Replacing the patient tabletop
SP
© Siemens AG 1996
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
English Print No.: RLL5-310.076.06.01.02 Release Date:06.96
Replaces: n.a.
Page 2 of 8 This page intentionally left blank.
Revision
Letzte Seite des Kapitel
UROSKOP D RLL5-310.076.06 TD SP 1 Siemens AG Rev. 01 06.96 Medical Engineering
CAUTION
Replacing the patient tabletop 1

Safety information 1

When performing service work and tests, observe the product-
specific safety information contai ned in this document as well as the general safety information in Register 2 of the TI folder / ARTD part 2.
Observe all C-MOS guidelines.

Tolerance specifications 1

General tolerances for the lengths of screws according to ISO 2768
Page 3 of 8
Ranges of
length
permitted tolerance
These tolerances apply to all lengths spec ified in t his document, provided a d ifferent value is not expressly indicated.
A tolerance of ±10% is permitted for the torques indicated.
from
3 mm
to
6 mm
± 0.5 mm ± 1 mm ± 1.5 mm ± 2.5 mm ± 4 mm ± 6 mm ± 8 mm
from
6 mm
to
30 mm
from
30 mm
to
120 mm
from
120 mm
to
400 mm
from
400 mm
to
1000 mm
from
1000 mm
to
2000 mm

Replacing damaged or lost screws 1

Damaged or lost screws may be replaced only with steel screws as indicated in the draw­ings according to DIN 267 and having the tensile strengt h specified.
All Allen screws used must have a tensile strength of 8.8, unless otherwise s peci­fied in this doc u me n t.
from
2000 mm
to
4000 mm
Siemens AG TD SP 1 RLL5-310.076.06 UROSKOP D Medical Engineering Rev. 01 06.96