Sysmex CA-500 Instruction manual

Page 1
INSTRUCTIONS FOR USE
Automated Blood Coagulation Analyzer
CA-500 series
CHAPTER 1: Introduction CHAPTER 2: Safety Information CHAPTER 3: Design and Function CHAPTER 4: Installation Environment CHAPTER 5: Operation CHAPTER 6: Display and Processing of
Analysis Results CHAPTER 7: Output CHAPTER 8: Quality Control CHAPTER 9: Setting Standard Curve CHAPTER 10: Instrument Setup CHAPTER 11: Maintenance and Supplies
Replacement CHAPTER 12: Troubleshooting CHAPTER 13: Functional Description CHAPTER 14: Technical Information CHAPTER 15: Index CHAPTER 16: Appendix (A)
KOBE, JAPAN
Copyright © 2001-2004 by SYSMEX CORPORATION
All rights reserved. No part of this Instruction for Use may be reproduced in any form or by any means whatsoever without prior written permission of SYSMEX CORPORATION.
Code No. 461-2655-0 PRINTED IN JAPAN Date of Last Revision: April 2004 Software version 00-17 and onwards
Page 2
Table of Contents
1. Introduction .........................................................1-1
1.1 Introduction ........................... ............................................. .....1-1
1.2 Explanation of Signs ...............................................................1-3
1.3 Names .................................................................................... 1-4
1.4 Serial Number..........................................................................1-4
1.5 Revision History ......................................................................1-4
2. Safety Information ...............................................2-1
2.1 Specified Conditions of Use ................................................... 2-1
2.2 General Information ................................................................2-1
2.3 Installation Location ................................................................2-3
2.4 Avoidance of Infections .......................................................... 2-3
2.5 Handling of Reagents ............................................................. 2-4
2.6 Maintenance of the Instrument ............................................... 2-5
2.7 Disposal of Materials .............................................................. 2-5
2.8 Markings on the Instrument .................................................... 2-6
2.9 Personnel ............................................................................... 2-8
2.10 Storage Condition (Transportation)......................................... 2-8
3. Design and Function ........................................... 3-1
3.1 Overview ......................... ................ ................ ................ ........ 3-1
3.2 Operation Flow ....................................................................... 3-8
4. Installation Environment ....................................4-1
4.1 Installation and Relocation ..................................................... 4-1
4.2 Installation Location ................................................................4-1
4.3 Basic Instrument Settings .......................................................4-3
Revised November 2003 - 2.0_en
Sysmex CA-500 series I
Page 3
5. Operation ..............................................................5-1
5.1 Display Screens and Operation Keys .....................................5-1
5.2 Menu Tree ............................. ... ... ... ... .....................................5-3
5.3 Types of Alarm ................................... .... ... ... ... .... ....................5-5
5.4 Inspection before Turning ON the Power ...............................5-5
5.5 Turn ON the Power .................................................................5-7
5.6 Prepare Reagents ...................................................................5-8
5.7 Set Reaction Tubes ..............................................................5-13
5.8 Confirm Standard Curve .......................................................5-14
5.9 Execute Quality Control ........................................................5-15
5.10 Prepare Samples ..................................................................5-15
5.11 Set Sample Nos. ...................................................................5-19
5.12 Manual Inquiry ...................................................................... 5-22
5.13 Automatic Inquiry ................................................... ...............5-22
5.14 Start Analysis ......................... ... ... ... ... .... ... ... ... ......................5-24
5.15 Automatic Sensitivity Adjustment of the Detector
(for CA-530, CA-540, CA-550 and CA-560 only) ..................5-26
5.16 Display Analysis Result ........................................................5-27
5.17 Interrupt Analysis ..................................................................5-28
5.18 Add Samples ........................................................................5-29
5.19 Analyze STAT Sample .......................................................... 5-30
5.20 Emergency Stop ................................................................... 5-31
5.21 Shutdown ........................................................ ...................... 5-33
6. Display and Processing of Analysis Results ....6-1
6.1 List Display/Graphic Display ...................................................6-1
6.2 Search .................................................................................... 6-6
6.3 Sort in Sequence of Sample ID Nos. and Analyses ........... ... . 6-9
6.4 Select Display ......................................................................... 6-9
6.5 Edit ID No. ............................................................................6-11
6.6 Deletion ................................................................................. 6-12
7. Output ...................................................................7-1
7.1 Automatic Printout of Analysis Data .......................................7-1
7.2 Output of Analysis Data ..........................................................7-1
7.3 Example of Printout ................................................................7-3
8. Quality Control .....................................................8-1
8.1 Quality Control Methods .........................................................8-1
8.2 QC File Setting .......................................................................8-1
8.3 Execute Quality Control ..........................................................8-5
8.4 Display QC Charts .................................................................. 8-5
8.5 Delete QC File ........................................................................8-7
8.6 Delete QC Data ......................................................................8-8
8.7 Print QC data .......................................................................... 8-9
II Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 4
9. Setting Standard Curve ...................................... 9-1
9.1 Display Standard Curve ..................... ... .... ... ... ... ... .... ... ... ... .... . 9-1
9.2 Standard Curve Analysis ................ ...... .... ... ... ... ... .... ... ... ... .... . 9-3
9.3 INR Manual Dilution Analysis ................................................ . 9-6
9.4 Manual Entry ..........................................................................9-8
9.5 Set Reagent Information ........................... ................ .............. 9-9
9.6 Set Calculation Parameters .................................................. 9-10
9.7 Print Standard Curve ............................................................ 9-13
10. Instrument Setup ...............................................10-1
10.1 General Information ..............................................................10-1
10.2 Setup of Automatic Transfer/Printout ................................... 10-2
10.3 Judgment on Analysis Result ............................................... 10-4
10.4 Replication Range .......... ... ... ... .... ... ... ... .... ... ... ... ... .... ...... ... ... 10-6
10.5 Report Limit ....... ... ... .... ... ... ................................................... 10-7
10.6 Setup of Test Name ..............................................................10-8
10.7 Reagent Name ..................................................................... 10-9
10.8 Test Protocol ............................................... ... ... ... .... ...... ... . 10-10
10.9 Replication ............................................................ ..............10-19
10.10 Setup of Test Group ........................................................... 10-20
10.11 Reagent Holder ..................................................................10-21
10.12 Setup of Reagent Volume Monitoring ................ ................. 10-23
10.13 Setting of Conversion Formula ........................................... 10-23
10.14 Devices to be Connected ...................................................10-24
10.15 Setup of System ................................................................. 10-26
10.16 Password Settings .............................................................. 10-28
10.17 Printout of Settings ............................... .... ... ... ... ... .... ... ... ... . 10-29
10.18 Addition of New Analysis Parameters ................................10-30
10.19 Reagent Name/Holder List ................................................. 10-31
Revised January 2003 - 2.0_en
Sysmex CA-500 series III
Page 5
11. Maintenance and Supplies Replacement ........11-1
11.1 Maintenance Schedule ...................... .... ... ... ....... ... ... ... ... .... .. 11-1
11.2 Clean Sample Probe .............................................................11-2
11.3 Discard Used Reaction Tubes .............................................. 11-3
11.4 Dispose of Waste ......................................................... ... ......11-4
11.5 Remove Dew from Reagent Rack
(for CA-530, CA-540, CA-550 and CA-560 only) ..................11-5
11.6 LED Calibration .....................................................................11-6
11.7 Replace Rinse Filter .............................................................11-9
11.8 Supply Printer Paper ............................................................. 11-9
11.9 Replace Fuse ......................................................................11-11
11.10 Check and Drain Trap Chamber ..... .................................... 11-11
11.11 Prime Rinse Solution to Hydraulic Line ..............................11-12
11.12 Clean Instrument ................................................................11-13
11.13 Replenish Reagent ...................................... ... .... ... ...... ... .... 11-14
11.14 Replenish Reaction Tubes .................................................. 11-16
11.15 Replenish Rinse Solution .................................................... 11-18
11.16 Supply Parts List .................................................................11-19
12. Troubleshooting ................................................12-1
12.1 Introduction ..................................................... ...................... 12-1
12.2 Error Corrective Procedure ...................................................12-2
12.3 Analysis Data Error ............................................................. 12-13
12.4 Cycle Counter .....................................................................12-14
12.5 Sysmex Menu .....................................................................12-15
12.6 Special Operation ...............................................................12-16
13. Functional Description ......................................13-1
13.1 Detection Principle of Coagulation Method (PT, APTT,
Fbg, TT, PCcl, BXT, LA1*, LA2*, Factor Deficiency) ............13-1
13.2 Detection Principle of Chromogenic Method (AT3, APL*, Plg*,
PC, Hep: CA-530, CA-540, CA-550 and CA-560 only) .........13-4
13.3 Detection Principle of Immunology Method
(D-Dimer, P-FDP*: CA-550 and CA-560 only) ......................13-5
13.4 Analysis Mechanism .............................................................13-7
13.5 Analysis Flow ........................................................................ 13-7
13.6 Reference Procedures ........................................................13-16
14. Technical Information .......................................14-1
14.1 Instrument Specifications ...................................................... 14-1
14.2 Installation ....................................................... .... ... ... ... ... ......14-8
14.3 Serial Interface for Host Computer ..................................... 14-17
14.4 Text Format ........................................................................14-24
14.5 ID Barcode ............................. ... ... ... ... .... ... ..........................14-34
15. Index ...................................................................15-1
16. Appendix (A) ......................................................16-1
16.1 Maintenance CheckList ........................................................ 16-1
16.2 Reagents .............................................................................. 16-3
Revised January 2003 - 2.0_en
IV Sysmex CA-500 series
Page 6
1. Introduction ..............................................................1-1
1.1 Introduction .................................................................................1-1
1.2 Explanation of Signs ...................................................................1-3
1.3 Names ................................ .................................... ..................... 1-4
1.4 Serial Number..............................................................................1-4
1.5 Revision History ..........................................................................1-4
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 7

1. Introduction

1.1 Introduction

Introduction
The Sysmex® Automated Blood Coagulation Analyzer CA-500 series is a compact fully-automated instrument capable of 5-parameter random analysis for In Vitro Diagnostic use.
This instrument incorporates latest technologies as represented by micro­computers, thus enabling analysis of multiple parameters with increased flexibility. Of PT, APTT, Fbg, Thrombin Time (Coagulation Method), and Antithrombin III (Chromogenic Method: Can be analyzed only with CA-530, CA-540, CA-550 and CA-560), D-Dimer (Immunology Method: Can be analyzed only with CA-550 and CA-560), etc., this instrument is able to analyze 5 parameters simultaneously. In addition, it has a number of functions including preferential processing of STAT samples and a built-in quality control function. Moreover, it allows analyzed data to be displayed and printed out together with reaction curves, thus making it possible to obtain highly reliable analysis results.

Analysis Parameters and Detection Principles

Parameter Test name Applied
Prothrombin Time PT Coagulation Method Activated Partial Thromboplastin Time APTT Coagulation Method Fibrinogen Fbg Coagulation Method Thrombin Time TT Coagulation Method Protein C coagulometric PCcl Coagulation Method Batroxobin BXT Coagulation Method LA1 Screening* LA1 Coagulation Method LA2 Confirmation* LA2 Coagulation Method Factor Assay** II, V, VII, VIII, IX,
Antithrombin III AT3 Chromogenic Method α2-Antiplasmin* APL Chromogenic Method Plasminogen* Plg Chromogenic Method Protein C chromogenic BCPC Chromogenic Method Heparin Hep Chromogenic Method D-Dimer Plus*, Advanced D-Dimer*** DDPl, AdDD Immunoassay Method P-FDP**** PFDP Immunoassay Method
Coagulation Method
X, XI, XII
(*) Not available in the USA. (**) Data evaluated for factors VII and VIII only. (***) Only available for use in the USA. (****) Only available for use in Asia.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 1-1
Page 8
Introduction

Manufacturer

SYSMEX CORPORATION 1-5-1 Wakinohama-Kaigandori Chuo-ku, Kobe 651-0073 Japan

European Representative

SYSMEX EUROPE GmbH Bornbarch 1 D – 22848 Norderstedt Tel.: +49 40 5 27 26-0 Fax: Tel.: +49 40 5 27 26-100

Ordering of Supplies and Replacement Parts

If you need to order supplies or replacement parts, please contact your local Sysmex representative.

Service and Maintenance

Training Courses

CE-Mark

Please contact the Service Department of your local Sysmex representa­tive.
For further information please contact the Sysmex representative in your country.
The IVD system described in this manual is marked with a CE mark which confirms the observance of the essential requirements of the fol­lowing European directives:
-98/79/EC in-Vitro Diagnostics Directive.
CA-500 series instruments with serial numbers shown below or smaller numbers only conform to the 89/336/EEC electromagnetic compatibility.
CA-510 A1342 CA-520 A1017 CA-530 A1902 CA-540 A3883 CA-550 A1051 CA-560 A1110
1-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 9

1.2 Explanation of Signs

Introduction
This manual carries a variety of illustrations to make sure that the prod­uct can be used safely and correctly, thus preventing users and others from suffering injuries and damage to property.
The illustrations and meaning are described in the following. Do understand what they mean before proceeding to the text of the
MANUAL.
Risk of Infection
Indicates the presence of a biohazardous material or condition.
Warning
If this sign is ignored and the instrument is operated incorrectly, there is a potentially hazardous situation which could result in death or serious injury of operator, or grave property damage.
Revised November 2003 - 2.0_en
Caution
If this sign is ignored and the instrument is operated incorrectly, there is a potentially hazardous situation which may result in injury of operator, adverse effect on results, or may cause property damage.
Important
Indicates what we would like you to know to maintain instrument performance and prevent its damage.
Note
Indicates information which will come handy in operating the instrument.
Sysmex CA-500 series 1-3
Page 10
Introduction

1.3 Names

Sysmex is a registered trademark of SYSMEX CORPORATION in
the USA, in Germany and other countries.
CA CLEAN I and CA CLEAN II are trademarks of SYSMEX COR-
PORATION in the USA, in Germany and other countries.
Actin, Ci-Trol, Data-Fi and Innovin are registered trademarks of
Dade Behring Inc in the USA, in Germany and other countries.
Dade is a registered trademark of Dade Behring Inc.
Pathromtin and Thromborel are registered trademarks of Dade
Behring Marburg GmbH in Germany and other countries and are trademarks of Dade Behring Marburg GmbH in the USA.
VACUTAINER is a registered trademark of Becton, Dickinson and
Company.
VACUETTE is a registered trademark of Greiner Bio-One GmbH.

1.4 Serial Number

1.5 Revision History

Other trademarks referenced are property of their respective owners.
CA-500 series instruments with serial numbers below or smaller
CA-510 A1300 CA-520 A1018 CA-530 A1805 CA-540 A3405
will not have the capability to manage more than 7 Assays/Standard curves and they are limited to use 7 × 6 QC files.
Version Date Changes
Manual Software
1.0 00-13 October 2001 Launch Version
2.0 00-15 January 2003 Second Edition
2.0 00-15 May 2003 Minor Correction
2.0 00-15 September 2003 Minor Correction
2.0 00-15 November 2003 To conform with IVD Directive
2.0 00-17 April 2004 Minor Correction
1-4 Sysmex CA-500 series
Revised April 2004 - 2.0_en
Page 11
2. Safety Information ...................................................2-1
2.1 Specified Conditions of Use ........................................................2-1
2.2 General Information ....................................................................2-1
2.3 Installation Location ..................... ... ... .... .....................................2-3
2.4 Avoidance of Infections ...............................................................2-3
2.5 Handling of Reagents ............................. ... ... ... ... .... .....................2-4
2.6 Maintenance of the Instrument ....................................................2-5
2.7 Disposal of Materials ...................................................................2-5
2.8 Markings on the Instrument .........................................................2-6
2.9 Personnel .................................. .................................................. 2-8
2.10 Storage Condition (Transportation) .. ... .... ... ... ... ... .... ... ... ....... ... ... ..2-8
Revised November 2003 - 2.0_en
Sysmex CA-500 series
Page 12

2. Safety Information

Before operating this instrument, carefully read this manual, and strictly follow the instructions given in them.

2.1 Specified Conditions of Use

The instrument shall only be used for in vitro analysis of human blood or artificial control blood. It may not be used for any other purpose.
Only reagents and cleaning solutions mentioned in this manual are per­mitted for use.
By observing the specified conditions of use, the frequency of cleaning and maintenance work can be reduced.

2.2 General Information

Safety Information
Warning
Keep long hair, fingers and clothing away from rotat­ing parts of the instrument.
During analysis, do not open the light shield cover and put in hands or fingers. This could cause injury. When the light shield cover is opened during analysis, the alarm sounds and the operation stops.
In the event that the instrument emits an abnormal odor or any smoke, turn off its power supply immedi­ately and pull out the power plug from the wall socket. If the instrument is used continuously in that state, there is a hazard that fire, electrical shock, or injury may result. Contact your local service representative for inspection.
Take care not to spill blood or reagent, or drop wire staples or paper clips into the instrument. These might cause short circuit or smoke emission. If such problem should occur, turn off the power supply immediately and pull out the power plug from the wall socket. Then contact your local service representa­tive for inspection.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 2-1
Page 13
Safety Information
Warning
Do not touch the electric circuits inside the instru­ment. Especially with wet hands there is a risk of electric shock.
Never put the power plug in any socket other than that specified. When installing the instrument, be sure to ground it. Otherwise, fire or electrical shock will result.
Take care not to damage the power cord, put a heavy thing on it, or pull it forcibly. Otherwise, the wire may become shorted or break, causing fire or electrical shock.
When connecting the instrument to a peripheral (host computer, etc.), be sure to switch off the power sup­ply beforehand. Otherwise, fire or electrical shock may result.
Use the check-digit as much as possible. If the check-digit cannot be used, the potential of the incor­rect reading of the barcode label may be increased.
Caution
Read this manual carefully to operate the instrument by the proper method. Keep it securely in a specified location for future reference.
This instrument must only be operated as instructed in this manual.
2-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 14

2.3 Installation Location

Safety Information
Caution
Install in a place which is not subject to water splash.
Install in a place which is not subject to adverse
effects of high temperature, high humidity, dust, direct sunlight, etc.
Do not give the instrument a strong vibration or impact.
Install at a place which is well ventilated.
Do not install near devices that cause signal noise,
such as radios and centrifugal machines.
Do not install near chemicals storage or in a place where gas is generated.

