CHAPTER 1:INTRODUCTION
CHAPTER 2:SAMPLE ANALYSIS
CHAPTER 3: DISPLAY AND PROCESSING OF ANALYSIS RESULTS
CHAPTER 4: MAINTENANCE AND SUPPLIES REPLACEMENT
CHAPTER 5: QUALITY CONTROL
CHAPTER 6: CALIBRATION
CHAPTER 7: TROUBLESHOOTING
CHAPTER 8: ADJUSTMENT
CHAPTER 9: FUNCTIONAL DESCRIPTION
CHAPTER 10: INSTRUMENT SETUP
APPENDIX A: INSTALLATION
APPENDIX B: TECHNICAL INFORMATION
INDEX
SYSMEX CORPORATION
KOBE, JAPAN
Copyright 1997 - 2000 by SYSMEX CORPORATION
All rights reserved. No part of this Operator’s Manual may beCode No. 461-2261-1
reproduced in any form or by any means whatsoever withoutPRINTED IN JAPAN
prior written permission of SYSMEX CORPORATION.Date of Last Revision: January 2000
• Sysmex is a registered trademark of SYSMEX CORPORATION.
• CELLCLEAN, CELLPACK, EIGHTCHECK-3WP, STROMATOLYSER-WH are
trademarks of SYSMEX CORPORATION.
• Cubitainer is a registered trademark of Hedwin Corporation.
• Teflon is a registered trademark of E.I. du Pont de Nemours & Co., Inc.
• VENOJECT is a registered trademark of Terumo Corporation.
• Other trademarks referenced are property of their respective owners.
• It is prohibited to reproduce part or all of the contents of this Manual without permission.
• The display screens carried in this Manual may in some cases differ from actual screens.
• We reserve the right to make further improvements and incorporate them in our products,
which then will have some points that differ from descriptions in this Manual.
• Patient names and doctor names are entered for information and illustration purposes only,
and do not imply real specific persons.
Sysmex KX-21 Operator’s Manual -- Revised October 1998
I
RECEIVING INSTRUCTIONS
The KX-21 has been thoroughly tested before shipment, and has been packaged
carefully to prevent damage from shipping and handling. Reagents and options have
also been sent and will arrive at approximately the same time as the analyzer. Follow
these guidelines when the system arrives:
•Check to see that the arrows on the sides of the packages are pointing up. If the
arrows do not point up, remark this information on the bill of lading.
•Visually inspect the outside of the package for rips, dents, or possible shipping
damage. Document any sign of damage on the bill of lading, regardless of how
insignificant it may appear. This is for your protection!
•Notify your service representative that the KX-21 system and its components
have arrived.
•Wait for your service representative to unpack the system and open the
packages.
•Follow the unpacking and storage instructions provided on the outside of the
package. Special requirements such as refrigeration are clearly marked on the
outside of the carton and will be included in the unpacking instructions and
package inserts.
WARRANTY INFORMATION
All instruments manufactured by Sysmex® are warranted against defective materials
or workmanship for a period of one year commencing on the installation date at the
customer's required location.
This Warranty does not cover any defect, malfunction, or damage due to:
1.Accident, neglect or willful mistreatment of the product
2.Failure to use, operate, service, or maintain the product in accordance with the
applicable Sysmex Operator's Manual
3.Failure to use the appropriate reagents or chemicals specified for the product
Sysmex KX-21 Operator’s Manual -- Revised October 1998
I
ENSURE SAFE OPERATION OF THE INSTRUMENT
Before operating this instrument, carefully read the "Ensure Safe Operation of the Instrument"
and OPERATOR’S MANUAL, and strictly follow the instructions given in them.
This manual carries a variety of illustrations to make sure that the product can be used safely
and correctly, thus preventing you and others from suffering injuries and damage to property.
The illustrations and meaning are described in the following.
Do understand what they mean before proceeding to the text of the MANUAL.
Meaning of Signs
WARNING
CAUTION
•If this sign is ignored and the instrument is operated incorrectly, there is
a potentially hazardous situation which could result in death or serious
injury of an operator, or grave property damage.
•If this sign is ignored and the instrument is operated incorrectly, there is
a potentially hazardous situation which may result in injury of an
operator, adverse effect on output results, or will cause property
damage.
Caution on Diagnosis
CAUTION
•This product is a clinical instrument for screening of abnormalities.
Clinical judgments by physicians should take into account results from
clinical examinations and other test results besides hematology result.
