The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design are
reserved.
English
Print No.: SPL1-130.840.02.03.02 Doc. Gen. Date:04.05
Replaces: SPL1-130.840.02.02.02
0 - 2Revision Level
ChapterPageRev.
allall03
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
MODULARIS Uro PlusSPL1-130.840.02Page 2 of 4Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
9 _______Changes to the previous version__________________________________ 9 - 1
Siemens AGSPL1-130.840.02Page 5 of 6MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
0 - 6Table of Contents
Page
MODULARIS Uro PlusSPL1-130.840.02Page 6 of 6Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
General1
Safety information and protective measures1
• Adhere to the safety measures described
1 - 1
WAR NING
Dangerous X-rax radiation and/or dangerous electrical voltage
during test and adjustment operations!
Disregarding safety precautions can result in death or serious
bodily injury.
The following points must be observed:
• Disconnect the power cable when working on the system.
• Ensure compliance with the general safety requirements when
working with the system under power.
• Switch off the power prior to replacing modules or boards.
• After all work has been completed and all covers have been
installed, perform the protective conductor test according to
ARTD-002.731.17...
• The protective conductor resistance must not exceed 0.2 ohms.
• When performing service work on the power-on module,
(replacing the power-on module or replacing the power cable),
the equivalent leakage current must be measured and recorded.
Checks and settings that need to be performed under X-ray radiation are identified by the radiation warning symbol .
When performing adjustments labeled as shown here, appropriate radiation protection measures must be taken (see ARTD,
Part 2).
Scope of applicability and regulations for the subsidiaries1
Equivalent leakage current measurement
The equivalent leakage current must be measured where applicable under the requirements of DIN VDE 0751, Part 1.
Where DIN VDE 0751 does not apply, the subsidiaries should comply with the following
regulations (refer to ARTD - 002.731.17, as well as the safety technical regulations for
installation and maintenance).
The local national regulations apply primarily for the subsidiaries.
In the event that there are no existing local regulations, the following provisions should be
adhered to in the interest of the safety of customers, patients, employees and third parties
as well as the company.
Siemens AGSPL1-130.840.02Page 1 of 2MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
1 - 2General
Initial measured value
The equivalent leakage current measurement was performed at the factory and the value
measured was entered in test protocol 1. The measurement was made at the line voltage
and line frequency indicated in test protocol 1.
If the line voltage or the line frequency on-site deviates from the information indicated
upon delivery of the Modularis Uro Plus, the values given are invalid. The values should
be marked invalid (crossed out with the comment "invalid values" and the service engineer should sign and date this copy).
In this case, the equivalent leakage current must be measured again. The value may not
exceed 1 mA according to DIN VDE 0751, Part 1.
The initial value measured must be documented.
Repeat measurements
When service or repair work is performed on the primary power supply circuit (e.g. repairs
to the power-on circuit), the equivalent leakage current test must be repeated. The values
measured in the subsequent test may not exceed the threshold value of 1 mA as specified
in VDE 0751, Part 1. In addition, they may not exceed the initial measured value by more
than 50%. If the value exceeds this threshold, the system must be repaired. The value
measured must be documented.
Tools and measurement devices required1
NOTE
• Standard service equipment
• Service PC (see Intranet Service Laptop for CSE’s
• PC connection cable, 5 m99 00 440 RE999
• ESD equipment
• Protective conductor meter44 15 899 RV090
• Digital multimeter, e.g. FLUKE 18799 94 831
• Shock wave pressure test unit30 95 408 J1008
• Adapter for C-head98 17 347 J1008
The specified articles are listed in the SCT (Catalog Service Tools)
if nothing else is indicated (the STC is a component of the Spare
Parts Catalog).
Auxiliary devices and documents required1
• SIREMOBIL Iso-C Service InstructionsSPR2-230.061.01..
• Service Instructions Shock Wave Pressure ControlSPL2-120.074.01..
MODULARIS Uro PlusSPL1-130.840.02Page 2 of 2Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
LITHOSTAR MODULARIS Error List2
2 - 1
NOTE
Error messages from the SIREMOBIL Iso-C have not been listed.
(Refer to the SIREMOBIL Iso-C service documents)
Info, E00 and E992
E00Error Text:System was switched on.
Possible
cause/action:
E99Error Text:The system was switched off without deleting the therapy data.
Possible
cause/action:
Press the Reset button.
