oscilloscope.
The main power supply module for these oscilloscopes is a wide input range
universal supply. It can provide about 25 W of power for the oscilloscope, while
allowing the input to run from about 90 V to 264 V. Input frequency ranges from
47 Hz to 440 Hz, which allows operation in virtually all countries in the world and
in several off-grid environments, such as power on military aircraft.
The secondary supplies from the power supply and the approximate current draws
are listed in a table with the associated connector pins on J101. (See Table 20.)
plies
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual25
Theory of Operation
Display M odule
Front Panel
Two-Chan
Oscilloscopes
nel
The P2 connecto
For information on voltages used in the oscilloscope, refer to the Troubleshooting
section of thi
The color di
The unit is patterned with 320 columns x 3 sub-columns by 240 rows and has
the associated drivers and backlight. The LED backlight output is about 75 mA
at 15.5 V.
You can m anipulate all of the sw itches, and position encoders on the Front-Panel
board of 2-channel oscilloscopes. Several LEDs are used to indicate when the
Multipurpose knob is active, when Autorange is active, and when a Save action
is in pr
your product.
An IC o
signals to the Main board. Two s ignals and a sense line are provided by the Main
board to support the front panel. One of these lines resets the scan; a second line
clocks the scan to the next position; and the sense line receives the current state of
the selected switch or encoder position.
ogress. For more information on the LEDs, refer to the user manual for
n the Front-Panel board provides buffering and multiplexing of switch
r uses 4 positions and 3 wires, w h ich are +6 V, +3.3 V, and GND.
smanual.
splay module is a standard passive liquid crystal display (LCD).
Four-Channel
Oscilloscopes
For the encoders, some amount of debouncing occurs inside the front-pan el IC.
All key debouncing is handled in the display
The LEDs are controlled by latching the value of the Channel 1 - 2 front-panel
scan counter when the appropriate scan value is set.
The Front-Panel board of the 4-channel units is effectively two panels in parallel.
The left side of the board is largely handled by the display FPGA for channels
1 and 2. The right side of the board is handled by the channel 3 and 4 ASIC.
Separate front-panel ICs support these data paths.
FPGA on the Main board.
26TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Performance Verification
This chapter contains performance verification procedures for the specifications
marked with the check mark. The following equipment, or a suitable equivalent,
is required t
Required Equipment
o complete these procedures.
Table 12: P
Descripti
DC Voltage
Leveled S
Time Mark
50Ω BNC CableBNC male to BNC male, ≈ 1 m (36 in) long
50Ω BNC CableBNC male to BNC male, ≈ 25 cm (10 in) long
50Ω Feedthrough TerminationBNC male and female connectors
Dual Banana to BNC AdapterBanana plugs to BNC female
BNC T AdapterBNC male to dual BNC female connectors
Splitter, PowerFrequency range: DC to 4 GHz.
pter (four required)
Ada
Adapter
Leads, 3 Black
ads, 2 Red
Le
erformance verification
on
Source
ine Wave Generator
Generator
Minimum re
17.5mVto7
50 kHz and
10 ms period, ±10 ppm accuracy
king: >2.0%
Trac
e N-to-female BNC
Mal
Fem
acking Banana Plug Patch Cord, ≈ 45 cm
St
(18 in) long
Stacking Banana Plug Patch Cord, ≈ 45 cm
(18 in) long
quirements
V, ±0.5% accuracy
200 MHz, ±3% amplitude accuracy
ale N-to-male BNC
Examples
Wavetek 91
System with Oscilloscope Calibration
Module (Option 250)
Tektronix part number 012-0482-XX
Tektronix part number 012-0208-XX
Tektronix part number 011-0049-XX
Tektronix part number 103-0090-XX
Tektronix part number 103-0030-XX
Tektronix part number 015-0565-XX
Tektronix part number 103-045-XX
Tektronix part number 103-0058-XX
mona #B-18-0
Po
Pomona #B-18-2
00 Universal Calibration
0A Multi-product Calibrator with
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual27
External Edge Trigger SensitivityStable trigger
Channel 1 Vertical Position Accuracy,
33.6 m V36.4 m V
1.358 V1.442 V
13.58 V14.42 V
33.6 m V36.4 m V
1.358 V1.442 V
13.58 V14.42 V
33.6 m V36.4 m V
1.358 V1.442 V
13.58 V14.42 V
V
33.6 m
V
1.358
8V
13.5
V
2.12
2V
2.1
2V
2.1
12 V
2.
2divs
-
Stable trigger
Stable trigger
0
36.4 m
1.442
2V
14.4
2
—
2
—
2
—
2
—
2divs
+
3
—
3
—
3
—
3
—
3
—
—
V
V
Minimum margin
Channel 2 Vertical Position Accuracy,
0
—
Minimum margin
Channel 3 Vertical Position Accuracy1,
0
—
Minimum margin
Channel 4 Vertical Position Accuracy1,
0
—
Minimum margin
1
Channels 3 and 4 are only available on four channel oscilloscopes.
2
The bandwidth test does not have a high limit.
3
Thelimitsvarybymodel. Checkthe procedure for the correct limits.
28TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Performance Verification
Performance V
Self Calibration
erification Procedures
Before beginning these procedures, two conditions must be met:
The oscilloscope must have been operating continuously for twenty minutes
within the operating temperature range specified in the Environmental
Performance table. (See Table 9.)
You must perform the S elf Calibration operation described below. If the
ambient temperature changes by more than 5 °C, you must perform the Self
on operation again.
Some procedures use hazardous voltages. To prevent electrical
ges are displayed before continuing with this procedure.
Self Test
Calibrati
The time required to complete the entire procedure is approximately one hour.
WAR N ING.
shock, always set voltage source outputs to 0 V before making or changing any
interconnections.
This internal procedure is automatically performed every time the oscilloscope
is powered on. No test equipment or hookups are required. Verify that no error
messa
The self calibration routine lets you quickly optimize the oscilloscope signal path
for maximum measurement accuracy. You can run the routine at any time, but you
ld always run the routine if the ambient temperature changes by 5 °C or more.
shou
heck DC Gain Accuracy
C
1. Disconnect all probes and cables from the channel input connectors (channels
, 3, 4).
1, 2
2. Push the Utility button and select the Do Self Cal option to start the routine.
e routine takes approximately one minute to complete.
Th
3. Verify that self calibration passed.
This test checks the DC gain accuracy of all input channels.
1. Set the DC voltage source output level to 0V.
2. Set up the oscilloscope using the following table:
Push menu buttonSelect menu optionSelect setting
Default Setup
Channel 1
AcquireAverage16
Measure
——
Probe1X
SourceChannel under test
TypeMean
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual29
Performance Verification
3. Connect the osc
illoscope channel under test to the DC voltage source as
shown in the following figure:
4. For each vertical scale (volts/division) setting in the following table, perform
the follow i ng steps:
a. Set the DC voltage source output level to the positive voltage listed and
then record the mean measurement as V
pos
.
b. Reverse the polarity of the DC voltage source and record the mean
measurement as V
c. Calculate V
diff
=V
.
neg
pos-Vneg
and compare V
to the accuracy limits in
diff
the follow ing table:
Vertical Scale
(volts/div) setting
5 mV/div
200 mV/div
2V/div
DC voltage
source output levelsAccuracy limits for V
+17.5 mV, -17.5 mV33.6 mV to 36.4 mV
+700 mV, -700 mV1.358 V to 1.442 V
+7.00 V, -7.00 V13.58 V to 14.42 V
5. Set DC voltage source output level to 0V.
6. Disconnect the test setup.
7. Repeat steps 1 through 3 for all input channels.
diff
30TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Performance Verification
Check Bandwidth
This test check
s the bandwidth of all input channels.
1. Set up the oscilloscope using the following table:
Push menu buttonSelect menu optionSelect setting
Default Setup
Channel 1
AcquireAverage16
Trig Menu
Measure
——
Probe1X
Coupling
SourceChannel under test
TypePk-Pk
Noise Reject
2. Connect the oscilloscope channel under test to the leveled sine wave generator
as shown in the following figure:
3. Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
4. Set the oscilloscope Horizontal Sc ale (seconds/division) to 10 ms/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-peakmeasurement is between 2.98 V and 3.02 V.