2.4 Avoidance of Infections

Risk of Infection
In principle, all parts and surfaces of the instrument must be regarded as infective.
Never touch waste, or parts having been in contact with waste, with bare hands.
Should you inadvertently come in contact with poten­tially infective materials or surfaces, immediately rinse skin thoroughly with plenty of water, then follow the antiseptic regulations of your laboratory.
Be careful when handling samples. Always wear latex or non latex examination gloves; otherwise con­tamination could result. If a sample happens to enter your eye or a cut, wash it off with plenty of water, and immediately visit a physician.
Control plasma may also be infective. Wear latex or non latex examination gloves during QC process. If plasma happens to enter your eye or a cut, wash it off with plenty of water, and immediately visit a physi­cian.
Use care when handling waste liquid. If it adheres to the skin or clothing, wash it off using an antiseptic solution.
Revised November 2003 - 2.0_en
Sysmex CA-500 series 2-3
Page 15
Safety Information

2.5 Handling of Reagents

Warning
Avoid direct contact with reagents. Reagents can cause irritation of the eyes, skin and mucous mem­branes.
If a reagent happens to adhere to the hand or the skin of another area, wash it off immediately using plenty of water.
If a reagent happens to enter your eye, wash it off immediately using plenty of water, and take medical treatment at once.
If you should swallow it inadvertently, call for a physi­cian immediately, drink a large volume of water, and then induce vomiting.
CA CLEAN I is a strong alkaline cleaning material. It should not come in contact with skin or clothing. If it happens nevertheless, rinse skin or clothing with plenty of water to avoid injury or damage.
Caution
Read the instructions described on the reagent con­tainers.
After unpacking, be sure not to allow dust, dirt, or bacteria into the reagents.
Do not use reagents that have passed the expiration period.
Handle a reagent gently to prevent formation of bub­bles. Shake carefully if necessary. Do not use directly after transportation.
Take care not to spill a reagent. If it has spilled, wipe it off immediately using a wet cloth or something.
Prepare a sufficient volume of reagent which takes into consideration the minimum sample volume required. When the volume of the reagent is insuffi­cient, sample may not be analyzed accurately.
The CA CLEAN I rinse solution contains sodium hypochlorite. If the material makes contact with the instrument's surfaces, it will affect the surface finish. There is a danger of corrosion. Immediately wipe up CA CLEAN I with a damp cloth.
A reagent is a chemical substance employed for external diagnosis and cannot be used for any medi­cal treatment.
Revised January 2003 - 2.0_en
2-4 Sysmex CA-500 series
Page 16

2.6 Maintenance of the Instrument

Always wear latex or non latex examination gloves when performing maintenance work or inspection. Also use the specified tools and parts. After work is over, wash the hands in an antiseptic solution. There is a possibility that those areas of the hand which came in contact with blood could suffer infection.
When carrying out inspection and maintenance of the instrument, use only the specified tools and parts. Never use substitute parts or modify parts. Such actions are dangerous and prohibited by the Pharmaceutical Affairs Law.
Safety Information
Risk of Infection
Warning

2.7 Disposal of Materials

Risk of Infection
When discarding waste liquid, instrument consumables and instrument, take proper steps to dispose of them as medical, infective and industrial wastes. If they are contaminated with blood, there is a possibility of bacterial infection occurring.
Revised November 2003 - 2.0_en
Sysmex CA-500 series 2-5
Page 17
Safety Information

2.8 Markings on the Instrument

Warning
DO NOT place your fingers and hands inside while ana­lyzing. Pipette can move in any direction.
Fast Stop (Mechanical Stop)
Warning
DO NOT press on the sampler unit to prevent damage.
Risk of Infection
In principle, all parts and surfaces of the instrument must be regarded as infective.
Warning
When the arm is to be moved while the pipette is in the down position, pull up the pipette to the same height as catcher, and move the arm.
Revised November 2003 - 2.0_en
2-6 Sysmex CA-500 series
Page 18
Safety Information
Warning
To avoid risk of electrical shock, disconnect the power cord before replacing the fuse.
Warning
Proper use of the appropriate power cord assures ade­quate grounding for the system. Failure to properly ground the instrument bypasses important safety fea­tures and may result in an electrical hazard.
RS-232C Serial Port (HC Connector)
Risk of Infection
To avoid contact with biohazardous materials, gloves must be worn when handling the tube trash and used reaction tubes. Wash your hands with an antimicrobial solution after completing the procedure.
Warning
To avoid dusts or contaminants entering the rinse bottle, ensure the float switch does not touch any surfaces when replenishing distilled water. Dusts or contaminants will cause malfunction of the solenoid value.
Revised April 2004 - 2.0_en
Sysmex CA-500 series 2-7
Page 19
Safety Information

2.9 Personnel

Risk of Infection
The waste container and contents should be considered potentially biohazardous. Do not handle the waste con­tainer without proper protective equipment. Wear gloves when handling. Wash your hands with an antiseptic solu­tion after completing the procedure.
Caution
Those who have no or only limited experience in using the instrument are recommended to have guid­ance or assistance from those with sufficient experi­ence.
If the instrument has developed a problem by any chance, a person in charge of it should take steps within the range specified in Instructions For Use. As to problems other than those mentioned, contact your local service representative for repair.
Instrument unpacking, installation, and confirmation of initial operation must be done by your local service representative.
Warning
This instrument is clinical laboratory equipment for screening. When making clinical judgment based on analysis results, the doctor must also consider clinical conditions and other inspection results for an overall judgment.

2.10 Storage Condition (Transportation)

Ambient Temperature: -10ºC to +60ºC
Relative humidity: 95% or less
(Non condensing / Keep dry)
2-8 Sysmex CA-500 series
Revised November 2003 - 2.0_en
Page 20
3. Design and Function ...............................................3-1
3.1 Overview ..................................................................................... 3-1
3.2 Operation Flow ....................................................... .....................3-8
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 21

3. Design and Function

3.1 Overview

Front

Design and Function
1. Light Shield Cover
2. LCD (Liquid Crystal Display)
3. LCD Contrast Adjustment Knob
4. Mechanical Stop Switch
5. Built-in Printer
6. Sampler
1. Light Shield Cover Prevents photoelectric detection from being affected by scattered light from external sources.
Ensure this cover is shut before proceeding to analyze any samples. Analysis cannot be started if this cover remains open.
Important
Do not open the Light Shield Cover while analysing. Opening the cover will sus­pend analysis and beep the alarm. Also, opening the cover and inserting your hand may cause injury.
2. Liquid Crystal Display (LCD) Displays analysis results, reaction curves, sample numbers, test conditions, etc. The LCD functions as a touch-sensitive control panel. The operator can execute various opera-
tions and enter settings by lightly touching keys displayed on the LCD.
3. LCD Contrast Adjustment Knob Controls LCD contrast. This knob is located inside the printer cover.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 3-1
Page 22
Design and Function
4. Mechanical Stop Switch Used to immediately stop the mechanical unit in the instrument in the event of an emergency.
Note that the power of the instrument is not turned off even when this switch is turned on. If emergency stop of the instrument is required due to power failure of the laboratory, immedi­ately turn off the power of the instrument.
If there is a sample that has been already dispensed, this sample has to be reana­lyzed from the start.
5. Built-in Printer Setting conditions, error messages and analysis results are printed out on the thermal paper of
the graphic printer. The LCD contrast adjustment knob is located inside the built-in printer.
6. Sampler The sampler has a load capacity of one sampler rack with 10 sample tubes. The sampler racks
are specific for Sysmex instruments. One rack can be set on the sampler at a time. Pull out the Sampler toward you to load a rack. Once the rack is loaded, the sampler will oper-
ate without the need for intervention by the operator.
Important
Note
The sampler unit is locked while sampling and dispensing. Once the status has become
ready to set samples, the sampler lock is released. You can pull out the sampler to set samples on available positions on the rack in use, or to place a next rack to allow contin­uous analyses.
The sampler unit can also be pulled out by the STAT sample analysis procedure to
allow an analysis of a STAT sample.
3-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 23

Front Interior (When Opening Light Shield Cover)

888
10
10 10
8
10
97
88
1. Reaction Tube Rack
2. Reagent Cooler Holder
3. Reagent Rack
Design and Function
7. Detection Well (For Immunology Method)
8. Sample Incubation Wells
9. Detection Well (For Chromogenic Method)
10. Detection Well (For Coagulation
Method)
11. Probe
12. Reaction Tube Trash
13. Probe Rinse Cup (Outside)
14. Probe Rinse Cup (Inside)
15. CA CLEAN I Holder (default)
16. Buffer Holder (default)
4. Sampler
5. ID Barcode Reader
6. STAT Sample Rack
1. Reaction Tube Rack Two reaction tube racks can be set. One rack holds up to 30 reaction tubes (SU-40). Tube posi-
tion numbers are assigned from the right rear position of the right rack (No. 1) and count upwards moving toward the front.
2. Reagent Cooler Holders (For CA-530, CA-540, CA-550 and CA-560 only) Can hold up to 4 vials with cooler function.
3. Reagent Rack Reagent vials, whose outer diameter is 22 mm and he ight is 40 mm , can be set directly. Use
sample cups or optional holders to place any vial with other outer diameters.
Caution
If any vial higher than 40 mm is used, the Probe will be damaged permanently.
4. Sampler Can hold one sample rack.
5. ID Barcode Reader (Optional on CA-510, CA-530 and CA-550) ID Barcode Reader moves in front of the rack and reads the barcoded label automatically.
6. STAT Sample Rack Place a STAT sample collection tube or sample cup here. If a sample collection tube is placed,
use optional holders to make the tube diameter fit the rack.
7. Detection well (For Immunology Method: CA-550 and CA-560 only) Detection well for the immunologic sample. The number of wells is one. The detector is always
kept at 37.0ºC ± 1.0ºC.
Revised September 2003 - 2.0_en
Sysmex CA-500 series 3-3
Page 24
Design and Function
8. Sample Incubation Wells Six incubation wells are provided, and these wells are kept at 37.0ºC ± 1.0ºC.
9. Detection Well (For Chromogenic Method: CA-530, CA-540, CA-550 and CA-560 only) Detection well for the chromogenic sample. The number of wells is one. The detector is always
kept at 37.0ºC ± 1.0ºC.
10. Detection Wells (For Coagulation Method) Four scattered light detection wells are provided, and these wells are kept at 37.0ºC ± 1.0ºC.
11. Probe This pipette is used to aspirate samples and reagents. It is kept at 37.0ºC ± 1.0ºC.
12. Reaction Tube Trash Used reaction tubes are disposed into this trash.
13. Probe Rinse Cup (Outside) The outside of the probe is rinsed with the rinse fluid kept in this rinse cup.
14. Probe Rinse Cup (Inside) The inside of the probe is rinsed in this rinse cup.
15. CA CLEAN I Holder CA CLEAN I detergent is set in the vial, whose outer diameter is 22 mm or less, and height is
50 mm or less.
Caution
Use the provided vials to hold the CA CLEAN I detergent. If any vial higher than 50 mm is used, the Probe will be damaged permanently.
16. Buffer Holder Buffer diluent used for sample dilution is set in the vial, whose outer diameter is 22 mm or less,
and height is 50 mm or less.
Caution
Use the provided vials for the container to keep the Buffer. If any vial higher than 50 mm is used, the Probe will be damaged permanently.
3-4 Sysmex CA-500 series
Revised April 2004 - 2.0_en
Page 25

Left Side

Design and Function
2. Cover Sensor Switch
1. Power Switch
3. Cooler Unit
1. Power Switch Turns the power ON or OFF.
Caution
Please allow at least 5 seconds between turning the instrument OFF and back ON, or the fuse may be blown.

Right Side

2. Cover Sensor Switch This monitors if the Light Shield Cover is closed.
3. Cooler Unit This keeps the reagent cool.
8. Solenoid Valve for Sample Aspiration
9. Reaction Tube Trash Drawer
1. X-Y Drive Motor This motor drives the Probe unit in the X-axis and Y-axis directions.
1. X-Y Drive Motor
2. Host Computer Serial Port
3. Memory Card Cover
4. Pressure Pump
5. DIP Switch
6. Vacuum Pump
7. Solenoid Valve for Probe Rinse
2. Host Computer Serial Port For connecting to an external host computer.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 3-5
Page 26
Design and Function
3. Memory Card Cover This card has PROM chips to load the CA-500 program in to RAM memory. (In tended for your
local service representative use only)
4. Pressure Pump Supplies the pressure. The instrument cannot function properly at lower pressures.
5. DIP Switch Changes the system settings. (Intended for your local service representative use only)
6. Vacuum Pump Supplies vacuum. The instrument cannot function properly if the vacuum level is lower.
7. Solenoid Valve for Probe Rinse This valve controls the supply of the rinse solution to the rinse cup unit.
8. Solenoid Valve for Sample Aspiration This valve controls the aspiration of sample plasma with high accuracy.
9. Reaction Tube Trash Drawer

Rear

Used for storing used reaction tubes.
1. Trap Chamber
6. Fuse Holder
2. Float Switch Connector for Waste Bottle
3. Float Switch Connector
for Rinse Bottle
4. Pressure Supply Nipple
for Rinse Bottle
5. Rinse Aspiration Nipple
8. Vacuum Nipple for Waste Bottle
9. Waste Outlet Nipple
7. Power Connector
1. Trap Chamber Prevents the waste fluid from flowing back to affect the vacuum pump, in the event of an
abnormality with the instrument.
2. Float Sensor Connector for Waste Bottle (“WASTE”) For connecting the float sensor switch, located on the waste container, for detecting the waste
fluid level.
3-6 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 27
Design and Function
3. Float Sensor Connector for Rinse Bottle (“RINSE”) For connecting the float sensor switch, located on the rinse container, for detecting rinse water
level.
4. Pressure Supply Nipple for Rinse Bottle (Colored Black) To be connected via a tube with the Rinse Bottle.
Important
When the Rinse Bottle is to be opened, disconnect this tubing first to release the pressure accumulated inside the Rinse Bottle. Failing to do this will splash the pres­surized rinse fluid.
5. Rinse Aspiration Nipple (Colored Blue) For aspirating the rinse water from the Rinse Bottle. To be connected via a tube to the Rinse
Bottle.
Important
When the Rinse Bottle is to be opened, disc onnecting this tubing first will splash the pressurized rinse fluid. Disconnect the black tubing first.
6. Fuse Holder Two time-lag type fuses are installed in this Fuse Holder. Replace with the correct type of fuse
(supplied). The rating will be different depending on the instrument specification as below.
Specification Part No. Description Fuse Type Location
117 VAC 266-5106-0 Fuse 250V 6.3A ST4-6.3A-N1
Time Lag Rear Panel
(N.Amer)
220-240 VAC 266-5293-0 Fuse 250V 3.15A No. 19195
(Europe)
Time Lag Rear Panel
Warning
To avoid risk of electrical shock, disconnect the power cord before replacing the fuses.
For continued protection against risk of fire, replace only with a fuse of the spec­ified type and current ratings.
7. Power Connector For connecting the main power supply (via the supplied power cable).
8. Vacuum Nipple for Waste Bottle (Colored Green) To be connected via a tube with the Trap Chamber.
9. Waste Outlet Nipple (Colored Red) For draining waste fluids. Must be connected via a tube to the Waste Bottle.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 3-7
Page 28
Design and Function

3.2 Operation Flow

Manual Order Registration On-Line Order Registration
(Manual Inquiry)
Inspection Before Turning On Power
Turn On Power
Self Check
Ready
Prepare Reagents
Register Analysis
(Manual Registration)
Prepare Samples
Operation After Completion of Analysis
: Indicates actions performed by the operator.
Prepare Samples Register Analysis
Press [HC] key
Press [Start] key
Execution of Analysis
Completion of Analysis
Ready
Turn Off Power
On-Line Order Registration
(Auto Inquiry)
Prepare Samples
: The message "Ready" will appear on the LCD screen, indicating that analysis, setting, dat a
processing and other operations can be executed.
Revised January 2003 - 2.0_en
3-8 Sysmex CA-500 series
Page 29
4. Installation Environment .........................................4-1
4.1 Installation and Relocation ........................ ... ... ... .... ... ... ... .... ... ... ..4-1
4.2 Installation Location ..................... ... ... .... ... ... ... ... .... ... ... ... .... ........4-1
4.3 Basic Instrument Settings ...................... ... ... ... ... .... ... ... ... .... ... .....4-3
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 30

4. Installation Environment

4.1 Installation and Relocation

Installation of the instrument must be conducted by your local service representative. If it is necessary to relocate the instrument, contact your local service representative.
It is to be noted that if problems should develop as a result of relocation conducted by a customer, it will void the warranty even if the instrument is in the warranty period.

