This manual applies to the following audio analyzers with their options:
R&S®UPV
R&S®UPV66
The software contained in this product uses several valuable open source software packages. For information, see the "Open Source Acknowledgment" document, which is available for download from the R&S
UPV product page at www.rohde-schwarz.com/de/firmware/upv/.
Rohde & Schwarz would like to thank the open source community for their valuable contribution to embedded computing.
Subject to change – Data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
1146.2078.62 | Version 10 | R&S®UPV
The following abbreviation is used throughout the manual: R&S®UPV is abbreviated as R&S UPV. Dolby is
a registered trademark of Dolby Laboratories.
1171.0000.42 - 09 Page 1
Basic Safety Instructions
Always read through and comply with the following safety instructions!
All plants and locations of the Rohde & Schwarz group of companies make every effort to keep the safety
standards of our products up to date and to offer our customers the highest possible degree of safety. Our
products and the auxiliary equipment they require are designed, built and tested in accordance with the
safety standards that apply in each case. Compliance with these standards is continuously monitored by
our quality assurance system. The product described here has been designed, built and tested in
accordance with the EC Certificate of Conformity and has left the manufacturer’s plant in a condition fully
complying with safety standards. To maintain this condition and to ensure safe operation, you must
observe all instructions and warnings provided in this manual. If you have any questions regarding these
safety instructions, the Rohde & Schwarz group of companies will be happy to answer them.
Furthermore, it is your responsibility to use the product in an appropriate manner. This product is designed
for use solely in industrial and laboratory environments or, if expressly permitted, also in the field and must
not be used in any way that may cause personal injury or property damage. You are responsible if the
product is used for any purpose other than its designated purpose or in disregard of the manufacturer's
instructions. The manufacturer shall assume no responsibility for such use of the product.
The product is used for its designated purpose if it is used in accordance with its product documentation
and within its performance limits (see data sheet, documentation, the following safety instructions). Using
the product requires technical skills and, in some cases, a basic knowledge of English. It is therefore
essential that only skilled and specialized staff or thoroughly trained personnel with the required skills be
allowed to use the product. If personal safety gear is required for using Rohde & Schwarz products, this
will be indicated at the appropriate place in the product documentation. Keep the basic safety instructions
and the product documentation in a safe place and pass them on to the subsequent users.
Observing the safety instructions will help prevent personal injury or damage of any kind caused by
dangerous situations. Therefore, carefully read through and adhere to the following safety instructions
before and when using the product. It is also absolutely essential to observe the additional safety
instructions on personal safety, for example, that appear in relevant parts of the product documentation. In
these safety instructions, the word "product" refers to all merchandise sold and distributed by the Rohde &
Schwarz group of companies, including instruments, systems and all accessories. For product-specific
information, see the data sheet and the product documentation.
Safety labels on products
The following safety labels are used on products to warn against risks and dangers.
Symbol
Meaning
Symbol
Meaning
Notice, general danger location
Observe product documentation
ON/OFF Power
Caution when handling heavy equipment
Standby indication
Danger of electric shock
Direct current (DC)
Basic Safety Instructions
1171.0000.42 - 09 Page 2
Symbol
Meaning
Symbol
Meaning
Caution ! Hot surface
Alternating current (AC)
Protective conductor terminal
To identify any terminal which is intended for
connection to an external conductor for
protection against electric shock in case of a
fault, or the terminal of a protective earth
Direct/alternating current (DC/AC)
Earth (Ground)
Class II Equipment
to identify equipment meeting the safety
requirements specified for Class II equipment
(device protected by double or reinforced
insulation)
Frame or chassis Ground terminal
EU labeling for batteries and accumulators
For additional information, see section "Waste
disposal/Environmental protection", item 1.
Be careful when handling electrostatic sensitive
devices
EU labeling for separate collection of electrical
and electronic devices
For additional information, see section "Waste
disposal/Environmental protection", item 2.
Warning! Laser radiation
For additional information, see section
"Operation", item 7.
Signal words and their meaning
The following signal words are used in the product documentation in order to warn the reader about risks
and dangers.
Indicates a hazardous situation which, if not avoided, will result in death or
serious injury.
Indicates a hazardous situation which, if not avoided, could result in death or
serious injury.
Indicates a hazardous situation which, if not avoided, could result in minor or
moderate injury.
Indicates information considered important, but not hazard-related, e.g.
messages relating to property damage.
In the product documentation, the word ATTENTION is used synonymously.
These signal words are in accordance with the standard definition for civil applications in the European
Economic Area. Definitions that deviate from the standard definition may also exist in other economic
areas or military applications. It is therefore essential to make sure that the signal words described here
are always used only in connection with the related product documentation and the related product. The
use of signal words in connection with unrelated products or documentation can result in misinterpretation
and in personal injury or material damage.
Basic Safety Instructions
1171.0000.42 - 09 Page 3
Operating states and operating positions
The product may be operated only under the operating conditions and in the positions specified by the
manufacturer, without the product's ventilation being obstructed. If the manufacturer's specifications are
not observed, this can result in electric shock, fire and/or serious personal injury or death. Applicable local
or national safety regulations and rules for the prevention of accidents must be observed in all work
performed.
1. Unless otherwise specified, the following requirements apply to Rohde & Schwarz products:
predefined operating position is always with the housing floor facing down, IP protection 2X, use only
indoors, max. operating altitude 2000 m above sea level, max. transport altitude 4500 m above sea
level. A tolerance of ±10 % shall apply to the nominal voltage and ±5 % to the nominal frequency,
overvoltage category 2, pollution degree 2.
2. Do not place the product on surfaces, vehicles, cabinets or tables that for reasons of weight or stability
are unsuitable for this purpose. Always follow the manufacturer's installation instructions when
installing the product and fastening it to objects or structures (e.g. walls and shelves). An installation
that is not carried out as described in the product documentation could result in personal injury or
even death.
3. Do not place the product on heat-generating devices such as radiators or fan heaters. The ambient
temperature must not exceed the maximum temperature specified in the product documentation or in
the data sheet. Product overheating can cause electric shock, fire and/or serious personal injury or
even death.
Electrical safety
If the information on electrical safety is not observed either at all or to the extent necessary, electric shock,
fire and/or serious personal injury or death may occur.
1. Prior to switching on the product, always ensure that the nominal voltage setting on the product
matches the nominal voltage of the mains-supply network. If a different voltage is to be set, the power
fuse of the product may have to be changed accordingly.
2. In the case of products of safety class I with movable power cord and connector, operation is
permitted only on sockets with a protective conductor contact and protective conductor.
3. Intentionally breaking the protective conductor either in the feed line or in the product itself is not
permitted. Doing so can result in the danger of an electric shock from the product. If extension cords
or connector strips are implemented, they must be checked on a regular basis to ensure that they are
safe to use.
4. If there is no power switch for disconnecting the product from the mains, or if the power switch is not
suitable for this purpose, use the plug of the connecting cable to disconnect the product from the
mains. In such cases, always ensure that the power plug is easily reachable and accessible at all
times. For example, if the power plug is the disconnecting device, the length of the connecting cable
must not exceed 3 m. Functional or electronic switches are not suitable for providing disconnection
from the AC supply network. If products without power switches are integrated into racks or systems,
the disconnecting device must be provided at the system level.
5. Never use the product if the power cable is damaged. Check the power cables on a regular basis to
ensure that they are in proper operating condition. By taking appropriate safety measures and
carefully laying the power cable, ensure that the cable cannot be damaged and that no one can be
hurt by, for example, tripping over the cable or suffering an electric shock.
Basic Safety Instructions
1171.0000.42 - 09 Page 4
6. The product may be operated only from TN/TT supply networks fuse-protected with max. 16 A (higher
fuse only after consulting with the Rohde & Schwarz group of companies).
7. Do not insert the plug into sockets that are dusty or dirty. Insert the plug firmly and all the way into the
socket provided for this purpose. Otherwise, sparks that result in fire and/or injuries may occur.
8. Do not overload any sockets, extension cords or connector strips; doing so can cause fire or electric
shocks.
9. For measurements in circuits with voltages V
rms
> 30 V, suitable measures (e.g. appropriate
measuring equipment, fuse protection, current limiting, electrical separation, insulation) should be
taken to avoid any hazards.
10. Ensure that the connections with information technology equipment, e.g. PCs or other industrial
computers, comply with the IEC 60950-1 / EN 60950-1 or IEC 61010-1 / EN 61010-1 standards that
apply in each case.
11. Unless expressly permitted, never remove the cover or any part of the housing while the product is in
operation. Doing so will expose circuits and components and can lead to injuries, fire or damage to the
product.
12. If a product is to be permanently installed, the connection between the protective conductor terminal
on site and the product's protective conductor must be made first before any other connection is
made. The product may be installed and connected only by a licensed electrician.
13. For permanently installed equipment without built-in fuses, circuit breakers or similar protective
devices, the supply circuit must be fuse-protected in such a way that anyone who has access to the
product, as well as the product itself, is adequately protected from injury or damage.
14. Use suitable overvoltage protection to ensure that no overvoltage (such as that caused by a bolt of
lightning) can reach the product. Otherwise, the person operating the product will be exposed to the
danger of an electric shock.
15. Any object that is not designed to be placed in the openings of the housing must not be used for this
purpose. Doing so can cause short circuits inside the product and/or electric shocks, fire or injuries.
16. Unless specified otherwise, products are not liquid-proof (see also section "Operating states and
operating positions", item 1). Therefore, the equipment must be protected against penetration by
liquids. If the necessary precautions are not taken, the user may suffer electric shock or the product
itself may be damaged, which can also lead to personal injury.
17. Never use the product under conditions in which condensation has formed or can form in or on the
product, e.g. if the product has been moved from a cold to a warm environment. Penetration by water
increases the risk of electric shock.
18. Prior to cleaning the product, disconnect it completely from the power supply (e.g. AC supply network
or battery). Use a soft, non-linting cloth to clean the product. Never use chemical cleaning agents such
as alcohol, acetone or diluents for cellulose lacquers.
Operation
1. Operating the products requires special training and intense concentration. Make sure that persons
who use the products are physically, mentally and emotionally fit enough to do so; otherwise, injuries
or material damage may occur. It is the responsibility of the employer/operator to select suitable
personnel for operating the products.
Basic Safety Instructions
1171.0000.42 - 09 Page 5
2. Before you move or transport the product, read and observe the section titled "Transport".
3. As with all industrially manufactured goods, the use of substances that induce an allergic reaction
(allergens) such as nickel cannot be generally excluded. If you develop an allergic reaction (such as a
skin rash, frequent sneezing, red eyes or respiratory difficulties) when using a Rohde & Schwarz
product, consult a physician immediately to determine the cause and to prevent health problems or
stress.
4. Before you start processing the product mechanically and/or thermally, or before you take it apart, be
sure to read and pay special attention to the section titled "Waste disposal/Environmental protection",
item 1.
5. Depending on the function, certain products such as RF radio equipment can produce an elevated
level of electromagnetic radiation. Considering that unborn babies require increased protection,
pregnant women must be protected by appropriate measures. Persons with pacemakers may also be
exposed to risks from electromagnetic radiation. The employer/operator must evaluate workplaces
where there is a special risk of exposure to radiation and, if necessary, take measures to avert the
potential danger.
6. Should a fire occur, the product may release hazardous substances (gases, fluids, etc.) that can
cause health problems. Therefore, suitable measures must be taken, e.g. protective masks and
protective clothing must be worn.
7. Laser products are given warning labels that are standardized according to their laser class. Lasers
can cause biological harm due to the properties of their radiation and due to their extremely
concentrated electromagnetic power. If a laser product (e.g. a CD/DVD drive) is integrated into a
Rohde & Schwarz product, absolutely no other settings or functions may be used as described in the
product documentation. The objective is to prevent personal injury (e.g. due to laser beams).
8. EMC classes (in line with EN 55011/CISPR 11, and analogously with EN 55022/CISPR 22,
EN 55032/CISPR 32)
Class A equipment:
Equipment suitable for use in all environments except residential environments and environments
that are directly connected to a low-voltage supply network that supplies residential buildings
Note: Class A equipment is intended for use in an industrial environment. This equipment may
cause radio disturbances in residential environments, due to possible conducted as well as
radiated disturbances. In this case, the operator may be required to take appropriate measures to
eliminate these disturbances.
Class B equipment:
Equipment suitable for use in residential environments and environments that are directly
connected to a low-voltage supply network that supplies residential buildings
Repair and service
1. The product may be opened only by authorized, specially trained personnel. Before any work is
performed on the product or before the product is opened, it must be disconnected from the AC supply
network. Otherwise, personnel will be exposed to the risk of an electric shock.
Basic Safety Instructions
1171.0000.42 - 09 Page 6
2. Adjustments, replacement of parts, maintenance and repair may be performed only by electrical
experts authorized by Rohde & Schwarz. Only original parts may be used for replacing parts relevant
to safety (e.g. power switches, power transformers, fuses). A safety test must always be performed
after parts relevant to safety have been replaced (visual inspection, protective conductor test,
insulation resistance measurement, leakage current measurement, functional test). This helps ensure
the continued safety of the product.
Batteries and rechargeable batteries/cells
If the information regarding batteries and rechargeable batteries/cells is not observed either at all or to the
extent necessary, product users may be exposed to the risk of explosions, fire and/or serious personal
injury, and, in some cases, death. Batteries and rechargeable batteries with alkaline electrolytes (e.g.
lithium cells) must be handled in accordance with the EN 62133 standard.
1. Cells must not be taken apart or crushed.
2. Cells or batteries must not be exposed to heat or fire. Storage in direct sunlight must be avoided.
Keep cells and batteries clean and dry. Clean soiled connectors using a dry, clean cloth.
3. Cells or batteries must not be short-circuited. Cells or batteries must not be stored in a box or in a
drawer where they can short-circuit each other, or where they can be short-circuited by other
conductive materials. Cells and batteries must not be removed from their original packaging until they
are ready to be used.