Water system, E10 – E122
LITHOSTAR MODULARIS circuit diagram J1048, sheets 13 and 14.
E10Error Text:Cooling system losing water.
Possible
cause/action:
E11Error Text:Maximum pressure in the cooling circuit exceeded.
Possible
cause/action:
E12Error Text:Minimum pressure in the cooling circuit exceeded.
Possible
cause/action:
Is water running out of the hoses?
-switch o.k.?
P
min
Board D3.X3 connection o.k.?
Fill the cooling circuit.
Pressure switch P
Are the hoses blocked, kinked?
P
-switch o.k.?
max
Board D3.X3 connection o.k?
Is the cooling unit operating?
Check the fill level in the cooling circuit.
Pressure switch P
Is water running out of the hoses and hose connections?
Air in the cooling system?
responded.
max
responded.
min
Check the fill level in the cooling circuit.
Siemens AGSPL1-130.840.02Page 1 of 14MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
2 - 2LITHOSTAR MODULARIS Error List
Water system, E14 – E152
E14Error Text:Water level too low in the water reserve tank.
Possible
cause/action:
E15Error Text:Coupling pressure not o.k.
Possible
cause/action:
Not enough water in the tank.
Float switch in the tank o.k.?
Board D3.X3 connection o.k.?
Completely fill the coupling bellows, deselect automatic pres-
sure regulation, and drain the water.
Repair any leaks.
Refill the water in the reserve tank.
Pressure sensor on the shock head o.k.?
If water can still be pumped out, the sensor is probably defec-
tive (refer to "Working with the terminal program: show diagnostic data").
Suction hose correctly inserted in the coupling bellows?
Return hose kinked?
Board D3.X8 connection o.k.?
Fill the coupling circuit.
If the error only occurs with obese patients, the suction hose in
the coupling bellows may be kinked.
Controls, E20 – E212
If the control panel is dark: check fuse F5 on board D3.
E20Error Text:No contact with the control unit (this error is visible only when
reading out the error log with the service PC).
Possible
cause/action:
E21Error Text:X-ray button on the control unit activated with switch on.
Possible
cause/action:
Connection cable X5 correctly connected to the control unit?
Check fuse F5 on board D3.
If the error occurs after replacing board D3 - is plug X35 pro-
perly connected?
X-ray button stuck?
Control unit o.k.?
Cable o.k.?
MODULARIS Uro PlusSPL1-130.840.02Page 2 of 14Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
LITHOSTAR MODULARIS Error List2 - 3
Controls, E22 – E252
E22Error Text:QRS signal from ECG unit missing, i.e. no ECG triggering pos-
sible.
Possible
cause/action:
E23Error Text:Motor for C-arm movement in LITHOSTAR MODULARIS
Possible
cause/action:
Trigger cable connected to ECG unit?
ECG triggering selected on the control unit?
Trigger cable plugged in on D3.X70?
draws too much current.
C-arm moving freely?
SIREMOBIL Iso-C cable connected?
Brake in SIREMOBIL Iso-C can be released, i.e. the motor is
not locked?
Brake connection cable in SM Iso-C o.k.?
(D3.X38 - SM Iso-C)
Motor connection cable o.k. (board D3.X36)?
Board D3?
Motor connection cable o.k. (board D3.X36)?
Settings for the SIREMOBIL Iso brakes o.k.?
(Angulation and orbital in 0°. LITHOSTAR MODULARIS not
coupled. Place spring scale at the handle near the I.I. Release
angulation brake, pull with spring scale, the C-arm must move
when the spring balance indicates a value from 20N, 30N, the
closer to 20 N the better. Settings are made at the position
where the angulation motor is coupled. Wrench size 30 mm,
coupling must be removed).
Refer also to SPR2-230.061.01... Chapter 7.
Board D3?
E24Error Text:C-arm blocked.
No change on the actual value potentiometer in LITHOSTAR
MODULARIS.
Possible
cause/action:
E25Error Text:C-arm potentiometer not working.
Possible
cause/action:
Siemens AGSPL1-130.840.02Page 3 of 14MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
C-arm moving freely?
Potentiometer o.k.?
Correct tension on the belt?
Connection cable to potentiometer o.k.? (D3.X38)
Check the setting of the SIREMOBIL angulation brake (see
E23)
Potentiometer connected?
Check plug D3.X30!
Connect the cable to the potentiometer.
