Stryker TP600, TP700 User manual

T/PUMP®
Series Localized Temperature
Therapy System
TP600 & TP700
Service Manual
2014/06 E.3 101176 REV E www.stryker.com

SYMBOLS

Symbols and Definitions

General Warning
Warm Water Fill Line
Cold Water Fill Line
Attention, consult Operator’s Manual
Water Levels
Water Flow
Dangerous Voltage
Type BF Applied Part
Do not penetrate with sharp object
Medical Electrical Equipment with Respect to Electrical Shock, Fire, and Mechanical hazards in ac­cordance with UL 60601-1, and CAN/CSA C22.2 No. 601.1, ASTM F 2196-2002
Model or Catalog Number
Manufacturer
Table of Contents
Symbols and Definitions ....................................................................3
Symbols ............................................................................3
Safety Precautions ........................................................................5
Introduction .............................................................................6
Before You Begin......................................................................6
Receiving Inspection ...................................................................6
To Return Pumps To Factory For Repair Or Exchange...........................................6
Introduction ..........................................................................7
Connecting the Pads ...................................................................7
Catalog Descriptions ...................................................................8
Features................................................................................9
Pump Features .......................................................................9
Keypad Features TP600 Series ..........................................................10
Keypad Features TP700 Series ..........................................................11
Keypad Features TP702 Series ..........................................................12
Specifications...........................................................................13
Operating Instructions.....................................................................14
Clik-Tite Connectors...................................................................14
Quick Disconnects....................................................................15
Start-Up Procedure ...................................................................16
Shutdown Procedure ..................................................................16
Cleaning...............................................................................17
Storage / Cleaning....................................................................17
Institutional (Hospital) Cleaning Instructions .................................................18
Pads / Accessories ...................................................................18
Water Temperature Control .................................................................19
Theory of Operation ......................................................................20
Functional Check ........................................................................21
Disassembly / Reassembly .................................................................26
Troubleshooting .........................................................................28
Troubleshooting .........................................................................31
Parts Illustration .........................................................................32
Parts Illustration .........................................................................33
Replacement Parts List....................................................................34
Warranty ..............................................................................36
Warranties..........................................................................36
Pump .............................................................................36
Pad, Single Patient Use ................................................................36
Pad, Reusable .......................................................................36
Parts ..............................................................................36
4 101176 REV E www.stryker.com

Safety Precautions

WARNING
• Risk of explosion. Do not use in the presence of flammable anesthetics.
• Risk of electric shock. Disconnect power before servicing the T/Pump.
• Do not attach the power cord to any moving parts of the bed frame or any location that may be a pinch point or
cause power cord damage. If routed improperly, the power cord may become pinched and could cause electrical shock.
• This device pumps temperature controlled water through a pad. Set the pad temperature only as directed by and
under the guidance of the prescriber.
• Check the skin integrity of the body surface to which therapy is applied. Evaluate patient response to temperature
application.
• Check patient’s skin for adverse reactions every 30 minutes or as directed by the prescriber. Patients at greatest
risk are those who are unconscious, diabetic, and/or incapacitated.
• Failure to adhere to these warnings could result in patient injury.
• The following Groups/Conditions require additional surveillance:
Group/Condition at risk Potential Injury
Patients with impaired circulation Ischemia
Areas of application under pressure Ischemia
• Only qualified medical service personnel should repair the T/Pump. Improper repair may result in death or serious
injury, equipment damage, or malfunction.
• UseT/PumpTP600&TP700seriescontrolswithMul•T•Pads®. For catalog numbers and descriptions, see page
8.
• Do not place additional heat sources between the patient and pad. Skin damage may result.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to the patient or operator. Modifying the product aslo voids the product warranty.
• Do not use accessories other than those specified by Stryker. Using competitor accessories may result in
unpredictable operation resulting in injury to the patient or operator or damage to the product.
• To avoid the risk of shock, unplug the unit before filling the reservoir.
CAUTION
• U.S. Federal law restricts this device to sale by or on the order of a physician.
• Set fluid temperature and monitor patient’s temperature and skin condition as directed by a physician to prevent
injury.
• Additional surveillance required for pediatric patients and patients with impaired circulation.
• Closely monitor areas of application under pressure as well as with use with topical solutions and other heat
sources to prevent injury.
• The use of quaternary disinfectant products may cause damage to the product.
• Frequent or prolonged exposure to higher concentration disinfectant solutions may result in damage to the product.
• Do not cover the control unit with blankets, pillows or other insulating materials. Air flow is required to maintain
system performance.
www.stryker.com 101176 REV E 5