4. Cells and batteries must not be exposed to any mechanical shocks that are stronger than permitted.
5. If a cell develops a leak, the fluid must not be allowed to come into contact with the skin or eyes. If
contact occurs, wash the affected area with plenty of water and seek medical aid.
6. Improperly replacing or charging cells or batteries that contain alkaline electrolytes (e.g. lithium cells)
can cause explosions. Replace cells or batteries only with the matching Rohde & Schwarz type (see
parts list) in order to ensure the safety of the product.
7. Cells and batteries must be recycled and kept separate from residual waste. Rechargeable batteries
and normal batteries that contain lead, mercury or cadmium are hazardous waste. Observe the
national regulations regarding waste disposal and recycling.
8. Follow the transport stipulations of the carrier (IATA-DGR, IMDG-Code, ADR, RID) when returning
lithium batteries to Rohde & Schwarz subsidiaries.
Transport
1. The product may be very heavy. Therefore, the product must be handled with care. In some cases,
the user may require a suitable means of lifting or moving the product (e.g. with a lift-truck) to avoid
back or other physical injuries.
2. Handles on the products are designed exclusively to enable personnel to transport the product. It is
therefore not permissible to use handles to fasten the product to or on transport equipment such as
cranes, fork lifts, wagons, etc. The user is responsible for securely fastening the products to or on the
means of transport or lifting. Observe the safety regulations of the manufacturer of the means of
transport or lifting. Noncompliance can result in personal injury or material damage.
Basic Safety Instructions
1171.0000.42 - 09 Page 7
3. If you use the product in a vehicle, it is the sole responsibility of the driver to drive the vehicle safely
and properly. The manufacturer assumes no responsibility for accidents or collisions. Never use the
product in a moving vehicle if doing so could distract the driver of the vehicle. Adequately secure the
product in the vehicle to prevent injuries or other damage in the event of an accident.
Waste disposal/Environmental protection
1. Specially marked equipment has a battery or accumulator that must not be disposed of with unsorted
municipal waste, but must be collected separately. It may only be disposed of at a suitable collection
point or via a Rohde & Schwarz customer service center.
2. Waste electrical and electronic equipment must not be disposed of with unsorted municipal waste, but
must be collected separately.
Rohde & Schwarz GmbH & Co. KG has developed a disposal concept and takes full responsibility for
take-back obligations and disposal obligations for manufacturers within the EU. Contact your
Rohde & Schwarz customer service center for environmentally responsible disposal of the product.
3. If products or their components are mechanically and/or thermally processed in a manner that goes
beyond their intended use, hazardous substances (heavy-metal dust such as lead, beryllium, nickel)
may be released. For this reason, the product may only be disassembled by specially trained
personnel. Improper disassembly may be hazardous to your health. National waste disposal
regulations must be observed.
4. If handling the product releases hazardous substances or fuels that must be disposed of in a special
way, e.g. coolants or engine oils that must be replenished regularly, the safety instructions of the
manufacturer of the hazardous substances or fuels and the applicable regional waste disposal
regulations must be observed. Also observe the relevant safety instructions in the product
documentation. The improper disposal of hazardous substances or fuels can cause health problems
and lead to environmental damage.
For additional information about environmental protection, visit the Rohde & Schwarz website.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 8
Instrucciones de seguridad elementales
¡Es imprescindible leer y cumplir las siguientes instrucciones e informaciones de seguridad!
El principio del grupo de empresas Rohde & Schwarz consiste en tener nuestros productos siempre al día
con los estándares de seguridad y de ofrecer a nuestros clientes el máximo grado de seguridad. Nuestros
productos y todos los equipos adicionales son siempre fabricados y examinados según las normas de
seguridad vigentes. Nuestro sistema de garantía de calidad controla constantemente que sean cumplidas
estas normas. El presente producto ha sido fabricado y examinado según el certificado de conformidad
de la UE y ha salido de nuestra planta en estado impecable según los estándares técnicos de seguridad.
Para poder preservar este estado y garantizar un funcionamiento libre de peligros, el usuario deberá
atenerse a todas las indicaciones, informaciones de seguridad y notas de alerta. El grupo de empresas
Rohde & Schwarz está siempre a su disposición en caso de que tengan preguntas referentes a estas
informaciones de seguridad.
Además queda en la responsabilidad del usuario utilizar el producto en la forma debida. Este producto
está destinado exclusivamente al uso en la industria y el laboratorio o, si ha sido expresamente
autorizado, para aplicaciones de campo y de ninguna manera deberá ser utilizado de modo que alguna
persona/cosa pueda sufrir daño. El uso del producto fuera de sus fines definidos o sin tener en cuenta las
instrucciones del fabricante queda en la responsabilidad del usuario. El fabricante no se hace en ninguna
forma responsable de consecuencias a causa del mal uso del producto.
Se parte del uso correcto del producto para los fines definidos si el producto es utilizado conforme a las
indicaciones de la correspondiente documentación del producto y dentro del margen de rendimiento
definido (ver hoja de datos, documentación, informaciones de seguridad que siguen). El uso del producto
hace necesarios conocimientos técnicos y ciertos conocimientos del idioma inglés. Por eso se debe tener
en cuenta que el producto solo pueda ser operado por personal especializado o personas instruidas en
profundidad con las capacidades correspondientes. Si fuera necesaria indumentaria de seguridad para el
uso de productos de Rohde & Schwarz, encontraría la información debida en la documentación del
producto en el capítulo correspondiente. Guarde bien las informaciones de seguridad elementales, así
como la documentación del producto, y entréguelas a usuarios posteriores.
Tener en cuenta las informaciones de seguridad sirve para evitar en lo posible lesiones o daños por
peligros de toda clase. Por eso es imprescindible leer detalladamente y comprender por completo las
siguientes informaciones de seguridad antes de usar el producto, y respetarlas durante el uso del
producto. Deberán tenerse en cuenta todas las demás informaciones de seguridad, como p. ej. las
referentes a la protección de personas, que encontrarán en el capítulo correspondiente de la
documentación del producto y que también son de obligado cumplimiento. En las presentes
informaciones de seguridad se recogen todos los objetos que distribuye el grupo de empresas
Rohde & Schwarz bajo la denominación de "producto", entre ellos también aparatos, instalaciones así
como toda clase de accesorios. Los datos específicos del producto figuran en la hoja de datos y en la
documentación del producto.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 9
Señalización de seguridad de los productos
Las siguientes señales de seguridad se utilizan en los productos para advertir sobre riesgos y peligros.
Símbolo
Significado
Símbolo
Significado
Aviso: punto de peligro general
Observar la documentación del producto
Tensión de alimentación de PUESTA EN
MARCHA / PARADA
Atención en el manejo de dispositivos de peso
elevado
Indicación de estado de espera (standby)
Peligro de choque eléctrico
Corriente continua (DC)
Advertencia: superficie caliente
Corriente alterna (AC)
Conexión a conductor de protección
Corriente continua / Corriente alterna (DC/AC)
Conexión a tierra
El aparato está protegido en su totalidad por un
aislamiento doble (reforzado)
Conexión a masa
Distintivo de la UE para baterías y
acumuladores
Más información en la sección
"Eliminación/protección del medio ambiente",
punto 1.
Aviso: Cuidado en el manejo de dispositivos
sensibles a la electrostática (ESD)
Distintivo de la UE para la eliminación por
separado de dispositivos eléctricos y
electrónicos
Más información en la sección
"Eliminación/protección del medio ambiente",
punto 2.
Advertencia: rayo láser
Más información en la sección
"Funcionamiento", punto 7.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 10
Palabras de señal y su significado
En la documentación del producto se utilizan las siguientes palabras de señal con el fin de advertir contra
riesgos y peligros.
Indica una situación de peligro que, si no se evita, causa lesiones
graves o incluso la muerte.
Indica una situación de peligro que, si no se evita, puede causar
lesiones graves o incluso la muerte.
Indica una situación de peligro que, si no se evita, puede causar
lesiones leves o moderadas.
Indica información que se considera importante, pero no en relación
con situaciones de peligro; p. ej., avisos sobre posibles daños
materiales.
En la documentación del producto se emplea de forma sinónima el
término CUIDADO.
Las palabras de señal corresponden a la definición habitual para aplicaciones civiles en el área
económica europea. Pueden existir definiciones diferentes a esta definición en otras áreas económicas o
en aplicaciones militares. Por eso se deberá tener en cuenta que las palabras de señal aquí descritas
sean utilizadas siempre solamente en combinación con la correspondiente documentación del producto y
solamente en combinación con el producto correspondiente. La utilización de las palabras de señal en
combinación con productos o documentaciones que no les correspondan puede llevar a interpretaciones
equivocadas y tener por consecuencia daños en personas u objetos.
Estados operativos y posiciones de funcionamiento
El producto solamente debe ser utilizado según lo indicado por el fabricante respecto a los estados
operativos y posiciones de funcionamiento sin que se obstruya la ventilación. Si no se siguen las
indicaciones del fabricante, pueden producirse choques eléctricos, incendios y/o lesiones graves con
posible consecuencia de muerte. En todos los trabajos deberán ser tenidas en cuenta las normas
nacionales y locales de seguridad del trabajo y de prevención de accidentes.
1. Si no se convino de otra manera, es para los productos Rohde & Schwarz válido lo que sigue:
como posición de funcionamiento se define por principio la posición con el suelo de la caja para
abajo, modo de protección IP 2X, uso solamente en estancias interiores, utilización hasta 2000 m
sobre el nivel del mar, transporte hasta 4500 m sobre el nivel del mar. Se aplicará una tolerancia de
±10 % sobre el voltaje nominal y de ±5 % sobre la frecuencia nominal. Categoría de sobrecarga
eléctrica 2, índice de suciedad 2.
2. No sitúe el producto encima de superficies, vehículos, estantes o mesas, que por sus características
de peso o de estabilidad no sean aptos para él. Siga siempre las instrucciones de instalación del
fabricante cuando instale y asegure el producto en objetos o estructuras (p. ej. paredes y estantes). Si
se realiza la instalación de modo distinto al indicado en la documentación del producto, se pueden
causar lesiones o, en determinadas circunstancias, incluso la muerte.
3. No ponga el producto sobre aparatos que generen calor (p. ej. radiadores o calefactores). La
temperatura ambiente no debe superar la temperatura máxima especificada en la documentación del
producto o en la hoja de datos. En caso de sobrecalentamiento del producto, pueden producirse
choques eléctricos, incendios y/o lesiones graves con posible consecuencia de muerte.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 11
Seguridad eléctrica
Si no se siguen (o se siguen de modo insuficiente) las indicaciones del fabricante en cuanto a seguridad
eléctrica, pueden producirse choques eléctricos, incendios y/o lesiones graves con posible consecuencia
de muerte.
1. Antes de la puesta en marcha del producto se deberá comprobar siempre que la tensión
preseleccionada en el producto coincida con la de la red de alimentación eléctrica. Si es necesario
modificar el ajuste de tensión, también se deberán cambiar en caso dado los fusibles
correspondientes del producto.
2. Los productos de la clase de protección I con alimentación móvil y enchufe individual solamente
podrán enchufarse a tomas de corriente con contacto de seguridad y con conductor de protección
conectado.
3. Queda prohibida la interrupción intencionada del conductor de protección, tanto en la toma de
corriente como en el mismo producto. La interrupción puede tener como consecuencia el riesgo de
que el producto sea fuente de choques eléctricos. Si se utilizan cables alargadores o regletas de
enchufe, deberá garantizarse la realización de un examen regular de los mismos en cuanto a su
estado técnico de seguridad.
4. Si el producto no está equipado con un interruptor para desconectarlo de la red, o bien si el
interruptor existente no resulta apropiado para la desconexión de la red, el enchufe del cable de
conexión se deberá considerar como un dispositivo de desconexión.
El dispositivo de desconexión se debe poder alcanzar fácilmente y debe estar siempre bien accesible.
Si, p. ej., el enchufe de conexión a la red es el dispositivo de desconexión, la longitud del cable de
conexión no debe superar 3 m).
Los interruptores selectores o electrónicos no son aptos para el corte de la red eléctrica. Si se
integran productos sin interruptor en bastidores o instalaciones, se deberá colocar el interruptor en el
nivel de la instalación.
5. No utilice nunca el producto si está dañado el cable de conexión a red. Compruebe regularmente el
correcto estado de los cables de conexión a red. Asegúrese, mediante las medidas de protección y
de instalación adecuadas, de que el cable de conexión a red no pueda ser dañado o de que nadie
pueda ser dañado por él, p. ej. al tropezar o por un choque eléctrico.
6. Solamente está permitido el funcionamiento en redes de alimentación TN/TT aseguradas con fusibles
de 16 A como máximo (utilización de fusibles de mayor amperaje solo previa consulta con el grupo de
empresas Rohde & Schwarz).
7. Nunca conecte el enchufe en tomas de corriente sucias o llenas de polvo. Introduzca el enchufe por
completo y fuertemente en la toma de corriente. La no observación de estas medidas puede provocar
chispas, fuego y/o lesiones.
8. No sobrecargue las tomas de corriente, los cables alargadores o las regletas de enchufe ya que esto
podría causar fuego o choques eléctricos.
9. En las mediciones en circuitos de corriente con una tensión U
eff
> 30 V se deberán tomar las medidas
apropiadas para impedir cualquier peligro (p. ej. medios de medición adecuados, seguros, limitación
de tensión, corte protector, aislamiento etc.).
10. Para la conexión con dispositivos informáticos como un PC o un ordenador industrial, debe
comprobarse que éstos cumplan los estándares IEC60950-1/EN60950-1 o IEC61010-1/EN 61010-1
válidos en cada caso.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 12
11. A menos que esté permitido expresamente, no retire nunca la tapa ni componentes de la carcasa
mientras el producto esté en servicio. Esto pone a descubierto los cables y componentes eléctricos y
puede causar lesiones, fuego o daños en el producto.