2 - 4LITHOSTAR MODULARIS Error List
Controls, E26 – E292
E26Error Text:ECG option not installed.
Possible
cause/action:
E27Error Text:Control unit.
Possible
cause/action:
E28Error Text:C-arm swivels in the wrong direction.
Possible
cause/action:
E29Error Text:Shock wave release button sticks.
Possible
cause/action:
Warning message only.
Key sticks.
Replace control panel.
Cables connected wrong at the potentiometer or the motor?
Correct tension on the belt?
Check the setting of the SIREMOBIL angulation brake
(see E23).
Check control unit. Circuit diagram J1048 – 25.
MODULARIS Uro PlusSPL1-130.840.02Page 4 of 14Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
MODULARIS Uro PlusSPL1-130.840.02Page 6 of 14Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
LITHOSTAR MODULARIS Error List2 - 7
Monitoring, E49 – E542
E49Error Text:Reference potentiometer defective.
Possible
cause/action:
E50Error Text:-5 volt power supply defective.
Possible
cause/action:
E52Error Text:Battery voltage for battery-buffered RAM and real-time clock
Possible
cause/action:Replace the lithium battery on board D3.
E53Error Text:Incorrect data in the battery-buffered RAM.
Possible
cause/action:
Short circuit in potentiometer for C-arm movement or potentiometer for ultrasound application (only if option is available)?
Connection cable D3.X38, D3.X28 (ultrasound option) o.k.?
Circuit diagram J1048 – 16/17.
Use the Service PC to read the current potentiometer
(menu "d").
For troubleshooting: the ultrasound option is available; remove
the board for the ultrasound option and cables X28 and X29
from board D3.
Replace board D3.
too low.
Normal occurrence at initial start-up or after replacing the
lithium battery.
Check the lithium battery!
Replace board D3.
Perform a software download!
E54Error Text:Clock stopped (this message is also displayed when the battery
is replaced)
Possible
cause/action:
Connection cable D3.X31 - XP7 present under the cover?
Set the clock!
Siemens AGSPL1-130.840.02Page 7 of 14MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
2 - 8LITHOSTAR MODULARIS Error List
High voltage circuit, E60 – E672
LITHOSTAR MODULARIS Circuit diagram J1048 - 12.
E60Error Text:Charging current too low during the charging process.
Possible
cause/action:
Check the line voltage!
Replace the charging / energy unit!
E61Error Text:Overvoltage in the charging unit.
Possible
cause/action:
System OFF
- 10 minute pause System ON.
Replace the charging / energy unit!
E62Error Text:Overtemperature in the charging unit.
Possible
cause/action:
Heat in the LITHO MODULARIS ?
System OFF
- 10 minute pause System ON.
Replace the charging / energy unit!
E63Error Text:Capacitor defective in the charging / energy unit (out of tole-
rance).
Possible
Replace the charging / energy unit!
cause/action:
E64Error Text:Default voltage not attained.
Current voltage after frequency converter (U
) is greater than
nom
selected voltage.
1
Possible
cause/action:
Check the fiber optic cable!
Replace the charging / energy unit!
)
E65Error Text:Charging voltage too high during the charging process.
Possible
Replace the charging / energy unit!
cause/action:
E66Error Text:Default voltage incorrect.
1
Possible
cause/action:
Check the fiber optic cable!
Replace the charging / energy unit!
)
E67Error Text:No ignition of spark gap after trigger pulse (several times).
Possible
cause/action:
Check the fiber optic cable!
Replace the charging/ energy unit!
1
)
MODULARIS Uro PlusSPL1-130.840.02Page 8 of 14Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
LITHOSTAR MODULARIS Error List2 - 9
High voltage circuit, E68 – E74 2
E68Error Text:Charging current too high during charging process.
Possible
cause/action:
E69Error Text:Repeated self-ignition of the spark gap.
Possible
cause/action:
E70Error Text:Charging current too low after ignition!
Possible
cause/action:
E71Error Text:Charging current too high after ignition.
Possible
cause/action:
E72Error Text:No communication possible with the charging / energy unit.
Possible
cause/action:
Replace the charging / energy unit!
Replace the charging/ energy unit!
Check the connections of the high voltage cable!
Check the high voltage connector!
Replace the shock wave head!
Check the high voltage cable!
Check the high voltage connector!
Replace the shock wave head!
Check the line voltage at the charging / energy unit!
Replace the fuse in the charging / energy unit!