Introduction

BEFORE YOU BEGIN

Read and understand this T/PUMP OPERATOR’S MANUAL and all SAFETY PRECAUTIONS (see page 5) prior to using the T/Pump.
Only qualified medical service personnel should repair or perform function tests on the T/Pump. Contact your dealer or Stryker’s Technical Service Department for assistance:
Telephone: 1-800-327-0770

RECEIVING INSPECTION

Check the shipping carton for damage immediately upon receipt. If package damage is discovered, the device should be unpacked with the carrier’s agent present. Any claims for shortage or damage must be filed with the delivering carrier by the purchaser. Do not return pumps that were damaged in shipment to Stryker without contacting our Technical Service Department for advice (see phone numbers below). If damaged goods are returned to Stryker without notifying the carrier, Stryker will assume the repairs will be made at the customer’s expense.

TO RETURN PUMPS TO FACTORY FOR REPAIR OR EXCHANGE

Merchandise returned to Stryker must be accompanied by a Return Goods Number (RG#), issued by Stryker, authorizing goods to be returned. Contact Customer Service at:
Customer Service: 1-800-327-0770
6 101176 REV E www.stryker.com

Introduction

INTRODUCTION
WARNING
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to the patient or operator. Modifying the product aslo voids the product warranty.
• Do not use accessories other than those specified by Stryker. Using competitor accessories may result in
unpredictable operation resulting in injury to the patient or operator or damage to the product.
CAUTION
• U.S. Federal law restricts this device to sale by or on the order of a physician.
• Do not cover the control unit with blankets, pillows or other insulating materials. Air flow is required to maintain
system performance.
Heat therapy is effective in the dilation of blood vessels, thereby increasing the blood flow to the heated area. Heat therapy has a variety of uses, the most common being treatment of aches and pains in joints and muscles.
Cooling therapy assists in vasoconstriction, decreasing blood flow and decreasing the metabolism in the affected area. Cooling therapy is applied in the acute phase of injury minimizing blood loss, inflammation of the tissue, and can be effective in pain management.
The Stryker T/Pump® Localized Temperature Therapy System provides therapy by warming or cooling the enclosed
water,andcirculatingitthroughtheStrykerMul•T•Pad.ThepadisconnectedtotheStrykerT/Pumpwith
easy-to-use Clik-Tite® or quick disconnects.
TheMul•T•Pad® provides the interface for delivering the temperature therapy. The unique button design allows water
to flow and provides trouble-free operation when the pad is folded to form a customized fit. This reduces the number of pads your facility must keep in inventory. The pads are applied to the part of the body requiring therapy, and the circulating water maintains the pad at the setpoint temperature. The setpoint temperature (TP700 series only) can be locked to prevent tampering.

CONNECTING THE PADS

TheMul•T•Pads® can be interconnected using Clik-Tite® connectors to provide therapy to more than one body site at
a time.
Figure 1A: Localized Temperature Therapy System with Single Pad with the TP650 and TP700.
www.stryker.com 101176 REV E 7
Figure 1B: Localized Temperature Therapy System with Multiple Pads.
Introduction

CATALOG DESCRIPTIONS

Catalog # Product Name
TP650 Classic Control Unit
TP700/TP702 Professional Control Unit
TP22B* Mul•T•Pad®: 15"w x 22"l (38cm x 56cm)
All Polymer. 10 per carton
TP22C Mul•T•Pad®: 15"w x 22"l (38cm x 56cm)
Nonwoven fabric on one side, pliable polymer on the other side. 1 per carton
TP3E* Mul•T•Pad®: 3"w x 23"l (8cm x 58cm)
Nonwoven fabric on one side, pliable polymer on the other side. 10 per carton
TP12E Mul•T•Pad®: 13"w x 18"l (33cm x 46cm)
Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton
TP22E Mul•T•Pad®: 15"w x 22"l (38cm x 56cm)
Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton
TP26E Mul•T•Pad®: 18"w x 26"l (46cm x 66cm)
Nonwoven fabric on one side, pliable polymer on the other side. 10 per carton
TP22G Mul•T•Pad®: 15"w x 22"l (38cm x 56cm)
Heavy Polymer, Reuseable. 10 per carton
TP650C Classic Control Unit with quick disconnects
TP700C Professional Control Unit with quick disconnects
TP612E Mul•T•Pad®: 13"w x 18"l (33cm x 46cm)
Nonwoven fabric on each side with quick disconnects. 10 per carton
TP622E Mul•T•Pad®: 15"w x 22"l (38cm x 56cm)
Nonwoven fabric on each side with quick disconnects. 10 per carton
TP626E Mul•T•Pad®: 18"w x 26"l (33cm x 46cm)
Nonwoven fabric on each side with quick disconnect. 10 per carton
*Non-CE Products
To order any of these products contact your dealer or Stryker’s Customer Service Department at:
Telephone: 1-800-327-0770
Or, visit our website at www.stryker.com
8 101176 REV E www.stryker.com