12. Si un producto se instala en un lugar fijo, se deberá primero conectar el conductor de protección fijo
con el conductor de protección del producto antes de hacer cualquier otra conexión. La instalación y
la conexión deberán ser efectuadas por un electricista especializado.
13. En el caso de dispositivos fijos que no estén provistos de fusibles, interruptor automático ni otros
mecanismos de seguridad similares, el circuito de alimentación debe estar protegido de modo que
todas las personas que puedan acceder al producto, así como el producto mismo, estén a salvo de
posibles daños.
14. Todo producto debe estar protegido contra sobretensión (debida p. ej. a una caída del rayo) mediante
los correspondientes sistemas de protección. Si no, el personal que lo utilice quedará expuesto al
peligro de choque eléctrico.
15. No debe introducirse en los orificios de la caja del aparato ningún objeto que no esté destinado a ello.
Esto puede producir cortocircuitos en el producto y/o puede causar choques eléctricos, fuego o
lesiones.
16. Salvo indicación contraria, los productos no están impermeabilizados (ver también el capítulo
"Estados operativos y posiciones de funcionamiento", punto 1). Por eso es necesario tomar las
medidas necesarias para evitar la entrada de líquidos. En caso contrario, existe peligro de choque
eléctrico para el usuario o de daños en el producto, que también pueden redundar en peligro para las
personas.
17. No utilice el producto en condiciones en las que pueda producirse o ya se hayan producido
condensaciones sobre el producto o en el interior de éste, como p. ej. al desplazarlo de un lugar frío a
otro caliente. La entrada de agua aumenta el riesgo de choque eléctrico.
18. Antes de la limpieza, desconecte por completo el producto de la alimentación de tensión (p. ej. red de
alimentación o batería). Realice la limpieza de los aparatos con un paño suave, que no se deshilache.
No utilice bajo ningún concepto productos de limpieza químicos como alcohol, acetona o diluyentes
para lacas nitrocelulósicas.
Funcionamiento
1. El uso del producto requiere instrucciones especiales y una alta concentración durante el manejo.
Debe asegurarse que las personas que manejen el producto estén a la altura de los requerimientos
necesarios en cuanto a aptitudes físicas, psíquicas y emocionales, ya que de otra manera no se
pueden excluir lesiones o daños de objetos. El empresario u operador es responsable de seleccionar
el personal usuario apto para el manejo del producto.
2. Antes de desplazar o transportar el producto, lea y tenga en cuenta el capítulo "Transporte".
3. Como con todo producto de fabricación industrial no puede quedar excluida en general la posibilidad
de que se produzcan alergias provocadas por algunos materiales empleados ―los llamados alérgenos (p. ej. el níquel)―. Si durante el manejo de productos Rohde & Schwarz se producen
reacciones alérgicas, como p. ej. irritaciones cutáneas, estornudos continuos, enrojecimiento de la
conjuntiva o dificultades respiratorias, debe avisarse inmediatamente a un médico para investigar las
causas y evitar cualquier molestia o daño a la salud.
4. Antes de la manipulación mecánica y/o térmica o el desmontaje del producto, debe tenerse en cuenta
imprescindiblemente el capítulo "Eliminación/protección del medio ambiente", punto 1.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 13
5. Ciertos productos, como p. ej. las instalaciones de radiocomunicación RF, pueden a causa de su
función natural, emitir una radiación electromagnética aumentada. Deben tomarse todas las medidas
necesarias para la protección de las mujeres embarazadas. También las personas con marcapasos
pueden correr peligro a causa de la radiación electromagnética. El empresario/operador tiene la
obligación de evaluar y señalizar las áreas de trabajo en las que exista un riesgo elevado de
exposición a radiaciones.
6. Tenga en cuenta que en caso de incendio pueden desprenderse del producto sustancias tóxicas
(gases, líquidos etc.) que pueden generar daños a la salud. Por eso, en caso de incendio deben
usarse medidas adecuadas, como p. ej. máscaras antigás e indumentaria de protección.
7. Los productos con láser están provistos de indicaciones de advertencia normalizadas en función de la
clase de láser del que se trate. Los rayos láser pueden provocar daños de tipo biológico a causa de
las propiedades de su radiación y debido a su concentración extrema de potencia electromagnética.
En caso de que un producto Rohde & Schwarz contenga un producto láser (p. ej. un lector de
CD/DVD), no debe usarse ninguna otra configuración o función aparte de las descritas en la
documentación del producto, a fin de evitar lesiones (p. ej. debidas a irradiación láser).
8. Clases de compatibilidad electromagnética (conforme a EN 55011 / CISPR 11; y en analogía con EN
55022 / CISPR 22, EN 55032 / CISPR 32)
Aparato de clase A:
Aparato adecuado para su uso en todos los entornos excepto en los residenciales y en aquellos
conectados directamente a una red de distribución de baja tensión que suministra corriente a
edificios residenciales.
Nota: Los aparatos de clase A están destinados al uso en entornos industriales. Estos aparatos
pueden causar perturbaciones radioeléctricas en entornos residenciales debido a posibles
perturbaciones guiadas o radiadas. En este caso, se le podrá solicitar al operador que tome las
medidas adecuadas para eliminar estas perturbaciones.
Aparato de clase B:
Aparato adecuado para su uso en entornos residenciales, así como en aquellos conectados
directamente a una red de distribución de baja tensión que suministra corriente a edificios
residenciales.
Reparación y mantenimiento
1. El producto solamente debe ser abierto por personal especializado con autorización para ello. Antes
de manipular el producto o abrirlo, es obligatorio desconectarlo de la tensión de alimentación, para
evitar toda posibilidad de choque eléctrico.
2. El ajuste, el cambio de partes, el mantenimiento y la reparación deberán ser efectuadas solamente
por electricistas autorizados por Rohde & Schwarz. Si se reponen partes con importancia para los
aspectos de seguridad (p. ej. el enchufe, los transformadores o los fusibles), solamente podrán ser
sustituidos por partes originales. Después de cada cambio de partes relevantes para la seguridad
deberá realizarse un control de seguridad (control a primera vista, control del conductor de
protección, medición de resistencia de aislamiento, medición de la corriente de fuga, control de
funcionamiento). Con esto queda garantizada la seguridad del producto.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 14
Baterías y acumuladores o celdas
Si no se siguen (o se siguen de modo insuficiente) las indicaciones en cuanto a las baterías y
acumuladores o celdas, pueden producirse explosiones, incendios y/o lesiones graves con posible
consecuencia de muerte. El manejo de baterías y acumuladores con electrolitos alcalinos (p. ej. celdas de
litio) debe seguir el estándar EN 62133.
1. No deben desmontarse, abrirse ni triturarse las celdas.
2. Las celdas o baterías no deben someterse a calor ni fuego. Debe evitarse el almacenamiento a la luz
directa del sol. Las celdas y baterías deben mantenerse limpias y secas. Limpiar las conexiones
sucias con un paño seco y limpio.
3. Las celdas o baterías no deben cortocircuitarse. Es peligroso almacenar las celdas o baterías en
estuches o cajones en cuyo interior puedan cortocircuitarse por contacto recíproco o por contacto con
otros materiales conductores. No deben extraerse las celdas o baterías de sus embalajes originales
hasta el momento en que vayan a utilizarse.
4. Las celdas o baterías no deben someterse a impactos mecánicos fuertes indebidos.
5. En caso de falta de estanqueidad de una celda, el líquido vertido no debe entrar en contacto con la
piel ni los ojos. Si se produce contacto, lavar con agua abundante la zona afectada y avisar a un
médico.
6. En caso de cambio o recarga inadecuados, las celdas o baterías que contienen electrolitos alcalinos
(p. ej. las celdas de litio) pueden explotar. Para garantizar la seguridad del producto, las celdas o
baterías solo deben ser sustituidas por el tipo Rohde & Schwarz correspondiente (ver lista de
recambios).
7. Las baterías y celdas deben reciclarse y no deben tirarse a la basura doméstica. Las baterías o
acumuladores que contienen plomo, mercurio o cadmio deben tratarse como residuos especiales.
Respete en esta relación las normas nacionales de eliminación y reciclaje.
8. En caso de devolver baterías de litio a las filiales de Rohde & Schwarz, debe cumplirse las
normativas sobre los modos de transporte (IATA-DGR, código IMDG, ADR, RID).
Transporte
1. El producto puede tener un peso elevado. Por eso es necesario desplazarlo o transportarlo con
precaución y, si es necesario, usando un sistema de elevación adecuado (p. ej. una carretilla
elevadora), a fin de evitar lesiones en la espalda u otros daños personales.
2. Las asas instaladas en los productos sirven solamente de ayuda para el transporte del producto por
personas. Por eso no está permitido utilizar las asas para la sujeción en o sobre medios de transporte
como p. ej. grúas, carretillas elevadoras de horquilla, carros etc. Es responsabilidad suya fijar los
productos de manera segura a los medios de transporte o elevación. Para evitar daños personales o
daños en el producto, siga las instrucciones de seguridad del fabricante del medio de transporte o
elevación utilizado.
3. Si se utiliza el producto dentro de un vehículo, recae de manera exclusiva en el conductor la
responsabilidad de conducir el vehículo de manera segura y adecuada. El fabricante no asumirá
ninguna responsabilidad por accidentes o colisiones. No utilice nunca el producto dentro de un
vehículo en movimiento si esto pudiera distraer al conductor. Asegure el producto dentro del vehículo
debidamente para evitar, en caso de un accidente, lesiones u otra clase de daños.
Instrucciones de seguridad elementales
1171.0000.42 - 09 Page 15
Eliminación/protección del medio ambiente
1. Los dispositivos marcados contienen una batería o un acumulador que no se debe desechar con los
residuos domésticos sin clasificar, sino que debe ser recogido por separado. La eliminación se debe
efectuar exclusivamente a través de un punto de recogida apropiado o del servicio de atención al
cliente de Rohde & Schwarz.
2. Los dispositivos eléctricos usados no se deben desechar con los residuos domésticos sin clasificar,
sino que deben ser recogidos por separado.
Rohde & Schwarz GmbH & Co.KG ha elaborado un concepto de eliminación de residuos y asume
plenamente los deberes de recogida y eliminación para los fabricantes dentro de la UE. Para
desechar el producto de manera respetuosa con el medio ambiente, diríjase a su servicio de atención
al cliente de Rohde & Schwarz.
3. Si se trabaja de manera mecánica y/o térmica cualquier producto o componente más allá del
funcionamiento previsto, pueden liberarse sustancias peligrosas (polvos con contenido de metales
pesados como p. ej. plomo, berilio o níquel). Por eso el producto solo debe ser desmontado por
personal especializado con formación adecuada. Un desmontaje inadecuado puede ocasionar daños
para la salud. Se deben tener en cuenta las directivas nacionales referentes a la eliminación de
residuos.
4. En caso de que durante el trato del producto se formen sustancias peligrosas o combustibles que
deban tratarse como residuos especiales (p. ej. refrigerantes o aceites de motor con intervalos de
cambio definidos), deben tenerse en cuenta las indicaciones de seguridad del fabricante de dichas
sustancias y las normas regionales de eliminación de residuos. Tenga en cuenta también en caso
necesario las indicaciones de seguridad especiales contenidas en la documentación del producto. La
eliminación incorrecta de sustancias peligrosas o combustibles puede causar daños a la salud o
daños al medio ambiente.
Se puede encontrar más información sobre la protección del medio ambiente en la página web de
Rohde & Schwarz.
Customer Support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz product, contact our customer supportcenter. A team of
highly qualified engineers provides support and works with youto find a solution to your query on any
aspect of the operation, programming or applicationsof Rohde & Schwarz products.
Contact information
Contact our customer support center at www.rohde-schwarz.com/support, or follow this QR code:
Although audio signals are mainly processed digitally nowadays, analog technology remains a viable alternative that is continuously being enhanced. Therefore,
both analog and digital measurements must be performed. The R&SUPV audio
analyzer is designed precisely for this purpose.
1.1Key Features
●
Suitable for all interfaces (analog, digital and combined)
●
Simultaneous display of multiple measurement functions
●
Sampling rate up to 400 kHz
●
All test signals and measurement functions in a single box
●
Largest variety of interfaces offered in a single instrument
●
Convenient operation throughout
1.2For Your Safety
The R&S UPV is designated for use in industrial, administrative and laboratory
environments. Use the R&S UPV only for its designated purpose. Observe the
operating conditions and performance limits stated in the data sheet.
The product documentation helps you to use the R&S UPV safely and efficiently.
Keep the product documentation in a safe place and pass it on to the subsequent
users.
Safety information is part of the product documentation. It warns you about the
potential dangers and gives instructions how to prevent personal injury or damage caused by dangerous situations. Safety information is provided as follows:
●
In the "Basic Safety Instructions", safety issues are grouped according to subjects. For example, one subject is electrical safety. The "Basic Safety Instructions" are delivered with the R&S UPV in different languages in print.
●
Throughout the documentation, safety instructions are provided when you
need to take care during setup or operation. Always read the safety instructions carefully. Make sure to comply fully with them. Do not take risks and do
5Getting Started 1146.2078.62 ─ 10
R&S®UPV
Documentation Overview
not underestimate the potential danger of small details such as a damaged
power cable.
Introduction
1.3Documentation Overview
This section provides an overview of the R&S UPV user documentation and on
how to access it.
Included in delivery
●
Getting started and basic safety instructions
The printed getting started manual and the basic safety instruction are delivered together with the instrument.
The getting started describes how to set up and start working with the instrument, it includes basic operations and general information, e.g. safety instructions,
●
User manual for the instrument and its options
The user manual is provided as context-sensitive online-help on the instrument.
In addition to the chapters of the quick start guide, the user manual contains
the description of all instrument functions and the remote control of the instrument. Furthermore, it features notes for the preventive maintenance of the
R&S UPV and for locating errors based on the warnings and error messages
issued by the instrument.
As download on the Internet via the download section of the instrument
page
●
Getting started
In addition to the printed manual, the getting started is provided as download
on the Internet (PDF format).