I
Sysmex KX-21 Operator’s Manual -- Revised October 1998
I
WARNING
• In the event the instrument emits abnormal odor or any smoke, turn off the power
immediately and disconnect the power plug from the wall socket.
If the instrument is used continuously in that state, there is a hazard that fire, electrical shock, or
injury may result.
Contact your Sysmex service representative for inspection.
• Take care not to spill blood or reagent, or drop wire staples or paper clips into the
instrument.
Those might cause short circuit or smoke emission. If such trouble should occur, turn off the power
supply immediately and pull off the power plug from the wall socket. Then contact Sysmex service
representative for inspection.
• Do not touch the electrical circuits inside the cover. Especially if your hands are wet,
there is a hazard that electrical shock may result.
• Always wear rubber gloves when performing maintenance work or inspection.
Use specified tools and parts.
After work is over, wash your hands with disinfectant.
There is a possibility that those areas of the hand which came in contact with blood could suffer
infection, electrical shock, or burn.
• Be careful when handling samples.
Always wear rubber gloves; otherwise infection by bacteria could result. If bacteria happen to enter
your eye or a cut, wash it off with plenty of water, and immediately see a doctor.
• When discarding waste liquid, or disassembling/assembling the related parts, do not
touch the waste liquid.
If it is contaminated with blood, infection of bacteria may result. If you should touch the waste liquid
inadvertently, wash it off with disinfectant first, then wash it off with soap.
When Handling Reagent
• If a reagent happens to enter your eye, wash it off immediately using plenty of water, and
take medical treatment at once.
• If you should swallow it inadvertently, call for a doctor immediately, drink plenty of water,
and throw up.
• If it happens to adhere to the hand or the skin of other area, wash it off using plenty of
water.
• When discarding waste liquid and instrument consumable, take proper disposing steps
as medical, ineffective, and industrial wastes.
If they are contaminated with blood, infection of bacteria may result.
• Do not modify the instrument. (Modification is prohibited by Pharmaceutical Affairs Law in
Japan.) Check the regulations that apply to laboratories in your country.
Sysmex KX-21 Operator’s Manual -- Revised October 1998
II
I
WARNING
Power Supply, Connection, and Grounding
• Never put the power plug in any socket other than the specified voltage.
Otherwise, fire or electrical shock will result.
• When installing the instrument, be sure to ground it.
Otherwise, fire or electrical shock will result.
Handling Power Supply Cord
• Take care not to damage the power cord, place a heavy device on it, or pull it forcibly.
Otherwise, the wire may break causing fire or electrical shock.
• When connecting the instrument to a peripheral (host computer), be sure to switch off the
power supply beforehand.
Otherwise, electrical shock or instrument failure may result.
V
Sysmex KX-21 Operator’s Manual -- Revised October 1998
CAUTION
y
g
Use of Reagents
• After unpacking, be sure not to allow dust, dirt, or bacteria to come in touch with
the reagent.
• Do not use reagents which are out of the expiration date.
• Handle a reagent gently to prevent formation of bubbles.
• Take care not to spill a reagent. If it spills, wipe it off immediately using a wet
cloth or the like.
• Follow other instructions described on the Package Insert on each reagent.
Use of Instrument
• When performing maintenance work or inspection, use specified tools and parts.
Do not use substitute parts, or modify the instrument. It is hazardous.
• Do not bring your body or clothes close to the instrument.
• Those who have no or only limited experience in using reagents are
recommended to have guidance or assistance of those with sufficient
experience.
• If the instrument has developed a trouble by any chance, a person in charge of it
should take steps within the range specified in the OPERATOR’S MANUAL. As to
troubles other than mentioned in it, contact Sysmex service representative for
repair.
• Unpacking, installation, and confirmation of initial operation must be done b
Sysmex service representative.
Environment for Use
• Install the instrument in a place which is not subject to water splash.
• Install the instrument in a place which is not subject to adverse effects of high
temperature, high humidity, dust, direct sunlight, etc.
• Do not give the instrument a strong vibration or impact.
• Do not install the instrument near a chemical storage or a place where a gas is
enerated.
Sysmex KX-21 Operator’s Manual -- Revised October 1998
Sysmex KX-21 Operator’s Manual -- Revised January 2000
V
STRUCTURE OF THIS MANUAL
Read this manual carefully so you will be able to use the instrument to its full extent, and
operate it correctly. This manual contains ten Chapters and APPENDICES as listed in the
following:
Chapter 1:Introduction
Chapter 2:Sample Analysis
Chapter 3: Display and Processing of
Analysis Results
Chapter 4: Maintenance and Supplies
Replacement
Chapter 5: Quality ControlDescribes execution procedures of the X control and L-
Describes overview of this instrument, including the
outline of the instrument, operation procedure, and
messages, and cautions during installation.
