Stryker GOBED FL17E, GOBED 2500 Operation Manual

Product Number for the United States: 2500
TM
The GOBED
ELECTRIC ACUTE CARE BED
Product Number: FL17E
TECHNICAL ASSISTANCE AND PARTS
1 800 327- 0770 (US) 1 800 428- 5025 (Canada) E-Mail (Canada): service@bertec.strykercorp.com
Manufactured by Stryker Bertec Medical Inc
72-0171 R1.1.doc September 2002 F15-43-E-A Printed in Canada
TABLE OF CONTENTS
1. INTRODUCTION.................................................................................................................... 4
1.1 Technical Specifications................................................................................................. 4
1.2 Technical Support........................................................................................................... 4
1.3 Warning / Caution / Note Definitions............................................................................... 5
1.4 Safety Tips and Guidelines............................................................................................. 5
1.5 Warranty......................................................................................................................... 7
Limited Warranty............................................................................................................ 7
To Obtain Service and/or Parts....................................................................................... 7
Return Authorization....................................................................................................... 7
Damaged Merchandise .................................................................................................. 7
1.6 Bed Position Pictograms................................................................................................. 7
1.7 Set up Procedure............................................................................................................. 8
Checklist......................................................................................................................... 8
1.8 Bed Cleaning and Preventative Maintenance................................................................. 9
Bed Cleaning and Mattress Care.................................................................................... 9
Preventative Maintenance Program.............................................................................. 10
1.9 Bed Illustration.............................................................................................................. 11
2. OPERATION GUIDE............................................................................................................ 12
2.1 Switching On Power..................................................................................................... 12
2.2 Applying the Brake ....................................................................................................... 12
2.3 Moving the Bed ............................................................................................................ 12
2.4 Positioning Siderails..................................................................................................... 13
2.5 Head and Foot Boards.................................................................................................. 13
2.6 Siderail Function Guide................................................................................................ 14
2.7 Foot Board Control Panel Guide................................................................................... 15
2.8 Foley Bag Hooks.......................................................................................................... 17
2.9 Foot Prop Rod.............................................................................................................. 17
2.10 Balkan Frame Installation........................................................................................... 17
2.11 Patient Restraint Strap Locations ............................................................................... 17
2.12 CPR Emergency Release (Optional) .......................................................................... 18
2.13 Night Light (Optional).................................................................................................. 18
2.14 Nurse Call (Optional).................................................................................................. 18
2.15 120V Auxiliary Power Outlet (Optional Only with 120V GOBED) ................................ 18
3. OPTIONAL ACCESSORIES ............................................................................................... 19
ANNEX A: FIXING BOARD PERMANENTLY.......................................................................... 20
3
The GOBED Operations Manual
1. INTRODUCTION
This manual is designed to assist you with the operation of the model FL17E GOBED. Read it thoroughly before operating the bed. Hospital staff should be able to refer to this guide at all time when using the bed.
This Operations Manual is an integral part of the bed and should be included if the bed is sold or transferred.
1.1 TECHNICAL SPECIFICATIONS *
Safe lifting Load 500 lb (227 kg ) including 100 lb (45.5 kg) of
boards and accessories.
Overall Length/Width: :
- with steel siderails up
- with steel siderails down
- with plastic siderails up
- with plastic siderails down
93" x 41 9/16" (236.2 cm x 105.5 cm) 93" x 38 3/8" (236.2 cm x 97.5 cm) 93" x 42 9/16" (236.2 cm x 108.1 cm)
93" x 38 3/4" (236.2 cm x 98.4 cm) Overall Weight (w/o boards and accessories) 385 lb - 175 kg Sleep surface 35" (89 cm) x 80" (203 cm) extendable to 82"
(208 cm) and 84" (213 cm) Minimum/Maximum Bed Height 14" to 29" - 35.5 cm to 73.7 cm Sound Level
< 58 dBa Fowler Angle 0° to 65° Knee Gatch w/o Contour Positioning Knee Gatch w/Contour Positioning
0° to 32°
0° to 24° Trendelenburg/Reverse Trendelenburg -14° to +-14° Electrical Requirements 100 V~, 50-60 Hz, 7.5 A
120 V~, 50-60 Hz, 4.8 A
120 V~, 50-60 Hz, 9.8 A w/auxiliary outlet
200 V~, 50-60 Hz, 3.2 A
220 V~, 50-60 Hz, 2.9 A
240 V~, 50-60 Hz, 2.7 A
* Stryker Bertec pays special attention to product improvement and reserve the right to change specifications without notice.