4.2 Installation Location

Grounding

The power cord of each instrument uses the 3P plug. When the power supply socket is 3P (with ground) type, simply plug it to the socket. The type of cord and plug supplied depends on the source voltage for the sys­tem.
Installation Environment

Installation Space

Warning
Proper use of the appropriate power cord assures ade­quate grounding for the system. Failure to properly ground the instrument bypasses important safety fea­tures and may result in an electric hazard.
Note
The number of power supply sockets required is one.
To ensure optional instrument performance properly, install it at an appropriate location.
Select a place where the power supply is located close.
Be sure to use the supplied bottles to collect rinse solution and waste.
Keep a space for maintenance and service. Giving consideration to
heat radiation by the instrument, provide at least 50 cm distance from the wall to sides, rear, and top panels.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 4-1
Page 31
Installation Environment
The dimensions of the instrument are shown below. The power cord is
1.8 m long.
Width (mm) Depth (mm) Height (mm) Weight (kg)
Main unit 540 470 487 Approx. 45
487
540
470
Caution
Be sure to place the rinse bottle and waste bottle on the base on which the instrument is set. Do not place them on the instrument. They may cause the instrument to break down or to fail to produce correct results.
4-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 32

Installation Environment

Use the instrument at an ambient temperature of 15 - 35ºC.
Use it at a relative humidity range of 30 - 85%.
When the ambient temperature and humidity are not appropriate,
Avoid using the instrument in a place where the temperature can
Avoid using the instrument in a place where it may be frozen.
Avoid using the instrument where it can be exposed to direct sun-
Select a well-ventilated place.
Avoid using it at a place close to a wireless telegraph, communica-

4.3 Basic Instrument Settings

Installation Environment
control by air conditioning.
become extremely high or low.
light.
tion equipment, etc. which may emit high-frequency waves or inter­fere with radio waves.

Contrast Adjustment for LCD Screen

Remove the printer cover, and adjust LCD screen contrast (shade) using the contrast adjust dial on the left side of the printer.
Dark
Light
Turning up the dial makes the screen darker and turning down makes it lighter.

Setup of System (Date/Time)

Set date and time.
The instrument has a built-in clock, so there is no need to set the date and time every day. Should the power be turned off, the built-in clock is pow­ered by an internal battery.
1. Press [Special Menu] key on the Root Menu screen.
2. Press [Settings] key on the Root Menu screen.
The contents of the Root Menu will change over.
The Setting Menu screen will appear.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 4-3
Page 33
Installation Environment
Ready
Sysmex
Settings
General Set Up
Date/Time Date 12/01/2001
 Time 16:35:38
Replace Rack? YES!
Auto Val/Out
Data Check
Analysis Settings
I/O Setting
General Set Up
Ready
Sysmex
Replace Rack? YES!
Date/Time
Date Format
Password Setting
Ready
Sysmex
Replace Rack? YES!
HC IP
Print Settings
HC IP
HC IP
7
4
1
0
C
Main Menu
Return
Date
89
5
6
3
2
/
Enter
Quit
3. Press [General Set Up] key on the Setting Menu screen.
The General Set Up Menu screen will appear.
4. Press [Date/Time] key on the General Set Up Menu screen.
The Date/Time Setting screen will display current date and time.
5. Using [] and [] keys, move the cursor to select Date or Time.
6. Using the numeric keys, set Date and Time, and press [Enter] key.
The parameter in the cursor position is set and the cursor will move to the next parameter.
Important
When entry is made in the wrong format, the setting
is not executed.
If the number of day or month is a single digit, enter it
with a 0 preceding it.
Enter the time in a 24-hour clock system.
7. When setting is completed, press [Quit] key.
The Renew Confirmation screen will appear.
Sysmex
Replace Rack? YES!
HC IPReady
8. Press [FIX] key, [Continue] key, or [Cancel] key.
[FIX] key: Changes to the renewed setting and returns to the
General Set Up Menu screen.
RENEW SETTING ?
[Continue] key: Returns to the Date/Time Setting screen and allows
continued operation.
[Cancel] key: Cancels the renewed setting and returns to the Gen-
ContinueCancel FIX
4-4 Sysmex CA-500 series
eral Set Up Menu screen.
Revised January 2003 - 2.0_en
Page 34
5. Operation ..................................................................5-1
5.1 Display Screens and Operation Keys .........................................5-1
5.2 Menu Tree ...................................................................................5-3
5.3 Types of Alarm ............................................................................5-5
5.4 Inspection before Turning ON the Power ....................................5-5
5.5 Turn ON the Power .....................................................................5-7
5.6 Prepare Reagents .......................................................................5-8
5.7 Set Reaction Tubes ...................................................................5-13
5.8 Confirm Standard Curve ...........................................................5-14
5.9 Execute Quality Control ............................................................5-15
5.10 Prepare Samples ......................................................................5-15
5.11 Set Sample Nos. .......................................................................5-19
5.12 Manual Inquiry ................................. ... .... ... ... ... ... .... ... ... ... .... ... ...5-22
5.13 Automatic Inquiry ......................................................................5-22
5.14 Start Analysis ............................................................................5-24
5.15 Automatic Sensitivity Adjustment of the Detector
(for CA-530, CA-540, CA-550 and CA-560 only) ......................5-26
5.16 Display Analysis Result .............................................................5-27
5.17 Interrupt Analysis ......................................................................5-28
5.18 Add Samples ... ... ... ... .... ... ... ... .... ... ... ..........................................5-29
5.19 Analyze STAT Sample ..............................................................5-30
5.20 Emergency Stop ................................. .... ... ... ... ... .... ... ... ... .... ... ...5-31
5.21 Shutdown .................................................................................. 5-33
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 35

5. Operation

5.1 Display Screens and Operation Keys

The instrument displays all information including the instrument status, analysis results, etc. on the LCD screen. The LCD screen is divided into system status area, data processing area, and menu processing area. Press a key pad, and the function indicated on the key will work.
Operation

System Status Area

(1) (2)
Ready
Sysmex
Stored
Data
Replace Rack? YES!
QC
Main Menu 1 PT 2 APTT 3 Fbg
Rack ID No.
01-01
01-02  01-03  01-04  01-05
(3)
4 TT 5 AT3
12345
Repeat
Standaed
Curve
(4)
HC IP
Group 1
ID No.
Entry
Test
Group
(5)
(6)
Start
Prev
Next
HC
Special
Menu
System Status
Data Processing
Menu Processing
The System Status Area displays [Sysmex] key, error message, analysis status, and the status of externally connected instruments.
1. [Sysmex] key
Press this key to display Sysmex menu showing Error List, Tempera­ture, and Paper Feed. When the instrument develops an error, causing the alarm to sound, [ALARM RESET] key appears. Press [Error List] key to display Error History. Press [Temperature] key to dis­play temperatures of various units. Press [P. FEED] key to feed printer paper.
For Sysmex Menu, refer to “12. Troubleshooting”.
2. Analysis Status This indicates analysis status with the current instrument. “Ready”,
“Analyzing”, “Waiting” will appear.
3. Rack Replacement This indicates whether the sample rack can be replaced or not.
4. HC (Host Computer) The host computer is set to “Connected”. “HC” appears when sample
data for automatic output becomes available.
5. IP (Internal Printer) “IP” appears when sample data for automatic output to the internal
printer becomes available.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-1
Page 36
Operation

Data Processing Area

Menu Processing Area

6. [Start]/[INTERR] key
Press this key to start/stop the sampler analysis. For STAT sample analysis, [Start STAT] key appears.
The data processing area displays analysis progress status, work list, stored data list, reaction curve, quality control data, standard curve data, instrument setup status, etc.
When the power supply is turned on, the work list screen (Root Menu screen) appears.
The menu processing area always displays the menu for function selec­tion.
In selecting a menu, touch a key that shows a menu you want to see. After power supply turn-on, when system check is completed, the Root
Menu appears. The Root Menu is the basic menu for selecting functions of this instrument.
5-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 37

5.2 Menu Tree

Root Menu 5.5 5.11
5.11
Repeat 5.11
ID No. Entry 5.11
HC 5.12
Stored Data 6.1 Top 6.2
QC 8.5 Settings 8.3
Operation
Bottom 6.2 Search 6.2 ID No./seq 6.3
Mark Select Display 6.4 Graph Delete 6.6 Prev Edit ID No. 6.5 Next Output 7.2 More Marked All Clear 6.6
Change Scale 8.5 (*1)
8.5 (*1) 8.5 (*1)
Delete 8.7 (*1) Print 8.8 Select File 8.5 Select Test 8.5
Standard Curve 9.1 Next 9.1 (*2)
Test Group 5.11 Set Reagents 5.6
Special Menu
Sysmex 12.5 Error List
Temperature P.FEED Main Menu
Start 5.14 INTERR 5.17
Prev 5.16 Next 5.16
Standard Analysis 9.2 (*2) Manual Entry 9.4 (*2) Select Param. 9.6 Select Test 9.1 Graph 9.1 (*2) Print 9.7 (*2) Lot No. Entry 9.5
Settings 10.1 Rinse Probe 11.2 Special Operate 12.6
: If set, password has to be entered. *1: This item is displayed only when QC data is st ored. *2: This item is displayed only when calculat ion
parameter is set.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-3
Page 38
Operation
Root Menu Special Menu Settings Auto Val/Out 10.2
Data Check Mark Limits 10.3
Analysis Settings Set Test Name 10.6
I/O Settings 10.14 Host Comp ute r
General Set Up Date/Time 10.15
Print Settings 10.17 Auto Val/Out
Replic. Limits 10.4 Report Limits 10.5
Set Reagent Name 10 .7 Test Protocol 10.8 Set Replication 10.9 Test Group 10.10 Reagents Holder 10.11 Alarm Settings 10.12 Conversion 10.13
Barcode Scanner
Date Format 10.15 Password Setting 10.16
Data Check Analysis Settings I/O setting General Set Up All Set Data
Special Operate Rinse Prepa r e 11.11
System Tests 12.6 LCD
Cycle Counter 12.4
LED Calibration 11.6
: If set, password has to be entered.
Touch Screen Host Computer Printer Barcode Scanner Sensor Status
5-4 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 39

5.3 Types of Alarm

Operation
The instrument alarm emits 4 different sounds:
Key entry sound (pip)
Sounds about 0.1 sec. when a touch panel key is pressed.
Rack replacement sound (pip, pip, pip)
Sounds when sampling and dispensing of all set samples are com­pleted and the system becomes ready to add orders or replace racks.
Analysis completion sound (pip, pip, peep)
Sounds when analysis of all registered samples is completed.
Instrument error sound (beep)
Sounds when some error has occurred in the instrument. This sound continues until [ALARM RESET] key is pressed. While the alarm is sounding, the [ALARM RESET] key is dis-
played in place of [Sysmex] key. This sound is emitted also when the sample rack has been lifted after
the system ran out of sample tubes or reagents or when they were being replenished.
This alarm sound stops when the [Conf.] key is pressed or when the sample rack is set correctly.

5.4 Inspection before Turning ON the Power

Inspect Rinse Bottle

When the rinse solution level is found low, replenish the rinse bottle with distilled water. As to the procedure for replenishing rinse solution, refer to “11.15 Replenish Rinse Solution”.
Caution
When analysis is made with the rinse bottle laid down, there is a possibility that correct analysis result may not be obtained. Make sure the rinse bottle is standing upright.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-5
Page 40
Operation

Inspect Waste bottle

When waste liquid has collected in the waste bottle, discard the contents. Regarding how to dispose of waste liquid, refer to “11.4 Dispose of Waste”.
Risk of Infection
When disposing of waste liquid, always wear latex or non latex examination gloves. After work is over, wash the hands in anti-septic solution. If hands are contaminated with blood, which could result in infection by pathogenic organisms, give careful con­sideration to the hazards of medical or infective waste materials.
Important
When analysis is made with the waste bottle laid down, waste may flow back into the vacuum pump, causing the pump to fail. Make sure the waste bottle is standing upright.

Check Power Cord

Check Connection Cord

Check Tube Trash Drawer

Check Printer Paper

Check Light Shield Cover

Check to see the power cord is securely plugged in the socket.
When the instrument is connected with the host computer, check to see the connection cord is securely connected.
When used reaction tubes remain in the tube trash drawer, discard them. As to discarding, refer to “11.3 Disc ard Used Reaction Tubes”.
Check to see the internal printer has enough paper to handle the number of samples expected that day.
Open the cover and check to see there are no obstacles for analysis.
5-6 Sysmex CA-500 series
Revised November 2003 - 2.0_en
Page 41

5.5 Turn ON the Power

Turn ON the Power

Automated Blood Coagulation Analyzer
CA-500
Check in Progress
Operation
1. Turn on the power switch on the left side of the instrument. The system automatically performs a roughly 10-second self check,
and the Root Menu screen will appear.
Ready
Sysmex
Stored
Data
Replace Rack? YES!
QC
Main Menu 1 PT 2 APTT 3 Fbg
Rack ID No.
01-01
01-02  01-03  01-04  01-05
4 TT 5 AT3
12345
Repeat
Standaed
Curve
HC IP
Group 1
ID No.
Entry
Test
Group
Special

Confirm Automatic Output

Ready
Sysmex
Stored
Data
Replace Rack? YES!
QC
Main Menu
1 PT 2 APTT 3 Fbg
Rack ID No.
01-01
01-02  01-03  01-04  01-05
4 TT 5 AT3
12345
Repeat
Standaed
Curve
HC IP
Group 1
ID No.
Entry
Test
Group
Special
Start
Prev
Next
HC
Menu
Start
Prev
Next
HC
Menu
2. When the detector and cooler reach an analysis-permitting tempera­ture, the Root Menu screen displays “Ready”.
Note
The detector and cooler reach an analysis-permitting tem-
perature in about 5 - 30 minutes after power turn-on.
When the system is waiting for an analysis-permitting
temperature, the message “Not Ready” is displayed and “Start” key is not displayed.
When automatic output to the internal printer or host computer becomes necessary, make this confirmation.
1. Confirm that the System Status area displays “HC” and “IP”.
2. Confirm the settings for automatic transfer/printout. Refer to “10.2 Setup of Automatic Transfer/Printout”.
Note
[HC] key is displayed only when the system is set for manual inquiry to the host computer. For detail, refer to “5.12 Manual Inquiry”.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-7
Page 42
Operation

5.6 Prepare Reagents

Prepare Reagents

Prepare coagulation reagents needed for analysis, Owren’s Veronal Buffer, and rinse solution. Refer to the package insert of each reagent for more information.
Parame-
ter
PT PT Reagent 100 µL APTT APTT Reagent 50 µL
Calcium Chloride Solution (0.025 mol/L) 50 µL
Fbg Thrombin Reagent 50 µL
Owren’s Veronal Buffer 90 µL TT Thrombin Clotting Time Reagent 100 µL PCcl Protein C Deficient Plasma 45 µL
Protein C Activator 50 µL
PC APTT Reagent 50 µL
Calcium Chloride Solution 50 µL BXT Batroxobin Reagent 100 µL LA1* LA1 Screening Reagent 100 µL LA2* LA2 Confirmation Reagent 100 µL AT3 Berichrom° Antithrombin III (A)
Thrombin Reagent 125 µL
Substrate 33 µL
Owren’s Veronal Buffer 83 µL BCPC Berichrom° Protein C
Protein C Activator 125 µL
Substrate 30 µL APL* Berichrom° α2-Antiplasmin
Antiplasmin Reagent 125 µL
Substrate 25 µL
Owren’s Veronal Buffer 112 µL Plg* Berichrom° Plasminogen
Streptokinase Reagent 125 µL
Substrate 25 µL
Owren’s Veronal Buffer 112 µL
Reagent Consumption per
Test
5-8 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 43
Operation
Parame-
ter
Hep Berichrom° Heparin
AT3 Reagent 20 µL FXa Reagent 125 µL Heparin Substrate 40 µL
DDPl* D-Dimer PLUS
Accelerator 25 µL Latex Reagent 150 µL
AdDD** Advanced D-Dimer
Accelerator 25 µL Latex Reagent 150 µL
PFDP*** Latex Test BL-2 P-FDP
Reagent 66 µL Latex 94 µL P-FDP Diluents 112 µL
Rinse Solution
Distilled Water (Rinse Bottle) Max. 24 mL per
CA CLEAN I 10 µL more than
Reagent Consumption per
Test
test
the required rea­gent volume.
(*) Not available for use in the USA. (**) Only available for use in the USA. (***) Only available for use in Asia.
The amount of Owren’s Veronal Buffer per test includes the amount used in dilution for each analysis parameter.
When analyzing the parameters that require two reagents, a rinse opera­tion is performed three times in total after dispensing samples and rea­gents.
Thus the amount used per test is approximately 24 mL. However, the number of rinse operations can be changed. (Refer to “10.8 Test Proto­col”.)
CA CLEAN I is used after the reagent is dispensed. The amount used for one rinse is the amount of the reagent + 10 µL.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-9
Page 44
Operation
Caution
Prepare each reagent taking into consideration the analysis parameters and the number of the samples to be analyzed. Prepare extra volumes as shown below, in addition to the required volumes for analy­sis:
Each coagulation reagent: Approx. 0.6 mL Distilled water: Approx. 500 mL Owren’s Veronal Buffer: Approx. 0.9 mL CA CLEAN I: Approx. 0.9 mL
The amount of reagent used in the initial operation (rinse operation) after analysis starts is as follows:
Distilled water: Approx. 20 mL CA CLEAN I: Approx. 125 µL Owren’s Veronal Buffer: Approx. 200 µL (*) (*)only when parameters are analyzed for the Chro-
mogenic Method or Immunology Method
The amount of extra reagents required vary according to the containe r used.
Sample cup conical (4 mL): Approx. 0.2 mL
®
Behring 4 mL vial: Approx. 1.0 mL
Dade
®
Dade
Behring 5 mL vial (GW5): Approx. 0.8 mL
TTO, NT 3 mL vial: Approx. 0.4 mL Push Vial PV-10 (22 mm OD × 40 mm high): Approx. 0.9 mL SLD Vial: Approx. 0.4 mL
The extra volumes shown are the maximum which may be required. There may be variation due to differences in fluid viscosity and slight vial to vial variation.
Caution
Prepare a sufficient volume of reagent which takes into consideration the minimum sample volume required. When the volume of the reagent is insufficient, sample may not be analyzed accurately.
5-10 Sysmex CA-500 series
Revised April 2004 - 2.0_en
Page 45
Operation
Dade Behring
ThromboplastinC Plus
[ ]
[ ]
[ ]
Holder No. 11
1. Prepare Reagents. Prepare reagents as per the document supplied with each reagent.
Caution
Strictly follow the instructions as given in the package
[ ]
insert supplied with each reagent. Otherwise, you will fail to obtain correct analysis result.
2. Set the reagents on the rack. Reagent bottles (each measuring 22 mm OD and 40 mm height), or
optional reagent holders and sample cups can be set in the reagent rack. (Refer to “10.19 Reagent Name/Holder List”.)
Caution
Always set the reagents at the specified positions to obtain correct results.
Confirm that reagents contain no bubbles. Other­wise, correct analysis results cannot be obtained.
Be sure to set CA CLEAN I in Holder No. 11 (inner right side).
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-11
Page 46
Operation

