Operating instructions/Consult instructions for use
General warning
Caution
Warning; electricity
Fuse rating
Do not drill
Hydraulic oil pressure
Pinch/crush hazard
Non-ionizing radiation
Catalogue number
Serial number
European medical device
CE mark
5900-009-002 Rev A.0EN
Authorized representative in the European Community
For US Patents see www.stryker.com/patents
Manufacturer
Page 4
Safe working load
~
≤2m ≥18m
87VL
Mass of equipment
Maximum patient weight
Adult patient biometrics
Alternating current
Direct current
Duty cycle of product
Unit provides terminal for connection of a potential equalization conductor. The potential
equalization conductor provides direct connection between the unit and potential equalization
busbar of the electrical installation.
Protective earth ground
IIPPXX44
Protection from liquid splash
Type B applied part
Medical Equipment Classified by Underwriters Laboratories Inc. With Respect to Electric Shock,
Fire, and Mechanical Hazards Only in Accordance with ANSI/AAMI ES60601-1:2005/(R)2012
and A1:2012 C1:2009/(R)2012 and A2:2010/(R)2012, CAN/CSA-C22.2 No. 60601-1:14, IEC
60601-2-52:2009/A1:2015, CAN/CSA-C22.2 No. 60601-2-52:11 with Amendment 1:2017.
In accordance with European Directive 2012/19/EU on Waste Electrical and Electronic
Equipment (WEEE) as amended, this symbol indicates that the product should be collected
separately for recycling. Do not dispose of as unsorted municipal waste. Contact local distributor
for disposal information. Ensure infected equipment is decontaminated prior to recycling.
Summary of safety precautions ................................................................................................................................4
Indications for use ...................................................................................................................................................5
Expected service life ...............................................................................................................................................5
Contact information .................................................................................................................................................8
Serial number location.............................................................................................................................................9
Maintenance menu ................................................................................................................................................. 20
Set clock .............................................................................................................................................................. 20
Test modes ..........................................................................................................................................................24
Service ....................................................................................................................................................................25
Protecting against electrostatic discharge (ESD) .....................................................................................................25
Main control board replacement .............................................................................................................................26
Footboard main cable replacement ........................................................................................................................28
Siderail replacement, head end..............................................................................................................................33
IV pole attachment ................................................................................................................................................34
Top level assembly .................................................................................................................................................36
Deck pan assembly, seat deck, patient right.......................................................................................................... 54
5900-009-002 Rev A.01EN
Page 6
Deck pan assembly, seat deck, patient left ............................................................................................................ 55
Deck assembly, head section ................................................................................................................................. 56
Deck pan assembly, head deck, patient right......................................................................................................... 65
Deck pan assembly, head deck, patient left ...........................................................................................................66
Lock assembly, head deck position........................................................................................................................ 67
Deck pan assembly, knee deck, patient right.........................................................................................................72
Deck pan assembly, knee deck, patient left ...........................................................................................................73
Deck pan assembly, foot deck, patient right...........................................................................................................81
Deck pan assembly, foot deck, patient left .............................................................................................................82
HILO frame assembly ............................................................................................................................................. 83
Rail assembly, head ...............................................................................................................................................93
Timing link assembly ............................................................................................................................................ 108
Rail arm assembly, power/communication........................................................................................................... 109
Rail arm assembly, nurse call/speaker................................................................................................................. 110
Rail arm assembly, no cable................................................................................................................................. 111
Latch lock assembly, left head/right foot - 5900200023....................................................................................... 113
Latch lock assembly, right head/left foot - 5900200024....................................................................................... 114
Power supply cord ................................................................................................................................................ 115
Accessory cord, type B 120V................................................................................................................................ 116
EMC information ................................................................................................................................................... 123
The words WWAARRNNIINNGG, CCAAUUTTIIOONN, and NNOOTTEE carry special meanings and should be carefully reviewed.
WWAARRNNIINNGG
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
CCAAUUTTIIOONN
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the product or other property. This includes special care necessary for the safe and effective
use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
NNoottee -- Provides special information to make maintenance easier or important instructions clearer.
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WWAARRNNIINNGG
• Always use this product for bariatric use only per the details on the product labeling and specifications from this manual.
• Do not use this product for psychiatric, pediatric, or home health care use.
• Do not use this product in an oxygen rich environment.
• Always use Stryker approved mattresses that have been tested for compatibility with the product frame to avoid the risk
of patient entrapment.
• Do not lower the litter too low or the litter may not be supported.
• The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
• Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be no closer than 12 inches (30 cm) to any part of MMVV33, including cables specified by the manufacturer.
• Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the products. If
such use is necessary, carefully observe stacked or adjacent equipment to make sure that they are operating properly.
CCAAUUTTIIOONN
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Always make sure that the product is clear of obstacles before you use motion functions.
• Always calibrate the product angles when you replace the main control board or any actuator or product damage may
occur.