1.2 TECHNICAL SUPPORT
For questions regarding this product, contact one of the following Technical Service departments or your local representative:
In Canada: In the United States: Stryker Bertec Stryker Medical Service in English: 1 800 428-5025 1 800 327-0770 E-mail (Canada): service@bertec.strykercorp.com 6300, South Sprinkle Road 70, 5th Avenue, P.O. Box 128 Kalamazoo, MI 49001-9799
L’Islet (Québec), G0R 2C0, Canada USA
4
Introduction Chapter 1
WWAARRNNIINNGG
WWAARRNNIINNGG
CCAAUUTTIIOON
N
1.3 WARNING / CAUTION / NOTE DEFINITIONS
The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
The personal safety of the patient or user may be involved. Disregarding this information could result in injury to the patient or user.
These instructions point out special procedures or precautions that must be followed to avoid damaging the equipment.
E
NNOOTTE
Notes provide special information to make maintenance easier or important instructions clearer.
1.4 SAFETY TIPS AND GUIDELINES
Before operating the GOBED, it is important to read and understand all information in this manual. Carefully read and strictly follow the safety guidelines listed on this page.
It is important that all users have been trained and educated on the inherent hazards associated with the usage of electric beds.
The GOBED is not intended for pediatric use.
Serious injury can result if caution is not used when operating the bed. Operate the bed only
when all persons and equipment are clear of the electrical and mechanical systems.
This bed is equipped with a hospital grade plug for protection against shock hazard. It must be plugged directly into a properly grounded receptacle. Grounding reliability can be achieved only when a hospital grade receptacle is used.
To help reduce the number and severity of falls by patients, always leave the bed in the lowest position when the patient is unattended.
Leave the siderails fully up and locked when the patient is unattended. When raising the siderails, be sure you hear the click that signals the up and locked position. Pull firmly on the siderail to ensure it is locked into position.
Always apply the brakes when a patient is on the bed (except during transport). Serious injury could result if the bed moves while a patient is getting on or off the bed. After the brake pedal is engaged, push on the bed to ensure the brakes are securely locked.
Keep siderails in the fully raised position and the sleep surface horizontal in its lowest position when the patient is sleeping unless the patient's medical conditions dictates otherwise.
Siderails, with or without their padded covers or nets, are not intended to serve as restraint devices to keep patient from exiting the bed. Siderails are designed to keep a patient from inadvertently rolling off the bed. It is the responsibility of the attending medical personnel to determine the degree of restraint necessary to ensure a patient will remain safely in bed. Failure to utilize the siderails properly could result in serious patient injury.
To avoid injury to the patient, ensure that the sleep surface is in the lowest position with siderails fully raised and securely locked when moving the bed with a patient in it.
5
The GOBED Operations Manual
To avoid injury to the patient and/or user, do not attempt to move the bed laterally with the steer pedal engaged. The fifth steer wheel cannot swivel.
When the sleep surface sections are articulated, ensure that all patient's extremities are within the raised siderails to avoid patient injury.
When a patient's condition requires greater safety measures for his security, use the lockout switches in the foot board control panel to deactivate the siderail or pendant control commands and install protective pads on the siderails.
The instant CPR release is for emergency use only. Before activating the CPR release, all people and equipment must be removed from the area below and around the head and foot sections of the bed or serious personal injury or damage to equipment could occur.
Possible fire hazard exists when this bed is used with oxygen administering equipment other than nasal, mask type or half bed-length tent type. It is recommended to unplug the bed power cord from the wall when oxygen-administering equipment is used. When using a half bed-length tent type, ensure the siderails are outside the oxygen tent and oxygen tent should not extend below the mattress support level.