●
User manual for the instrument and its options
In addition to the online help, the user manual is provided as download (PDF
format) and as online manual (HTML format) on the Internet.
●
Instrument security procedures
The instrument security procedures manual deals with security issues when
working with the R&S UPV in secure areas.
●
Data sheet and brochure
6Getting Started 1146.2078.62 ─ 10
R&S®UPV
Documentation Overview
The data sheet contains the technical specifications of the R&S UPV. It lists
the options, their order numbers and optional accessories. The brochure provides an overview of the instrument and deals with the specific characteristics.
●
Release notes and open source acknowledgment
Release notes and open source acknowledgment are provided together with
the corresponding firmware.
The release notes describe the installation of the firmware, new and improved
functions, problems solved and last-minute changes to the documentation.
The corresponding firmware version can be seen on the cover sheet of the
release notes.
The open source acknowledgment document provides verbatim license texts
of the used open source software. The acknowledgment is also provided on
the instrument.
●
Application notes
Application notes deal with special applications or background information on
particular topics.
Introduction
Link to the product page on the Internet: www.rohde-schwarz.com/product/
upv.html.
As download for registered customers on GLORIS, the Global Rohde &
Schwarz Information System.
●
Service manual
The service manual contains all the necessary information to maintain the
R&S UPV by replacing modules and to expand its functionality by installing
options.
Link to GLORIS: gloris.rohde-schwarz.com.
7Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
2Startup
The R&S UPV audio analyzer is available in two models. The R&S UPV standard
model and an R&S UPV66 model which is specifically designed for use in systems and has no display, no front panel controls and no CD/DVD drive. Except for
the operation via front panel, the R&S UPV66 variant is largely identical to the
standard model with respect to functionality. This manual therefore describes both
instruments in parallel, and any differences are indicated at the appropriate points
in the text.
This chapter describes the controls and ports of the R&S audio analyzer UPV
based on the front and rear view and shows how to put the instrument into operation. It also describes the connection of peripherals such as printer, keyboard,
mouse and monitor. Specifications for the interfaces can be found in the data
sheet.
The introduction to Chapter 3.1, "Introduction - Getting Started", on page 91
provides an overview of the functions and the operating concept of the audio analyzer. Detailed operating instructions and an overview of the menus are provided
in Chapter 4.1, "Introduction - Manual Operation", on page 140.
The instrument help (and the PDF and online manual on the Internet) comprise
the entire manual: The individual menus and functions of the instrument including
the associated remote-control commands are explained in detail in chapter
"Instrument Functions". Information on remote control of the instrument is provided in the chapters "Remote Control - Fundamentals" and "Remote Control Commands Description". A detailed description of the instrument interfaces can
be found in the chapter "Maintenance and Instrument Interfaces".
8Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
The audio analyzer has a Windows operating system. No special knowledge of the operating system used is required for operating the instrument.
Basic PC knowledge, such as what a file, a directory, etc. is, or how data is
transferred using Windows Explorer® is assumed and, therefore, not discussed in any detail.
The R&S UPV audio analyzer can be operated via the controls on the front
panel. It is, however, recommended - and for certain functions required - to
operate the instrument using an external keyboard and mouse. Operation
follows the same basic rules which apply to other Windows programs in use
today. It is also assumed that users are familiar with these basic rules. This
information is therefore not specified in detail here.
The R&S UPV66 model does not feature a front panel keyboard and has no
display or CD/DVD drive; but similar to the R&S UPV standard model, it can
be operated by means of a keyboard or mouse after an external monitor is
connected.
Startup
2.1Description of Front View
This section provides an overview of the controls and ports on the front of the
R&S UPV/R&S UPV66.
9Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Description of Front View
Figure 2-1: Front Panel R&S UPV
Figure 2-2: Front Panel R&S UPV66
10Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
Startup
2.1.1Screen with softkeys
The large screen shows all of the important settings and states of the audio analyzer in a clearly organized display. 5 displays ("screens") are provided to ensure
clearer organization of the many possible panels and display windows.
The screen is divided into the following sections:
●
menu bar
●
operating and display area
●
WINBAR with softkey labeling
●
Softkeys
The menu bar
●
contains a row of pull-down menus which can be used to, for example, load
instrument settings, select panels and screens or call up help functions.
●
appears when the MENU key is pressed or the mouse pointer is moved to the
left of the screen. It otherwise remains hidden in order to save space.
In the operating and display area
●
the Audio Analyzer R&S UPV is operated using numerous panels which can
be selected by the user
●
the measurement results are displayed numerically and/or graphically in the
format selected by the user
The Softkey/WINBAR bar at the bottom of the screen has two functions:
11Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
●
In normal operating mode, the softkey bar is visible; it contains the functions
available in the selection windows of the panels; the softkeys permit rapid
selection without needing to use the mouse.
●
When the WINBAR key is pressed, the WINBAR containing the names of all
opened panels / graphical windows is displayed in place of the softkey bar.
Pressing the associated softkey puts the focus on the corresponding panel /
graphical window, i.e. it becomes active. At the same time, the system returns
to normal operating mode (see above).
●
It is possible to toggle between the softkey bar and WINBAR at any time by
pressing the WINBAR key.
The softkeys
●
corresponds to the label on the softkey bar and WINBAR. The softkeys can
also be operated using the function keys on the external keyboard or by clicking the associated button.
Startup
2.1.2Navigation keys
The navigation keys consist of 4 rocker buttons that are used for switching forward and back:
[SCREEN]
Switches between the individual display settings (the "screens").
[WINDOW]
Switches the focus to each of the panels open in the visible screen in sequence
(the panel with the focus is active).
[SCAN]
Changes between the individual traces ("scans") in graphical displays so that, for
example, measured values can be read off from the traces using the cursor.
[PAGE]
Scrolls back and forward in the tables one page at a time.
12Getting Started 1146.2078.62 ─ 10
R&S®UPV
2.1.3Data entry keys
[0 ... 9]
Entry of numeric values
Startup
Description of Front View
[.]
Entry of decimal point
[+/-]
Entry of sign
abc[]
Entry of letters for file names and/or directory names in the file selector in the
mobile phone method, provided that it appears in the R&S UPV style.
[Blank]
Entry of a space
[* ... #]
Entry of special characters
A <---> a
Switching between uppercase and lowercase letters
[MENU]
Displays the menu bar on the screen
13Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
[<--- / UNDO]
Deletes the character to the left of the cursor
Undoes the action last performed
[ESC]
Closes the open window; the old value or parameter is retained
[ENTER]
Completes the data entry; the new value is adopted
Confirms (OK) and closes open entry windows
Pressing the rotary knob has the same effect
2.1.4CONTROL panel
Startup
The CONTROL panel features the following control functions:
●
Starting measurements / sweeps
●
Adjusting the loudspeaker
●
Printing out the screen contents
●
Shutting down the operating system
●
Switch between remote / manual operation
●
Switch off outputs
[START]
Starts continuous measurements and sweeps (the LED is on); resets Min/Max
values for bargraph displays, averaging for measurement results, and limit overshoots
[SINGLE]
Starts a single measurement or a single sweep (the LED is on during execution)
[STOP / CONT]
Measurement/sweep is stopped or continued (toggle function)
14Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
[ON]
Switches the loudspeaker or connected headphones on and off (toggle function)
Volume
Opens a window for adjusting the audio monitoring volume
[H COPY]
Depending on the settings under Utilities, printout of the screen contents on the
connected printer is started or the screen contents are stored in a file
[SHUTDOWN]
Stores the current settings on the hard disk and then shuts down the Windows
operating system
[LOCAL]
Switches from remote control to manual operation (the LED is on with remote
control)
Startup
[OUTPUT]
Switches all outputs of the measuring instrument off or on again (the LED is on
when the outputs are switched off)
2.1.5VARIATION panel
Rotary Knob
Moves the focus in the panels, selection windows and tables
Varies the entry value at the cursor position
15Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
Moves the active cursor in the graphical windows
Pressing the rotary knob has the same effect as pressing the ENTER key
[Cursor Up / Down]
Move the focus in the panels, selection windows and tables
Varies the entry value at the cursor position
[Cusor Left / Right]
Moves the active cursor inside the graphical windows
Moves the cursor position in entered values
2.1.6USB Sockets
Startup
USB (Universal Serial Bus) interfaces of type A (host USB) for:
●
Connecting peripherals such as mouse, keyboard, printer
●
Connecting a USB stick for transferring files
●
There are two additional USB interfaces (type A) at the rear of the instrument.
2.1.7CD/DVD Combo drive
The integrated combined CD/DVD drive is used to
●
install software updates on the R&S UPV
●
install sequence programs and macros
●
import data and instrument settings from other R&S UPVs
●
store data on a CD
16Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
2.1.8Ground socket
Ground socket that is connected with the instrument housing.
2.1.9Headphone Connector
Startup
Connection of headphones for audio monitoring of the analysis or generator signal
2.1.10WINDOW / SETUP panel
[WINBAR]
The WINBAR key is used to toggle between the softkey bar and WINBAR at any
time:
●
In normal operating mode, the softkey bar is visible; it contains the functions
available in the selection windows of the panels; the softkeys permit rapid
selection without needing to use the mouse.
●
When the WINBAR key is pressed, the WINBAR containing the names of all
opened panels or graphical windows is displayed in place of the softkey bar.
Pressing the associated softkey puts the focus on the corresponding panel or
graphical window, i.e. it becomes active. At the same time, the system returns
to normal operating mode (see above).
17Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
[MODIFY]
Opens a selection of functions on the softkey bar; these functions can be used in
conjunction with the rotary knob to move or resize the active panel or graphical
window
[HIDE]
Minimizes ("hides") the active panel or graphical window
[CLOSE]
Closes the active panel or graphical window
[HELP]
Pressing this key displays a context-sensitive help text
[SAVE]
Opens a window for storing instrument settings
Startup
[LOAD]
Opens a window for loading predefined instrument settings
[PRESET]
Loads the factory default settings of the instrument or carries out a Master
PRESET.
●
Factory setting:
Pressing the button briefly displays the security prompt "Do you want to preset
the UPV?". If you answer this question with "Yes", the factory default settings
are restored immediately.
●
Master PRESET:
With the Master PRESET, you can easily get out of an unsuitable screen and
monitor configuration and back to an operable instrument: Switch the instrument off and on again. During the software start, the OUTPUT Off-LED is illuminated briefly in yellow. As soon as this LED is illuminated in yellow, press
the PRESET button briefly. The LED flashes for confirmation. Within the next
5 seconds, press the PRESET key again briefly. The Master PRESET is now
carried out and forces a switchover to the external monitor with a screen resolution of 800 x 600 pixels and the factory default settings.
2.1.11Analog Audio Interfaces
Output and input sockets of the analog generators and analyzers
18Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
Possible instrument damage due to incorrect input voltages
The instrument complies with measuring category I; make sure that the
input voltage at the sockets of the analog analyzers does not exceed 110 V
(RMS, sine wave) and 160 V (peak value).
Startup
Do not use the instrument in measuring categories II, III and IV.
Explanation: Measuring circuits as defined in section 6.7.4 of EN61010-1:
Measuring category I is intended for measurements on circuits which are
not connected to the high‑voltage current system.
Meaning of the LEDs: Generator OVLD: Generator is overloaded. Analyzer
OVLD: The low-impedance input resistors (300 Ohm or 600 Ohm) are overloaded and have been switched off.
The output stage can be damaged by an unacceptably high external signal
feed. The frontend can be damaged by an unacceptably high external signal feed from a low-impedance source in the analyzer.
2.1.12Digital Audio Interfaces
Inputs and outputs for connecting digital consumer instruments and professional
studio equipment.
Requires the R&S UPV‑B2 option (digital audio interfaces)
19Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
[UNBAL]
BNC sockets
[OPTICAL]
Optical interface conforming to EIAJ CP-340, TOSLINK system
Startup
[BAL]
XLR female and male connector
2.1.13Power Switch
The ON/OFF switch switches the R&S UPV on and starts the booting process.
When switched off, there is immediate bipolar disconnection of the R&S UPV
from the mains supply.
20Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of Front View
Risk of data loss
There is a risk of data loss if the R&S UPV is switched off without first shutting down the operating system.
It is urgently recommended to shut down the operating system before
switching off the instrument ([SHUT-DOWN] key or via mouse operation).
Shutting down the operating system ensures that open files are closed
properly.
Startup
2.1.14Status LEDs
With the R&S UPV66 audio analyzer, the instrument’s state is indicated via LEDs.
The individual LEDs have the following meaning:
[POWER ON]
The instrument is switched on
[SWITCH OFF]
The operating system has been shut down; you can now switch off the instrument.
[LAN]
The instrument is connected to a local area network.
[READY]
The measurement instrument is ready for operation
[BUSY]
Lights as long as a measurement is running
[REMOTE]
The instrument is in remote operation
21Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of the Rear Panel
[OUTPUT OFF]
All outputs of the R&S UPV66 audio analyzer are switched off
[GEN OVLD]
Generator is overloaded
[ANLR OVLD]
The low-impedance input resistors (300 Ohm or 600 Ohm) are overloaded and
have been switched off.
[UNBAL]
BNC sockets
Startup
2.2Description of the Rear Panel
This section provides an overview of the ports on the rear panel of the R&S UPV /
R&S UPV66.
22Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of the Rear Panel
2.2.1Remote control connectors
[USB IN]
USB device socket for remote operation of the audio analyzer (with the
R&S UPV‑K4 option)
[LAN]
LAN interface for
Startup
●
Integration of the audio analyzer in a network
●
Remote operation of the audio analyzer from a remote computer
●
Remote control of the audio analyzer (with the R&S UPV‑K4 option)
2.2.2PC Interfaces
[COM]
RS 232-C interface, 9-pin D-Sub connector (not with internal control computer
FMR9. For R&S UPV, the FMR9 is installed for serial number 103000 or higher;
for R&S UPV‑66, it is installed for serial number 101000 or higher; see Chap-
ter 2.6.1, "Retrieving Computer Hardware Information and Windows Operating
System", on page 41).