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
• Always perform this procedure in a secure location with the floor jack secure to the base frame before you remove the
caster to prevent the risk of injury.
• Do not put your hands or feet below the caster when you remove or attach the caster to prevent the risk of injury.
• Always leave the bolts loose in the support holes, so the bolt head is out far enough to stop the lift mechanism slide
block.
EN45900-009-002 Rev A.0
Page 9
IInnttrroodduuccttiioonn
This manual assists you with the operation or maintenance of your Stryker product. Read this manual before operating or
maintaining this product. Set methods and procedures to educate and train your staff on the safe operation or maintenance
of this product.
CCAAUUTTIIOONN
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
NNoottee
• This manual is a permanent part of the product and should remain with the product even if the product is sold.
• Stryker continually seeks advancements in product design and quality. This manual contains the most current product
information available at the time of printing. There may be minor discrepancies between your product and this manual. If
you have any questions, contact Stryker Customer Service or Technical Support at 1-800-327-0770.
PPrroodduucctt ddeessccrriippttiioonn
The Stryker model 5900 MMVV33™™ bariatric hospital bed is an AC-powered adjustable hospital bed designed to be utilized with
a patient support surface matching the deck size of the bed frame. MMVV33 consists of a Fowler and knee Gatch to aid in the
adjustment of the surface contour of the bed, moveable and latchable siderails, and electronic controls located in the
footboard and siderails. The
positioning and provides visual alerts when those parameters set by the health care professional (HCP) are altered.
ii
BBeedd®® Awareness system allows users to set various bed parameters to monitor bed
IInnddiiccaattiioonnss ffoorr uussee
WWAARRNNIINNGG
• Always use this product for bariatric use only per the details on the product labeling and specifications from this manual.
• Do not use this product for psychiatric, pediatric, or home health care use.
• Do not use this product in an oxygen rich environment.
The MMVV33 bariatric bed is intended to provide a patient support surface for medical purposes and to provide a method of
transporting patients within a healthcare facility. It is intended to be used with bariatric, adult, non-behavioral health patients
with a BMI of 30 kg/m² or greater and weighing 250 lb (113.4 kg) or more.
CClliinniiccaall bbeenneeffiittss
Patient treatment, patient positioning, and diagnostic
EExxppeecctteedd sseerrvviiccee lliiffee
The 5900 MMVV33 bariatric bed has an eight year expected service life under normal use conditions and with appropriate
periodic maintenance.
The backup batteries have a one year expected service life under normal use conditions.
DDiissppoossaall//rreeccyyccllee
Always follow the current local recommendations and/or regulations governing environmental protection and the risks
associated with recycling or disposing of the equipment at the end of its useful life.
5900-009-002 Rev A.05EN
Page 10
CCoonnttrraaiinnddiiccaattiioonnss
None known.
SSppeecciiffiiccaattiioonnss
WWAARRNNIINNGG -- Always use Stryker approved mattresses that have been tested for compatibility with the product frame to
avoid the risk of patient entrapment.
Safe working load
NNoottee:: Safe working load indicates the sum of the
occupant, accessories, and mattress weight.
1102 lb
500 kg
Maximum patient weight
Product weight
Scale system capacity maximum1102 lb500 kg
± 4.4 lb (2 kg) for patients weighing 225 (102 kg)
to 882 lb (400 kg)
Scale system accuracy
± 6.6 lb (3 kg) for patients weighing 886 (402 kg)
to 1102 lb (500 kg)
92 in. x 38.3 in.233.7 cm x 97.2 cm
Overall length and width
Patient sleep surface
Siderails stowed
Siderails at low,
intermediate, and high
position
36 in. (91.4 cm) deck
position
42 in. (106.7 cm) deck
position
92 in. x 42 in.233.7 cm x 106.7 cm
92 in. x 48 in.233.7 cm x 121.9 cm
92 in. x 40.5 in.233.7 cm x 102.9 cm
92 in. x 46.5 in.233.7 cm x 118.1 cm
92 in. x 52.5 in.233.7 cm x 133.4 cm
80 in. x 34.5 in.
80 in. x 40.5 in.
1027 lb
902 lb
203.2 cm x 87.6 cm
203.2 cm x 102.9 cm
465.8 kg
409.1 kg
48 in. (121.9 cm) deck
position
Bed height to top of seat litter
Knee Gatch position0° to 20° ± 2°
Fowler position
Foot position
Trendelenburg and reverse Trendelenburg
Electrical requirements
NNoottee -- Class I Electrical Equipment: Protection against electrical
shock relies on connection to protective earth of an appropriately
rated hospital grade outlet.
Hospital grade GFCI auxiliary outlet
EN65900-009-002 Rev A.0
80 in. x 46.5 in.
