International Warranty Clause ............................................................19 1
672-1048E MM FL23SE REV Aw ww.stryker.com
Page 7
Introduction
RESTRAINT STRAP
LOCATIONS
HALF
-
LENGTH
SIDERAIL
FOOT SECTION
FOOTBOARD CONTROL
PANEL
(
OPTIONAL
)
EMERGENCY CRANK
OPENINGS
FOOT END
CONTROL PANEL
DRAINAGE BAG
HOOK
BRAKE
/
STEER
PEDAL
5
TH WHEEL
(
OPTIONAL
)
WHEEL COVER
(
OPTIONAL
)
SIDERAIL RELEASE
LEVER
MAIN SW ITCH
(
POWER
CONNECTOR
)
STORED
EMERGENCY
CRANK
(
OPTIONAL
)
BUMPER
HEAD SECTION
CPR RELEASE
HANDLE
OPTIONAL
FIXED SIDERAIL
CONTROL PANEL
(
OPTIONAL
)
INTENDED USE
This manual is designed to assist you with the maintenance of Stryker Model FL23SE Electrical MedSurgBed. Carefully
read this manual thoroughly before using the equipment or beginning maintenance on it. To ensure safe operation of this
equipment, it is recommended that methods and procedures be established for educating and training staff on the safe
operation of this bed.
50° to 104°F
5 to 95% without condensation
700 to 1060 hPa
100V~, 50-60Hz, 7.5A - Two 250V, 10A Fast Acting Fuses
120V~, 50-60Hz, 4A (9.8A with 120V Optional Auxiliary
200V~, 50-60Hz, 3.2A - Two 250V, 6.3A Slow Blow Fuses
220V~, 50-60Hz, 2.9A - Two 250V, 6.3A Slow Blow Fuses
240V~, 50-60Hz, 2.7A - Two 250V, 6.3A Slow Blow Fuses
500 lbs227 kg
229,6 x 104,9 cm
229,6 x 101,3 cm
or 203,2 cm
-40° to 70°C
10° to 40°C
Outlet) - Two 250V, 10A Fast Acting Fuses
Stryker reserves the right to change specifications without notice.
Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations.
*This device has a 10% duty cycle.
Note
The Safe Working Load for the BNQ beds is of 390 lbs (177 kg).
Return To Table of Contents
872-1048E MM FL23SE REV Aw ww.stryker.com
Page 9
Introduction
WARNING / CAUTION / NOTE DEFINITION
The words WARNING, CAUTION, and NOTE carry special meanings and should be carefully reviewed.
WARNING
Alerts the reader about a situation, which if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
CAUTION
Alerts the reader of a potentially hazardous situation, which if not avoided, may result in minor or moderate injury to the
user or patient or damage to the equipment or other property. This includes special care necessary for the safe and
effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or
misuse.
Note
This provides special information to make maintenance easier or important instructions clearer.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A9
Page 10
Symbols
~
10A 250V
6.3A 250V
Warning, Refer to Accompanying Documentation
Alternating Current
Type B Equipment: Equipment providing a particular degree of protection against electric shock,
particularly regarding allowable leakage current and reliability of the protective earth connection.
Class 1 Equipment: Equipment in which protection against electric shock does not rely on basic insulation
only, but which includes an additional safety precaution in that means are provided for the connection of
the equipment to the protective earth conductor in the fixed wiring of the installation in such a way that
accessible metal parts cannot become live in the event of a failure of the basic insulation.
IPX4
Protection from liquid splash
Dangerous Voltage Symbol
Protective Earth Terminal
Potential Equalization Symbol
Medical Equipment Classified by Underwriters Laboratories Inc. with respect to Electric Shock, Fire,
Mechanical and Other Specified Hazards Only in Accordance with UL 60601-1, First Edition (2003) and
CAN/CSA C22.2 No. 601.1-M90 with updates 1 and 2.
Safe Working Load Symbol
Fuse rating for bed with 100V~ and 120V~ electric systems.
Fuse rating for bed with 200V~, 220V~ and 240V~ electric systems.
In accordance with European Directive 2002/96/EC on Waste Electrical and Electronic Equipment
(WEEE), this symbol indicates that the product must not be disposed of as unsorted municipal waste,
but should be collected separately. Refer to your local distributor for return and/or collection systems
available in your country.
Return To Table of Contents
1072-1048E MM FL2 3SE REV Awww.stryker.com
Page 11
www.stryker.com72-1048E MM FL2 3SE REV A11
CONNECTION DIAGRAM
Electrical Service Information
Return To Table of Contents
Page 12
1272-1048 E MM FL 23SE REV Awww.stryker.com
Return To Table of Contents
CONNECTION DIAGRAM (CONTINUED)
Electrical Service Information
Page 13
Preventative Maintenance
WARNING
Use only replacement parts provided by Stryker.
ANNUAL CHECKLIST
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
All fasteners secure.
Inspect for excessive wear the oil-impregnated bronze shoulder spacers found at the bed hinge points.
Replace as needed. Do not lubricate these spacers.
On both sides of the bed, depress fully down the side of the pedal identified with a red sticker and ensure
that the brakes are applied and the bed immobilized. Toggle the pedal to neutral and ensure the brakes
are released.
On both sides of the bed, depress fully down the side of the pedal identified with a green sticker and ensure
that the steer mode (fifth wheel (optional) or bed wheel) is engaged. Toggle the pedal to neutral and ensure
that the steer mode disengages.
Siderails move smoothly and latch properly in high position.
Ensure that each control of the inner and outer head siderail control panels (optional) and those of the
removable pendant control (optional) are working properly.
All controls of the foot end panel and those of the foot board panel (optional) working properly, including
Trendelenburg and lockout LED’s.
Fowler and Knee Gatch (if raised) flatten and the Fowler control motor resets itself automatically when the
two CPR handles (optional) are pulled until Fowler is flattened. Wait about 30 seconds, the time for the
Fowler motor to reset itself, and then raise the Fowler to ensure that the resetting of the motor has indeed
occurred.
Verify the Fowler, Knee Gatch and Hi-Lo movements to ensure that the motion interrupt switch integrated
to the four electric actuators is operating properly.
Auxiliary outlet (option available only with 120V beds) working properly.
Night light (optional) working properly.
Head end bumpers tightly secured to frame and working properly.
No rips or cracks in mattress cover.
Power cord is not frayed. Replace if the protective sleeve is cut or torn.
No cables worn or pinched.
All electrical connections tight.
All grounds secured to the frame.
All casters roll properly. Check caster for cuts, wear, etc.
Measure current leakage and grounding continuity of the bed and the optional auxiliary outlet. Check with
our Technical Service department for the acceptable values.
Do not use harsh cleaners, solvents or detergents. Equipment damage could occur. •Do not steam clean, hose off
or ultrasonically clean the bed. Do not immerse any part of the bed. The bed electrical parts may be damaged
by exposure to water.
Germicidal disinfectant, used as directed, and/or Chlorine Bleach products are not considered as mild detergents. •
These products are corrosive in nature and may cause damage to your bed if used improperly. If these types of
products are used, ensure the beds are wiped with clean water and thoroughly dried following cleaning. Failure to
properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly causing premature
corrosion of critical components. Failure to follow the above directions when using these types of cleaners may void
the product’s warranty.
Procedure:
Hand wash all surfaces of the bed with a soft cloth moistened with a solution of lukewarm water and a mild detergent.
Wipe the bed clean and dry thoroughly to avoid build up of cleaning solution.
MATTRESS CARE
WARNING
Inspect the mattress after each use. Discontinue use if any cracks or rips, which may allow fluid to enter the mattress,
are found in the mattress cover. Failure to properly clean the mattress, or dispose of it if defective, may increase the risk
of exposure to pathogenic substances and may bring about diseases to the patient and user.
Inspection
Implement local policies to address regular care, maintenance, and cleaning of mattresses and covers. The cover •
cleaning procedure can be found below and on the bed label.
Inspect mattress cover surface, interior and exterior, as well as zip fasteners and cover inner surface if mattresses •
have zip fasteners, regularly for signs of damage. If the mattress cover is heavily stained or soiled, or is torn, remove
the mattress from service.
