Seca 335 LC, 334 LC Service Manual

Variants:
3341321008 3341321368
Service Manual
for seca 334 LC / 335 LC
3357021094 3357021129 3357021159
Description: Baby scale with 4 load cell technology
When servicing or repairing seca devices according to this service manual, always take note of the instruction manual of th e product. To download the latest version of the inst ruction manual and EMC
Recommendation go to
www.seca.com
Contents:
Description electronic 30-34-00-864 Cableplan Replacement
08-02-06-036 a 30-34-00-752 a
17-05-01-342-b
Valid as of: 01.04.2013
.
Spare parts Spare parts Spare parts Spare parts drawing EMC Recommendation 30-35-00-219 Instruction manual mod. 334 LC Instruction manual mod. 335 LC 17-10-07-552 Instruction manual mod. 335 LC China 17-10-07-555
Manual number: 17-05-01-342-b
334 1321008 334 1321368 335 7021159 30-34-00-749 b
17-10-06-335
Service Instructions LC Electronics seca
Service Instructions Electronics (LC)
MODEL
[334|335|382|383|384|385|769|799|780|833|834|835|869|899|874|
875|876|877|878|952|956]
03.04.2013/JEG 1 (24) 30-34-00-864
Contents seca
Contents
1 General Information....................................................................................................................... 3
1.1 Rating plates..................................................................................................................... 3
1.2 Structure and function of scale section ............................................................................ 4
1.2.1 Structure .............................................................................................................. 4
1.2.2 Function............................................................................................................... 4
2 Calibration of Scales ..................................................................................................................... 6
2.1 Manual calibration of scale [334|335|382|383|384|385|833|834|835].............................. 6
2.2 Manual calibration of scale [874|875|876|877|878|952|956]............................................ 9
2.3 Manual calibration of scale [769|780|799|869|899]........................................................ 11
2.3.1 Linear calibration ............................................................................................... 11
2.3.2 Calibration of corner loading.............................................................................. 13
3 Troubleshooting........................................................................................................................... 15
3.1 Error symptoms scale..................................................................................................... 15
3.2 Scale error messages..................................................................................................... 16
3.3 Error messages [334|335|382|383|384|385|833|834|835|874|875|876|877|878|952|956]16
3.4 Error messages [769|799|780|869|899]......................................................................... 17
4 Measurements for Fault Localisation .......................................................................................... 19
4.1 Power supply.................................................................................................................. 19
4.2 Load cell ......................................................................................................................... 19
4.2.1 Testing load cells with a multimeter................................................................... 19
5 Replacing Electronics..................................................................................................................21
6 Appendices.................................................................................................................................. 23
03.04.2013/JEG 2 (24) 30-34-00-864
General Information seca
1 General Information
These service instructions are for technicians responsible for the maintenance and repair of devices. These persons must be familiar with all regulations governing electrical engineering and comply with them in all cases. These instructions are not intended for users without specialist skills.
These instructions describe servicing for the devices [334|335|382|383|384|385|769|799| 780|833|834|835|869|899|874|875|876|877|878|952|956]. The individual sections indicate the devices to which they apply in each case.
1.1 Rating plates
To identify the exact model see the information and serial number on the underside or base frame of the device (see Figure 1). Please take note of this data so you have it to hand when making queries or ordering spare parts.
Figure 1: Rating plates
03.04.2013/JEG 3 (24) 30-34-00-864
General Information seca
1.2 Structure and function of scale section
1.2.1 Structure
The scale basically consists of the following parts:
Base frame with up to four load cells Display unit with keypad Power supply (batteries, optional power pack)
.
1.2.2 Function
The effect of weight on the weighing surface causes elastic deformation of the load cell(s) underneath. They produce an electronic analogue signal which shows linear variation in relation to the application of force. This signal is measured by the weight calculation system, analysed and displayed as a weighing value. Figure 2 shows the function chart for scales with separate electronic modules for weight calculation and display. Figure 3 shows the function chart for scales with one electronic module for weight calculation and display.
LCD module
08-06-18-147
Keypad
(Modular cable)
LC weight calculation
DC power
08-06-18-149
Load
cell
03.04.2013/JEG 4 (24) 30-34-00-864
Load
cell
Figure 2: Function chart [769|799|780|869|899]
Load
cell
Load
cell
General Information seca
LC compact modules (weight
calculation and display)
08-06-18-142 or 08-06-18-160
Keypad
DC supply
Load
cell
Figure 3: Function chart [334|335|382|383|384|385|833|834|835|874|875|876|877|878|952|956]
Load
cell
Load
cell
Load
cell
03.04.2013/JEG 5 (24) 30-34-00-864
Calibration of Scales seca
2 Calibration of Scales
The scale is equipped with a calibration option to compensate for linear measurement errors as for example, those occurring with geographical variation in the earth's gravitational force. This calibration procedure must also be carried out whenever a load cell is replaced.