Describes procedures for start-up of the instrument,
measurement of the sample, and shut-down.
Summarizes the operation method for analyzing the
sample.
Describes the displayed content of the analysis data,
and the process of the latest sample/stored data.
Also describes the manual analysis and the external
output.
Describes scheduled maintenance and replacement
method of supplies such as reagent.
J control for quality control.
Describes procedures for automatic/manual calibration.
Describes error messages and troubleshooting.
Describes the method of pressure adjustment based on
the troubleshooting.
Describes the analyzing principles of this instrument
and names of the components.
Describes the system environment setting such as
date, time, and unit, and the data analysis setting such
as data error judgment.
Describes the installation method of the KX-21.
Appendix B:Technical Information
VI
Technical reference data describing the specification for
host output format.
(The host output function is an option.)
Sysmex KX-21 Operator’s Manual -- Revised October 1998
INTRODUCTION
Thank you for purchasing the Sysmex® Automated Hematology Analyzer KX-21.
Carefully read the OPERATOR'S MANUAL for correct use of the unit.
Keep this MANUAL handy after reading. It will continue to be of your service in finding
specific information about this instrument.
Sysmex KX-21 Operator’s Manual -- Revised October 1998
VII
PREMISES FOR SIGNS
Meaning of Signs
WARNING
CAUTION
CAUTION:•Indicates what we would like you to know to maintain instrument
NOTE:•Indicates information which will come handy in operating the instrument.
•If this sign is ignored and the instrument is operated incorrectly, there is
a potentially hazardous situation which could result in death or serious
injury of an operator, or grave property damage.
•If this sign is ignored and the instrument is operated incorrectly, there is
a potentially hazardous situation which may result in injury of an
operator, adverse effect on output results, or will cause property
damage.
performance and prevent its damage.
Document Conventions
In explaining operation, this manual uses the conventions as shown below.
• The keys on the panel keyboard are expressed within square brackets.
For example: [SELECT], [ENTER], [
]
• The display on LCD appears within quotation marks.
For example: "Stand-by," "WB"
• The name of menu appears within quotation marks.
For example: "2: Quality Control," "6: Settings"
NOTE:• LCD and printing described in this manual may differ from that in
practice.
• Due to the improvement of the product, the content of this manual may
not conform with the product.
VIII
Sysmex KX-21 Operator’s Manual -- Revised October 1998
14. MENU TREE......................................................................1-18
Sysmex KX-21 Operator’s Manual -- Revised October 1998
INTRODUCTION
1.INTRODUCTION
The Sysmex KX-21 is an automatic multi-parameter blood cell counter for in vitro
diagnostic use in clinical laboratories.
The KX-21 processes approximately 60 samples an hour and displays on the LCD screen the
particle distribution curves of WBC, RBC, and platelets, along with data of 18 parameters,
as the analysis results.
Chapter 1 introduces the overview of the instrument, analysis procedure, etc. that we
recommend you to read before using the KX-21. The main contents of Chapter 1 are as
follows:
Overview of Instrument
The important functions of the KX-21 and the options for efficient operation are explained.
Main Points of Analysis Procedure
Explanation is given on the procedure for implementing each analysis mode, descriptions and
functions of the keys on panel keyboard, and the messages on the LCD screen.
Precautions at Time of Installation
Explanation is given on the matters that need to be confirmed before installation, such as
installation space, required equipment, environmental conditions.
Instrument Specifications
The instrument specifications are described.
Menu Tree
This chapter describes the KX-21 menu tree and the corresponding chapters which explain
the usage of the menus.
Sysmex KX-21 Operator’s Manual -- Revised October 19981-1
INTRODUCTION
2.OVERVIEW OF INSTRUMENT
The KX-21 performs speedy and accurate analysis of 18 parameters in blood and detects the
abnormal samples. To assure easy sorting of abnormal samples in the laboratory, the
instrument displays abnormal analysis data with abnormal marks attached on the LCD
screen. Thus displayed analysis data allows detecting those samples which are outside the
tolerance and need further analysis and reconsideration.
The KX-21 employs three detector blocks and two kinds of reagents for blood analysis. The
WBC count is measured by the WBC detector block using the DC detection method. The
RBC count and platelets are taken by the RBC detector block, also using the DC detection
method. The HGB detector block measures the hemoglobin concentration using the noncyanide hemoglobin method.