7. Set the leveled sine wave generator frequency to:
40 MHz if you are checking a TDS1001C-EDU
50 MHz if you are checking a TDS2001C
60 MHz if you are checking a TDS1002C-EDU
70 MHz if you are checking a TDS2002C or TDS2004B
100 MHz if you are checking a TDS1012C-EDU, TDS2012C, or
TDS2014C
200 MHz if you are checking a TDS2022C or TDS2024C
8. Set the oscilloscope Horizontal Scale (seconds/division) to 10 ns/div.
9. Check that the peak-to-peak measurement is ≥ 2.12 V.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual31
Performance Verification
Check Sample R
ate
Accuracy and Delay Time
Accuracy
10. Disconnect the
test setup.
11. Repeat steps 1 through 10 for all input channels.
This test checks the time base accuracy.
1. Set up the oscilloscope using the following table:
Push menu bu
Default Se
Channel 1
tup
tton
Select menu
——
Probe1X
option
Select sett
ing
2. Connect the oscilloscope to the time mark generator as shown in the following
figure:
3. Set the time mark generator period to 10 ms.
4. Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
5. Set the oscilloscope Main Horizontal Sc ale (seconds/division) to 1ms/div.
6. Push Set To 50%.
7. Use the Ve r ti ca l Pos i ti on c ontrol to center the test signal on screen.
8. Us
etheHorizontal Position control to set the position to 10.00 ms.
9. Set the oscilloscope H o rizontal Scale (seconds/division) to 250 ns/div.
32TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Performance Verification
10. Check that the r
ising edge of the marker crosses the center horizontal graticule
line within ±2 divisions of the vertical center graticule line, as shown in the
following figure:
NOTE. On
e division of displacem ent from graticule center corresponds to a
25 ppm time base error.
Check
Edge Trigger
Sensitivity
11. Disconnect the test setup.
This test checks the edge trigger sensitivity for all input channels.
1. Set up the oscilloscope using the following table:
menu button
Push
ault Setup
Def
nnel 1
Cha
Trig MenuModeNormal
Acquire
Measure
ct menu option
Sele
——
Probe1X
Sample
SourceChannel under test
TypePk-Pk
Sele
—
ct setting
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual33
Performance Verification
2. Connect the osc
illoscope channel under test to the leveled sine wave generator
as shown in the following figure:
3. Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
4. Set the oscilloscope Horizontal Scale (seconds/division) to 25 ns/div.
5. Set the leveled sine wave generator frequency to 10 MHz.
6. Set the leveled sine wave generator output level to approximately 500 mV
so that the measured amplitude is approximately 500 mV.(Themeasured
amplitude can fluctuate around 500 mV.)
7. Push Set To 50%. Adjust Trigger Level as necessary and then check that
triggering is stable.
p-p
8. Set the leveled sine wave generator frequency to:
40 MHz if you are checking a TDS1001C-EDU
50 MHz if you are checking a TDS2001C
60 MHz if you are checking a TDS1002C-EDU
70 MHz if you are checking a TDS2002C, or TDS2004C
100 MHz if you are checking a TDS1012C-EDU, TDS2012C, TDS2014C,
TDS2022C, or TDS2024C
9. Set the oscilloscope Horizontal Scale (seconds/division) to 5ns/div.
10. Set the leveled sine wave generator output level to approximately 750 mV
so that the measured amplitude is approximately 750 mV.(Themeasured
p-p
amplitude can fluctuate around 750 mV.)
11. Push Set To 50%. Adjust Trigger Level as necessary and then check that
triggering is stable.
12. For the TDS2022C and TDS2024C models, set the frequency to 200 MHz,
and increase the amplitude to 1 V
. Verify stable triggering.
p-p
13. Set the oscilloscope Horizontal Scale (seconds/division) to 2.5 ns/div.
14. Change the oscilloscope setup using the following table:
34TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Performance Verification
Push menu buttonSelect menu optionSelect setting
Trig Menu
Slope
Falling
15. Push Set To 50%. Adjust Trigger Level as necessary and then check that
triggering is stable.
16. D isc onnect the test setup.
17. Repeat steps 1 through 16 for all input channels.
Check External Edge
Trigger Sen sitivity
This test checks the edge trigger sensitivity for the external trigger.
1. Set up the oscilloscope using the following table:
2. Connect the oscilloscope to the leveled sine wave generator as shown in the
following figure, using channel 1 and Ext Trig.
3. Set the oscilloscope Vertical Scale (volts/division) to 100 mV/div.
4. Set the oscilloscope Horizontal Scale (seconds/division) to 25 ns/div.
5. Set the leveled sine wave generator frequency to 10 MHz.
6. Set the sine wave generator output level to approximately 300 mV
power splitter. This is about 200 mV
on channel 1 of the oscilloscope.
p-p
into the
p-p
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual35
Performance Verification
The Ext Trig inp
ut will also be receiving approximately 200 mV
deviations from the nominal 200 mV
oscilloscope display are acceptable.
p-p
. Small
p-p
7. Set the leveled sine wave generator frequency to:
40 MHz if you are checking a TDS1001C-EDU
50 MHz if you are checking a TDS2001C
60 MHz if you are checking a TDS1002C-EDU
70 MHz if you are checking a TDS2002C, or TDS2004C
100 MHz if you are checking a TDS1012C-EDU, TDS2012C, TDS2014C,
TDS2022C, or TDS2024C
8. Set the oscilloscope Horizontal Scale (seconds/division) to 5ns/div.
9. Push Set To 50%. Adjust Trigger Level as necessary and then check that
triggering is stable.
10. For TDS2022C and TDS2024C models, set the sine wave generator output
level for a channel 1 display to approximately 350 mV
p-p
.
11. For TDS2022C and TDS2024C models, set the sine wave generator frequency
to 200 MHz.
12. Set the oscilloscope Horizontal Scale (seconds/division) to 2.5 ns/div.
Check Vertical Position
Accuracy
13. Push Set To 5 0%. Adjust Trigger Level as necessary and then check that
triggering is stable.
14. Change the oscilloscope setup using the following table:
Push menu buttonSelect menu optionSelect setting
Trig Menu
Slope
Falling
15. Push Set To 5 0%. Adjust Trigger Level as necessary and then check that
triggering is stable.
16. Disconnect the test setup.
The results of this test and the DC Gain Accuracy test together define the D C
Measurement Accuracy of the oscilloscope. The DC Measurement Accuracy
specification encompasses two different ranges of operation over two different
attenuator settings.
DC Gain Accuracy: Ide
ntifies err ors, mostly from the A/D converter,
when the v ertical position (known as offset in these oscilloscopes) is set to
0 divisions (or a grounded input will show screen center)
Vertical Position Accuracy: Identifies errors, mostly from the position control,
made when the vertical position is set to a non-zero value
36TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Performance Verification
The two attenua
tor settings operate identically, so verification of the attenuation
range from -1.8 V to 1.8 V also verifies the attenuation range of -45 V to 45 V.
1. Set up the oscilloscope as shown in the following table:
The test operates without a trigger. To maintain uniformity and to avoid false triggering on noise, the Ext trigger is
the recommended source.
——
Probe1X
1
Source
ModeAuto
Sample
SourceChannel under test
TypeMean
Ext
—
2. Make a spreadsheet approximately as shown in the example in Appendix A.
You only need to enter the values for column A and the equations. The values
incolumnsB,C,D,E,F,andGareexamplesofthemeasuredorcalculated
values.
The PDF versio n of the service manual (which you can download from
www.tektronix.com/manuals), includes an empty spreadsheet for your
convenience. To access and save the test spreadsheet, see the instructions
in Appendix A: Example of a Vertical Position Accuracy Test Spreadsheet
on page A-1.
3. Connect the oscilloscope, power supply, and voltmeter as shown in the
following figure:
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual37
Performance Verification
4. Set the power su
Test Voltage.
5. Adjust the ver
center of the screen.
6. Enter the vo
in the appropriate columns, B and C.
