Rohde&Schwarz UPV-K9, UPV-K91 Operating Manual

PAD-T-M: 3574.3259.02
/01.00
/CI/1/EN
R&S®UPV-K9/-K91 LTE/UMTS/GSM Mobile Phone Tests
Operating Manual
Version 3.3.1.84
Operating Manual
1402.0043.12 – 10
© 2020 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29 - 0 Fax: +49 89 41 29 12 164 E-mail: info@rohde-schwarz.com Internet: http://www.rohde-schwarz.com Subject to change – Data without tolerance limits is not binding. R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners.
The following abbreviations are used throughout this manual: R&S®UPV-K9/-K91 is abbreviated as R&S UPV-K9/-K91
R&S UPV-K9/-K91 Contents
Operating Manual 1402.0043.12 - 10 3

Contents

1 Overview ................................................................ .............................. 9
2 Preparation and Start of the Application Software ........................ 10
2.1 Required Measuring Instruments and Accessories ...............................................10
2.2 Installing the Software ..............................................................................................15
2.3 Verification of the Installation ..................................................................................17
2.4 Test Setup ...................................................................................................................18
2.5 Starting the Application Software ............................................................................22
2.6 Options (General settings) ........................................................................................27
2.6.1 Mouth Calibration Frequency Range ...........................................................................27
2.6.2 Standard ......................................................................................................................28
2.6.3 Select Standard at Startup ...........................................................................................28
2.6.4 Release ........................................................................................................................28
2.6.5 Select Release at Startup ............................................................................................28
2.6.6 Ear Simulator ...............................................................................................................29
2.6.7 Select Ear Simulator at Startup ...................................................................................29
2.6.8 Artificial Mouth .............................................................................................................29
2.6.9 Select Artificial Mouth at Startup ..................................................................................29
2.6.10 System Simulator .........................................................................................................29
2.6.11 Hands-free Settings .....................................................................................................29
2.6.12 Activation Signal for Distortion Tests ...........................................................................29
2.6.13 Show Operator Instructions .........................................................................................30
2.6.14 R&S CMU Remote Control ..........................................................................................32
2.6.15 R&S CMU Subsystem .................................................................................................32
2.6.16 R&S CMW Remote Control .........................................................................................32
2.6.17 Noise Calibration Configuration ...................................................................................33
2.6.18 R&S UPP Remote Control ...........................................................................................40
2.6.19 Input Switcher ..............................................................................................................41
2.6.20 Report settings .............................................................................................................41
2.6.21 Generate Temporary Export Files ...............................................................................41
2.6.22 Generate Temporary Image Files ................................................................................42
2.6.23 Store Results of Further Measurements ......................................................................42
2.6.24 Do not Change Scale for Further Meas. ......................................................................42
R&S UPV-K9/-K91 Contents
Operating Manual 1402.0043.12 - 10 4
2.6.25 Store Loaded Curve Data to Results ...........................................................................42
2.6.26 Enable Remote Control ...............................................................................................42
3 Calibration ......................................................................................... 43
3.1 Calibration Devices....................................................................................................43
3.2 Microphone Calibration .............................................................................................45
3.3 Calibration of Artificial Ear .......................................................................................47
3.3.1 Calibration of Ear Type 1 .............................................................................................47
3.3.2 Calibration of Ear Type 3.2 Low Leakage ...................................................................48
3.3.3 Calibration of Ear Type 3.2 High Leakage ...................................................................48
3.3.4 Calibration of Ear Type 3.3 ..........................................................................................49
3.3.5 Calibration of Ear Type 3.4 ..........................................................................................51
3.4 Calibration of Artificial Mouth for Handset Tests ...................................................51
3.5 P.50 Speech Spectrum Calibration ..........................................................................53
3.6 P.501 Speech Spectrum Calibration ........................................................................54
3.7 Calibration of CMU Speech Codec...........................................................................55
3.8 Calibration of CMW Speech Codec ..........................................................................56
3.9 Calibration of Ambient Noise Field ..........................................................................57
3.10 Calibration of Noise Field for “Speech Quality in Presence of Background Noise”
Test ..............................................................................................................................60
3.10.1 Connections .................................................................................................................60
3.10.2 Establishing the Remote Control Connection between R&S UPV and R&S UPP ......61
3.10.3 Prerequisites ................................................................................................................61
3.10.4 Configuration ................................................................................................................61
3.10.5 Switcher Support..........................................................................................................67
3.10.6 Selection of Background Noise Signals .......................................................................70
3.10.7 Starting the Calibration Process ..................................................................................70
3.10.8 Stopping and Resuming the Calibration Process ........................................................71
3.10.9 Preparatory Measurements for Delay Measurement ...................................................73
3.10.10 Delay Measurement .....................................................................................................74
3.10.11 Preparatory Measurements for Equalization ...............................................................77
3.10.12 Shared Equalization Method ........................................................................................80
3.10.13 Individual Equalization Method ....................................................................................80
3.10.14 Level Adjustment .........................................................................................................81
3.10.15 Equalization .................................................................................................................81
3.10.16 Final Test of All Ambiances .........................................................................................84
3.10.17 Calibration Sections and Resuming Points .................................................................84
3.10.18 Calibration Report ........................................................................................................86
R&S UPV-K9/-K91 Contents
Operating Manual 1402.0043.12 - 10 5
3.11 Synchronization of CMW clock ................................................................................89
3.11.1 Preparations .................................................................................................................89
3.11.2 Starting the Synchronization Process ..........................................................................90
3.11.3 Stopping the Synchronization Process ........................................................................91
3.11.4 Steps of Synchronization Process ...............................................................................91
3.11.5 Options after Termination of Synchronization .............................................................95
3.11.6 Synchronization Report ...............................................................................................96
4 Data Entry for Reporting .................................................................. 98
4.1 Operator ......................................................................................................................98
4.2 Test object ..................................................................................................................98
5 Measurements ................................................................................... 99
5.1 General ........................................................................................................................99
5.2 Starting Measurements .............................................................................................99
5.3 Functionality and Control of the Measurement Macros ......................................101
5.3.1 Zooming .....................................................................................................................101
5.3.2 Changing the Scale of the Graph ..............................................................................102
5.3.3 Cursor ........................................................................................................................102
5.3.4 Data Point Size ..........................................................................................................102
5.3.5 Entering a Comment ..................................................................................................102
5.3.6 Storing a Hardcopy of the Graph ...............................................................................102
5.3.7 Making Additional Measurements ..............................................................................103
5.3.8 Storing and Loading Curves ......................................................................................103
5.3.9 Storing Curves as Limit Curves .................................................................................104
5.3.10 Deleting Additional Curves in the Measurement Graph ............................................104
5.3.11 Creating a Report.......................................................................................................104
5.3.12 Closing the Measurement Window ............................................................................104
5.4 Applicability of Measurements and Equipment Depending on the Release .....104
5.5 Sending Frequency Response and Loudness Rating .........................................105
5.5.1 Sending Frequency Response ..................................................................................105
5.5.2 Sending Loudness Rating ..........................................................................................105
5.5.3 Sending Tests Using Artificial Voice According to ITU-T P.50 or Single-Talk Speech
according to ITU-T P.501 as Test Signal ...................................................................107
5.6 Receiving Frequency Response and Loudness Rating.......................................107
5.6.1 Receiving Frequency Response ................................................................................108
5.6.2 Receiving Loudness Rating .......................................................................................109
5.6.3 Receiving Tests Using Artificial Voice according to ITU-T P.50 or Single-Talk Speech
according to ITU-T P.501 as Test Signal ...................................................................110
R&S UPV-K9/-K91 Contents
Operating Manual 1402.0043.12 - 10 6
5.7 Sidetone Masking Rating (STMR) ..........................................................................111
5.7.1 STMR Tests Using Artificial Voice according to ITU-T P.50 or Single-Talk Speech
according to ITU-T P.501 as Test Signal ...................................................................112
5.8 Sidetone Delay .........................................................................................................113
5.9 Roundtrip Delay .......................................................................................................114
5.10 Receiving Impairments ...........................................................................................116
5.11 Jitter Buffer Management Behaviour .....................................................................118
5.12 Echo Loss (TCLw)....................................................................................................119
5.13 Stability Margin ........................................................................................................120
5.14 Stability Loss ............................................................................................................121
5.15 Echo Control Characteristics .................................................................................122
5.16 Sending Distortion ...................................................................................................125
5.17 Receiving Distortion ................................................................................................129
5.18 Idle Channel Noise Sending ...................................................................................132
5.19 Idle Channel Noise Receiving .................................................................................134
5.20 Ambient Noise Rejection ........................................................................................136
5.21 Speech Quality in Presence of Ambient Noise .....................................................138
5.21.1 Setup ..........................................................................................................................138
5.21.2 Connections ...............................................................................................................139
5.21.3 Switcher Support........................................................................................................139
5.21.4 Prerequisites ..............................................................................................................141
5.21.5 Measurement .............................................................................................................141
5.21.6 Continuing Aborted Measurements ...........................................................................143
5.21.7 Typical Problems during measurement .....................................................................143
5.21.8 Additional operations when finished ..........................................................................145
5.21.9 Extended Speech Quality Measurements .................................................................145
5.21.10 Continuing Aborted Extended Measurements ...........................................................146
5.22 UE Send Clock Accuracy ........................................................................................147
5.23 Notes on Hands-Free Measurements ....................................................................148
5.23.1 General Remarks .......................................................................................................148
5.23.2 Test Setup ..................................................................................................................148
5.23.3 Acoustic Calibration for Hands Free Tests ................................................................150
5.24 “Utilites” Measurements .........................................................................................151
5.24.1 Sidetone Distortion.....................................................................................................151
5.24.2 Gain Variation Tests ..................................................................................................152
5.24.3 Delay Measurements .................................................................................................153
5.24.4 Background Noise Measurements .............................................................................154
R&S UPV-K9/-K91 Contents
Operating Manual 1402.0043.12 - 10 7
6 Customizing Measurements .......................................................... 156
6.1 Editing Parameters ..................................................................................................157
6.2 Editing Limit Curves ................................................................................................158
7 Measurements with Electric Connections .................................... 159
7.1 Introduction ..............................................................................................................159
7.2 Calibration Values for Electric Connections .........................................................159
7.2.1 Electric Connection Replacing Artificial Ear ..............................................................159
7.2.2 Electric Connection Replacing Artificial Mouth ..........................................................160
7.2.3 Electric Connection Replacing Encoder ....................................................................160
7.2.4 Electric Connection Replacing Decoder ....................................................................160
7.3 Performing the Measurements ...............................................................................160
8 Automatic Test Sequences ............................................................ 161
8.1 Creating and Editing a Sequence ...........................................................................161
8.2 Remote Control of R&S CMU200 within a Sequence ...........................................162
8.3 Remote Control of R&S CMW500 within a Sequence ..........................................163
8.4 Starting an External Executable within a Sequence ............................................164
8.5 Opening an Existing Sequence ..............................................................................164
8.6 Running a Sequence ...............................................................................................165
8.7 Running a Single Measurement out of a Sequence .............................................165
8.8 Reporting on Sequence Results ............................................................................165
9 Reporting, Storing, Loading and Deleting Results ...................... 166
9.1 Result Files ...............................................................................................................166
9.2 Report Settings ........................................................................................................166
9.3 Generating a Single Report ....................................................................................167
9.4 Generating a Sequence Report ..............................................................................167
9.5 Selection Report ......................................................................................................167
9.6 Preview Window.......................................................................................................168
9.7 Storing and Loading Curves ...................................................................................169
9.8 ASCII Result Files ....................................................................................................169
9.9 Deleting Results .......................................................................................................169
10 Remote Controlled Start of Testcases via GPIB .......................... 170
10.1 Preparations .............................................................................................................170
10.2 Starting a Measurement ..........................................................................................170
10.3 Reading the Results ................................................................................................170
R&S UPV-K9/-K91 Contents
Operating Manual 1402.0043.12 - 10 8
10.4 Determining the Termination of a Measurement ..................................................173
11 Terminating the Application .......................................................... 174
12 Appendix A Settings on the Radio Communication Tester
R&S CMU200 ................................................................................... 175
12.1 Settings for GSM:.....................................................................................................175
12.2 Settings for UMTS WCDMA FDD: ...........................................................................178
13 Appendix B Settings on the Radio Communication Tester
R&S CMW500 .................................................................................. 182
13.1 Settings for LTE: ......................................................................................................182
13.2 Settings for UMTS WCDMA FDD: ...........................................................................185
13.3 Settings for GSM:.....................................................................................................187
14 Appendix C Troubleshooting ................................ ......................... 189
14.1 Error Message 2908 During Installation ................................................................189
14.2 Damaged Setup File is Reported ............................................................................189
14.3 Damaged Results File is Reported.........................................................................189
14.4 A Test is Not Starting Properly ...............................................................................189
14.5 A Calibration Value is Missing or No Device is Selected for a Required
Calibration Type .......................................................................................................189
14.6 The Receiving Noise Test Produces an Overrange Error ...................................190
14.7 ARL for the Sending Distortion Test Cannot be Adjusted ..................................190
14.8 A Measurement Using a Custom Limit Curve Produces an Error ......................190
14.9 Other Problems of Unknown Reason ....................................................................190
R&S UPV-K9/-K91 Overview
Operating Manual 1402.0043.12 -10 9

1 Overview

The acoustic transmission and reproduction quality of a mobile phone is its most important characteristic in everyday use. Even the most visually appealing design and wonderfully sophisticated operating concept are not much use, when you cannot or can hardly understand what is being said at the other end. Improvements in voice quality achieved with wideband and super-wideband codecs as well as with modern noise suppression and echo cancelling techniques have to be verified. In voice calls over IP-based IMS connections like Voice over LTE, the processing delay in the mobile phone as well as its ability to cope with packet delay jitter and packet loss is important and has to be tested.
Instruments and methods for measuring acoustic characteristics are therefore essential tools for assessing the quality and suitability of a mobile phone.
These tests are based on standards for 3GPP mobile phones. The test methods are stipulated in 3GPP TS 26.132 and the values to be attained in 3GPP TS 26.131.
As of release 4 of the GSM 51.010 standard (successor to GSM 11.10), even GSM mobile phones may be tested to 3GPP TS 26.132.
The R&S® UPV-K9 and R&S® UPV-K91 options (LTE/UMTS/GSM Mobile Phone Tests) of the Audio Analyzer R&S® UPV, called herein below “UPV”, are now available for measuring the acoustic characteristics of LTE, UMTS and GSM mobile phones. The measurements are in line with 3GPP TS 26.131, TS 26.132 and TS 51.010 and have been validated in conjunction with Wideband Radiocommunication tester R&S® CMW500 by an independent test house for conformance testing on LTE, UMTS and GSM mobile phones.
The current version of the software supports 3GPP TS 26.131 and TS 26.132 up to Release 15.
From version 2.3.1.47 of the software, a new “Release” menu is available which allows
you to determine and start the test cases applicable to a given combination of speech codec bandwidth, type of tested device (handset, headset or handheld, desktop or vehicle-mounted hands-free) and release number of the test specifications.