Figure 1-2-1: Overview of KX-21
1-2Sysmex KX-21 Operator’s Manual -- Revised October 1998
INTRODUCTION
y
(
y
3.OPTION UNITS
This instrument offers several option units to ensure its efficient operation. The options that
can be used with the KX-21 are:
• Serial interface:Outputs analysis data and QC data to the host computer. Ordering
Information: OUTPUT UNIT NO. 1 KX-21 (Part No. 973-4531-1)
4.OUTLINE OF OPERATION
Two analysis modes are available with the KX-21: whole blood mode and pre-diluted mode.
Analysis procedures in these modes are listed below.
Whole blood modePre-diluted mode
Check before turning ON the power.
Turn ON the power.
• Self-check
• Background check
Read
Select whole blood mode.Select pre-diluted mode.
Set sample No.Prepare analysis samples in pre-diluted mode
1:26 dilution).
Set sample No.
Set sample to the sample probe.
Press the start switch.
• Execute analysis.
• End analysis.
Read
Check after analysis.
Execute shutdown.
Turn off the power.
Table 1-4-1: Outline of Operation
• The work by an operator is shown in shaded cells.
• : "Ready" is displayed on the LCD screen. In this state, various operations
Ready
including analysis, settings, and data processing can be performed.
5.AUTOMATIC STOP FUNCTION OF PNEUMATIC UNIT
The KX-21, when left non-operating for 15 minutes, automatically stops the pneumatic unit.
This function saves power consumption and extends component service life. In addition,
with this function, analysis-ready status can be resumed faster than by turning on the power.
It is also possible to manually stop the pneumatic unit through the Select Menu screen.
Sysmex KX-21 Operator’s Manual -- Revised October 19981-3
Sysmex KX-21 Operator’s Manual -- Revised January 2000
INTRODUCTION
6.ANALYSIS PARAMETERS
This instrument analyzes the following parameters using three detector blocks and two kinds
of reagents:
1)Whole WBC (white blood cell) (Analysis principle: DC detection method)
WBC count in 1 µL of whole blood
2)LYM% [W-SCR] (WBC-Small Cell Ratio)
Ratio (%) of lymphocytes (small cells) to whole WBC
3)MXD% [W-MCR] (WBC-Middle Cell Ratio)
Ratio (%) of the summation of basophils, eosinophils and monocytes (middle cells) to
whole WBC
4)NEUT% [W-LCR] (WBC-Large Cell Ratio)
Ratio (%) of neutrophils (large cells) to whole WBC
5)LYM# [W-SCC] (WBC-Small Cell Count)
Absolute count of lymphocytes (small cells) in 1 µL of whole blood
6)MXD# [W-MCC] (WBC-Middle Cell Count)
Absolute count of the basophils, eosinophils and monocytes (middle cells) in 1 µL of
whole blood
7)NEUT# [W-LCC] (WBC-Large Cell Count)
Absolute count of neutrophils (large cells) in 1 µL of whole blood
8)RBC (red blood cell) (Analysis principle: DC detection method)
RBC count in 1 µL of whole blood
9)HGB (Hemoglobin) (Analysis principle: Non-Cyanide hemoglobin analysis method)
Volume (gram) of hemoglobin in 1 dL of whole blood
10)HCT (Hematocrit value) (Analysis principle: RBC pulse height detection method)
Ratio (%) of whole RBC volume in whole blood
11)MCV (Mean RBC volume)
Mean RBC volume (fL) in whole blood, which is calculated by Hct/RBC.
12)MCH (Mean RBC hemoglobin)
Mean hemoglobin volume (pg) per RBC, which is calculated by Hgb/RBC.
13)MCHC (Mean RBC hemoglobin concentration)
Mean hemoglobin concentration (g/dL), which is calculated by Hgb/Hct.
1-4Sysmex KX-21 Operator’s Manual -- Revised October 1998
INTRODUCTION
14)RDW-CV (RBC distribution width - CV)
RBC distribution width (%) calculated from the points defining 68.26% of the entire
area spreading from the peak of the RBC particle distribution curve.
15)RDW-SD (RBC distribution width - SD)
The distribution width (fL) at the height of 20% from the bottom when the peak RBC
particle distribution curve is taken as 100%.
16)PLT (Platelet) (Analysis principle: DC detection method)
Platelet count in 1 µL of whole blood
17)PDW (Platelet distribution width)*
The distribution width (fL) at the height of 20% from the bottom with the peak of
platelet particle distribution curve taken as 100%.