It is recommended to use an extremely well shielded cable for connecting the
R&S UPV via the RS 232‑C interface.
23Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of the Rear Panel
[LPT]
Parallel printer interface, 25-pin D-Sub socket (not with internal control computer
FMR9. For R&S UPV, the FMR9 is installed for serial number 103000 or higher;
for R&S UPV‑66, it is installed for serial number 101000 or higher; see Chap-
ter 2.6.1, "Retrieving Computer Hardware Information and Windows Operating
System", on page 41).
To connect a printer, use a well-shielded cable.
[MONITOR]
Connector for an external monitor, 15-pin D-Sub socket, 3-tier.
Use an extremely well shielded cable with ferrite cores to connect the external
monitor to the R&S UPV.
[DVI-D]
Startup
Connector for an external digital monitor, 24-pin D-Sub socket, 3-tier (only with
internal control computer FMR9. For R&S UPV, the FMR9 is installed for serial
number 103000 or higher; for R&S UPV‑66, it is installed for serial number
101000 or higher; see Chapter 2.6.1, "Retrieving Computer Hardware Information
and Windows Operating System", on page 41).
Use an extremely well shielded cable with ferrite cores to connect the external
digital monitor to the R&S UPV.
2.2.3USB Sockets
USB (Universal Serial Bus) interfaces of type A (host USB).
●
Connecting peripherals such as mouse, keyboard, printer
●
Connecting a USB stick for transferring files
There are additional USB interfaces on the front of the instrument.
24Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of the Rear Panel
Startup
2.2.4BNC Phone Out
The signal that is output on these BNC sockets is the same as the signal on the
headphone output on the front panel.
Note
Requires the R&S UPV-U2 option (BNC audio monitoring outputs).
2.2.5Trigger, Aux ports
[TRIG IN]
Input for triggering measurements
[TRIG OUT]
Trigger output and clock output
[ANLG AUX OUT]
Additional analog output for special applications
●
DC output, e.g. for supplying power to hearing aids
●
Output for the analog generator signal via an integrated amplifier for the direct
actuation of small loudspeakers
25Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of the Rear Panel
Startup
2.2.6Digital synchronization and expansion interfaces
Inputs and outputs for reference and sync signals for the R&S UPV‑B2 option
(digital audio interfaces)
Note
These interfaces are only available if the R&S UPV‑B2 option is installed.
[SYNC IN]
Synchronization input for word-clock signals
[SYNC OUT]
Output for synchronizing external digital devices with the word clock or bitphase
clock of the R&S UPV
[AUX IN]
Input for a digital audio reference signal (DARS)
[AUX OUT]
Output for a digital audio reference signal (DARS) generated by the R&S UPV
2.2.7IEC/IEEE-bus socket
IEC bus socket (IEC 625/IEEE 488) for remote control of the R&S UPV
26Getting Started 1146.2078.62 ─ 10
R&S®UPV
Note
Requires the R&S UPV‑K4 option (remote control).
2.2.8Mains Voltage Connector
Startup
Description of the Rear Panel
Mains voltage connector with indication of the set nominal mains voltage
Caution
The R&S UPV must be set to the nominal voltage of the AC system used.
Before switching on the instrument for the first time, check whether the correct mains voltage is set.
2.2.9Slots for extension units
Two slots are provided on the rear panel of the R&S UPV for extension units (see
Chap. 4 and data sheet). The following can be installed in addition to or as an
alternative to the UPV‑B2 option (digital audio interfaces AES/EBU, S/P DIF) on
the front panel:
●
1 x UPV-B41 (I²S interfaces)
●
1 x UPV-B42 (Universal Serial Interface)
●
2 x UPV-B48 (Eight-Channel Analog Inputs).
If both slots are occupied by the UPV-B48, 16 analog input channels are available.
27Getting Started 1146.2078.62 ─ 10
R&S®UPV
Description of the Rear Panel
The left plug-in slot at the rear panel of the R&S UPV is designated as Slot 1, and
the right one as Slot 2. The installation manual included with the option refers to
this designation.
Startup
2.2.10I2S interface
The R&S UPV‑B41 option, which is an I²S interface, is shown here as an example
of how an extension unit can be installed in addition to or as an alternative to the
UPV‑B2 (digital audio interfaces AES/EBU, S/P DIF).
[I²S]
25-contact D-Sub connector for connecting the input and output lines of the I²S
interfaces. The interface assignment is described in Chapter 4. To connect the
DUT, use a well-shielded cable.
[TX CLK IN]
Additional BNC connector for connecting an external master clock signal. Details
are provided in Chapter 4.
2.2.11LAN RESET Pushbutton
Note
This pushbutton is present only on the R&S UPV66 model.
The LAN RESET pushbutton has three functions:
●
Resets the LAN default settings:
To do this, you have to press the pushbutton for 5 seconds until the READY
LED switches off on the front panel of the R&S UPV66.
28Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
●
Shuts down the operating system:
If the operating system of the R&S UPV66 is to be shut down manually without using a mouse, this is done by pressing the LAN RESET pushbutton three
times within 5 seconds.
●
Master PRESET:
With the Master PRESET, you can easily get out of an unsuitable screen and
monitor configuration and back to an operable instrument: Switch the instrument off and on again. During the software start, the OUTPUT Off-LED is illuminated briefly in yellow. As soon as this LED is illuminated in yellow, press
the LAN RESET pushbutton briefly. The LED flashes for confirmation. Within
the next 5 seconds, press the LAN RESET pushbutton again briefly. The Master PRESET is now carried out and forces a switchover to the external monitor
with a screen resolution of 800 x 600 pixels and the factory default settings.
Startup
2.3Putting into Operation
The following section describes how to put the instrument into operation, connect
external devices such as printer and monitor, and a connection of the audio analyzer with a network. It contains general safety instructions for instrument operation.
For information about installing options, see Chapter 2.9, "Installing Options",
on page 59.
For information about the firmware update, see Chapter 2.11, "Firmware Update",
on page 81.
Safety precautions
Be absolutely sure to follow the instructions in the sections below to prevent
injury to people or damage to the instrument. This is particularly important if
the instrument is used for the first time. The general safety instructions at
the beginning of this manual must also be observed.
29Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
Startup
2.3.1Unpacking the instrument
The instrument is delivered together with the accessories in one box. Proceed as
follows to unpack its contents:
1. Remove the instrument from its packaging and check the shipment for completeness using the shipping document and the accessory lists for the various
items.
2. First, pull off the protective plastic pads from the rear feet of the instrument
and then from the front handles.
3. Remove the protective cover made of corrugated board from the rear of the
instrument.
4. Carefully unthread the corrugated cardboard cover at the front that protects
the instrument handles and remove it.
5. Check the instrument for any damage. If there is damage, immediately contact
the carrier who delivered the instrument. In this case, do not discard the box
and packaging material.
Packaging material
It is recommended to retain the packaging. It is advisable to keep the original packing material in order to prevent control elements and connectors
from being damaged in case the instrument is to be transported or shipped
at a later date.
2.3.2Setting Up the Instrument
The instrument is intended for indoor use. It can either be set up as a separate
unit or mounted in a 19" rack.
The following ambient conditions at the location must be ensured:
●
The ambient temperature must comply with the range specified in the data
sheet.
●
All fan openings must be clear, and the airflow at the openings of the side
walls must not be impeded. The distance to the wall must be at least 10 cm.
30Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
Possible damage from electrostatic discharge
Electrical discharges can damage components of the instrument or a connected device under test.
For this reason, the instrument may be operated only at a workplace that is
protected against electrostatic discharge.
The following methods can be used separately or in combination to protect
against electrostatic discharges:
●
Protective armband with grounding cable
Startup
●
Conductive floor covering combined with use of a heel grounder
EMI Suppression
To suppress electromagnetic interference (EMI), the instrument may be
operated only if it is closed and all covers are in place. The EMC class is
listed on the data sheet.
The following conditions must be maintained to avoid interference:
●
Use suitable double-shielded cables that are no longer than 3 m.
●
Use only USB cables that are no longer than 1 m.
●
Use only USB devices that adhere to the permissible EMC values.
●
Terminate the outputs and cables with 50 Ω.
●
Use only the cables R&S UP-Z2, R&S UP-Z3 and R&S UP-Z4 for the
digital interfaces.
●
Use the cable R&S UP-Z8A for the analog interface of option R&S UPPB8, and cable R&S UP-Z8D for the digital interface.
●
Use only doubled shielded CAT6 or CAT7 cables for the LAN interface.
31Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
2.3.2.1Separate installation
The R&S UPV can be set up on a level surface in horizontal position and with
unfolded instrument feet.
Risk of injury with unfolded instrument feet
The instrument feet can suddenly fold back if the instrument is being moved
or the feet are not completely folded out. This can lead to personal injuries
or damage to the instrument.
Instrument stability, and thus safe operation, is only ensured when the
instrument's feet are completely folded out. Do not move the instrument and
avoid performing work under it when the feet are folded out. Secure the
instruments from slipping (e.g. by locking the instrument feet to the top of
the front-panel frame).
Startup
If excessive loads are applied, the feet might break. The uniformly applied
load to the feet when folded out must not exceed 500 N (including the
weight of the instrument itself along with any equipment set on top of it).
2.3.2.2Rack installation
A rack adapter (refer to data sheet for Order No.) is required for installation in a
19" rack. The mounting instructions are supplied with the adapter.
32Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
Possible damage to the instrument from overheating
Restricted airflow at the side wall openings can cause the instrument to
overheat.
To ensure sufficient air supply, all fan openings must be clear and the airflow at the openings of the side walls must not be impeded. The distance to
the wall must be at least 10 cm.
Startup
33Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
2.3.3Connecting the R&S UPV to the AC Supply
Possible damage to the instrument
Before connecting and switching on the instrument, observe the following
items to avoid any damage to the instrument:
●
The tube for the housing has to be attached and screwed in place.
●
The ventilation openings must be clear; ensure that air can freely exit at
the holes on the sides. The minimum distance to the wall should therefore be at least 10 cm.
●
The instrument must be dry.
●
The instrument should only be operated in the horizontal position on an
even surface.
Startup
●
The ambient temperature must not exceed the range specified in the
data sheet.
●
It must be ensured that none of the voltage levels at the inputs exceed
permissible limits.
●
The instrument complies with measuring category I; the input voltage at
the connectors of the analog analyzers should not exceed 110 V (effective, sinusoidal) and 160 V (peak value). Do not use the instrument in
measuring categories II, III and IV.
Explanation:
Measuring circuits as defined in section 6.7.4 of EN61010-1
Measuring category I is intended for measurements on circuits which
are not connected to the high-voltage current system.
●
The outputs of the instrument should not be overloaded and correct
polarity must be ensured.
The AC supply connector socket is at the rear of the instrument. The instrument
can be set up for connection to AC supply systems of 100 V, 120 V, 220 V and
230 V and can be operated with ±10 % tolerance and an AC supply frequency of
47 Hz to 63 Hz. The instrument is also suitable for operation with power supply
systems with the other nominal voltages specified in the table below
34Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Putting into Operation
Nominal AC supply voltageVoltage selector settingTolerance of instrument
110 V100 V+4 %
–18 %
120 V+20 %
–6 %
115 V120 V+15 %
–10 %
127 V120 V+4 %
–18 %
240 V230 V+10 %
–14 %
2.3.3.1Setting the Supply Voltage
Possible damage to the instrument from incorrect AC supply voltage
An incorrect AC supply voltage setting can lead to damages to the instrument when the instrument is connected to the AC supply.
For this reason, check whether the correct AC supply voltage is set before
switching on the instrument for the first time.
The AC supply voltage can be changed as follows:
1. Disconnect the power cable.
2. Open the flap covering the voltage selector using a small screwdriver (or similar).
3. Remove the cylinder labeled with the nominal voltages and reinsert it so that
the value visible through the hole in the cover flap when refitted corresponds
to the desired nominal voltage. If the desired voltage is not printed on the cylinder, select the nearest suitable value according to the table above.
4. Close the flap.
35Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
The R&S audio analyzer UPV is operated with the same AC fuses for all
specified rated AC supply voltages. The precise designation of the fuses is
indicated on the rear of the instrument.
2.3.3.2Switching on the Instrument
1. Connect the audio analyzer to the AC supply by means of the supplied power
cable.
The R&S UPV is in compliance with safety class EN61010-1, it should only be
connected to a socket with grounding contact.
Startup
2. Press the power switch on the front of the instrument. The instrument starts
the booting process and all modules are supplied with power.
2.3.3.3Start Screen and Booting of the R&S UPV
After the instrument is turned on, the BIOS version that is installed and the computer characteristics are displayed on the screen for several seconds.
Afterwards, the Windows operating system boots first followed by the instrument
firmware. A self-test is carried out while instrument firmware is booted. The most
recent instrument settings saved automatically at power-off are then loaded so
that the instrument is in the same state as it was before power-off. After booting is
completed, the audio analyzer screen is displayed and the instrument can be
operated manually or via remote control.
36Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
2.3.3.4Restarting the R&S UPV
If the R&S UPV firmware was closed without shutting down the operating system
(e.g. by clicking on the icon in the title bar or ALT F4 on the keyboard), the program can be restarted with the mouse in two ways:
●
Clicking on the "R&SUPV" icon on the desktop
●
Clicking on the Start button in the taskbar, and then selecting the folders "All
Programs, Rohde & Schwarz, UPV" and clicking on the "R&S UPV" icon
If the software should crash unexpectedly and the instrument is operated
via the front panel, it can only be rebooted by pressing the power switch.
When using an external keyboard and a mouse, we recommend carrying
out a warm start first using the shortcut Ctrl+Alt+Del (restart) or trying the
Restart icon with the green arrow.
Startup
2.3.3.5Switching Off
Risk of data loss
There is a risk of data loss if the R&S UPV is shut down without shutting
down the operating system first. This may cause trouble during the boot
process the next time the instrument is switched on.