7. Repeat step
8. Swap the connections to the positive term inal of the power supply with those
at the n ega
pply to the 1.8 V value shown in column A, the Approximate
tical position k nob for the DC line to position the line in the
ltage on the voltmeter and on the oscilloscope into the spreadsheet
s 4 through 6 for the values of 1.76 V through 0 V.
tive terminal as shown in the following figure:
9. Repeat steps 4 through 6 for the values of -0.04 V through -1.8 V.
10. Enter the Minimum Margin number (cell I16) for the channel tested in the
test record.
11. Repeat steps 1 through 10 for all input channels.
38TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Performance Verification
Data verificati
columns D, E, and F. Verify that no error values (the blue line in the center) go
above the yellow line (upper line), or below the purple line (lower line). For
calculations involved in this example, refer to the data in the previous table (see
step 1).
on. To verify data, set the spreadsheet to present a line graph of
Figure 3: Example of a line graph for the Vertical Position Accuracy test
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual39
Performance Verification
40TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Adjustment Procedures
This section contains adjustment procedures for the TDS2000C and
TDS1000C-EDU series oscilloscopes.
Only qualified personnel should perform service procedures. Read the ServiceSafety Summary and the General Safety Summary at the beginning of this
manual before p erform ing any service p rocedures. Refer to the user manual for
information about using these oscilloscopes.
NOTE. The voltage references inside the oscilloscopes are very stable over time
and do not require routine updates. Do not perform any procedure in this section
unless the oscilloscope does not meet specifications. Refer to the Performance
Ver ification section to check the oscilloscope performance to specifications.
Required Equipment
The equipment described in the following table, or a suitable equivalent, is
required to com plete these procedures. (See Table 14.)
Table 14: Required equipment
DescriptionMinimum requirements sExample
DC Voltage Source-20 V to 20 V, ± 0.1% accuracy
Leveled Sinewave Generator5 kHz and 200 MHz, ±0.6% amplitude accuracy
50 Ω BNC CableBNC male to BNC male, ≈ 1 m (36 in) long
50 Ω BNC Cable (seven)BNC male to BNC male, ≈ 25 cm (10 in) long
50 Ω Feedthrough TerminationBNC male and female connectors
Dual Banana to BNC AdapterBanana plugs to BNC female
BNC T (three)One m ale and two female BNC connectors
Wavetek 9100 Universal Calibration
System with Oscilloscope Calibration
Module (Option 250)
Fluke 5500A Multi-product Calibrator with
Oscilloscope Calibration Option (Option
5500A-SC)
Tektronix part number 012-0482-XX
Tektronix part number 012-0208-XX
Tektronix part number 011-0049-XX
Tektronix part number 103-0090-XX
Tektronix part number 103-0030-XX
Adjustment Procedure
The adjustment procedure consists of 70 steps for the 2-channel oscilloscope, and
106 steps for the 4-channel model. Each step requires an external voltage source
for new calibration constants.
Before performing the adjustment procedure, warm up the oscilloscope for at least
ten minutes in an ambient temperature between 20 °C and 30 °C. Adjustments
performed before warm up or outside this temperature range may result in poor
performance.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual41
Adjustment Procedures
Enable the Service Menu
If all steps in t
displayed and the new calibration constants take effect. If any step fails, the
procedure is aborted and the current calibration is not affected. You can choose
to abort the procedure at a ny step, without affecting the current calibration, by
selecting FCAL ABORT.
NOTE. While performing the adjustment procedure, keep in mind that some of the
steps take a
Equipment setups are shown in the following figure. (See Figure 4 on page 43.)
Signal re
on page 45.)
Before performing the adjustment procedure, enable the Service menu as follows:
1. Power on the oscilloscope.
2. Push the front-panel Measure button to access the Measure menu.
3. Push the top option button to access the Measure 1 menu.
4. Push and hold the front-panel Single button.
he procedure are completed successfully, a "Pass" message is
significant amount of time (up to 5 minutes) to complete.
quirements for each step are shown in the following table. (See Table 15
5. Push
6. Wait at least two seconds.
7. Release the Single button.
8. Release the AutoSet button. A message appears in the lower left corner of the
9. Push the front-panel Utility button. The last item in the Utility m enu is now
At completion of the Adjustment procedure, disable the "Service" menu through
the Utility front panel button, the "Service" option button, and the "Ser vice"
Mode Off" option button.
and hold the front-panel AutoSet button.
screen stating "Service mode ON."
"Service."
42TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Adjustment Procedures
Figure 4: Adjustment setups
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual43
Adjustment Procedures
Adjustment Procedure
Use this proced
uretoloadnewcalibrationconstants.
1. Enable the Service menu. (See page 42, Enable the Service Menu.)
2. Push the Service option button.
3. Push the Factory Cal option button.
4. Push the Warmup Timer option button.
5. Push the Start Timer option button to allow a 10-minute warm up. After
10 minutes, the oscilloscope displays a large black or green cross.
6. Press the front-panel Utility button and push the Service option button.
7. Push the Factory Cal option button.
8. Push the F
CAL INIT option button to initiate the factory calibration routine.
You are now in step 1 of the procedure.
9. As indic
ated by step 1, connect a 0.0 V
source to the channel inputs. (See
DC
Table 15.)
10. Press F
CAL STEP to load the calibration constant for step 1.
11. Follow t h e screen prompts and the Adju stm ent setup table for the remainder
routine, applying the proper signal for each step. (See Table 15.)
of the
NOTE. During some steps, the oscilloscope may appear to be idle for several
minutes while it is processing information internally. You must wait for the screen
prompt to continue.
If any step in the procedure fails, the procedure terminates. Failure will result
if an incorrect signal source is connected. If this happens, you must start the
procedure over from the beginning.
If the procedure completes succ es sfully, a "Pass" message is d isplayed and the
new adjustment takes effect.
12. Compensate the signal path by p erfo rm ing the self-calibration routine.
13. Disconnect all probes and cables from the channel input connectors.
14. Press the Utility button and select Do Self C al to confirm that you are ready
to proceed.
15. Disable the Service menu option by pushing the Utility front-panel button and
the Service and ServiceModeOffoption buttons.
44TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
9
0
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 1
DC Voltage 2
DC Voltage 2
DC Voltage 2
DC Voltage 2
0.0 V
-20.0 V
-2.0 V
-1.6 V
-1.2 V
-0.8 V
-0.4 V
-0.32 V
-0.2 V
-0.16 V
-0.08 V
-0.04
3V
-0.0
2V
-0.0
015 V
-0.
15 V
0.0
02 V
0.
03 V
0.
.04 V
0
.08 V
0
0.16 V
0.2 V
0.32 V
0.4 V
0.8 V
1.2 V
1.6 V
2.0 V
20.0 V
0.0 V
-5.0 V
-1.0 V
1.0 V
5.0 V
V
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual45
5kHz,1V
1 MHz, 100 mV
1MHz,1V
50 kHz, 1 V
20 MHz, 1 V
BWL, 25 mV
BWL, 50 mV
BWL, 100 mV
BWL, 250 mV
BWL, 500 mV
BWL, 1 V
5kHz,2.5V
1MHz,2.5V
50 kHz, 2.5 V
20 MHz, 2.5 V
BWL, 2.5 V
BWL, 5 V
BWL, 1.6 V
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
1 kHz, 0 to -800 mV
5kHz,1V
1 MHz, 100 mV
1MHz,1V
50 kHz, 1 V
20 MHz, 1 V
BWL, 25 mV
BWL, 50 mV
BWL, 100 mV
BWL, 250 mV
BWL, 500 mV
BWL, 1 V
5kHz,2.5V
1MHz,2.5V
50 kHz, 2.5 V
20 MHz, 2.5 V
BWL, 2.5 V
BWL, 5 V
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual47
Adjustment Procedures
48TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Preventing ESD
This chapter contains the informationneededtodoperiodicandcorrective
maintenance on the oscilloscope, as well as repackaging instructions to use if you
need to return the oscilloscope to Tektronix for service.
Before servicing this product, read the Safety Summary and Introduction at the
front of the manual and the electrostatic discharge (ESD ) information below.
CAUTION. Static discharge can damage any semiconductor component in this
oscilloscope.
When performing any service w hich requires internal access to the oscilloscope,
adhere to the following precautions. These precautions help you avoid damaging
internal modules and their components due to electrostatic discharge (ESD):
Minimize handling of static-sensitive modules.