12 in. to 28 in.30.5 cm to 71.1 cm
0° to 60° ± 2°
0° to 13° ± 2°
+12° to -12°
120 VAC, 60 Hz, 8A230 VAC, 50 Hz, 4A
120 VAC, 60 Hz, 6A
203.2 cm x 118.1 cm
230 VAC, 50 Hz, 3A
Page 11
Battery voltage
95 °F
(35 °C)
50 °F
(10 °C)
140 °F
(60 °C)
-40 °F
(-40 °C)
80%
20%
95%
10%
106 kPa
70 kPa
106 kPa
50 kPa
NNoottee -- Always replace with Stryker approved batteries.
12 VDC (x2) (Stryker part number: 5900280025)
Duty cycle
2 minutes ON, 18 minutes OFF
Application environments1, 2, 3, and 5 per IEC 60601-2-52
Maximum acoustic sound pressure
< 60 dBA
CCoommppaattiibbllee mmaattttrreessss
Bariatric non-powered
support surface
Length
WidthThickness
Bolster width,
patient right and
patient left
Bolster
thickness,
patient right and
patient left
(28850555001)
79.5
in.
201.9cm45 in.114.3cm7 in.17.8
cm
6 in.15.2
cm
7 in.17.8
NNoottee -- Minimum mattress firmness: ILD at 50% - minimum of 108 lbf
Stryker reserves the right to change specifications without notice.
Specifications listed are approximate and may vary slightly from product to product or by power supply fluctuations.
Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA
EN85900-009-002 Rev A.0
Page 13
NNoottee -- The user and/or the patient should report any serious product-related incident to both the manufacturer and the
A
Competent authority of the European Member State where the user and/or patient is established.
To view your operations or maintenance manual online, see https://techweb.stryker.com/.
Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support.
Include the serial number in all written communication.
SSeerriiaall nnuummbbeerr llooccaattiioonn
You can find the serial number (A) below the headboard at the head of the bed (Figure 2).
Remove the product from service before you perform the preventive maintenance inspection. Make sure that all items listed
during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by qualified personnel.
NNoottee -- Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
All welds
All fasteners are secure
Casters lock with brake pedal applied
Casters are secure and swivel
Casters are free of wax and debris
Center light LED and bumper LEDs flash when brakes are released
Steer function works
Siderails move, latch, and stow
Deck expands, retracts, and locks in all positions
CPR release operates
IV pole is intact and operates (option)
Foley bag hooks intact
No cracks or splits in headboard, footboard or siderail panels
No rips or cracks in mattress cover
All functions on head end siderails operate (includes LEDs)
All functions on footboard operate (includes LEDs)
No cracks or damage to control overlays
Scale system calibrated
Night light operates
Power cords and plug not frayed or damaged
No cables worn or pinched
No damage to nurse call connections
All electrical connections tight
All ground strap cables are secure to the frame
Ground impedance not more than 200 mΩ (milliohms)
Current leakage not more than 300 µA (microamps)
Apply grease to the litter grease points (see maintenance manual for locations)
Ground chains intact
Trendelenburg/reverse Trendelenburg angle accuracy is minimum +12° to -12°
Fowler angle accuracy is 0° - 60° ± 2°
Fowler holds position at 30° with weight
Siderail switches operate (iiBBeedd Awareness)
Center light LED and bumper LEDs operate (iiBBeedd Awareness)
Inspect footboard control labels for signs of degradation
EN105900-009-002 Rev A.0
Page 15
Inspect hi-lo actuators for oil leaks
Inspect footboard connector housing for cracks or damage
All motions function
Nurse call functions
Auxiliary outlets function (test ground fault interrupter)
Replace batteries (one year expected service life)
Set clock to local date and time
Product serial number:
Completed by:
Date:
5900-009-002 Rev A.011EN
Page 16
EErrrroorr ccooddeess
Error codes are detailed in the table below.
