Stryker Crossfire Footswitch User Manual

Crossfire™ Footswitch
0475000100
User Guide
EN-1
Warnings and Cautions .................................................... 3
Product Description and Intended Use .................. 4
System Overview ....................................................................4
Indications/Contraindications .................................................6
Package Contents ................................................................... 6
Compatible Accessories .........................................................6
Setup ........................................................................................... 7
Operation .................................................................................. 8
Cleaning and Maintenance .......................................... 11
Cleaning ................................................................................11
Maintenance .......................................................................... 11
Technical Specifications ............................................... 12
Environmental Specications ................................................12
Symbols and Definitions ................................................ 13
EN-3
Warnings and Cautions
Please read this manual and follow its instructions carefully. e words warning, caution, and note carry special meanings and should be carefully reviewed:
Warning Warnings indicate risks to the safety of the patient or user.
Failure to follow warnings may result in injury to the patient or user.
Caution Cautions indicate risks to the equipment. Failure to follow
cautions may result in product damage.
Note Notes provide special information to clarify instructions or
present additional useful information.
To avoid potential serious injury to the user and the patient and/or damage to this device, the user must obey the following warnings. e warranty is void if any of these warnings is disregarded.
1. Read this operating manual thoroughly and be familiar with its contents prior to setting up or using this equipment.
2. Carefully unpack the unit and check for any damage that may have occurred during shipment. If damage is detected, refer to the Service and Claims section in this manual.
3. Use of this device is restricted to qualied health care professionals who have complete knowledge of the use of this equipment and the devices to which it can be connected.
4. DO NOT use the Crossre™ system with non-conductive irrigants (e.g. sterile water, air, gas, glycine, etc.). Use only conductive irrigants, such as saline or Ringer’s lactate, in order for the system to function properly.
5. Test this equipment prior to using it in a surgical procedure. is unit was fully tested at the factory before shipment; however, it must be tested for proper function in the environment and setup conguration in which it will be used.
6. Attempt no internal repairs or adjustments not specically detailed in this operating manual. Refer any readjustments, modications, and/or repairs to Stryker Endoscopy or its authorized representatives.
7. Pay close attention to the care and cleaning instructions in this manual. Failure to follow these instructions may result in product damage.
8. Install this device in an operating room that complies with all applicable IEC, CEC, and NEC requirements for safety of electrical devices.
EN-4
Product Description and Intended Use
e Crossre™ Footswitch is a foot-operated remote control used in conjuction with the Crossre Integrated Arthroscopy system. It enables the user to activate and control the powered shaver handpiece and radio frequency (RF) electrosurgical probe.
System Overview
e footswitch is part of the Crossre™ Integrated Arthroscopy System, which is a combination powered shaver system/electrosurgical generator, intended for use in arthroscopic and orthopedic procedures.
Illustrated below, the Crossre™ system consists of the following components: