
ISOFLEX®
Pressure Management Mattress
Instructions for Use
Read and understand the operating instructions and all safety precautions prior to use.
Symbols
Consult
Instructions for
Use
Do not tumble dry
Foot end of
mattress
Latex free
Damp wipe only
10%
Chlorinated
Bleach
SAFETY PRECAUTIONS
Warning:
• ISOFLEX mattress covers (top and bottom) should be inspected for tears, punctures, excessive wear, and
misaligned zippers each time the covers are cleaned. If a mattress cover becomes compromised, the
mattress should be removed from service immediately to prevent cross contamination.
Caution:
• When utilizing an X-Ray plate while the patient is on ISOFLEX, take preca utions to insure that no sharp
edges from the X-Ray plate come in contact with the mattress cover. It is recommended that a pillow case,
or some other protective device, is placed on the plate prior to placement under the patient. If a mattress
cover becomes compromised, the mattress should be removed from service immediately to prevent cross
contamination.
• When cleaning the underside of the mattress, insure that no liquid is allowed to seep in to the zipper area
and watershed cover barrier, fluids allowed to come in contact with the zipper may leak in to the mattress.
• Check patient’s skin regularly. Consult physician if any redness or skin break occurs. Serious injury could
result if the patient’s skin condition is left untreated.
• The patient’s skin should be checked and assessed every 8 hours, or once per shift, to insure proper
mattress performance.
Do not iron
Do not dry clean
Allow to completely air dry
Intended Use
ISOFLEX mattresses provide pressure redistribution for patients with existing, or who are at risk of developi ng,
pressure ulcers.
Unpacking Instructions:
1. Remove the mattress from the box.
2. Lay the mattress on its side.
3. Using scissors, carefully cut the sleeve that holds the folded mattress. Be
careful not to inadvertently cut the mattress.
4. Remove and discard the sleeve.
5. Remove and discard the plastic bag and carton. Keep plastic away from children.

Set up Instructions:
1. Place the ISOFLEX mattress on the bed frame. Insure that the mattress properly fits the bed frame on which
it is being placed. Using an ISOFLEX mattress on a bed frame of a larger or smaller width than the stated
ISOFLEX width may result in the mattress sliding and patient injury.
2. Insure that the foot end symbol (
) is located at the foot end of the bed frame. ISOFLEX
mattresses are equipped with a dedicated sloped heel section and must be installed with this section at the
foot end of the bed frame.
3. Place the linens on the mattress leaving them as loose as possible to avoid creating a “hammocking” effect
and increasing interface pressures.
4. Place the patient on the mattress.
Cleaning Instructions:
1. Gaymar mattress surfaces are designed for long-lasting comfort and reliability. In order to maintain these
products, they should be cleaned regularly using soap and water and a hospital grade disinfectant. Gaymar
recommends these disinfecting agents: Ecolab
®
Airkem® A-33 Quaternary Solution, 3M® Phenolic Disinfectant
Cleaner Concentrate Solution Product #18, or a 10% Chlorinated Bleach Soluti on.
2. Alcohol is not recommended because it can discolor and degrade the cover.
3. Allow to completely dry. The solution contact time is what makes the disinfectant process effective.
4. Wipe down the mattress with a clean dry cloth to remove any excess disinfectant.
NOTE: The cover must be dry prior to storage or application of linens. Failure to remove excess dis infectant
could eventually degrade the cover.
5. An antimicrobial agent has been used to help prevent bacteria and fungus from destroying the coating of the
cover. Should unsightly stains, discoloration, brittleness, stickiness, or unpleasant odors become noticeable, the
antimicrobial agent may have become ineffective and the product should be replaced.
Distributed by:
Stryker Medical
3800 East Centre Avenue
Portage, MI 49002 USA
Copyright:
©2010. Stryker Medical, Inc. All rights reserved.
Trademark Information:
Stryker is a registered trademarks of Stryker Medical.
ISOFLEX is a trademark of Gaymar Industries, Inc.
Ecolab and Airkem are registered trademarks of Ecolab Inc.
3M is a registered trademarks of 3M Corporation.
This product is made with Intelli-Gel
Gelastic™* polymeric columned material.
*EdiZONE, LLC of Alpine, UT.
Patent Information:
This product is covered by one or more of the following US patents:
7666341, 7076822, 7060213, 6843873, 6797765, 6767621, 6447865,
6413458, 6099951, 6026527, 5749111 and other US and/or foreign patents
and pending patent applications.
®
* polymeric columned material and
P/N 101392000 Rev A 6/10