Rotronic Hygrolab C1 Specifications

Page 1
Laboratory device for water activity measurements
HYGROLAB
Benefits
The reliable AW Quick mode for quick results (typical 4 - 5 min)
4 probe inputs for simultaneous or asynchrony measurements
Standalone laboratory measuring device with 7” touch screen
Remote access with PC or Tablet
Firmware updates over the internet
Applications
Food Industr y
Pharmaceutical Industr y
Chemical Industry
Cosmetics Industr y
DESCRIPTION
Measurement
• AW Quick functions for quick results within 4 to 5 minutes
• Normal AW measurements, which takes from 30 to 60 minutes
Internet access
• Easy updates of the HygroLab over the Internet
• Remote access by a normal browser with the PC or Tablet
• Sharing protocols via E-Mail or downloading it with the PC
• Optional possibilit y for additional WiFi stick
Connections
• 4 x HC2-connectors for HC2-AW and HC2A-S probes, which can measure simultaneity or asynchrony
• 2 x USB 2.0 and 2x USB 3.0 (Type A) for connecting a Mouse or an additional Keyboard
• Ethernet (L AN) port for Internet access
• Power supply port for the included AC adapter with 2 plugs for Nor th America and Europa (incl. UK )
Stable housing
• Use of robust materials such as Aluminum, ABS and Stainless steel
• Magnetic attachment in the back for the included pen
HC2-AW
HygroLab
WP -14-S
Pen
AC adapter
1
www.rotronic.com
59096E/2020-04
Page 2
PROBES
Water activity probe (HC2-AW)
Operating range: -40…85 °C (-40…185 °F) / 0…100 %RH
Probe type: Standard water activity probe
Humidity probe (HC2A-S)
Operating range: -50...100 °C (-50…212 °F ) / 0…100 %RH
Probe type: Standard probe for humidity & temperature
ACCESSORIES
Sample holders and sample containers
The sample holders provide excellent sample containment and optimum temperature stability. The WP-40TH can be attached to a water bath for additional temperature control.
The disposable sample containers ensure the optimum sample volume is introduced into the sample holders. They prevent the sample holders from coming into direct contact with the product being tested, thereby preventing soiling or cross contamination. The sample containers also provide a convenient means of collecting and storing samples.
Sample Holders
Order code WP -14 -S WP-40 WP-40TH
Type Use with Height Material Weight
Order code PS -14 PS-40
Type Use with Height Packaging unit
Sample holder PS -14 PS-14 / PS-40 14 mm 40 mm V2A steel Brass, nickel-plated 350 g 1250 g 1550 g
Order code AW-KHS
Sample container WP-14-S / WP-40 / WP-40TH WP-40 / WP-40TH 14 mm 40 mm 100 pc .
Clamp sealing mechanism
In the case of very dry or ver y moist samples additional mechanical sealing of the Aw measurement probe and sample holder may be necessary to prevent external conditions inf luencing the sample.
2
www.rotronic.com
Use with
Weight
WP-40 WP-40TH
1100 g
59096E/2020-04
Page 3
214,40 mm
TECHNICAL INFORMATION
78,88 mm
Dimensions
146,07 mm
214,40 mm
78,88 mm
General specifications
Order code HygroLab Operating range 0…40 °C (32…104 °F ) / 0…80 %RH Parameters shown AW, %RH, °C, °F Dimensions 215 ×146 ×79 mm Accurac y @ 23 °C ± 5 %
(HC2-AW) Long-term stability (HC2-AW) < 1 %rh / year (Normal conditions) AW Quick Mode Yes Display 7” LCD capacitive touch screen Power button Push but ton (with RGB light indicator) Trend display Yes Clock Real time clock provides a time stamp
Storage capacit y 16 GB Power supply 12VDC 2.0A 24W Max (AC adapter) Power supply plugs CH (EU & UK) / US Probe connections 4 x HC2 station probe
USB (A) types 2x 2.0 and 2x 3.0 USB power supply 5V / 500 mA Measurement interval 15 s Start-up time 30 s Software update Direct with the HygroLab over the internet Protocols (values) Date, time, AW / RH, °C / °F, AW Setpoint,
HW4 compatible No Acoustic alarm No Calculations No psychrometric calculations available Delivery package • HygroLab
Housing / Mechanics
Housing material Aluminum, ABS, stainless steel Weight 1.02 kg (2.25 lbs)
Conformity with standards
CE confor m CE 2014/30/EU IP protection rating IP30
± 0.008 aw / 0.8 %rh / ±0.1 K
for ever y measurement
(HC2A-S or HC2-AW)
Serial No., etc.
• Power adapter
• Pen
• Short instruction manual
• USB WiFi stick (optional)
Subject to technical change without notice. Printing and other errors reserved.59096E/2020-04