Check the fiber optic cable!
1
)
Replace the charging / energy unit!
E73Error Text:Message: Constancy limit reached.
Possible
cause/action:
E74Error Text:Constancy limit reached.
Possible
cause/action:
1
) Before replacing the charging / energy unit, check the fiber optic cable (LWL):
- Disconnect the fiber optic cable on both ends and check by holding it up to the light
or
- place it on the Clk (U5) of board D3 and check whether light is visible.
For user information only.
Approx. 10-15 treatments still possible, then error 74 appears.
The constancy limit has been reached.
Replace the following parts:
- shock wave head
In addition, replace the water in the entire system!
Siemens AGSPL1-130.840.02Page 9 of 14MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
2 - 10LITHOSTAR MODULARIS Error List
High voltage circuit, E75 – E782
E75Error Text:Chip card terminal error (documented in error log only).
Possible
cause/action:
E76Error Text:Chip card expired
Possible
cause/action:
E77Error Text:Chip card communication error
Possible
cause/action:
E78Error Text:Charging process too long.
Possible
cause/action:
E79Error Text:Call Service
Check execution level for board D3:
AS 00 - 02: replace board D3.
AS 03 and higher: replace chip card reading device.
Wrong card or card incorrectly inserted.
Insert a new card.
Check execution level for board D3:
AS 00 - 02: replace board D3.
AS 03 and higher: replace chip card reading device
Cable?
If the error recurs after replacing board D3, is plug X33 properly
connected?
Line voltage o.k.?
Replace the charging / energy unit!
Possible
cause/action:
1
) Before replacing the charging / energy unit, check the fiber optic cable (LWL):
- Disconnect the fiber optic cable on both ends and check by holding it up to the light
or
- place it on the Clk (U5) of board D3 and check whether light is visible.
HV-cable ok?
Short circuit?
MODULARIS Uro PlusSPL1-130.840.02Page 10 of 14Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
LITHOSTAR MODULARIS Error List2 - 11
Internal errors, E80 – E892
E80Error Text:Flash memory defective.
Possible
cause/action:
E81Error Text:CMOS pulse counter incorrect.
Possible
cause/action:
E82Error Text:Pulse counter backup incorrect
Possible
cause/action:If this error occurs repeatedly, board D3 must be replaced.
E83Error Text:CMOS chip card data incorrect
Possible
cause/action:
E85Error Text:Measurement acquisition incorrect.
Possible
cause/action:
E86Error Text:Flash memory defective
Possible
cause/action:
Not applicable for the user.
Read out the memory and perform a reset at the next opportu-
nity.
If this error occurs repeatedly, board D3 must be replaced.
Not applicable for the user.
Read out the memory and perform a reset at the next opportu-
nity.
E89Error Text:Watchdog responded.
Possible
cause/action:
If this error occurs repeatedly, board D3 must be replaced.
Appears each time a reset is done with switch S3 on board D3.
This error may occur if the system is switched off during initiali-
zation.
Siemens AGSPL1-130.840.02Page 11 of 14MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
2 - 12LITHOSTAR MODULARIS Error List
Ultrasound, E90 – E932
E90Error Text:
Possible
cause/action:No connection from LITHOSTAR to SONOLINE
E91Error Text:
Possible
cause/action:
E92Error Text:The two potentiometers have different values.
Possible
cause/action:
E93Error Text:Erroneous US values.
Possible
cause/action:
Potentiometer skips.
Cable from potentiometer to board D3 ok?
Potentiometer defect?
5
Plug at US holder
Value between 1 -3 and 2-5
must be identical.
3
2
1
Dampness on the potentiometer.
C-MOS has lost values. Perform potentiometer adjustment
(Point "U" in the service menu).
Check default values (see SPL1-130.038.01...).
MODULARIS Uro PlusSPL1-130.840.02Page 12 of 14Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
LITHOSTAR MODULARIS Error List2 - 13
Displays on board D32
Once the system is turned on, different symbols are displayed chronologically on the display of board D3.
Display Possible cause/action
0System o.k.
(only with the ultrasound localization option and if a communication was set
up to the ultrasound device.)
dSoftware download.
1Data transfer, the decimal point is flashing.
4The receiving file is being checked.
5Software download time too long.
7Everything o.k.
(only if no ultrasound localization option is available or if the ultrasound device
is turned off.)
-The service switch is set to normal mode - only for boards that still have no
software.
