DFC DIGITAL FORCE CONTROLLER
QUICK REFERENCE GUIDE
PKG08812-QRGDFC 3
Page 4
TABLE OF CONTENTS
PAGE
Introduction 6
Keypad and Navigation 8
On/Off/Menu Key 8
Zero Key 8
Softkey 1 8
Softkey 2 8
Controller Display Formats 11
Using Tolerance Limits 12
Saving Results 12
Clear Results from Memory 13
View Statistics 13
Export Results 13
Gage Setup 13
Charging Battery 14
4 PKG08812-QRGDFC
Page 5
DFC DIGITAL FORCE CONTROLLER
THIS IS A STARRETT QUICK REFERENCE GUIDE
FOR THE DFG DIGITAL FORCE CONTROLLER.
ALL SPECIFICATIONS IN THIS DOCUMENT ARE CORRECT AT TIME OF PRODUCTION AND ARE SUBJECT TO CHANGE. PLEASE CONTACT
STARRETT FOR FURTHER INFORMATION.
PKG08812-QRGDFC 5
Page 6
ENGLISH
Thank you for choosing the DFC Force Controller. This Quick Reference Guide is an overview of the basic
functions available with your instrument. For detailed instructions on instrument setup and operation, please refer to the electronic user manual. You may download a copy of the user manual at http://www.Starrett.com/u?dfc-um.
INTRODUCTION
The DFC may be used as a classic digital force gage. It may also be used as a digital controller when connected to the FMM Digital Force Tester.
The DFC, as a controller, is the only gage available that lets you setup an automated test, including the operating parameters of the force tester. Configure speed and distance limits through the DFC.
STARRETT DFC ADVANCED FORCE GAGE
SPECIFICATION DFC
Accuracy, Full Scale 0.1%
Data Sampling (Hz) 25,000
Display Resolution 10,000:1
Safe Overload, Full Scale 200%
Maximum Tare 10%
COMMUNICATIONS
®
Bluetooth
USB 2.0 Yes
RS-232 Yes
Digital I/O 2 channels
Memory, saved results max. 99
OPERATING MODE
Machine Control
Real Time Yes
Peak Compression Yes
Peak Tension Yes
Load Limit Yes
Break Limit Yes
Load Average Yes
Load-Time Average Yes
Cyclic Count, Max. 99,999
Cyclic Duration, Max. 27 Hrs.
Hold Duration, Max. 27 Hrs.
Coefficient of Friction Yes
POWER, ENVIRONMENTAL
Battery Type Lithium Ion
Battery Life, @ 20% brightness >30 hours
Charge Time, using 110/240V Mains <3 hours
Display (OLED) 320 x 240
Operating Temperature
Thread, for adapters (Metric) M6, M10
Instrument Weight (approx.)
CE COMPLIANCE
EN61010-1 Safety for Electrical Equipment Yes
EN61000-6-3 EMC Generic Emissions Yes
EN61000-6-1 EMC Generic Immunity Yes
NOTES
1. Machine control is exclusive to the DFC. When connected to the FMM digital force tester, configuration of force gage and tester is performed through the gage.
1
Yes
Yes
40°F to 110°F
(4°C to 43°C)
3 lbs
(1.36 kgs)
6 PKG08812-QRGDFC
Page 7
ENGLISH
0.91 (23.0)
7.50 (190.5)
DFC - ADVANCED FORCE CONTROLLER
MODEL
DFC-2 10 1 2 32 900 0.013 0.33 M6 x 1-6H SPK-FG-A
DFC-5 25 2 5 80 2,200 0.007 0.18 M6 x 1-6H SPK-FG-A
DFC-10 50 5 10 160 5,000 0.006 0.15 M6 x 1-6H SPK-FG-S
DFC-20 100 10 20 320 9,000 0.008 0.20 M6 x 1-6H SPK-FG-S
DFC-50 250 25 50 800 23,000 0.015 0.39 M6 x 1-6H SPK-FG-S
DFC-100 500 50 110 1,600 45,000 0.024 0.60 M6 x 1-6H SPK-FG-S
DFC-200 1,000 100 225 - - 0.021 0.54 M6 x 1-6H SPK-FG-M
DFC-500 2,500 250 550 - - 0.028 0.70 M10 x 1.5-5H SPK-FG-L
NOTES
Load measurement accuracy is ±0.1% of load cell capacity. Display resolution is 10,000:1.