Register Reagent Volume

7
CaCl2
0.0mL
0.0mL
8
HC IP
9
10 12
11
CleanI
OVB
Ready
Sysmex
PT
5.0mL
4.8mL
2
Fbg
AT3Thro
Replace Rack? YES!
3
APTT
0.0mL
0.0mL
4
AT3Subs
0.0mL
0.0mL
Enter Reagent Volume
51
0.0mL
0.0mL
6
0.0mL
0.0mL
0.0mL
0.0mL
0.0mL
0.0mL
Main
Menu
1. Press [Special Menu] key on the Root Menu screen.
The contents of the Root Menu will change over.
2. Press [Set Reagents] key in the Root Menu.
The Reagent Volume screen will display reagent level for each rea­gent holder.
3. Press the key of the reagent holder to be registered. The Reagent Volume Entry Screen will display the numeric keys to
enter reagent volume.
Ready
Sysmex
Sysmex
1
PT
5. 0 m L
5. 0 m L
Replace Rack? YES!
Ready Replace Rack? YES!
RENEW SETTING ?
Enter Reagent Volume
Enter Reagent Volume
IP HC
7
4
1
0
C
HC IP
89
5
6
3
2
Enter
.
Quit
4. Enter the reagent volume set on the reagent holder and press [ENTER] key.
The value entered at the cursor position will be displayed. The sec­ond line will automatically display the available reagent volume to be used for the analyses.
By pressing [] and [] keys, movement can be effected to the pre­ceding or following reagent holder No.
Note
If the Reagent Volume Monitoring is set to “Valid”, each
time the reagent is dispensed, the volume for the test will be subtracted from the entered value.
Refer to “10.12 Setup of Reagent Volume Monitoring” for
the setting procedures.
Refer to “10.11 Reagent Holder” for confirmation of vial
type.
5. Press [Quit] key on the numeric keys. The Reagent Volume Screen will reappear. Confirm the entered value.
6. Press [Main Menu] key on the Reagent Volume screen. The Renew Check screen is displayed. Press [Cancel] key, [FIX]
key, or [Continue] key to return to the Root Menu screen.
Important
The reagent amount which was input is erased when the
FIX ContinueCancel
5-12 Sysmex CA-500 series
power is turned OFF.
Revised January 2003 - 2.0_en
Page 47

5.7 Set Reaction Tubes

Operation
Reaction tube
30
Reaction tube rack
1
Set the reaction tubes for analysis on the reaction tube rack. Set the reaction tube rack on the specified location on the analysis table.
Caution
When setting reaction tubes on the reaction tube rack, be careful not to drop sweat or the saliva into the reaction tube. This will alter the analysis result.
Reaction tubes are for single use only or incorrect results may occur.
Important
Set the reaction tube rack securely to prevent the tubes from unseating and rising; otherwise, the probe might be damaged.
The reaction tubes are set successively from the first position (top on extreme right-hand column). There­fore, make sure no position is left empty.
Make ready some extra reaction tubes in addition to the quantity needed for analysis.
Be sure to use the supplied reaction tubes (SU-40). The reaction tubes (SUA-400A) for CA-6000, etc. or those manufactured by other companies cannot be used.
Note
A maximum of 30 reaction tubes can be mounted on a reaction tube rack. Since two reaction tube racks can be used, the max­imum number of reaction tubes that can be set is 60.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-13
Page 48
Operation

5.8 Confirm Standard Curve

Ready
Sysmex
%
50.0
25.0
12.5
6.3
3.1   
sec
11.4
17.4
27.9
52.6
11.4
1.73
Replace Rack? YES!
Cal Date 12/01/2001
PT Apc.63
Ref.
0.0
0.0
Standard
Analysis
Standard Curve PT PT%
100.0
Normal
ISI
HC IP
Lot No.EXP.
Manual
Entry
12/31/2001
12/31/2001
Select
Param.
Confirm before performing analysis that the standard curve is correctly set.
Caution
Unless the standard curve is properly set, percent activ­ity, concentration, and other calculation parameters can­not be reported.
1. Press [Standard Curve] key on the Root Menu screen.
The Standard Curve screen will be displayed.
2. Press [Select Test] key on the Standard Curve screen.
The Select Test screen will appear.
Select
Sysmex
Standard Curve
PT
VIII
Sysmex
Standard Curve PT PT%
%
100.0
50.0
25.0
12.5
Normal
ISI
Select
Graph
Test
Ready Replace Rack? YES!
APTT Fbg TT
AT3 BCPC Hep
Ready Replace Rack? YES!
sec
PT Apc.63
11.4
17.4
27.9
52.6
0.0
6.3
0.0
3.1 
11.4
1.73
Graph
Test
Print
Cal Date 12/01/2001
Ref.
Standard
Analysis
Print
Lot No.
Entry
HC IP
II
DDPl
HC IP
Lot No.EXP.
Manual
Entry
Lot No.
Entry
12/31/2001
12/31/2001
Main
Menu
Cancel
Select
Param.
Main
Menu
3. Press the analysis parameter key to be confirmed. The standard curve data of the parameter selected will appear on the
standard curve screen.
4. Press [Graph] key. Check the standard curve.
5. Press [Main Menu] key. The standard curve setting program is now completed.
Repeat the above Step-(2 - 3) to confirm the standard curve of each anal­ysis parameter. For detail, refer to “9.2 Standard Curve Analysis”.
5-14 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 49

5.9 Execute Quality Control

5.10 Prepare Samples

Operation
To maintain the reliability of analyzed data, quality control has to be per­formed. With the instrument, when QC File No. (QC01 - QC06) is registered for ID No., and QC sample (control plasma, pooled plasma, etc.) is ana­lyzed, then analysis data is kept in the QC File. It is by processing this analysis data with the QC Program that instrument stability that varies from-time- to-time is monitored.
For detail of the QC Program, refer to “8. Quality Control”.
Set sample tubes or dispensed sample cups on the sampler rack.
Caution
If samples are left at room temperature for a long time, they may deteriorate. Set samples to the sampler just before analysis starts.
A
75mm18mm
1. Prepare plasma.
1) Add 1 part of 3.8%, 3.2% or 3.13%* sodium citrate solution as anticoagulant to 9 parts of venous blood, and mix the contents thoroughly.
2) Centrifuge the mixture at 3000 rpm for 15 minutes to separate
plasma components from blood components.
*Not for use in the USA.
3) Affix Barcode Label (Option).
To ensure correct reading of a barcode, a barcode label has to be
affixed at the proper position.
4) Set, in the supplied sample rack, the centrifuged blood tube itself
or the plasma which has been removed and put into another test tube. Insert the test tube securely to the bottom of the rack. When an optional sample barcode scanner is used, set barcode labels so as to face the scanner.
Warning
Affix the barcode label so that the bars on the label would become horizontal when the rack is placed on the sampler. If the barcode label is affixed slanted, the potential of the incorrect reading of the barcode label will be increased.
Revised September 2003 - 2.0_en
Sysmex CA-500 series 5-15
Page 50
Operation
Anticoagulant 3.8% sodium citrate solution
3.2% sodium citrate solution
3.13% sodium citrate solution*
Useable OD: 10 - 15 mm
HT: 65 -100 mm (Tubes less than 8 mm in ID cannot be used)
*Not for use in the USA.
Caution
Cautions about handling plasma:
As its container, use a plastic or silicone-coated glass tube.
As anticoagulant, use 3.8%, 3.2% or 3.13%* sodium citrate solution. When any other anticoagulant than this solution is used, it will cause a white precipitation and lead to incorrect analysis results.
Mix blood and sodium citrate solution in an accurate ratio of 9 parts to 1 part, respectively. As the mixing ratio varies, coagulation time also varies, occasion­ally leading to incorrect analysis results.
Samples must be analyzed within 4 hours of collect­ing if stored cool.
Those samples that had more than 4 hours pass after collecting and those kept in improper storage condition will not give correct analysis results.
*Not for use in the USA.
Dead volume of plasma-only sample
Plasma
ID
12
0.3 mL or more
ID 13
0.4 mL or more
0.2 mL or more
ID
8.0 - 10.6
ID 11
0.3 mL or more
Height between tube bottom and plasma surface: Approx. 7 mm
Dead volume of centrifuged sample
Plasma layer
ID 13
0.6 mL or more
0.3 mL or more
Blood layer
ID
8.0 - 10.6
11
ID
0.5 mL or more
ID 12
0.5 mL or more
Height of plasma layer: Approx. 4.5 mm
ID 14
ID
14
0.5 mL or more
0.7 mL or more
Revised September 2003 - 2.0_en
5-16 Sysmex CA-500 series
Page 51
Operation
OD x Length Blood Volume
13 mm x 75 mm 1.8*, 2.4, 2.7**, 3.0**, 3.5**, 4.5 mL
13.2 mm x 78 mm 1.8, 2.7, 4.5 mL
12.8 mm x 75 mm 1.8, 2.7, 3.6 mL
12.7 mm x 75 mm 2, 3 mL
* Does not include VACUTAINER Plus Plastic Citrate Tube 1.8
mL (Becton Dickinson).
** Except for the following tubes with a double wall structure any
other tubes with a double wall structure which are not applica­ble:
- VACUTAINER Plus Plastic Citrate Tube, 2.7 mL
- VACUETTE Sandwich Coagulation Tube, 3.0 mL/3.5 mL
Caution
The blood volumes shown above are the dead vol­umes. (Prepare an extra volume for the parameter to be analyzed.)
If there is plasma only in the test tubes and the sam­ple volume is lower than the dead volume, air may get aspirated and/or a “Probe Crash” or “Sampling Error” may occur, preventing correct analysis results from being obtained.
If there is plasma only in the tubes, it is not recom­mended to use the following tubes with a double wall structure because their narrow inner diameters and higher bottom positions cause a higher possibility of a short sample error.
- VACUTAINER Plus Plastic Citrate Tube, 13 mm × 75 mm, 2.7 mL
- VACUETTE Sandwich Coagulation Tube, 13 mm × 75 mm, 3.0 mL/3.5 mL
If the sample volume of centrifuged samples is lower than the dead volume, blood cells may get aspirated, preventing correct analysis results from being obtained.
Revised April 2004 - 2.0_en
Sysmex CA-500 series 5-17
Page 52
Operation
Caution
When using test tubes with an outside diameter of 15 mm, remove the 13 mm tube adapters from the sam­ple rack beforehand. When using test tubes with an outside diameter of 10 mm, remove the adapters from the sample rack and mount the optional holder No. 113 (φ10 mm test tube adapter) beforehand.
When using sample cups, set the optional holder No. 70 beforehand.
When using sample cups, avoid setting samples which are low in plasma volume. Such samples will cause the error of “Sample Probe Crash”. For samples, prepare the required volume plus 100 µL.
Sampler
Sampler Rack Sensor
2. Pull out the sampler.
3. Set the sample rack in the sampler. Only one rack (10 s amples) can be set.
Important
Unless the sample rack is correctly set, instrument fail­ure will result. Press in the sample rack sensor (lower left) so that the sample rack is level.
5-18 Sysmex CA-500 series
Revised September 2003 - 2.0_en
Page 53

5.11 Set Sample Nos.

Set Sample Nos. and Analysis Parameters

With the instrument, all samples are analyzed according to the analysis order.
Analysis information for 10 samples (1 rack filled) can be set at a time. There are four procedures for setting sample ID Nos. and analysis
parameters:
Manual setting with the numeric keys and analysis parameter
Receiving sample ID Nos. and analysis parameters collectively
Manually entering sample ID Nos. so that analysis orders will be
Using sample ID Nos. that have been read with the optional bar-
Operation
keys.
from the host computer (Refer to “5.12 Manual Inquiry”.)
automatically received from the host computer (Refer to “5.13 Automatic Inquiry”.)
code scanner so that analysis information will be automatically received from the host computer. (Refer to “5.13 Automatic Inquiry”.)

Setting of Sample ID Nos.

Ready
Sysmex
Main Menu
1 PT 2 APTT 3 Fbg
-01
01-01 123-456-78901
01-02
01-03 01-04 01-05
Replace Rack? YES!
ID No. 123-456-78901 123-456-78902 123-456-78903 123-456-78904 123-456-78905
Repeat
4 TT 5 AT3
12345
HC IP
Group 1
7
4
1
0
C
ID No.
Q C
Start
89
5
6
3
2
Enter
Quit
Important
The instrument does not store analysis information. The information is erased when power is turned off.
1. Specify the rack to be set on the Root Menu screen.
Press [] and [] keys to move the cursor to the desired rack posi­tion.
2. Press [ID No. Entry] key.
The numeric keys screen will be displayed.
3. Enter sample ID No. and press [Enter] key. By pressing [C] key, you can erase one letter entered (back space
function). When entering a sample ID No. for QC, press [QC] key following a
numeral between 01 - 06.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-19
Page 54
Operation
Important
When an ID No. is not registered, this instrument automatically assigns it. After power turn-on, it begins with 000000000000001 (15 digits), which is incremented by 1 at a time.
Analysis results will not be stored in a QC File if a Standard Curve is not set and QC parameters are set as calculation parameters.
Note
To completely erase a registered sample ID No., position the cursor over the number and press first [C] key then [Enter] key.