Text on the Console2
"Chip card invalid" 2
This message is displayed when:
- the gold card is still logged on a different LITHOSTAR MODULARIS
- board D3 was replaced and the "init counter" on the gold card is already set at 3.
- the card is defective.
In normal mode, check whether the LED 174 on board D3 blinks temporarily and then
remains lit when turning it on.
LITHOSTAR MODULARIS SIEMENS2
This message is displayed when too many errors occur. In this case, use the service PC
to generate an error log.
Missing Iso-center cross on the ultrasound system2
There is no communication between the ultrasound system and the LITHOSTAR
MODULARIS.
Siemens AGSPL1-130.840.02Page 13 of 14MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
2 - 14LITHOSTAR MODULARIS Error List
This page intentionally left blank.
MODULARIS Uro PlusSPL1-130.840.02Page 14 of 14Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
Service software3
(Hyper) Terminal program3
The Windows program (Hyper) Terminal is used to retrieve error codes, settings, shock
wave counts, etc. from the LITHOSTAR MODULARIS and to save them to diskette.
The error log and therapy data can only be deleted from the CPU board using this (Hyper)
Terminal program.
This program is usually located under Accessories in WINDOWS.
The following provides a short overview of the connection procedure.
The standard cable (99 00 440 RE999) is used as the connection cable between the ser-
vice PC and the LITHOSTAR MODULARIS.
3 - 1
NOTE
In case of problems, use the Help function in the (Hyper) Terminal program.
The font "Courier New" is recommended for reading data stored on the diskette.
Due to the numerous versions and languages for WINDOWS, the
connection procedure is described in German and English.
Siemens AGSPL1-130.840.02Page 1 of 18MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
3 - 2Service software
German WINDOWS 95 / 98 / NT3
Generate and switch on connection between Service PC and the D3.X9 (com 3 Service) board
Open the Hyper Terminal Program (located under "Windows" - "Zubehör")
in LITHOSTAR MODULARIS
Set switch S2 on board D3 to service
First time
connection ?
Click on Hypterm.exe
In the window" Beschreibung der Verbindung"
for the name: enter Litho / for the icon: select one and click OK
In the window "Rufnummer", select, com1 or com2 for the
Enter the following settings in the window:
Parität: keine, Stopbits: 2, Protokoll: kein and click OK
connection, and click OK
Bits pro sekunde: 9600, Daten bits: 8
1
New
board?
Save
the data to
drive a: or c:
Yes
No
Perform software download - page 3-13
Yes
Menu: Click on "Übertragung" and select "Text aufzeichnen"
In the window "Datei aufzeichnen" accept the file path indicated or
search for the desired file path with "Browse"... then click "Start"
NoYes
Open the connection, e.g. Litho
Press the "H" key, the text is displayed
Perform the desired tasks (refer to pages 3-4 to 3-10)
Then in the "Übertragung" Menu , select "Text aufzeichnen" and select "Beenden"
In the "Datei" menu, click on exit
Enter the appropriate response to the other queries
Set switch S2 on board D3 back to normal and remove the connection cable
End
*Additional information: In order to save the information from LITHOSTAR MODULARIS, it is best to save
the desired file as a text file. Basic procedure: open Windows Explorer. Open the folder, e.g. Temp, then click on
MODULARIS Uro PlusSPL1-130.840.02Page 2 of 18Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
the Datei menu and click "Neu / Textdatei". Enter the name for the text file.
Service software3 - 3
English Windows 95 / 98 / NT3
Generate and switch on connection between Service PC and the D3.X9 (com 3 Service) board
Open the Hyper Terminal Program (located under "Windows" - "Accessories")
in LITHOSTAR MODULARIS
Set switch S2 on board D3 to service
First time
connection ?
Click on Hypterm.exe
for the name: enter Litho ; for the icon: select one and click OK
In the window "Connect To " select : com1 or com2 and click OK
In the window" Connection Description "
Enter the following settings in the window:
Bits per second: 9600, Data bits: 8, Parity: none,
Stop bits:2, Flow control: None, and click OK
1
Save the data
to drive c: or a:
New
board?