R&S UPV-K9/-K91 Preparation and Start of the Application Software

Operating Manual 1402.0043.12 -10 10
2 Preparation and Start of the Application
Software

2.1 Required Measuring Instruments and Accessories

The R&S® UPV audio analyzer with the options as described below is required for the measurements.
The mobile phone under test is connected via the RF interface using Wideband Radio Communication Tester R&S® CMW500, called herein below R&S® CMW500. Legacy support for the Universal Digital Radio Communication Tester R&S® CMU200, called herein below CMU200,is also provided. These testers simulate a base station for the mobile phone so that a speech call can be set up.
The R&S® CMW500 must be equipped with the options R&S® CMW-B400B (Audio Board) and R&S® CMU-B405A (Speech Board) in addition to the options required for the signaling of the respective radio access technology. For detailed information on the configuration of the R&S® CMW500 please contact your R&S sales office.
The firmware version in the R&S CMW500 must be the following or later:
Base
3.7.170
Audio
3.7.21
LTE Signaling
3.7.70 (SUA only), 3.7.10
Data Application
3.7.50
WCDMA Signaling
3.7.22
GSM Signaling
3.7.30
If using newer versions, make sure that the installed versions are compatible with each other.
For GSM, the R&S® CMU200 must be equipped with the options R&S® CMU-B21 (signaling unit), R&S® CMU-B52 (speech coder/decoder) and the appropriate software options for the GSM band used.
For WCDMA (UMTS), the option R&S® CMU-B69 is required. Wideband-AMR tests require option R&S® CMU-K46 for both GSM and/or WCDMA (UMTS).
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 11
The firmware version in the R&S® CMU200 must be 5.04 or higher.
Acoustic devices such as an artificial mouth, artificial ear and other accessories are required for the measurements. The following equipment from Brüel & Kjær or G.R.A.S. is normally used:
Table 1: Equipment for mobile phone tests
Device
Description
Type (examples)
Telephone test head (up to Release 8)
Device for fixing the DUT in the prescribed position
B&K 4602B
Wideband ear simulator (up to Release 8)
IEC 711 type occluded ear simulator with adapters for connection to the ear piece of the DUT
B&K 4195 (type 3.2)
Artificial mouth (up to Release 8)
Special loudspeaker for simulation of the mouth
B&K 4227 or G.R.A.S. 44AB or 44AA (with power amplifier)
Head and torso simulator
Head and torso simulator with artificial ear (type 3.3) and artificial mouth, alternative to the abovementioned devices, mandatory since Release 9
B&K 4128D (NB and WB only) B&K 5128-D-111 (SWB, but not according to ITU-T
P.58, two artificial ears) Head Acoustics HMS II.3 (SWB)
Driver amplifier for artificial mouth (required for B&K 4227 and
B&K 4128)
Buffer amplifier to deliver the
required current into the 5 Ω load
impedance of the loudspeaker
B&K 2735
Second (left) artificial ear for head and torso simulator B&K 4128
Required for hands-free tests and adjustment of the background
noise field for “speech quality in presence of ambient noise”
B&K 4159C
Handset fixture for Head and Torso simulator
Required to position the handset on the artificial head, and to apply a defined force with the earpiece against the pinna
B&K 4606 for B&K 4128 and B&K 5128 (included with B&K 4128D and B&K 5128D)
Head Acoustics HHP II.1 or HHP IV for HMS II.3
Measurement microphone
½” or ¼” measurement
microphone for measurement of artificial mouth output during calibration. This microphone can use the preamplifier of the artificial ear type 3.2 if existing.
Use a ¼” measurement
microphone for calibration of the HATS mouth.
B&K 4938 with preamplifier B&K 2670 for calibration of the HATS mouth and as reference microphone for background noise tests
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 12
Device
Description
Type (examples)
Acoustic calibrator
Sound level calibrator for calibrating the measuring microphone
B&K 4231 with ¼” adapter DP0775
Microphone power supply
Power supply and preamplifier for the measuring microphone
B&K 2829, 5935L or 2690A0S2 or G.R.A.S. 12AD or 12AA
With the amplifier set to 0 dB, the microphone power supply B&K 2690A0S2 produces too much noise for measuring idle noise and distortion. It is therefore advisable to set a gain of 20 dB.
From Release 9 of TS 26.132, the use of the HATS is mandatory for handset and headset measurements. From Release 10, only artificial ear type 3.3 (anatomically shaped soft pinna) may be used. For Super-Wideband testcases the artificial ears and artificial mouth of the HATS must support the SWB frequency range of approximately 100 Hz to 14 kHz.
A cable with a BNC connector and special small or angled banana plugs is required for connecting the P.51 artificial mouth, as the space between the mouth connectors and the LRGP test head (B&K 4602B) is too small for common banana plugs.
The transformer supplied with option R&S UPV-K9 can only be used with the MMS test signal. It must be connected between generator output 1 of Audio Analyzer R&S UPV and the connector of the artificial mouth. The transformer matches the impedance of the loudspeaker in the artificial mouth to that of the generator output of the R&S UPV. Without this transformer, the available power is too low for driving the artificial mouth.
Alternatively, a power amplifier, preferably with a voltage gain of approx. 0 dB, can be connected between generator output and mouth instead of the transformer. In this case, the gain set must be kept absolutely stable after calibration.
For tests using artificial voice according to ITU-T P.50 or real speech according to ITU-T P.501, a power amplifier is required. It may also be required for high-level activation signals with high crest factor.
For connection to the audio input and output of the radio communication tester R&S® CMW500, use BNC cables and the UP-Z1 adapters (XLR to BNC) supplied with R&S® UPV-K9. Connect R&S UPV generator output 2 to CMW AF1 IN and CMW AF1 OUT to R&S UPV analyzer input 2.
For connection to the "Speech" connector of the Digital Radio Communication Tester R&S® CMU200 a cable with male (analyzer) and female (generator) XLR connector is supplied.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 13
UPV Analyzer
Generator
5
1
Figure 2-1: Assignment of 9-contact speech connector on R&S CMU front panel
This cable is configured for connection to link handler #1 in the R&S CMU200. Depending on the R&S CMU hardware, link handler #2 in the R&S CMU200 can also be used for GSM; in this case, the supplied adapter must be inserted between the cable and the speech connector on the R&S® CMU200.
The R&S® CMU200 connects pin 1 and 3 of the R&S® UPV generator output to the equipment grounding conductor of the mains. If an external power amplifier is used, care must be taken that the external power amplifier does not connect R&S® UPV generator output pin 2 to the equipment grounding conductor of the mains. If available, a balanced connection to the power amplifier should be preferred.
If the CODEC calibration fails, the adapter has most likely to be inserted.
An external power amplifier should be switched on after the UPV-K9x program has been started, and switched off before the program is ended. This prevents the artificial mouth from undue power loading by setups, which are loaded external to the application.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 14
An external USB keyboard and a mouse must be connected to the R&S UPV.
R&S UPV-K9x does not support the use of headphones. Do not connect headphones to the R&S UPV during mobile phone tests. High-level signals may be present at the headphone connector.
The R&S UPV audio analyzer must be equipped as follows:
R&S UPV firmware version 4.0.4.159 or higher.
License Key R&S UPV-K9 installed.
License Key R&S UPV-K91 installed.
For use of test signal according to ITU-T P.50 and distortion, tests with CSS or
customer specific activation signal License Key UPV-K9101 or higher must be installed.
For the use of requirements and test, methods according to Release 10 of TS
26.131 and TS 26.132, License Key UPV-K9102 or higher must be installed.
For the use of requirements and test, methods according to Release 11 of TS
26.131 and TS 26.132, License Key UPV-K9103 or higher must be installed.
For the use of requirements and test, methods according to Release 12 of TS
26.131 and TS 26.132, License Key UPV-K9104 or higher must be installed.
For the use of requirements and test, methods according to Release 13, 14 or 15 of TS 26.131 and TS 26.132, License Key UPV-K9105 or higher must be installed.
License Key UPV-K9105 includes UPV-K9104, UPV-K9103, UPV-K9102 and UPV-K9101.
For testcases 7.11, 8.11, 9.11 and 10.11 (“echo control characteristics”), option R&S UPV-B3 (second analog generator) is required.
For testcases 7.10.4, 8.10.4, 9.10.4 and 10.10.4 (“delay and speech quality in conditions with packet arrival time variations and packet loss”) and for testcases
7.13, 8.13, 9.13 and 10.13 (“Jitter buffer management behavior”), option R&S® UPV-K63 (listening quality analysis POLQA) is required in addition.
For testcases 7.12 and 8.12 (“speech quality in the presence of ambient noise”), the following additional equipment and options are required:
Description
Instrument Type
Ordering Number
R&S UPV-K98
Background noise control software acc. ETSI ES 202 396-1
1424.2003.02
R&S UPV-K101
Measurements using background noise acc. ETSI TS 103 106 and EG 202 396-3
1424.2203.02
R&S UPP200
Audio Analyzer two channels
1411.1003.02
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 15
R&S UPP-B8
Eight-channel generator
1411.2700.02
R&S UP-Z8A
8-channel analog cable for R&S UPP-B8
1411.3206.02
The following equipment not supplied by Rohde & Schwarz is also required:
Device
Description
Type (examples)
2 Pairs of full-range speakers
Frequency response at least 100 Hz to 10 kHz
Behringer Truth B2031A
Subwoofer
Frequency response at least 40 Hz to 120 Hz
M-Audio SBX-10
5 Speaker cables
XLR male – XLR female length approx. 10m
4 Speaker stands
The base plane of the speaker should be adjusTable such that the tweeter of the speaker is at the same height as the artificial ears of the head-and-torso simulator. With the B2031A speakers, the tweeter is about 31 cm above the base. With the Bruel & Kjaer HATS, the entrance of the artificial ear is about 60 cm above the base.

2.2 Installing the Software

The application program requires license keys R&S UPV-K9 and R&S UPV-K91 to be installed. The application program and the license keys are installed together with license key UPV-K9105 in the factory in case a new R&S UPV is ordered together with these options. If the options are ordered separately, the license keys as well as the installation instructions are part of the delivery.
The program required and the associated files are in the folder “UPV-K9x Software” on the installation CD supplied with the R&S UPV-K91 option. It is recommended to copy the files MCRInstaller.exe, CRRuntime_32bit_13_0_12.msi, UPV-
K9x_33184.msi, and dotNetFx40_Full_x86_x64.exe from the folder “UPV-
K9x Software” on the installation CD to drive D: on the R&S UPV hard disk drive, e.g. to a folder named “D:\R&S Software\UPV-K9x\Version 3.3.0”. Run the file UPV-
K9x_33184.msi to start the installation. Follow the instructions of the installer on the screen. If prompted to do so, also run dotNetFx40_Full_x86_x64.exe. If this runtime library has to be installed on a UPV with Windows XP, copy installer wic_x86_enu.exe” to the UPV hard disk and run it before installing the .Net 4.0 runtime library.
Subsequently, run also MCRInstaller.exe and
CRRuntime_32bit_13_0_12.msi.exe.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 16
Note (1) MCRInstaller.exe must be run, unless UPV-K9x software version 3.0.3.56 or higher or UPV-K9y software version 3.0.0.65 (or higher) is already installed on the R&S UPV.
Note (2) If a UPV-K9y software version below 3.0.0.65 is installed it is recommended to update to the latest version (currently 3.1.0.69).
Note (3) Prior to an upgrade of UPV-K9x software from version 2.3.1 or below the program
“MATLAB
®
Component Runtime” must be de-installed using the windows tool “Add or
Remove programs” (XP) or “Programs and Features” (Win7) respectively. Note (4)
In case it is desired to run an older version of the UPV-K9y software, the old version of MCRInstaller.exe (from the older version UPV-K9y installation CD) must be re­installed after installation of the new version.
To have full control over minimized windows, it is recommended to set the windows
taskbar to “Auto-hide”, on top of other windows. Right-click on the Windows Start button and click “properties”.
Figure 2-2: Context menu for taskbar
In the properties window, click tab “Taskbar” and activate “Auto-hide the taskbar” and “Keep the taskbar on top of other windows”. Click “Apply” and close the window.
UPV-K101 does not support super-wideband and fullband evaluation of speech quality in background noise according to ETSI TS 103 281. For this measurement an upgrade license (ACOPT 35) to the batch calculator has to be obtained directly from Head Acoustics GmbH in Herzogenrath, Germany (www.head-acoustics.com).
For an update of UPV-K101 to SWB / FB evaluation, use version 2.1.200 of the Head Batch Processor. Newer versions are not tested with this version of UPV-K91.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 17
Figure 2-3: Properties window for taskbar

2.3 Verification of the Installation

After the installation, check the existence of the subfolder “UPV-K9x Mobile Phone Tests” in “C:\Program Files\Rohde&Schwarz”. This subfolder must contain 90
files plus 2 more subfolders: “General_Data” containing 33 files and “3GPP_Data
with subfolder ”No_Overwrite” containing 124 files and subfolder “Overwrite containing 990 files.
If the software reports a missing key code at the first start, delete folder “D:\3GPP” (if
existing) and install the missing key code before starting the software again.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 18

2.4 Test Setup

Figure 2-4: Test setup and connection of external components with R&S CMU200
Figure 2-5: Test setup and connection of external components with R&S CMW500
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 19
Figure 2-6: Test setup and connection of external components with input switcher R&S UPZ and R&S CMU200
Figure 2-7: Test setup and connection of external components with input switcher R&S UPZ and R&S CMW500
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 20
Figure 2-8: Test setup and connection of external components for the adjustment of the background noise field
Generator channel 10 is the reference channel for the delay measurement. If no switcher is used, it has to be connected according to the instructions from the K9x software.
Figure 2-9: Test setup and connection of external components with input switcher UPZ for the adjustment of the background noise field
Figure 2-10: Test setup and connection of external components with CMU200 for measurements with background noise field
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 21
Figure 2-11: Test setup and connection of external components with R&S CMW500 for measurements with background noise field
Figure 2-12: Test setup and connection of external components with R&S CMU200 and input switcher R&S UPZ for measurements with background noise field
Figure 2-13: Test setup and connection of external components with R&S CMW500 and input switcher R&S UPZ for measurements with background noise field
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 22
In case of hum noise in one or more of the speakers, it may be necessary to insert an isolating audio transformer between the output of UPP-B8 and the input of the speaker. It may also help to connect the speaker to the same power outlet as the UPP.
Switching of connections between tests is required for binaural measurements (handsfree tests and tests of binaural headsets) and for calibration, in particular for the adjustment of the background noise reproduction. In the setups in Figure 2-4, Figure 2-5, Figure 2-8, Figure 2-10 and Figure 2-11, connections have to be changed manually between tests and even during the background noise adjustment. This can be avoided by using an R&S UPZ12 input switcher as shown in Figure 2-6, Figure 2-7, Figure 2-9, Figure 2-12 and Figure 2-13.
Table 2: Assignment of R&S UPZ switcher connectors
UPZ Input
Connected Device
1
Second (left) artificial ear
2
First (right) artificial ear
3
Decoder output
4
Free field microphone
5
Diffuse field microphone
6
R&S UPP-B8 output 10
A
R&S UPV analyzer input 1
B
R&S UPV analyzer input 2