Cleaning
Stains: Wash with lukewarm water using a mild detergent. Rinse with water and let dry. For tough stains, use bleach
diluted with ten parts of water.
Return To Table of Contents
1472-1048E MM FL23 SE REV Awww.stryker.com
Page 15
Troubleshooting Guide
Problem/FailureRecommended Action
No power to bedIs the bed power switch at the head end of the bed •
turned on?
Is the power cord connected to the power connector •
and plugged into the wall outlet?
Is the power cord severed? Replace if so. •
Are the two fuses inside the power connector still •
operational (see pg. 51)?
Verify power at the wall outlet.•
No bed up or down motion when:
The removable pendant control (optional) is used. •
The foot board command is used.•
No Fowler up or down motion when:
The siderail (optional) or removable pendant control •
(optional) is used.
The foot board command is used.•
The Fowler does not fully raise.This situation arises when the CPR emergency release
No Knee Gatch up or down motion when:
The siderail (optional) or removable pendant control •
(optional) is used.
The foot board command is used.•
Is the Hi-Lo lockout activated (LED off)? If so, deactivate •
it (LED on).
Is the cable of the removable pendant control (optional) •
properly connected to the bed connector under the
mattress support?
Check points of the “No power to bed” problem •
described above.
Is the Fowler lockout activated (LED off)? If so, deactivate •
it (LED on).
Is the cable of the siderail control panel (optional) or •
of the removable pendant control (optional) properly
connected to the bed connector under the mattress
support?
Check points of the “No power to bed” problem •
described above.
(optional) is used to partially lower the Fowler. The use
of the CPR for this purpose creates a situation where the
Fowler motor is out of sync with the actual position of the
Fowler. To correct the situation:
Lower the Fowler •completely using a CPR handle or
the Fowler electric control to enable the Fowler motor
to reset itself and thus synchronize its course with the
actual position (flatten) of the Fowler.
Is the Knee Gatch lockout activated (LED off)? If so, •
deactivate it (LED on).
Is the cable of the siderail control panel (optional) or •
of the removable pendant control (optional) properly
connected to the bed connector under the mattress
support?
Check points of the “No power to bed” problem •
described above.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A15
Page 16
Troubleshooting Guide
Problem/FailureRecommended Action
The positioning mechanisms of the Fowler, and/or
the Knee Gatch, and/or the bed height operate with
difficulty.
No Auto Contour (optional) motion.Is the Knee Gatch lockout activated (LED off)? If so, •
Improper operation of the CPR emergency release
(optional): Knee Gatch does not lower and/or the Fowler
actuator does not reset.
Check the four actuator screws to verify their lubrication •
condition and apply grease if needed (see pg. 20).
deactivate it (LED on).
Check and replace if needed the two Auto Contour limit •
switches (see pg. 52).
Check and replace if needed the two CPR mechanism •
limit switches (see pg. 54).
Return To Table of Contents
1672-1048E MM FL23SE REV Awww.stryker.com
Page 17
Quick Reference Replacement Parts List
Note
The parts and accessories listed on this page are all currently available for purchase. Some of the parts identified on
the assembly drawing parts in this manual may not be individually available for purchase. Please call Stryker Customer
Service USA: 1-800-327-0770 (Option 2) or Canada: 1-888-233-6888 for availability and pricing.
Part NamePart Number
Electric/Electronic Components
Motor Control BoardQDF23-0315
Micro Switch1325P003 / QDF9159
Fuse - Fast Acting 10A, 250V (100/120V)QDF8078
Fuse - Slow Blow 6.3A, 250V (200/220/240V)QDF8068
Power Cord with Straight North American Molded PlugQDF8066
Night Light (Optional)QDF9539
5A Circuit Breaker for Auxiliary Outlet (optional)QDF9025
Hi-Lo Actuator23-0702
Head Section Actuator23-0335
Thigh Section Actuator23-0336
Toroidal Transformer - International Series BedsQDF14-1160
Stand-Off Pins: Motors Control BoardQP23-0735
Strain-Relief BushingQDF9541
Support for the Foot End Control Panel MembraneQP17-0127-07
Foot End Control Panel MembraneQDF17-0693 / QDF17-0180
Siderail Components
Protective CapQDFP1514
Transfer Plate - Half-Length Siderail23-0777G / 23-0778G
23-0817G / 23-0818G
Left Inner Membrane without Nurse Call
Right Inner Membrane without Nurse Call
Right Inner Membrane with Nurse Call
Left Inner Membrane with Nurse Call
Left Outer MembraneQDF23-0405
Right Outer MembraneQDF23-0406
PVC Protector for Painted Head Half-Length Siderails23-0476
QDF23-0392 / QDF23-0393
QDF23-0396 / QDF23-0397
PVC Protector for Painted Foot Half-Length Siderails23-0478
PVC Protector for Painted Full-Length Siderails23-0641
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A17
Page 18
Quick Reference Replacement Parts List
Part NamePart Number
Mattress Support Components
Lateral Mattress RetainerQP14034-07
I.V. Pole HolderQP17-0121
Caster
6” with Lock MechanismRD25038807
6” without Lock MechanismRD25049307
6” with Steer MechanismRD25049607
5” with Lock MechanismRD25049407
5” without Lock MechanismRD25039107
5” with Steer MechanismRD25049507
Fifth Wheel Mechanism 5” Caster (Optional)RL5
Base Components
Rue Ring Cotter Diameter 3/8”QDF7878
Green Sticker - Fifth WheelQE71-0496
Red Sticker - BrakeQE71-0511
Wheel Cover - Right (Optional)QP28-0299-07
Wheel Cover - Left (Optional)QP28-0300-07
Brake/Steer PedalQP23-0627
Velcro Fastener25-0550
Miscellaneous
Grey Spray PaintDDCAP-GSP
Shoulder SpacerQDF17-0020
Nylon Cable TieQDF9518
Return To Table of Contents
1872-1048E MM FL23SE REV Awww.stryker.com
Page 19
GROUNDING DIAGRAM
BED
Operation Guide
MAINTENANCE PROCEDURES
WARNING
Always unplug the bed power cord when cleaning or servicing the bed.•
To prevent injury when working under the bed with the bed in the high position, always place blocks under the Hi-Lo •
levers and apply the brakes. Always unplug the bed power cord from the wall outlet when servicing or cleaning the
bed.
STATIC DISCHARGE PRECAUTIONS
The electronic circuits of the bed are protected from static electricity damage only while the bed is assembled. It is
extremely important that all service personnel always use adequate static protection when servicing the electronic
components of the bed.
STATIC PROTECTION EQUIPMENT
The necessary equipment for a proper static protection is:
1 static wrist strap •
1 grounding plug •
1 test lead with a banana plug on one end and an alligator clip on the other.•
STATIC PROTECTION PROCEDURE
Unplug the bed power cord from the wall outlet. 1.
Insert the grounding plug into a properly grounded hospital grade wall receptacle. Plug the banana plug of the test 2.
lead into the receptacle on the grounding plug. Connect the alligator clip on the other end of the test lead to a
ground point on the bed.
Place the static control wrist strap on your wrist. Connect the clip at the other end of the wrist strap cord to a ground 3.
point on the bed.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A19
Page 20
Service Information
C
B
Screw Covers Shown Are Optional
A
WARNING
Always unplug the bed power cord when cleaning or servicing the bed.•
To prevent injury when working under the bed with the bed in the high position, always place blocks under the Hi-Lo •
levers and apply the brakes. Always unplug the bed power cord from the wall outlet when servicing or cleaning the
bed.
ACTUATOR LUBRICATION
Note
In compliance with the service life requirement of the BNQ 6641-120/2003 standard, this unit was designed maintenance
free with regard to lubrication. However, if the bed is used improperly or cleaned without following the recommendations
set forth in the Maintenance Manual (presence of cleaning detergent on the actuator screws), a lubrication of the
actuator could be necessary.
Completely raise the bed and apply the brakes. 1.
Completely raise the head section and flatten the thigh section to expose maximum screw threads on both actuators. 2.
