ISOFLEX® 
Pressure Management Mattress 
Instructions for Use 
Read and understand the operating instructions and all safety precautions prior to use. 
Symbols    
Consult 
Instructions for 
Use 
Do not tumble dry 
Foot end of 
mattress 
Latex free   
Damp wipe only  
10% 
Chlorinated 
Bleach 
SAFETY PRECAUTIONS 
Warning: 
•  ISOFLEX mattress covers (top and bottom) should be inspected for tears, punctures, excessive wear, and  
misaligned zippers each time the covers are cleaned. If a mattress cover becomes compromised, the 
mattress should be removed from service immediately to prevent cross contamination. 
Caution: 
•  When utilizing an X-Ray plate while the patient is on ISOFLEX, take preca utions to insure that no sharp  
edges from the X-Ray plate come in contact with the mattress cover. It is recommended that a pillow case, 
or some other protective device, is placed on the plate prior to placement under the patient. If a mattress 
cover becomes compromised, the mattress should be removed from service immediately to prevent cross 
contamination. 
•  When cleaning the underside of the mattress, insure that no liquid is allowed to seep in to the zipper area  
and watershed cover barrier, fluids allowed to come in contact with the zipper may leak in to the mattress. 
•  Check patient’s skin regularly. Consult physician if any redness or skin break occurs. Serious injury could  
result if the patient’s skin condition is left untreated. 
•  The patient’s skin should be checked and assessed every 8 hours, or once per shift, to insure proper  
mattress performance. 
Do not iron 
Do not dry clean 
Allow to completely air dry 
Intended Use 
ISOFLEX mattresses provide pressure redistribution for patients with existing, or who are at risk of developi ng, 
pressure ulcers. 
Unpacking Instructions: 
1. Remove the mattress from the box. 
2. Lay the mattress on its side. 
3. Using scissors, carefully cut the sleeve that holds the folded mattress. Be 
careful not to inadvertently cut the mattress. 
4. Remove and discard the sleeve. 
5. Remove and discard the plastic bag and carton. Keep plastic away from children. 
 

Set up Instructions: 
1. Place the ISOFLEX mattress on the bed frame. Insure that the mattress properly fits the bed frame on which 
it is being placed. Using an ISOFLEX mattress on a bed frame of a larger or smaller width than the stated 
ISOFLEX width may result in the mattress sliding and patient injury. 
2. Insure that the foot end symbol ( 
 ) is located at the foot end of the bed frame. ISOFLEX 
mattresses are equipped with a dedicated sloped heel section and must be installed with this section at the 
foot end of the bed frame. 
3. Place the linens on the mattress leaving them as loose as possible to avoid creating a “hammocking” effect 
and increasing interface pressures. 
4. Place the patient on the mattress. 
Cleaning Instructions: 
1. Gaymar mattress surfaces are designed for long-lasting comfort and reliability. In order to maintain these 
products, they should be cleaned regularly using soap and water and a hospital grade disinfectant. Gaymar 
recommends these disinfecting agents: Ecolab
®
 Airkem® A-33 Quaternary Solution, 3M® Phenolic Disinfectant 
Cleaner Concentrate Solution Product #18, or a 10% Chlorinated Bleach Soluti on.  
2. Alcohol is not recommended because it can discolor and degrade the cover. 
3. Allow to completely dry. The solution contact time is what makes the disinfectant process effective. 
4. Wipe down the mattress with a clean dry cloth to remove any excess disinfectant. 
NOTE: The cover must be dry prior to storage or application of linens. Failure to remove excess dis infectant 
could eventually degrade the cover. 
5. An antimicrobial agent has been used to help prevent bacteria and fungus from destroying the coating of the 
cover. Should unsightly stains, discoloration, brittleness, stickiness, or unpleasant odors become noticeable, the 
antimicrobial agent may have become ineffective and the product should be replaced.
Distributed by: 
Stryker Medical 
3800 East Centre Avenue 
Portage, MI 49002 USA 
Copyright: 
©2010. Stryker Medical, Inc. All rights reserved. 
Trademark Information: 
Stryker is a registered trademarks of Stryker Medical. 
ISOFLEX is a trademark of Gaymar Industries, Inc. 
Ecolab and Airkem are registered trademarks of Ecolab Inc. 
3M is a registered trademarks of 3M Corporation. 
This product is made with Intelli-Gel 
Gelastic™* polymeric columned material.  
*EdiZONE, LLC of Alpine, UT. 
Patent Information: 
This product is covered by one or more of the following US patents: 
7666341, 7076822, 7060213, 6843873, 6797765, 6767621, 6447865, 
6413458, 6099951, 6026527, 5749111 and other US and/or foreign patents 
and pending patent applications. 
®
* polymeric columned material and 
P/N 101392000 Rev A  6/10