We strongly recommend shutting down the operating system before switching off the instrument ([SHUTDOWN] button or via mouse operation). Shutting down the operating system ensures that open files are closed properly.
Model R&S UPV
1. Press the [SHUTDOWN] button on the front of the instrument.
The R&S UPV stores the current settings on the hard disk and then shuts
down the operating system.
37Getting Started 1146.2078.62 ─ 10
R&S®UPV
Putting into Operation
2. Power switch [POWER] on the front of the instrument
The instrument is disconnected from the AC supply, and power is removed
from all modules.
Alternatively, the operating system can also be shut down in the same way
as other Windows applications by using the mouse: Start, Shut Down, OK
or the red Shutdown icon.
Model R&S UPV66
1. Press the [LAN RESET] button at the rear of the instrument three times within
5 seconds.
The R&S UPV stores the current settings on the hard disk and then shuts
down the operating system.
Startup
2. Wait until the operating system has shut down and the [SWITCH OFF] LED is
on.
3. Press the [POWER] switch on the front of the instrument.
The instrument is disconnected from the AC supply, and power is removed
from all modules.
For remote control, you can also shut down the operating system by using
the SYSTem:SHUTdown SCPI command.
Alternatively, the operating system can also be shut down in the same way
as other Windows applications by using the mouse: Start, Shut Down, OK
or the red Shutdown icon.
38Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Presets
2.4Function Check
The audio analyzer monitors the most important instrument functions automatically when it is switched on.
2.5Presets
Pressing the [PRESET] key loads a set of defined instrument settings.
The preset function sets all parameters and settings, even those of inactive operating modes.
Possible malfunction of the instrument function due to external software
The instrument operates under the Windows operating system. This allows
external software programs to be installed on the instrument. The use and
installation of commercial off-the-shelf (COTS) software may impair the
instrument's functionality.
For this reason, we recommend that you only execute programs tested by
Rohde & Schwarz with regard to their compatibility with the instrument software. In certain cases, the use of these programs may also impair the performance of the instrument.
Startup
Windows Operating System
The drivers and programs used in the instrument under Windows have
been adapted to the measuring instrument. Existing instrument software
must only be modified with update software released by Rohde & Schwarz.
Only the settings described below may be performed.
In its delivery state, the configuration of Windows is optimized to the functions of
the audio analyzer. Changes in the system setup are only required when peripherals like a keyboard or printer are installed or in the event that the network configuration does not correspond to the default settings (see sections below). After
switching on the audio analyzer, the operating system boots and the instrument
firmware is started automatically without a password prompt (auto login).
Auto login is performed with the user name and the password "instrument".
The user has administrator rights so that printer and network installations
are possible.
The operating system can only be accessed if an external keyboard is connected.
The Windows key on the external keyboard (next to the [STRG] key) opens the
Windows start menu; the Windows programs can then be called from this menu.
Connecting a mouse makes Windows easier to use. The system settings under
Windows are entered in the menu "Start – Control Panel" (see the description of
Windows and the hardware for information on the necessary settings). The R&S
UPV does not feature a diskette drive. Data is exchanged using a USB stick
40Getting Started 1146.2078.62 ─ 10
R&S®UPV
Windows Operating System
which is inserted into one of the USB interfaces. The USB stick is automatically
assigned a free drive letter and the data can be transferred in Windows Explorer.
Startup
2.6.1Retrieving Computer Hardware Information and Windows
Operating System
The function of some interfaces depends on which internal computer hardware
the R&S UPV has been equipped with. This description makes reference to the
computer hardware at some points. With firmware versions 3.x and older, the
Windows version and the computer hardware information can be retrieved as
follows: Call up the system settings under Windows via the menu "Start – Control
Panel – System". This opens a window with the name "System Properties". The
desired information can be found under the "General" tab:
With firmware version 4.x, the computer hardware information can be
checked directly in the firmware via "Help" – "About" under "CPU type:", and the
Windows version can be checked under "Windows:".
41Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connecting External Keyboard and Mouse
2.7
Connecting External Keyboard and Mouse
2.7.1Connecting an External Keyboard.
The R&S UPV allows connecting a commercially available, external keyboard
with USB interface. A keyboard simplifies the input of list entries, comment texts,
file names, etc. and is required for operating Windows.
The keyboard is connected to one of the USB interfaces at the front or rear of the
instrument.
The keyboard is detected automatically when it is connected. The US keyboard
assignment is the default setting. A change of the keyboard language assignment
and special settings, such as the repeat rate etc., can be performed in the Windows menu "Start - Control Panel - Keyboard" or "Regional and Language
Options". The menu is opened by pressing the Windows key on the external keyboard.
2.7.2Connecting an External Mouse
The R&S UPV allows connecting a commercially available mouse with USB interface. The mouse makes it easier to move and resize the panels / graphical windows on the screen and provides alternative operating modes. This is recommended for convenient operation of Windows.
42Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
The mouse is connected to one of the USB interfaces on the front or rear of the
instrument.
The mouse is detected automatically when it is connected. Special settings, such
as the speed of the mouse cursor etc., can be performed in the Windows menu
"Start - Control Panel - Mouse". The menu is opened by pressing the Windows
key on the external keyboard.
Startup
2.8Connecting an External Monitor
Possible Impairments to Instrument Function
Make sure that the instrument is switched off before connecting the monitor.
Otherwise, the monitor or the R&S UPV may be damaged.
Do not modify the screen driver (Display type) and the display configuration
since this impairs instrument operation.
For instruments with computer hardware FMR9 and FMR11, the R&S UPV provides the possibility of connecting an external monitor. The monitor can be connected with the analog interface to the MONITOR port or with digital interface to
the DVI‑D socket on the rear of the instrument (see Chapter 2.6.1, "Retrieving
Computer Hardware Information and Windows Operating System", on page 41).
Connection
To prevent damage to the monitor and R&S UPV, the instrument must be
switched off before an external monitor is connected. After connection, the external monitor is detected when the instrument is started (but not if it is connected
while the instrument is in operation). The screen contents with all elements, e.g.
panels, measurement displays, softkey bar, etc., appear on the external monitor.
No further settings are needed.
43Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
Startup
2.8.1Changing Screen Resolution of External Monitor under
Windows XP
The R&S UPV audio analyzer is equipped with an internal computer. Up to February 2007, the instruments were supplied with a computer bearing the internal
Rohde & Schwarz designation FMR6, from March 2007 with FMR7, from March
2012 with FMR9 and from February 2015 with FMR11 (see also Chapter 2.6.1,
"Retrieving Computer Hardware Information and Windows Operating System",
on page 41). Instruments with FMR11 support Windows 7 only; see the next
chapter. The procedure for setting the screen resolution when using an external
monitor is different for the three computer models. The procedure for each is
described separately below.
Devices with FMR6
Setting external monitor resolution to higher than 800 x 600 pixels
1. Switch off R&S UPV.
2. Connect the external monitor, keyboard and mouse.
3. Switch on R&S UPV.
4. After starting the R&S UPV, exit or minimize the UPV program by clicking the
buttons on the right in the title bar.
5. Right-click on the desktop to open the following context menu.
6. Select "Graphics properties…".
7. Click the "Monitor" icon.
44Getting Started 1146.2078.62 ─ 10
R&S®UPV
8. Place the focus on the "True Color" field.
Startup
Connecting an External Monitor
9. Select the hidden button (press the [Tab] key three times, then press the
spacebar and click the "Apply" button).
The internal LC screen (digital display) is switched off.
10. Click "OK" to confirm the change, then on "OK" to close the window.
You can now set the external monitor to a resolution higher than 800 x 600
pixels.
11. Right-click on the desktop to open the following context menu again:
45Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
12. Select "Graphics properties…".
13. In the Screen Area field, select the desired resolution, e.g. "1024 by 768".
Startup
14. Click "Apply".
15. Click "OK" to confirm the change, and then click "OK" again to close the window.
Resetting the External Monitor to a Resolution of 800 x 600 Pixels
The external monitor, mouse and keyboard are connected to the R&S UPV, the
R&S UPV is switched on.
1. Exit or minimize the UPV program by clicking the buttons on the right in
the title bar.
2. Right-click on the desktop to open the following context menu again:
46Getting Started 1146.2078.62 ─ 10
R&S®UPV
3. Select "Graphics properties…".
4. Click on the "Twin View" icon.
Startup
Connecting an External Monitor
5. Place the focus on the "Monitor" field as shown in the figure below.
47Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connecting an External Monitor
6. Select the hidden button (press the [Tab] key three times), then press the
spacebar and click the "Apply" button.
The resolution is reset to 800 x 600 pixels. This resolution is the same as for
the internal LC display, which is switched on again.
7. Click on "OK" to confirm the change, then on "OK" to close the window.
Devices with FMR7
Setting external monitor resolution to higher than 800 x 600 pixels
1. Switch off R&S UPV.
2. Connect the external monitor, keyboard and mouse.
3. Switch on R&S UPV.
4. After starting the R&S UPV, exit or minimize the UPV program by clicking the
buttons on the right in the title bar.
48Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
5. Right-click on the desktop to open the following context menu.
6. Select "Monitor".
Switch off the internal LC screen and switch on the external monitor if it was
previously switched off.
You can now set the external monitor to a resolution higher than 800 x 600
pixels.
Startup
7. Right-click on the desktop to open the following context menu.
8. Select "Graphics Properties ...".
9. Select "Display Settings" and then choose the desired resolution, e.g.
"1024 x 768".
49Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
10. Click on "Apply".
11. Click "OK" to confirm the change, and then click "OK" again to close the window.
Startup
Resetting the External Monitor to a Resolution of 800 x 600 Pixels
The external monitor, mouse and keyboard are connected to the R&S UPV, the
R&S UPV is switched on.
1. Exit or minimize the UPV program by clicking the buttons on the right in
the title bar.
2. Right-click on the desktop to open the following context menu again:
3. Select "Digital Display + Monitor".
The resolution is reset to 800 x 600 pixels. This resolution is the same as for
the internal LC display, which is switched on again.
50Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
Switching Off the External Monitor
The external monitor, mouse and keyboard are connected to the R&S UPV, the
R&S UPV is switched on.
1. Right-click on the desktop to open the following context menu again:
Startup
2. Select "Digital Display".
The external monitor is switched off.
Devices with FMR9
Special features with the FMR9:
The Windows XP operating system used is only able to support a maximum of
two monitors simultaneously. One of the two monitors is the internal LC display.
This means that only one external monitor can be controlled even if the R&S UPV
with FMR9 has an analog VGA and a digital DVI port. The default setting
assumes that the internal monitor and one external analog VGA monitor is used.
The connected external monitor determines the behavior during the boot procedure depending on the selected setting under Windows: If an external analog
VGA monitor is connected and "Notebook + Monitor" has been selected under
Windows, both screens show the same content during the boot procedure and
when Windows is running. If a digital DVI monitor is connected, the internal display remains dark during the boot procedure until Windows becomes active,
regardless of its Windows setting. Only after the boot procedure both displays
show the same content in parallel if "Notebook + Digital Display" was selected
under Windows. Otherwise only the internal display becomes active under Windows.
51Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
Setting the External Monitor Resolution to Higher than 800 x 600 Pixels
1. Switch off R&S UPV.
2. Connect the external monitor, keyboard and mouse.
3. Switch on R&S UPV.
4. After starting the R&S UPV, exit or minimize the UPV program by clicking the
buttons on the right in the title bar.
5. Right-click on the desktop to open the following context menu.
Startup
6. Select "Monitor".
Switch off the internal LC screen and switch on the external monitor if it was
previously switched off.
You can now set the external monitor to a resolution higher than 800 x 600
pixels.
7. Right-click on the desktop to open the following context menu.
8. Select "Graphics properties…".
9. Select "Display Settings", then select the desired resolution, e.g. "1024 by
768".
52Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
10. Click "Apply".
11. Click "OK" to confirm the change, and then click "OK" again to close the window.
Startup
Resetting the External Monitor to a Resolution of 800 x 600 Pixels
The external monitor, mouse and keyboard are connected to the R&S UPV, the
R&S UPV is switched on.
1. Exit or minimize the UPV program by clicking the buttons on the right in
the title bar.
2. Right-click on the desktop to open the following context menu again:
3. Select "Notebook + Monitor".
The resolution is reset to 800 x 600 pixels. This resolution is the same as for
the internal LC display, which is switched on again.
On some monitors, it is possible to utilize only a limited part of the screen –
and not the entire screen – after switching the resolution. To expand the contents to match the entire screen area, the following settings must be performed: Right-click the desktop to open the following context menu.
53Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
Select ""Graphics Properties"". Under "Display Settings" and the "Monitor" tab,
select a sampling frequency of 72 Hz in the "Display Expansion" field.
Startup
Clickon "Apply", then click "OK" twice to apply the change. The content is then
expanded to the entire screen area.
Switching Off the External Monitor
The external monitor, mouse and keyboard are connected to the R&S UPV, the
R&S UPV is switched on.
1. Right-click on the desktop to open the following context menu again:
54Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
2. Select "Notebook".
The external monitor is switched off.
Setting the External Monitor Resolution (Digital DVI Monitor) to Higher Than
800 x 600 Pixels
1. Switch off R&S UPV.
2. Connect the external monitor, keyboard and mouse.
3. Switch on R&S UPV.
4. After starting the R&S UPV, exit or minimize the UPV program by clicking the
buttons on the right in the title bar.
5. Right-click on the desktop to open the following context menu.
Startup
6. Select "Digital Display".
Switch off the internal LC screen and switch on the external monitor if it was
previously switched off.
You can now set the external monitor to a resolution higher than 800 x 600
pixels.
7. Right-click on the desktop to open the following context menu.
55Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
8. Select "Graphics properties…".
9. Select "Display Settings", then select the desired resolution, e.g. "1024 by
768".
10. Click on "Apply".
Startup
11. Click on "OK" to confirm the change, then on "OK" to close the window.
Resetting the External Monitor to a Resolution of 800 x 600 Pixels
The external monitor, mouse and keyboard are connected to the R&S UPV, the
R&S UPV is switched on.