When large fluid spills occur in the area of the circuit board, cables and motors, immediately unplug the bed. Remove the patient from the bed and clean up the fluid. Have maintenance completely check the bed. Fluids can have an adverse effect on operational capabilities of any electrical product. DO NOT put the bed back into service until it is completely dried and has been thoroughly tested for safe operation.
Do not steam clean, hose off or ultrasonically clean the bed. Do not immerse any part of the bed. The internal electrical parts may be damaged by exposure to water. Hand wash regularly all surfaces of the bed with warm water and a mild detergent. Wipe cleaned surfaces dry to avoid build up of cleaning substance. Inspect the mattress after each use. Discontinue use if any cracks or rips are found in the mattress cover that may allow fluid to enter the mattress. Failure to properly clean mattress or dispose of defective mattress will increase the risk of exposure to pathogenic substances and may cause injury to the patient and/or user.
Preventive maintenance should be performed periodically to ensure all bed features are functioning properly. Ensure that any bed malfunction is promptly reported to service personnel for immediate attention.
Always unplug the bed power cord from the wall outlet when cleaning or servicing the bed. When working under the bed with the bed in the high position, always place blocks under the frame and lock the casters to prevent injury in case the "Bed Down" switch is accidentally pressed.
To avoid damage to the siderail mechanisms, do not move the bed using the raised siderails. Use the push/pull handles integrated to the boards to move the bed.
When using the emergency crank during a power failure, unplug the power cord from the wall receptacle so that unexpected resumption of power will not rotate handle. Remove and store the crank before reconnecting the bed.
When servicing use only identical replacement parts provided by Stryker Bertec.
NOTE
Throughout this operations manual, the words “right” and “left” refer to the right and left sides of a patient lying face up on the bed
6
Introduction Chapter 1
POSITION
POSITION
POSITION
1.5 WARRANTY
LIMITED WARRANTY
All Stryker Bertec products are guaranteed against material or manufacturing defects, improper operation of mechanisms, and premature wear of bed components under normal use conditions.
For questions regarding warranty, please contact Stryker Bertec Technical Service department (see section 1.2) or your local representative.
TO OBTAIN SERVICE AND/OR PARTS
For an on-site diagnosis and/or repair of a bed malfunction by a Stryker Field Service Representative or to order replacement parts (see section 1.4, "To Order Parts" in the GOBED Maintenance Manual), simply contact Stryker Bertec Technical Service department or your local
representative.
RETURN AUTHORIZATION
Merchandise cannot be returned without approval from the Stryker Bertec Technical Service department. An authorization number will be provided which must be clearly printed on the returned merchandise. Stryker Bertec reserves the right to charge shipping and restocking fees on returned items.
DAMAGED MERCHANDISE
Claims for damaged merchandise must be made with the carrier within fifteen (15) days of receipt of merchandise. DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT. Upon prompt notification, Stryker Bertec will file a freight claim with the appropriate carrier for damages incurred. Claims will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker Bertec within the fifteen (15) days period following the delivery of the merchandise, or the damage was not noted on the delivery notice at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Claims for any short shipment must be made within 5 days of invoice.
1.6 BED POSITION PICTOGRAMS
The following pictograms illustrate the bed basic positions. Not illustrated here is the Cardiac Chair position that is obtained through combining the Reverse
Trendelenburg and the Auto Contour positions.
FOWLER ELEVATION
TRENDELENBURG
VASCULAR
KNEE GATCH ELEVATION
BED Hi-Lo MOTION
REVERSED TRENDELENBURG
AUTO CONTOUR
POSITION
Figure 1.6
7
The GOBED Operations Manual
1.7 SET-UP PROCEDURE
CHECKLIST
It is important to ensure that the bed is working properly before putting it into service. The following list will help ensure that each part of the bed is verified.
____Install the foot and head boards on the bed. Insert the foot board carefully so that the board
connector connects smoothly to the foot end casing connector.
____Plug the power cord into a properly grounded hospital grade wall receptacle and ensure
that the Power LED light at the foot end of the bed comes on when the "On/Off" switch is turned on.
WWAARRNNIINNG
G
The bed is equipped with a hospital grade plug for protection against shock hazard. It must be plugged directly into a properly grounded receptacle. Grounding reliability can be achieved only when a hospital grade receptacle is used.