EErrrroorr ccooddeeCCaauussee
0.02Error on non volatile memory
(nvm)
0.03Watchdog switch in wrong position or broken
(comWatchdogDisabled)
0.04Software version mismatched or missing board
(softwareVersionMismatch)
0.05Brake switch malfunction
(brake)
0.06Obstruction switch malfunction
(obstruction)
0.07CPR switch/cable malfunction
(cprHandle)
0.08FSC_KPD board is not responding
(FSC_KPD_missing)
TTrroouubblleesshhoooottiinngg
1. Reboot product
2. Replace MCU
1. Toggle watchdog switch on MCU
1. Reboot product
2. Check for a missing board error, if missing take
corrective action on cable or board
1. Check brake cable and switch
1. Check each of the four obstruction switches and
cables
1. Check CPR switch, cable, and head actuator
1. Check cable
2. Check AVB board (electrical issue)
0.09FSC_GUI board is not responding
(FSC_GUI_missing)
0.1SRK PL board is not responding
(SRK_PL_missing)
0.11SRK PR board is not responding
(SRK_PR_missing)
0.12PSI board is not responding
(PSI_missing)
1. Check cable
2. Check FSC_KPD
•Software in bootloader mode
•Software incorrect
•Electrical issue
3. AVB board (electrical issue)
1. Check cables
2. Check junction boards
3. Check SRK board
4. Check SRC (software and electrical)
1. Check cables
2. Check junction boards
3. Check SRK board
4. Check SRC (software and electrical)
1. Check cables
2. Check junction boards
EN125900-009-002 Rev A.0
Page 17
EErrrroorr ccooddeeCCaauussee
TTrroouubblleesshhoooottiinngg
0.13ACB HILO foot board is not responding
(ACB_HILO_FT_missing)
0.14ACB HILO head board is not responding
(ACB_HILO_HD_missing)
0.15ACB foot board is not responding
(ACB_FT_missing)
0.16ACB head board is not responding
(ACB_HD_missing)
0.17ACB knee board is not responding
(ACB_KN_missing)
0.18SRC foot patient left board is not responding
(SRC_FT_PL_missing)
0.19SRC head patient left board is not responding
(SRC_HD_PL_missing)
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
0.2SRC foot patient right board is not responding
(SRC_FT_PR_missing)
0.21SRC head patient right board is not responding
(SRC_HD_PR_missing)
0.22DPM foot board is not responding
(DPM_FT_missing)
0.23DPM head board is not responding
(DPM_HD_missing)
0.24DPM MID board is not responding
(DPM_MID_missing)
0.25SMB foot board is not responding
(SMB_FOOT_missing)
0.26SMB head board is not responding
(SMB_HEAD_missing)
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
1. Check cables
2. Check junction boards
5900-009-002 Rev A.013EN
Page 18
EErrrroorr ccooddeeCCaauussee
TTrroouubblleesshhoooottiinngg
0.27AVB board is not responding
(AVB_missing)
0.29Real Time Clock is not responding
(RTC chip fail)
0.3CPR handle switch stuck
(StuckCPRhandle)
1.0524V is out of range
(24V)
1.08Battery not connected to the system or completely
depleted
(batteryMissing)
1.09Damaged battery
(batteryShort)
1. Check cables
2. Check junction boards
1. Reboot product
2. Replace MCU, if 0.02 is present, also issue with
PMB charger chip
3. Replace PMB
1. Check CPR handle
2. Check pull cable
3. Check head actuator
1. Power cable issue
2. PMB ciruit error, replace PMB
1. Check cables
2. Replace battery
1. Replace batteries
1.1Unable to communicate with accelerometer
(Accelerometer error)
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
NNoottee -- Always ship back circuit boards to Stryker in the same antistatic bags that the new boards were originally shipped in.
The electronic circuits in the product are completely protected from static electricity damage when factory assembled.
Always use adequate static protection when servicing the electronic systems of the product. All service personnel must use
static protection whenever they are touching wires.
Sample antistatic protection equipment includes:
•1 antistatic wrist strap
•1 grounding plug
•1 test lead with a banana plug on one end and an alligator clip on the other end
Make sure that you follow the ESD manufacturer’s instructions for appropriate protection against static discharge.
BBaatttteerryy rreeppllaacceemmeenntt
TToooollss rreeqquuiirreedd::
•5/32” hex driver
•#2 Phillips screwdriver
•3/8” nut driver
PPrroocceedduurree::
1. Raise the product to the highest height position.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Push down on the brake pedal to apply the brake.
6. Using a 5/32” hex driver, remove the two screws and flat washers that secure the motion interrupt pan to the litter
weldment. Remove the motion interrupt pan.
7. Using a #2 Phillips screwdriver, remove the two screws from the power supply mounting bracket closest to the center of
the bed.
8. Using a #2 Phillips screwdriver, remove the two screws from the power supply mounting bracket closest to the outside
of the bed.
9. Hinge the batteries down.
10.Unplug the two battery supply cables from the main control board.
11.Unplug the USB cable from the main control board.
12.Using a 3/8” nut driver, remove the nut that secures the ground cable to the battery mounting plate.
13.Lift up to remove the battery mounting plate assembly from the litter.
14.Reverse steps to reinstall.
15.Verify proper operation of the product before you return it to service.
5900-009-002 Rev A.025EN
Page 30
CCaasstteerr rreeppllaacceemmeenntt
CCAAUUTTIIOONN
• Always perform this procedure in a secure location with the floor jack secure to the base frame before you remove the
caster to prevent the risk of injury.
• Do not put your hands or feet below the caster when you remove or attach the caster to prevent the risk of injury.
TToooollss rreeqquuiirreedd::
•#2 Phillips screwdriver
•Small flat head screwdriver
•5 mm hex wrench
•Small floor jack
PPrroocceedduurree::
1. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
2. Highlight SShhuuttddoowwnn and press EEnntteerr.
3. Unplug the power cord from the wall outlet.
4. Using a small flat head screwdriver, remove and save the four base cover screw caps.
5. Using a #2 Phillips screwdriver, remove and save the four screws that secure the base cover to the base frame.
Remove and save the base cover.
NNoottee -- Set the brakes to the neutral position.
6. Using a small floor jack, raise the base frame high enough to remove the caster.
7. Using a 5 mm hex wrench, remove and save the caster locating screw. Remove and discard the caster.
8. Reverse steps to reinstall.
NNoottee -- The caster stem has a red and green half. Face the red half of the caster stem toward the foot end of the bed.
9. Verify proper operation of the product before you return it to service.
1. Raise the product to the highest height position.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Push down on the brake pedal to apply the brake.
6. Using a 5/32” hex driver, remove the two screws and flat washers that secure the motion interrupt pan to the litter
weldment. Remove the motion interrupt pan.
7. Using a #2 Phillips screwdriver, remove the two screws from the power supply mounting bracket closest to the center of
the bed.
EN265900-009-002 Rev A.0
Page 31
8. Using a #2 Phillips screwdriver, remove the two screws from the power supply mounting bracket closest to the outside
of the bed.
9. Hinge the batteries down.
CCAAUUTTIIOONN
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
10.Unplug the two battery supply cables from the main control board.
11.Unplug the USB cable from the main control board.
12.Using a 3/8” nut driver, remove the nut that secures the ground cable to the battery mounting plate.
13.Lift up to remove the battery mounting plate assembly from the litter.
14.Unplug all cables from the main control board. Make note of how the cables are labeled and plugged in.
15.Using a #2 Phillips screwdriver, remove the two screws that secure the main control board assembly to the litter.
16.Grasp the main control board mounting bracket where you removed the two screws and pull down.
17.Reverse steps to reinstall.
18.Verify proper operation of the product before you return it to service.
AAVVBB bbooaarrdd rreeppllaacceemmeenntt
This procedure applies to both the head end and foot end AVB board.
TToooollss rreeqquuiirreedd::
•#2 Phillips screwdriver
•ESD system
PPrroocceedduurree::
1. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
2. Highlight SShhuuttddoowwnn and press EEnntteerr.
3. Unplug the power cord from the wall outlet.
4. Push down on the brake pedal to apply the brake.
5. Using a #2 Phillips screwdriver, remove and save the three screws that secure the AVB board assembly to the litter
frame.
6. Remove the AVB board assembly from the litter frame.
CCAAUUTTIIOONN
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
7. Disconnect all cables from the AVB board assembly.
8. Using a #2 Phillips screwdriver, remove and save the two screws that secure the beeper to the AVB board assembly.
Remove and save the beeper.
9. Discard the AVB board assembly.
10.Reverse steps to reinstall.
11.Verify proper operation of the product before you return it to service.
1. Raise the product to the highest height position.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Push down on the brake pedal to apply the brake.
6. Raise both head end siderails to the up and locked position.
7. Pull the CPR release handle to lower the fowler down.
8. Using needle nose pliers, remove both rue ring retaining clips from the clevis pins that attach the actuator to the litter
weldment and fowler weldment.
9. While supporting the actuator with one hand, remove the head end clevis pin and then lower the actuator down.
NNoottee -- Apply a thin coat of Syn-Tech grease to the clevis pin when you reinstall.
10.Using a small flat screwdriver, unlock the gray CPR pull retainer from the red CPR pull to loosen the CPR pull retainer.
11.Using a small flat screwdriver, remove the gray CPR pull retainer from the red CPR pull. A white spacer between the
CPR pull and the CPR cable retaining bracket will fall out. Save the white spacer.
12.Remove the CPR cables from the CPR cable retaining bracket.
13.Using a small flat screwdriver, release the retaining ring from the power and sensor cable quick connects and
disconnect the cables from the actuator. When re-connecting the cables, fully seat and lock the cable quick connector
retaining ring.
14.While supporting the actuator with one hand, remove the foot end clevis pin and then lower the actuator down.
NNoottee -- Apply a thin coat of Syn-Tech grease to the clevis pin when you reinstall.
15.Reverse steps to reinstall.
16.Verify proper operation of the product before you return it to service.
1. Raise the product to the highest height position.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Push down on the brake pedal to apply the brake.
6. Raise both foot end siderails to the up and locked position.
7. Using a small flat screwdriver, release the power/sensor cable quick connect and disconnect the cable from the
actuator. When re-connecting the cables, fully seat and lock the cable quick connector retaining ring.
8. Using needle nose pliers, remove both rue ring retaining clips from the clevis pins that attach the actuator to the litter
weldment and fowler weldment.
9. While supporting the actuator with one hand, remove the head end clevis pin and foot end clevis pin and then lower the
actuator down.
NNoottee -- Apply a thin coat of Syn-Tech grease to the clevis pin when you reinstall.
10.Reverse steps to reinstall.
11.Verify proper operation of the product before you return it to service.