1. Exit or minimize the UPV program by clicking on the
in the title bar.
2. Right-click on the desktop to open the following context menu again:
3. Select "Notebook + Digital Display".
buttons on the right
The resolution is reset to 800 x 600 pixels. This resolution is the same as for
the internal LC display, which is switched on again.
56Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
Switching Off the External Monitor
The external monitor, mouse and keyboard are connected to the R&S UPV, the
R&S UPV is switched on.
1. Right-click on the desktop to open the following context menu again:
2. Select "Notebook".
The external monitor is switched off.
Startup
Internal Display Not Activated
The graphics driver of the FMR7 and FMR9 computer modules in the R&S
UPV saves the screen configuration that was last selected and detects
whether an external monitor is connected.
If the R&S UPV was previously operated only with an external monitor and
was switched off in that state, the internal display will remain deactivated
the next time the R&S UPV is switched on and no external monitor is connected.
Since no external monitor is detected in this case when the instrument is
switched on, the monitor output remains deactivated even if a monitor is
connected later .
An external keyboard is required to reactivate the screen display:
You can switch to the external monitor by using the key combination CTRL
+ ALT + F1 on an external keyboard.
Before March 2008: You can switch on the internal display by using the
key combination CTRL + ALT + F4 on an external keyboard.
Since March 2008, the R&S UPV has been shipped with a new LCD that
can be switched on using the shortcut CTRL + ALT + F3.
57Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connecting an External Monitor
A special feature of the R&S UPV66 is that it has no internal display. It is therefore possible to control an analog VGA and a digital DVI monitor simultaneously.
The following settings must be performed to enable this feature:
Using the shortcut CTRL + ALT + F1 on an external keyboard switches on the
external VGA monitor; the shortcut CTRL + ALT + F4 switches on the external
DVI monitor if it is to be activated subsequently.
Startup
2.8.2Changing Screen Resolution of External Monitor under
Windows 7
The R&S UPV audio analyzer has a built‑in computer. Up to February 2007, the
instruments were supplied with a computer bearing the internal Rohde & Schwarz
designation FMR6, from March 2007 with FMR7, from March 2012 with FMR9
and from February 2015 with FMR11. Instruments with FMR6 do not support Windows 7. The procedure for setting the screen resolution when using an external
monitor is the same for the different computer models and is described below.
The default factory setting supports clone mode of the internal and, if connected,
external VGA screen and means that both screens show the same picture. In this
mode, the screen resolution cannot be changed and remains fixed at the resolution of the internal screen, i.e. 800 x 600 pixels. Instruments with FMR9 andFMR11 support an external monitor with analog interface at the VGA port or with
digital interface at the DVI‑D port on the rear of the instrument. Switchover takes
place automatically.
Setting external monitor resolution to higher than 800 x 600 pixels
1. Switch off R&S UPV.
2. Connect the external monitor, keyboard and mouse.
58Getting Started 1146.2078.62 ─ 10
R&S®UPV
Installing Options
3. Switch on R&S UPV.
4. After starting the R&S UPV, exit or minimize the UPV program by clicking the
buttons on the right in the title bar.
5. Right‑click the desktop to open the context menu and then select "Graphics
Options" and "Output To".
6. If the external monitor is connected to the VGA output: Select "Monitor". If the
external monitor is connected to the DVI‑D output: Select "Digital Display".
Switch off the internal LC screen and switch on the external monitor if it was
previously switched off.
You can now set the external monitor to a resolution higher than 800 x 600
pixels.
7. Right-click the desktop to open the context menu and then select "Screen resolution". A Windows dialog box opens in which the screen resolution can be
changed in the "Resolution" field.
Startup
8. Click "Apply" and then "OK" to apply the new screen resolution.
Switching on internal screen
Proceed as described in the section "Setting external screen resolution to higher
than 800 x 600", point 1 to point 5, then:
1. If only the internal display is required, select "Built-In Display". The external
monitor is switched off.
2. If clone mode is required, select "Clone Display" and then select "Built-in Display + Monitor" for an external monitor at the VGA port or "Built-in Display +
Digital Display" for the DVI‑D port. The internal screen and the external screen
are switched on.
2.9Installing Options
The R&S UPV can be equipped with a few options. These options can be divided
into two basic categories:
59Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
Software optionsThese can be installed by the user alone; installation instructions
and a unique activation code supplied with the option are required.
Hardware optionsThese are installed in the R&S UPV audio analyzer, can be retro-
fitted only by a Rohde & Schwarz service facility. "Exception: The
plug-in cards for the R&S UPV-B41, -B42 and -B48 options can
also be installed by the user by following the information listed in
the installation instruction since they do not require that the base
unit be opened."
2.10Connect R&S UPV to a Network (LAN)
The R&S UPV has a network connector and can be connected to a LAN (Local
Area Network).
Network administrators with the appropriate rights can transfer files across the
network and use network resources such as the network directories or printers.
The instrument can also be remote-controlled and manually operated in the network.
The manual remote allows operating the R&S UPV from any remote external
computer. For example, the user can operate one or several R&S UPVs from his
workplace that are integrated inn a test setup in another part of the building.
The R&S UPV is shipped with activated Windows firewall. File transfer via LAN or
manual remote operation using the remote desktop requires only a partial exception in the firewall. It is advisable to coordinate the configuration of the firewall
with the network administrator.
The remote control of the instrument via the LAN interface is described in the
operating manual on the documentation CD.
60Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
Startup
2.10.1Connection to the Network
Possible malfunction of the network operation
It is recommended that the connection of the instrument to the network be
coordinated with the network administrator. Connection errors may affect
the entire network.
Do not connect or disconnect the network cable until the instrument is
switched off (standby). Only then can the network connection be reliably
detected and impairments to instrument operation can be avoided.
The instrument is connected to the network using a commercially available RJ-45
cable connected to the LAN interface at the rear of the instrument. In order to
supress electromagnetic interference, it is recommended to use suitable shielded
CAT6 or CAT7 cables of high quality.
2.10.2Configuration of R&S UPV for Network Operation
The network interface operates with 100-MHz Ethernet IEEE 802.3u. The TCP/IP
network protocol and the associated network services are preconfigured.
In order to exchange data in a local network (LAN), each connected computer or
instrument must have a unique IP address or a unique computer name. In addition, network access of the different users is organized by the assignment of
access rights.
The access rights determine which of the available network resources, e.g. file
storage systems, are available for the R&S UPV.
Networks with DHCP
The R&S UPV is preconfigured for networks using DHCP (dynamic host configuration protocol). In these networks, the R&S UPV is automatically assigned an
unused IP address.
61Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
In this case, a unique computer name is used for identification in the network.
Every R&S UPV is assigned an individual computer name at the factory. This
name can be queried in the Windows menu "Start - My Computer" and changed
(see Chapter 2.10.2.3, "Querying the computer name", on page 66).
Networks with fixed IP addresses
In networks using fixed IP addresses, the addresses are mostly assigned by the
network administrator. The fixed IP address must be entered in the Windows
menu "Start - Control Panel" (see Chapter 2.10.2.2, "Entering the IP Address",
on page 63).
Point-to-point connections
To set up a simple network – a LAN connection between an R&S UPV and a
computer without integration in a larger network – an IP address must be
assigned for the R&S UPV and the computer. The IP addresses 192.168.xxx.yyy
are available, whereby both xxx and yyy can have the values 1 to 254; the value
for the subnet mask is 255.255.255.0.
Startup
In this case, a commercially available crossover RJ-45 cable must be used for the
connection. In order to supress electromagnetic interference, it is recommended
to use suitable shielded cables of high quality.
User ID
For the R&S UPV, the user instrument has been defined. The user is used for
the auto login when the instrument is started and for manual remote operation.
The password is also instrument. By assigning the respective rights, the network
administrator decides which directories and resources in the network can be
accessed by the R&S UPV.
2.10.2.1 Preparations
The instrument is configured for network operation in the menus of the Windows
operating system. The operating system can only be accessed if an external keyboard is connected; connection of a mouse is recommended for easy operation.
The instrument should be switched off before the keyboard and mouse are connected. This will ensure that the keyboard and mouse are correctly detected by
the operating system.
62Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
2.10.2.2 Entering the IP Address
By default, "Obtain an IP address automatically " (DHCP (Dynamic Host Configuration Protocol) is preset.
When using Windows XP:
1. "Click on Start – Settings – Control Panel".
Startup
2. "Click on Network and Internet Connections" and then click on "Network Connections" at the bottom right in this menu.
3. Right-click in the menu "Network Connections Local Area Connection" and
then click on "Properties".
63Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
4. On the "General" tab in the field "This connection uses the following items:",
mark the "Internet Protocol (TCP/IP)" selection and then click on "Properties".
Startup
5. In the "Internet Protocol (TCP/IP) Properties" menu, enter the IP address in
the "Use the following IP address" field (the complete data are available from
the network administrator). Complete the entry with "OK" in all menus.
64Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
When using Windows 7:
1. Open the Control Panel, then select "Network and Sharing Center".
2. Click on "Change adapter settings" and then right-click on "Local Area Connection"
65Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
The "Networking" tabs opens.
3. Select "Internet Protocol Version 4 (TCP/IPv4)" and then click on "Properties".
4. Select "Use the following IP address:" and enter the IP address.
2.10.2.3 Querying the computer name
When using Windows XP:
1. "Click on Start – Control Panel: Performance and Maintenance"
2. "Click on System" and select "Computer Name" in the tab menu.
The computer name is displayed under "Full Computer Name". It can be
changed in the "Change" submenu.
66Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
When using Windows 7:
► "Click on Start – Control Panel - System"
The computer name is displayed and can be changed.
2.10.2.4 Accessing directories in the network
Access to network drives depends on access rights and whether the drives in
question have been enabled. The complete integration of R&S UPV in a larger
network with the required rights assignment is very complex and is generally performed by a network administrator.
Access from R&S UPV to the hard disk of a computer that is also connected to
the network can, however, be performed relatively easy. The directory which the
R&S UPV is to access simply has to be enabled on the remote computer. Access
from R&S UPV to this directory is done with the help of a Windows search function.
This procedure is also of importance for a point-to-point connection, for instance
in order to start a firmware update for which the files have been stored on the
hard disk of the remote computer.
67Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
Both the computer and the audio analyzer must have a computer name and
an IP address (see Chapter 2.10.2, "Configuration of R&S UPV for Network
Operation", on page 61).
The menu name may deviate from the name specified in the following operating sequence, depending on the language and on the operating system
used on the computer.
Enabling the desired directory on the remote computer
1. On the computer, select the directory to be enabled in Windows Explorer and
right-click to call up the "Properties" menu.
2. In the "Sharing" panel, activate the "Share this folder" check box.
3. Record the computer name of the PC (see Chapter 2.10.2.3, "Querying the
computer name", on page 66).
Startup
At the R&S UPV, access the enabled directory
1. Use the Windows key to call up the "Start" menu.
2. In the menu, select "Search Computers or People", followed by "A Computer
on the Network".
3. In the input window with the query "Which computer are you looking for?",
enter the name of the computer and start the search with Enter.
The PC is listed with its computer name as the search result.
4. Click on the computer name. The enabled directory is displayed and the files
stored in it can be used in the R&S UPV.
Note: If a user name and password are requested when you click the computer, the login name and password used on the computer must be entered.
68Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
Startup
2.10.3Configuration via LXI
Be default, LXI is only available to the R&S UPV66, but it can also be installed on the R&S UPV at a later time, if required.
The LXI functionality for the R&S UPV66 requires at least the Windows XP
operating system with Service Pack 2 which may not yet be installed on
older computers. If this is the case, please contact your Rohde & Schwarz
subsidiary. The version of the operating system can be checked under
"Instrument Properties" in the "General" menu.
LAN eXtensions for Instrumentation (LXI) is an instrumentation platform for measuring instruments and test systems that is based on standard Ethernet technology. LXI is intended to be the LAN-based successor to GPIB, combining the
advantages of Ethernet with the simplicity and familiarity of GPIB. This standard
provides a new way for configuring virtual front panel instruments by using a network and a Web interface. The main tool for this configuration method is a
browser, such as Microsoft Internet Explorer, but it is also possible to use other
browsers such as Firefox or Netscape.
2.10.3.1 LXI classes and LXI functionality
LXI-compliant instruments are divided into three classes, A, B and C, with the
functionality of the classes hierarchically based one upon the other:
●
Class C instruments are characterized by a common LAN implementation,
including an ICMP ping responder for diagnostics. The instruments can be
configured via a web browser; a LAN Configuration Initialize (LCI) mechanism
resets the LAN configuration. LXI instruments of class C must also support
automatic detection in a LAN via the VXI-11 discovery protocol and programming by means of IVI drivers.
69Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
●
Class B adds IEEE 1588 Precision Time Protocol (PTP) and peer-to-peer
communication to the base class C. IEEE 1588 allows all instruments on the
same network to automatically synchronize to the most accurate clock available and then provide time stamps or time-based synchronization signals to all
instruments with exceptional accuracy.
●
Class A-instruments are additionally equipped with the eight-channel hardware trigger bus (LVDS interface) defined in the LXI standard.
Instruments of classes A and B can generate and receive software triggers via
LAN messages and communicate with each other without involving the controller.
The R&S UPV66 meets the general requirements of LXI Class C. In addition to
the properties described above, the instrument offers the following LXI-related
functions:
●
Integrated display of the current instrument settings
●
Possibility of remote controlling the instrument within the WEB browser.
Startup
2.10.3.2 LXI LAN Indicator
The green LAN LED on the front panel of the R&S UPV66 shows the LAN status
of the instrument. The LAN LED indicator has three states.
●
LAN LED off
–The instrument does not have valid IP address. This may be due to one of
the following reasons:
1) the instrument is configured with DHCP, but no DHCP router is connected,
2) the instrument is manually configured and an IP address conflict is present,
3) the instrument lost its IP address that was configured with DHCP and
switched to Auto IP.