____Depress the red "TOTAL BRAKE" side of the pedal at either side of the bed to set the
brakes. Make sure the bed is immobilized by pushing on it. Ensure brakes disengage when the pedal is toggled to neutral position.
____Depress the green "AXIAL STEER" side of the pedal on either side of the bed to engage
the 5th wheel. Ensure the 5th wheel is functional. Ensure the 5th wheel disengages when
the pedal is toggled to neutral position. ____Ensure the siderails raise and lower smoothly and lock in the up position. ____Run through each function on the foot board control panel and ensure that each function is
working properly. ____Ensure all functions are working properly on the siderail control panels (inner and outer
sides). Also check the optional nurse call function. ____Raise the bed fully up and activate the Trendelenburg function (see "Trendelenburg Switch
(E)", page 16). Ensure the head end lowers to the full down position.
____Raise the bed fully up and activate the reverse Trendelenburg function (see
"Trendelenburg Switch (E)", page 16). Ensure that the foot end lowers to the full down
position. ____Ensure that the optional Auto Contour function works properly (see "Auto Contour", page
16).
____Leaving the bed in the Auto Contour position, pull the instant CPR release handle (optional
equipment) and ensure the Fowler and Knee Gatch will drop with minimal effort. Wait and listen for the typical noise indicating that the Fowler motor has reset itself (during the resetting period, no bed functions are available).
____Verify the following optional equipment for proper operation: 120V auxiliary outlet,
photoelectric night light, two or three function pendant control, etc. If any problems are found during bed set-up, contact our Technical Service department (see
section 1.2).
8
Introduction Chapter 1
CCAAUUTTIIOON
N
WWAARRNNIINNG
G
1.8 BED CLEANING AND PREVENTATIVE MAINTENANCE
WWAARRNNIINNG
G
Always unplug the bed power cord from the wall outlet when cleaning or servicing the bed.
BED CLEANING AND MATTRESS CARE
Do not use harsh cleaners, solvents or detergents. Do not steam clean, hose off or ultrasonically clean the bed. Do not immerse any part of the bed. The bed electrical parts may be damaged by exposure to water.
Germicidal disinfectant, used as directed, and/or Chlorine Bleach products are not considered mild detergents. These products are corrosive in nature and may cause damage to your bed if used improperly. If these types of products are used, ensure the beds are rinsed with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly causing premature corrosion of critical components. Failure to follow the above directions when using these types of cleaners may void this product warranty.
Bed Cleaning
Hand wash all surfaces of the bed with a soft cloth moistened with a solution of lukewarm water and a mild detergent.
Wipe the bed clean and dry thoroughly to avoid build up of cleaning solution.
Mattress Care
Inspect the mattress after each use. Discontinue use if any cracks or rips are found in the mattress cover, which may allow fluid to enter the mattress. Failure to properly clean the mattress, or dispose of it if defective, may increase the risk of exposure to pathogenic substances and may cause injury to the patient and/or user.
Inspection
Implement local policies to address regular care, maintenance, and cleaning of mattresses and covers. The cover cleaning procedure can be found below and on the bed label.
Inspect mattress cover surface (also zip fasteners and cover inner surface if mattresses have zip fasteners) regularly for signs of damage. If the mattress cover is heavily stained or soiled, or torn, remove the mattress from service.
Cleaning
Stains: Wash with lukewarm water using a mild detergent. Rinse with water and let dry. For tough stains, use bleach diluted with ten parts of water.
9
The GOBED Operations Manual
PREVENTATIVE MAINTENANCE PROGRAM Annual Checklist
____ Inspect for excessive wear the oil-impregnated bronze shoulder spacers found at the bed
hinge points. Replace as needed. Do not lubricate these spacers.
____ Inspect and lubricate when needed the bed lubrication points described in section 2.2 of the
GOBED Maintenance Manual. ____ Inspection of all bolt, locknut and screw tightening, tighten if necessary. ____ Engage the brake pedal on both sides of the bed (see page 12) and push on the bed to ensure
it is completely immobilized.