2.5 Starting the Application Software

After installation, the program can be started by double-clicking the Icon “UPV-K9x Mobile Phone Tests” or by clicking “R&S UPV Applications UPV-K9x Mobile Phone Tests” in the “Programs” menu.
At the first start of the program, selection windows appear for the standard according to which the measurements should be made, and for the artificial ear and artificial mouth used.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 23
Figure 2-14: Query window for selection of applied standard
If “none” is selected in the standard selection window, all measurements appear in the “Measurement” menu. If a standard is selected, all measurements appear in the “Measurement” menu, but compliant measurements are checked in the menu. If “Allow only selected measurements” is checked, non-compliant measurements are suppressed in the Measurement menu. “Non-standard handset” and “non-standard hands free” select handset or hands-free tests, respectively, which can be customized.
The next window is for selection of relevant measurements according to tested device type, codec bandwidth, test signal and Release of the 3GPP standards TS 26.131 and TS 26.132. All test cases relevant for the selected combination will be offered in the “Release” menu.
Figure 2-15: Query window for the Release menu
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 24
Figure 2-16: Query window for selection of artificial ear
Figure 2-17: Query window for selection of artifical mouth
When the checkbox “Do not show this dialog again” is checked, the corresponding selection window will not appear at the program start in future. However, the selection
is still available in the “Options” menu (see below). After the last of the four selection
windows has been closed, the main window of the R&S UPV-K9 opens.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 25
Figure 2-18: Main window
Initially the data grid in the center of the screen, which shows that the result overview is empty. The screenshot in Figure 2-18 shows an example after a number of measurements have been made.
By clicking with the right mouse button on the data grid, a context menu opens which allows you to create a report of a result marked by a solid triangle in the respective row header, to edit the comment for a marked result, to delete the marked result or to export one of the curves in the marked result graph to an ASCII format.
Figure 2-19: Context menu of the result data grid
Figure 2-20: A triangle in the row header indicates a marked row
The “Report Selection” submenu allows you to report on or delete a selection of results, which have been marked in the “Select” column of the data grid. To select or unselect a result, click the checkbox in the “Select” column of the respective row.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 26
If the results have been sorted alphabetically according to one of the columns by clicking on the respective column header, the chronological order can be restored using the context menu item “Sort Chronologically”.
Operating Concept R&S 3GPP mobile phone tests consist of the main user interface window, which allows
general settings, calibration routines, data handling, automatic sequencing and reporting tools, and of test macros for basic measurement types defined in the standards. Each measurement type provides a set of parameters (R&S UPV setup files, limits etc.) which are defined in separated measurement definition files, one for each test case.
Do not attempt to modify files with extensions “set”, “sup”, “xml”, “cal” “seq” or “mdf”,
using a text editor. Any change with a text editor may make the files unusable and cause malfunction of the software.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 27

2.6 Options (General settings)

Figure 2-21: Options menu
The “Options” menu in the main window allows changing a set of general settings like ear type and mouth type used, which are valid for all or at least a plurality of the tests.

2.6.1 Mouth Calibration Frequency Range

For narrowband and wideband tests, the mouth frequency response is equalized between 90 Hz and 9000 Hz. For super-wideband and full band tests, a wider frequency range from 90 Hz to 17800 Hz is required. As not all types of artificial mouths support this wider frequency range, “Mouth calibration frequency range" allows you to select the desired frequency range before the start of the calibration of the artificial mouth.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 28
Figure 2-22: Mouth calibration frequency range
The frequency range for super-wideband and fullband tests is identical, as in 3GPP TS
26.132 the test definitions for fullband refer to those for super-wideband.
As the wideband and narrowband frequency ranges are included with the super­wideband frequency range, there is only one mouth calibration required for each artificial mouth in use.

2.6.2 Standard

Standardallows selecting one of the available standards. Measurements belonging to this standard will be checked in the measurement menu. If “Allow only standard measurements” is checked, it is not possible to start single measurements, which do
not belong to the standard.

2.6.3 Select Standard at Startup

If Select standard at startupis checked, the selection window for the standard is opened at each start of the program.

2.6.4 Release

Release opens the window to select type of tested device, codec bandwidth, test signal and Release of 3GPP standards TS 26.131 and 26.132 to be applied. Any change in these settings will re-build the “Release” menu with the applicable test cases.

2.6.5 Select Release at Startup

If Select release at startup” is checked, the selection window for the “Release” menu is opened at each start of the program.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 29

2.6.6 Ear Simulator

Ear simulator allows selecting the type of ear simulator used. For details on the handling of calibration data, see section 3 Calibration below.

2.6.7 Select Ear Simulator at Startup

If Select ear simulator at startupis checked, the selection window for the ear simulator is opened at each start of the program.

2.6.8 Artificial Mouth

Artificial mouth allows selecting the type of artificial mouth used. For details on the handling of calibration data, see section 3 Calibration below.

2.6.9 Select Artificial Mouth at Startup

If Select artificial mouth at startupis checked, the selection window for the artificial mouth is opened at each start of the program.

2.6.10 System Simulator

System simulator allows selecting the type of system simulator used. For details on the handling of calibration data, see section 3 Calibration below.

2.6.11 Hands-free Settings

Hands-free settingsallows selecting the acoustic instruments used for hands free testing. It is possible to use an ITU-T P.51 artificial mouth together with a free field microphone, a HATS with one artificial ear or a HATS with two artificial ears. If two artificial ears are used, the decoder has to be disconnected from analyzer input 2 and the second artificial ear has to be connected to it for receiving measurements. Do not forget to re-connect the decoder for any measurement in sending direction.

2.6.12 Activation Signal for Distortion Tests

In the sub-menu to “Activation signal for distortion tests”, signals can be imported to the test system. If more than one imported signal is available, it is possible to choose one of them for actual use.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 30
Figure 2-23: Window for import of activation signals
Use the “Browse” button to select a wave file, which fulfills the conditions shown on top of the window. The selected signal is analyzed and checked for appropriateness. Note that a high crest factor of the signal may overdrive the codec or artificial mouth. If
“Equalize for Selected Artificial Mouth” is activated, the file is immediately equalized for the selected artificial mouth when the “Ok” button is clicked. Otherwise or when the
mouth calibration has been re-done or a different artificial mouth has been selected in the meantime, the signal is equalized automatically immediately before the sending distortion test starts.
If “Immediately select this activation signal” is checked, the signal is selected for use when the “Ok” button is clicked.

2.6.13 Show Operator Instructions

If “Show operator instructionsis checked, instructions to the operator are displayed in a window before the measurement starts. The operator may be prompted to position the mobile in a defined way or to set the volume to a certain setting.
Auxiliary delay measurements For the alignment of the acquisition with the arrival of the test signal at the analyzer
input, an auxiliary end-to-end delay measurement is performed before the main measurement.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 31
The item “Auxiliary delay measurements → Configure” allows you to select the measurement method (sine burst or cross correlation), the test signal for the cross correlation method (composite source signal or real speech word “five”) and the re-use of a delay value which has been measured for the same device under test on the same path for a previous test. A maximum age has to be specified for the use of existing delay values to avoid the use of outdated values.
Figure 2-24: Window for configuration of auxiliary delay measurements
The recommended method for auxiliary delay measurement is cross-correlation using the real speech word “five” as test signal. Re-use of existing delay results speeds up the tests, which is particularly helpful in automatic test sequences. It avoids changing of the test setup when uplink and downlink delay values are required for echo loss and stability loss tests. However, if changes are done in the setup, which could have an influence on the end­to-end delay, like changing the device under test, the codec or codec rate etc., the auxiliary delay measurement has to be re-run. If a different device under test is selected in menu item “Data → Test object → Select …” the memorized auxiliary delay values are automatically invalidated. In all other cases, the operator has to invalidate them using menu item “Auxiliary delay measurements → Invalidate all existing values”.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 32

2.6.14 R&S CMU Remote Control

If it is intended to remote control a R&S CMU200 from a sequence of this program, menu item “CMU remote control” can be used to select the communication interface (GPIB or RS-232) and, if multiple devices are found, to select one of them.
Figure 2-25: Window for configuration of the R&S CMU remote control
For remote control via RS-232, an external USB-to-serial converter has to be connected and installed. For remote control via GPIB, the R&S VISA has to be installed. The installer for the R&S VISA is distributed with the Option R&S UPV-K91. For details of the remote control, see 8 Automatic Test Sequences and the operation manuals of the R&S CMU200.

2.6.15 R&S CMU Subsystem

In “CMU subsystem”, one of the communication standards / GSM frequency bands can
be selected. This allows sequences to be run in different GSM bands without changing the subsystem in every single CMU control instance in the sequence.

2.6.16 R&S CMW Remote Control

For clock adjustment in Voice over LTE (VoLTE), and for roundtrip delay measurements using the wideband radiocommunication tester R&S CMW500 as system simulator, the R&S CMW500 has to be remote controlled from the R&S UPV­K9 program. Available remote control interfaces are TCP/IP (LAN), GPIB and USB.
The item “CMW Remote Control” in the options menu allows selecting the device and
the interface to be used. For TCP/IP connections, enter the device name ("CMW50050-", followed by the serial number) or the current IP address of the R&S CMW500. For GPIB connection, use the connector labelled “IEEE 488 CH1” on the backplane of the R&S CMW500.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 33

2.6.17 Noise Calibration Configuration

In “Noise Calibration Configuration”, parameters for the background noise field used for the “Speech Quality in Presence of Ambient Noise” test can be set.
Figure 2-26: Window for configuring the background noise calibration
The configuration of the noise field calibration comprises
Choice of the calibration method
Speaker configuration
Bandwidth and tolerance settings
Other settings
The configuration is saved in the settings of the R&S UPV-K9x program, so that it has to be done at least once and can then be re-used as long as the speaker configuration and calibration methods do not need to be changed.
Equalization Method
Three different equalization methods are available with different trade-off between required measurement time and achieved accuracy:
1. Individual equalization and level alignment for each ambient noise condition
The complete equalization procedure is repeated for each ambient noise condition. This allows equalizing individual responses of each noise condition and obtains best matching equalization results. The required measurement time is the highest of all three alternatives and requires at least eight hours.
Figure 2-27: Setting for the individual speaker equalization of each ambient noise condition
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 34
2. Equalization using pink noise and individual level alignment for each ambient noise
condition The room response is measured using stochastic pink noise. The obtained
equalization is used for all ambient noise conditions. The stochastic nature of the pink noise ensures that the energy is distributed evenly of the complete bandwidth. Required calibration time is about three hours.
Figure 2-28: Configuration of pink noise for speaker equalizations
The level of the pink noise can be chosen between 40 and 70 dBspl(A). For obtaining a similar level as with the noise fields, it should be chosen as 60 dBspl(A) or higher.
The pink noise duration (and averaging time for the measurement) can be chosen between 10 s and 120 s. Recommended duration is 30 s, which equals the averaging time for equalization with real ambient noise condition. An increased averaging time increases the accuracy in particular at the low frequency end.
Alternative to the default pink noise file, a user-defined pink noise file can be loaded. It must be located in the working directory (i.e. “D:\3GPP”) and be at least 120 seconds long.
3. Equalization using a selected one of the ambient noise conditions and individual
level alignment for each ambient noise condition. This method determines the equalization using one of the available ambient noise
conditions as test signal. This equalization is then used for all ambient noise conditions. Only the level is aligned individually for each ambient noise condition. The measurement time is about 3 hours.
Figure 2-29: Selection of ambient noise for the speaker equalization
As the result of this method depends on the spectral distribution of the energy in the chosen noise condition, this method cannot be generally recommended.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 35
Speaker Configuration
Currently only the following two configurations are supported:
1. Four full-range speakers without subwoofer
Figure 2-30: Settings for using four speakers without subwoofer
2. Four full-range speakers with one subwoofer
Figure 2-31: Settings for using four speakers with one subwoofer
The speakers have to be connected as follows:
Full-range speaker front left: R&S UPP generator channel 3
Full-range speaker rear left: R&S UPP generator channel 4
Full-range speaker front right: R&S UPP generator channel 5
Full-range speaker rear right: R&S UPP generator channel 6
The subwoofer (if existing) can be connected to one of the channels 7 … 9
according to the configuration setting.
Bandwidth and Tolerance
The permissible tolerance of the equalized spectrum is stipulated by the standards to be ±3 dB and cannot be changed. For special cases it is, however, possible to raise the lower end of the tolerance template if the room has acoustic drawbacks (standing waves, structure-borne noise) which cannot be improved with acoustic means.
Figure 2-32: Bandwidth and tolerance settings
“Bandlimit low” limits the transmission frequency range of the speaker setup by
applying a highpass filter. This sets the lower end of the equalized frequency range.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 36
It is recommended to suppress frequencies below the 50 Hz third octave, which means to set the bandlimit low to about 44.5 Hz. For other third octaves, the recommended lower band limit is fc / 1.122 with fc being the center frequency of the third octave. The value of “Flatness start at” is automatically adapted to an increased lower band limit if required.
“Bandlimit high” sets the upper frequency limit for the equalization. The default value of
20,000 Hz should not be altered.
“Flatness starts at” sets the lower frequency limit for the tolerance check. The standard
ETSI EG 202 396-1 stipulates a value of 50 Hz.
“Flatness ends at” sets the upper frequency limit for the tolerance check. The standard
ETSI EG 202 396-1 stipulates a value of 10,000 Hz.
Ear Equalization
Combobox “Ear Equalization” allows setting the equalization for both artificial ears to the equalization applied with the recording of the ambient noise condition.
Figure 2-33: Selection of ear equalization
Besides diffuse field equalisation (“DF”) and free field equalization (“FF”), a so-called “independent of direction” (“ID”), equalization can be applied. This equalization curve must be imported for each artificial ear in use with menu item “Calibration → ID (Independent of Direction) Equalization → ... ”.
Figure 2-34: Import of ID equalization data
An average ID equalization curve is provided with file ID_equalization.veq. Alternatively, an individually determined curve can be loaded.
If an equalization curve of the chosen type is missing for one of the artificial ears in use, the calibration procedure is aborted with an appropriate error message.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 37
Speaker Distance Range
The following controls allow specifying a restricted range for the distance between each speaker and the artificial ears of the HATS:
Figure 2-35: Setting the range for the speaker distance
This allows detecting for example swapped ear connections, because the delay from the speaker to the contra-lateral ear takes a detour over wall reflections. The default values are appropriate for the room size specified in standard ETSI EG 202 396-1.
The distance between full-range speakers and artificial ears is determined by delay measurements. The distance between the subwoofer and the HATS must be measured by the operator and entered in the configuration window.
Level Tolerance
The control
Figure 2-36: Control for maximum allowable level deviation
Defines the maximum allowable tolerance for the level adjustment. Higher accuracy usually requires more iteration steps (up to 5), and is possibly not even adjusTable with this maximum number of iterations. Is the required tolerance not reached, the operator is prompted in a message box whether to start more iterations or accept the tolerance. If the adjustment process is to run automatically without operator interaction, this message can be deactivated with checking tick box “Accept single level deviation”.
Figure 2-37: Deactivation of window prompting to accept a level of tolerance
Harmonic Distortion Limit
A violation of the harmonic distortion specified in “Max THD” can be a sign for clipping
in a speaker due to overrange.
Figure 2-38: Entry of THD limit
If a speaker generally has higher distortion, this value can be increased. If a THD value above the specified limit is detected, the operator has the following options:
Reduce the gain settings in the speaker and repeat the measurement.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 38
Cancel the calibration process, reduce the setting in “Maximum output level” and restart the calibration process.
Figure 2-39: Entry of THD limit
Accept the observed THD value and continue without modification of parameters.
Additional Delay Modification for Each Speaker
According to ETSI ES 202 396-1, in all test rooms each loudspeaker signal should be delayed additionally with an individual value in order to increase the diffusity of the noise field and avoid comb filter effects. The operator can modify the individual delays, which are added to the delays determined in the delay measurement step. Maximum allowed value is 50 ms.
Figure 2-40: Entry of additional speaker delay
It is also possible to disable the additional delay. Then only the delays determined in the delay measurement step are applied.
Maximum Iteration Count for Equalization Steps
The following control can be used to set the number of iteration steps for the equalization of individual speakers and of speaker pairs.
Figure 2-41: Entry of maximum iteration count
Additional iteration steps improve the equalization accuracy on the cost of increased measurement time. The iteration process continues until either a flatness better than 1 dB is achieved or the maximum number of iteration steps id reached.
More than three iteration steps cannot be expected to improve the equalization significantly. Maximum input number is 5 iterations. A value of 1 disables iteration altogether. In this case, the equalization function is calculated from the first measured frequency response.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 39
Final Spectrum Check
If the following tick box is activated, the equalization of all ambiences is finally checked, and the flatness and absolute level difference is determined.
Figure 2-42: Checkbox for final equalization test of all ambient noise files
This measurement takes approximately 20 minutes and can also be started after the calibration has been finalized.
The result of this check is stored in the database and appears in the report of the
calibration process. A “Fail” in a single ambience only has influence on the total verdict
if the ambiences had been chosen to be equalized individually (see above).
Refinement of Total Spectrum Equalization
Activating the following tick box enables an algorithm, which tries to improve the result of the total equalization into a tolerance scheme of ±3 dB.
Figure 2-43: Activating refinement of the total spectrum equalization
Figure 2-44: Selection of the total spectrum equalization method
This is attempted by a re-equalization of single speakers. It is possible to choose between the algorithm recommended in ES 202 396-1 and a proprietary algorithm developed by R&S. The standard algorithm requires at least 1 hour in addition to the preceding adjustment process and is not in all cases successful. The R&S algorithm is faster and has so far been successful in more cases. Alternatively, it is recommended to optimize the speaker positioning in the room and re-run the adjustment process.
Minimizing Operator Interaction
After the initial plausibility check and delay measurements it is desirable to run the calibration process without operator supervision. The settings required for this must be made before starting the calibration process.
A “Fail” of the final check of the total equalization may be caused by an unexpected
background noise in the test room. In this case, it is possible to repeat the check to avoid an immediate fail.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 40
The combo box “Action on final flatness deviation” allows you to select the action to be taken in case of a failing final check of the total equalization.
Figure 2-45: Selection of the action performed if the final equalization check fails
“Prompt” leads in case of a fail to a message box where the operator can choose
to repeat the final check
to cancel the calibration process in order to improve the acoustic properties of the
speaker setup in the room, or
continue the measurement with this check being finally failed.
Choose “Retry once” or “Retry twice” for unattended operation.
Report of Intermediate Results
For thorough investigation of a failed adjustment, it may be helpful to get reports of the results of individual and pair-wise equalization of the loudspeakers. This can be activated with the checkbox “Report intermediate steps”.
Figure 2-46: Selection of the addition of intermediate adjustment results to the report