1. Raise the product to the highest height position.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Push down on the brake pedal to apply the brake.
6. Raise both foot end siderails to the up and locked position.
7. Using needle nose pliers, remove the rue ring retaining clip from the head end clevis pin that attaches the actuator to the
litter weldment.
8. Using needle nose pliers, remove the circle ring retaining ring from the foot end clevis pin that attaches the actuator to
the thigh weldment.
9. While supporting the actuator with one hand, remove the head end clevis pin and the foot end clevis pin and then lower
the actuator down.
NNoottee -- Apply a thin coat of Syn-Tech grease to the clevis pin when you reinstall.
10.Using a small flat screwdriver, release the power/sensor cable quick connect and disconnect the cable from the
actuator. When re-connecting the cables, fully seat and lock the cable quick connector retaining ring.
11.Reverse steps to reinstall.
NNoottee -- Route the power/sensor cable over the top of the actuator when you reinstall the actuator.
12.Verify proper operation of the product before you return it to service.
1. Run the actuator up or down to remove the litter load from the actuator.
2. Raise both siderails to the up and locked position at either the head end or foot end.
3. If the activator still runs:
a. Raise the product to the highest height position to expose the base weldment support holds on both sides.
CCAAUUTTIIOONN -- Always leave the bolts loose in the support holes, so the bolt head is out far enough to stop the lift
mechanism slide block.
b. Insert two 3/8” bolts through the support holes in the base weldment from the inside out and then thread the nuts
onto the bolts to keep the bolts from falling out.
WWAARRNNIINNGG -- Do not lower the litter too low or the litter may not be supported.
c. Lower the litter until the lift mechanism slides touch the 3/8” bolt heads. You should be able to wiggle the actuator by
hand after you remove the litter load from the litter surface. If not, lower the litter more until the actuator is loose.
4. If the activator does not run:
a. Put the bed into the Maintenance mode (limits bypassed) and then go to the Actuator Maintenance Menu. Select the
actuator that needs to be moved and see if it will run.
b. See step 2a.
5. If the actuator will not run with the limits bypassed in step 4a, use a hoist, with a strap around and under the litter, to
remove the litter load from the actuator and support the litter surface on the end that has failed.
6. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
7. Highlight SShhuuttddoowwnn and press EEnntteerr.
8. Unplug the power cord from the wall outlet.
9. Using the cable quick connector release, unlock the cable from the motor control box so that it is loose.
10.Using needle nose pliers, remove both rue ring retaining clips from the clevis pins that attach the actuator to the litter
weldment and lift weldment.
11.While supporting the actuator with one hand, remove the head end clevis pin and foot end clevis pin and then lower the
actuator down.
NNoottee -- Apply a thin coat of Syn-Tech grease to the clevis pin when you reinstall.
12.Reverse steps to reinstall.
13.Verify proper operation of the product before you return it to service.
1. Push down on the brake pedal to apply the brake.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Using a small flat head screwdriver, remove and save the four base cover screw caps.
6. Using a #2 Phillips screwdriver, remove and save the four screws that secure the base cover to the base frame.
Remove and save the base cover.
7. Using a #2 Phillips screwdriver, remove and save the two screws that secure the load cell board assembly to the base
weldment. Remove the load cell board assembly.
CCAAUUTTIIOONN
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
8. Disconnect the communication cable and the two load cells from the load cell board assembly. Discard the load cell
board assembly.
9. Reverse steps to reinstall.
10.Recalibrate the scale (
11.Verify proper operation of the product before you return it to service.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Using a small flat head screwdriver, remove and save the four base cover screw caps.
6. Using a #2 Phillips screwdriver, remove and save the four screws that secure the base cover to the base frame
weldment. Remove and save the base cover.
7. Using a small floor jack, near the load cell that you need to replace, push downward on the jack’s lever to raise the lift
support tube upward.
8. Using needle nose pliers. remove and save the rue clip and pin from the base frame weldment.
9. Using a ratchet and 1/2” socket, remove and save the two bolts that secure the load cell assembly to the base frame
weldment..
NNoottee -- When you reinstall the load cell assembly, orient the load cell assembly up, so you can secure the bolts by hand
to start.
10.Using needle nose pliers, remove the load cell assembly from the base frame weldment.
EN325900-009-002 Rev A.0
Page 37
11.Using a #2 Phillips screwdriver, remove and save the two screws that secure the inspection cover to the base frame
weldment. Remove and save the inspection cover.
12.Using a #2 Phillips screwdriver, remove and save the two screws that secure the board assembly to the base frame
weldment. Remove and save the board assembly.
13.Disconnect the load cell assembly and remove the cable from the base frame weldment. Discard the load cell assembly.
NNoottee -- Note the routing of the load cell cable for reinstallation.
14.Reverse steps to reinstall.
15.Recalibrate the scale (
16.Verify proper operation of the product before you return it to service.