18)MPV (Mean platelet volume)
Mean volume of platelet (fL)
19)P-LCR (Large platelet ratio) *
Ratio (%) of large platelet volume exceeding 12 fL to the platelet volume
NOTE:• When analyzing in the pre-diluted mode, only the CBC8 parameters
are output.
• The RBC distribution width is the switchover type between RDW-SD
and RDW-CV.
* For Investigational Use Only in the United States of America.
Not a reportable parameter.
Sysmex KX-21 Operator’s Manual -- Revised October 19981-5
INTRODUCTION
7.PANEL KEYBOARD
The KX-21 is provided with the 22-key panel keyboard.
6
C
SAMPLE No.
ENTER
SELECT
MODE
HELP
SHUTDOWN
7
4
1
0
89
5
23
-/.
Figure 1-7-1: Panel Keyboard
NameFunction
SAMPLE No.Used to set a sample No. and QC File No.
ENTERUsed to fix a sample No., selected menu, etc.
SELECTUsed to select a menu. Press this key to display the Select Menu screen. When
you press it while the Select Menu screen is displayed, the Analysis screen returns.
MODEUsed to changeover analysis mode (whole blood mode/pre-diluted mode).
HELPUsed when an error has occurred.
SHUTDOWNUsed to execute shutdown program.
0 - 9Used to enter numerics such as a sample No., QC File No., and set value.
−/.Used to enter "−" (hyphen) of a sample No., the decimal point of a set value.
CUsed to delete characters when entering numerics, and stop the alarm.
, Used to select a menu. Each time a key is pressed, the cursor moves to the
previous or the next item.
, Used to select conditions when setting, and to select Manual Discriminator.
Table 1-7-1: Functions of Panel Keyboard
CAUTION:• When the alarm is sounding after an error, etc. occurred, press [C] key
to stop it or press [HELP] key to display the HELP screen. By
pressing [HELP] key, the alarm stops and the HELP screen appears.
While the alarm is sounding, the keys other than [C] key and [HELP]
key cannot be used.
1-6Sysmex KX-21 Operator’s Manual -- Revised October 1998
NameFunction
Next Sample No.Displays the sample No. to be analyzed next.
Next Sample
Analysis Mode
Analysis Progress
Status
Status DisplayStatus of the unit is displayed.
Menu Display AreaUseable menus for each screen are displayed.
Analysis Result
Display Area
Displays the analysis mode for the sample to be analyzed next.
WB: whole blood mode
PD: pre-diluted mode
With the start-to-end analysis progress classified into six steps, the
present progress status is displayed.
The area for displaying analysis result
Table 1-8-1: Contents of Display
Menu Display Area
Sysmex KX-21 Operator’s Manual -- Revised October 19981-7
INTRODUCTION
8.2 Status Display Messages
The KX-21 displays instrument status message on the LCD screen.
This section gives the meanings of status messages displayed on the LCD screen.
Display MessageMeaning
The instrument is making self-check. When
the power is turned ON, the instrument
checks itself automatically. When the power
is turned ON, [00-01] appears briefly in the
right lower corner, which indicates the version
of the unit control program.
Sysmex KX-21
[00-01]
*Auto Rinse*
Please wait.
Please wait.
The hydraulic system is undergoing Auto
Rinse. When the power is turned ON, the
instrument starts self-check and motor check,
followed by auto rinse of the hydraulic
system. Auto rinse is also performed when
"5: Auto Rinse" in the Select menu is
executed and when [SHUTDOWN] key is
pressed. When an automatic background
check shows a background count exceeding
the permissible background count level, rinse
sequence is automatically extended.
1-8Sysmex KX-21 Operator’s Manual -- Revised October 1998
The instrument is ready for pre-diluted mode
analysis. "No." in the left upper corner shows
the sample No. to be analyzed next.
The instrument is ready for X control. "FILE
No. 1" in the left upper corner shows QC file
No.
The instrument is ready for L-J control. "FILE
No. 1" in the left upper corner shows QC file
No.
Sysmex KX-21 Operator’s Manual -- Revised October 19981-9
INTRODUCTION
Change Mode
*
Display MessageMeaning
*
The whole blood mode is turning into the prediluted mode, or the pre-diluted mode is
turning into the whole blood mode.
Please wait.
No.123457 WB
No.123457 PD
No.123457 WB
No.123457 WB
No.123457 WB
No.123457 WB
Aspirating
Aspirating
Analyzing
Rinsing
Replenish Diluent
Not Ready
PU Sleeping
Not Ready
Momentary power failure occurred.