Lower the four siderails.
Unplug the power cord from the wall outlet. 3.
Lift and fold back the foot section toward the head end of the bed. Attach it to the bed using bungee cords. 4.
Using a no. 2 Phillips screwdriver, remove the four screws (A) holding the cover plate to the frame and remove the 5.
plate. Remove the optional night light if present.
a. Using a 1/4” ratchet (with 6” extension) and a 5/16” socket, remove the two screws (B) holding the dust tube 6.
of each knee gatch actuator. Push the tube to free it.
b. For the head actuator, remove the two screws, spring washer (C) holding the dust tube of the actuator. Push
the tube to free the threads.
Using a brush, apply grease on the screw threads. Make sure the grease reaches the bottom of the threads.7.
Replace the dust tubes. 8.
Using the electric controls, raise and lower several times the Fowler and Knee Gatch to spread the grease evenly. 9.
Replace the cover plate.10.
Hi-Lo Actuators
Lower the bed completely. 1.
Completely raise the head and foot sections, then lift and fold back the foot section toward the head end of the 2.
bed.
Unplug the power cord from the wall outlet. 3.
Using a brush, apply grease on the threads of the bed lift actuators screws. Optional dust tubes may be present, if 4.
so, simply apply grease through the dust tube side openings.
Using the electric controls, raise and lower the bed several times to spread the grease evenly.5.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A21
Page 22
I
I
J
A
SIDERAIL REPLACEMENT
E
H
G
K
E
F
A
D
C
J
I
HALF-LENGTH RAIL REPLACEMENT
Tools Required:
3/8” Drive Ratchet•
Small Slotted Head Screwdriver •
Long Nose Pliers •
Retaining Ring Pliers •
Rubber Mallet•
7/16” Socket•
Service Information
Right Head Siderail Shown
Return To Table of Contents
2272-1048E MM FL2 3SE REV Awww.stryker.com
Figure 4.1 A
Page 23
Service Information
SIDERAIL REPLACEMENT (CONTINUED)
HALF-LENGTH SIDERAIL
Procedure:
Completely raise the bed and apply the brakes. 1.
Position all the mattress support sections to flat. 2.
Unplug the bed power cord from the wall outlet. 3.
Raise the siderail needing repair. 4.
Using a small slotted head screwdriver and long nose pliers, remove the two protective caps (A). Be careful while 5.
using the tools not to damage the siderail paint.
Note
Replace the caps using a rubber mallet. The removal of the caps may damage them; make sure you have some on
hand.
a. For a head siderail, using the retaining ring pliers, remove the retaining ring on one arm and using the 7/16” 6.
socket, remove the nut, spring washer and spacer holding the rail in place.
b. For a foot siderail, remove using the 7/16” socket the nut, the spring washer and the spacer holding
the siderail in place (B).
Note
Inspect the self-lubricating bearings for wear and replace them if need be.
Reverse the above steps to install the replacement rail. 7.
Verify the siderail movement for proper operation before returning the bed to service.8.
Unless otherwise indicated, refer to Figure 4.1 A on pg. 22 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Depending on the position of the siderail needing repair, place the mattress support sections as follows: 2.
Head Siderail: Completely raise the head section. •
Foot Siderail: Lower completely the head section. Completely raise the thigh section, then lift and fold back the •
foot section toward the head end of the bed.
Unplug the bed power cord from the wall outlet. 3.
Raise the siderail needing repair. 4.
Using diagonal pliers, cut, if applicable, the cable ties holding the siderails fixed control cable (optional). Loosen the 5.
lock ring (D) and unplug the cable.
Note
Note the position and the gathering of the cable before cutting the cable ties to make sure the replacement cable will
be connected correctly.
Using the 7/16” socket and 7/16” combination wrench, remove the two locknuts and bolts (E) holding the two half 6.
bushings (F), the stopper (G) and the siderail mechanism to each of the two siderail support (H). Remove the
half bushings and the mechanism from their location, hold the bushings tightly to prevent them from falling to the
ground.
Note
First, remove the two locknuts and bolts at one end of the assembly and, while holding the assembly, finish with the ones
at the opposite end.
Reverse the above steps to install the replacement siderail assembly. 7.
Verify the siderail for proper operation before returning the bed to service.8.
Unless otherwise indicated, refer to Figure 4.1 A on pg. 22 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Depending on the siderail position, place the mattress support sections as follows: 2.
Head Siderail: Completely raise the head section. •
Foot Siderail: Lower completely the head section. Completely raise the thigh section, then lift and fold back the •
foot section toward the head end of the bed.
Unplug the bed power cord from the wall outlet. 3.
Lower the siderail needing repair. 4.
Using the 7/16” socket and 7/16” combination wrench, remove the two locknuts, bushings, washers and bolts (I) 5.
holding the transfer plate (J) to the siderail mechanism.
Note
At assembly, tighten the two locknuts to a 75 in/lb torque using the torque wrench.
Reverse the above steps to install the replacement transfer plate. 6.
Verify the siderail for proper operation before returning the bed to service.7.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A25
Page 26
2672-1048E MM FL23SE REV Awww.stryker.com
A
D
F
J
C
F
C
E
F
A
B
D
F
G
H
I
Return To Table of Contents
Service Information
Figure 4.1 B
Page 27
Service Information
FULL-LENGTH SIDERAIL PIVOT TUBE REPLACEMENT
Tools Required:
1/2” Combination Wrench •
Rubber Mallet •
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.1 B on pg. 26 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Lower the siderail needing repair. 3.
Using the 1/2” combination wrench, remove the two bolts and washers (A) holding the pivot tube ends to the siderail 4.
mechanism arms. Use a rubber mallet to free the tube ends from the arms.
Slide the tube out of its location.5.
Note
Apply grease on a 4” distance from both ends of the tube.
Reverse the above steps to install the replacement pivot tube. 6.
Verify the siderail movement before returning the bed to service.7.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A27
Page 28
Service Information
FULL-LENGTH RAIL REPLACEMENT
Tools Required:
1/2” Combination Wrench •
Rubber Mallet •
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.1 B on pg. 26 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Raise the siderail needing repair. 3.
Loosen, if applicable, the lock ring (B) and unplug the removable pendant control (optional) cable. Remove the 4.
pendant control from the siderail by sliding it out its supports.
Using the 1/2” combination wrench, remove the four bolts and washers (A, C) holding the rail to the siderail 5.
mechanism arms. Use a rubber mallet to free the tube ends from the arms.
Note
Hold the rail firmly when removing the last bolts.
Reverse the above steps to install the replacement rail. 6.
Verify the siderail movement before returning the bed to service.7.
Return To Table of Contents
2872-1048E MM FL23SE REV Awww.stryker.com
Page 29
Service Information
FULL-LENGTH SIDERAIL ASSEMBLY REPLACEMENT
Tools Required:
No. 2 Phillips Screwdriver•
Two 1/2” Combination Wrenches•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.1 B on pg. 26 for the following replacement procedure.
Completely raise the bed and raise the siderail needing repair. 1.
Unplug the bed power cord from the wall outlet.2.
Remove the head and foot boards. 3.
Using two 1/2” combination wrenches, remove the two locknuts and bolts (K, Figure 4.3 C, 4. pg. 43) holding the two
bumpers to their supports at the head end of the bed.
Using the no. 2 Phillips screwdriver, remove the seven screws (Q, Figure 4.3 C, 5. pg. 43) holding the cover to the head
end casing. Remove the cables connected to the power connector after having noted their location and remove the
cover.
Using the no. 2 Phillips screwdriver, remove the 10 screws (A, Figure 4.2 A, 6. pg. 32) holding the cover and the I.V.
pole holders to the foot casing. Lift and remove the cover while taking care to disconnect the control panel cable
(B, Figure 4.2 A, pg. 32) from the control board and, if applicable, the foot end casing connector cable (C, Figure
4.2 A, pg. 32) (optional).
Using the no. 2 Phillips screwdriver, remove the four screws (D) holding together the two parts of both siderail 7.
mechanism covers. Remove the covers.