2.6.18 R&S UPP Remote Control

This dialog allows you to detect and connect an R&S UPP with option R&S UPP-B8 on the LAN, to be used as playback system for the background noise generation.
Figure 2-47: Window for connecting to an R&S UPP on the LAN
A click on the “find devices...” button searches the local area network for audio analyzers and fills the combobox next to the button with the computer names and IP addresses of the found devices. Select the appropriate instrument in the combobox and click “Ok”.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 41

2.6.19 Input Switcher

Input Switcher” allows configuring the use of an R&S UPZ input switcher connected to the R&S UPV RS-232 port, to switch analyzer input 2 between sending tests and binaural receiving tests.
If “Use UPZ input switcher for all connections” is activated, all audio signal sources including reference microphones and reference channel 10 of UPP-B8 are connected via the switcher.
Figure 2-48: Switcher window

2.6.20 Report settings

Report settings allows defining the type of information, which should appear in the reports. For details on reporting, see section 8 Automatic Test Sequences below.

2.6.21 Generate Temporary Export Files

If Generate temporary export files is checked, some of the measurement macros create result.exp with result value and verdict, and, if applicable, curve.exp or abscurve.exp and relcurve.exp files with curve data in ASCII format. This function is provided for compatibility with R&S UPL-B9.
R&S UPV-K9/-K91 Preparation and Start of the Application Software
Operating Manual 1402.0043.12 -10 42

2.6.22 Generate Temporary Image Files

Item Generate temporary image files causes the measurement macros to store a screenshot of the measurement window after completion of the test to a file Image.TIF for use by a remote controlling host.

2.6.23 Store Results of Further Measurements

If “Store results of further measurementsis checked, curves and calculated values like loudness ratings are also stored for additional measurements started with the “Add Measurement” key. They will appear in reports of the respective measurements.

2.6.24 Do not Change Scale for Further Meas.

Menu item “Do not change scale for further meas.” deactivates the automatic Y-axis scaling if the curve of an additional measurement leaves the plot area partially or totally.

2.6.25 Store Loaded Curve Data to Results

If “Store loaded curve data to resultsis checked, curves loaded from file with the “Load Curve” softkey are stored to the results database and will appear in reports of
the respective measurements.

2.6.26 Enable Remote Control

Menu item “Enable remote control” activates the interface of the program for remote
control using the client ControlK9.exe.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 43

3 Calibration

3.1 Calibration Devices

R&S UPV-K9 allows you to simultaneously storing calibration values for multiple individuals of the same type. This allows switching devices without the requirement for re-calibration. However, regular re-calibration is recommended in order to assure the correct function of the used devices.
Figure 3-1: Calibration menu
For every device to be calibrated, an entry must be created using “Calibration New device” in the main menu. This menu item opens an entry window to specify the
calibrated device. A category of device must be chosen from the combo box on top. Type, manufacturer and serial number are entered into the text boxes below. The
entries are confirmed by clicking on the “Save and close” button. If the box “Immediately select this device” is checked, the device is selected for immediate use with the “Save, select and close” button.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 44
For ease of use, a “default” device is present and selected for each device type
(category) upon first start of the program.
Figure 3-2: Input window for information about calibrated device
Before a device can be calibrated or used, it must be selected. This can be done from
the “Calibrated Device” input window if the checkbox “Immediately select this device” is checked. Existing devices may be selected with “Calibration Select device” in the
main menu. Clicking on this item opens a window with a Table of all entered devices.
Figure 3-3: Window for selection of calibrated devices
In the combo box lower left, a category of devices must be selected. Subsequently the Table is reduced to available devices in this category. At first startup, there is only a
“Default” device for each category, but when more devices have been created with the “New device” function, a choice will be available in this view. In this state, a row in the
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 45
Table can be marked with a mouse click on its left end. A mouse click on the button “Select” selects this device for the associated usage.
Figure 3-4: Selection window for particular device type
Subsequently this procedure may be repeated for other categories. Finally, all selections are confirmed by clicking the “Ok” button.
A calibrated device may be deleted, using menu item “Calibration Delete device”. The device to be deleted must not be selected for use. Another device of the same type must be selected before the device can be deleted. Select the device to be deleted the same way as in the preceding paragraph the device to be used. Note that all calibration information is deleted for this device except result data required for reporting previous test results.
All calibration data are stored on the R&S UPV hard disk and are therefore automatically available again after every restart. The calibration values for the R&S UPV-K91 option are stored independently of other options. Calibration values for 3GPP tests and CDMA2000 tests are handled separately.
All calibration routines below require as pre-requisite that a device for the respective usage has been generated and selected.
To simplify the procedure for the case that it is not intended to use multiple devices of the same category nor to include information about the used devices into a report,
there is a “Default” device selected for each device type or category after the first start
of the program. An overview of all selected calibration devices together with the calibration values is
given under the menu item “Calibration Show selected devices”.

3.2 Microphone Calibration

1. Insert the measuring microphone fully into the adapter of the sound level calibrator
and switch on the calibrator. A ½” adapter or ¼” adapter is required according to
the microphone diameter.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 46
After inserting the microphone, wait about 10 s to allow for static pressure equalization.
2. Call the test routine with “Calibration Microphone Free field” or “Calibration Microphone Diffuse field” from the main menu, depending on the microphone.
Figure 3-5: Initial window of microphone calibration
For calibrators providing a sound pressure level of 114 dB (10 Pa), the checkbox “Calibration level increased by 20 dB” must be checked. In all other cases, the checkbox must remain unchecked! If a calibrator with a frequency other than 1000 Hz is used, make sure that the “Selective” item in the Calibration menu is unchecked.
The output voltage of the microphone is measured and the sensitivity displayed with reference to 1 Pa. If a mere power supply without gain is used, the displayed sensitivity must approximately match the value in the calibration certificate of the microphone cartridge (typical value for microphone capsule 4134 of artificial ear 4185 is approx. 12 mV/Pa). If a conditioning amplifier with 20 dB gain (recommended value) is used, the displayed sensitivity must be about 10 times higher (e.g. 120 mV/Pa). If the voltage measured is below 3 mV, an error message is displayed. Possible error sources are, for example, a switched-off microphone power supply or a disabled calibrator. In this case, the program requests that the test be repeated. After switching on the microphone power supply, wait approx. 20 s before starting the calibration again.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 47

3.3 Calibration of Artificial Ear

Prior to the measurements, the absolute sensitivity of the microphone in the artificial ear must be determined using a sound level calibrator such as the Brüel & Kjær 4231 with a sound pressure level of 94 dBSPL or a sound pressure of 1 Pa at 1 kHz.

3.3.1 Calibration of Ear Type 1

1. Switch off the microphone power supply.
The 200 V polarization voltage of the microphone may cause a slight electric shock. The current is harmless, but the microphone preamplifier may be damaged.
2. Remove the microphone from the artificial ear.
3. Screw back the microphone capsule and switch on the operating voltage.
4. Insert the microphone fully into the adapter of the sound level calibrator and switch on the calibrator.
After inserting the microphone, wait about 10 s to allow for static pressure compensation.
5. Select “Calibration Artificial ear Type 1” from the main menu.
For calibrators providing a sound pressure level of 114 dB (10 Pa), the checkbox “Calibration level increased by 20 dB” must be checked. In all other cases, the checkbox must remain unchecked!
The output voltage of the microphone is measured and the sensitivity displayed with reference to 1 Pa. If a mere power supply without gain is used, the displayed sensitivity must approximately match the value in the calibration certificate of the microphone cartridge (typical value for microphone capsule 4134 of artificial ear 4185 is approx. 12 mV/Pa). If a conditioning amplifier with 20 dB gain (recommended value) is used, the displayed sensitivity must be about 10 times higher (e.g. 120 mV/Pa). If the voltage measured is below 3 mV or fluctuating by more than 0.2 dB, an error message is displayed. Possible error sources are, for example, a switched-off microphone power supply or a disabled calibrator. In this case, the program requests that the test be repeated. After switching on the microphone power supply, wait approx. 20 s before restarting the calibration.
The reference value measured is stored in a file on the hard disk and used for all subsequent measurements when the same artificial ear type 1 is used as currently selected.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 48

3.3.2 Calibration of Ear Type 3.2 Low Leakage

1. Connect the sound level calibrator tightly to the artificial ear using the adapter DP0939 and switch on the calibrator.
2. Select “Calibration Artificial ear Type 3.2 Low Leak” from the main menu.
For calibrators providing a sound pressure level of 114 dB (10 Pa), the checkbox “Calibration level increased by 20 dB” must be checked. In all other cases, the checkbox must remain unchecked!
The output voltage of the microphone in the ear is measured and the sensitivity displayed with reference to 1 Pa. If the voltage measured is below 3 mV or fluctuating by more than 0.2 dB, an error message is displayed. Possible error sources are, for example, a switched-off microphone power supply or a disabled calibrator. In this case, the program requests that the test be repeated. After switching on the microphone power supply, wait approx. 20 s before restarting the calibration.
The measured reference value is stored in a file on the hard disk and used for all subsequent measurements when the same ear type 3.2L is used as currently selected.
Reading the DRP-ERP Correction Data of the Artificial Ear of Type 3.2L:
The frequency response of the artificial ear of type 3.2L is supplied on a floppy together with the artificial ear. The data is used for transforming the measurement values from the drum reference point to the ear reference point.
1. Connect a USB floppy disk drive with the calibration disk to the R&S UPV, copy the file OES_LL.ADA from the manufacturer’s calibration disk to a USB stick or a CD-ROM or make this file available on a network drive via LAN.
2. Call the routine “Calibration DRP to ERP Correction Type 3.2 Low leak”. Browse to the file OES_LL.ADA and click “Open”.
The calibration file is read. The modified data is stored on the R&S UPV hard disk. This procedure needs only be repeated after a change of the calibration data, e.g. after a
recalibration of the ear by the manufacturer, or when the “3GPP” directory has been
renamed or removed

3.3.3 Calibration of Ear Type 3.2 High Leakage

1. Connect the sound level calibrator tightly to the artificial ear using the adapter DP0939 and switch on the calibrator.
2. Select “Calibration Artificial ear Type 3.2 High Leak” from the main menu.
For calibrators providing a sound pressure level of 114 dB (10 Pa), the checkbox “Calibration level increased by 20 dB” must be checked. In all other cases, the checkbox must remain unchecked!
The output voltage of the microphone in the ear is measured and the sensitivity displayed with reference to 1 Pa. If the voltage measured is below 3 mV or fluctuating by more than 0.2 dB, an error message is displayed. Possible error sources are, for example, a switched-off microphone power supply or a disabled calibrator.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 49
In this case, the program requests that the test be repeated. After switching on the microphone power supply, wait approx. 20 s before restarting the measurement.
The measured reference value is stored in a file on the hard disk and used for all subsequent measurements when the same ear type 3.2H is used as currently selected.
Reading the DRP-ERP Correction Data of the Artificial Ear of Type 3.2H:
The frequency response of the artificial ear of type 3.2H is supplied on a floppy together with the artificial ear. The data is used for transforming the measurement values from the drum reference point to the ear reference point.
1. Connect a USB floppy disk drive with the calibration disk to the R&S UPV, copy the file OES_HL.ADA from the manufacturer’s calibration disk to a USB stick or a CD-ROM or make this file available on a network drive via LAN.
2. Call the routine “Calibration DRP to ERP Correction Type 3.2 High leak”. Browse to the file OES_HL.ADA and click “Open”.
The calibration file is read. The modified data is stored on the R&S UPV hard disk. This procedure needs only be repeated after a change of the calibration data, e.g. after a
recalibration of the ear by the manufacturer, or when the “3GPP” directory has been
renamed or removed.

3.3.4 Calibration of Ear Type 3.3

Calibration of artificial ear type 3.3 and 3.4 is possible with head-and-torso simulators (HATS) 4128 from Brüel & Kjær and HMS II from Head Acoustics. As the calibration level depends on the HATS type, this has to be selected on start of the calibration. For HMS II the calibrator type to be used has to be selected as well.
Follow the instructions of the manufacturer of the HATS for the calibration of the artificial ear. Here the procedure is described for B&K 4128.
1. Remove the pinna from the artificial ear according to the manufacturer’s instructions
2. Connect the sound level calibrator tightly to the artificial ear using the adapter UA­1546 and switch the calibrator on.
3. Select “Calibration Artificial ear HATS Type 3.3” from the main menu.
The type 3.3 ear calibration requires the calibrator with adapter UA-1546 to be held manually against the ear.
For Head Acoustics HMS II, calibrator B&K 4231 must be used in conjunction with the short steel calibration adapter.
If the HATS is installed remote from the R&S UPV, the checkbox “Delay start by …
seconds” can be checked and a time interval can be entered by which the start of the
calibration measurement is delayed. The countdown is displayed on the R&S UPV screen.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 50
Figure 3-6: Ear calibration window for Head Acoustics HMS II with delayed start
The output voltage of the microphone in the ear is measured and the sensitivity displayed with reference to 1 Pa. If the voltage measured is below 3 mV or fluctuating by more than 0.2 dB, an error message is displayed. Possible error sources are, for example, a switched-off microphone power supply or a disabled calibrator. In this case, the program requests to repeat the test. After switching on the microphone power supply, wait approx. 20 s before restarting the calibration.
The measured reference value is stored in a file on the hard disk and used for all subsequent measurements when the same ear type 3.3 is used as currently selected.
Reading the DRP-ERP Correction Data of the Artificial Ear of Type 3.3:
Call the routine “Calibration  DRP to ERP Correction Type 3.3 From ITU-T P.57 Table 2b”. This procedure needs only be repeated when the “3GPP” directory has been renamed or removed.
Alternatively, it is also possible to load individual DRP-ERP correction data from a calibration disk. This option is not conformant with the standards.
Reading the Diffuse Field Equalization Data of the Artificial Ear of Type 3.3:
From Release 10 of 3GPP TS 26.132, diffuse field equalization is applied to receiving
frequency response results with handset and headset UEs. Call the routine “Calibration
Diffuse field equalization Type 3.3 From ITU-T P.58 Table 3”. This procedure needs to be repeated only when the “3GPP” directory has been renamed or removed.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 51
Alternatively, it is also possible to load individual diffuse field equalization data from a calibration disk. This option is not conformant with the standards.
Reading the Free Field Equalization Data of the Artificial Ear of Type 3.3:
For hands-free tests, free field equalization is applied to receiving frequency response
results. Call the routine “Calibration Free field equalization Type 3.3 artificial ear From ITU-T P.58 Table 2”. This procedure needs only be repeated when the “3GPP”
directory has been renamed or removed. Alternatively, it is also possible to load individual free field equalization data from a
calibration disk. This option is not conformant with the standards.

3.3.5 Calibration of Ear Type 3.4

The calibration of the HATS ear with the alternative type 3.4 pinna is done likewise, similar to the type 3.3 ear calibration.

3.4 Calibration of Artificial Mouth for Handset Tests

Before a mobile phone can be tested, the absolute sensitivity and frequency response of the artificial mouth have to be measured and corrected with the aid of a previously calibrated free field or diffuse-field (pressure-field) microphone. The measuring microphone removed from artificial ear type 1 can be used for this purpose or an additional microphone capsule is screwed to the microphone preamplifier. The standard microphone is used as a reference for determining the frequency response of the mouth. The frequency response of the microphone can be ignored in the test frequency range (100 Hz to 8 kHz) (see also calibration certificate of microphone capsule).
Since interfering sound falsifies the corrections, the artificial mouth must be calibrated in the anechoic and isolated test chamber. In order to reject any noise present in the test chamber, it is recommended to activate item “Selective” in the calibration menu.
First, a calibrated measuring microphone has to be selected.
1. Select a reference microphone type with “Calibration Artificial mouth Select reference mic”.
For the selected microphone type, a device must be selected and calibrated (see above). This device must be connected to R&S UPV analyzer input 1 via power supply/conditioning amplifier.
For calibration of a P.51 type artificial mouth using a diffuse field type microphone (e.g. B&K 4131 or 4134) or a microphone from a Type1 artificial ear, fit the microphone at right angles to the mouth at the mouth reference point (MRP) using the gauge supplied with the mouth (positioning at right angles is necessary because diffuse field or pressure-calibrated microphones have a flat frequency response to sound from random incident direction and therefore exhibit an emphasis on high frequencies with frontal sound incidence).
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 52
For calibration of a P.51 type artificial mouth using a free field microphone, the microphone must be mounted in the axis of the sound outlet of the artificial mouth.
For calibration of the HATS (P.58 type) artificial mouth, a ¼” microphone is clamped in the fixture attached to the HATS. For the super-wideband frequency range a diffuse field or pressure field microphone is required.
2. Call the calibration routine with “Calibration Artificial mouth LRGP ITU-T P.51 Without Reference Spectrum”, “Calibration Artificial mouth LRGP ITU-T P.51 with P.50 Ref. Spectrum” or “Calibration Artificial mouth LRGP ITU-T P.51 with P.501 NB/WB ST Speech Ref. Spectrum” or “Calibration  Artificial mouth LRGP ITU-T P.51 with P.501 SWB/FB ST Speech Ref. Spectrum”, “Calibration Artificial mouth HATS ITU-T P.58 Without Reference Spectrum” or “Calibration  Artificial mouth HATS ITU-T P.58 with P.50 Ref. Spectrum” or “HATS ITU-T P.58 with P.501 NB/WB ST Speech Ref. Spectrum” or “HATS ITU-T P.58 with P.501 SWB/FB ST Speech Ref. Spectrum” from the main menu.
The sound pressure generated at the MRP is set to exactly -4.7 dBPa in an automatic measurement routine at 1 kHz. The generator voltage required is stored in a file on the hard disk and used as a reference for all subsequent settings with the same artificial mouth. If the sound pressure cannot be adjusted to -4.7 dBPa, an error message is displayed with a request to check the connection of the artificial mouth and to repeat the measurement. A possible error source would be that the transformer supplied is not connected between the generator and the artificial mouth.
The uncorrected frequency response of the artificial mouth is measured and displayed. Next, the frequency response is measured with the inverse frequency response correction automatically selected in the generator (equalization). Residual errors caused by nonlinearities of the speaker in the mouth are measured and taken into account in the final equalization file as fine correction.
To verify the results, the absolute sound pressure versus frequency is measured at a sound pressure of 4.7 dBPa (reference value for most of the measurements). The absolute sound pressure at each frequency must be within a tolerance band of 0.2 dB. Correct calibration without interfering sound yields an almost straight line in the middle between the two limit lines.
If mouth calibration with reference spectrum measurement is chosen, the test signal is subsequently filtered with the inverse mouth frequency response, and the resulting spectrum at the MRP is recorded as reference for sending frequency response tests.
If a mouth calibration is performed without reference spectrum measurement, previously recorded reference spectra are invalidated. A sending measurement using artificial voice according to ITU-T P.50 can only performed after an additional reference spectrum calibration (see below).
To calibrate the artificial mouth for all different sending tests available, proceed as follows:
1. Choose the largest required frequency range in “Calibration Artificial Mouth Frequency Range”. Smaller frequency ranges are included in larger ones and do not have to be calibrated separately.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 53
2. Perform the artificial mouth calibration for hands-free tests, which includes the MMS reference spectrum calibration and mouth calibration for handset tests.A separate calibration for handset tests is not required.
3. Perform the sending reference spectrum calibration for artificial voice according to ITU-T P.50 and for real voice according to ITU-T P.501. The reference spectrum calibration has to be performed separately for handset tests and hands-free tests, and for NB / WB frequency range and SWB / FB frequency range, respectively.

3.5 P.50 Speech Spectrum Calibration

For tests using artificial voice according to ITU-T P.50 as test signal, the spectrum of the test signal has to be measured and stored as reference spectrum for the transfer function (gain) calculation. For sending and sidetone tests, the signal has to be filtered in addition with the inverse frequency response of the artificial mouth. A calibrated reference microphone must be placed at the Mouth Reference Point (MRP) for this purpose. It is recommended to perform the reference spectrum calibration directly after the mouth calibration, using sub-menu item “Calibration Artificial Mouth With P.50 Ref. Spectrum”.
The reference spectrum calibration for the receiving direction does not require external equipment or wiring. The reference signal at the output from the R&S UPV generator to the speech input of the system simulator is measured via internal connection to the R&S UPV analyzer input. For narrow-band tests in receiving direction, a band-limited version of the test signal is used. Each version of the test signal has to be calibrated separately.
For ambient noise rejection according to Release 10 of 3GPP TS 26.132, a separate speech reference spectrum calibration is necessary because the active speech level at the MRP is specified to be +1.3 dBPa instead of -4.7 dBPa for this test. The test routines can be found in sub-menu “Calibration __> P.50 Speech Ref. Spectrum Sending for ANR”.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 54
Figure 3-7: Submenu for P.50 reference speech spectrum calibration

3.6 P.501 Speech Spectrum Calibration

From Release 11 of 3GPP TS 26.132 real speech from ITU-T P.501 is used for frequency response and loudness rating tests. Like the artificial voice according to ITU-T P.50 the test signal has to be equalized according to the frequency response of the artificial mouth, and the spectra at the mouth reference point and at the encoder input to the system simulator have to be measured and stored as reference. All respective calibration functions can be found in the sub-menu of menu item “Calibration --> P.501 ST Speech Ref. Spectrum.
For the receiving direction, the test signal has to be filtered according to the codec bandwidth. Therefore separate reference spectra have to be stored for narrowband, wideband, super-wideband and fullband in receiving direction.
In sending direction, the signal bandwidth is wideband for narrowband and wideband tests, and super-wideband for super-wideband and fullband tests. Furthermore, separate reference spectra have to be stored for handset and headset tests on one side and hands-free test on the other side, as the level adjustment is different for both cases. Note that the wideband reference spectrum is required for sidetone tests, even with super-wideband and fullband codecs.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 55
Figure 3-8: Submenu for P.501 reference speech spectrum calibration

3.7 Calibration of CMU Speech Codec

The calibration of the CMU speech coder and decoder is necessary to be able to calculate absolute loudness. Calibration has to be performed only once and must be repeated only if the R&S CMU200 used is replaced. If the R&S CMU200 is equipped with model 14 of the Link Handler R&S CMU-B21, GSM as well as WCDMA use the same paths. Calibration of the coder is therefore identical for both operating modes and need not be repeated when switching from GSM to WCDMA or vice versa.
If GSM and WCDMA use different link handlers, separate calibration devices have to be generated for the two link handlers (i.e. with and without the supplied adapter).
Note that encoder and decoder are handled as separate devices but calibrated with one routine.
Auxiliary settings required for calibration can be found in the R&S CMU200 under Bit Stream (for GSM) and under BS Signal, Dedicated Channel, Voice settings (for WCDMA) (firmware version 4.52 or higher). Call the calibration routines with “Calibration → Codec → CMU/CBT” from the main menu.
The following information is displayed:
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 56
Figure 3-9: Message box during decoder calibration
Set up a call to the mobile phone. Set bit stream or voice setting on the R&S CMU200 to "Decoder Cal" and then click the “OK” button.
The actual voltage at the decoder output of the R&S CMU200 is now measured for a digital full-scale signal and the required correction value is calculated and saved in the R&S UPV. The following request is then displayed:
Figure 3-10: Message box during encoder calibration
After the “OK” button has been clicked, the input sensitivity of the speech coder is
measured and the input voltage required for digital full scale is measured at the speech coder and saved in the R&S UPV.