Calibration can be performed in one of two ways:
The scales are equipped with an internal manual calibration feature, which is
regulated by the control elements on the device.
Calibration is performed with the help of seca serva
2.1 Manual calibration of scale [334|335|382|383|384|385|833|834|835]
Manual calibration is activated by plugging a jumper on the printed circuit board (see Figure 4). These contacts are only accessible after breaking the seal and opening the scale. After plugging the jumper, the scale has to be switched on to start manual calibration.
Figure 4: Jumper contacts to activate manual calibration
03.04.2013/JEG 6 (24) 30-34-00-864
Calibration of Scales seca
The following steps are necessary to perform manual calibration of the scale:
1. Switch on the scale and check the battery charge (see note). Calibration must not continue if the batteries are low!
2. Then switch the scale off and unscrew the lower service lid after having removed the seal.
3. Connect the two jumper contacts using a jumper.
4. Then set up the scale properly again and switch on.
5. Start-up of the scale is followed by the segment test and then a flashing display showing the weight to be placed on the scale.
6. Place the calibrated weight in the middle of the weighing surface in one, or at most two steps of equal size. The calibration weight required may vary depending on the model (see Table 1).
Model Calibration weight
335, 382, 384 10 kg
334, 384, 833, 834 15 kg
835 20 kg
383, 385 40 kg
Table 1: Calibration weights required
7. Once a correct, stable weight value is identified, the LCD display will stop flashing, and a calibration value will be calculated and stored. The scale then switches off automatically shortly afterwards. If the LCD display does not stop flashing, the measured weight is either above or below the weight to be expected, or a stable weight value cannot be established (due to vibration / shocks, etc.).
03.04.2013/JEG 7 (24) 30-34-00-864
Calibration of Scales seca
8. Then remove the weight from the scale, detach the jumper from the PCB and replace the service lid.
9. Now set up the scale again and switch on. Next, place a calibrated weight on the scale and check the display for the correct weight value. If manual calibration was successful, reseal the service lid according to the national regulations.
Note: Automatic monitoring of the battery voltage is not possible during manual calibration. It
should therefore be checked before and after every calibration procedure. This can be carried out by switching on the scale and observing the display. If the battery voltage is insufficient, either the battery warning symbol or the "bAtt" display will be activated. If one of these displays appears before or immediately after calibration, the batteries should be replaced and the calibration procedure repeated.
Manual calibration can be aborted before placing the required weight on the scale by switching off the device.
03.04.2013/JEG 8 (24) 30-34-00-864
Calibration of Scales seca
2.2 Manual calibration of scale [874|875|876|877|878|952|956]
Manual calibration is activated using a sliding switch (SJ1) on the PCB (see Figure 5). It is only accessible after breaking the seal and opening the scale. After changing the switch position to 2 ("CAL"), the scale has to be switched on to start manual calibration.
Figure 5: Slide switch to activate manual calibration
The following steps are necessary to perform manual calibration of the scale:
1. Switch on the scale and check the battery charge (see note). Calibration must not continue if the batteries are low!
2. Then switch the scale off and open the integrated display by releasing the four screws after having removed the seals.
3. The position 2 ("CAL") must be selected with the slide switch on the PCB, which is now accessible.
4. Then set up the scale properly and switch on.
5. Start-up of the scale is followed by the segment test and then a flashing display showing the weight to be placed on the scale.
03.04.2013/JEG 9 (24) 30-34-00-864
Calibration of Scales seca
6. Place the calibrated weight in the middle of the weighing surface in one, or at most two steps of equal size. The calibration weight required may vary depending on the model (see Table 2).
Model Calibration weight
874, 875, 876, 877, 878,
952, 956
Table 2: Calibration weights required
120 kg
7. Once a correct, stable weight value is identified, the LCD display will stop flashing, and a calibration value will be calculated and stored. The scale then switches off automatically shortly afterwards. If the LCD display does not stop flashing, the measured weight is either above or below the weight to be expected, or a stable weight value cannot be established (due to vibration / shocks, etc.).
8. Then remove the weight from the scale, move the slide switch on the PCB back to its original position and refit the display.
9. Now set up the scale again and switch on. Next, place a calibrated weight on the scale and check the display for the correct weight value. If manual calibration was successful, reseal the scale according to the national regulations.
Note: Automatic monitoring of the battery voltage is not possible during manual calibration. It
should therefore be checked before and after every calibration procedure. This can be carried out by switching on the scale and observing the display. If the battery voltage is insufficient, either the battery warning symbol or the "bAtt" display will be activated. If one of these displays appears before or immediately after calibration, the batteries should be replaced and the calibration procedure repeated.
Manual calibration can be aborted before placing the required weight on the scale by switching off the device.
03.04.2013/JEG 10 (24) 30-34-00-864
Loading...
+ 25 hidden pages