Using the 1/2” combination wrench, remove the four nuts, washers and bolts (E) holding the siderail release handle 8.
to the seat section.
Using the 1/2” combination wrench, remove the two nuts, washers and bolts (F) holding each end of the full-length 9.
siderail to the head end and foot end casings.
Reverse the above steps to install the replacement siderail assembly. 10.
Verify the siderail for proper operation before returning the bed to service.11.
Unless otherwise indicated, refer to Figure 4.1 B on pg. 26 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Raise both siderails. 3.
Using needle nose pliers, remove the return spring (G). 4.
Attach the siderail needing repair to the other siderail using a bungee cord. 5.
Using the No. 1 Phillips screwdriver, remove the six screws (H) holding both parts of the release handle. Hold the 6.
lower part to prevent it from falling.
Using the 7/16” combination wrench, remove the upper part of the handle by turning the lock rod (I) towards the 7.
floor.
Note
Use the same procedure to insert the upper part of the new handle on the rod, but this time turn the lock rod up.
Reverse the above steps to install the replacement release handle. 8.
Verify the release handle for proper operation before returning the bed to service.9.
Return To Table of Contents
3072-1048E MM FL23SE RE V Awww.stryker.com
Page 31
Service Information
FULL-LENGTH SIDERAIL LOCKING HOOK REPLACEMENT
Tools Required:
No. 2 Phillips Screwdriver•
Bungee Cord•
Two 1/2” Combination Wrenches•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.1 B on pg. 26 for the following replacement procedure.
Completely raise the bed and raise the siderail needing repair.1.
Unplug the bed power cord from the wall outlet. 2.
If the hook is at the head end of the bed, follow steps 3 to 6, then go to step 8.
If the hook is at the foot end of the bed, go directly to step 7.
Head End
Remove the head board.3.
Raise the defective siderail.4.
Using the two 1/2” combination wrenches, remove the two locknuts and bolts (K, Figure 4.3 C, 5. pg. 43) holding the
bumpers to their support.
Using the No. 2 Phillips screwdriver, remove the seven screws (Q, Figure 4.3 C, 6. pg. 43) holding the cover of the head
box. Remove cables connected to power connector taking note of their position then remove cover.
Foot End
Using the No. 2 Phillips screwdriver, remove the 10 screws (A, Figure 4.2 A, 7. pg. 32) holding the cover as well as
the two I.V. poles to the foot box. Raise and remove the cover taking care to disconnect the electronic board first,
the cable (B, Figure 4.2 A, pg. 32) of the control panel and, if applicable, the cable (C, Figure 4.2 A, pg. 32) of the
connector from the foot box (optional).
Using the No. 2 Phillips screwdriver, remove the four screws (D) holding the two parts of the plastic cover over the 8.
siderail mechanism. Remove cover.
Using the two 1/2” combination wrenches, remove locknut, shoulder spacer and bolt (J) holding the locking hook 9.
to the plate. Remove hook.
Reverse the preceding steps to install the new hook.10.
Verify siderail for proper operation before returning the bed to service.11.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A31
Page 32
A
A
A
B
C
D
E
F
Service Information
FOOT END CONTROL PANEL REPLACEMENT
Tools Required:
No. 2 Phillips Screwdriver •
Note
The control panel is made of a membrane and its support. Both parts should be ordered for this replacement procedure.
Refer to the “Quick Reference Replacement Parts List” section on pg. 17.
Procedure:
Figure 4.2 A
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board. 3.
Using the No. 2 Phillips screwdriver, remove the 10 screws (A) holding the cover and the I.V. pole holders to the foot 4.
casing.
Use the Static Discharge Precautions procedure (see 5. pg. 19).
Lift and remove the cover taking care to disconnect the electronic board, the cable (B) from the control panel and, 6.
if applicable, the foot casing connector cable ((optional), C). Lay the cover on a workbench.
Using the No. 2 Phillips screwdriver, remove the three screws, washers and spacers (D) holding the panel to the 7.
cover.
Reverse the above steps to install the new control panel. 8.
Verify all the controls of the foot end control panel before returning the bed to service.9.
Unless otherwise indicated, refer to Figure 4.2 A on pg. 32 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board. 3.
Using the No. 2 Phillips screwdriver, remove the 10 screws (A) holding the cover and the two I.V. pole holders to the 4.
foot casing.
Lift and remove the cover taking care to disconnect the electronic board, the control panel cable (B) and the foot 5.
casing connector cable (C). Lay the cover on a workbench.
Using the 3/8” combination wrench, remove the two locknuts, shoulder spacers and screws (E) holding the connector 6.
and the seal (F) to the cover. Remove the defective connector using the 3/8” combination wrench and No. 2 Phillips
screwdriver. Keep the seal for the replacement connector.
Reverse the above steps to install the new connector. 7.
Verify all the controls of the foot end control panel for proper operation before returning the bed to service.8.
The control panel is made of a membrane and its support. Both parts should be ordered for this replacement procedure.
Refer to the “Quick Reference Replacement Parts List” section on pg. 17.
Figure 4.2 B
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board and lay it on a workbench. 3.
Using the No. 2 Phillips screwdriver, remove the seven screws (A) holding the support cover to the inner face of the 4.
foot board and remove the cover.
Disconnect the foot board panel cable (B). 5.
Using the No. 2 Phillips screwdriver, remove the last screw (C) holding the control panel to the foot board. Remove 6.
the panel. A small part of the panel cable is stuck to the inner face of the panel with a small double-sided tape. Do
the same at reassembly.
Reverse the above steps to install the replacement foot board panel. 7.
Verify all the controls of the foot board panel for proper operation before returning the bed to service. 8.
Return To Table of Contents
3472-1048E MM FL23SE REV Awww.stryker.com
Page 35
Service Information
FOOT BOARD CONNECTOR REPLACEMENT (OPTIONAL)
Tools Required:
No. 2 Phillips Screwdriver •
9 mm Combination Wrench •
Procedure:
Note
This connector is present when the bed is equipped with a foot board control panel.
Unless otherwise indicated, refer to Figure 4.2 B on pg. 34 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board and lay it on a workbench. 3.
Using the No. 2 Phillips screwdriver, remove the seven screws (A) holding the support cover to the inner face of the 4.
foot board and remove the cover.
Disconnect the foot board connector cable (B). 5.
Using the 9 mm combination wrench and No. 2 Phillips screwdriver, remove the two locknuts, washers, sleeves and 6.
screws (D) holding the connector to its support. Remove the defective connector.
Reverse the above steps to install the replacement connector. 7.
Verify the foot board controls for proper operation before returning the bed to service.8.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A35
Page 36
A
C
B
B
A
Service Information
SIDERAIL CONTROL PANEL MEMBRANE REPLACEMENT
Tools Required:
No. 2 Phillips Screwdriver•
Note
The replacement of a membrane implies that the one facing it on the inner or outer control panel also has to be replaced.
Service kits, including both membranes are available.
Figure 4.2 C
Procedure:
Completely raise the bed and apply the brakes. Bring the head section to horizontal position. 1.
Unplug the bed power cord from the wall outlet. 2.
Raise the head siderail needing repair. 3.
Properly ground yourself (see 4. pg. 19).
Detach the two membranes (A) from their mounting plates and let them hang down. 5.
Using a No. 2 Phillips screwdriver, remove the eight screws (B) holding together the two sections of the control 6.
panel housing.
Open the housing, disconnect the defective membrane cable (C) and remove the membrane. 7.
Reverse the above steps to install the new control panel membranes.8.
Note
Make sure to connect the membrane cables before fastening together the two sections of the control panel housing.
Stick the membranes to the support at the end of the replacement procedure.
Verify all the siderail functions for proper operation before returning the bed to service.9.
Completely raise the bed and apply the brakes. Flatten all sections of the mattress support. 1.
Unplug the bed power cord from the wall outlet. 2.
Lower the siderails. 3.
Lift and fold back the foot section toward the head end of the bed. Secure the section to the bed using a bungee 4.
cord.
Using the No. 2 Phillips screwdriver, remove the two lateral mattress retainers (A). 5.