Yes
Yes
NoYes
Open the existing connection, e.g. Litho.ht
Perform software download - page 3-13
In the Transfer menu, select Capture Text..*
In the window " Capture Text " accept the file path indicated or
search for the desired file path with Browse.. and click Start
No
Press the "H" key to display the text
Perform the desired tasks (refer to pages 3-4 to 3-10 )
Then in theTransfer menu, select Capture Text and select Stop
In the File menu, click Exit
Enter the appropriate response to the other queries
Set switch S2 on board D3 back to normal and remove the connection cable
End
*Additional Information: In order to save the information from LITHOSTAR MODULARIS, it is best to save the
desired file as a text file. Basic procedure: open Windows Explorer, open the folder e.g. Temp, then click on the
Siemens AGSPL1-130.840.02Page 3 of 18MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
"File" menu and click on New / Text file and enter the name for the text file.
3 - 4Service software
Working with the Terminal program3
Once you create a connection between the service PC and the LITHOSTAR
MODULARIS, press the "h" key to display the following text:
You will be notified if the software version is not compatible with the hardware version.
Refer to document SPL1-130.038.01... (Intranet) for default values.
select option:
==============
h: show this help screen
p: show pulse counters
d: show diagnostic data
e: show error log 100 errors present
t: show therapy data
c: show chipcard data
a/b: adjust c-arm turn-back (a=-, b=+) now: 1200 ms
i: install/remove option ECG-trigger now: not installed
u: adjust ultrasound localization
r: set pulse rate now: variable 1/1.5/2 Hz
s: download new software
o: change orientation of treatment table. Now: ------ ----- : \_/ :
(X) = Physician's place for (X) : :
endourolog. therapy : :
---------------
The following will be displayed if the software version is not compatible with the hardware
version:
--------------< SIEMENS LITHOSTAR MODULARIS >------------- software version: V_ _ _ hardware:_ _ - _ _
** Error: software is not compatible with hardware and/or EPLD-version **
** System will not work ! Please download appropriate software. **
select option:
==============
h: show this help screen
p: show pulse counters
...
...
o: change orientation of treatment table. Now: ------ ----- : \_/ :
(X) = Physician's place for (X) : :
endourolog. therapy : :
---------------
MODULARIS Uro PlusSPL1-130.840.02Page 4 of 18Siemens AG
Rev. 0304.05CS PS 24Medical Solutions
Service software3 - 5
select option: h3
• Press the "h" key to display the input menu.
select option: p3
• Press the "p" key to display the following text:
=> P
number of pulses:
system: 3627
(1) shock head: 3627
(2) capacitor: 3627
(3) spark gap: 3627
select 1..3 to delete pulse counter, any other key to abort
please note values or save to file before deleting data !
The total number of shock waves released are displayed.
The counters for the shock head / capacitor / spark gap can be reset to zero.
Always perform this procedure when replacing one of these parts, i. e. the "spark gap"
counter must be reset to zero if the charging unit is replaced.
select option: d3
Press the "d" key to display the following text
Ö
D
'x'=exit
sensors HVP-bellow C-arm potiUS poti1US poti2
(C-grade)(V)(mV(mV) (mV) (mV)
24.15 23.97 20 75651006300 6250
24.21 24.00 20 74951006300 6250
sensors (C-grade)Displays the temperature at the shock wave head in celsius.
The difference between both temperature values may not exceed a
maximum of 3° C.
HV (V)Displays the voltage at the high-voltage capacitor.
If there is no connection to the charging unit (i.e. fiber optic cable
removed), the value is 20460.
P-bellow (mV)Displays the pressure in the coupling bellows.
enter 'A' to list all data: 40 patients, approx. 20 kByte of data
or
enter date of one file to read, press 'space' to stop display
format mm/dd/yyyy (x=exit):
Displays all days for which therapy data exists, as well as the number of patients.
To select the day: enter month/day/year.
The following data is displayed for each treatment (max 40):
- each increase in energy level
- the number of shock waves for each of the energy levels
(energy levels between 0.1 and 1.0 will be summarized in steps of 0.1)
- any errors that occurred
- the text output can be temporarily stopped at any time with the "Space" key.
Entering "A" is only recommended for saving all data to a diskette.
The data cannot be deleted until they have been read (on a workday).
Selecting delete will erase all therapy data.
Siemens AGSPL1-130.840.02Page 7 of 18MODULARIS Uro Plus
Medical SolutionsRev. 0304.05CS PS 24
3 - 8Service software
Content of Therapy log (selected day):
Patient no. 0001 start of therapy: 03/03/1999 17:03.15
---------- table of energy-level (*10) : pulses -----------