____ Engage the steer pedal on both sides of the bed (see page 12) and ensure the 5th wheel
operates properly. ____ Siderails move, latch and stow properly (see page 13). ____ All functions on the foot board control panel (see page 15) working properly (give special
attention to lockout switch LED's).
____ All functions of the siderail outer/inner control panels (see page 14) working properly. Ensure
the nurse call signal (if present) reaches the nurse station.
____ Optional instant CPR release handles working properly (see page 18) . The motor resets itself
once the Fowler is down.
____ Verify the Fowler and Knee Gatch movements to ensure the motor course is properly adjusted.
Refer to Caution following step 11 of the "Thigh actuator" and the "Head actuator" replacement
procedures found at page 38 and 39 respectively of the GOBED Maintenance manual. ____ Optional 120 volt auxiliary outlet working properly. ____ Optional photoelectric night light working properly. ____ Optional Auto Contour working properly. ____ Foot prop rod working properly when Knee Gatch or Auto Contour function is activated. ____ No cracks or splits in head and foot boards. ____ On/Off switch and associated LED working properly. ____ Head end bumpers tightly secured to frame and working properly. ____ No rips or cracks in mattress cover. Remove from service if damaged. ____ Power cord not frayed. ____ No cables worn or pinched. ____ All electrical connections tight. ____ All ground secure to the frame. ____ All casters roll properly. Verify caster tire for cuts, wear. ____ Ground chain intact and in place. ____ Measure current leakage and grounding continuity of the bed and the optional 120V auxiliary
outlet. Verify with our Technical Service department (see section 1.2) for the acceptable values for this bed.
E
NNOOTTE
Preventive maintenance may need to be performed more frequently based on the usage level of the bed.
Bed serial number: _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________
Completed by:___________________________________ Date:________________
10
Introduction Chapter 1
HOLDERS
SIDE MATTRESS
RETAINERS
FOOT MATTRESS
RETAINERS
HEAD SIDERAIL
CONTROL PANEL
SIDERAIL LATCH
Figure 1.9
1.9 BED ILLUSTRATION (Bed may differ from illustration)
PUSH/PULL HANDLE
HEAD BOARD
SIDERAIL OUTSIDE
CONTROL PANEL
LEVER
FIFTH WHEEL
BRAKE/STEER
PEDAL
SIDERAIL INSIDE
HEAD SECTION
CENTER SECTION
THIGH SECTION
FOOT SECTION
FOOT SIDERAIL
FOOT BOARD
CONTROL PANEL
FOOT BOARD
ON/OFF SWITCH
EMERGENCY
CRANK STORED
HEAD IV POLE
RESTRAINT STRAP
LOCATIONS
PATIENT HELPER
MOUNTING SOCKETS
HEAD END
CASING
SIDE MATTRESS
RETAINERS
RESTRAINT STRAP
LOCATIONS
FOOT IV POLE
HOLDERS
EMERGENCY
CRANK OPENINGS
FOLEY BAG HOOKS
FOOT END
CASING
11
The GOBED Operations Manual
WWAARRNNIINNG
G
WWAARRNNIINNG
G
2. OPERATION GUIDE
2.1 SWITCHING ON POWER
The main power switch, located beneath the foot board in the middle of the foot end case, enables the bed functions to be activated or deactivated at any time. When the switch is turned on, its green status indicator LED will light up indicating that the bed electric commands are available. Note that siderail controls are automatically shut off when the main power switch is turned off.
2.2 APPLYING THE BRAKE
The GOBED is equipped with a central locking system activated by a brake/steer pedal (see page 11, "Bed Illustration") located at the midpoint of the bed on both sides.
Always apply the brakes when a patient is on the bed (except during transport) or when entering/exiting the bed. Serious injury could result if the bed moves while a patient is getting on or off the bed. After the brake pedal is engaged, push on the bed to ensure the brakes are securely locked.
Brake Pedal Operation
To engage the wheel brakes, press the red "TOTAL BRAKE" side of the brake/steer pedal fully down on either side of the bed.
To disengage the wheel brakes, toggle the pedal to neutral position.
2.3 MOVING THE BED
The GOBED is equipped with a 5th wheel steer mechanism (see page 11, "Bed Illustration") activated by a brake/steer pedal located at midpoint of the bed on both sides.