3.8 Calibration of CMW Speech Codec

The R&S CMW500 does not provide the “Decoder Cal” and “Encoder Cal” functionalities. Instead, the maximum peak input and output voltage can be set in the user interface.
Figure 3-11: Setting of the full-scale peak input and output values in the CMW user interface
“Calibration → Codec → CMW” opens windows where these values can be entered in
unit mV in the mobile phone test program.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 57

3.9 Calibration of Ambient Noise Field

This section applies only to the pink noise required for the “ambient noise rejection test”. For calibration of noise ambiences for the “speech quality in presence of background noise” test, see section 3.10.
Figure 3-12: Measurement window for ambient noise field calibration
Please refer to application note 1GA51, available on the R&S download web site, for suggestions how to generate the noise field for the ambient noise rejection test. For the calibration of the noise field for ambient noise rejection tests, all obstacles (test head or HATS, Telephone etc.) have to be removed from the test chamber. For the generation of the noise field, additional equipment is necessary. The noise field should have sufficient homogeneity (sound pressure level independent of place) and diffuseness (randomly incident sound at the place of the microphone). This can be achieved either in a reverberation room with omni-directionally radiating sound source or in an anechoic room with a number of uncorrelated noise sources. Limited homogeneity can be achieved in the centre between two speakers. For good diffuseness a minimum of four speakers distributed in different spatial directions are required.
A diffuse field or pressure field microphone must be positioned at the spot of the mouth reference point of the (removed) test head or HATS.
The menu Item Calibration Ambient noise field opens the window shown in Figure 2-27.
The button “Adjust Spectrum” starts a continuous spectrum measurement, which allows you to adjust the spectrum of the noise, e.g. using equalizers. The template is centered around the curve irrespective of the absolute level.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 58
Figure 3-13: Measurement window for spectrum adjustment
The continuous spectrum measurement can be aborted by clicking “Stop Adjustment”. The button “Adjust level” starts a continuous level measurement. If a multitude of
sources is used, each source must produce a sound pressure level of -24 dBPa – 3 * LOG n dB, whereby n is the number of noise sources.
Figure 3-14: Measurement window for level adjustment
The left thermometer column gives a coarse overview. The right column has an enlarged scale for fine adjustment. The numeric field in the center shows the numeric value.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 59
The continuous level measurement can be aborted by clicking “Stop Adjustment”. Once the noise field has been adjusted, the result can be documented by pressing
“Start Measurement”. Although there is no calibration value to be used in the ambient
noise rejection measurement, the level is stored and can be included for reference in the measurement report.
Figure 3-15: Result of ambient noise calibration
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 60

3.10 Calibration of Noise Field for “Speech Quality in Presence of Background Noise” Test

3.10.1 Connections

Figure 3-16: Test setup and connection of external components for the adjustment of the background noise field
Connect
1. LAN socket of R&S UPV to LAN socket of R&S UPP
2. R&S UPP trigger output to R&S UPV trigger input
3. R&S UPP-Z8A output 3 to input of front left active speaker
4. R&S UPP-Z8A output 4 to input of rear left active speaker
5. R&S UPP-Z8A output 5 to input of front right active speaker
6. R&S UPP-Z8A output 6 to input of rear right active speaker
7. R&S UPP-Z8A configured subwoofer output (by default 7) to input of active subwoofer
8. Right artificial ear to R&S UPV analyzer input 1 via microphone power supply
9. Left artificial ear to R&S UPV analyzer input 2 via microphone power supply For the delay measurements, you will be prompted to modify the input connections: a
connection has to be made from R&S UPP-Z8A output 10 to R&S UPV input 1, and both artificial ears have to be connected subsequently to R&S UPV analyzer input 2.
Optional an R&S UPZ input switcher can be used; see chapter 5.21.3 Switcher Support.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 61

3.10.2 Establishing the Remote Control Connection between R&S UPV and R&S UPP

After power up of both instruments and the interconnection with a LAN switch cable, first an IP connection has to be established between R&S UPV and R&S UPP.
1. Start the R&S UPV-K9x program and open window “Options → UPP remote
control”.
2. Once the “LAN” LED on the R&S UPP is on, press button “CASCADE / LAN
RESET” once and wait until the “CASCADE” LED is blinking.
3. Press the “CASCADE / LAN RESET” button again.
4. Click the “Find” button in the R&S UPP remote control window on the R&S UPV.
5. Now the computer name on the R&S UPP appears in the R&S UPP remote
control window. Select it and close the window with “Ok”.

3.10.3 Prerequisites

The calibration of both artificial ears of the HATS is required for performing the calibration of the background noise field.

3.10.4 Configuration

The configuration of the noise field calibration comprises
Choice of the calibration method
Speaker configuration and setup
Bandwidth and tolerance settings
Switcher usage
Other settings
It is done in the 'Noise Calibration Configuration' window. This window is opened with item 'Noise calibration configuration' in the “Options” menu. The Figure below shows the default settings; if any item has been changed you can retain the default settings by selecting the button "Set to default".
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 62
Figure 3-17: Noise Calibration Configuration Window, Default state
The configuration is saved in the settings of the R&S UPV-K9x program, so that any modification will be effective and re-used as long as the speaker configuration and calibration methods do not need to be changed.
3.10.4.1 Choice of Calibration Method
Three different calibration methods are available that differs in time consumption and accuracy:
Pink Noise Equalization (recommended by Rohde & Schwarz) equalizes speaker and room with an uncorrelated, non-periodical Pink Noise signal and uses this equalization for all ambiances. It offers the best compromise between calibration time and accuracy. After room equalization with the pink noise signal, only an individual level correction has to be done for each ambiance.
Figure 3-18: Configuration Window, calibration method Pink Noise
Although it is not recommended to change the default settings, you can select level and duration of the Pink Noise; specifying a longer duration results in longer calibration time but might increase the accuracy especially at low frequencies. Instead of using the recommended WAV file, you can provide an own noise file.
Equalization with one shared ambiance (recommended by ETSI ES 202 396) equalizes speaker and room with one of the ambiances and uses this equalization for
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 63
all other ambiances, too. Because the real ambiances do not contain the same energy in each frequency band, the equalization is optimal only for the selected ambiance.
Figure 3-19: Configuration Window, calibration method shared ambiance
This method takes the same calibration time as Pink Noise but normally offers worse accuracy. After room equalization, the individual level correction has to be done for each ambiance.
Individual equalization of each ambiance results in best equalization accuracy but makes the calibration process more than 3 times longer (about 8 hours). It could be considered if the room and speaker setup will never be changed.
Figure 3-20: Configuration Window, calibration method individual ambiance
Speaker Configuration and Setup
Currently two different speaker configurations are allowed for background noise generation:
Figure 3-21: Configuration Window, speaker configuration
4 full range speakers (2 pairs) with 1 subwoofer or 4 full range speakers without subwoofer. A subwoofer must be used if the full range speakers do not pass frequencies down to 50 Hz. The full range speakers are connected to R&S UPP channel 3.6; the optional subwoofer is connected to channel 7. R&S UPP channel 10 is reserved for connection to R&S UPV output.
If a subwoofer is used then the cross over frequency between full range speakers and subwoofer can be adjusted:
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 64
Figure 3-22: Configuration Window, speaker crossover frequency
To avoid overdriving the speakers' (or amplifiers') maximum input voltage, you can adjust the R&S UPP maximum (RMS) output level:
Figure 3-23: Configuration Window, speaker maximal output level
At the beginning of the calibration process checks for each speaker are performed to achieve a minimum sound preasure level and to detect speaker overdrive by measuring THD. You are instructed to increase the speaker (or amplifier) gain if the required sound level could not be reached. The required Sound Preasure Level could be set in the range from 80 to 99 dBspl:
Figure 3-24: Configuration Window, speaker Min Sound Pressure Level
For speakers that generate higher harmonics than 10% (-20 dB) the tolerable THD can be increased (to e.g. -10 dB):
Figure 3-25: Configuration Window, speaker Max THD
To detect left/right speaker connection faults the calibration process need to know the rough speaker-to-microphone distances. If your room is much larger or smaller than the recommended size, you have to adjust the minimum or maximum speaker distance:
Figure 3-26: Configuration Window, expected HATS to speaker distance
Bandwidth and Tolerance Settings
The bandlimit specifies the frequency range where the equalization will be performed.
Figure 3-27: Configuration Window, Bandlimit
The 'low' value additionally enables a highpass filter. To include the lowest third octave (50 Hz) the lower bandlimit should be lower or equal to a half-third octave (f / 1.122) below. To exclude the 40 Hz third octave from equalization the bandlimit low should be set to 44.5 Hz
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 65
Flatness check is limited to the frequency range between start and end.
Figure 3-28: Configuration Window, Flatness range
According to ETSI ES 202 396 the equalization must be flat to +/-3 dB, thus the flatness width must not be greater than 6 dB.
Figure 3-29: Configuration Window, Flatness width
To achieve an even better flatness this value might be reduced by 1 or 2 dB, if room and speaker setup are able to fulfil such narrow flatness.
Switcher Usage
To reduce interaction during calibration and measurement, an R&S Input switcher R&S UPZ can be used. This allows hard wired connection to be established between each signal source and the R&S UPZ while the calibration process automatically switches the required R&S UPZ input channels to R&S UPV input
Figure 3-30: Configuration Window, Use Input Switcher R&S UPZ
Other Settings
All other configuration settings normally need not to be changed. Modify these settings only to change the default behaviour of the calibration.
Figure 3-31: Configuration Window, ear equalization
Instead of using ID ear equalization (Independent of Direction) you can select 'free field (FF)' or 'diffuse field (DF)' ear equalization.
Figure 3-32: Configuration Window, Modify speaker delay
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 66
To avoid coherent superposition of sound from each speaker the measured speaker delay is individually increased. Only experienced users should modify these settings to fix acoustical problems!
Figure 3-33: Configuration Window, Max iterations during equalization
To achieve flat frequency response several iterations of equalization are performed for single and pair of speakers. Iteration is terminated if flatness is better than 1 dB or 'Max iterations during equalization' is reached. This value can be selected between 1 (no iteration) and 5.
Figure 3-34: Configuration Window, Action on final flatness deviation
This item specifies how the calibration process should behave if – after all iteration steps – the flatness deviation is higher than 'Flatness width'. By default it prompts you how to proceed (retry, ignore or abort). To avoid this interruption of calibration process you can predefine if (and how often) the measurement should be retried.
Figure 3-35: Configuration Window, Max level deviation
During level adjustment, the sound pressure level of each speaker is iterated to the requested value. This iteration terminates if the level deviation is below the selected 'Max level deviation'. To achieve higher level accuracy this value can be decreased; however, this might increase the count of iteration steps and total calibration time.
Figure 3-36: Configuration Window, Ignore uncritical warnings
To avoid interruption of calibration process by default, only errors are reported; warnings about uncritical deviations (e.g. noise floor, flatness of unequalised room, level deviation during single speaker adjustment) are suppressed, but entered in report.
To get uncritical warning as soon as they appear this tick box has to be deactivated.
Figure 3-37: Configuration Window, Use All speaker Equalization
If the sound super positioning of all speakers degrades the flatness to more than 6 dB (or the selected 'flatness width') a post equalization process should be used to improve this flatness. Otherwise, the calibration would fail!
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 67
Instead of using the R&S recommended algorithm the post equalization can be done according to ES 202 396-1.
Figure 3-38: Configuration Window, Final test of all ambiances
After successful calibration a final level and equalization test of all ambiances should be performed. Additionally this test adjusts the total level setting to minimum deviation.
Disabling this test reduces the total calibration time. The final test of successfully calibrated noise field (but without level optimization) can
be performed separately from the calibration process by executing the calibration and skipping all calibration steps.
Figure 3-39: Configuration Window, report intermediate steps
In case of problems during the calibration process, you can activate a detailed reporting that logs all intermediate steps.
Selecting 'Report intermediate steps' does not increase total calibration time but the pages of report.

3.10.5 Switcher Support

If 'Input Switcher UPZ' is enabled in 'Noise Calibration Configuration', the calibration process will try to connect all required signal sources automatically to R&S UPV input instead of prompting you to change cabling. This presumes:
R&S UPV is connected to R&S UPZ via serial cable (RS232) OR USB cable (do
not connect USB if you want to use RS232!);
USB-to-serial device driver is installed (only if using USB connection);
R&S UPZ is powered ON (green 'ON'-LED on switcher front panel)
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 68
Figure 3-40: Test setup and connection of external components with input switcher UPZ for the adjustment of the background noise field
When starting the calibration process, a dialogue is displayed to instruct you how to connect the input switcher's analogue input and output connections. The connections to R&S UPZ input channel 3, 4 and 5 are not necessary for the calibration, but will be needed for later measurement (note: only 1 microphone need to be connected to input channel 4 or 5, respectively).
Figure 3-41: Switcher support, analogue connection instructions
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 69
Or – if switcher could not be detected – how to establish the control connection from R&S UPV to R&S UPZ:
Figure 3-42: Switcher support, control connection instructions
If the switcher cannot be enabled at this point, then 'Cancel' will disable switcher for the current calibration; cabling has to be done manually. The corresponding setting in configuration window will not be changed. On next start of calibration, the switcher will be tried to use again.
The calibration process checks the connection to R&S UPZ whenever the switcher needs to be operated. If switcher control fails while calibration is running (e.g. by unplugging the control or power cable), you are prompted to establish the analogue connections manually. This allows the calibration to terminate correctly:
Figure 3-43: Switcher support, error message if switcher failed during calibration
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 70

3.10.6 Selection of Background Noise Signals

According to 3GPP TS 26.132 different sets of background noise signals have to be used for handset tests and handheld handsfree tests, respectively. Background noise signals for handheld devices are distributed with ETSI standard ES 202 396-1. For test of handheld handsfree devices the signal files distributed with ETSI TS 103 224 have to be used.
As each noise signal is prepared and adjusted individually during the calibration process, the signals to be adjusted have to be selected with the start of the calibration process.
Sub-item For Handset only (ES 303 396-1)” starts calibration for handset tests with the signals from ES 202 396-1. Sub-item “For HHHF only (TS 103 224)” starts calibration for handheld handsfree tests with the signals from TS 103 224. Sub-item For Handset and HHHF” starts calibration of all noise signals.
Note that with the successful completion of a new calibration all previous calibration results are void. It is not possible to add calibration for more signals to an existing calibration result. In order to calibrate different sets of noise signals separately, use separate calibration devices (see section 3.1). Generate new devices with item “New
Device” of the Calibration menu. Switch to the desired device with menu item “Select Device” in the calibration menu before starting the calibration process, and before
starting the tests with which the respective calibration result is to be used.

3.10.7 Starting the Calibration Process

”Select Reference Mic” allows to choose between free-field microphone and diffuse­field microphone to be used during the test for recording the noisy signal at the microphone input of the UE under test. The software for test execution will use the calibration value of the respective microphone type to select the analyzer range. For SWB tests this microphone is not required, as the modified evaluation method does not use the noisy signal as input.
For the calibration process this selection has no influence, since the adjustment is performed with the signals from the artificial ears of the head-and torso simulator.
Calibration is started by selecting the appropriate sub-item of 'Office Room':
Figure 3-44: Starting the background noise calibration process
For handset tests eight noise scenarios from ETSI ES 303 396-1 are calibrated. The first scenario (“Pub Noise”) is also used for the test of RLR in presence of background noise. For tests of hands-free devices in background noise, the standard specifies five scenarios provided with ETSI TS 103 224. Therefore a complete calibration for handset and HHHF comprises thirteen noise scenarios. For reasons of calibration time it is possible to select sub-sets for handset, handheld hands-free and RLR in presence of background noise only.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 71
After selecting the desired set of noise scenarios to be calibrated, a window with cabling instructions appears first:
Figure 3-45: Cabling instructions for the speakers
After establishing and checking the connections, click “OK” to start the calibration process.