1. Raise the product to the highest height position.
2. Raise the Fowler to the full upright position.
3. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
4. Highlight SShhuuttddoowwnn and press EEnntteerr.
5. Unplug the power cord from the wall outlet.
6. Push down on the brake pedal to apply the brake.
7. Expand all litter decks to the 48 in. position.
8. Raise the siderail to the up and locked position.
9. Using a T15 Torx driver, remove the two screws that secure the junction board assembly to the siderail carrier. Remove
the junction board assembly.
10.Unplug the two siderail cables from the junction board assembly.
11.Using a T27 Torx driver, remove the screw that secures the guide rod to the litter.
12.Push the siderail release button to lower the siderail to the lowest stowed position.
13.Using needle nose pliers, from under the litter surface, slide each glide rod out.
NNoottee -- Save the two rubber washers on each glide rod, one on both sides of the siderail carrier, to reinstall. Use SynTech grease to lube the glide rod, if required.
14.While supporting the rail with one hand, remove both glide rods and remove the siderail assembly.
15.Reverse steps to reinstall.
16.Verify proper operation of the product before you return it to service.
1. Raise the product to the highest height position.
2. On the footboard control panel, press EEnntteerr to access the SSeettttiinnggss menu.
3. Highlight SShhuuttddoowwnn and press EEnntteerr.
4. Unplug the power cord from the wall outlet.
5. Push down on the brake pedal to apply the brake.
6. Expand all litter decks to the 48 in. position.
7. Raise the siderail to the up and locked position.
8. Using a T15 Torx driver, remove the two screws that secure the junction board assembly to the siderail carrier. Remove
the junction board assembly.
9. Unplug the two siderail cables from the junction board assembly.
10.Using a T27 Torx driver, remove the screw that secures the guide rod to the litter.
11.Push the siderail release button to lower the siderail to the lowest stowed position.
12.Using needle nose pliers, from under the litter surface, slide each glide rod out.
NNoottee -- Save the two rubber washers on each glide rod, one on both sides of the siderail carrier, to reinstall. Use SynTech grease to lube the glide rod, if required.
13.While supporting the rail with one hand, remove both glide rods and remove the siderail assembly.
14.Reverse steps to reinstall.
15.Verify proper operation of the product before you return it to service.
IIVV ppoollee aattttaacchhmmeenntt
TToooollss rreeqquuiirreedd::
•3/8” ratchet
•3/16” hex driver
•1/2” socket
•9/16” socket
PPrroocceedduurree::
1. Using a 3/16” hex driver, remove and save the power cord retainer shoulder bolt and strap on both sides of the product.
2. Using a 1/2” socket, secure the bracket with the IV pole assembly to the left side of the product with the two supplied
bolts and washers.
3. Using a 1/2” socket, secure the bracket with the rest block to the right side of the product with the two supplied bolts and
washers.
4. Using a 3/16” hex driver, secure the power cord retainer shoulder bolt and strap on both sides of the product through
both brackets.
5. Verify proper operation of the product before you return it to service.
PPaattiieenntt hheellppeerr aattttaacchhmmeenntt
TToooollss rreeqquuiirreedd::
•3/8” ratchet
EN345900-009-002 Rev A.0
Page 39
•3/8” hex driver
•7/16” socket
PPrroocceedduurree::
1. Using a 3/8” hex driver, attach the patient helper mounting bracket with the four supplied screws.
NNoottee -- Make sure that the armrest pin faces upward.
2. On the floor, place the patient helper lifting pole, patient helper cap, and patient helper trapeze hanger together.
3. Using a 7/16” socket, secure the patient helper lifting pole, patient helper cap, and trapeze hanger together with the
supplied bolt.