The 5th wheel helps guiding the bed along a straight line and also for pivoting at corners.
To reduce risk of injury, ensure the sleep surface is horizontal and in the lowest position with the siderails fully raised and locked when moving the bed with a patient in it.
To avoid injury to the patient and/or user, do not attempt to move the bed laterally with the steer pedal engaged. The fifth steer wheel cannot swivel.
To avoid damage to the siderail mechanisms, do not move the bed using the raised siderails. Use the push/pull handles integrated to the boards to move the bed
Steer Pedal Operation
To engage the 5th wheel, press the green "AXIAL STEER" side of the brake/steer pedal fully down on either side of the bed.
To disengage 5th wheel, toggle the pedal to neutral position.
12
Operation Guide Chapter 2
WWAARRNNIINNG
G
CCAAUUTTIIOON
N
2.4 POSITIONING SIDERAILS
To help reduce the number and severity of falls by patients, always leave the bed in the lowest position and siderails fully up when the patient is unattended. After raising a siderail, pull firmly on the siderail to ensure it is securely locked into position
Siderails, with or without their padded covers or nets, are not intended to serve as restraint devices to keep patient from exiting the bed. Siderails are designed to keep a patient from inadvertently rolling off the bed. It is the responsibility of the attending medical personnel to determine the degree of restraint necessary to ensure a patient will remain safely in bed. Failure to utilize the siderails properly could result in serious patient injury.
Keep siderails in the fully raised position and the sleep surface horizontal in its lowest position when the patient is sleeping unless the patient's medical conditions dictates otherwise.
To avoid damage to siderail mechanisms, do not pull up a lowered siderail tucked away. To engage a stored siderail, remove it first from its stored position by pulling it towards you.
To engage a head end siderail, grasp the rail, pull it towards you and swing it upward towards the head end of the bed until it latches in the up position.
To engage a foot end siderail, the same procedure is required as for the head end siderail; however, the siderail swings to the foot end of the bed.
To disengage a siderail, lift it up slightly, push in the latch lever located under the sleep surface (see page 11, "Bed Illustration") and hold it as it rotates down to low position. Push the siderail against the sleep surface to store it.
2.5 Head and Foot Boards
The head board and the foot board slide down into two mounting sockets located at each end of the bed.
Removing Boards
Seize both ends of the board and lift it up until posts come out of the mounting sockets.
Replacing Boards
Head board: Insert board with the laminated finish facing the outer side of the bed. Foot board: Insert the board with the laminated finish facing the inner side the bed. Be cautious
when inserting the foot board to avoid damaging the board and foot end casing connectors.
The boards can be mounted permanently. See annex A for the permanent installation procedure.
E
NNOOTTE
If the the bed electric functions have to remain available when the foot board is removed, ensure the siderail controls are accessible by activating all three lockout switches before removing the foot board.
13
The GOBED Operations Manual
B D F A C E E C A F D B
E C A A C
D B B D
E
2.6 SIDERAIL FUNCTION GUIDE
E
NNOOTTE
Siderail control panel may differ in shape depending on the type of siderail equipping the bed.
OUTER CONTROL PANEL
Right siderail Left siderail
Figure 2.14A
A: Push to raise Fowler C: Push to raise sleep surface E: Push to raise Knee Gatch B: Push to lower Fowler D: Push to lower sleep surface F: Push to lower Kne Gatch
INNER CONTROL PANEL
Right Siderail Left Siderail
A: Push to raise Fowler C: Push to raise Knee Gatch B: Push to lower Fowler D: Push to lower Knee Gatch E: Push for nurse assistance (optional Nurse Call function, see section 2.14, page 18)
14
Figure 2.14B
Operation Guide Chapter 2
RAIL CONTROL
A1 B1 C1 D1 F E A2 A3 B2 C2 D2
2.7 FOOT BOARD CONTROL PANEL GUIDE
SIDE
LOCKOUT
SWITCHES
CONTOUR
WHEN ON
SLEEP SURFACE POSITION SWITCHES
HEAD THIGH HI-LO
AUTOMATIC CONTOUR
PICTOGRAM (OPTIONAL)
TRENDELENBURG
SWITCH
Lockout Switches (A1 to A3) These three switches enable the selective lock out of the bed functions available to patient
and nursing staff through the siderail control panels (inside and outside).
