Stryker HydroSet Step-by-step Manual

HydroSet Mixing Instructions
A Step-By-Step Guide
Indications for Use
HydroSet is a self-setting calcium phosphate cement indicated to fill bony voids or gaps in the skeletal system (i.e. extremities, craniofacial, spine and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bony. HydroSet is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure.
HydroSet cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K. wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support media and is
not intended to provide structural support during the healing process. HydroSet is a self-setting calcium phosphate cement intended for
use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects as well in the augmentation or restoration of bony contour in the craniofacial skeleton.
Note
For optimal performance store HydroSet between 18-22˚C for at least 1.5 hours prior to surgery . Please see the pack insert for detailed information.
Step 1
Firmly click the Cement Delivery
Syringe into
place in the inner
blister tray so it is at an angled
position with the finger loops
resting on the
shelves of the tray.
Step 2
Inject all the liquid from the glass syringe into the mixing bowl.
Step 3
Vigorously mix the powder and solution in a circular motion for 45 seconds, ensuring that all of
the solution has been distributed
throughout the powder.
Step 4
Transfer the HydroSet to the
syringe by
pouring it into the side
of the funnel. Use
spatula to assist.
Step 5
Remove pre-attached funnel by turning counter-clockwise.
Step 6
Attach the supplied cannula by turning clockwise. Please remove the plastic cannula cover.
Step 7
Screw the plunger rod into the plunger at the base of the syringe, while keeping the cannula elevated.
Step 8
Position the syringe so the cannula is pointing up and push the plunger rod to remove excess air.
STEP
TOTAL TIME
1
34
56
78
2
Mix 45 sec
Load Syringe 1 min 45 sec
Inject & Sculpt 2 min
Harden 4 min
45 sec
2 min 30 sec
*4 min 30 sec
8 min 30 sec
*Sculpting beyond 4 min
30 sec may inhibit the cement setting properties
The information presented in this brochure is intended to demonstrate a Stryker product. Always refer to the package insert, product label and/or user instructions before using any Stryker product. Surgeons must always rely on their own clinical judgment when deciding which products and techniques to use with their patients. Products may not be available in all markets. Pr oduct availability is subject to the regulatory or medical practices that govern individual markets. Please contact your Stryker representative if you have questions about the availability of Stryker products in your area.
This document is not for distribution in the USA Stryker Corporation or its subsidiary owns the registered trademark: Stryker
Stryker Corporation or its subsidiary owns, uses or has applied for the following trademarks: HydroSet™ Literature Number 90-01607
LOT A4908 Copyright © 2009 Stryker
Printed in France
Stryker Leibinger GmbH & Co. KG Bötzinger Straße 41 D-79111 Freiburg Germany
www.osteosynthesis.stryker.com