3.30 (83.8)
Thread
0.31 (8.0)
2.25 (57.2)
LOAD CAPACITY FULL SCALE DEFLECTION THREAD ACCESSORY
N KGF LBF OZF GF IN MM MM KIT
1.63 (41.3)
0.50 (12.7)
A
A
1.70 (43.2)
2.00 (50.8)
0.55 (14.0)
2.00 (50.8)
M4x0.7-6H 0.30
Ø0.20 0.18
PKG08812-QRGDFC 7
Page 8
ENGLISH
KEYPAD AND NAVIGATION
The DFC keypad is multi-functional. There are four keys.
ON/OFF/MENU KEY
This key is used to power the gage on/off. Press to power the gage on. Press and hold for 3 seconds to power the gage off.
This key is also used to access the Setup Menu. When the gage is powered ON, press to access the Setup Menu.
This key is also used to navigate UP when in the Setup Menu. p
ZERO KEY
This key is used to zero the displayed values. Press to zero load and if connected to the FMM digital force tester, zero distance.
This key is also used to navigate DOWN when in the Setup Menu. q
SOFTKEY 1
This is the left arrow key. It is used to move out of a setup when in the Setup Menu. t
This key may also be mapped to a specific function. The Setup Menu has a Key setup function where you
may assign how Softkey 1 performs. For example, you can assign a SAVE function to the key. When pressed, the measured values are "saved" to memory.
SOFTKEY 2
This is the right arrow key. It is used to move in to a setup when in the Setup Menu. u
Like the Softkey 1, it also may be mapped to a specific function. For example, you can assign a MODE function to the key. Pressing the key can change the operating modes of the gage. You can switch from Real Time to Peak Tension by pressing Softkey 2.
DISPLAY LAYOUT
The force gage features a high-resolution OLED color display with adjustable backlight. The backlight may be adjusted from a setting in the Main Menu.
The DFC features a primary and secondary display window.
DISPLAY TYPES
PRIMARY DISPLAY SECONDARY DISPLAY
Real Time Peak Measurement
Peak Tension Real Time Measurement
Peak Compression Real Time Measurement
Average
Average (Functional Capacity) Peak Measurement Time Duration
Static COF Kinetic COF
Kinetic COF Static COF
Peak Measurement Preload to Start Average
8 PKG08812-QRGDFC
Page 9
ENGLISH
lbf
7
4
7
lbf
lbf
7
Real time load being placed on load cell
Compression Direction
Shown: Real Time view showing Peak Compression
Peak/Max Compression
0.00
16.89
16.84
Real Time Mode Indicator
Real Time value
lbf
5.66
5.71
Peak Tension Mode Indicator
Shown: Peak Tension view with Tolerance result
Peak/Max Tension
22.29
Tensile Direction
Shown: Peak Tension view
Peak/Max Compression
Compression Direction
21.10
Peak Compression Mode Indicator
Shown: Peak Compression view
Real Time value
19.24
Peak Tension Mode Indicator
Real Time value
21.06
PKG08812-QRGDFC 9
Page 10
ENGLISH
Battery Status Shows remaining battery life.
Primary Result
Shows load direction using a symbol for Tension or Compression.
Shows functions mapped to softkeys 1 and 2. Menu function is always center key.
Softkey 1 Associated with above Function Use to navigate LEFT and move out of setup option.
Zero key Use to zero the display User to tare the force gage Use to navigate DOWN
Softkey 2 Associated with above Function Use to navigate LEFT and move out of setup option.
Shows functions mapped to softkeys 1 and 2. Menu function is always center key.
Shows Load measured by load cell Shows Distance measured by encoder on the FMM Test Frame.
Unit of Measure Shows unit of measure for displayed results.
Shown: Display in Classic mode
Secondary Display
Load Bargraph Shows active load on sensor. Shows direction of load.
Softkey 2 Associated with above Function Use to navigate RIGHT and INTO setup option
On/Off/Menu key Use to power gage ON/OFF Use to access the SETUP MENU Use to navigate UP
Softkey 1 Associated with above Function Use to navigate RIGHT and INTO setup option
Shows setpoint target that completes the test method.
Shows Load unit of measure. Shows Distance unit of measure.