A1:: Push to give access to Fowler adjustment from both interior and exterior siderail control
panels. The LED of the lockout switch used will light up.
A2: Push to give access to Knee Gatch adjustment from both interior and exterior siderail
control panels. The LED of the lockout switch used will light up.
A3: Push to give access to bed Hi-Lo control at the exterior siderail control panel. The LED
of the lockout control used will light up.
E
NNOOTTE
When a lockout switch is used to inhibit a specific function of the siderail control panels, the function is inhibited on both sides of the siderail control panel. The foot board controls are not affected by the lockout switches. Lockout settings are automatically saved in the event of a power failure and restored following resumption of power.
Fowler Control Switches (B1 and B2) These two switches enable the adjustment of the Fowler angle.
B1: Push to raise Fowler. B2: Push to lower Fowler.
Knee Gatch Control Switches (C1 and C2)
These two switches enable the adjustment of the Knee Gatch angle.
C1:Push to raise Knee Gatch. C2: Push to lower Knee Gatch.
15
The GOBED Operations Manual
Hi-Lo Switches (D1 and D2)
These two switches enable adjustment of the sleep surface height (Hi-Lo).
D1: Push to raise the sleep surface D2: Push to lower the sleep surface.
Trendelenburg Switch (E)
This switch, when activated, enables both Trendelenburg positions through the Hi-Lo switches. A lighted green LED on the pictogram indicates that Trendelenburg positioning is available.
E: Push to activate the Trendelenburg positions, which then become available through the
Hi-Lo switches. Trendelenburg: Push switch D1 to lower head end/raise foot end of the bed. Reverse Trendelenburg: Push switch D2 to lower foot end/raise head end of the bed.
E
NNOOTTE
To replace the sleep surface in horizontal position after Trendelenburg positioning, simply press the Trendelenburg switch to deactivate the function (green LED will go out) and lower or raise the sleep surface to its limit. Then position the bed to the desired height.
Auto Contour (optional) Pictogram (F)
This figure, illustrating the Contour position, points to the lockout switch to activate in the
foot board control panel to position the bed automatically to the Contour position.
Contour positioning automatically raises the Knee Gatch as the Fowler is raised. This
position prevents the patient from slipping towards the foot of the bed when the Fowler is
inclined.
F: As indicated by the pictogram arrow, push the Knee Gatch lockout switch (A2) to enable
the Auto Contour positioning.
E
NNOOTTE
To obtain the Auto Contour position when raising the Fowler from a siderail control panel, both Knee Gatch and Fowler functions must be made available to siderail control panels through activating their respective lockout switches (A1 and A2) in the foot board control panel.
16
Operation Guide Chapter 2
WWAARRNNIINNG
G
2.8 FOLEY BAG HOOKS
Foley bag hooks (see page 11, "Bed Illustration") are found at three locations on both sides of the bed, under the edges of the mattress support head, seat and foot sections.
E
NNOOTTE
The Foley bag hooks move when the Fowler is raised or lowered. Fowler motion must be locked out when using these hooks to avoid inadvertent movement of the hooks. To do so, use the Fowler lockout switch at the foot end control panel.
2.9 FOOT PROP ROD
A foot prop rod, integrated to the foot section, is automatically engaged when the Knee Gatch is raised. The foot prop rod maintains the foot section nearly horizontal as the Knee Gatch raises, thus positioning the sleep surface into the vascular position (see page 7, "Bed Positions").
Since the foot prop rod automatically engages when the Knee Gatch is raised, it must be disengaged manually whenever a simple Knee Gatch raise is required. To do so, perform the following steps:
Reach under the foot section, lift the prop rod and hold it while raising the Knee Gatch using the control located on the foot board control panel.
As the Knee Gatch raises, the rod will come to pass the catches. Release it then and continue to raise the Knee Gatch. The foot section will then simply follow the move of the Knee Gatch without being propped.
2.10 BALKAN FRAME INSTALLATION
A standard Balkan frame can be installed on the bed using the IV sockets (see page 11, "Bed Illustration") located on all four corners of the bed. The GOBED IV pole sockets enable the use of IV poles in conjunction with a Balkan frame.