Group Selection

4. Press [Quit] key on the numeric keys.
The screen returns to the Root Menu.
A combination of analysis parameters, selected from three menus, can be set. The analysis parameters that can be set are limited to those in the selected group.
1. Press [Test Group] key on the Root Menu.
The Group Setting screen will be displayed.
Note
When the cursor is on a rack which has been or is being analyzed, [Test Group] key is not displayed if an analysis parameter is already set.
2. Press [Group] key. Select a group from among three groups ([Group 1] - [Group 3]) by
using [] key, or [] key. For the group setting program, refer to “10.10 Setup of Test Group”.
5-20 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 55

Setting of Analysis Parameters

Operation
3. Press [Return] key.
The Root Menu will return and the analysis parameter will change to the parameter selected with the group setting program .
Note
Using this [Test Group] function, Fbg with alternative dilu­tion ratios can be displayed.
+ Fbg is a parameter that is analyzed with Fbg diluted to
1:20 (a half of the usual concentration).
- Fbg is a parameter that is analyzed with Fbg diluted to
1:5 (2 times of the usual concentration).
Ready
Sysmex
Stored
Data
Replace Rack? YES!
12345

Repeat

Standaed
QC
Curve
Main Menu 1 PT 2 APTT 3 Fbg4 TT 5 AT3
Rack ID No.
02-01 123-456-78901
01-01 123-456-78901 
01-02 123-456-78902 01-03 123-456-78903 01-04 123-456-78904 01-05 123-456-78905
Repeat
Ready
Sysmex
Stored
Data
Replace Rack? YES!
12345
Repeat
Standaed
QC
Curve
Main Menu 1 PT 2 APTT 3 Fbg4 TT 5 AT3
Rack ID No.
02-01 123-456-78901
01-01 123-456-78901 
01-02 123-456-78902 01-03 123-456-78903 01-04 123-456-78904 01-05 123-456-78905
HC IP
Group 1
ID No.
Entry
Test
Group
HC IP
Group 1
ID No.
Entry
Test
Group
Start
Prev
Next
HC
Special
Menu
Start
Prev
Next
HC
Special
Menu
1. Specify the samples to be set on the Root Menu screen
Press [] and [] keys to move the cursor to a sample to be set.
2. Set an analysis parameter by pressing analysis parameter keys.
Each time an analysis parameter key ([PT], [APTT], [Fbg], [TT], [AT3]) is pressed, signs “- (Not analyze)” and “O (Analyze)” change alternately.
Assigns consecutive sample ID Nos. for one rack (10 samples), with the cursor positioned at the top, and copies analysis parameter setting for all subsequent samples.
1. Using [] and [] keys, specify a sample you want to repeat.
2. In the Root Menu screen or ID No. Entry screen, press [Repeat] key.
Important
When sample ID No. is “0” or it is for QC or Standard Curve, [Repeat] cannot be executed.
Note
Sample ID No. cannot be carried beyond “-”. For example, the number following 1-99 will be 1-00.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-21
Page 56
Operation

5.12 Manual Inquiry

5.13 Automatic Inquiry

When the instrument is bidirectionally connected with the host compu­ter, analysis information for one rack (10 samples), ba sed on rack Nos., can be received prior to analysis. For this purpose, the host computer needs to be set in the conditions shown below. For detail, refer to “10.14 Devices to be connected”.
Connection: Connected Class: Class B Inquiry: Manual
1. Press [HC] key on the Root Menu. The instrument makes inquiries to the host computer for a specified
rack No. and receives analysis information. Each time analysis infor­mation is received for one sample, it is displayed on the screen.
2. Analysis information received is confirmed on the screen.
The following automatic inquiry process is performed depending on bar­code scanner connection status and order entry status.
Sample ID No. With manually entered order Without manually entered order
Without barcode scanner No inquiry to host computer
Analysis with automatically-
assigned sample ID No. and manu­ally entered order
With barcode scanner
Reading normal No inquiry to host computer
Analysis with sample ID No. from
barcode and manually entered order of barcode
Reading error No inquiry to host computer
Analysis with error No. and manu-
ally entered order

Automatic Inquiry (without barcode scanner)

When the instrument is bidirectionally connected with the host compu­ter, the instrument can receive, in advance, analysis information for one rack (10 samples) based on manually entered sample ID Nos.
The host computer must be set in the conditions given below. For detail, refer to “10.14 Devices to be connected”.
Host Computer Status: Connected
Barcode scanner: Not connected
No inquiry to host computer
No analysis
Inquiry made to host computer
using sample ID No. from barcode
Analysis with sample ID No. and
order from host computer
No inquiry to host computer
No analysis
Class: Class B Inquiry: Auto
1. Manually set sample ID Nos. Refer to “5.11 Set Sample Nos.”.
5-22 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 57
Operation
Ready
Sysmex
Main Menu 1 PT 2 APTT 3 Fbg
Rack ID No.
01-01 123-456-78901
01-02 123-456-78902 01-03 123-456-78903 01-04 123-456-78904 01-05 123-456-78905
Replace Rack? NO!
HC Inquiry Please wait.
4 TT 5 AT3
12345
HC IP
Group 1
Start
Prev
Next
2. Press [Start] key.
Instrument makes inquiries to the host computer using the manually entered ID No. for one sample at a time. It then displays the analysis information on the screen, and starts analysis when the inquiries for 10 samples are completed.

Automatic Inquiry (with barcode scanner)

When the instrument is bidirectionally connected with the host compu­ter, and the optional barcode scanner is fitted, the instrument can receive analysis information for one rack (10 samples), based on the ID Nos. read with the barcode scanner.
The host computer must be set in the conditions shown below. For detail, refer to “10.14 Devices to be connected”.
Host Computer Status: Connected
Barcode scanner: Connected
1. Set the samples in the sample rack.
Important
Inquiry is not performed for those samples on which
both ID Nos. and analysis parameters have been set.
Inquiry is not performed either for those samples on
which ID Nos. have not yet been set.
Class: Class B Inquiry: Auto
Revised January 2003 - 2.0_en
Set the samples with the barcode labels facing toward you for read­ing.
2. Pull out the sampler, set the sample rack, and push it in.
3. Press [Start] key.
The sample ID Nos. for one rack are read from the barcode labels, and inquiries are made to the host computer for those ID Nos. When the analysis information is received, it is displayed on the screen and analysis is started.
Caution
Remove any foreign matters, if exists, on barcode drive mechanism.
Note
Should any error occur in reading barcode labels, the sam-
ple ID Nos. become “ERR0000000001” and increase in sequence. Inquiry is not performed.
Sysmex CA-500 series 5-23
Page 58
Operation

5.14 Start Analysis

Note
Regarding the rack positions for those ID Nos. that have
been manually set, you can make inquiries to the host computer using the manually set ID Nos, without reading barcodes.
Inquiries are not made for samples with registered analysis
parameters.
With completion of analysis preparation and registration of analysis information, the instrument is now ready to start analysis.
1. Check the system status display of the instrument. Make sure that the Root Menu screen displays “Ready”.
2. Press [Start] key.
Ready
Sysmex
Sysmex
Stored
Data
Replace Rack? NO!
Continue
Analyzing Replace Rack? NO!
12345
Repeat
Group 1
Select Tube Position
Main Menu 1 PT 2 APTT 3 Fbg 4 TT 5 AT3
Rack ID No.
02-01 123-456-78901
02-02 123-456-78902 02-03 123-456-78903 02-04 123-456-78904 02-05 123-456-78905
HC IP
First Tube
HC IP
ID No.
Entry
INTERR
Prev
Next
HC
The screen confirming the first tube's initial position will appear.
3. Press [Continue] key or [First Tube] key. [Continue] key: Starts with the reaction tube that follows the last
used tube in the previous analysis.
[First Tube] key: Starts with the upper extreme-right tube in the
right-hand reaction tube rack.
Important
When you have pressed [First Tube] key inadvertently, causing the instrument to stop by error, then Continue information for the reaction tubes is lost. In this case, set the reaction tubes again from the first position, and press [Start] key.
Note
The first time the analysis is started after power switch on,
[Continue] key is not displayed.
5-24 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 59
Ready
Sysmex
Select Tube Position
Replace Rack? NO!
Rinsing Probe program was forced to run due to long interval.
HC IP
Operation
Note
When analysis begins, if 24 hours have been passed from
the previous run, following will be displayed and rinsing program will run automatically to replace rinse and buffer solutions in the hydraulic lines. It will take approximately 3 minutes.
Automatic sensitivity adjustment of the detector is per-
formed when the first analysis starts after power-ON, or when analysis starts after each interval of 24 hours. (Refer to “5.15 Automatic Sensitivity Adjustment of the Detector (for CA-530, CA-540, CA-550 and CA-560 only)”.)
4. When all analyses are over, the alarm sounds “pip, pip, peep”. If you want to continue analysis, when the message “Replace Rack?
No!” changes to “Replace Rack? YES!”, then samples can be set to the next sample rack or the same sample rack. (Refer to “5.18 Add Samples” for sample addition.)
Even during analysis, when the [INTERR] key changes to the [Start] key, it will be possible to analyze samples in the next rack.
Warning
During analysis, do not insert hands or fingers, either through the gap in the light shield cover or by opening the cover. This is to avoid the risk of injury. If you open the light shield cover during operation, the alarm will beep and the instrument will stop.
Caution
When the power switch is turned off during operation, the instrument will fail. Make sure the analysis has been completed and the instrument status display is “Ready” before turning off the power switch.
Important
During operation, the sampler cannot be pulled out as the lock mechanism is activated to prevent injury and damage to the instrument.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-25
Page 60
Operation
Note
RESTRICTIONS DURING OPERATION
Only [Stored Data], [↑], [↓], [Repeat], [ID No. Entry],
[HC], [Prev], [Next] in the Root Menu can be used during
operation.
Once registered, analysis information cannot be changed.

5.15 Automatic Sensitivity Adjustment of the Detector (for CA-530, CA-540, CA-550 and CA-560 only)

At the start of analysis at either of the times below, sensitivity adjustment of the detector is performed automatically.
Start of first analysis after power-ON
Start of analysis after each interval of 24 hours
Ready
Sysmex
Replace Rack? NO!
The Intensity of the LED in the  Immunoassay channel  is becoming weak. However analysis is still possible. Please contact your local SYSMEX representative for  replacement of the LED. Please press "Start" to continue  the analysis.
Start Cancel
IP
Note
When Chromogenic Method parameters are included in the parameters selected for group settings, adjustment is per­formed for the Chromogenic Method detectors. When Immu­nology Method parameters are included, adjustment is per­formed for the Immunology Method detectors.
First, 200 µL Owren’s Veronal Buffer is aspirated and dispensed in the reaction tube. Then, using the buffer, automatic sensitivity adjust­ment is performed in the detectors for the Chromogenic Method and Immunology Method (for CA-550 and CA-560 only) respectively.
If the results of sensitivity adjustment indicate that the detector (LED) must be replaced, then a message will appear prompting the operator to replace the detector. Contact your local service represent­ative.
The message varies as shown below according to the detector.
For the Chromogenic Method detector:
“The Intensity of the LED in the Chromogenic channel is becom­ing weak.”
For the Immunology Method detector :
“The Intensity of the LED in the Immunoassay channel is becoming weak.”
For both the Immunology Method detector and the Chromogenic
Method detector: “The Intensity of the LED in the Chromogenic and Immunoassay channels are becoming weak.”
5-26 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 61
Analysis can still be performed after the replace-detector mes­sage has appeared.
Press [Start] key to continue with analysis.
To cancel analysis, press [Cancel] key.

5.16 Display Analysis Result

Analysis Status Display (Root Menu screen)

Analysis status of each sample is displayed on the Root Menu screen.
A sample is distinguishable by ID No. and rack position.
Analyzing
Sysmex
Stored
Data
Replace Rack? NO!
Main Menu 1 PT 2 APTT 3 Fbg
Rack ID No.
02-01 123-456-78901
02-02 123-456-78902 02-03 123-456-78903 02-04 123-456-78904 02-05 123-456-78905
4 TT 5 AT3
12345
Repeat
HC IP
Group 1
ID No.
Entry
INTERR
Prev
Next
HC
1. Press [Prev] key, [Next] key, [] key, or [] key to search for sam-
ples whose analysis status you want to check.
2. Current analysis status of each of the applicable samples is displayed as follows:
Meanings of the analysis status symbols:
– : Analysis is not ordered for the parameter.
: Analysis is requested.
: Analysis is in progress.
Operation
Note

Display and Printout of Sample Data

HC IP
All Seq 0/142(142)
PT%APTT
88.1
88.1
88.1
88.0
87.9
88.0
87.9
More
Mark
Main
Menu
Revised January 2003 - 2.0_en
Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
Graph
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
Prev Next
PT
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0
: Analysis is completed.
× : Analysis is not completed due to interruption or error.
1. Press [Stored Data] key on the Root Menu screen.
Analysis data will be displayed in the list format.
2. To display coagulation curves, press [Graph] key.
Move the cursor to the sample whose coagulation curve you wish to display, and press [Graph] key.
Sysmex CA-500 series 5-27
Page 62
Operation
Ready
Sysmex
0
Change
Scale
Replace Rack? YES!
60
Edit
ID No.
Output Delete
Stored Data 0/142(142) ID No. 123-456-78901 36.8ûC 01-01 12/01 17 : 28 (PT : 100.0%)
HC IP
123.4 sec(50)
123.4% 
--.--
--.--
 mg/dl
--.-­ERR [ 8] bH [ 56] dH [ 123] CH No. [1] 
Info
Return

5.17 Interrupt Analysis

Analyzing
Sysmex
Stored
Data
Replace Rack? NO!
Main Menu 1 PT 2 APTT 3 Fbg
Rack ID No.
02-01 123-456-78901
02-02 123-456-78902 02-03 123-456-78903 02-04 123-456-78904 02-05 123-456-78905
4 TT 5 AT3
12345
Repeat
HC IP
Group 1
ID No.
Entry
INTERR
Prev
Next
HC
For detail, refer to “6. Display and Processing of Analysis Results”.
The analysis process can be interrupted at any time.
1. Press [INTERR] key.
While analysis is being performed, [Start] key at the upper right cor­ner of the screen is replaced by [INTERR] key.
The Interruption Confirmation screen will be displayed.
Analyzing
Sysmex
Replace Rack? NO!
Interrupt sampling ?
AddReag AddOrder STAT
HC IP
Cancel
2. Press [AddReag] key, [AddOrder] key, [STAT] key, or [Cancel] key.
[AddReag] key: The interruption process starts and no new analy-
sis is performed.
[AddOrder] key: The order addition process starts.
[STAT] key: The process moves to the STAT sample process-
ing. Refer to “5.19 Analyze STAT Sample”.
[Cancel] key: The screen returns to the original one.
Note
While the Interruption Confirmation screen is on, the instru­ment is performing analysis. When all analyses are completed with this screen on, the Root Menu screen returns and “Ready” is displayed.
5-28 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 63
Ready
Sysmex
Replace Rack? YES!
1 PT 2 APTT 3 Fbg
Rack ID No. 02-01 123-456-78901 02-02 123-456-78902 02-03 123-456-78903
02-04 123-456-78904
02-05
Order remains. OK?
Start Cancel
4 TT 5 AT3
12345
HC IP
Group 1
Operation
3. During the analysis interruption process, [Resume] key is displayed
at the upper right corner of the screen. When you want to interrupt analysis, do not press [Resume] key. When [Resume] key is pressed:
Analysis restarts.
When [Resume] key is not pressed:
When analysis for dispensed samples is completed, and there are
still parameters remaining to be analyzed, a screen is displayed to confirm analysis continuation. If reagent is to be added, add it at this time.
4. When you want to interrupt analysis, press [Cancel] key.
[Start] key: Analysis restarts. [Cancel] key: Analysis is canceled and the screen returns to the Root
Menu.
Note
The samples whose analysis has been canceled have “X” dis­played in the Work List.

5.18 Add Samples

Sysmex
Main Menu Group 1
1 PT 2 APTT 3 Fbg
Rack ID No. 02-01 123-456-78901
02-02 123-456-78902 02-03 123-456-78903 02-04 123-456-78904
02-05 000-00001
Waiting Replace Rack? YES!
4 TT 5 AT3
12345
Repeat
HC IP
ID No.
Entry
Resume
AddOrder
Cancel
Next
HC
1. Press [INTERR] key.
The Interruption Confirmation screen will be displayed.
2. Press [AddOrder] key on the Interruption Confirmation screen.
Analysis of the new sample will be interrupted, and the Sample Addition screen will be displayed.
3. When the analysis interruption process is completed, “Replace Rack? YES!” is displayed.
Then the sampler can be pulled out.
4. Pull out the sample rack only, without opening the light shield cover, and register the parameters to be added.
Important
Registered sample ID Nos. and orders cannot be modified.
Additional orders can only be registered to a position after the final rack position of the registered orders.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-29
Page 64
Operation
5. Register sample ID Nos. and analysis parameters for additional orders.
When entering sample ID Nos. manually, press [ID No. Entry]
key, and input sample ID Nos. Refer to “5.11 Set Sample Nos.”, or “Automatic Inquiry (without barcode scanner)” of “5.13 Automatic Inquiry”.
Press [HC] key to receive sample ID Nos. and analysis parame -
ters from the host computer. Refer to “5.12 Manual Inquiry”. This registration process is not required when sample ID Nos. are read using an optional barcode scanner, and analysis parameters are sent automatically from the host computer.
Note
To cancel addition of orders, press [AddOrderCancel] key.
6. Press [Resume] key. Analysis restarts.