Load Bargraph Shows active load on sensor Shows direction of load
10 PKG08812-QRGDFC
Shown: Display in Controller mode used with FMM Test Frame
Page 11
ENGLISH
9
5
5
5
Average Load Average Load
lbf
3.65
N
7.15
4
Shown: Functional Capacity result
Start Load
8.71
1.00
Average Mode IndicatorShown: Average view
Countdown TimerAverage Load
7.15
8.1
Average Mode Indicator
Kinetic COF
Static COF
33.45
CONTROLLER DISPLAY FORMATS
As a Controller, the DFC will display both load and distance information and results. The gage will serve is the universal interface for the "system" and will be used to setup the force gage and the tester.
The DFC force gage will display load information, as measured by the load cell.
The DFC force gage will display distance information, as measured by the FMM digital force tester.
Start/Stop Test SoftkeyLoad Limit Target
100.00N
Active Load
L D
Shown: Load Limit Test results
Active Load
L8.80 D
16.55
16.64
Distance Limit Target Start/Stop Test Softkey
8.78
N
mm
50.00mm
N
mm
Active Distance
Active Distance
Shown: COF result
µ µ
k
s
33.450
17.055
5
Average Mode Indicator
Shown: Distance Limit Test results
Break Limit Target
Active Load
L 27.45 D
Shown: Break Limit Test results
28.28
PKG08812-QRGDFC 11
Start/Stop Test Softkey
50%
N
Active Distance
mm
Page 12
ENGLISH
4
USING TOLERANCE LIMITS
Use tolerances to setup "pass" and "fail" measurements. You may specify a Limit 1 and a Limit 2 to create a tolerance band. Measured results that equal or fall within the range created by the two limits are considered "pass" results. If the measured result falls outside the band created by the two Limits, the result is considered a "fail" result. A "fail" results displays in RED.
PressqMove to TOLERANCE PressuSelect ENABLE
q
Press Press Press PresstMove out of Limit 1 Press Press PresstMove out of LIMIT 2 PresstMove out of TOLERANCE
Move down to LIMITS
q
Move down to LIMIT 1
u
Select Limit 1 value using qp
q
Move down to LIMIT 1
u
Select Limit 2 value using qp
Real Time value
lbf
5.66
5.71
Peak Tension Mode Indicator
Shown: Peak Tension view with Tolerance result
SAVING RESULTS
The DFC feature an internal memory for saving results for the purpose of calculating and displaying statistics. You may save up to 99 individual results in memory.
To save results to memory, you must configure a softkey with the SAVE function. Press the SAVE softkey to save the displayed results to memory.
Results saved to memory MUST be of the same type. You cannot have mixed type results. For example, you cannot save tension results with COF results. Results MUST be of the same type.
Results obtained using the TESTS feature are saved in pairs. For each test, there is a Load and Distance result (L-D). You cannot save results from different types of tests. For example, you cannot save results from a Load Limit test with results from a Break Limit test. Each load-distance pair of results must have come from the same test type: Load, Distance or Break.
RESULT TYPE MEMORY TYPE
MODES
Real Time Real Time
Tension Peak Tension Peak
Compression Peak Compression Peak
Average Average
Static COF Static COF
Kinetic COF Kinetic COF
TESTS
LOAD LIMIT TEST
Distance @ Limit Distance @ Limit
Load Limit Load Limit
DISTANCE LIMIT TEST
Load @ Limit Load @ Limit
Distance Limit Distance Limit
BREAK LIMIT TEST
Load Max (Peak Load) Load Max (Peak Load)
Distance @ Load Max Distance @ Load Max
12 PKG08812-QRGDFC
Page 13
ENGLISH
Management of the DFC memory is important to ensure correct statistical analysis. Always clear old results that are from a different test method.
CLEAR RESULTS FROM MEMORY
Results in memory may be cleared individually or collectively. To clear an individual result from memory, go to the STATS view. Select the results you want to clear. Select the "X" (delete) key.
To clear all results from memory, go to the Memory setup and select CLEAR. This will erase all results for the gage's memory.
VIEW STATISTICS
You must configure a softkey with STATS to view statistics. When results are saved to memory, you press the STATS softkey to view results in memory and the statistics for those results.
EXPORT RESULTS
You must configure a softkey with SEND to export to an external device via Bluetooth, RS232 or USB. Press the SEND softkey to export the displayed result.