WB
Ready
The sample is being aspirated in whole blood
mode analysis.
The sample is being aspirated in pre-diluted
mode analysis.
Samples are being aspirated, diluted, and
counted in whole blood mode (or pre-diluted
mode).
Hydraulic line in the whole blood mode (or
pre-diluted mode) is being rinsed.
An error occurs, making analysis impossible.
(When an error occurs, the error message
appears in reversed display in the right upper
area of the LCD screen.)
When operation has been suspended over
15 minutes, the compressor power is turned
OFF and this message is displayed. Press
the [START] switch to get ready for analysis.
When the power is turned off without making
shutdown process or when power outage
caused the instrument to stop, this message
appears when the unit is started next time.
When data is automatically output to the host
*
computer, " * " mark lights up in the right
upper corner of the screen. This applies only
when an optional serial interface is
incorporated.
1-10Sysmex KX-21 Operator’s Manual -- Revised October 1998
INTRODUCTION
8.3 LCD Brightness Adjustment
Open the front cover of the instrument. LCD
brightness can be adjusted using the brightness
adjustment knob under the panel keyboard.
Turning to the right brightens the light and turning
to the left dims it.
Light
9 .EMERGENCY STOP PROCEDURE
When there arises the need to stop the instrument urgently because of power outage, etc., in
the laboratory, turn off the instrument power switch.
10. ALARM SOUNDS
Dark
The KX-21 indicates different situations by five kinds of alarm sounds:
1.Key sound (single beep)
Sounds about 0.1 second when a key on the panel keyboard is pressed.
2.Input error sound (short beep)
Sounds about 1 second when a wrong key is pressed on the panel keyboard.
3.Analysis error sound (long beep)
Sounds when an error occurs in the instrument and continues until you press [C] key
or [HELP] key on the panel keyboard.
4.Sound at sample aspiration
Usually: A single "beep" when the start switch is pressed; "beep, beep" when
aspiration is finished.
Sample No. "0" or in pre-diluted mode: Beep sounds (beep, beep, ...) continue from
pressing the start switch until aspiration ends.
11. CONTENTS OF PACKAGE
The instrument is fully inspected before leaving the factory and carefully packed to withstand
shocks in transit. Upon arrival, check the package to see that there is no external damage.
Sysmex service representative will unpack, install the instrument, and make initial settings.
To confirm the contents of the package, refer to "Appendix A: Installation."
Sysmex KX-21 Operator’s Manual -- Revised October 19981-11
INTRODUCTION
12. INSTALLATION ENVIRONMENT
12.1Installation and Relocation
The KX-21 is installed by Sysmex Service representative. In case relocation becomes
necessary after installation, contact your Sysmex service representative.
Pay careful attention to this because if relocation, etc. of the instrument should be conducted
by the customer, resulting in any trouble, warranty would not be applied even in the
warranty period.
1 2 .2Grounding
The instrument power supply cord uses the 3-prong plug. When the power supply socket is
3-prong with grounding, simply plug it to the socket.
117 VAC Spec.220 VAC Spec.240 VAC Spec.
Figure 1-12-1: Plugs
WARNING
NOTE:• The number of power supply sockets required is one.
• Make sure to ground the instrument. Inadequate grounding could cause
electrical shocks.
1-12Sysmex KX-21 Operator’s Manual -- Revised October 1998
INTRODUCTION
12.3Installation Space
To ensure that the instrument fulfills its function, it is important to install it at an appropriate
place:
• Choose a place that is close to the power supply and proper waterway.
• Secure a place spacious enough for maintenance and service. Giving consideration to
heat radiation by the instrument, provide at least 50 cm distance from the wall to side,
rear, and top panels.
• Choose a place where reagent is easy to handle.
The instrument dimensions are shown below. The power supply cord is 1.8 m long, the
waste tube up to 6 m long, and the diluent (CELLPACK) tube up to 2 m long.
Width (mm)Depth (mm)Height (mm)Weight (kg)
Main Unit42035548028
480
420
355
Figure 1-12-2: Instrument Dimensions
CAUTION:• When the diluent (CELLPACK) tube is more than 2 m long, it may be
impossible to have reagent aspirated. Avoid setting CELLPACK at a
level higher than the instrument; otherwise, reagent may flow into the
vacuum line, possibly damaging the instrument.
Sysmex KX-21 Operator’s Manual -- Revised October 19981-13
Loading...
+ 276 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.