5.19 Analyze STAT Sample

Waiting
Sysmex
Sysmex
Replace Rack? NO!
12345
Waiting Replace Rack? YES!
4 TT 5 AT3
12345
Main Menu Group 1
1 PT 2 APTT 3 Fbg 4 TT 5 AT3
Rack ID No.
01-07 123-456-78901 01-08 123-456-78902 01-09 123-456-78903 01-10 123-456-78904
01-00 000-00001
Main Menu
1 PT 2 APTT 3 Fbg
Rack ID No.
01-06 123-456-78901 01-07 123-456-78902 01-08 123-456-78903 01-09 123-456-78904
01-00 000-00001
HC IP
ID No.
Entry
HC IP
Group 1
ID No.
Entry
Cancel
STAT
HC
Start
STAT
Cancel
STAT
HC
1. Press [INTERR] key. The Interruption Confirmation screen will come on.
2. Press [STAT] key. Analysis of the new sample will be interrupted.
3. When the analysis interruption process is over, “Replace Rack? YES!” is displayed, and the sampler can now be pu lled out.
4. Set a STAT sample in a STAT sample rack.
5. Register ID Nos. and analysis parameters for STAT samples. When entering sample ID Nos. manually, press [ID No. Entry] key,
and input sample ID No. Refer to “5.11 Set Sample Nos.”. Press [HC] key to receive sample ID Nos. and analysis parameters
from the host computer. Refer to “5.12 Manual Inquiry”.
Note
To cancel registration of STAT samples, press [Cancel STAT] key.
5-30 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 65
Waiting
Sysmex
Replace Rack? YES!
1 PT 2 APTT 3 Fbg
Group 1
4 TT 5 AT3
HC IP
Operation
6. Press [Start STAT] key.
Analysis of a STAT sample starts.
Rack ID No. 01-06 123-456-78901 01-07 123-456-78902 01-08 123-456-78903 01-09 123-456-78904
01-00 000-00001
Replace STAT Sample. OK?
Start Cancel
12345

5.20 Emergency Stop

Note
When [Start STAT] key is pressed without pulling out the sampler, the message “Replace STAT Sample. OK?” will appear. Press [Cancel] key, and STAT Sample Order Setting screen returns without starting STAT analysis.
7. When STAT analysis is completed, analysis of the original rack con­tinues.
Note
Rack Nos. 01-00, 02-00, 03-00, and 99-00 of STAT sam-
ples are displayed and 99-00 is followed by 01-00.
While STAT sample is being dispensed, [INTERR] key is
not displayed.
By pressing the Mechanical Stop switch, analysis operation can be stopped immediately.
Revised January 2003 - 2.0_en
Important
Should the analyzer need to be shut down in an emergency, such as an unexpected outage of power to the laboratory, immediately turn off the power switch to the main unit.
Note that the above-mentioned switch differs from the emergency mechanical stop switch (red switch) on the front of the instrument. When the red mechan­ical stop switch is pressed, the mechanical system comes to a stop but the power supply is not turned off.
Sysmex CA-500 series 5-31
Page 66
Operation
Mechanical Stop Switch
1. Press the Mechanical Stop switch (red switch). Analysis operation will stop immediately.
2. When the Mechanical Stop switch is pressed during photo detection, the screen for confirming whether to stop photo detection comes on.
Analyzing
Sysmex
Yes
Sysmex
Replace Rack? NO!
Analyzing Replace Rack? NO!
Mechanical Stop
Mechanical Stop
HC IP
Stop Analysis?
Please wait until analysis is completed
HC IP
Discard/Start rinse operation?
3. To stop analysis, press [Yes] key. When the sample tube is not in place at the detector, the following
screen comes on:
Caution
When photo detection is stopped and discard/rinse operation starts, the sample data being analyzed is discarded. Reanalyze sample data marked with “×” in the Work List.
If the sampler table is pulled out when the sample probe is lowered in the tube, the sample probe will be permanently damaged.
4. To start discard/rinse operation, press [Yes] key.
The screen will display the message “Discard/Rinse in progress. Please wait.”.
The discard/rinse operation will be executed.
Note
No
Yes
In what case is rinse operation not possible The system is stopped and starting operation could damage the sample probe.
Analyzing
Sysmex
Mechanical Stop
Replace Rack? NO!
HC IP
5. When rinse operation cannot be performed, press [No] key.
6. When an unanalyzed order remains, the screen for choosing whether to continue analysis comes on.
Discard/Rinse in progress Please wait.
5-32 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 67
Operation
Ready
Sysmex
Replace Rack? YES!
1 PT 2 APTT 3 Fbg
Rack ID No. 01-06 123-456-78901 01-07 123-456-78902 01-08 123-456-78903 01-09 123-456-78904 01-00 000-00001
Order remains. OK?
Start Cancel
4 TT 5 AT3
12345
Group 1
HC IP
7. Press [Start] key or [Cancel] key.

5.21 Shutdown

Turn OFF the Power

Confirm that the instrument status is “Ready” before turning off the power switch.

Operation after Analysis Completion

At the conclusion of the day's analyses or after the instrument has been run for at least 24 hours, perform the following daily maintenance:
[Start] key: Starts analysis. [Cancel] key: Interrupts analysis and returns the screen to Root
Menu.
1) Discard the used reaction tubes.
2) Dispose of waste fluids.
3) Remove dew from the reagent holders.
4) Clean the sample probe.
For detail, refer to “11. Maintenance and Supplies Replacement”.
Caution
If the power is turned off when the rinse bottle or waste bottle are lying flat, the solution may flow back into the instrument. Before turning off the power switch, confirm that the bot­tles are not lying flat.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 5-33
Page 68
6. Display and Processing of Analysis Results ........6-1
6.1 List Display/Graphic Display .......................................................6-1
6.2 Search ................................... ....................................... ............... 6-6
6.3 Sort in Sequence of Sample ID Nos. and Analyses ....................6-9
6.4 Select Display .............................................................................6-9
6.5 Edit ID No. .................................................................................6-11
6.6 Deletion .............................. ....................................................... 6-12
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 69
Display and Processing of Analysis Results

6. Display and Processing of Analysis Results

The instrument displays analysis results and information that helps inter­pret analysis results, and outputs them to external devices. This chapter describes the processing of analysis results, including display of stored data and output to external devices.

6.1 List Display/Graphic Display

The instrument can store analysis results of up to 300 samples (1500 tests) and reaction curves of the latest 600 tests. The stored data is kept even when the power switch is turned off and will be displayed unless erased.
Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
PT
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
HC IP
All Seq 0/142(142)
PT%APTT
88.1
88.1
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0
Mark
88.1
88.0
87.9
88.0
87.9
1. Press [Stored Data] key on the Root Menu screen.
The analysis result will appear in List Display together with Local Menu of Stored Data.
By pressing [More] key, Local Menu can be changed-over. When a desired Local Menu is not displayed, press [More] key.
Graph
Prev Next

List Display

More
Main
Menu
When [Stored Data] key on the Root Menu is pressed, the Analysis Result screen appears. While the Graphic Display screen is on, press [Return] key on Local Menu to view the analysis results.
Note
Using [Select Display] and [ID No./Seq] keys, you can change a stored data to be displayed and its chronological sequential arrangement. When the power switch is turned off or the display is returned to the Main Menu, however, all stored data return to the previous list in chronological analysis sequence.
The Stored Data List Display screen can display up to 8 samples per screen. When the first List Display screen appears after the power switch is turned on, data of the last 8 samples is displayed. When latest analyses are completed, those sample data are automatically added to the last of the List. The List Display screen is composed of the analysis results, showing 3 data parameters at a time, and the sample information. Use [] and [] keys to switch between pages.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 6-1
Page 70
Display and Processing of Analysis Results
Ready
Sysmex
Replace Rack? YES!
ID No.
123-456-78901 123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906
123-456-78907
123-456-78908
Graph
Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
Graph
sec
11.5 88.232.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
Prev Next
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
Prev Next
HC IP
All SeqStored Data
PT
All Seq 0/142(142) PT
2/142(142)
PT%APTT
88.1
88.1
88.1
88.0
87.9
88.0
87.9
More
HC IP
PT%APTT
88.1
88.1
88.1
88.0
87.9
88.0
87.9
More
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0 Mark
Main
Menu
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0
Mark
Main
Menu
Key operation on the List Display screen is as follows:
[Prev] key: Scrolls down one display screen (8 samples). [] key: Moves the cursor up one sample position.
When the cursor is at the top of the screen, the List scrolls down.
[] key: Moves the cursor down one sample position.
When the cursor is at the bottom of the screen, the List scrolls up.
[Next] key: Scrolls the List up one screen (8 samples). [] [] key: Turns over a page (scrolls sideways). [Mark] key: Attaches or deletes marking.
When you press [Mark] key on the List Display screen, a mark ( ) can be put on an analysis data in the current cursor position. The mark will appear at left side of the data.
When an analysis data is already attached with a mark, pressing [Mark] key causes the mark to be deleted.
Analysis Result Screen
When [Stored Data] key on the Root Menu screen is pressed, the Analy­sis Result screen appears first. This screen also appears when you press [] key on the Sample Information screen.
The Analysis Result screen displays the results for calculation para mete r and coagulation time that were selected in the test protocol and standard curve parameter setting. Analysis parameters are shown with 3 parameters on each page. [] and [] keys can be used to change parameters.
The following contents are displayed on the Analysis Result screen. ID No.:
Sample ID Nos. entered by operator, transmitted from host computer, and read by barcode scanner are displayed.
Calculation parameters, coagulation time, ∆OD/min: The calculation results and coagulation time or OD/min for each
parameter are displayed. If the calculation did not take place nor­mally for any parameter, one of the following is displayed.
***.* : Analysis data was not obtained due to an error or other cause.
- - -.- : The calculation parameter could not be calculated. +++.+ : The c alculated value was large and exceeded the number of
available display digits.
6-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 71
Display and Processing of Analysis Results
      
Abnormal Flag:
Abnormal flags displayed on the right or left of an analysis data indi­cate the following:
m (on right of ID No.): The data is a mean. The parameter ha s been
calculated from mean data of seconds or OD/min.
! (on right of ID No.): A dilution ratio other than 100% was used for
the data.
* (on left of data): Some kind of error has occurred, or there is
deviation in repeated analysis. > (on left of data): A data that exceeds the Upper Report Limits < (on left of data): A data that exceeds the Lower Report Limits + (on right of data): A data that exceeds the Upper Mark Limits
- (on right of data): A data that exceeds the Lower Mark Limits
Note
“*” has a higher display priority than “>” and “<” flags.
Ready
Sysmex
Replace Rack? YES!
DATE TIME RACKID No.
12/31
12:55
12/31
12:57
12/31
12:58
12/31
12:59
12/31
12:56
12/31
12:57
12/31
12:58
12/31
12:59
Graph
Prev Next
SEQ
123-456-78901
1
123-456-78902
2
123-456-78902
3
123-456-78902
4
123-456-78905
5
123-456-78906
6
123-456-78907
7
123-456-78908
8
All
HC IP SeqStored Data
More
0101
0102 0103 0104 0105 0106 0107 0108
0/142(142)
OUT
HP
HP HP HP HP HP HP HP
Mark
Main Menu
Sample Information Screen
When [] key is pressed on the first page of the Analysis Result Screen, the Sample Information screen will be displayed.
Contents of Sample Information Screen DATE: The date the analysis was conducted. TIME: The time the analysis result was obtained. SEQ: A sequential sample number counted from power turn-on. ID No.:
The sample No. entered by operator, transmitted from host computer, or read by barcode scanner.
RACK:
The rack number and tube position number are displayed as follows:
Rack No. Tube Position No.
Normal Sample, QC Data 01 - 99 01 - 10 STAT Sample 01 - 99 00 Standard Curve 00 01 - 06
OUT:
Data output flag: H: When output to the host computer, the flag disappears. P: When output to the built-in printer, the flag disappears.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 6-3
Page 72
Display and Processing of Analysis Results

Graphic Display

Using the cursor, specify a sample to be graphically shown on the List Display screen, then press [Graph] key on the Local Menu. This enables displaying the coagulation curve of the sample chosen from the stored data.
While Graphic Display is on, press [] key to display a sample one posi­tion above on the List; and press [] key to display a sample one position below.
By pressing [] and [] keys, a parameter for an identical sample can be shifted to another parameter.
(1) (2) (3)
Stored Data 0/142(142)
(4)
ID No. 123-456-78901 36.8˚C 01-01 12/01/2001 17 : 28 (PT : 100.0%)
(5) (6) (7) (8)
Sysmex
x 4
0
Ready Replace Rack? YES!
ERR [ 8] dH [ 123] bH [ 56] CH No. [1]
60
HC IP
123.4 sec(50)
123.4% 
--.--
--.--
 mg/dl
--.--
Info
(11) (12)
(13) (14)
(15) (16)
(9)
(10)
Change
Scale
Edit
ID No.
Output Delete
Return
The following contents are displayed on the Graphic Display screen.
1. Analysis Time Time of day the analysis result was obtained
2. ID No. Sample ID No. entered by operator, transmitted from host computer,
or read by barcode scanner
3. Analysis Date Date the analysis was conducted
4. Rack No./Tube Position No. Rack No. and tube position No. where sample was placed: 00 - 99: Rack No. (00 indicates standard curve analysis.) 00 - 10: Tube position No. (00 indicates STAT sample holder.)
5. Analysis Parameter Name/Dilution Ratio Analysis parameter and dilution ratio of the displayed reaction curve
6. Scale of Scattered Light Intensity If an enlarged graph is available, an enlarged graph with the scale
such as “x4” is displayed when [Change Scale] key is pressed.
6-4 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 73
Display and Processing of Analysis Results
7. Reaction curve A graph is plotted with time on the horizontal axis and the scattered
light intensity or light absorbance variation on the vertical axis. Time scale is shown at right end. When the reaction curve graph area is pressed, the Graphic Enlarge-
ment window will be displayed. Refer to “Graphic Enlargement Window” described later.
8. Error code Error code when an error occurs in analysis result When there is no error, 0 appears.
9. dH Increase of scattered light intensity in reaction process
10. CH No. Detection channel in which analysis was conducted
11. Abnormal Flag Abnormal flags displayed on the right or left of an analysis data.
12. Coagulation Time or Light Absorbance Variation Time taken for coagulation or change in light absorbance
13. Analysis Temperature Temperature of the detector at the start of analysis
14. Coagulation detection point Coagulation detection point set It is not displayed in case of analyses by Chromogenic Method and
Immunology Method.
15. Calculated Parameters Activity percentage, PT ratio, INR (PT), (Fbg concentration), etc.
16. bH Scattered light intensity at the start of analysis
Note
Reaction curve is not displayed when some error occurred dur­ing analysis and no analysis data was obtained. The curve is also not displayed for the mean data of repeated analysis.

Error Detail Window

When an analysis error has occurred, its detail can be displayed in the window.
1. Press [Info.] key on the Graphic Display screen to display the Error Detail window.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 6-5
Page 74
Display and Processing of Analysis Results
Ready
Sysmex
Stored Data ID No. 123-456-78901 01-01 12/01/2001 17:28 <PT :100.0%>
Replace Rack? YES!
0 60
HC IP
0/142(142)
33.0˚C
23.4 sec(50)
Temp Error
Return

Graphic Enlargement Window

Ready
Sysmex
Stored Data ID No. 123-456-78901 01-01 12/01/2001 17:28 <PT :100.0%>
Replace Rack? YES!
HC IP
0/142(142)
33.0˚C
23.4 sec(50)
2. Press [Return] key of Error Detail Window.
The Error Detail window will be closed.
The reaction curve graph on the Graphic Display screen can be enlarged on the window.
1. Press an area of the reaction curve graph on the Graphic Display screen.
The Graphic Enlargement window will be displayed.
2. Press an area of reaction curve graph on the Graphic Enlargement window.
The Graphic Enlargement window will be closed.
060

6.2 Search

With this program, a data can be moved to the top data point or the bot­tom data point in the List Display, and samples that agree with specified conditions can be searched from stored data.

Top Data/Bottom Data

Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
Top Bottom Search Seq
Edit
ID No.
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
Output
HC IP
All Seq 0/142(142) PT
PT%APTT
sec
Marked
AllClear
88.1
88.1
88.1
88.0
87.9
88.0
87.9
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0
Select
Display
ReturnDelete
The cursor in the screen can be moved to the top of the list or the end of the list in the List Display.
[Top] key: Displays the top of analysis data at the head of the List. [Bottom] key: Displays the bottom of analysis data at the tail of the List.
6-6 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 75

Search by ID No.

Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
All Seq 0/142(142) PT
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
HC IP
PT%APTT
sec
88.1
88.1
88.1
88.0
87.9
88.0
87.9
32.1
32.1
32.1
32.0
32.2
31.9
32.0
Display and Processing of Analysis Results
Analysis data of a specified sample ID No. can be searched from stored data.
1. Press [Search] key on the Stored Data List Display screen.
The Local Menu for search will appear.
2. Press [Search ID No.] key.
The Search by ID No. screen will appear.
ID No.
ID No.
12345678902
Search
Date
Ready Replace Rack? YES!
Ready Replace Rack? YES!
Search
Sysmex
Stored Data Search by ID No.
Sysmex
Stored Data Search by ID No.
Not found
HC IP
7
4
1
0
C
HC IP
Return
ID No.
89
5
2
Enter
Q C
Quit
3. Enter sample ID No. Using the ID No. numeric keys on the Search by ID No. screen, enter
a sample ID No. to be searched.
6
3
If [C] key is pressed when there is no entry, [QC] key will change to [STD] key.
4. Press [Enter] key on the ID No. numeric keys. Search by ID No. begins from the sample ID No. in the current cur-
sor position toward the latest sample. To cancel search by ID No., press [Quit] key on the ID No. numeric keys.
5. A List is displayed in which analysis data of the sample ID No. entered is listed.
When analysis data of the specified ID No. is found, a list of analysis data the List will indicate the analysis data at the head of the screen. The cursor is on the analysis data corresponding to the ID No.
When analysis data corresponding to an entered ID No. is not found, the Confirmation Message screen will appear.
Press [Conf.] key to return to the Search by ID No. screen.
Conf.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 6-7
Page 76
Display and Processing of Analysis Results

Search by Date

The first analysis data of a specified date can be displayed at the head of the screen.
1. Press [Search] key on the Stored Data List Display screen.
The Local Menu for search will appear.
2. Press [Search Date] key.
The Search by Date screen will appear.
Ready
Sysmex
MM/DD/YY
12/01/2001
Sysmex
Replace Rack? YES!
Ready Replace Rack? YES!
Stored Data Search by Date
Stored Data Search by Date
HC IP
Date
89
7
4
5
1
2
0/
C
HC IP
6
3
Enter
Quit
3. Enter Date. Using the date numeric keys in the Search by Date screen, enter a
date by which to search.
4. Press [Enter] key on the date numeric keys. Search by date begins from the sample in the current cursor position
toward the latest sample. To cancel search by date, press [Quit] key on the date numeric keys.
5. A List is displayed in which analysis data of the date entered is listed. When analysis data of the specified date is found, the a list of analy-
sis data will be displayed. The cursor is on the first analysis data cor­responding to the date entered.
When analysis data corresponding to the date entered is not found, the Confirmation Message screen will appear.
Press [Conf.] key to return to the Search by Date screen.
Not found
Conf.
6-8 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 77
Display and Processing of Analysis Results

6.3 Sort in Sequence of Sample ID Nos. and Analyses

With this program, stored analysis data can be listed in a specified sequence.
The Stored Data List Display screen displays [Seq] key when the data are listed in the sequence of ID No. and [ID No.] key when the data are listed in the chronological analysis sequence.
Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
Top Bottom Search Seq
Edit
ID No.
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
Output
HC IP
All ID No. 0/142(142)
PT
PT%APTT
88.1
88.1
88.1
88.0
87.9
88.0
87.9
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0
Select
Display
ReturnDelete
Marked
AllClear
[ID No.] key: The analysis data are listed as rearranged in the sample
ID No. sequence.
[Seq] key: The analysis data are listed as rearranged in the chrono-
logical analysis sequence.
Note
While analysis is in progress, the program is automatically run in the analysis sequence.