GAGE SETUP
MODES
p u
for Menu for Modes
t to exit setup
Press Press Select your Mode, press
UNITS
p q
for Menu for Units
t to exit setup
Press Press Select your Units, press
MEMORY
p
Press
for Menu
Pressqfor Memory
u
Press Press Press Select your Units, press
*You must configure a softkey to display memory and statistics.
to Display results from memory
u
to Clear results from memory
u
to Export results from memory
t to exit setup
TESTS
p
Press Press Press Press Press
*Refer to the User Manual for detailed instructions on using the TESTS feature in your DFC.
for Menu
q
for Tests
u
to Enable Tests
q
to your Test Method
u
to enter specific Test Method options
TOLERANCE
p
Press Press Press
for Menu
q
for Tolerance
u
to Enable Tolerance Pressuto setup Limit 1, select Limit 1 value pq Pressuto setup Limit 2, select Limit 2 value pq Presstto exit setup
KEYS
p
Press Press Press
for Menu
q
for Keys
q
to specify a Sound for key presses Pressqto specify Softkey 1 function
q
Press Press
*Certain gage functions require a softkey to be setup in order to use the function- Memory, Statistics, Tests, Export (Send)
to specify Softkey 2 function
t
to exit setup
SETTINGS
p
Press Press
for Menu
q
for Settings, select Settings Type
COMMUNICATIONS
Pressqfor Communications (Comms)
Press
Press Press Press Press Press
results: USB, RS-232, or BT (Bluetooth®)
q
to select RS-232
q
to select RS-232 Baud Rate
q
to setup direction sign (- default for Compression)
q
to transmit Units
q
to transmit Tolerance Limits
to specify the Data Channel method for exporting
u
Presstto exit Comms setup
DISPLAY
q
Press Press Press
Press
Press Press
for Display
u
to setup Auto Off
u
to setup Backlight Brightness to setup Flip orientation (required for
u
TESTS feature)
u
to setup Radix
t
to exit Display setup
FILTERS
q
Press Press Press
for Filters
u
to specify filter rate
t
to exit Filter setup
ABOUT
q
Press
Press
Press
*The About setup is read-only.
for About to view the gage’s characteristics, serial
u
number, overload history and more.
t
to exit setup
PKG08812-QRGDFC 13
Page 14
ENGLISH
MEMORY
q
Press
u
Press
pq
Press
t
Press
LANGUAGE
Press Press Press
for Password to Enable a Password to select the Passcode to exit setup
q
for Language
q
to your preferred display language
t
to exit setup
CHARGING BATTERY
Connect the USB cable supplied with your DFC to the USB connector at the base of the gage. Connect the opposite end of the cable to the USB connector on a source device, i.e., personal computer, USB adapter.
If the DFC is connected to an FMM test frame and power to the frame is ON, the DFC battery is continuously charged.
Communicating with External Devices The DFC can communicate using Bluetooth, USB 2.0
and RS-232. Select the Data Channel type to be used.
PROTOCOL WHERE TO USE
Bluetooth
USB 2.0
RS-232
RS-232USB 2.0
Gage Setup Menu Structure
MODES UNITS MEMORY TESTS TOLERANCE KEYS SETTINGS
Real Time ozf Display Enable Enable Enable Comms Peak Tension lbf Clear Load Limit Limit 1 Softkey 1 Data Channel Peak Compression gf Export Type Limit 2 Send RS-232 Averaging kgf Target Sound Units Xmit Comp ­ Time Avg N Speed Save Xmit Units COF User Distance Limit Mode Xmit TOL
Type Stats Display Target Start/Stop Auto Off Speed Return 0 Backlight Break Limit Softkey 2 Flip Type Send Radix Minimum Break Units Filter % Drop Save About Auto 0 Mode Password Auto Return Stats Language Auto Save Start/Stop English
Control
Print via Bluetooth printer
®
Export data to an external device Charge DFC Battery Upload firmware and new features to DFC from a PC Export data to a printer Export data to a PC Communicate with a serial printer Communicate with a serial external computer or hard drive Controls the crosshead direction and velocity the FMM from the DFC gage
Return 0 Deutsch
Espanol Portugues Francais Italiano Chinese Russian Polski Czech
14 PKG08812-QRGDFC
Page 15
NORTH AMERICA
WWW.STARRETT.COM
ATHOL, MA, USA, 01331-1915
EUROPE & ASIA
WWW.STARRETT.CO.UK
JEDBURGH, SCOTLAND, TD8 6LR
SOUTH & CENTRAL AMERICA
WWW.STARRETT.COM.BR
13306-900, ITU, SP, BRASIL
CNPJ 56.994.700/0001-01
CHINA
WWW.STARRETT.COM.CN
USER MANUALS AVAILABLE ONLINE
AT STARRETT.COM
PKG08812-QRGDFC 15
Page 16
Starrett.com
Loading...