Transport and store static-sensitive modules in their static protected containers
or on a metal rail. Label any package that contains static-sensitive modules.
Inspection and Cleaning
Inspection and cleaning are done a s preventive maintenance. Preventive
maintenance, when done regularly, can prevent oscilloscope malfunction and
enhance its reliability.
Preventive maintenance consists of visually inspecting and cleaning the
osc
maintenance depends on the severity of the environment in which you use the
oscilloscope. A good time to perform preventive maintenance is just before
adjustment of the oscilloscope.
Discharge the static voltage from your body by wearing a grounded antistatic
wrist strap while handling these modules.
Service static-sensitive modules only at static-free workstations. Do not allow
anything capabl e of generating or holding a static charge on the workstation
surface.
Handle circuit boards by the edges when possible.
Do not slide the modules over any surface.
Avoid handling modules in areas th
capable of generating a static charge.
illoscope, and using general care when operating it. How often you do
at have a floor or work-surface covering
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual49
Maintenance
General Care
Interior Cleaning
For best perfor
Protect the oscilloscope from adverse weather conditions. The oscilloscope is
not waterproo
Do not store or leave the oscilloscope w here the liquid crystal display (LCD)
is exposed t
The front and rear cases help keep dust out of the oscilloscope and must be in
place durin
To avoid damage to the oscilloscope or probes, do not expose them to any
sprays, li
Use a dry, low-velocity stream of air to clean the interior of the chassis. Use a
soft-bristle, non-static-producing brush for cleaning around components. If you
must use
and rinse with deionized water.
WARNING. To avoid injury, power o ff the instrument and disconnect it from line
voltage before performing any procedure that follows.
mance, follow these recommendations:
f.
o direct sunlight or high humidity for long periods of time.
g normal operation.
quids, or solvents.
a liquid for minor interior cleaning, use a 75% isopropyl alcohol solution
Exterior Cleaning
ION. Avoid the use of chemical cleaning agents, which might damage the
CAUT
plastics used in this oscilloscope. Use only deionized water when cleaning the
menu buttons or front-panel buttons. Use a 75% isopropyl alcohol solution as a
cleaner and rinse with deionized water. Bef ore u sing any other type of cleaner,
consult your Tektronix Service Center or representative.
Clean the exterior surfaces of the chassis with a dry lint-free cloth or a soft-bristle
brush. If any dirt remains, use a cloth or swab dipped in a 75% isopropyl alcohol
solution. Use a swab to clean narrow spaces around controls and connectors.
o not use abrasive compounds on any part of the chassis that m ay damage the
D
chassis.
CAUTION. To prevent getting moisture inside the oscilloscope during e xternal
cleaning, use only enough liquid to dampen the cloth or applicator.
Clean the On/Standby switch using a dampened cleaning towel. Do not spray or
wet the switch directly.
50TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
If the display i
s very dirty, moisten the wipe with distilled water or a 75%
isopropyl alcohol solution and gently rub the display surface. Avoid u sing excess
force or you may damage the plastic display surface.
Exterior Inspection
Inspect the o
utside of the oscilloscope for damage, wear, and missing part. (See
Table 16.) Immediately repair defects that could cause personal injury or lead
to further damage to the oscilloscope.
Table 16: External inspection checklist
ItemInspect forRepair action
Front and rear case
Front-panel knobsMissing, damaged, or loose knobs
Connectors
Carrying handle, and cabinet feetCorrect operationRepair or replace defective module
Accessories
Cracks, scratches, deformations, damaged
hardware
Broken shells, cracked insulation, and
deformed contacts. Dirt in connectors
Missing items or parts of items, bent pins,
broken or frayed cables, and damaged
connectors
Repair or replace defective module
Repair or replace missing or defective knobs
Repair or replace defective modules. Clear
or wash out dirt.
Repair or replace damaged or missing
items, frayed cables, and defective modules
Interior Inspection
To access the inside of the oscilloscope for inspection and cleaning, refer to the
val Procedures in this section.
Remo
Inspect the internal portions of the oscilloscope for damage and wear. (See
le 17.) Repair any defects immediately.
Tab
If any circuit board is repaire d or replaced, check the table to see if it is necessary
adjust the oscilloscope. (See Table 17.)
to
CAUTION. To prevent damage from electrical arcing, make sure that circuit
boards and components are dry before applying power to the oscilloscope.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual51
Firmly sea
modules with defective wires or cables.
Straighten, repair, or replace defective
hardware.
t connectors. Repair or replace
To clean the oscilloscope interior, perform the following steps:
1. Blow off dust with dry, low-pressure, d eionized air (approximately 9 psi).
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl
alcohol (75% solution) and then wipe with a lint-free cloth dampened with
warm deionized water. A cotton-tipped applicator is useful for cleaning in
narrow spaces and on circuit boards.
lace damaged circuit board.
STOP. If, after performing steps 1 and 2, a module is clean on inspection, skip
steps 3, 4, and 5. Proceed with step 6.
3. If steps 1 and 2 do not remove all the dust or dirt, all modules in the
oscilloscope except the Display Module and Front Panel Module may be
ay washed using a solution of 75% isopropyl alcohol.
spr
During cleaning of the Acquisition board (Main board), make sure that
residue is left in the attenuators. Debris or residue can cause several
no
problems with signal fidelity that cannot be corrected by calibration.
When handling the board, avoid handling the attenuators s ince skin oils
may cause problems.
Extensive cleaning of the Display Module is difficult because water must
not be trapped between the display shield and the display. The polarizers
integrated into the display degrade with long-term exposure to water. The
display shield and display must be replaced as an assembly if significant
moisture or debris become trapped inside. Cotton tipped applicators and
52TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
lint-free rags
the best cleaning with the least opportunity for dam age to the assembly.
The Front-Pan
inside the knob encoders. To avoid this, be careful not to drip liquid or
blow debris into the encoders.
4. Spray wash dirty parts with the 75% isopropyl alcohol solution and wait
60 seconds for the majority of the a lcohol to evaporate.
5. Use hot (49 °C to 60 °C / 120 °F to 140 °F) deionized water to thoroughly
rinse the parts.
6. Dry all parts with low-pressure, deionized air.
7. Dry all co
low-temperature (52 °C to 66 °C / 125 °F to 150 °F) circulating air.
Lubrication. Periodic lubrication is not required for the oscilloscope.
mponents and assemblies in an oven or drying compartment using
moistened in isopropyl alcohol and/or deionized water offer
el module can be damaged by deposition of film or debris
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual53
Maintenance
Removal and Installation Procedures
WARNING. Only qualified personnel should perform service procedures. Before
doing this or a
Summary and Service Safety Summary found at the beginning of this manual. To
prevent damage to electrical components, refer to Preventing ESD. (See page 49,
Preventing ESD.)
Always disconnect the power cord before disassembling any part of the
oscilloscope.
Refer to the module locator drawings to find the modules that you want to remove
or install. (See Figure 5 on page 55.) (See Figure 6.)
Refer to the Inspection and Cleaning procedure if you are disassembling the
instrument for cleaning. (See page 49, Inspection and Cleaning.)
Refer to the Replaceable Parts List for a list of all replaceable modules and their
part numbers.
ny other procedure in this manual, read the General Safety
Summary of Procedures
Refer to the Exploded View Diagrams for each procedure, as indicated in the table.
Complete disassembly is best achieved by following the procedures in the order
n in the following table:
give
ProcedureBegins on page
Feet
Rear
p Feet
Fli
nt-Panel Knobs
Fro
Rear Case(See page 58, Rear Case.)
ont Feet
Fr
Power Supply Module(See page 60, Power Supply Module.)
nternal Assembly
I
Front-Panel Cable(See page 62, Front-Panel Cable.)
Main Board Module
Display Module
Front-Panel Module
Keypad
(See page 56, Rear Feet.)
(See page 57, Flip Feet.)
(See page 57, Front-Panel Knobs.)
(See page 59, Front Feet.)
(See page 61, Internal Assembly.)
(See page 63, Main Board Module (Serial number C030000or later).)
(See page 66, Display Module.)