6.4 Select Display

All Data

Sysmex
Stored Data
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
Ready Replace Rack? YES!
Mean CancelAll
All Seq 0/142(142)
PT
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
Not
Output
HC IP
PT%APTT
sec
88.1
88.1
88.1
88.0
87.9
88.0
87.9
32.1
32.1
32.1
32.0
32.2
31.9
32.0
With this program, the type of data to be list-displayed can be selected from all stored data.
As the condition for displaying a list, All Data can be specified.
1. Press [Select Display] key on the Stored Data List Display screen.
The Local Menu for selecting a display will appear.
2. Press [All] key on the Local Menu.
All analysis data stored are displayed in the sequence chosen from ID No. sequence and analysis sequence.
To cancel selection, press [Cancel] key. The Stored Data List Dis­play screen will return.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 6-9
Page 78
Display and Processing of Analysis Results

Mean Data

Ready
Sysmex
Stored Data Select Display
Replace Rack? YES!
No Data
Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901 123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907
123-456-78908
Top Bottom Search Seq
Edit
ID No.
All Seq 0/142(142) PT
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
Output

Not Output

HC IP
HC IP
PT%APTT
88.1
88.1
88.1
88.0
87.9
88.0
87.9
Marked
AllClear
Conf.
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0
Select
Display
ReturnDelete
As the condition for displaying a list, Mean Data can be specified.
When there is no sample to be displayed, the screen as shown at the left will appear.
1. Press [Select Display] key on the Stored Data List Display screen.
The Local Menu for selecting a display will appear.
2. Press [Mean] on the Local Menu.
From among the stored data, only mean data are displayed in the sequence chosen, either ID No. sequence or analysis sequence.
To cancel selection, press [Cancel] key. The Stored Data List Dis­play screen will return.
As the condition for displaying a list, the data that have not yet output can be specified.
Ready
Sysmex
IP Not Output HC Not Output
Sysmex
Graph
12:55
12:57 12:58 12:59 12:56 12:57 12:58 12:59
Replace Rack? YES!
Ready Replace Rack? YES!
Prev Next
SEQ
123-456-78901
1
123-456-78902
2
123-456-78902
3
123-456-78902
4
123-456-78905
5
123-456-78906
6
123-456-78907
7
123-456-78908
8
All Seq 0/142(142)Stored Data
Stored Data Select Display
DATE TIME RACKID No.
12/31
12/31 12/31 12/31 12/31 12/31 12/31 12/31
1. Press [Select Display] key on the Stored Data List Display screen.
The Local Menu for selecting a display will appear.
HC IP
2. Press [Not Output] key on the Local Menu.
The Not Output Data Selection screen will be displayed.
To cancel the selection, press [Return] key. The Stored Data List Display screen will return.
v
3. On the Not Output Data Selection screen, press the key to select the Not Output data to display. The selected data is indicated with a “v”.
Return
To cancel the selection, press [Return] key. The screen will return to the Stored Data List Display.
HC IP
0101
0102 0103 0104 0105 0106 0107 0108
Mark
OUT
HP
HP HP HP HP HP HP HP
The “v” alternately appears or disappears each time the key is pressed.
IP Not Output: Stored analysis data that have not yet output to the
built-in printer are displayed in the sequence cho­sen, either ID No. sequence or analysis sequence.
HC Not Output: Stored analysis data that have not yet output to the
host computer are displayed in the sequence cho-
Main
More
Menu
sen, either ID No. sequence or analysis sequence.
6-10 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 79

6.5 Edit ID No.

Display and Processing of Analysis Results
Note
When an output was performed with the display condition
“Not output data” selected, the output flag will be changed but the data will remain in the “Not output data” display even though the data has been output.
In this case, exit from the “Not output data” display by
pressing [Main Menu] key, or change the display condi­tion, to update the contents of the “Not output data” dis­play.
With this program, ID No. of cursor-specified analysis data can be changed.
1. Press [Stored Data] key on the Root Menu screen.
Ready
Sysmex
Stored Data Edit ID No.
Replace Rack? YES!
Date/Time
12/01 20 : 28
Rack
3401
Current ID No.
000000000000035
New ID No.
000000000000035
HC IP
7
4
1
0
C
ID No.
89
5
2
Enter
Q C
Quit
2. Press [] key or [] key to select data.
3. Press [More] key.
4. Press [Edit ID No.] key on the Stored Data List Display screen.
The Edit ID No. screen will display Date/Time, Rack position, and sample ID No.
5. Enter a new sample ID No. on the ID No. numeric keys.
6. Press [Enter] key or [Quit] key on the ID No. numeric keys. [Enter] key: Changes sample ID No. and returns you to the Stored
6
3
[Quit] key: Cancels editing sample ID No. and returns you to the
Data List Display screen.
Stored Data List Display screen.
Caution
There should be thorough control of sample ID Nos. within the facility, so that when the sample ID No. is changed, the new No. can be used to identify the pa tient.
Note
When ID No. sequence is chosen instead of analysis sequence, change in sample ID No. will not automatically cause change in arrangement of the samples. By pressing [Seq] key and then [ID No.] key on the Stored Data List Display screen, the sam­ple arrangement will return to normal.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 6-11
Page 80
Display and Processing of Analysis Results

6.6 Deletion

With this program, Stored Data can be deleted. Data to delete can be specified on the Local Menus for Current Data,
Marked Data, and All Data.
Once deleted, data cannot be restored. Check carefully in advance before deleting data.

Current Data

From the analysis data, the data in the current cursor position can be deleted.
1. Press [Delete] key on the Stored Data List Display screen.
The Local Menu for deletion will appear.
Ready
Sysmex
Stored Data
Replace Rack? YES!
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
HC IP
All Seq 0/142(142) PT
PT%APTT
sec
88.1
88.1
88.1
88.0
87.9
88.0
87.9
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0
2. Press [Current] key on the Local Menu.
The Deletion Confirmation screen will appear.
Note
Sysmex
Stored Data
Marked All CancelCurrent
Ready Replace Rack? YES!
Current data will be deleted.
HC IP
3. Press [Set] key or [Cancel] key on the Deletion Confirmation screen. [Set] key: Deletes analysis data in the current cursor position. [Cancel] key: Cancels deletion and returns you to the Local Menu
for deletion.
CancelSet
6-12 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 81

Marked Data

Ready
Sysmex
Replace Rack? YES!
ID No.
123-456-78901 123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906
123-456-78907
123-456-78908
HC IP
All SeqStored Data PT
PT%APTT
sec
11.5 88.232.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
88.1
88.1
88.1
88.0
87.9
88.0
87.9
2/142(142)
sec
32.1
32.1
32.1
32.0
32.2
31.9
32.0 Mark
Display and Processing of Analysis Results
From the analysis data, the marked data can be deleted.
1. Press [Mark] key on the Stored Data List Display screen and make a
mark. When [Mark] key is pressed, the analysis data in the current cursor
position will be marked ( ). If the mark is already there, it will disap­pear.
Press [Marked All Clear] key to remove all marks.
Graph
Sysmex
Stored Data

All Data

Sysmex
Stored Data
Prev Next
Ready Replace Rack? YES!
Marked data will be deleted.
Ready Replace Rack? YES!
All data will be deleted.
More
HC IP
HC IP
Main
Menu
CancelSet
2. Press [More] key.
3. Press [Delete] key on the Stored Data List Display screen.
The Local Menu for deletion will appear.
4. Press [Marked] key on the Local Menu.
The Deletion Confirmation screen will appear.
5. Press [Set] key or [Cancel] key on the Deletion Confirmation screen.
[Set] key: Deletes marked analysis data. [Cancel] key: Cancels deletion and returns you to the Local Menu
for deletion.
All data can be deleted.
1. Press [Delete] key on the Stored Data List Display screen.
The Local Menu for deletion will appear.
2. Press [All] key on the Local Menu.
The Deletion Confirmation screen will appear.
3. Press [Set] key or [Cancel] key on the Deletion Confirmation screen.
[Set] key: Deletes all analysis data. [Cancel] key: Cancels deletion and returns you to the Local Menu
for deletion.
CancelSet
Revised January 2003 - 2.0_en
Sysmex CA-500 series 6-13
Page 82
7. Output .......................................................................7-1
7.1 Automatic Printout of Analysis Data ............................................7-1
7.2 Output of Analysis Data ..............................................................7-1
7.3 Example of Printout .....................................................................7-3
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 83

7. Output

7.1 Automatic Printout of Analysis Data

The instrument can automatically output analysis data to the built-in printer or the host computer. Refer to “10.2 Setup of Automatic Transfer/ Printout”.

7.2 Output of Analysis Data

The instrument can output analysis data to the built-in printer or the host computer.
Sysmex
Stored Data
ID No.
123-456-78901
123-456-78902 123-456-78902 123-456-78902 123-456-78905 123-456-78906 123-456-78907 123-456-78908
Ready Replace Rack? YES!
sec
11.5 88.2 32.0
11.6
11.4
11.5
m
11.7
11.6
11.8
11.5
HC IP
All Seq 0/142(142)
PT
PT%APTT
sec
88.1
32.1
88.1
32.1
88.1
32.1
88.0
32.0
87.9
32.2
88.0
31.9
87.9
32.0
The data to be output to an external device can be specified via Local Menu of Current Data, Marked Data, and All Data. To display this menu, press [Output] key on the Stored Data List Display screen.
Output
Marked All CancelCurrent

Current Data

Ready
SysmexSysmex
Stored Data Select Device
Replace Rack? YES!
IP Graph
IP List
Host Computer
HC IP
Return
The data in the cursor position can be output to the built-in printer or the host computer.
1. Press [Current] key on the Local Menu for external output.
The Select Device screen will appear.
2. Select the device on the Select Device screen.
[IP Graph] key: Analysis data (with graph) at the cursor
position is output to the built-in printer.
[IP List] key: Analysis data (without graph) at the cursor
position is output to the built-in printer.
[Host Computer] key: Analysis data at the cursor position is output
to the host computer. (Displayed only when “Connected” is selected.)
[Return] key: Cancels the device selection and returns the
screen to the Local Menu for External Out­put.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 7-1
Page 84
Output
Important
Referring to “IP Graph”, when Automatic Transfer/ Printout Setting is formatted for printout with graph, a graph will be attached; when it is formatted for data analysis printout, analysis result is printed out. Refer to “10.2 Setup of Automatic Transfer/Printout” for the procedures.
However, analysis result will be lost if power has been turned OFF then ON, and the data with graph will be printed instead.
When the graph is not stored, the data without graph will be printed even if the data with graph is selected. (The graph is stored for the latest 600 tests.)
The scale of the coagulation curve automatically changes depending on the error.
Referring to “IP Graph” and “IP List”, when Automatic Transfer/Printout Setting is formatted for auto change depending on the error, an analysis data will be attached.
Ready
Sysmex
Stored Data Output
Replace Rack? YES!
Output in progress
Marked Data
HC IP
3. Output of analysis data in the current cursor position begins. The screen of output in progress will appear. If [Cancel] key is pressed on the screen of output in progress, the
output stops and the Select Device screen returns.
Cancel
The marked data can be output to the built-in printer or the host compu­ter.
1. Press [Mark] key on the Stored Data List Display screen to make a mark.
When [Mark] key is pressed, the analysis data in the current cursor position will be marked ( ). When the mark is already there, it will disappear.
Press [Marked All Clear] key to remove all the marks.
2. Press [More] key.
7-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 85

All Data

Output
3. Press [Output] key on the Stored Data List Display screen.
The Local Menu for external output will appear.
4. Press [Marked] key on the Local Menu to display the Select Device
screen.
5. Select the device on the Select Device screen.
6. Output of marked analysis data begins. The screen of output in progress will appear. If [Cancel] key is pressed on the screen of output in progress, the
output stops and the Select Device screen returns.
All data can be output to the built-in printer or the host computer.
1. Press [All] key on the Local Menu for external output. The Select Device screen will appear.
2. Select a device on the Select Device screen.

7.3 Example of Printout

3. Output of all analysis data begins. The screen of output in progress will appear. If [Cancel] key is pressed on the screen of output in progress, the
output stops and the Select Device screen returns.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 7-3
Page 86
8. Quality Control .........................................................8-1
8.1 Quality Control Methods ......................... ... ... ... ... .... ... ..................8-1
8.2 QC File Setting ............................................................................8-1
8.3 Execute Quality Control ..............................................................8-5
8.4 Display QC Charts ......................................................................8-5
8.5 Delete QC File ........................... ... ... ... .........................................8-7
8.6 Delete QC Data .. ... ... ........................................................... ... ... ..8-8
8.7 Print QC data .................. ... ... .... ... ... ... .........................................8-9
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 87

8. Quality Control

8.1 Quality Control Methods

Quality Control
Quality control is performed to obtain high-reliability data over long periods and also to constantly monitor the status of the instrument to pre­vent problem in advance.
This instrument analyses control plasma and other standard samples (QC samples) and performs statistical control of the results.
With the CA-500, two QC Programs are available: X
control: Uses an average data of two consecutive analyses made on
a QC sample (control plasma or pooled plasma). Because an average is used as control data, this control method suf­fers virtually no effects of reproducibility during analysis.
L-J control: Uses data of a single analysis on a QC sample. The L-J
control is susceptible to effects of reproducibility so that its control range is wider than X
control.

8.2 QC File Setting

Replicates Sets the times of quality control analysis. Parameter Sets the parameters for analysis results that are the object of quality control.
Upper Stop If a QC data exceeds this upper limit, the instrument stops analysis. Upper Flag If a QC data exceeds this upper lim it, the instrument flags the data as an error. Target Target value of QC data
Limit
Lower Flag If a QC data exceed s this lower lim it, the instrument flags the data as an error.
Lower Stop If a QC data exceeds this lower limit, the instrument stops analysis. Lot No. Sets a lot No. for samples to be used in quality control. EXP. Sets an expiry date on samples to be used in quality control.
To execute the QC Program, it is necessary to set replication, control parameters, control limits, Lot No., and expiry date.
The parameters to be set for QC File are:
Revised January 2003 - 2.0_en
Sysmex CA-500 series 8-1
Page 88
Quality Control
Ready
Sysmex
Quality Control File No. PT QC01 Mean
 SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
Replace Rack? YES!
%
100.0 
3.3 %
3.3
60
Settings
Change
Scale
HC IP
Lot. No TR-463 EXP. 12/31/2001
Cursor
100.0% 12/01/2001 15:09
110 107 100 93 90
1. On the Quality Control screen, display the QC Chart of the file to set parameters for.
For how to display QC Chart, refer to “Select QC Chart” of “8.4 Display QC Charts”.
2. Press [Settings] key on the Quality Control screen. The current settings of the selected file are displayed on the Quality
Control Setting screen.
Delete
Sysmex
QC Settings File No. PT
Replicates Parameter Limit Upper Stop Upper Flag Target Lower Flag Lower Stop Lot No. EXP.
Select
Print
Ready Replace Rack? YES!
File
QC01
2
PT% 
110.0%
107.0%
100.0%
93.0%
90.0% TPC747 12/31/2001
Auto
Calc.
Select
Test
HC IP
Next
Option
Main
Menu
Return
Using [] and [] keys, move the cursor to select the parameters to be set.
3. Set replicates using [Next Option] key. Each time [Next Option] key is pressed, “1” and “2” are changed
alternately.
4. Set a control parameter using [Next Option] key. Each time [Next Option] key is pressed, a parameter set as a stand-
ard curve parameter is changed over.
Note
As control parameters, you can select the analysis parame-
ters set in Test Name and Standard Curve Parameter set­ting.The parameters linked to the standard curve are selected from the parameters which were set at link source.
When changing control parameters, delete all control data
of a QC File to be changed.
Important
Analysis results will not be stored in a QC File if a Stand­ard Curve is not set and QC control parameters are set as calculation parameters.
8-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 89
Quality Control
Ready
Sysmex
QC Settings File No. PT QC01 
Replicates Parameter Limit Upper Stop Upper Flag Target Lower Flag Lower Stop Lot No. EXP.
QC Settings File No. PT QC01 
Replicates Parameter Limit Upper Stop Upper Flag Target Lower Flag Lower Stop Lot No. EXP.
Replace Rack? YES!
Ready
Sysmex
Replace Rack? YES!
Not correct limit value.
1 PT% 
110.0%
107.0%
100.0%
93.0%
90.0% TPC747 12/31/2001
1 PTact 
15.0%
107.0%
100.0%
93.0%
90.0% TPC747 12/31/2001
Auto
Calc.
HC IP
Upper Stop
7
89
4
5
1
2
0.
C
HC IP
Next
Return
Option
Enter
Quit
5. Enter control limits using the numeric keys.
When the “Limit” parameter is selected, [Next Option] key will be changed to [Numeric] key. Press [Numeric] key to display the
6
3
numeric keys.
Important
In setting control limits, the following conditions have to be met.
Lower Stop Lower Flag Upper Flag ≤ Upper Stop Lower Stop < Upper Stop
If the Quality Control Setting screen is quit without satisfying the above conditions, the cursor will be positioned at the upper control limit and the error message “Not correct limit value” will appear. Since the cursor will not move, enter proper set­tings.
Revised January 2003 - 2.0_en
Note
Press [Auto Calc.] key, and the limit values will be automati­cally calculated from QC data stored in the file, as in the fol­lowing.
Target: Mean of control data
Upper Flag: Mean of control data + 2SD of control data
Lower Flag: Mean of control data - 2SD of control data
Upper Stop: Mean of control data + 3SD of control data
Lower Stop: Mean of control data - 3SD of control data
6. Press [Enter] key on the numeric keys.
The setting is changed to the entered numeral and the cursor moves to the next parameter. For a parameter whose setting is not to be
changed, change the cursor position with [] and [] keys.
7. Press [Quit] key.
Setting by the numeric keys is completed.
Sysmex CA-500 series 8-3
Page 90
Quality Control
Ready
Sysmex
A
G
M
S
X
Sysmex
Replace Rack? YES!
TRC747
BC
H
N
T
Y
Ready Replace Rack? YES!
1 PT% 
110.0%
107.0%
100.0%
93.0%
90.0% TPC747
12/31/2001
QC Settings File No. PT QC01 
QC Settings File No. PT QC01 
Replicates Parameter Limit Upper Stop Upper Flag Target Lower Flag Lower Stop Lot No. EXP.
I
O
U
Z
Lot.No
D
J
P
V
BS
HC IP
EF
K
L
R
Q
Enter
W
Quit
CHG.
HC IP
EXP.
89
7
4
5
1
2
/
0
C
6
3
Enter
Quit
8. Set Lot No. By moving the cursor to Lot No., [Next Option] key will be changed
to [Lot No] key. Press [Lot No.] key to display the Lot No. Entry Screen. Each time the [CHG.] key is pressed, the enter mode changes from
capital alphabet small alphabet the numeric keys in this order. Lot No. up to 12 digits can be set.
9. Press [Enter] key on the Lot No. Entry screen. The setting is changed to the Lot No. entered and the cursor moves to
the next parameter. To cancel setting, press [Quit] key, and the screen will return to the Quality Control screen without changing the setting.
10. Enter Expiry date using the numeric keys. When the cursor is moved to “EXP.”, [Next Option] key will be
changed to [Date] key. Press [Date] key to display the numeric keys. Using the numeric keys, enter the date in accordance with System
Setting - Date Format.
11. Press [Enter] key on the numeric keys. The setting will be changed to the expiry date entered and the cursor
will move to the next parameter. To cancel setting, press [Quit] key, and the screen will return to
Quality Control screen without changing the setting.
Sysmex
QC Setting
Ready Replace Rack? YES!
RENEW SETTING ?
HC IP
Note
When the entered date is in a wrong format, the expiry
date will not be corrected by pressing [Enter] key. The cursor also does not move. Enter again in a correct format.
If you attempt to use the QC File that has passed its expiry
date, then the confirmation screen with the error message will appear immediately after [Start] key is pressed.
12. Press [Return] key on the Quality Control Setting screen. The Renew Confirmation screen will appear.
13. Press [FIX] key, [Continue] key, or [Cancel] key. [FIX]: Renews the setting and returns to the Quality Control
screen.
[Continue]: Returns to the Quality Control Setting screen and
allows continuous operation.
[Cancel]: Cancels the renewed setting and returns to the Quality
ContinueCancel FIX
Control screen.
Revised January 2003 - 2.0_en
8-4 Sysmex CA-500 series
Page 91