4. With another person, lift and insert the patient helper assembly into the patient helper mounting bracket.
5. Verify proper operation of the product before you return it to service.
5900-009-002 Rev A.035EN
Page 40
TToopp lleevveell aasssseemmbbllyy
10
4
2X
5
2X
7
2X
TORQUE TO
150-180in/lbs
5
2X
4
2X
6
2X
14
2X
15
2X
TORQUE TO
150-180in/lbs
TORQUE TO
150-180in/lbs
11
16
2X
9
2X
17
3
2X
2
2X
NOTE ORIENTATION OF SLIDE BLOCKS
(COUNTERBORE HOLES FACE OUTWARD)
IItteemmNNuummbbeerrNNaammeeQQuuaannttiittyy
15900240042Deck weldment, head section1
25900200068Deck pan assembly, head deck, patient
1
right
35900200069Deck pan assembly, head deck, patient
left
1
45900600007Bushing, pinion shaft cover2
55900500072Pinion shaft, head deck1
65900560003Pinion gear, expandable deck4
75900200051Lock assembly, head deck position1
85900520103PCover pan, head deck electrical top
1
plated
95900520104PCover pan, head deck electrical bottom
1
plated
105900620007Cover, head deck1
115900370022Grommet, push-in locking 5/8" hole8
125900340105Screw, plastite pan head torx #8-16 x
2
1.25"lg zinc
5900-009-002 Rev A.063EN
Page 68
IItteemmNNuummbbeerrNNaammeeQQuuaannttiittyy
135900340101Screw, rd hd flg hex drive 1/4-unc x 3/8lg
1
zinc
145900310020Nutsert, #10-unf thick sheet hex body -
4
zinc
155900340114Screw, plastite pan head ph #10-14 x 1/
8
2" lg znpl
165900340017Screw, pan ph #8-unc x 0.31254
175900340090Screw, sems pan ph #10-32unf x 3/4lg
12
int wash zinc
185900600035Mattress keeper, short6
190025-120-000Rivet, 1/4 x 0.600" lg blind alum12
205900400051PCBA, jctn hd/ft pl-in/pr-out1
215900400052PCBA, jctn hd/ft pl-out/pr-in1
225900340097Screw, sems pan ph #6-32x1-1/2lg int
4
wash zinc pl
235900200159Handle assembly, deck expansion2
245900200160Handle assembly, CPR2
255900320021Washer, flat 1.375 x 0.75 x 0.125 acetal 2
265900340081Screw, flat Torx head 5/16-18 x 3/4 tri
105900600035Mattress keeper, short2
110025-120-000Rivet, 1/4 x 0.600" lg blind alum4
125900340090Screw, sems pan ph #10-32unf x 3/4lg
4
int wash zinc
135900200159Handle assembly, deck expansion2
145900370022Grommet, push-in locking 5/8" hole4
155900340017Screw, pan ph #8-unc x 0.31252
165900340114Screw, plastite pan head ph #10-14 x 1/
4
2" lg ZnPL
175900340081Screw, flat Torx head 5/16-18 x 3/4 tri
M LOBE ORIENTATION (IDENTIFICATION MARKS FACE OUTWARD)
FLEX CABLE TO BE INSTALLED WITH
RED LABEL INDICATOR FLUSH WITH
END OF CASTING. ( 1/4")
ROUTE ARM CABLES UNDER
POSTS & CONNECT TO PCBA
CASTING CUT-AWAY
FOR CLARITY
5900200130 Rev C (right) (Reference only)
5900200131 Rev C (left) (Reference only)
5900-009-002 Rev A.093EN
Page 98
21
2X
23
2X
5
2X
6
2X
32
6X
30
2X
4
CENTERLIN
E
OF
ARM
C
ENTERLINE OF ARM
INSTALL GEARS WITH ARROWS POINTING TO TOP SCREW
RELATIVE TO ARM CENTERLINE. TYPICAL BOTH ARMS.
START SCREWS IN THREADS - DO NOT TIGHTEN AT THIS TIME.
40
2X
24
4X
CONNECT RAIL POSITION HARNESS TO CORRESPONDING
SWITCHES, ORIENT AND INSTALL AS SHOWN.
CONNECT LOCK HARNESS
CONNECTOR DURING LATCH
ASSEMBLY INSTALLATION.
CONNECT TO 6-PIN
HEADER ON PCBA
EN945900-009-002 Rev A.0
Page 99
192X28
2X
18
2X
10
16
29
17
7
APPLY NS-18191-G
GREASE TO LENGTH
OF RACK
APPLY NS-18191-G
GREASE TO LENGTH
OF RACK
TIGHTEN RACK MESH SET SCREW TO ENSURE GEAR
MESH IS TIGHT. FUNCTIONALLY CYCLE TEST TO ENSURE
CORRECT SETTING FOR GRAVITY FALL AND "HITCH"
ADJUST BOTH LEVER SET SCREWS AT THE SAME TIME.
RACK MESH
ADJUSTING
SET SCREW
AFTER MESH IS TIGHT, TORQUE ALL GEAR
SCREWS TO 45IN/LB (TWICE AROUND)
12
9
8
36
27
2X
26
2X
35
2X
11
26
2X
37
22
16X
8.00 0.5
0.25
PLACE TAPE APPROXIMATELY AS SHOWN ON BACKPLATE
1137
5900-009-002 Rev A.095EN
Page 100
25
4X
39
WIRE HARNESS ROUTING
ROUTE ALL HARNESSES AS SHOWN.
ROUTE HARNESSES UNDER
PLATE TABS AS SHOWN.
LOOP RAIL ARM HARNESS AS SHOWN,
ORIENT P-CLIP HORIZONTALLY.
CONNECT TO PCBA.
LOOP RAIL ARM HARNESS AS SHOWN,
ORIENT P-CLIP HORIZONTALLY.
CONNECT TO PCBA.
57
57
CABLE TIE WIRES NEATLY
TO BACKPLATE
CABLE TIE WIRES NEATLY
TO BACKPLATE
EN965900-009-002 Rev A.0
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.