3.10.8 Stopping and Resuming the Calibration Process

The calibration process can be stopped at any time. All of the calibration data recorded up to this point are then available and can be viewed in the report.
To abort a running calibration (e.g. to improve the loudspeaker positions) is possible in the following ways:
If a dialog box appears, click the "Abort" or "Cancel" button
Figure 3-46: Dialogue with 'Abort' button to abort calibration
Figure 3-47: Dialogue with 'Cancel' button to abort calibration
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 72
If measurement is running, click the "Cancel" button and confirm with 'Break':
Figure 3-48: Window during running measurement and 'Cancel' confirmation dialogue
After calibration had been stopped a dialog box informs you about the calibration section that was aborted.
Figure 3-49: 'Calibration aborted' message, e.g. when level adjusting Outside_Trafffic_Road
A calibration that had been aborted can be resumed at certain resuming points. When restarting a calibration that partly had been finished the calibration process prompts you at every resuming point
to resume the calibration at this point (recalibrate this and the following)
to skip this calibration section
to abort the calibration
By skipping all successfully terminated calibration sections, the calibration process will be resumed at the beginning of the section where it had been aborted. For details, see chapter 'Calibration Sections and Resuming Points'.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 73

3.10.9 Preparatory Measurements for Delay Measurement

Before the delay measurement is performed, additional plausibility checks are carried out at each speaker to ensure that the wiring is correct and that the sound pressure level is high enough:
Measurement of Sound Pressure
The unweighted sound pressure level should reach at least 80 dBspl (or the value selected as 'Min sound pressure level' in configuration window, if modified) at normal drive level.
Figure 3-50: Message for insufficient level
If this sound pressure is not reached, an error message appears which cannot be ignored. After the gain factor has been increased, the measurement can be repeated (by selecting 'Retry').
Measurement of Level Change at Microphone Amplifier Output
During this measurement, two different levels are successively applied to each speaker. The output at the microphone amplifier (measured by R&S UPV input channel
2) is expected to change by the same amount (5 dB).
Figure 3-51: Message for insufficient sound pressure level increase
If the amplifier output does not change with the R&S UPP output level, this may have the following causes:
the microphone amplifier is switched off
the HATS ear is not connected to the correct amplifier channel
the microphone amplifier is not connected or is incorrectly connected to R&S UPV
input channel 2
the desired speaker is switched off
the desired speaker is not connected to the correct R&S UPP output channel
the requested sound pressure level cannot be output by the speaker (speaker
gain to high or 'Maximum output level' to high),
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 74
After the error has been rectified, the measurement can be repeated by selecting 'Retry'. If the configuration has to be changed, then the calibration must be aborted ('Cancel' button).
Measurement of Level Change at Reference Input
During this measurement, two different levels are successively applied to R&S UPP output channel 10. The input level at UPV input channel 1 is expected to change by the same amount (5 dB).
Figure 3-52: Message for insufficient level change at reference input
If the reference input does not change with the R&S UPP output level, there is no connection between R&S UPV channel 1 and R&S UPP output channel 10.
After the error has been rectified, the measurement can be repeated by selecting 'Retry'.

3.10.10 Delay Measurement

The delay measurement determines the time it takes for the sound to travel from the speaker membranes to the left and right ear of the HATS. The travel time and therefore, the distance of the full-range speakers from the HATS is measured by means of cross-correlation of the acoustic (sound pressure) signal and the electrical reference signal.
Since the delay measurement is quite complicated owing to the required cabling modifications, it can be skipped if a valid delay calibration is already available and if the speaker layout and the HATS position have not been changed in the meantime. The latter is not recognizable by the calibration process; it is therefore the responsibility of you to repeat the delay calibration if necessary.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 75
Figure 3-53: Message querying the start of the delay calibration
To perform the delay measurement, the reference signal from channel 10 of the R&S UPP generator is connected to analyzer channel 1 of the R&S UPV, and the signal from the respective ear is connected to analyzer channel 2 (via a preamplifier). Which ear is currently being measured and therefore must be connected (to R&S UPV input channel 2) is shown in dialog boxes. If the R&S input switcher R&S UPZ is used, then the connection will be switched automatically without any interaction.
Figure 3-54: Instruction for connections during delay measurement
This box must be confirmed after reconnection of the R&S UPV analyzer; then the measurement and the generator signal are started.
Connection problems can be determined from the measured value obtained during the delay measurement:
Checking of the delay differences between the measurement channel and
reference channel
Figure 3-55: Plausibility check for the delay result
If the reference channel has a greater delay than the measurement channel, it is not connected to the reference signal but instead to the second channel of the microphone amplifier.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 76
Distance between the speaker and associated HATS ear is too great
Figure 3-56: Message for unexpectedly long delay
If the measured distance is greater than the intended maximum distance, the longer indirect sound path to the ear located opposite may possibly have been measured. In this way, it is possible to detect inadvertent swapping of the HATS ears. The 'Max distance HATS to speaker' can be adjusted in the configuration window after cancelling the calibration.
Distance between the speaker and associated HATS ear is too small
Figure 3-57: Message for unexpectedly short delay
If the distance is less than the intended minimum distance, the speaker is too close to the HATS. This warning can be ignored, if the measured distance matches the actual position of the speaker. The 'Min distance HATS to speaker' can be adjusted in the configuration window after cancelling the calibration. The measurement can be repeated after the cabling or the amplifier level has been corrected.
If the R&S Input switcher UPZ is used, then the connections are automatically switched without any interaction. Otherwise, the cabling must be changed during and after the delay measurement; you are is instructed to do this in dialog boxes.
Figure 3-58: Instructions for change of cabling during the calibration process
The distance between HATS and subwoofer cannot be measured but must be entered manually. This value is not critical for the measurement and can be given with an accuracy of 10 cm.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 77
Figure 3-59: Delay measurement, manual entry of subwoofer distance
After completion of the delay measurement, it is thus ensured that the HATS ears are correctly cabled and the output level is high enough. A list showing all of the determined distances is displayed before the actual equalization begins.
Figure 3-60: List of determined distances

3.10.11 Preparatory Measurements for Equalization

Equalization can only be performed successfully if the room acoustics meet certain requirements.
Noise Floor
The accepTable noise floor is max. 30 dB(A)spl. If it is above this value, equalization may fail in the case of weak noise fields.
Figure 3-61: Warning for exceeded noise floor limit
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 78
This warning is suppressed if 'Ignore uncritical level warnings' is activated in configuration window. After the soundproofing has been improved, the measurement can be repeated. "Ignore" continues the calibration process although this limit has been exceeded.
Frequency Response of Unequalized Speakers
The frequency response of each speaker before equalization is measured in the range of 40 Hz to 20 kHz (or the range selected as 'Bandwidth low/high' in configuration window, if modified).
The acceptable flatness deviation for full range speakers is max. ±9 dB in the range 50 Hz to 10 kHz (or the range selected as 'Flatness start/end at' in configuration window, if modified).
The graph displays the frequency range that is to be equalized, the Y axis is scaled from -20 to +20 dB. The limit curve shows the range that is checked for flatness.
Figure 3-62: Frequency response of unequalized full range speaker
If the flatness deviation is above this value, equalization is hard to achieve for the room and the room properties and the speaker layout should be improved before equalization is performed.
Figure 3-63: Warning for excessive equalization needs
'Ignore' continues the calibration process although this limit has been exceeded. This warning is suppressed if 'Ignore uncritical level warnings' is activated in
configuration window.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 79
The final curve for each speaker and the flatness result for each fullrange speaker are added to the report.
The subwoofer frequency response can be used to determine the cutoff frequency, if unknown. If it is below the default value of 120 Hz, then the actual cutoff frequency must be modified in the configuration window. If the subwoofer does not pass frequencies down to 40 Hz, then the lowest third octave should not be equalized.
In this case, set the Bandlimit Low in configuration window to 50 Hz. If it does not pass 50 Hz, then this subwoofer is not appropriate for this calibration.
Figure 3-64: Frequency response of unequalized subwoofer
Total Harmonic Distortion
Measurement of the total harmonic distortion is performed for each speaker at nominal level in order to prevent speakers from being overdriven. Since destructive interference of the fundamental can occur at discrete frequencies which then results in a poorer total harmonic distortion being determined, the measurement is repeated at different frequencies in the case of an error.
Figure 3-65: Warning for excessive total harmonic distortion of a speaker
If speakers with a high total harmonic distortion are used, the tolerance limit 'Max THD' must be set higher (e.g. -10 dB) in the configuration window.
Cabling Check
This measurement which was already performed during the delay measurement is intended to ensure that the cabling is correct if the delay measurement is skipped and
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 80
after the cabling has been changed. The results are the same as those under "Measurement of Level Change at Microphone Amplifier Output".

3.10.12 Shared Equalization Method

The fastest way to establish a valid calibration is to equalize room and speakers with only one signal – preferably Pink Noise – and use the equalization data for all ambiances. Afterwards the ambiances are level adjusted individually. This is the default setting; in configuration window you can select any of the ambiances as equalization signal alternatively (see chapter 3.10.4.1 Choice of Calibration Method).
At the beginning of equalization, a check is performed to determine whether a valid equalization data set is available. If this is the case, equalization can be skipped (by selecting 'No') and the calibration process continues with individual level setting:
Figure 3-66: Valid shared equalization, recalibration prompt
Once started the equalization as well as the subsequent level adjustment are largely performed without input. The only time input is required is if errors occur and you are is asked what should happen next.
During calibration process, the equalization result graph – if equalization had been performed – and the values of the level adjustment are stored and can be reviewed in the report after termination or abort of calibration.

3.10.13 Individual Equalization Method

Alternatively, the calibration process can generate individual equalization data for each ambiance. Depending on the room and speaker characteristic, this method could generate even better flatness, but it is very time consuming. Each ambiance will first be level adjusted and then equalized. Before starting the calibration process, select this calibration method in the configuration window (see chapter 3.10.4.1 Choice of Calibration Method).
For each ambiance, a check is performed to establish whether a valid calibration data set is available. If this is the case, equalization of this ambiance can be skipped by selecting 'No':
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 81
Figure 3-67: Valid individual equalization, recalibration prompt
Once started the level adjustment and equalization are largely performed without input. The only time input is required is if errors occur and you are asked what should happen next.
During calibration process each measured equalization result graph and the values of the level adjustment are stored and can be reviewed in the report after termination or abort of calibration.

3.10.14 Level Adjustment

Level adjustment is performed with a maximum of five iteration steps. Normally two steps are sufficient to achieve the desired level tolerance (< 0.3 dB, can be modified as 'Max level deviation' in the configuration window). This number may not be sufficient for the subwoofer, especially for ambiances with weak bass. A message is then displayed where you can choose whether to accept the actual level deviation or to perform a further iteration. This dialog box by default is disabled in the noise calibration configuration window:
Figure 3-68: Checkbox for suppressing the message caused by excessive single level deviation or other uncritical warnings
Level adjustment is performed before (in case of individual equalization) of after (in other cases) the single speaker equalization.

3.10.15 Equalization

Equalization is performed in 4 steps:
single speaker equalization
pairwise equalization of the two left and the two right speakers
flatness check of all speakers
all speaker post equalization (optional)
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 82
Each of the two steps is performed with a maximum number of iteration steps which is selected in the configuration dialog box; 2 or 3 iterations are usually enough to achieve a flatness of 1 dB. If additional iterations are expected (up to 5), you can modify this setting in the configuration window
If the all speaker equalization check does not result in a PASS, you can try to achieve a PASS by repeating this measurement. This may be successful if, for example, the tolerance violation is only slight or unwanted noise had impaired the measurement.
Figure 3-69: Message for failed verification of all speaker equalizations
This dialog can be suppressed by predefining the desired behavior in the configuration window:
Figure 3-70: Selection of action on failed verification of all speaker equalizations in the noise calibration configuration window
Additionally optional post equalization can be performed if the all speaker equalization check fails. Activated by default this part of the calibration slightly modifies the equalization of individual speakers to fulfil the flatness requirement.The result of each intermediate step is displayed as a trace during equalization. By selecting 'Report intermediate steps' in the configuration window, all these traces are stored in the report.
Figure 3-71: Display of intermediate equalization results
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 83
All Speaker Post Equalization
Following an unsuccessful all speaker equalization check, it is possible to adjust the all speaker equalization to within the ±3 dB tolerance band by subsequent correction of individual speakers.
Since the algorithm recommended by ES 202 396-1 is extremely time-consuming and not always successful, a much faster post equalization is selected by default. Alternatively, you can select the ETSI recommendation or deactivate the post equalization in the configuration window.
Figure 3-72: Checkbox in the noise calibration configuration window for deactivating the additional all speaker equalization step
Deactivating the 'Use All speaker Equalization' is not recommended because it will cause a final FAIL of calibration if the first all speaker check fails.
If 'All Speaker Post Equalization according to ES 202 396-1' is activated, these further equalization steps are performed:
Measurement of the crosstalk flatness separately for left and right. If at least one
of these is outside the tolerance, the crosstalk speaker, which impairs the frequency response most, is determined (by switching off the other crosstalk speaker). Starting with this speaker, the four crosstalk speakers are equalized again individually (next is the other speaker on the same side followed by the next worse speaker) until the interaction of all speakers is below the tolerance limit or repeated equalization of all four speakers was unsuccessful.
If this equalization step was unsuccessful, adjustment with the direct speakers is
attempted. Once again, the direct speaker, which impairs the frequency response most, is determined. Starting with this speaker, the four direct speakers are equalized again individually until the interaction of all speakers is below the tolerance limit or repeated equalization of all four speakers was unsuccessful.
If 'All Speaker Post Equalization Fast (recommended by R&S)' is activated, these further equalization steps are performed:
Equalization of each pair (left and right) with the attenuated inverse frequency
response of the opposite pair.
If this equalization step was unsuccessful, then equalization of each pair (left and
right) with the attenuated inverse frequency response of the same pair.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 84
If this equalization step is also unsuccessful, equalization has definitively failed and must be repeated with improved room properties. Otherwise, the calibration data now are valid and can be used for measurements in the presence of background noise.
Figure 3-73: Message of successful calibration or verification

3.10.16 Final Test of All Ambiances

This test performs a flatness and level check of each ambiance, with the A-weighting filter activated. It can be performed during the calibration process or separately after a calibration had been terminated successfully. Therefore, it allows checking a valid calibration to see if any modification had been done that requires a recalibration.
For each checked ambiance, a report entry is generated with the graph (documenting the flatness over frequency) and the numeric deviations of flatness and level for left and right channel. Note that the +/-3 dB tolerance limits in the graph are displayed centered to the highest left and right deviation.
If performed during the calibration process the final test section additionally adjusts the total level to minimum level deviation of both channels. To have this final level optimization it is strongly recommended not to deactivate the final test!
To perform the final test separately, start the calibration and skip all calibration sections. Note that the preparatory measurements cannot be skipped but has to be done to ensure proper cabling. The final test cannot be skipped but of course aborted at any time.