8.3 Execute Quality Control

Quality Control
To maintain the reliability of analyzed data, quality control has to be per­formed. When QC sample (control plasma, pooled plasma, etc.) is regis­tered with ID No. (QC01 - QC06) and then analyzed, the analysis data is kept in the QC File. It is by processing this analysis data with the QC Program that instrument and reagent stability that varies from time to time is monitored.
1. Set quality control samples on the sample rack.
2. Register the QC Sample Nos.
Press [ID No. Entry] key in the Root Menu screen, and using the numeric keys, enter ID Nos. for quality control ([QC] [0] [1] - [QC] [0] [6]).
3. Register the analysis parameters.
Specify samples in the Root Menu screen using [] and [] keys and analysis parameter keys. Mark “O” for the parameters to be analyzed.
4. Press [Start] key at the right upper corner in the screen.
Analysis data is automatically kept in the QC File.

8.4 Display QC Charts

Ready
Sysmex
Delete
100.0
3.3
3.3
Replace Rack? YES!
%
%
60
Settings
Print
Quality Control File No. PT QC01 Mean
 SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
Change
Scale
Select
File
HC IP
Lot. No TR-463 EXP. 12/31/2001
Cursor
100.0% 12/01/2001 15:09
Select
Test
Main
Menu
110 107 100 93 90
Note
Analysis data with “Slight Coagulation” and "Analysis Time Over" error will be also stored. Other error data cannot be stored.
5. Confirm the QC Chart. The instrument executes QC Program and confirms the QC Chart.
QC data is displayed in the Quality Control screen which is the initial screen of QC Program. When you press [QC] key on the Root Menu screen, the Quality Control screen will appear.
The QC Chart shows the latest 60-point data. By pressing [Change Scale] key, you can select the latest 60-point data or the stored 180-point data.
Note
[Change Scale] key will appear only when there is data in a file.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 8-5
Page 92
Quality Control
Ready
Sysmex
Quality Control File No. PT QC01 Mean
 SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
Replace Rack? YES!
%
100.0 
3.3 %
3.3
180
Settings
PrintDelete
HC IP
Lot No TR-463 EXP. 12/31/2001
Cursor
100.0% 12/01/2001 15:09
Change
Scale
File
Select
Test
Select
110 107 100 93 90
Main
Menu
The Quality Control screen displays the following QC data:
File No.: Parameter and File No. in the QC Chart being displayed Lot No.: Lot No. of the QC sample in use for Quality Control Exp.: Expiry date of the QC sample in use for Quality Control Mean: Mean of QC data SD: Standard deviation of QC data CV: Reproducibility coefficient of variation of QC data N: No. of QC data. When analysis is made twice, its mean
value is taken as the data, so the number of data is 1.
Stop_UL: When result of QC analysis exceeds this limit, subsequent
analyses are canceled.
Flag_UL: When result of QC analysis exceeds this limit, it is QC
error. TARGET: A target QC value Flag_LL: When result of QC analysis goes below this limit, it is QC
error. Stop-LL: When result of QC analysis goes below this limit, subse-
quent analyses are canceled.

Select QC Chart

Note
The point mark “ ” indicates data where the results of QC analysis exceed STOP-UL or are below STOP-LL.
Cursor: The position of control data being displayed. You can
move it to right or left by using [] and [] keys. When
there is no data in the file, neither [] key nor [] key
will be displayed. Cursor position data: Analysis data (calculated parameter) and analy-
sis date of the QC data in the cursor position
The instrument stores 6 different QC Files for each parameter (max 14 parameters).
To delete, set, or print a QC File, the QC Chart of the desired QC File has to be displayed.
1. Press [QC] key on the Root Menu screen.
The Quality Control screen will appear.
2. Press [Select Test] key on the Quality Control screen.
The Local Menu for Select Test will appear.
8-6 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 93
Quality Control
Ready
Sysmex
Quality Control
PT
II
Plg
Quality Control File No.PT
Replace Rack? YES!
APTT Fbg TT
V VIII AT3
Ready
Sysmex
Replace Rack? YES!
QC01 12/01
QC04 12/01
QC01 
QC02 12/01
QC05 QC06
HC IP
HC IP
IX
DDPl
Cancel
QC03 12/01
Cancel
3. Press the parameter key on the Local Menu. The Select File screen for the selected parameter will appear. To cancel the change of the parameter, press [Cancel] key. With the
parameter remaining unchanged, the Quality Control screen will return.
Note
When the current parameter is one which has the QC Chart displayed, press [Select File] key on the Quality Control screen.
4. Press the key of a file on Select File Screen. The QC Chart of the selected file will be displayed on the Quality
Control screen. To cancel the change of the file, press [Cancel] key. With the param-
eter remaining unchanged, the Quality Control screen will return.
Note
The key for each file will display the date of the latest analy­sis. Those files with no analysis date do not have any data.

8.5 Delete QC File

To start Quality Control anew, the existing data in the QC Files has to be deleted. Specifically, to change the type or lot of control plasma, it is necessary to delete the existing data in the files.
With this program, data can be deleted for each analysis parameter of an X
and L-J control file (QC01 - QC06).
Note
[Delete] key is not displayed when the file has no data.
1. Display the QC Chart of the desired file to be deleted. For how to display the QC Chart, refer to “Select QC Chart” of
“8.4 Display QC Charts”.
2. Press [Delete] key on the Quality Control screen. The Local Menu for deletion will appear.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 8-7
Page 94
Quality Control
Ready
Sysmex
Delete QC File
Sysmex
100.0
3.3
3.3
Replace Rack? YES!
%
%
50
Ready Replace Rack? YES!
File No. : PT QCO1 All data will be deleted.
Quality Control File No. PT QC01 Mean SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
Quality Control
Change Cursor
HC IP
Lot No TR-463 EXP. 12/31/2001
Cursor
100.0% 12/21/2001 15:09
HC IP

8.6 Delete QC Data

3. Press [Delete QC File] key on the Local Menu.
4. Press [Yes] key on the Local Menu.
110 107 100 93 90
CancelYes
The screen for confirming deletion of all data in a specified file will appear.
5. Press [Set] key or [Cancel] key. [Set]: Deletes all data in a specified file and returns to Quality
Control Screen.
[Cancel]: Cancels the deletion and returns to Quality Control
Screen.
CancelSet
Caution
Once data has been deleted, it cannot be recovered.
Ready
Sysmex
Delete QC File
Sysmex
Delete QC File
100.0
3.3
3.3
100.0
3.3
3.3
Replace Rack? YES!
%
%
50
Yes
Ready Replace Rack? YES!
%
%
50
Quality Control File No. PT QC01 Mean SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
Quality Control File No. PT QC01 Mean SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
HC IP
Lot No TR-463 EXP. 12/31/2001
Cursor
100.0% 12/01/2001 15:09
Change Cursor
HC IP
Lot No TR-463 EXP. 12/31/2001
Cursor
100.0% 10/10/2001 15:09
Change Cursor
110 107 100 93 90
Cancel
110 107 100 93 90
When this process is selected, two cursors will be displayed, and the data between those cursors can be deleted from the QC Chart.
1. Display the QC Chart from which you want to delete QC data. For how to display the QC Chart, refer to “Select QC Chart” of
“8.4 Display QC Charts”.
2. Press [Delete] key on the Quality Control screen. The Local Menu for deletion will appear and the cursor for setting
the range to be deleted will appear on the latest data.
3. Press [] key on the Local Menu. The cursor will move, forming two cursor lines, which are used to
highlight the data range to be deleted.
Yes
Cancel
8-8 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 95
Quality Control
Ready
Sysmex
Delete QC File
Sysmex
100.0
3.3
3.3
Replace Rack? YES!
%
%
50
Yes
Ready Replace Rack? YES!
File No. : PT QC01 10/10-11/15 will be deleted.
Quality Control File No. PT QC01 Mean SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
Quality Control
Lot No TR-463 EXP. 12/31/2001
Cursor
100.0% 11/15/2001 15:09
Change Cursor

8.7 Print QC data

HC IP
HC IP
110 107 100 93 90
Cancel
CancelEnter
4. Using [] and [] keys on the Local Menu, move the cursors and
set the data range to be deleted. Press the [Change Cursor] key to select the range-setting cursor. The cursor position data area displays the data for the range-setting
cursor.
5. Press [Yes] key on the Local Menu.
The screen confirming the range to be deleted will be displayed.
6. Press [Set] key or [Cancel] key on the Deletion Confirmation screen.
[Set]: Executes the deletion and quits the program. [Cancel]: Cancels the deletion and returns to the Quality Control
screen.
The QC data can be printed by the built-in printer by the following pro­cedure:
Ready
Sysmex
Graph
Sysmex
100.0
3.3
3.3
Replace Rack? YES!
%
%
50
Ready Replace Rack? YES!
Output in progress
Quality Control File No. PT QC01 Mean SD CV N Stop-UL Flag-UL  Flag-LL Stop-LL
Quality Control
HC IP
Lot No TR-463 EXP. 12/31/2001
Cursor
100.0% 12/21/2001 15:09
CancelList
HC IP
110 107 100 93 90
1. Display the QC Chart you want to print. For how to display the QC Chart, refer to “Select QC Chart” of
“8.4 Display QC Charts”.
2. Press [Print] key on the Quality Control screen. The Local Menu for print setting will be displayed.
3. Press [Graph] key or [List] key.
[Graph] key: Prints out QC data and QC chart. [List] key: Prints out QC data and data lists. [Cancel] key: Returns to the Quality Control screen without
printout.
When [Cancel] key is presse d on the “Output in progress” message screen, printing will be canceled and the Quality Control screen will return.
Cancel
Revised January 2003 - 2.0_en
Sysmex CA-500 series 8-9
Page 96
Quality Control
Note
The number of data that can be printed is all data for either
180 or 60 points, selected by [Change Scale] key.
A mark at the right edge of the printed list indicates that
quality control analysis results for that data exceeded STOP-UL or fell below STOP-LL.
Graphic printout
8-10 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Page 97
List printout
Quality Control
Revised January 2003 - 2.0_en
Sysmex CA-500 series 8-11
Page 98
9. Setting Standard Curve ...........................................9-1
9.1 Display Standard Curve ..............................................................9-1
9.2 Standard Curve Analysis .............................................................9-3
9.3 INR Manual Dilution Analysis ......................................................9-6
9.4 Manual Entry ...................... ... ......................................................9-8
9.5 Set Reagent Information .............................................................9-9
9.6 Set Calculation Parameters ......................................................9-10
9.7 Print Standard Curve .................................................................9-13
Revised January 2003 - 2.0_en
Sysmex CA-500 series
Page 99

9. Setting Standard Curve

The Standard Curve is a set of parameters used to determine each calcu­lation parameter, based on the coagulation time and OD/min in the analysis results.

9.1 Display Standard Curve

Description is given here on standard curve data and how to make its graphic display.

Data Display

Setting Standard Curve
Ready
Sysmex
Replace Rack? YES!
sec
11.4
17.4
27.9
52.6
11.4
1.73
0.0
0.0
(M)
PT OVB
Graph
Cal Date 12/01/2001
Lot No. EXP. Apc.63
Ref.
Standard
Analysis
Print
Standard Curve PT PT%
%
100.0
50.0
25.0
12.5
6.3
3.1
Normal
ISI
Next
Select
Test
HC IP
Manual
Entry
Lot No.
Entry
12/31/2001 12/31/2001
12/31/2001
Select
Param.
Return
1. Press [Standard Curve] key on the Root Menu screen. The PT Standard Curve Data screen will appear. The contents displayed on the Standard Curve Data screen are:
(M): Displayed when some of standard curve data is
manually input. Cal Date: The date on which the Standard Curve is set Reagent Name: Name of reagent Lot No.: Lot No. of reagent (up to 12 digits) EXP.: Expiry date of a reagent
<Standard Curve Data>
Activity/Concentration:
Percent activity or concentration Up to 6 points can be set for each parameter. (A lower-position point indicates a lower activity or concentra-
tion.) Displayed when activity or concentration is set in [Select Param.] key.
Coagulation Time - Reaction Speed:
Coagulation time or reaction speed for 6-point activity or concen­tration
Normal:
Normal value used for finding the ratio. Displayed when the ratio or INR is set by [Select Param.] key.
ISI:
International Sensitivity Index to calculate INR. Displayed when INR is set by [Select Param.] key.
Revised January 2003 - 2.0_en
Sysmex CA-500 series 9-1
Page 100
Setting Standard Curve
Caution
When analysis is made after expiry date has passed, coagulation activity (PT ratio, PT-INR, etc.) cannot be accurately calculated. In processing [Lot No. Entry], set a new reagent Lot No. and expiry date. When [Start] key is pressed after expiry date passed, the message “Check Reagent Expiry” appears, with the alarm sounding.
Note
[Next] key is displayed when dFbg or PT-INR Calibration is set by [Select Param.] key of standard curve for PT parame­ter.
Ready
Sysmex
Standard Curve
PT
VIII
Replace Rack? YES!
APTT Fbg TT
AT3 BCPC Hep

Graphic Display

Ready
Sysmex
Standard Curve PT
Expression Correlation rate offset
Replace Rack? YES!
Log P Log T=a∗Log P+b r =0.999 a =8.663e+00 b = 1.993e+02
HC IP
HC IP
II
DDPl
Cancel
Log T
Return
2. Press [Select Test] key on the Standard Curve Data screen.
Note
The key for analysis parameters which have been linked to the standard curve is masked.
3. Press the analysis parameter key to display the standard curve data of the selected analysis parameter.
When [Cancel] key is pressed, the screen will return to the original data's standard curve data screen.
1. Press [Graph] key on the Standard Curve Data screen. The contents displayed on the Graphic Display screen are: Standard Curve Graph: Plots activity or concentration on X axis and
coagulation time or reaction rate on Y axis.
Expression: Displays a expression such as
Log P = a * Log T + b from the plotted data. P: Coagulation activity (Concentration) T: Coagulation time (Reaction speed) Correlation: Correlation coefficient of approximate expression a, b: Constant derived from analysis results
9-2 Sysmex CA-500 series
Revised January 2003 - 2.0_en
Loading...