Features

PUMP FEATURES

Easy to Use Keypad See KEYPAD FEATURES on page 10 (TP600 series) and page 11 (TP700 series).
Attached hose 10 ft (305 cm) dual hose. Connectors allow pads to be connected to the pump (Figures 1A
and 1B, page 7).
Flow indicator Indicates no flow. Turns off heater if pump is tipped.
Warm/Cool Delivery Three setpoints on the TP600 series.
Four temperature setpoints on the TP700 series.
Therapy Cycles (TP700
Choose from 20-minute, 30-minute, or Continuous cycles.
series only)
On/Standby Button Indicates power is supplied to the unit.
Over Temp Safety
Limit thermostat shuts off heater if the high temperature limit is exceeded.
Thermostat
Self Check Automatic system check at startup.
Hose/cord Management Convenient and easy storage areas for hose and cord.
Comfortable
Designed for a more comfortable grip when moving the pump.
Handle Design
Dual Micro Processor Two electronic circuits, one over temperature sensing circuit.
Tethered Easy-Open
Prevents misplacing the cap. Only 1/4 of a turn is needed to remove or secure the cap.
Cap
Handle Vents The vents in the handle allow air flow to keep the motor and heater inside the unit cool.
Easy to Use Keypad
On/Standby Button
Comfortable Handle
Handle Vents
Flow Indicator
Tethered Easy-Open Cap
Cord storage
Temperature
Setpoints for warm
or cool delivery
Therapy Time Cycles
(TP700 series only)
Hose Connections
Hose storage
Figure 2: T/Pump Features
www.stryker.com 101176 REV E 9

KEYPAD FEATURES TP600 SERIES

Indicator/ Warning Light
Features
Figure 3: TP600 Series Keypad
See the Troubleshooting section on page 28.
Water Flow
Check hoses and clamps for kinks or occlusions.
Water Level
Check water level.
Setpoints Press the button at the bottom of the setpoint indicator to
toggle through the three setpoints. High Heat is 107°F (42°C). Low Heat is 100°F (38°C). Cooling is 50°F (10°C).
On/Standby Button
Green indicates the unit is on. Yellow indicates power is supplied to the unit but the unit is not on.
10 101176 REV E www.stryker.com

KEYPAD FEATURES TP700 SERIES

Indicator/ Warning Light
Features
Figure 4: TP700 Series Keypad
See the Troubleshooting section on page 28.
Water Flow
Check hoses or clamps for kinks or occlusions.
Water Level
Check water level.
Setpoints Press the button at the bottom of the setpoint indicator
to toggle through the four setpoints. Temperatures are identified in °C and °F.
Setpoint Lock Prevents tampering.
Press and hold for 2 seconds to lock or unlock the setpoint.
Therapy Cycles
Continuous cycle, 30-minute cycle or 20-minute cycle.
On/Standby Button
www.stryker.com 101176 REV E 11
Green indicates the unit is on. Yellow indicates power is supplied to the unit but the unit is not on.

KEYPAD FEATURES TP702 SERIES

Indicator/ Warning Light
Features
Figure 5: TP702 Keypad
See the Troubleshooting section on page 28.
Water Flow
Check hoses or clamps for kinks or occlusions.
Water Level
Check water level.
Setpoints Press the button at the bottom of the setpoint indicator
to toggle through the four setpoints. Temperatures are identified in °C.
Setpoint Lock Prevents tampering.
Press and hold for 2 seconds to lock or unlock the setpoint.
Therapy Cycles
Continuous cycle, 30-minute cycle or 20-minute cycle.
On/Standby Button
12 101176 REV E www.stryker.com
Green indicates the unit is on. Yellow indicates power is supplied to the unit but the unit is not on.