3.10.17 Calibration Sections and Resuming Points

The calibration process consists of the following sections, depending on the calibration method used:
For Pink Noise Equalization or Equalization with shared ambiance:
Delay Measurement
Preparatory Measurements
Pink Noise Frequency response before equalization
Equalization with Pink noise or shared Ambiance
Level adjustment of each ambiance
Final test of all ambiances
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 85
For Individual Equalization of each ambiance:
Delay Measurement
Preparatory Measurements
Pink Noise Frequency response before equalization
Level adjustment and Equalization of each ambiance
Final test of all ambiances
Resuming points are located after successful termination of
Delay Measurement
Figure 3-74: Resuming Point after delay measurement
Equalization with Pink noise or shared Ambiance
Figure 3-75: Resuming Point after (shared) equalization
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 86
Level adjustment (and Equalization) of each ambiance
Figure 3-76: Resuming Point after level adjustment or individual equalization of any ambiance, e.g. Pub_Noise_V2
Selecting 'Yes' always resumes the calibration at the current resuming point (with the current beginning calibration section), 'No' skips the current calibration section and goes to the next resuming point. If the calibration section following the current resuming point is not valid, then the calibration process will continue with this section.

3.10.18 Calibration Report

In the report list, an entry is available to open the report of the – failed, aborted or passed – calibration process. For each calibration that has been started, a new entry is generated.
The complete report includes
frequency response (unequalized) of all speakers
limit check is done for fullrange speakers only; no FAIL verdict
Figure 3-77: Report Content: frequency response of unequalised (fullrange) speaker
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 87
all speaker flatness of equalization method used (graph and values)
limit check for flatness and level deviation
Figure 3-78: Report Content: All speaker flatness after equalization
final test results of all ambiances (graph, flatness and level deviation)
limit check for flatness (FAIL verdict for individual equalization method only) and level (including PASS/FAIL verdict)
Figure 3-79: Report Content: final test of each ambiance
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 88
other numeric calibration results
including Delays (verdict for fullrange speakers only), Noise Floor (no FAIL verdict)
Figure 3-80: Report Content: other numeric values
General Settings
Test Parameters (configuration window and internal settings)
unused or irrelevant parameters are skipped
Figure 3-81: Report Content: Test parameters
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 89
Calibration data (of HATS) used
Figure 3-82: Report Content: HATS Calibration data
If the calibration process was aborted or partly skipped, then only a subset of these data is stored
If 'Report intermediate steps' is activated in configuration window, then additional graphs and numeric values are stored, that represents the iteration of single speaker equalization, pair of speaker equalization and all speaker equalization. To avoid this unnecessary information it is recommended to activate this option only for support purposes.

3.11 Synchronization of CMW clock

For packet switched connections such as VoLTE the clock of the system simulator has to be synchronized to the clock of the UE.
Unlike the calibration of mouth, ear or room the clock synchronization has to be repeated whenever the UE is replaced.

3.11.1 Preparations

The Clock Synchronization sets the "Clock Drift" value in R&S CMW. Therefore, a remote connection to R&S CMW must be established before starting the clock synchronization. The remote connection is selected in the "CMW remote control" dialogue that is opened from the "Options" menu
Figure 3-83: R&S CMW Remote Control dialogue
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 90
The R&S CMW can be controlled via GPIB, LAN or USB. For remote control via LAN either CMW name ("CMW50050-", followed by the serial number) or IP address can be entered.
The call between R&S CMW and UE should be established before running the clock synchronization. At the beginning of synchronization process, a check is performed to ensure that connection between R&S UPV and R&S CMW and call between R&S CMW and UE are established. The call yet can be established at this point; to establish the R&S CMW connection however, the synchronization must be aborted.
Figure 3-84: Missing R&S CMW connection

3.11.2 Starting the Synchronization Process

The synchronization is started by selecting the "CMW500 Clock Synchronization" in the Calibration Menu. There are different entries for Handset and Handsfree. When running the first synchronization for a UE the "… New" entry must be selected.
Figure 3-85: Starting the Clock Synchronization
As long as the same UE is used, the clock should be stable. To ensure this the "… Update" entry can be selected at any time. It checks the overall delay and compares it to the value of the last completed ("New") synchronization. If the delay did not significantly change since last adjustment, then a dialogue appears and the synchronization can be skipped by selecting "No".
Figure 3-86: Delay check of update synchronization
If the value differs by more than 2 ms or if you decide to re-synchronize (by selecting "Yes"), a new synchronization will be performed
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 91

3.11.3 Stopping the Synchronization Process

A new synchronization takes about 6 to 18 minutes, an update synchronization at least 4 minutes. The synchronization can be aborted at any point by selecting the "Cancel" button, but cannot be resumed.
Figure 3-87: Cancel button during measurement
Once aborted the synchronization process must be restarted from the beginning to get a valid result.

3.11.4 Steps of Synchronization Process

The requirement for maximum delay drift is 1 ppm. To ensure this requirement the synchronization process uses up to 3 iterations to adjust the CMW clock to a drift value of +/- 1 ppm. Normally 2 iteration steps are sufficient. If the resulting delay drift after 3 steps is still greater than 1 ppm, the synchronization will report a "Fail". Nevertheless, adjust the CMW "clock drift" value. In case of "Fail", a second "New" synchronization that starts with the best "clock drift" value can be performed, and should finally result in "Pass".
Each synchronization step takes about 6 minutes and consists of 5 parts:
Recording of the speech signal
The downlink signal at the ear is measured for 160 s while the R&S UPV generator plays the 2nd sentence of the first English female speaker 40 times to the audio input of R&S CMW.
Figure 3-88: measuring - recording speech signal
After recording is finished, the total peak level of the recorded speech signal is checked. If it is below 25 mV, a warning is displayed and you should check the measurement setup. Anyway, the measurement can be continued.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 92
Figure 3-89: Level check of recorded signal
Adjusting the R&S UPV delay measurement
If necessary an internal adjustment for the R&S UPV delay measurement is performed. This takes less than a second
Figure 3-90: measuring - adjusting delay measurement
Calculating the total overall delay
This step calculates the total delay between outgoing signal to R&S CMW and incoming signal at the ear. It covers delays up to 1 s; the calculation takes about 20 s. Subsequent measurements use this delay to reduce the calculation time.
Figure 3-91: measuring - calculating total delay
Calculating the delay of each 4 s interval
For each of the 40 intervals, the delay difference to total delay is calculated. Each measurement takes about 4 s. During these measurements the current numeric delay value (ms) and the delay-versus-time curve is displayed. The red limit lines display the maximum allowed delay variation of 1 ppm after 160 s.
Figure 3-92: measuring - calculating delay in each interval
In each interval, the level is checked to detect a "no signal" error (e.g. caused by a lost call). If the level falls 20 dB below then the synchronization process is aborted and must be restarted.
These – and all following steps – can be skipped for "update synchronization", if total delay did not significantly change.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 93
Calculation of delay drift and adjustment of CMW clock
The delay values versus time are median filtered to eliminate delay spikes and stitched (if necessary) to delete delay jumps caused by delay buffer over­/underrun).
Figure 3-93: Delay drift calculation, buffer underrun
Figure 3-94: Delay drift calculation, stitched
The measured curve should not cross any of the two red curves by the end of plot. Otherwise, the current iteration is failed and an additional iteration is required to
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 94
verify and possibly re-adjust the synchronization. At this point you can decide to abort the synchronization process by selecting "No". Skipping the verification would spare time (about 6 minutes) of the next iteration, but will cause a FAIL result. It is recommended to run all iteration steps until the delay drift is within tolerance (PASS result).
At the end of each iteration step, the delay drift is calculated, displayed and used for adjusting the CMW clock. If the delay drift is higher than the adjustment limits of CMW clock drift (+/- 100 ppm), you can decide to set the limited value (Yes) or to skip this result (No). Skipping is only recommended if the delay drift result obviously is unusable (e.g. high impairments or lost call).
Figure 3-95: Delay drift calculation, out of CMW limit
At the end of each iteration, the graph displays the following information:
Total Verdict PASS (green) or FAIL (red)
Calculated Delay Drift (in ppm)
Delay of last interval (in ms)
Dialogue with instruction how to proceed
If the calculated delay drift is within +/- 1 ppm, then a PASS result (and report entry) is generated. The last delay is stored for subsequent "Update" synchronizations.
Figure 3-96: PASS result
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 95
If the 3rd iteration had failed, then the synchronization could not properly be completed and a FAIL result (and report entry) is generated.
Figure 3-97: Final FAIL result
Even in case of FAIL the CMW clock had been adjusted as accurately as possible. Thus, a FAIL means that the correct clock adjustment had not be verified.

3.11.5 Options after Termination of Synchronization

If you abort synchronization during measurement, the window of the synchronization process is automatically closed.
If synchronization is terminated with a Fail or Pass result, then the window stays open, displaying the last iteration result.
3 Softkeys are available for further options:
"Close" closes the window of the synchronization process and returns to the test
result overview. The results now are stored and a new line is generated in the test result overview:
Figure 3-98: Result entry of clock synchronization
"Enter Comment" allows adding an individual comment for this synchronization. It
appears in the right column of the test result overview and in the report.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 96
"Create Report" creates a report of the current synchronization. The result can
also be created from the test result overview after closing the synchronization window

3.11.6 Synchronization Report

A report can be generated after termination of the synchronization process.
Report of the current synchronization run by using softkey "Create Report" from
the synchronization window.
Report of any synchronization run from the test result overview The result section of the report contains
3 graphs per iteration containing the raw, filtered and stitched delay values (ms)
versus time (s). They correspond to the graphs displayed during the synchronization process.
A list of numeric values
Figure 3-99: Synchronization report, numeric values
For each iteration, the "Values" section contains the following entries:
"Delay Drift measured" in ppm; PASS if within +/- 1 ppm; negative if the CMW
clock was to slow.
"Clock Drift Value for CMW" in ppm, calculated as sum of current CMW clock drift
value and the negated delay drift measured; PASS if within the CMW adjustment limit (+/- 100 ppm). Does not appear if out of CMW adjustment limit and you decided not to set the value.
"Clock Drift Value set on CMW" in ppm, rounded to full ppm values and limited to
the CMW adjustment limit (if necessary); no limit values and no verdict. Does not appear if out of CMW adjustment limit and you decided not to set the value.
Minimum and maximum of the delay in ms after filtering and stitching (if
necessary); no limit values and no verdict.
The last measured delay in ms; "Delay Last" of the last iteration will be used for all
following update synchronizations; no limit values and no verdict.
R&S UPV-K9/-K91 Calibration
Operating Manual 1402.0043.12 -10 97
The example above shows 3 iteration of a (fictive) synchronization process:
1st iteration is performed with a current CMW clock drift of 100 ppm, resulting in
adjustment value of 100.68 ppm that cannot be set in R&S CMW. If you decided not to set this value in R&S CMW, it does not appear in the report. This requires a second iteration.
2nd iteration is performed with the same, unmodified CMW clock drift, resulting in
adjustment value of 100.79 ppm that cannot be set in R&S CMW. This time you decided to store the limited value in R&S CMW. The delay drift value was in tolerance (Pass), but the CMW clock drift value was out of adjustment limit (Fail), requiring a third iteration.
3rd iteration is performed with the adjusted CMW clock drift. This time the CMW
clock drift value was in adjustment limit and could be set (Pass), but the delay drift value (1.04 ppm) was out of tolerance (Fail). An additional iteration would be necessary, but was not performed, because already 3 iterations had been done. The total verdict was FAIL.
As the delay drift is only slight out of tolerance, a new synchronization executed by you will result in a "Pass".
R&S UPV-K9/-K91 Data Entry for Reporting
Operating Manual 1402.0043.12 -10 98

4 Data Entry for Reporting

4.1 Operator

Under “DataOperator” an operator’s name can be entered which is stored in
association with all calibration and measurement results. If “Operator” is activated in the report settings, the name will appear in all reports about these measurements.

4.2 Test object

Under “Data  Test object New”, a description of the device under test can be entered.
Figure 4-1: Window for input of information about a test object
If the checkbox “Immediately select this test object” is checked, the test object
associated with the entered data is automatically selected upon closing the window. The data of this selected test object will be stored in association with all measurements
and appear on the associated reports if “DUT information” is activated in the report
settings. A previously entered test object can be selected with “Data Test object Select”.
R&S UPV-K9/-K91 Measurements
Operating Manual 1402.0043.12 -10 99

5 Measurements

5.1 General

Special problems caused by the coding and decoding algorithms of mobile phones are encountered when measuring acoustic characteristics. Voice codecs are used to attain the lowest possible data rate. Mobile phone voice codecs are optimized for transmitting human speech with low data rate. Not the actual voice signals but only the filter and fundamental parameters required for signal reconstruction are transmitted.
Purely sinusoidal tones normally used for audio measurements cannot be transmitted with such a system. Therefore, the coder and decoder had initially been excluded from the measurement, which required a specially prepared test mobile phone with a specific test interface.
In modern mobile phones, this interface is not available anymore. Measurements are generally performed via the air interface with the speech coder and decoder included. As mentioned above, measurements using sinusoidal tones cannot be performed because the static sinusoidal input signal becomes a more or less stochastic output signal as a result of coding, particularly in the medium and high audio frequency ranges.
Signals similar to voice therefore have to be used for the measurement, i.e. either artificial voice according to ITU-T P.50 or a multitone signal according to ITU-T P.501 is possible. At the same time, modulation of the signal in time must largely correspond to voice, since many modern mobile phones use algorithms for interference suppression, which use the modulation to distinguish the useful from the interfering signal.
The test routines in the R&S UPV-K91 use an amplitude-modulated multitone signal according to ITU-T P.501 as described in 3GPP TS 26.132 or alternatively artificial voice according to ITU-T P.50 (update key UPV-K9101 or UPV-K9102 required).

5.2 Starting Measurements

Measurements can be started from the “Measurement” menu. This menu contains
either all available measurements, or the measurements belonging to the selected
standard if the checkbox “Allow only selected measurements” in the “Options  Select standard” window is checked.
The menu “Standards” allows you to start any measurement, whereby the
measurements are structured into submenus according to the standards to which they belong. This allows a quick and easy access to all available measurements.
If “Sort” on top of the menu is checked, the standards are sorted according to the
devices Handset and headset, binaural headset, handheld handsfree and desktop and vehicle-mounted handsfree plus Utilities and a sub-menu “More” which contains all available testcases not assigned to any of the standards.
R&S UPV-K9/-K91 Measurements
Operating Manual 1402.0043.12 -10 100
Figure 5-1: "Standards" menu with submenu for narrow band handset tests according to 3GPP TS
26.132, using artificial voice according to ITU-T P.50
The menu “Release” offers all applicable test cases for a selected combination of UE
device type, speech codec bandwidth, test signal and release of the 3GPP test specifications. Item “Select Release” opens the selection window of Fig. 7. Sub-menu “Settings” has the same test case entries and allows you to open the editing window of Figure 3-20.
Figure 5-2: "Release" menu
A fourth possibility to start a single measurement is the button “Run highlighted as single measurement” in the sequence menu (see section 7 Measurements with Electric
Connections below).
Loading...