2.11 PATIENT RESTRAINT STRAP LOCATIONS
The GOBED is equipped with 12 separate locations for installing patient restraint straps. Ten of them are located on the mattress support edges directly across from each other whereas the remaining two are located on the top part of the head section, parallel to the head board (see page 11, "Bed Illustration").
Improperly adjusted restraint straps can cause serious injury to a patient. It is the responsibility of the attending medical personnel to determine proper use of restraint straps and restraint strap locations.
17
The GOBED Operations Manual
WWAARRNNIINNG
G
WWAARRNNIINNG
G
2.12 CPR EMERGENCY RELEASE (OPTIONAL)
The instant CPR release is for emergency use only. When activating the CPR release, all people and equipment must be removed from the area below and around the head and foot sections of the bed or serious personal injury or damage to equipment could occur.
If the Fowler and the Knee Gatch are raised and quick access to the patient is needed, pull outward one of the two emergency release handles, located under both top corners of the Fowler, to lower the Fowler and the Knee Gatch to flat position.
E
NNOOTTE
Once the Fowler is completely down, the Fowler motor will reset itself. During this short period of time, the bed electrical functions will not be available
2.13 NIGHT LIGHT (OPTIONAL)
The GOBED may be equipped with an optional photoelectric night light to illuminate the floor area around the bed. The night light turns on as the room lights dim. It is located on the left side of the frame at the foot end of the bed.
2.14 NURSE CALL (OPTIONAL)
The bed may be equipped with an optional Nurse Call function allowing the patient to call for nurse assistance by simply pressing the Nurse Call red button.
The Nurse Call module is integrated to both siderail panel controls, next to the control switches (see page 14).
Included with this option are two Ø1/4" phono jacks located on both sides of the head end case. The phono jacks enable the use of a nurse call cord which can be placed within reach of a patient who is not in bed.
An optional cord with a Ø1/4" phono plug enables the linking of the bed to the hospital nurse call system.
2.15 120V AUXILIARY POWER OUTLET (OPTIONAL ONLY WITH 120V GOBED)
The North American version (120V electrical system) of the GOBED may be equipped with an optional auxiliary power outlet with a 5A breaker. Located on the left side at the foot end of the bed, this feature provides nursing staff with a convenient power source for small electrical equipments.
E
NNOOTTE
Use only equipment with a current consumption of 5 amp or less.
Use only hospital grade electric equipments with the auxiliary power outlet. The use of normal electric equipments may bring the current leakage to a level unacceptable for an hospital equipment.
18
Optional Accessories Chapter 3
3. OPTIONAL ACCESSORIES
Listed below are all the accessories that may be attached to the GOBED as well as their safe working load when applicable.
5/8"Ø removable anodized aluminum IV pole.
1"Ø removable removable anodized aluminum IV pole.
1"Ø fixed anodized aluminium IV pole.
Two-function pendant control with or without hook (Product & Design).
Three-function pendant control with or without hook (Product & Design)
Emergency crank.
Mattress support extension.
10" cushion for the mattress support extension.
Overhead trapeze. Safe working load: 150 lb (68 kg).
Oxygen bottle holder. Safe working load: 75 lb (34 kg).
Flip type bed shelf. Safe working load: 40 lb (18 kg).
Bed cradle.
Chart holder.
Siderail protective pads - set of two (head siderails) or four (all four siderails).
19
The GOBED Operations Manual
A
Annex A: Fixing Board Permanently
Figure annex A
One or both boards can be fixed permanently to the bed frame. Two 1/4" flat washers, two 1/4" spring washers and two 1/4-20 x 1 3/4" hexagon bolts are
needed for the head board. Two 1/4" flat washers, two 1/4" spring washers and two 1/4-20 x 2 1/4" hexagon bolts are needed for the foot board.
Simply insert the appropriate bolt and the two washer types in the factory-installed nut accessible through holes (A) located on the inner sides of the foot end and the head end casings.
E
NNOOTTE
Do not permanently install boards when beds are used in care units where patient may require emergency treatments.
20
Loading...