Remove the foot end mattress retainer (B). 6.
Remove the foot prop rod (C). 7.
Remove the bungee cord and replace the foot section to horizontal position. 8.
Using the 1/2” combination wrench, remove the two locknuts, washers, shoulder spacers and bolts (D) linking the 9.
foot section to the thigh section and remove the defective section.
Reverse the above steps to install the new foot section.10.
Unless otherwise indicated, refer to Figure 4.3 A on pg. 37 for the following replacement procedure.
Completely raise the bed and apply the brakes. Flatten all sections of the mattress support. 1.
Unplug the bed power cord from the wall outlet. 2.
Lower the siderails. 3.
Lift and fold back the foot section toward the head end of the bed. Secure the section to the bed using a bungee 4.
cord.
Using needle nose pliers, remove the rue ring cotter, washers and clevis pin (D, Figure 4.3 C, 5. pg. 43) linking the thigh
section lever arms to the thigh actuator tube.
Note
Before linking back the thigh actuator tube to the thigh section lever arms, verify that the actuator course was preserved.
To do so, refer to the caution that follows step 16 of the “Knee Gatch Actuator Replacement Procedure” on pg. 45 and
proceed with step 17 of the procedure.
Remove the bungee cord and bring the foot section back to flat position. 6.
Using the 1/2” combination wrench, remove the four locknuts, washers, shoulder spacers and bolts (D, E) linking the 7.
thigh section to the foot and seat sections and remove the thigh section.
Reverse the above steps to install the replacement thigh section.8.
Unless otherwise indicated, refer to Figure 4.3 A on pg. 37 for the following replacement procedure.
Completely raise the bed and apply the brakes. Flatten the mattress support. 1.
Unplug the power plug from the wall outlet. 2.
Lower the siderails. 3.
Using the 1/2” socket, remove the four thread rolling bolts (F) holding the seat section to the frame.4.
Note
At reassembly, screw the bolts in manually and then tighten applying a 135 lb/in torque.
Using the 1/2” combination wrench, remove the four locknuts, washers, shoulder spacers and bolts (E, G) linking 5.
the seat section to the thigh and head sections and remove the seat section.
Using the Phillips No. 2 screwdriver turn the seat section up side down and remove the three screws (H) holding the 6.
protective plate to the seat section. Keep the plate for the replacement seat section.
Remove the four rubber caps (K) covering the screws (L) holding the stops (M) to the defective section and replace 7.
the parts on the replacement section.
Reverse the above steps to install the replacement seat section.8.
Return To Table of Contents
4072-1048E MM FL23 SE REV Awww.stryker.com
Page 41
Service Information
A
B
HEAD SECTION REPLACEMENT
Tools Required:
No. 2 Phillips Screwdriver •
Strap•
Two 1/2” Combination Wrenches•
Two 7/16” Combination Wrenches •
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.3 A on pg. 37 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
If the bed is equipped with full-length siderails, lower them. If equipped with half-length siderails, raise the head 2.
siderails and lower the foot ones.
Remove the head board. 3.
Completely raise the head section and attach it securely to the bed using a strap. 4. It must not move during the procedure.
Unplug the bed power cord from the wall outlet. 5.
Using the No. 2 Phillips screwdriver, remove the two lateral mattress retainers (I) and keep them for the replacement 6.
head section.
If the bed is not equipped with the CPR emergency release (optional), proceed with steps 7 to 14. Otherwise, proceed
with steps 15 to 24.
www.stryker.com72-1048E MM FL2 3SE REV A41
Figure 4.3 B
Return To Table of Contents
Page 42
Service Information
HEAD SECTION REPLACEMENT (CONTINUED)
Using two 1/2” combination wrenches, remove the locknut, spacer and bolt holding the upper part of the compression 7.
bar to the head section and lay it down.
Using two 1/2” combination wrenches, remove the locknut, shoulder spacers and bolt (B) linking the mattress 8.
support lever to the head section and remove the lever-compression bar assembly.
Remove the strap and manually lower the head section completely. 9.
If the bed is equipped with half-length siderails, lower the head siderails and use the No. 2 Phillips screwdriver 10.
to remove the four screws (K, Figure 4.1 A, pg. 22) holding the two siderail supports to the head section. Let the
support-siderail assembly rest on the frame.
Using the 1/2” combination wrench, remove the two locknuts, flat washers, shoulder spacers and bolts (G) linking 11.
the head section to the seat section. Remove the defective head section.
Using the no. 2 Phillips screwdriver, remove the two screws (J) holding the micro switch activator (optional) to the 12.
head section.
Note
This activator is present only if the bed is equipped with the optional Auto Contour positioning. Remember the position of
the micro switch activator before removing the screws. Properly positioned, the blade of the Auto Contour micro switch
is pushed in by the activator when the Fowler is in flat position.
Reverse the above steps to install the replacement head section. 13.
Verify the Auto Contour positioning (optional) for proper operation before returning the bed to service. 14.
End of the procedure for beds with Auto Contour.
Note
Unless otherwise indicated, refer to Figure 4.3 A on pg. 37 for the following replacement procedure.
Disconnect the two wires (A, Figure 4.3 F, 15. pg. 54) from the micro switch located on the head section lever. Note
their position before doing so.
Using two 7/16” combination wrenches, remove the locknut and bolt (O, Figure 4.3 F, 16. pg. 54) holding the cylinder
end fitting to the head section.
Using two 1/2” combination wrenches, remove the nut, shoulder spacers and bolt (P, Figure 4.3 F, 17. pg. 54) holding
the upper part of the section lever to the head section.
Using a 7/16” combination wrench, remove the two locknuts and bolts (Q, Figure 4.3 F, 18. pg. 54) holding each CPR
release handle to the head section.
Remove the strap and manually lower the head section completely. 19.
If the bed is equipped with half-length siderails, lower the head siderails and use the No. 2 Phillips screwdriver 20.
to remove the four screws (K, Figure 4.1 A, pg. 22) holding the two siderail supports to the head section. Let the
supports-siderails assembly rests on the frame.
Using the 1/2” combination wrench, remove the two locknuts, shoulder spacers, washers and bolts (G) holding the 21.
head section to the seat section. Remove the defective head section.
Using the No. 2 Phillips screwdriver, remove the two screws holding the activator (J) to the head section.22.
Note
Note the position of the activator before removing the screws. Properly positioned, the blade of the CPR and Auto
Contour (optional) micro switches are pushed in by the activator when the Fowler is in flat position.
Reverse the above steps to install the replacement head section. 23.
Verify the CPR emergency release and the Auto Contour positioning (optional) for proper operation before returning 24.
the bed to service.
End of the procedure for beds with CPR and Auto Contour.
Return To Table of Contents
4272-1048E MM FL23SE REV Awww.stryker.com
Page 43
A
D
B
C
E
F
G
H
I
H
I
J
N
O
P
FOWLER ACTUATOR REPLACEMENT
Q
Q
C
C
L
M
Tools Required:
Needle Nose Pliers•
No. 2 Phillips Screwdriver •
Diagonal Pliers •
5/16” Socket •
Small Regular Head Screwdriver •
Pliers•
Bungee Cord •
1/4” Drive Ratchet•
Procedure:
Service Information
Figure 4.3 C
www.stryker.com72-1048E MM FL2 3SE REV A43
Return To Table of Contents
Page 44
Service Information
FOWLER ACTUATOR REPLACEMENT (CONTINUED)
Completely raise the bed and apply the brakes. Completely raise the Knee Gatch section and flatten the Fowler 1.
section.
Unplug the bed power cord from the wall outlet. 2.
Raise the siderails. 3.
Manually lift and fold the foot section back toward the head end of the bed. Attach it securely to the bed using a 4.
bungee cord.
Using needle nose pliers, remove the rue ring cotter, washers and clevis pin (A) linking the head actuator tube to 5.
the section lever arms.
Remove the foot board. 6.
Using the No. 2 Phillips screwdriver, remove the 10 screws (A, Figure 4.2 A, 7. pg. 32) holding the cover and the two
I.V. pole holders to the foot casing.