(See page 67, Front-Panel Module.)
(See page 68, Keypad.)
54TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Figure 5: Locator for trim and cabinet removal (4-channel m odel shown)
Figure 6: Locator for internal modules (4-channel model shown)
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual55
Maintenance
Required Tools
Rear Feet
Use the followi
Torque-limiting screwdriver, long shank, 12 in-lb (1.3 N ·m) range with Torx®
T-15 tip; if yo
bit must be a long bit, at least 5 cm (2 in) overall length
Torque-lim
5/8 inch deep socket with wide center hole (S·K Hand Tool Corp., S-K
#40820, or e
Torque wrench to 40 in-lb (4.5 N·m)
Screwdriver with 1/4 inch flat blade
Pliers (all-purpose and needle-nose)
Sharp knife (X-ACTO®knife or equivalent)
3M Electrical Tape 92 with 1-mil polyimide Kapton backing, temperature
rated 180 °C, UL510 Flame Retarded
You w il
Refer to the exploded view diagram. (See page 96.)
l need a flat-blade screwdriver to remove the rear feet.
ng tools to remove and replace all modules in the oscilloscope:
u use a magnetic screwdriver with interchangeable bits, the T-15
iting screwdriver with Torx T-8 bit
quivalent)
Removal.
1. Place the oscilloscope face down on a so ft surface such as an anti-static mat.
2. Use a flat-blade screwdriver to pry out the foot, or twist the foot as you are
removing it so that it disengages from the guides. Use care not to mar the
surface of the rear case.
Installation.
1. Place the oscilloscope face down on a so ft surface such as an anti-static mat.
2. Align the slots of the rear foot with the guides in the rear case and press the
foot in.
56TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Flip Feet
Front-Panel Knobs
You will need a p
Refer to the exploded view diagram. (See page 96.)
Removal.
1. Place the os
2. Grip the flip foot with the pliers as close to the notches as possible and
squeeze gen
3. Twist slightly to swing one hinge free of the slot.
4. Remove the flip foot from the case.
Installa
1. Place the oscilloscope face down on a soft surface such as an anti-static mat.
2. Hold the flip foot with the smooth side facing down.
3. Gently squeeze the foot so that the hinge on one side fits into the slot, and
Refer to the exploded view diagram for the 2-channel models or the 4-channel
models. (Seepage89.) (Seepage92.)
tion.
twist slightly to snap into the slot on the other side.
air of needle-nose pliers to remove the flip feet.
cilloscope face down on a soft surface such as an anti-static mat.
tly on the foot.
Power Button
NOTE. Wrap the plier jaws with tape (such as e lectrical tape) to avoid scratching
the knob.
oval. Firmly grasp the knob (with pliers if necessary) and pull it away from
Rem
the front panel.
Installation. Align the keyed knob with the shaft and press the knob onto the shaft.
Refer to the exploded view diagram for either the 2-channel or the 4-channel
models. (Seepage89.) (Seepage94.)
NOTE. Wrap the plier jaws with tape (such as e lectrical tape) to avoid scratching
the b utton.
Removal. Firm ly grasp the button with pliers and p ull it straight out of the cabinet.
Installation. Press the button onto the power button shaft until it snaps into place.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual57
Maintenance
Rear Case
You will need a t
procedure.
Refer to the ex
Removal.
1. Remove the p
2. Place the oscilloscope face down on a so ft surface such as an anti-static mat.
3. Remove the two ESD protection covers from the rear case to expose two
screws.
4. Remove the five screws located on the rear case. Lift the handle to access the
top two screws.
5. Lift the rear case off of the oscilloscope.
lation.
Instal
1. Place the oscilloscope face down on a soft surface (such as an anti-static mat).
CAUTION. Be careful not to pinch any wires or other parts between the front
ar cases.
and re
orque-limiting Torx T-15 screwdriver and pliers for this
ploded view diagram. (See page 96.)
ower button. (See page 57, Power Button.)
2. Place the rear case o nto the assembled oscilloscope and verify that:
All of the edges align with the front case.
The Li-ION b attery is vertical in the battery location.
The slots a
CAUTION. Avoid cross-threading the screws that attach the rear case to the front
case. An easy way to avoid this is to slip the screw into the hole and, using slight
pressure, slowly rotate the screw counterclockwise until you feel the screw slip
into the screw threads. Then tighten the screw c l ockwise. A properly started screw
will exhibit very little resistance going in.
3. Install the five screws in the rear case. Tilt the handle to access the top two
holes.
4. Install the two ESD protection covers.
5. Install the power button. (See page 57, Power Button.)
lign near the USB connector and near the battery.
58TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Front Feet
You will need a t
procedure.
Refer to the ex
models. (Seepage89.) (Seepage92.)
Removal.
1. Remove the power button and rear case. (See p age 58, Rear Case.)
2. Slide the f
Installation. Slide the feet into the slots provided in the front case.
NOTE. If the front feet are not installed, but are available, they can be inserted
by carefully pinching them and pushing them into their slots. For example, you
may need to do this if the rear case has been removed, and a foot has been rubbed
off and u
nnoticed until the case is being assembled.
orque-limiting Torx T-15 screwdriver and pliers for this
ploded view diagram for the 2-channel models or the 4-channel
ront feet out of the front case.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual59
Maintenance
Power Supply Module
Use a torque-li
Refer to the exploded view diagram. (See page 94.)
Removal.
1. Remove the power button and rear case. (See page 58, Rear Case.)
2. Disconnect the following wires and cables:
The ground wire on the power supply module from the chassis ground lug.
The power supply cable on the Main board at J101. Press down on the
latch while pulling up on the cable.
3. Remove the five screws that connect the power supply module to the chassis
and lift the module from the internal assembly. The two screws for the AC
connector are longer than the others.
lation.
Instal
1. Align and place the power supply module into the internal assembly.
2. Install the two longer screws with thread lock into the IEC AC power
connector mounting holes.
miting Torx T-15 screwdriver and pliers for this procedure.
3. Install the three shorter screws through the power supply and into the chassis.
4. Connect the following wire and cables:
The ground wire on the power supply module to the chassis ground lug.
You may need to use pliers to secure the wire in place.
The power supply cable to the Main board at J101. Position the cable to
keep it from contacting the boards and chassis.
NOTE. Twist the power supply cable until it is tight; loops are okay. This keeps
the cable bundle together.
5. Assemble the oscilloscope by performing the installation procedure in reverse
order for each mod ule that was remove d.
60TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Internal Assembly
Youwillneedat
this procedure.
Refer to the ex
Removal.
1. Remove the f
2. Remove the power button and rear case. (See p age 58, Rear Case.)
3. Remove the five screws attaching the internal assembly to the front case.
4. Lift the entire internal assembly (consisting of the chassis, all of the boards,
and the display screen) out of the front case.
NOTE. The switch keypad will most likely remain inside the front case. You do not
need to r
Installation.
1. Install the switch keypad mat in the front case if it was removed.
2. Place the internal assembly in the front case.
emove it with the inter nal assembly.
orque-limiting Torx T-15 screwdriver and a pair of pliers for
ploded view diagram. (See page 96.)
ront-panel knobs. (See page 57, Front-Panel Knobs.)
3. Install the five screws to attach the internal assembly to the front case.
4. Asse
mble the oscilloscope by performing the installation procedure in reverse
order for each module that was removed.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual61
Maintenance
Front-Panel Cable
You will need a t
procedure.
Refer to the ex
Removal.
1. Remove the f
(See page 92.)
2. Remove the
3. Remove the internal assembly. (See page 61, Internal Assembly.)
4. Remove the four screws attaching the front panel board to the chassis.
5. Open the wire clamp to release the front-panel cable.
6. Disconnect the front-panel cable at J202 on the Main board by pulling straight
up from the connector.
7. Pull on the front-panel cable assembly to disconnect the front-panel cable
from J1 on the Front-Panel module.
Installation.
orque-limiting Torx T-15 screwdriver and pliers for this
ploded view diagram. (See page 94.)
ront-panel knobs. (See page 89.)
power button and rear case. (See page 58, Rear Case.)
1. Connect the front-panel cable to the Front-Panel module at J1 through the
appropriate slot on the chassis.