This state is called "error state" and normally can be resolved by the LCI
mechanism (LAN reset). This mechanism is required by the LXI standard
and can be invoked with the "LAN RESET" button (key press > 10 s) on
70Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
the rear side of the instrument (see Chapter 2.2.11, "LAN RESET Pushbut-
ton", on page 28).
●
LAN LED is on
–The instrument has a valid IP address and is reachable via TCP/IP.
●
LAN LED flashes
–The instrument is in the "Device Indicator" state initiated by the WEB inter-
face. This state is required by the standard and is used for finding instruments in racks (see Chapter 2.10.3.4, "LXI Browser Interface",
on page 71).
2.10.3.3 LXI Default State of Network Settings
After a LCI (LAN reset), the instrument automatically reboots and starts up with
following LAN settings:
ParameterSetting
HostnameRSUPV66vvv-xxxxxx
DescriptionAudio Analyzer
TCP/IP ModeDHCP + Auto IP Address
Dynamic DNSEnabled
ICMP PingEnabled
NegotiationAuto Detect
VXI-11 DiscoveryEnabled
Password for LAN ConfigurationLxiWebIfc
More information about the LXI standard is located on the LXI website
under
http://www.lxistandard.org
or in the article about LXI in
"News from Rohde & Schwarz, 2006/II - 190".
2.10.3.4 LXI Browser Interface
The only information necessary for performing configuration via the LXI/Web
interface is the name of the instrument (which means its Windows name) in net-
71Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
works that operate with DHCP enabled, or its IP address in networks that operate
with manual network configuration. In rare cases where neither one of these
names is known, it is also possible to derive the IP address from the MAC
address, which is printed on the rear panel of the instrument.
The instrument's LXI browser interface works correctly with all W3C compliant
browsers.
To start the instrument home page (welcome page), open a browser on the PC
and enter the instrument name or the IP address of R&S UPV66 in the address
field, e.g.
http://RSUPV66-100014 or http://10.215.1.107.
The instrument name for the R&S UPV66 is always RSUPV66-xxxxxx, whereby
xxxxxx represents the six-digit serial number of the instrument.
Startup
The instrument home page shows all important information in a compact form.
The MAC address and the (often dynamically assigned) IP address as well as the
complete resource string for the remote control via VISA are also located here.
72Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
The LAN indicator on the instrument will blink if Device Indicator is set to ACTIVE.
Clicking on "INACTIVE (press to toggle)" switches the status on and off. It can be
used to quickly identify an instrument that is mounted together with several other
instruments of the same type, such as in a 19" rack.
Clicking on "ACTIVE (press to toggle)" switches the flashing state off again.
The left side shows a navigation bar containing the important "LAN Configuration"
menu. This page of the LXI Web configuration website enables extensive configuration of all important LAN parameters, which is necessary because the configuration steps mentioned in the next sections cannot be performed as described
with a virtual front panel instrument.
Startup
Changing the LAN configuration is password-protected. The password
reads LxiWebIfc (notice upper and lower case characters). This password
cannot be changed in the current firmware version.
2.10.3.5 IP Configuration
●
TCP/IP Mode
73Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
–decides whether DHCP is switched on or not. The IP Address, Subnet
Mask, Default Gateway and DNS Server configuration fields are only
enabled if DHCP is disabled. Consult the network administrator if it is necessary to modify these settings or any other settings.
Switching from DHCP enabled mode to manual configuration mode may
trigger an instrument restart when the new setting is confirmed (depends on
the configuration). If incorrect network configuration settings have been
entered, LAN RESET on the rear panel of the instrument must be used to
restore access to the Web interface.
●
IP address
–Allows the IP address of the instrument to be modified if DHCP is disabled.
●
Subnet mask, default gateway
Startup
–are settings that may have to be changed at times compared to the dis-
played presets if DHCP is switched off. Consult the network administrator
for the proper values if it is necessary to modify these settings.
●
DNS server
–This is the IP address of the domain name server of the network segment.
Consult the network administrator if it is necessary to modify this setting
when operating with DHCP disabled.
●
Dynamic DNS
–Allows different DNS servers to be addressed in succession. Consult the
network administrator if it is necessary to modify this setting when operating with DHCP disabled.
2.10.3.6 Advanced LAN Configuration
"Advanced LAN Configuration" provides LAN settings that are not declared mandatory by the LXI standard. These advanced parameters should only be modified
by experienced users.
74Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
●
Negotiation
–The Negotiation configuration field provides different Ethernet speed and
duplex mode settings. In general, the Auto Detect mode is sufficient.
●
ICMP Ping
–ICMP Ping must be enabled to use the ping utility.
●
VXI-11 Discovery
–Enables or prevents discovering the instrument via the VXI-11 Discovery
protocol, which is a protocol jointly developed by National Instruments and
VISA for searching for devices on a LAN.
VXI-11 is the protocol that is used for discovery of the instrument in the
LAN. According to the standard, LXI instruments must use VXI-11 to provide a discovery mechanism; other additional search methods are permitted.
2.10.3.7 Ping Client
Ping is a utility that verifies the connection between the LXI-compliant instrument
and another instrument. The ping command uses the ICMP echo request and
echo reply packets to determine whether the LAN connection is functional. Ping is
useful for diagnosing IP network or router failures.
75Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
The ping utility is not password-protected.
To initiate a ping between the LXI-compliant instrument and a second connected
instrument:
1. "Activate ICMP Ping" on the "Advanced LAN Configuration" page (activated
after LCI).
2. Enter the IP address of the second device without the ping command and
without any further parameters into the Destination Address field (e.g.
10.113.10.203).
3. Click on "Submit".
2.10.4Configuration for manual remote operation
The R&S UPV can be operated manually from an external computer via a network connection. The Windows program Remote Desktop Connection is used for
operation, see Chapter 4.16, "Manual Remote Operation", on page 237.
Manual remote control of the audio analyzer is possible under the following conditions:
●
the Windows 95 operating system or higher and the Remote Desktop Connection program are installed on the external computer and a LAN interface is
configured for the network
76Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
●
the audio analyzer and the computer are connected via LAN (see Chap-
ter 2.10.2, "Configuration of R&S UPV for Network Operation", on page 61),
●
the Remote Desktop Connection program is activated on the audio analyzer
(see Chapter 2.10.4.1, "Activating the Remote Desktop Connection program
at the R&S UPV", on page 77),
●
on the external computer, the R&S UPV data have been entered in the
Remote Desktop Connection program (IP address or computer name of R&S
UPV in the network (see Chapter 2.10.2.3, "Querying the computer name",
on page 66),
●
the login at the external computer for R&S UPV was performed with the correct user name (instrument) and the correct password (instrument) (see Chap-
ter 2.10.4.2, "Starting manual remote operation at the external computer",
on page 78).
2.10.4.1 Activating the Remote Desktop Connection program at the R&S
UPV
Startup
Unauthorized access possible to audio analyzer
After activating Remote Desktop, access to this R&S UPV is possible for
any user in the network who knows the computer name and the login data
of the audio analyzer.
When using Windows 7 Remote Desktop Connection is enabled by default.
Activating Remote Desktop when using Windows XP:
1. "Start – Settings – Control Panel": Click on "Performance and Maintenance".
2. "Click on System" and select "Remote" in the tab menu.
3. Activate the "Allow users to connect remotely to this computer" check box.
77Getting Started 1146.2078.62 ─ 10
R&S®UPV
Startup
Connect R&S UPV to a Network (LAN)
2.10.4.2 Starting manual remote operation at the external computer
The Remote Desktop Connection program is already installed in the Windows
operating system (program name "mstsc.exe" in "C:\Windows\System32"). . For
all other Windows operating systems from Windows 95 and higher, the program
can be downloaded free-of-charge from the Internet (http://www.microsoft.com). It
can be loaded onto any external computer by following the instructions, which are
also provided on the Internet.
1. If necessary, install the Remote Desktop Connection program on the external computer
2. Start the program in the Windows menu "Start - All Programs - Accessories Communications".
78Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
Before the program can be used for the first time, the instrument and user ID
of the R&S UPV must be entered on the external computer. The instrument ID
- the computer name of R&S UPV - identify the R&S UPV in the network.
Every R&S UPV is supplied together with a computer name which can be
used for manual remote operation. For notes about querying the computer
name, see Chapter 2.10.2.3, "Querying the computer name", on page 66.
The user ID is necessary as it authorizes access to the R&S UPV. They are
preset on the R&S UPV and are "instrument" for both the user name and the
password. No entry is needed in the Domain field.
Startup
3. Enter the instrument and user ID on the "General "tab of the extended
"Remote Desktop Connection" menu (called up by pressing the "Options>>"
button).
79Getting Started 1146.2078.62 ─ 10
R&S®UPV
Connect R&S UPV to a Network (LAN)
4. The login data can be saved with the "Save as" button.
When saving as default.rdp file, the connection to R&S UPV is offered as
default at the start of the program. When saving under a different name, the
setting for the connection with R&S UPV is provided in the selection list which
is called up by clicking on the button in the "Computer:" entry line.
5. 5. The resolution for the display of the R&S UPV monitor is set to 800 x 600
on the Display tab.
Startup
6. Establish the connection by clicking on the "Connect" button.
After the connection has been established, the audio analyzer screen appears
on the external computer. Operation is possible with the mouse and/or the
keyboard. Buttons on the front panel that do not have a direct correspondence
to the keyboard can be replaced by shortcuts (see Chapter 4.3.4, "Operation
Using an External Computer Keyboard", on page 151).
If several R&S UPVs should be manually remote-controlled by one computer,
a separate Remote Desktop Control window must be opened for every R&S
UPV. This is possible by repeatedly starting the program on the external computer.
80Getting Started 1146.2078.62 ─ 10
R&S®UPV
Firmware Update
Startup
2.11Firmware Update
The firmware update is described in the Release Notes.
2.11.1Troubleshooting during a Firmware Update
The two most common problems that can occur when installing the software (or
during a major update) are these:
●
At the beginning, just before the progress bar starts, the progress window displays the text "InstallShield is preparing InstallScript". Occasionally you then
see an error message and installation is canceled.
This problem can usually be solved by restarting Windows and carrying out
the firmware installation again.
●
After the installer has installed the firmware, i.e., the progress bar is completely filled, the progress window displays the text "Removing Backup Files",
"Registering Components" or "Publishing Product Information". Here, too, it
occasionally happens that an error message is followed by a second message, and the firmware is subsequently not installed.
This problem can sometimes be solved by restarting Windows and carrying
out the setup again. If the error occurs again, this can be for two reasons:
–Installation of the front panel driver went wrong. In this case, it must be
installed using the Windows wizard; the required INI file is named
C:\UPV\fpnl_wdm.inf. You should now reinstall the firmware.
–The installation of additional Windows settings went wrong. In this case,
execute the files C:\UPV\ScanFirmwareVersions.exe and
C:\UPV\SetReg.exe in succession. You should now reinstall the firmware.
If neither of these steps is successful, the instrument needs to go in for service (or the Windows image program must be reloaded).
81Getting Started 1146.2078.62 ─ 10
R&S®UPV
Windows XP Recovery and Backup
Startup
2.12Windows XP Recovery and Backup
The R&S UPV provides an invisible backup and recovery partition. A backup of
the factory system partition (C:\) is saved by default and can be restored in
case of a system crash.
In addition, backups of up to 5 firmware versions can be stored on this partition. It
is e.g. possible to backup the current system partition prior to a firmware update
or to provide different system configurations for different environments.
For the R&S UPV, the use of an external keyboard is recommended.
The operation is described with respect to an external keyboard. It is recommended to connect a mouse that allows an intuitive operation!
During restore, the system partition (C:\) is deleted, formatted and rewritten.
The data partition (D:\) is not affected by this.
2.12.1Calling the Windows XP Recovery and Backup Partition
Menu
If several backups are already stored on the disk, the remaining space
might not be sufficient for an additional backup. In this case a warning is
indicated in the Windows XP Recovery and Backup Partition menu and the
user is requested to remove an old backup before making a new one.
1. Switch the instrument off and on again Press the cursor keys multiple times
during the boot phase to ensure that the Boot menu does not independently close after a short period of time.
2. In the Boot menu, use the cursor keys to select the"Backup/Recovery" line.
3. Open the "Windows XP Recovery and Backup Partition" menu with [ENTER].
The menu shows the available selections for the recovery and backup partition:
82Getting Started 1146.2078.62 ─ 10
R&S®UPV
Windows XP Recovery and Backup
2.12.2Backup Current System Partition
Startup
1. In the Windows XP Recovery and Backup Partition menu, use the cursor keys
to select the "Make Backup" button.
2. Open the "Make Backup" menu with the [ENTER] key.
The menu shows the current versions of the firmware and the software platform.
3. Select the "Make Backup" button with the TAB keys.
4. Start the backup by pressing the [ENTER] key.
After the backup, the "Windows XP Recovery and Backup Partition" menu
appears again.
83Getting Started 1146.2078.62 ─ 10
R&S®UPV
Windows XP Recovery and Backup
5. Select "Exit and Shutdown" with the cursor keys.
6. Exit the program and shut down the R&S UPV with the [ENTER] key.
Startup
2.12.3Restoring a Selected Version of the System Partition
1. Use the cursor keys to select "Restore Backup" in the "Windows XP Recovery
and Backup Partition" menu to select the desired version of the system partition.
2. Open the "Restore Backup" menu with the [ENTER] key.
The menu shows the versions of the firmware and the software platform of the
backup which is displayed in the "Select Backup" window.
3. Select the "Select Backup" window with the TAB keys.
4. Select the backup to be restored with the up/down cursor keys and the
[ENTER] key.
5. Select the "Restore" button with the TAB keys.
6. Start the restore process with the [ENTER] key.
The script that is performed during recovery is displayed.
7. After the restore, shut down and switch off the R&S UPV.
The new setting is now active.
84Getting Started 1146.2078.62 ─ 10