Lift up and remove the cover, taking care first to disconnect, from the control board, the control panel cable 8.
(B, Figure 4.2 A, pg. 32) and, if applicable, the foot casing connector cable ((optional), C, Figure 4.2 A, pg. 32).
Properly ground yourself (see 9. pg. 19).
Using cutting pliers, cut the cable tie holding the head actuator cable to the other cables. 10.
Disconnect the actuator cable (connector J8) from the control board and remove the cable from the foot end 11.
casing.
Using the 5/16” socket, remove the two screws (B) holding the retaining plate to the actuator support.12.
Remove the two pivot pins (C) holding the actuator to the support. To facilitate the removal of the pins, insert a small 13.
slotted head screwdriver into the opening at the end of the actuator and push out the pins.
Move the actuator toward the centre of the bed to remove it from its location. 14.
Reverse the above steps to install the replacement actuator. Take note of the following caution before hooking 15.
up the actuator tube to the head section lever arms.
CAUTION
It is of utmost importance that the course of the replacement head actuator be adjusted before hooking up its tube to
the head section lever arms. An improper adjustment can damage the head section structure.
To adjust the course of the replacement head actuator, proceed as follows: 16.
Make sure that the actuator cable is connected to the control board, and then connect the bed power cord. A.
Grab the replacement actuator tube and position its holes horizontally. While firmly holding the tube with one B.
hand to prevent it from rotating, press the Fowler up control for a few seconds using the other hand, then press
the down control until the actuator stops. This will be the lower limit of the Fowler actuator course.
Gently turn the tube in either direction to align the tube holes with those of the head section lever arms. Then, C.
still holding firmly the tube, raise again the Fowler a few inches, and lower it completely.
Check the alignment of the holes. If the tube holes are not aligned any more with those of the lever arms, repeat D.
steps B and C until they are. Once the holes are aligned, install the washers and the clevis pin. Slightly raise
the head section manually to ease the insertion of the clevis pin.
Before installing the rue ring cotter to finalize the head actuator replacement, raise and lower completely the E.
Fowler. Check that the actuator stops working as soon as the head section reaches the frame.
Return To Table of Contents
4472-1048E MM FL23SE RE V Awww.stryker.com
Page 45
Service Information
KNEE GATCH MOTOR REPLACEMENT
Tools Required:
Bungee Cord •
Needle Nose Pliers •
No. 2 Phillips Screwdriver •
Pliers •
Diagonal Pliers •
Small Regular Head Screwdriver •
5/16” Socket •
1/4” Drive Ratchet•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.3 C on pg. 43 for the following replacement procedure.
Completely raise the bed and apply the brakes. Flatten the mattress support. 1.
Unplug the bed power cord from the wall outlet. 2.
Raise the siderails. 3.
Manually lift and fold the foot section back toward the head end of the bed. Attach it securely to the bed using a 4.
bungee cord.
Using needle nose pliers, remove the rue ring cotter, washers and clevis pin (D) linking the thigh actuator tube to 5.
the thigh section lever arms.
Using the No. 2 Phillips screwdriver, remove the 10 screws (A, Figure 4.2 A, 6. pg. 32) holding the cover and the two
I.V. pole holders to the foot casing.
Lift up and remove the cover, taking care to first disconnect the control panel cable (B, Figure 4.2 A, 7. pg. 32) and, if
applicable, the foot casing connector cable ((optional), C, Figure 4.2 A, pg. 32) from the control board.
Properly ground yourself (see 8. pg. 19).
Using diagonal pliers, cut the cable tie holding the thigh actuator cable to the other cables. 9.
Disconnect the motor cable (connector J9) from the control board. 10.
Using pliers, squeeze the upper part of the black strain-relief bushing (E) holding the actuator cable to the foot 11.
casing and raise the bushing to disengage it from its location.
Disengage the cable from the strain-relief bushing and remove the cable from the foot casing. 12.
Using the 5/16” socket, remove the two screws (F) holding the retaining plate to the actuator support. 13.
Remove the two pivot pins (G) holding the motor to the support. To facilitate the removal of the pins, insert a small 14.
regular head screwdriver into the opening at the end of the actuator and push out the pins.
Push on the defective actuator toward the centre of the bed to remove it from its location. 15.
Reverse the above steps to install the replacement actuator. Take note of the following caution before hooking up 16.
the actuator tube to the thigh section lever arms.
CAUTION
It is of utmost importance that the course of the replacement thigh actuator be adjusted before hooking up its tube to
the thigh section lever arms. An improper adjustment can damage the thigh section structure.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A45
Page 46
Service Information
KNEE GATCH MOTOR REPLACEMENT (CONTINUED)
To adjust the travel of the replacement knee gatch motor, proceed as follows: 17.
Make sure that the motor cable is connected to the control board, then connect the bed power cord.A.
Grab the replacement motor tube and position its holes horizontally. While firmly holding the tube with one hand B.
to prevent it from rotating, press the Knee Gatch up control for a few seconds using the other hand, then press
the down control until the actuator stops. This will be the lower limit of the Knee Gatch actuator course
Gently turn the tube in either direction to align the tube holes with those of the thigh section lever arms. Then, C.
still holding firmly the tube, raise again the Knee Gatch a few inches, and lower it completely.
Check the alignment of the holes. If the tube holes are not aligned any more with those of the lever arms, repeat D.
steps B and C until they are. Once the holes are aligned, install the washers and the clevis pin. Slightly raise
the head section manually to ease the insertion of the clevis pin.
Before installing the rue ring cotter to finalize the head actuator replacement, raise and lower completely the E.
Knee Gatch. Check that the actuator stops working as soon as the thigh section reaches the frame.
In order to preserve the adjustment of the bed’s lowest position when replacing a bed lift motor, a special tool kit
designed for that purpose must be used. The kit includes alignment jigs. To obtain this kit, contact our Customer Support
department and order part number KR0121.
Unless otherwise indicated, refer to Figure 4.3 C on pg. 43 for the following replacement procedure.
Position the mattress support sections depending on the location of the bed lift motor to replace: 1.
At the A. foot of the bed: Completely raise the thigh section, then lift and fold back the foot section toward the
head end of the bed. Attach it securely to the bed using
a bungee cord. Finally completely raise the Fowler.
At the B. head of the bed: Completely raise the Fowler.
Position the alignment jigs on the floor, on a plane surface, 2.
right under the two transversal tubes supporting the link arms
of both bed lift levers. Lower the bed until the tubes come to
rest on the jigs (see opposite Figure). Use the 3/8” ratchet
wrench with the 1/2” socket provided in the kit to lower a
defective bed lift motor until the tube rests on the jig.
Unplug the bed power cord from the wall outlet. 3.
Raise the siderails. 4.
Disconnect the motor cable and cut, using diagonal pliers, 5.
the cable ties holding it to the frame.
Using the 5/16” socket, remove the two screws (H) holding 6.
the retaining plate to the actuator support.
Remove the two pivot pins (I) holding the motor to the support. To facilitate the removal of the pins, insert the small 7.
regular head screwdriver into the opening at the end of the actuator and push out the pins.
Using the 1/2” combination wrench, remove the two bolts and washers (J) holding the molded nut support to the 8.
Hi-Lo lever. Remove the molded nut support and keep it for the replacement actuator.
Note
Make sure that the support and the molded nut holes are aligned before screwing in the bolts. If resistance is felt, it
means that the holes are not aligned.
Remove the defective motor. 9.
Using the No. 2 Phillips screwdriver, remove the two screws holding together the two parts of the screw cover 10.
(optional) and keep it for the replacement actuator.
Reverse the above steps to install the replacement actuator. Take note of the following caution before hooking up 11.
the motor to the bed lift lever.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A47
Page 48
Service Information
HI-LO ACTUATOR REPLACEMENT (CONTINUED)
CAUTION
The course of the replacement motor must be adjusted prior to hooking it to the bed lift lever. An improper adjustment
can damage the bed lift mechanism.
To adjust the replacement motor, proceed as follows:12.
Once the replacement motor cable is connected, plug the bed power cord and press the bed down control until A.
the motor stops. This will be the lower limit of the actuator course.