2. Position the cable to keep it from contacting the boards and chassis, and
attach it using the cable tie.
NOTE. Twist the front-panel cable until it is tight; loops are okay. This keeps the
ble bundle together.
ca
3. Connect the front-panel cable at J202 on the Main board by pushing the cable
traight down into the connector.
s
4. Install the four screw s to attach the front panel m odule to the chassis.
5. Assemble the oscilloscope by performing the installation procedure for each
module that was removed.
62TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Main Board Module (Serial
number C030000 or later)
You will need a t
wrench, and a 5/8 inch deep socket for this procedure.
Refer to the ex
CAUTION. The Main board module is particularly sensitive to ESD. Read the
Preventing ESD section before removing or installing the Main board. (See
page 49, Preventing ESD.)
Removal.
1. Remove the power button and rear case. (See p age 58, Rear Case.)
Remove th
Assembly.)
2. Disconn
The power supply cable at J101. Press down on the latch while pulling
up on th
The display ribbon cable (J1501 for 2 Ch, J2002 for 4 Ch)
The backlight wire cable (J150 2 for 2 Ch, J2001 for 4 Ch)
orque-limiting Torx T-15 long shank screwdriver, a torque
ploded view diagram. (See page 94.)
e internal assembly from the front case. (See page 61, Internal
ect the following cables from the Main board:
ecable
The front-panel cable at J202
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual63
Maintenance
Figure 7: Instrument cable locations (Main Board SN C030000 or later)
3. Use the screwdriver to remove the three screws that attach the Main board
le to the internal assembly.
modu
4. Use the socket to remove the nuts and washers from the BNC connectors.
5. Pull the board out from the front slots of the internal assembly.
stallation.
In
1. Work from the bottom of the internal assembly and place the front of the Main
board module into the slots on the internal assembly near the BNC connectors.
2. Use the deep socket to attach the washers and nuts to each BNC connector so
that the Main board is securely attached to the chassis.
64TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Support the BNC
nuts to 40 in-lbs (4.5 N@m) using the torque wrench.
3. Install the th
assembly.
4. Connect the
The power supply cable at J101
The display ribbon cable (J1501 for 2 Ch, J2002 for 4 Ch)
The backlight wire cable (J150 2 for 2 Ch, J2001 for 4 Ch)
The front-panel cable at J202
5. Assemble the oscilloscope by performing the installation procedure in reverse
order for each module that was removed.
to prevent it from twisting the Main board, and torque the
ree screws attaching the Main board module to the internal
following cables on the Main board:
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual65
Maintenance
Display Module
You will need a t
T-8 screwdriver, and pliers for this procedure.
Refer to the ex
(See page 89.) (See page 92.)
CAUTION. Do not separate the display shield from the display module. Removing
the shield will destroy the attaching gasket.
Removal.
1. Remove th
2. Remove th e internal ass embly from the front case. (See page 61, Internal
Assembl
3. Disconnect the display backlight wire cable from the Main board. (See
page 64.)
4. Disconnect the display ribbon cable from the Main board. (See page 64.)
5. Remove the power supply module. (See page 60, Power Supply Module.)
6. Remove the four screws attaching the display module to the chassis. The
screws were exposed when the power supply module was removed.
orque-limiting Torx T-15 screwdriver, a torque-limiting Torx
ploded view diagram for the 2-channel or the 4-channel models.
e power button and rear case. (See page 96.)
y.)
7. Remove the display from the chassis. Handle this assembly carefully.
Installation.
1. Position the Display module on the chassis.
2. Use the Torx T-8 screwdriver and install the four screws to a ttach the display
module to the chassis. Torque the nuts to 3 in-lbs (0.33 N@m) using the
torque wrench (2.864 ±0.26 in-lbs).
3. Route the backlight and display cables through the appropriate openings in
the internal assembly. (See page 64.)
4. Connect the backlight wire cable to the Main board (J1502 for 2 Ch, J2001
for 4 Ch).
5. Connect the display ribbon cable to the Main board (J1501 for 2 Ch, J2002
for 4 Ch).
6. Assemble the oscilloscope by performing the installation procedure in reverse
order for each mod ule that was remove d.
66TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Front-Panel Module
You will need a t
for this procedure.
Refer to the ex
models. (Seepage89.) (SeeFigure13.)
Removal.
1. Remove the power button and rear case. (See page 96.)
2. Remove the
Assembly.)
3. Disconne
holder through which the cable is routed. (See page 64.)
4. Remove t
assembly.
5. Pull th
6. Pull on the front-panel cable assembly to disconnect the front-panel cablefrom J
Installation.
orque-limiting Torx T-15 screwdriver and a flat-blade screwdriver
ploded view diagram for the 2-channel models or the 4-channel
internal assembly from the front case. (See page 61, Internal
ct the front-panel cable from J202 on the Main board. Open the wire
he five screws attaching the Front-Panel module to the internal
e module away from the internal assembly.
1 on the Front-Panel module.
1. Conn
2. Pos
3. Ins
4. As
ect the front-panel cable to J1 on the Front-Panel module b y pushing the
cable straight down into the connector.
ition the Front-Panel module on the chassis; route the front-panel cable
through the appropriate opening in the chassis. (See page 64.)
tall the five screws to attach the Front-Panel module to the internal
assembly.
semble the oscilloscope by performing the installation procedure in reverse
order for each module that was removed.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual67
Maintenance
Keypad
You will need a t
Refer to the exploded view diagram for the 2-channel m odels or the 4-channel
models. (See p
Removal.
1. Remove the p
2. Remove th e internal ass embly from the front case. (See page 61, Internal
Assembly.
3. Lift the keypad switch out of the front case.
CAUTION.
and dust inhibit good electrical contact.
ation.
Install
1. Place the front case face down on a soft surface (such as an anti-static mat).
2. Align the switch keypad into the front case using the guide posts and the
holes in the keypad.
orque-limiting Torx T-15 screwdriver for this procedure.
age 89.)(See Figure 13.)
ower button and rear case. (See page 58, Rear Case.)
)
Keep the electrical contacts on the back of the keypad clean. Skin oils
CAUTION. Keep the electrical contacts on the back of the switch keypad clean.
oils and dust inhibit good electrical contact.
Skin
3. Check that all buttons are properly seated before proceeding.
4. Assemble the oscilloscope by performing the installation procedure in reverse
order for each mod ule that was remove d.
68TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Troubleshooting
Maintenance
This section contains information and procedures to help you isolate a defective
module in the oscilloscope.
WAR N ING. Before performing this or any other procedure in this manual, read the
General Safety Summary and Service Safety Summary found at the beginning of
this manual
. Also, to prevent possible injury or damage to electrical components,
read Preventing ESD. (See page 49, Preventing ESD.)
Adjustment After Repair
If you replace the Main board or power supply, you m ust adjust the oscilloscope
after repair. Refer to the chapter Adjustment Procedures.
Required Tools and
Use the f
ollowing tools and equipment to troubleshoot the oscilloscope:
Equipment
Table 18: Tools and equipment
Tools and equipmentExample
Oscilloscope with 1X/10X and 100X
probes
DMM3.5 or greater digit DMM
USB flash d rive
USB type A to type B cableTektronix part numbers: 174-4401-00, 0.91 meters (3 feet) 174-5194-00, 1.83 meters (6 feet)
PC with USB host post and
with TekVISA (OpenChoice PC
unications) installed
Comm
Troubleshooting Tree
Tektronix TDS2000C or TDS1000C-EDU series oscilloscope with standard accessory 10X
probe (TPP0201 or TPP0101) and optional 100X probe (P5100 or P5200)
onix part number 119-7276-00, 128 MB
Tektr
PC with USB communication port
solate a problem to a specific troubleshooting procedure, use the
To i
troubleshooting tree that is shown on the next four pages. Detailed procedures
follow the tree.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual69
Maintenance
gure 8: Oscilloscope troubleshooting tree (page 1 of 4)
Fi
70TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Figure 9: Oscilloscope troubleshooting tree (page 2 of 4)
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual71
Maintenance
Figure 10: Oscilloscope troubleshooting tree (page 3 of 4)
72TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual73
Maintenance
Probe Comp Output
When the oscill
oscope is on and operating properly, the Probe Comp output
should generate a squa re wave, approximately 5 V in amplitude, at a 1 kHz
frequency. Use the oscilloscope and use a 10X TPP0101 or TPP0201 probe to
probe this output.
As shown in the following table, there are two cases of defects. Either the 1 kHz
signal is on and the power supply and Main board are both active and functionally
capable of d
oing acquisition, or one of those two modules is not performing
correctly. While many different possible frequencies exist, all involve detected
failures on the Main board due to Main board failure or power s upply failure.
Table 19: Probe Comp output
Signal at Probe CompPossible problem
1 kHz signal, no display
Non-1 kHz signal
No signal
Refer to Troubleshooting the Display. (See page 76,Troubleshooting the Display.)
Refer to Troubleshooting the Main Board. (See page 80,
Troubleshooting the Main Board.)
Refer to Troubleshooting the Power Supply. (See page 75,
Troubleshooting the Power Supply.)
74TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Troubleshooting the Power
Supply
To troubleshoo
1. Remove the rear case. (See page 58, Rear Case.)
t the power supply, follow these steps:
2. Use the test oscilloscope to measure the voltages from the power supply
module at J101 on the Main board module. The following table lists the
voltages yo
u should expect to see.
Table 20: Troubleshooting the power supply
SupplyVoltage range
35 V
+6 V5.5 V to 6.5 V1.0 A43, 5, 8, 10+5 V
+3.3V3.0Vto3.6V.
-4 V-5.0 V to -3.5 V0.8 A93, 5, 8, 10-2.5 V
Line
trigger
1
30 V
DC
Pin 2 connected to +3.3 V on the Main board.
Requires minimum load to maintain regulation.
-2 V to 6 V open circuit.
±1 diode drop when attached to the Main board.
With 3 W minimum load; still functions with no load.
1
to 40 VDCfloating.
Maximum
current draw
15 mA12
1.8 A6, 73, 5, 8, 103.3 V, +2.5 V
1 mA113, 5, 8, 10Line trigger
J101 pow er
pins
J101 return
pins
Derived
supplies
+28 V LCD
3. If all of the voltages are present, the Main board is probably defective.
Replace it.
4. If all or some of the voltages are missing, turn off the oscilloscope and
disconnect the cable at connector J101 on the Main board module.
5. Turn on the oscilloscope and check the voltages at the loose end of the cable
disconnected from J101, checking for the same voltages as in step 2.
6. If all of the voltages are missing, check the line fuse on the power supply
module. If the fuse is defective, replace the power supply module.
7. If the fuse is okay, and all or some of the voltages are missing, the power
supply module is probably defective. Replace it.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual75
Maintenance
Troubleshooting the
Display
To troubleshoo
1. Remove the rear case. (See page 58, Rear Case.)
2. Remove th e internal ass embly from the front case. (See page 61, Internal
Assembly.)
3. Remove the Display module. (See page 66, Display Module.)
4. Reseat the display cable to the Display m odule, making sure that the c able is
fully seated in the connector.
5. Reconnect the Display module to the internal assembly.
6. Reattach the front panel.
7. Reconnect the backlight and display cables to the Main board.
8. Power on
local Tektronix Customer Service Center for assistance.
t a nonfunctional display, follow these steps:
the oscilloscope. If the display still does not work, contact your
76TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Troubleshooting the Front
Panel
To troubleshoo
t an improperly functioning front pane l, follow these steps:
1. To execute the internal diagnostic tests, remove all input signals from the
oscilloscope
, and enable the Service menu. (See page 42, Enable the Service
Menu.)
2. Push the UTI
LITY button and push the Service option button. The following
options appear:
Table 21: Troubleshooting the front panel – service
MenuComments
Factory CalSteps through the calibration routine steps.
Total Ini t
Service Diag.
Service Mode OffDisables the Service menu.
Writes factory default settings to all saved setups, writes factory
default d
and performs the same actions as the front-panel Default Setup
button.
Displays the service diagnostics menus.
ata to all reference waveforms, turns off Service Mode,
3. Select Service Diag. The following options appear:
22: Troubleshooting the front panel – Service Diag
Table
MenuComments
Mfg. TestsDisplays a front-panel simulation used to check that all front panel
controls are operational.
Loop Diag.
ek/Poke
Pe
ear Error Log
Cl
Opens a menu that enables you to rerun a subset of the power-on
ftests.
sel
For engineering use only.
moves all Error Log entries from memory and sets the power
Re
cycle count to zero.
4. Select Mfg. Tests to start the front-panel controls test. The following options
appear:
Table 23: Troubleshooting the front panel – Mfg. test
MenuComments
Test FP
Test ScreenChecks the display screen pixels with different patterns from each
ContrastAllows you to check the display contrast operation across 100%
Displays a front-panel simulation used to check the operation of
all front panel controls.
All buttons must be pushed and all knobs must be turned in both
directions before you can exit this test.
push of the button.
of the hardware range.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual77
Maintenance
Table 23: Troubleshooting the front panel – Mfg. test (cont.)
MenuComments
Warm-up Timer
Stop TestExits the front-panel test (Test FP) and the screen test (Test
Ensures a ten-minute warm-up time for the oscilloscope before
executing the calibration procedure.
Screen).
5. Select Te s t FP to start the front-panel controls test.
6. Push each button and rotate each knob to verify the operation of all controls.
If all buttons and knobs have passed the test, pushing the lower menu button
(EXIT) exits the test.
7. If any button fails to operate, inspect and/or clean the switch keypad a nd
the front panel board contacts. If cleaning does not solve the problem, the
switch
keypad, front panel board, or M ain board may be defective. Follow the
remaining steps in this procedure to isolate the problem.
8. Repla
ce the switch keypad and retest the buttons. If the problem persists,
go to step 9.
9. Use t
he test oscilloscope and the TPP0201 probe to probe the front-panel cable
connector at J202 on the Main board (2-channel models) or J1 (4-channel
models). The following table describes the signals you should expect to see.
Table 24: Expected signals at J202
Signal at J202Description
1 FPLE D_P
2 FPCLR_12_PClear Front Panel Scan line A 6.44 kHz. Low pulsing high with
3 FPCLK_12_PFront Panel Clock line A 405 kHz. High with 50 ns pulses going
4 FPDATA12P
5 GND0 V (ground)
6 FPTHERMSTR_PFront Panel Temp Sense About 2 VDCdepending on temperature,
7 FP _PWR5.0 V
8 FPTHERMGND_PTemp Sense Return Ground.
9 GND0 V (ground)
LED latch line 6.44 kHz. High with 120 ns low pulses. Location of
pulse relative to FPCLR_12_P determines which lights are lit.
50 ns pulses. Determines beginning of a scan sequence for first
half of front panel.
low. Each clock after FPCLR_12_P increments the scan location
by one.
Front P anel Scan Data line A No fixed pattern. Stable when
triggered from FPCLR_12_P. Pattern depends on rotation of
knobs being scanned.
contrast setting, and display type.
DC
78TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Table 24: Expected signals at J202 (cont.)
Signal at J202Description
10 FPDATA_34_P
1
Front Panel Scan Data line B No fixed pattern. Stable when
triggered from FPCLR_34_P. Pattern depends on rotation of
knobs being scanned.
11 FPCLR_34_P
2
Front Panel Clock line B 405 kHz. High with 50 ns pulses going
low. Each clock after FPCLR_34_P increments the scan location
by one.
12 FPCLK_34_P
2
Clear Front Panel Scan line B 6.44 kHz. Low pulsing high with
50 ns pulses. Determines beginning of a scan sequence for
second half of front panel.
1
Unused and open on 2-channel m odels.
2
Unused and tied low on 2-channel models.
11. If all of the signals are present, the Front-Panel board is probably defective.
Replace it.
12. If some or all of the signals are missing, continue with step 13.
13. Turn off the oscilloscope and disconnect the front-panel cable from the Main
board at connector J202.
14. Turn on the oscilloscope and check connector J202 for the same signals as in
step 9.