Link the motor to the bed lift lever. B.
The alignment jigs still in position, completely raise and lower the bed to ensure that the lower limit has been C.
preserved following the motor replacement.
Return To Table of Contents
4872-1048E MM FL23SE RE V Awww.stryker.com
Page 49
Service Information
TOROIDAL TRANSFORMER REPLACEMENT (INTERNATIONAL SERIES MODELS)
Unless otherwise indicated, refer to Figure 4.3 C on pg. 43 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the head board. 3.
Using a No. 2 Phillips screwdriver, remove the seven screws (L) holding the cover to the head casing. 4.
Remove the cables connected to the power connector after having carefully noted their position and remove the 5.
cover.
Disconnect the transformer cables. 6.
Using the 1/2” combination wrench, remove the nut, washer and bolt (M) holding the transformer to the head 7.
casing.
Note
Do not over tighten the nut.
Reverse the above steps to install the replacement transformer. 8.
Test all the bed electric controls for proper operation before returning the bed to service.9.
Unless otherwise indicated, refer to Figure 4.3 C on pg. 43 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board. 3.
Using the No. 2 Phillips screwdriver, remove the 10 screws (A, Figure 4.2 A, 4. pg. 32) holding the cover and the I.V.
pole holders to the foot casing.
Lift up and remove the cover, taking care to first disconnect the control panel cable (B, Figure 4.2 A, 5. pg. 32) and, if
applicable, the foot casing connector cable ((optional), C, Figure 4.2 A, pg. 32) from the control board.
Properly ground yourself (see 6. pg. 19).
Using diagonal pliers, cut the cable ties holding together the cables.7.
Note
Carefully note the position and the grouping together of the cables before clipping the cable ties.
Remove all cables connected to the control board. Note their location so they will be connected properly to the 8.
replacement board. Refer to the connection diagram in the “Electrical Service Information” section for the connecting
position of the cables on the motor control board.
Using the 3/8” combination wrench, remove the nut (N) holding the ground cable to the frame and remove the 9.
cable.
Using needle nose pliers, pinch the upper part of the stand-off pins (O) and gradually lift the board up and out.10.
Note
Verify that the DIP switches of the replacement control board are set as the ones of the defective board. They should
be as follows:
DIP switches
no. 1: off;
no. 2: off;
no. 3: on;
no. 4: on.
Using the 3/8” combination wrench and No. 2 Phillips screwdriver, remove the nut and screw (P) holding the ground 11.
cable to the board.
Keep the jumper (W1) that is located on the electronic board before replacing the board.12.
Reverse the above steps to install the replacement control board. 13.
Test all the bed electric functions before returning the bed to service.14.
Return To Table of Contents
5072-1048E M M FL23SE REV Awww.stryker.com
Page 51
Service Information
B
D
C
A
POWER CONNECTOR AND FUSE REPLACEMENT
Tools Required:
No. 2 Phillips Screwdriver •
No. 1 Phillips Screwdriver •
Small Regular Head Screwdriver •
Procedure:
Figure 4.3 D
Note
Fuses used in this bed, powered by the 120V electrical system, are of the fast acting type and their rating is 250V, 10A.
For beds powered by other electrical system, see the “Specifications” section on pg. 8 for the fuse ratings.
For the replacement of a fuse, proceed with steps 8 to 10.
Completely raise the bed and apply the brakes. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the foot board.3.
Raise the head section,using the emeregency crank.4.
Using the No. 2 Phillips screwdriver, remove the seven screws (Q, Figure 4.3 C, 5. pg. 43) holding the cover to the
head casing. Remove all the cables (B) connected to the power connector (D). Note their location so they will be
connected correctly to the replacement power connector. Remove the cover and lay it on a work bench.
Using the No. 1 Phillips screwdriver, remove the two screws (A) holding the power connector to the frame. 6.
Using the small regular screwdriver, remove the fuse holder. 7.
Remove the fuses from the fuse holder and keep them for the replacement power connector, taking care to check 8.
that they are still in good condition, or replace them if you are only proceeding with the replacement of the fuses.
Reverse the above steps to install the replacement power connector. 9.
Turn the power switch on and check that the bed is powered before returning it to service.10.
Completely raise the bed and apply the brakes. 1.
Deactivate the Auto Contour function by pressing the Knee Gatch lockout switch at the foot end of the bed. The 2.
lockout switch LED will go off.
Completely raise the Fowler. 3.
Unplug the bed power cord from the wall socket.4.
Head Section Micro Switch
Note
This switch signals to the motor control board that the Knee Gatch must be lowered. The signal is sent during the
lowering of the Fowler when the bed is in the Auto Contour position.
Return To Table of Contents
5272-1048 E MM FL 23SE REV Awww.stryker.com
Page 53
Service Information
AUTO CONTOUR MICRO SWITCH REPLACEMENT (OPTIONAL) (CONTINUED)
Remove the cable wires from the switch. Note their location so they will be connected correctly to the replacement 5.
switch.
Using the No. 2 Phillips screwdriver, remove the two screws (A) holding the support to the frame and remove the 6.
support.
Note
The mounting position of the support may be adjusted thanks to the oblong holes present on the frame. The proper
adjustment of the support is obtained when the switch blade is pushed in by the activator, located at the base of the
head section, when the Fowler is in flat position.
Using the No. 1 Phillips screwdriver, remove the two screws (B) holding the switch to the support and remove the 7.
defective switch.
Reverse the above steps to install the replacement micro switch. 8.
Test the Auto Contour positioning for proper operation before returning the bed to service.9.
Thigh Section Micro Switch
Note
This switch signals to the motor control board that the Knee Gatch has reached the maximum elevation angle in the
Auto Contour positioning context. As a consequence, the Knee Gatch stops raising.
Remove the cable wires from the switch. Note their location so they will be connected correctly to the replacement 1.
switch.
Using the No. 2 Phillips screwdriver, remove the two screws (C) holding the support to the frame and remove the 2.
support.
Note
The mounting position of the support may be adjusted thanks to the oblong holes present on the frame. The proper
adjustment of the support is obtained when the switch blade is pushed in by the thigh section outer lever arm, located
underneath the thigh section, when the thigh section is in flat position.
Using the No. 1 Phillips screwdriver, remove the two screws (D) holding the switch to the support. 3.
Reverse the above steps to install the replacement micro switch. 4.
Test the Auto Contour positioning for proper operation before returning the bed to service.5.
This switch informs on the state of the Fowler motor (engaged or disengaged) when the CPR emergency release is
activated.
Completely raise the bed and apply the brakes.1.
Raise the head siderails.2.
Raise the head section at an angle of 45°. 3.
Unplug the bed power cord from the wall outlet. 4.
Remove the wires (A) from the micro switch. Note the location of the wires so they will be connected correctly to 5.
the replacement micro switch.
Using the No. 1 Phillips screwdriver, remove the two screws (B) holding the micro switch to the fowler section lever. 6.
Keep the mounting plate (C). Remove the defective micro switch.
Reverse the above steps to install the replacement micro switch. 7.
Test the CPR emergency release before returning the bed to service.8.
Micro Switch Attached to the Frame
Note
This switch signals to the motor control board on one hand, that the resetting of the Fowler motor may begin because the
Fowler is completely lowered, and on the other hand, that the lowering to flat of the Knee Gatch may also begin.
Completely raise the Fowler. 1.
Unplug the bed power cord from the wall outlet. 2.
Remove the cable wires (D) from the micro switch. Note the location of the wires so they will be connected correctly 3.
to the replacement micro switch.
Using the regular screwdriver, remove the micro switch from its support.4.
Reverse the above steps to install the replacement micro switch. 5.
Test the CPR emergency release before returning the bed to service.6.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A55
Page 56
Service Information
CPR ACTIVATION CABLE REPLACEMENT (OPTIONAL)
Tools Required:
7/16” Combination Wrench•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.3 F on pg. 54 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Completely raise the Fowler and remove the head board. 2.
Unplug the bed power cord from the wall outlet. 3.
Using the 7/16” combination wrench, remove the nut, sleeve and bolt (G) holding both ends of the activation cable 4.
to the fixed lever.