15. If some or all of the signals are missing, the Main board is probably defective.
Replace it.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual79
Maintenance
Troubleshooting the Main
Board
To troubleshoo
1. Remove the rear case. (See page 58, Rear Case.)
t the Main board (acquisition module), follow these steps:
2. Use the test oscilloscope to measure the voltages from the power supply
module at J101 on the Main board module. The follow ing table lists the
voltages yo
ushouldexpecttosee.
Table 25: Troubleshooting the Main board
SupplyVoltag e range
35 V
+6 V5.5 V to 6.5 V1.0 A43, 5, 8, 10+5 V
+3.3 V3.0 V to 3.6 V. Requires minimum load to
-4 V-5.0 V to -3.5 V0.8 A93, 5, 8, 10-2.5 V
Line
trigger
1
With3W
30 V
DC
to +3.3 V on the Main board.
maintain regulation.
-2 V to 6 V open circuit.
±1 diode drop when attached to the Main
board.
minimum load; still functions with no load.
1
to 40 VDCfloating. Pin 2 connected
Maximum
current draw
15 mA12
1.8A6,73,5,8,103.3V,+2.5V
1 m A113, 5, 8, 10Line trigger
J101 power
pins
J101 return
pins
Derived
supplies
+28 V LCD
Running Diagnostics
Troubleshooting Input
nections
Con
3. If all of the voltages are present, the Main board is probably defective.
ce it.
Repla
The oscilloscope runs an extensive self-diagnostic routine at every power-on.
Running the diagnostics from the Service menu will provide no additional
rmation and therefore is not needed. The menu selections are only used during
info
manufacturing of the oscilloscope.
Follow these steps to troubleshoot the input connections only if the oscilloscope
pears to function normally in every way, but an input signal is not getting into
ap
the oscilloscope as expected.
80TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Maintenance
Troubleshooting the USB
Interface
Using the Error Log
1. Remove the rear
case. (Seepage58,Rear Case.)
2. Check that the coaxial connections to the back side of the BNC connectors are
intact. Use the DMM to measure continuity from the front side of the BNC
connector to the point where it attaches to the Main board.
NOTE. The ground side of the BNC has thre
e connections (two mounting lugs
and one signal wire) while the center connection has one. Incomplete or poor
connection of the ground side may cause AC signal quality problems without
affecting DC measurement capability.
3. If no problem is found and the trouble persists, the Main board is probably
defective in some way that the internal diagnostics cannot detect. Replace it.
To troubleshoot the USB interface, if the acquisition is working, and the USB is
not, replace the Main board.
Failures that occur during the power-up diagnostics or normal oscilloscope
operation place an entry in the error log.
Accessing the error log. The error log list can be accessed anytime by pushing the
UTILITY button and selecting Error Log.
The following table lists the error c odes and probable causes. (Se e Table 26.)
Table 26: List of error codes
Error codeD escriptionProbable cause
0 DIDiagnostics internal error
1DI
2DI
3DI
4DI
5DI
6DI
7DI
12 DI
13 DI
14 DI
15 DI
16 DI
17 DI
255
ADG522 diagnostics failedMain board or power supply failure
PROM diagnostics failedMain board or power supply failure
MEMORY diagnostics failedMain board or power supply failure
Ch 1 diagnostics failedMain board or power supply failure
Ch 2 diagnostics failedMain board or power supply failure
Line trigger diagnostics failedMain board or power supply failure
ADG421 diagnostics failedMain board or power supply failure
Ch 3 diagnostics failedMain board or power supply failure
Ch 4 diagnostics failedMain board or power supply failure
ADG421 #2 diagnostics failedMain board or power supply failure
ADG522 #2 diagnostics failedMain board or power supply failure
DRAM2 diagnostics failedMain board or power supply failure
USB diagnostics failedMain board or power supply failure
SoftwareContact Tektronix
Contact Tektronix
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual81
Maintenance
Clearing the er
steps:
1. Enable the Ser
2. Push the UTILITY button and push the Service option button to display
the Service
3. Push the Service Diag. option button to display the Diagnostics menu.
4. Push the Clear Error Log option button to clear the error log of past failures.
This also sets the Power Cycle count to zero.
ror log. To cle ar the error log after fixing a problem, follow these
vice menus. (See page 42, Enable the Service Menu.)
menu.
82TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Repackaging Instructions
Use these procedures to repackage the oscilloscope for shipment or storage.
Maintenance
Packaging
Storage
Use a corruga
(275 pounds), and with an inside dimension at least 15.25 cm (six inches) greater
than the oscilloscope dimensions. Make s ur e that the oscilloscope is adequately
padded with a cushioning material like foam, paper, or bubble wrap. The
oscilloscope should not move around inside the carton when the carton is sealed.
If the oscilloscope is being shipped to a Tektronix Service Center, enclose the
following information:
The owner address
Name and phone number of a contact person
Type and serial number of the oscilloscope
Reason
A complete description of the service required
Seal the shipping carton with an industrial stapler or strapping tape.
Mark the address of the Tektronix Service Center and also your own return
address on the shipping carton in two prominent locations.
The oscilloscope should be stored in a clean, dry environment. The following
environmental characteristics apply for both shipping and storage (oscilloscope
not operating):
ted cardboard shipping carton having a test strength of at least 125 kg
for returning
Temperature range: -40 °C to +71 °C (-40 °F to 159.8 °F)
Altitude: To 3,000 m (approximately 10,000 ft)
The chapter on Specifications lists the environmental characteristics. (See page 1,Specifications.)
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual83
Maintenance
84TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Diagrams
This section describes the electrical operation o f the TDS2000C and
TDS1000C-EDU series oscilloscopes using the major circuit blocks or modules.
The oscilloscope module interconnections are shown in the figure. (See Figure 11.)
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual85
Diagrams
The shaded area
4-channel oscilloscopes.
in the figure represents modules and circuits used only in
gure 11: TDS2000C and TDS1000C-EDU series block diagram
Fi
86TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Replaceable Parts
This chapter contains a list of the replaceable modules for the TDS2000C and
TDS1000C-EDU oscilloscopes. Use this list to identify a nd order replacement
parts.
Parts Ordering Information
Replacement parts are available through your local Tektronix field office or
representative.
Changes to Tektronix products are sometimes made to accommodate im proved
components as they become available and to give you the benefit of the latest
improvements. Therefore, when ordering parts, it is important to include the
following information in your order:
Part number
Instrument type or model number
Instrument serial number
Instrument modification number, if applicable
Part Number Revision
Level
Module Servicing
If you order a part that has been replaced with a different or improved part, your
local Tektronix field office or representative will contact you concerning any
change in part number.
Tektronix part numbers contain two digits that show the revision level of the
part. For some parts in this manual, you will find the letters XX in place of the
revision level number.
When you order parts, Tektronix will provide you with the most current part for
your product type, serial number, and modification (if applicable). At the time of
your order, Tektronix will determine the part number revision level needed for
your product, based on the information you provide.
Modules can be serviced by selecting one of the following two options. Contact
your local Tektronix service cente r or representative for repair assistance.
Module exchange. In some cases you may exchange your module for a
remanufactured module.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual87
Replaceable Parts
These modules c
ost significantly less than new modules and meet the same factory
specifications. For more information about the module exchange program, call
1-800-833-9200 and select option 2.
New modules. You m ay purchase replacement modules in the same way as other
replacement parts.
Using the Replaceable Parts List
This section contains a list of the mechanical and/or electrical components that are
replaceab
parts. The follow ing table describes each column in the parts list.
Table 27: Replaceable parts list
ColumnColumn nameDescription
1
2Tektro
3 and 4
5
6
Figure & Index NumberItems in this section are referenced by
Serial NumberColumn three indicates the serial number at
QtyThis indicates the quantity of parts used.
Name & DescriptionAn item name is separated from the
le for the oscilloscope. Use this list to identify and order replacement
nd index numbers to the exploded
figure a
view illustrations that follow.
nix Part Number
s part number when ordering
Use thi
replacement parts from Tektronix.
which the part was first effective. Column
ndicates the serial number at which the
four i
part was discontinued. No entries indicates
the part is good for all serial numbers.
cription by a colon (:). Because of space
des
limitations, an item name may sometimes
appear as incomplete. Use the U.S. Federal
alog handbook H6-1 for further item
Cat