Using the 7/16” combination wrench, remove the nut and washer (H) holding the defective cable jacket to the mo-5.
bile lever. Remove the cable from the mobile lever and keep the spring (I). Try not to move the other cable from its
position.
Unscrew completely the two locknuts (J) holding the defective cable to the CPR handle end. Remove the cable. 6.
Reverse the above steps to install the replacement cable. 7.
Note
Make sure there is no loose space at the CPR handle once the replacement cable is installed.
Test the CPR emergency release before returning the bed to service.8.
Return To Table of Contents
5672-1048E MM FL2 3SE REV Awww.stryker.com
Page 57
Service Information
CPR PNEUMATIC CYLINDER REPLACEMENT (OPTIONAL)
Tools Required:
Two 1/2” Combination Wrenches •
11/16” Combination Wrench•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.3 F on pg. 54 for the following replacement procedure.
Completely raise the bed and apply the brakes. 1.
Completely raise the fowler section. 2.
Unplug the bed power cord from the wall outlet. 3.
Using two 1/2” combination wrenches, remove the nut (K) from the bolt holding the lower end of the pneumatic 4.
cylinder.
Partially remove the bolt until the lower end of the cylinder is free. Keep the shoulder spacer and sleeve (L). 5.
Disengage the lower end of the cylinder from the sleeve still in place.
Using the 11/16” combination wrench, loosen the locknut (M) and unscrew the cylinder completely from the end 6.
fitting (N). Note the location of the locknut before loosening it so the replacement cylinder will be properly screwed
into the end fitting.
Reverse the above steps to install the replacement cylinder. 7.
Test the CPR emergency release for proper operation before returning the bed to service.8.
Completely raise the bed. 1.
Unplug the bed power cord from the wall outlet. 2.
Position the brake/steer pedal to neutral. 3.
Using the 1/2” socket, remove the two bolts (A) holding the fifth4.
(F, Figure 4.4 B, (without fifth wheel option) pg. 62) on the side of the defective pedal and lay the support down.
Keep the molded spacer (B).
Remove the pedal from the activation lever shaft. Use a rubber mallet if necessary.5.
Reverse the above steps to install the replacement pedal.6.
Return To Table of Contents
5872-1048E M M FL23SE RE V Awww.stryker.com
wheel support (optional) or the lever support
Page 59
Service Information
FIFTH WHEEL CASTER REPLACEMENT (OPTIONAL)
Tools Required:
Two 1/2” Combination Wrenches•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.4 A on pg. 58 for the following replacement procedure.
Completely raise the bed. 1.
Unplug the bed power cord from the wall outlet. 2.
Position the brake/steer pedal to neutral. 3.
Using two 1/2” combination wrenches, remove the nut and bolt (C) holding the caster to the swing arm. 4.
Reverse the above steps to install the replacement caster.5.
Unless otherwise indicated, refer to Figure 4.4 A on pg. 58 for the following replacement procedure.
Completely raise the bed. 1.
Unplug the bed power cord from the wall outlet. 2.
Position the brake/steer pedal to neutral. 3.
Using the 1/2” combination wrench, remove the two nuts, washers, shoulder spacers and bolts (E1, Figure 4.4 B, 4.
pg. 62) linking the end of the brake rod to the locking levers, on the side of the pedals. If the optional four-wheel
brake system is present on the bed, also remove the fasteners linking the end of the two other brake rods to the
same locking levers.
Using the 1/2” socket, remove the four bolts (A) holding both fifth wheel supports to the base. Keep the molded 5.
spacers (B). Remove the assembly.
Remove the two brake/steer pedals from the activation lever shafts. Use a rubber mallet if necessary. 6.
Remove and keep the two locking levers (G, Figure 4.4 B, 7. pg. 62).
Reverse the above steps to install the replacement fifth wheel assembly. 8.
Test the fifth wheel for proper operation before returning the bed to service.9.
Return To Table of Contents
6072-1048E MM FL23SE RE V Awww.stryker.com
Page 61
Service Information
FIFTH WHEEL SWING ARM ASSEMBLY REPLACEMENT (OPTIONAL)
Tools Required:
3/8” Drive Ratchet•
1/2” Socket •
Two 1/2” Combination Wrenches•
3/16” Allen Wrench•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.4 A on pg. 58 for the following replacement procedure.
Completely raise the bed. 1.
Unplug the bed power cord from the wall outlet. 2.
Position the brake/steer pedal to neutral. 3.
Using the 1/2” socket, remove the two bolts (A) holding the left or right fifth wheel support to the base. Lay the 4.
support down and keep the molded spacer (B).
Using two 1/2” combination wrenches, remove the nut, shoulder spacer and bolt (D) holding the torsion lever (E) to 5.
the lower part of the counter lever.
Using the 1/2” combination wrench and 3/16” Allen wrench, remove the two nuts and Allen screws (F) holding the 6.
torsion levers (E) to both ends of the torque shaft (G).
Note
At reassembly, screw down the Allen screws before installing the nuts.
Disengage the swing arm assembly from the torsion levers and remove it. 7.
Using two 1/2” combination wrenches, remove the nut and bolt (C) holding the caster to the swing arm and keep it 8.
for the replacement swing arm assembly.
Reverse the above steps to install the replacement swing arm assembly. 9.
Test the fifth wheel for proper operation before returning the bed to service.10.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A61
Page 62
BED CASTER REPLACEMENT
F
A
A
G
B
C
C
D
D
E1
E2
E1
E2
B
Tools Required:
Jack Stand •
Needle Nose Pliers •
1/2” Combination Wrench •
Procedure:
Service Information
WITH OPTIONAL FIFTH WHEEL
WITHOUT OPTIONAL FIFTH WHEEL
Figure 4.4 B
Completely raise the bed. 1.
Position the brake/steer to neutral.2.
Install a jack stand under the frame near the defective caster. 3.
If present, remove the wheel cover (optional). 4.
Lower the bed until the caster is six inches off the ground. 5.
Unplug the bed power cord from the wall outlet. 6.
Remove, using needle nose pliers, the rue ring, washer and locking axle (A) linking the caster shaft to the locking 7.
lever (B) and the brake rod (C) if the caster is part of the braking system.
Return To Table of Contents
6272-1048E MM FL23SE REV Awww.stryker.com
Page 63
Service Information
BED CASTER REPLACEMENT (CONTINUED)
Note
The shaft of a caster that is part of the braking mechanism must be properly
oriented before mounting it to the base. The opposite illustration shows the
side of the shaft that should be facing the inside of the bed: it is the side
showing the largest opening below the cam in neutral position.
Using the 1/2” combination wrench, remove the bolt and washers (D) 8.
holding the caster to the base.
Note
Before completely tightening the bolt (D), make sure that the caster shaft is completely inserted into the mounting
socket.
Reverse the above steps to install the replacement caster. 9.
If the replacement caster is part of the brake mechanism, apply the brakes and check that the caster is locked 10.
before returning the bed to service.
Return To Table of Contents
www.stryker.com72-1048E MM FL2 3SE REV A63
Page 64
Service Information
BRAKE ROD REPLACEMENT
Tools Required:
1/2” Combination Wrench•
Procedure:
Note
Unless otherwise indicated, refer to Figure 4.4 B on pg. 62 for the following replacement procedure.
Completely raise the bed.1.
Using a 1/2” combination wrench, remove the nut, shoulder spacer, washer and bolt (E1, E2) holding each end of 2.
the linking rod.
Remove the defective linking rod.3.
Reverse the above steps to install the replacement linking rod.4.
Check the braking mechanism before returning the bed to service.5.
Return To Table of Contents
6472-1048E MM FL23SE REV Awww.stryker.com
Page 65
For Reference Only: L23-006
1
3
2
Rev. 01
Base
Base (For Reference Only: L23-006G, Rev. 01)
Item Part No. Part Name Qty.
1 25-0108G Base S.A. 1
2 28-0469 Ground Chain 1
3 VV83A9G16 Tapping Screw Pan Type B
Phillips #10 x 1/2” z.p. 1